Designation: D 7418 07
INTRODUCTION
This practice describes the instrument set-up and operation parameters for using FT-IR spectrometers for in-service oil condition monitoring. The following parameters are typically monitored for
petroleum and hydrocarbon based lubricants: water, soot, oxidation, nitration, phosphate antiwear
additives, fuel dilution (gasoline or diesel), sulfate by-products and ethylene glycol. Measurement and
data interpretation parameters are standardized to allow operators of different FT-IR spectrometers to
obtain comparable results by employing the same techniques. Two approaches may be used to monitor
in-service oil samples by FT-IR spectrometry: (1) direct trend analysis and (2) differential (spectral
subtraction) trend analysis. The former involves measurements made directly on in-service oil
samples, whereas the latter involves measurements obtained after the spectrum of a reference oil has
been subtracted from the spectrum of the in-service oil being analyzed. Both of these approaches are
described in this practice, and it is up to the user to determine which approach is more appropriate.
2. Referenced Documents
2.1 ASTM Standards: 2
D 4057 Practice for Manual Sampling of Petroleum and
Petroleum Products
E 131 Terminology Relating to Molecular Spectroscopy
E 168 Practices for General Techniques of Infrared Quantitative Analysis
E 1421 Practice for Describing and Measuring Performance
of Fourier Transform Mid-Infrared (FT-MIR) Spectrometers: Level Zero and Level One Tests
E 1866 Guide for Establishing Spectrophotometer Performance Tests
E 2412 Practice for Condition Monitoring of Used Lubricants by Trend Analysis Using Fourier Transform Infrared
(FT-IR) Spectrometry
3. Terminology
3.1 Definitions:
3.1.1 For definitions of terms relating to infrared spectroscopy used in this practice, refer to Terminology E 131.
1
This practice is under the jurisdiction of ASTM Committee D02 on Petroleum
Products and Lubricants and is the direct responsibility of Subcommittee D02.96 on
In-Service Lubricant Testing and Condition Monitoring Services.
Current edition approved Dec. 1, 2007. Published February 2008.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standards Document Summary page on
the ASTM website.
--`,,`,```,,,``,`,``,,`,,,```,,`-`-`,,`,,`,`,,`---
1. Scope
1.1 This practice covers the instrument set-up and operation
parameters for using FT-IR spectrometers for in-service oil
condition monitoring for both direct trend analysis and differential trend analysis approaches.
1.2 This practice describes how to acquire the FT-IR spectrum of an in-service oil sample using a standard transmission
cell and establishes maximum allowable spectral noise levels.
1.3 Measurement and integrated parameters for individual
in-service oil condition monitoring components and parameters are not described in this practice and are described in their
respective test methods.
1.4 The values stated in SI units are to be regarded as the
standard. The values given in parentheses are for information
only.
1.5 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
Licensee=Chevron HIST ASTM 9 loc Richmond, CA/1000001103
Not for Resale, 06/16/2008 10:53:41 MDT
D 7418 07
standardized configuration for FT-IR instrumentation and operating parameters employed in in-service oil condition monitoring in order to obtain comparable between-instrument and
between-laboratory data.
5. Apparatus
5.1 Fourier Transform Infrared (FT-IR) SpectrometerAll
FT-IR instruments suitable for use in this practice must be
configured with a source, beamsplitter and detector suitable for
spectral acquisition over the mid-infrared range of 4000 to 550
cm-1. The standard configuration includes a room temperature
deuterated triglycine sulfate (DTGS) detector, an air-cooled
source, and a germanium-coated potassium bromide (Ge/KBr)
beamsplitter, although a zinc selenide (ZnSe) beamsplitter may
also be used. The FT-IR spectrometers IR source and interferometer should be in a sealed compartment to prevent
harmful, flammable or explosive vapors from reaching the IR
source.
NOTE 1Photoconductive detectors such as mercury cadmium telluride (MCT) should not be used owing to inadequate linearity of the
detector response.
3
The boldface numbers in parentheses refer to a list of references at the end of
this standard.
2
Licensee=Chevron HIST ASTM 9 loc Richmond, CA/1000001103
Not for Resale, 06/16/2008 10:53:41 MDT
D 7418 07
7.2 Sample PreparationFiltering the sample using a filter
described in 5.3 prior to loading the cell with the sample is
highly recommended.
respectively. The actual cell pathlength obtained can be determined from the interference fringes in the spectrum recorded
with an empty cell or by recording the spectrum of a check
fluid; details for calculating cell pathlength are presented in
Appendix X2. The reporting units of the various in-service oil
condition monitoring parameter test methods are based on a
pathlength of 0.100 mm (see the respective test methods).
Accordingly, all data must be normalized to a pathlength of
0.100 mm, either by multiplying all data points in the absorption spectra by a pathlength correction factor (spectral normalization) or by multiplying the results of the respective test
methods by a pathlength correction factor (see 10.2). The
normalization procedure is usually part of the software provided by instrument manufacturers.
NOTE 2For purposes of interlaboratory comparison of results, spectral normalization should be performed.
7. Sampling
7.1 Sample AcquisitionThe objective of sampling is to
obtain a test specimen that is representative of the entire
quantity. Thus, laboratory samples should be taken in accordance with the instructions in Practice D 4057.
4
5
Conