Concentrix FX, FB, SX and SW

High-Speed Handpieces

Manual Chuck Turbine - Concentrix FB - continued:

Concentrix SX Only

Concentrix SX and Concentrix SW


4-Line (F.O.) Swivel

ISO 4-Line
Connector

ISO 6-Pin
Connector Tubing
Coupler

6-Pin Swivel
Tubing
Coupler

Instruction Manual

Bearing
Assembly

Division of DentalEZ Group


1816 Colonial Village Lane
Lancaster, PA 17601 U.S.A.
Printed in the U.S.A.
2012 DentalEZ, Inc.

Caution: Federal law restricts this device to sale by or on the order of a dentist.
0344

Y O U R

P R A C T I C E

Wrench

866.DTE.INFO (383.4636)
www.DentalEZ.com

P E R F E C T

Turn tubing coupler


while holding swivel
in place

Coupler Nut

65705 A Date of Issue 3/1/2012

DO NOT USE FOR ORAL SURGERY

Tubing Coupler
ISO B (4-Line) Type Swivels

ISO C Type (6-Pin) Swivel

6. Connect handpiece to swivel until it clicks.


Note: This joint loosens over time, resulting in a loss of power. Check for tightness periodically.

Handpiece Operation
Contents
Specifications
Handpiece Installation
Handpiece Operation

Safety Rules
Cleaning & Lubricating

Sterilization
Troubleshooting

Turbine Replacement
Replacement Part Numbers
Warranty

Air and Water Adjustment


1. Air Pressure
WARNING: To prevent premature damage to the turbine, do not exceed 34 psi when using a HiFlo swivel.

Using air pressure gauge StarDental Part #262591, for 4-Line tubing and handpieces, if a gauge is available, set pressure to:

30-32 psi for the Concentrix FX and FB

32-34 psi for the Concentrix SX and SW
If more torque is desired for the Concentrix SX, increase air pressure to a maximum of 42 psi.
If a gauge is not available, set pressure at delivery head allowing 1 psi pressure loss for every foot of tubing.
(See Specifications for recommended operational pressure.)

Specifications

Concentrix FX and FB are conventional non-swivel High-Speed handpieces. Concentrix SX and SW High-Speed Handpieces
are designed to fit a StarDental 360 quick disconnect swivel. The Concentrix SX connects to both StarDental and
standard 4-Line ISO fiber optic (F.O.) tubing; and to standard 4/6-Line ISO fiber optic (F.O.) tubing. The Concentrix SW
connects to both StarDental and standard 4-Line ISO fiber optic (F.O.) tubing.

Note: For maximum turbine life, do not exceed recommended pressure.

Type:
Operating Air Pressure:
Air Consumption:
No Load Speed:
Connection:
Power:
Weight:

Concentrix FB

Type:
Operating Air Pressure:
Air Consumption:
No Load Speed:
Connection:
Power:
Weight:

Concentrix SX

Type:
Operating Air Pressure:
Air Consumption:
No Load Speed:
Connection:
Power:
Weight:

Concentrix SW

Type:
Operating Air Pressure:
Air Consumption:
No Load Speed:
Connection:
Power:
Weight:

NOTE: Insert and remove


burs without excessive
force.

Bur

Manual
Bur Tool

4. Disengage tool and tug on bur to verify that it is completely engaged and holding securely. If chuck does not hold bur securely,
clean or replace chuck.
CAUTION: Be careful to keep bur tool aligned with handpiece. If tool is side-loaded while engaged in chuck, tool may be damaged.
REMOVAL of Bur from the Manual Chuck
1. Use manual bur tool and turn counterclockwise one full turn, 360 rotation.
2. Pull bur straight out.
CAUTION: Never rest the end cap on any part of the oral anatomy during handpiece operation.

Safety Rules

Concentrix FX

Slide Open End of


Tool Down

Handle

Tubing

1. Grasp the swivel in front of the bearing assembly between the thumb and index finger as shown.
2. Slide tubing coupler until the end of the tubing connector is exposed; hold in hand as shown with fingers securing the tubing.
3. Align and insert swivel tubes into the tubing connector holes.
4. While applying slight pressure to hold fittings in the connector holes, slide coupler forward and engage threads. Continue threading
coupler clockwise until hand tight against the swivel.
5. Using the wrench as shown, tighten the swivel to the tubing coupler until snug. Be sure to turn the tubing coupler, not the swivel.
DO NOT OVER TIGHTEN.

DentalEZ Dental Products (GB) Ltd.,


Cleveland Way,
Hemel Hempstead, Hertfordshire
HP2 7DY, ENGLAND

M A K E S

Bearing
Assembly

Tubing

Manual Chuck Turbine Bur Changing System


1. Pull handle back on manual bur tool, StarDental Part #257629, and slide open end of tool onto square nut as shown; release spring
loaded handle and seat shaft of tool into the opening in the rear of turbine housing.
2. To unlock chuck, turn manual bur tool handle counterclockwise.
3. After inserting a 1.6 mm diameter friction grip bur fully into chuck, turn handle of tool clockwise to tighten chuck.

2.
3.
4.

High speed with a replaceable autochuck turbine


30 to 32 psi
1.9 SCFM @ 30 psi
400,000 rpm
4-line fixed ISO configuration
10 watts
43g

High speed with a replaceable manual chuck turbine


30 to 32 psi
1.9 SCFM @ 30 psi
400,000 rpm
4-line fixed ISO configuration
10 watts
43g

Water Pressure Adjust water volume at the delivery head to obtain a satisfactory spray mist.
StarDental 6-Pin Fiber Optic System(s) set up: refer to the instruction manual that was included with the system and/or tubing.
Non-StarDental 6-pin Fiber Optic System(s):
If the handpiece is connected to a non-StarDental 6-pin fiber optic (F.O.) system, consult your operators manual of F.O. system
for correct voltage. Non-StarDental F.O. systems must have a rated amperage output between 1.5 Amps DC and .73 Amps DC at
3.1 Vdc. Use StarDental Part # 262962 bulb for non-StarDental F.O. systems that meet these requirements. Adjust the voltage per
instructions with the 3.1 Vdc bulb.
If the non-StarDental 6-pin F.O. system has a rated amperage output of at least 1.5 Amps dc at 3.3 Vdc, then the StarDental Part
#262522 bulb may be used. Adjust the voltage per instructions with the 3.3 Vdc bulb.

Autochuck Turbine - Concentrix FX, Concentrix SX and Concentrix SW:


WARNING: Use burs with an overall length of less than 24mm and a shank length of at least 7.3mm.
Use standard length concentric diamond and carbide burs with .0628" (1.6mm) nominal diameter
friction grip shanks designed for high speed handpieces. Every StarDental Diamond Bur meets
specifications below for recommended lengths. Contact your dealer for a complete list of
StarDental Diamond Burs. StarDental autochuck turbines have the following bur requirements:
Use only burs with hardened, tempered steel shanks or carbide shanks. Every StarDental Bur
maintains required hardness. Before use, verify concentric rotation by operating handpiece
outside the oral cavity in a safe place.

High speed fiber optic with a swivel and replaceable autochuck turbine (Swivel sold separately.)
34 to 42 psi
1.9 SCFM @ 34 psi
400,000 rpm
360 quick-disconnect swivel with 4-Line and 6-Pin ISO configuration and fiber optics
12 watts
49g

Note: Do not use burs that produce excessive noise or do not run concentrically.
Do not use burs that are modified, eccentrically designed, or have bent or soft
(non-heat treated) shanks.

Autochuck Bur Changing System


Maximum
24.0 mm
OVERALL LENGTH

Minimum
7.3 mm
Shank Length

Autochuck Bur Changing System


1. Insert bur into autochuck until resistance is felt.
2. Depress autochuck end cap button and continue to feed-in bur until it clicks and stops.
3. Release end cap button. Now, apply light inward force to the dental bur. This will help set
the chucking mechanism for initial start up. (Always tug on bur after it has been installed
to confirm that it is secured).

High speed with a swivel and replaceable autochuck turbine (Swivel sold separately.)
30 to 32 psi
1.9 SCFM @ 30 psi
400,000 rpm
360 quick-disconnect non fiber optic swivel with 2/3 line or 4-Line ISO configuration
12 watts
47g
135C

For example, if you are using the Concentrix FX that you want to operate at 32 psi and you are using either 5 feet or 12 feet of
tubing, then the pressure should be set at the delivery head to compensate for the air pressure losses.

5 feet tubing
32 psi + 5 psi = 37 psi

or 12 feet tubing
32 psi + 12 psi = 44 psi

1. To remove, grasp handpiece firmly so your thumb touches the end cap and push until a
slight click is noticed.
2. Pull bur straight out.
3. Release end cap.

Sterilizable up to 135C

CAUTION: Never depress the autochuck end cap button or rest the end cap against any part of the oral anatomy during handpiece
operation. Depressing the end cap while a bur is rotating can activate the chuck mechanism and result in end cap heatup and/or loss of bur retention.

Handpiece Installation

Manual Chuck Turbine - Concentrix FB:

Note: Filter air and water supplies, recommended filtration: 25 microns. Soften and deionize water. Be sure filters
operate efficiently.
Concentrix FX and Concentrix FB
1. Slide tubing coupler until the end of the ISO connector is exposed; hold
in hand as shown with smaller fingers securing the tubing.
2. Align handpiece tubes into the ISO connector holes.
3. While applying slight pressure to hold handpiece tubes in the connector
holes, slide coupler forward and engage threads. Continue threading
coupler clockwise until hand tight against the handpiece.

ISO 4-Line Connector


Tubing
Coupler

Tubing

WARNING: Use burs with an overall length of less than 24mm and a shank length of at least 10.8mm.
Autochuck Bur Changing System
Use standard length concentric diamond and carbide burs with .0628" (1.6mm) nominal
Maximum
diameter friction grip shanks designed for high speed handpieces. Every StarDental Diamond
24.0 mm
Bur meets specifications below for recommended lengths. Contact your dealer for a complete
OVERALL LENGTH
list of StarDental Diamond Burs. StarDental manual chuck turbines have the following bur
requirements:
Use only burs with hardened, tempered steel shanks or carbide shanks. Every StarDental Bur
maintains required hardness. Before use, verify concentric rotation by operating handpiece
outside the oral cavity in a safe place.
Note: Do not use burs that produce excessive noise or do not run concentrically.
Do not use burs that are modified, eccentrically designed, or have bent or soft
(non-heat treated) shanks.

Minimum
10.8 mm
Shank Length

PLEASE OBSERVE SAFETY RULES


1. Always tug on bur after chuck is loaded to confirm that the bur is secured.
2. Before use, verify concentric rotation by operating the handpiece outside the oral cavity in a safe place. Do not use burs that do not
run concentrically or produce excessive noise.
3. Never operate handpiece without a bur installed.
4. Always use burs with an overall length of less than 24 mm. The shank length must be at least 7.3 mm. This includes most short
shank burs.
5. When using handpieces with autochuck turbine - Concentrix FX, SX and SW - never depress end cap while bur is rotating.
6. Use only recommended burs to prevent damage to handpiece turbine and/or chuck and in some cases to prevent burs from
improper or incorrect placement. An improperly clamped bur may dislodge during use, causing injury to you or your patient. A
sudden high-pitched sound or increased vibration while handpiece is operating alerts the operator to the possibility that the chuck
is not clamped sufficiently to retain the bur or a bent bur is inserted.

WARNING: Always insert the bur full length of the chuck. If you attempt to extend it from the chuck, the bur may dislodge
during use and cause injury to you or your patient.
CAUTION: Replace the turbine when the handpiece fails to achieve full speed and/or increases sharply in noise level.This
is an indication of a failing turbine. A failing turbine will reduce the efficiency of the handpiece and may result
in end cap heat up. See "Turbine Replacement".

Cleaning and Lubricating


The Concentrix FX, FB, SX and SW Handpieces contain a turbine with steel bearings that are lubricated and sealed to keep the dirt out,
therefore, never clean ultrasonically or immerse in solvent that would dilute the lubricant. Lubrication of the turbine is required daily and prior
to sterilization. DentaLube II is required to ensure extended life of the Concentrix Handpieces.
CLEANING The Chuck
Cleaning the Manual Chuck (Concentrix FB):
Clean the chuck once a day and prior to sterilization (after each patient). Remove bur from
chuck; place three drops of DentaLube II, StarDental Part #262539, into chucking mechanism.
Aerate with syringe air.
Monthly Cleaning of the Manual Chuck
1. Loosen chuck with bur tool; remove bur.
2. Turn bur tool knob counterclockwise until chuck emerges from end cap.
3. Remove bur tool; turn chuck with fingers to free.
4. Clean chuck with alcohol
5. Run paper through 3 slots of chuck; remove debris.
6. Reinstall chuck by reversing steps 1-3.

Chuck

Chuck Slot

Cleaning the Autochuck (Concentrix FX, SX and SW):


PLEASE NOTE: to ensure that your High-Speed chuck continues to be in optimum working condition, clean once a week with Junk-Out
High-Speed Chuck Cleaner Kit P/N 263790:
Shake well before use
2
-Install brush to handle. See diagram at right.
-Use one new brush per cleaning application.
-Do not reuse brushes.
1
-Remove bur from chuck.
-Dip brush into bottle.
-Remove excess cleaner from the brush.
3
-Insert brush into bur end of turbine.
-Clean gently by rotating and inserting the brush in and out of the turbine.
-Repeat process if necessary.
Reseal cap on bottle immediately after use to reduce evaporation.
CLEANING Optical Surfaces
Every exposed optical surface of the Concentrix SX Fiber Optic Handpiece system accumulates debris which ultimately decreases light
output.To maintain maximum intensity, clean each surface with an alcohol soaked cotton swab. Use firm rubbing motion to remove debris
and mineral deposits. See the illustration below denoting each of the surfaces to clean.
For swivels: Disconnect handpiece from swivel; disconnect swivel from tubing by rotating coupler in the opposite direction shown in
Handpiece Installation. Depress foot control; verify water exits tubing.
CAUTION: Prior to cleaning, let lamp cool a few minutes, then remove.After cleaning, let lamp dry completely before reinstalling.

= Fiber Optic Surfaces


Swivel
Bearing/
Spring
Assy.

Xenon-Halogen Fiber
Optic Lamp
Clean

Water
Tube

Threaded
Coupler
Rear Seal
Duckbill
Valve Assy.

6-Pin HiFlo Swivel

NOTE: Do not attempt to disassemble handpiece or swivel.

KryptonHalogen F.O.
Lamp

CLEANING Burs
NOTE: Burs Must Be Cleaned and Sterilized Before First Use and After Each Patient.
Cleaning Diamond Burs:

- After each use, clean the diamond bur by running it against a wet diamond cleaning stone.

- To mechanically clean, the diamond bur should be held in a solution of either enzymatic presoak or water.

- Rinse the bur thoroughly and allow it to dry.

- Sterilize See Sterilization
CLEANING the Swivel
1. Disconnect handpiece from swivel; disconnect swivel from tubing by rotating coupler in the opposite direction as shown in Handpiece
Installation. Depress foot control; verify water exits tubing.
2. Connect swivel to tubing by following steps in Handpiece Installation. Depress foot control; if
almost no water exits front of swivel, clean as follows.
3. Remove swivel and find water tube on front face of swivel. See illustration.
4. Insert cleaning wire, StarDental Part #256719, into water tube. For 6-pin Fiber Optic and LED 4-line HiFlo
swivels, remove duckbill valve assembly from swivel to avoid damage. Move wire back and forth to F.O. Swivel
loosen debris at blockage.
Water Tube
5. Remove wire and apply a blast of air into water tube with syringe air.
CLEANING the Mating Parts of the Swivel
To optimize operation of the swivel disconnect mechanism, swab all internal surfaces of handpiece with a cotton swab soaked in
isopropyl alcohol until clean. Dry using syringe air.
Clean area from bearing/spring assembly at rear of swivel using isopropyl alcohol and a toothbrush. Dry using syringe air.
Bearing/
Spring
Assy.

Clean

Internal Surfaces

LUBRICATING the Steel Bearing Turbine


For maximum turbine life do not exceed recommended pressure.
Note: Lubricate handpiece daily. Use only DentaLube II, StarDental Part #262539.
1. Disconnect handpiece.
2. Assemble the conical red tip applicator, StarDental Part #202954, into the DentaLube II bottle cap.
3. Find drive air tube at rear of handpiece and place two drops of DentaLube II.
(See illustrations for more detail).
4. With bur in place connect handpiece to tubing, operate for 20 seconds to expel excess lubricant.
Concentrix SX and Concentrix SW

Problem:
Cause:
Correction:
Cause:
Correction:

Replacement Part Numbers

Lack of power, continued


Worn out turbine
Refer to Turbine Replacement.
Handpiece or swivel not threaded fully onto tubing
Refer to Handpiece Installation.

Problem: water leakage


Cause: Brittle handpiece or swivel silicone rubber seal
Correction: Replace silicone rubber seal:
For 4-line fixed back end handpieces and 4-line non HiFlo,non LED swivels - StarDental Part #202897 (Green)
For 6-line non HiFlo, non LED swivels - StarDental Part #262520 (Green)
For 4-line HiFlo fiber optic swivel - StarDental Part #263734 (Black)
For 6-pin HiFlo and LED swivels - StarDental Part #263770 (Black)
Swivel O-rings
Cause: Swivel O-rings worn
at the front of
Correction: Replace both O-rings on all non HiFlo swivels with StarDental Part #259884; or, for HiFlo the swivel
and LED swivels, use StarDental Part #263771. Refer to O-ring and Rear Seal Replacement
section in applicable swivel instruction manual.
Problem: swivel will not disengage from handpiece
Cause: Build-up of debris
Correction: Place two drops of DentaLube II into swivel/tubing connector area. Rotate swivel to circulate lube. Disengage
handpiece. Contact dealer if unable to disengage swivel.To prevent reoccurrence, refer to Cleaning.
Problem: improper spray
Cause: Clogged filters, tubes, or water ports
Correction: Inspect and clean water filter in delivery head. Disconnect handpiece and depress foot
control. Verify water exits the swivel. If water does not come from swivel, refer to Cleaning.
Cause: Brittle handpiece or swivel silicone rubber seal
Correction: Replace silicone rubber seal
Cause: Swivel O-rings worn
Correction: Replace O-rings. (See replacement O-rings.)
Problem:
Cause:
Correction:
Cause:
Correction:

Spray
Tube

bur locks OR SLIPS in autochuck


Debris build-up in chucking mechanism
Clean Chuck, see 'CLEANING The Chuck' under 'Cleaning and Lubricating'.
Use of non-ISO burs
Always use ISO burs.

Note: Always use ISO burs to prevent locking and slipping of burs in turbine chuck. If problems of locking and/or slipping
continues, replace worn Turbine immediately. Replace Autochuck Turbine with StarDental Part #260296, Autochuck Turbine
Replacement Kit. Replace Manual Chuck Turbine with StarDental Part #258661, Manual Chuck Turbine Replacement Kit.
Red Tip
Applicator

If any problem persists, contact dealer for further assistance; or, contact StarDental Technical Service at 1-866-DTE-INFO.

Drive Air Tube

Concentrix FX and Concentrix FB


Drive Air

Coolant Air
Water
Drive Air

Turbine Replacement

Autochuck Turbine - Concentrix FX, Concentrix SX and Concentrix SW:

Sterilization

1. Remove old end cap and turbine assembly, use end cap wrench shown.
2. Inspect housing before cleaning. Remove any obvious debris to prevent damage and to insure performance of turbine.
3. Connect handpiece and run air through turbine housing without turbine installed for 5 seconds.
4. Clean housing thoroughly with cotton swab and isopropyl alcohol. Dry with syringe air.
5. Place turbine assembly into housing and gently insert as shown.

CAUTION: Use steam sterilization exclusively for complete sterility. Never dry heat sterilize, ultrasonically
clean, or expose handpiece, bur or swivel to chemical disinfection or cold sterilization by:
Immersion techniques or Surgical milk
Surface sprays
Surface wipes, Hand soaps, or Detergents Sterilization bags containing disinfectants

Troubleshooting
Problem:
Cause:
Correction:
Cause:
Correction:
Cause:
Correction:
Cause:
Correction:

No Light or intermittent light


Electrical connections
Verify connections: wall outlet- power pack, power pack- lamp control, and lamp control-tubing connector.
Crimped signal air line to lamp control
Straighten signal air line to allow air flow.
Tubing corrosion on lamp contact pins
Remove lamp from swivel connector and clean contact pins with isopropyl alcohol.
Swivel lamp burned out
Replace lamp - See Replacement Part Numbers section of this manual.
When operating with non-StarDental 6-Pin F.O. systems, use StarDental Replacement Lamp Part #262962 and consult
your operators manual for correct voltage; Note: See Handpiece Installation, Non-Star 6-Pin F.O. system.

Autochuck Turbine

Wrench Opening

Housing

6. Position and engage threads of end cap into the housing by hand until snug.
7.Tighten end cap with the combination end cap wrench employing the wrench opening marked AC as shown above. Position wrench
with marked side facing end cap for best wrench control.

Manual Chuck Turbine - Concentrix FB:


1. Remove old end cap and turbine assembly, end cap wrench shown.
2. Inspect housing before cleaning. Remove any obvious debris to prevent damage and to insure performance of turbine.
3. Connect handpiece and run air through turbine housing without turbine installed for 5 seconds.
4. Clean housing thoroughly with cotton swab and isopropyl alcohol. Dry with syringe air.
5. Place turbine assembly into housing and gently insert as shown.
Manual Chuck Turbine
End Cap Wrench
Manual Chuck Turbine

Problem: Yellowish light


Cause: Debris on optical surfaces
Correction: Clean all exposed optical surfaces, refer to Cleaning.
Problem: reduced light
Cause: Low voltage
Correction: Refer to Fiber Optic Electrical System or Tubing Instruction Manual or contact StarDental customer service at
1-866-DTE-INFO.
Cause: Swivel lamp burned out
Correction: Replace lamp - See Replacement Part Numbers section of this manual.
When operating with non-StarDental 6-Pin F.O. systems, use StarDental Replacement Lamp Part #262962 and consult
your operators manual for correct voltage; Note: See Handpiece Installation, Non-Star 6-Pin F.O. system.
Problem:
Cause:
Correction:
Cause:
Correction:
Cause:
Correction:

Lack of power
Clean
Air pressure too low
debris from
Verify proper handpiece air pressure. Refer to Handpiece Installation.
inside of
housing
Crimped tubing
Straighten tubing inside and outside delivery head
Bearings
Debris build-up
Clean inside of housing and end cap with swab and alcohol. Lubricate with DentaLube II after reassembly.
Note: Before cleaning, check for and remove any debris from turbine housing to prevent damage and to ensure
turbine performance.

Concentrix FB Handpiece with ISO 4-Line Fixed Connection and Manual Chuck Turbine........................................................ 265635
Concentrix SX Fiber Optic Handpiece with Autochuck Turbine................................................................................................. 265632
Concentrix SW Handpiece with Autochuck Turbine................................................................................................................... 265633
HiFlo Swivel Assembly: 4-Line Fiber Optic ................................................................................................................................. 263758
HiFlo Swivel Assembly: ISO/C (6-pin 3.3 Vdc Fiber Optic)........................................................................................................... 263772
LED Swivel.................................................................................................................................................................................... 265400
HiFlo Swivel Assembly: ISO/C (6-pin 3.1 Vdc) (for non-Star F.O. System)..................................................................................... 263773
Steel Bearing Autochuck Turbine Replacement Kit for Lubricated High-Speed Handpieces....................................................... 260296
End Cap, Steel Bearing Autochuck Turbine................................................................................................................................... 260294
End Cap Wrench, Steel Bearing Autochuck Turbine..................................................................................................................... 203048
Steel Bearing Manual Chuck Turbine Replacement Kit for Lubricated High-Speed Handpieces (For Concentrix FB)................ 258661
Chuck, For Manual Chuck Turbine............................................................................................................................................... 258659
End Cap, Steel Bearing Manual Chuck Turbine............................................................................................................................. 254204
End Cap Wrench, Steel Bearing Manual Chuck Turbine............................................................................................................... 258739
Bur Changing Tool, Manual Chuck Turbine.................................................................................................................................. 257629
Replacement LED Lamp for LED Swivel (Package of 1)............................................................................................................... 265399
Replacement LED Lamp for halogen-style 6-Pin swivel 3.3Vdc only, swivel Part#263772 (Package of 1)................................... 265385
Replacement Halogen Lamp with spring (Package of 2)(For ColoRight Tubing, 5.5 Vdc)........................................................... 260986
Replacement Halogen Lamp, 6-pin 3.3 Vdc for swivel connected to StarDental F.O.Tubing (Package of 2)................................ 262522
Replacement Halogen Lamp, 6-Pin 3.1 Vdc for swivel connected to non-StarDental F.O.Tubing (Package of 2)......................... 262962
Rear Seal, HiFlo 4-Line Fiber Optic Swivel (Package of 3)............................................................................................................ 263734
Rear Seal, 4-Line for fixed back end handpieces and non HiFlo,non LED swivels (Package of 3) .............................................. 202897
Rear Seal, HiFlo 6-Pin Fiber Optic and LED Swivels (Package of 3).............................................................................................. 263770
Rear Seal, 6-Line for non HiFlo, non LED swivels (Package of 3) . ............................................................................................... 262520
Swivel O-Ring Kit for HiFlo Swivels (2 Large, 2 Small)................................................................................................................. 263771
Swivel O-Ring Kit for all non HiFlo Swivels (2 Large, 2 Small) .................................................................................................... 259884
Air Pressure Gauge (ISO 4/6-Line)................................................................................................................................................ 262591
Cleaning Wire (Gauge: 0.008, Package of 2)................................................................................................................................. 256719
DentaLube II . ............................................................................................................................................................................. 262539
ISO Tubing Set Screw (Package of 2)............................................................................................................................................ 261355
Allen Wrench .035........................................................................................................................................................................ 280830
Junk-Out High-Speed Chuck Cleaner........................................................................................................................................... 263790

Note: Replace the turbine when the handpiece fails to achieve full speed and/or increases sharply in noise level. The replacement
cartridge is available from your dealer and is easily replaced in the office.When ordering an Autochuck Turbine replacement, be
sure to order StarDental Autochuck Turbine Replacement Kit, StarDental Part #260296.When ordering a Manual Chuck Turbine
replacement, be sure to order StarDental Manual Chuck Turbine Replacement Kit, StarDental Part #258661.

Note: Never operate handpiece without bur in place.

Handpieces and burs must be cleaned, lubricated and sterilized before first use and after each patient.
Note: To Prevent Damage To the Bur, Use an Autoclavable Bur Block.
A wrapped handpiece or bur should be steam sterilized at 132 C for a minimum of 10 minutes or at 121 C for a minimum of 50
minutes. Do not exceed 135 C.
An unwrapped handpiece or bur should be steam sterilized at 132 C for a minimum of 8 minutes or at 121 C for a minimum of 30
minutes. Do not exceed 135 C.
Note: An unwrapped handpiece or bur needs to be used immediately or sterility is not maintained.
Do not exceed the sterilizer manufacturers guidelines for load capacity. Use only paper or paper/plastic bags. Always use complete
dry cycle of sterilizer. Always allow product to cool in bag to room temperature prior to using. Contact your sterilizer manufacturer
if moisture remains inside bag and on product after drying.

Always insist on StarDental Parts for best performance.


Note: Swivel assemblies sold separately.
Description
Item Number
Concentrix FX Handpiece with ISO 4-Line Fixed Connection and Autochuck Turbine.............................................................. 265634

Housing

6. Position and engage threads of end cap into the housing by hand until snug.
7.Tighten end cap with the manual chuck turbine end cap wrench.

Important Notice About


Returning Product To The DentalEZ Group
Anyone wishing to return any DentalEZ Group product to a DentalEZ facility MUST obtain the proper RETURN AUTHORIZATION NUMBER from
the Customer Service Department. The RA Number must appear clearly marked on both the outside and inside of the returned product carton.
When calling for a return authorization number you will be asked to provide the following information:
To Obtain A
Return Authorization
Number Call
1-866-DTE-INFO

Product Name, Model Number or Part Number


Serial Number of Product
Dealer Return Purchase Order Number
Reason For Return
Copy of Invoice Showing Purchase

Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation with this procedure will help us expedite
processing of any returns.

For new products, manuals and technical information, call


1-866-DTE-INFO or visit our web site @ www.dentalez.com
Limited Warranty
O-Rings and Seals are specifically excluded from this warranty.
DentalEZ warrants the Concentrix Handpiece to be free of defects in material and workmanship, under normal usage, under the following
terms:
StarDental Products:
Warranty Period:
Handpieces:
Concentrix FX
1 Year from date of purchase*
Concentrix FB
1 Year from date of purchase excluding turbine (turbine 6 months)*
Concentrix SX
1 Year from date of purchase*
Concentrix SW
1 Year from date of purchase*
Handpiece Parts:
Manual Bur Changing Turbine
6 Months from date of purchase*
Steel Bearing Autochuck Turbine
1 Year from date of purchase*
Please note the following additional terms of our warranty and return policy:
Warranties cover manufacturing defects only and do not cover defects resulting from abuse, improper handling, cleaning, care or
maintenance, normal wear and tear or non-observance of operating, maintenance or installation instructions. Failure to use authorized
parts or an authorized repair facility voids this warranty.
Liability is limited to repair or replacement of the defective product at our sole discretion. All other liabilities, in particular liability for
damages, including, without limitation, consequential or incidental damages are excluded.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO EMPLOYEE, REPRESENTATIVE OR DEALER IS AUTHORIZED
TO CHANGE THIS WARRANTY IN ANY WAY OR TO GRANT ANY OTHER WARRANTY.
WARRANTY REPAIRS: Parts repaired or replaced on a product that is in warranty will be warranted for the duration of that product's
original warranty.
NON-WARRANTY REPAIRS: The warranty on parts either repaired or replaced on an out-of-warranty product will cover the repaired part
only and will be for the time frame of a new parts warranty period.
PRODUCT RETURN: Opened products or product returns more than a year old cannot be returned for credit.There will be a 15% ($25.00
minimum) restocking charge on all items authorized for return.
*Provided conditions defined in the Instruction Manual are met.

Warranty validation card must be completed and returned within 15 days of purchase. StarDental will, without charge, repair or
replace (at its sole discretion) and return any defective part(s) covered in this warranty, providing the above conditions are met and
the product is returned though an authorized StarDental dealer.You must have a copy of your original invoice to qualify. StarDental is
not responsible for shipping damages.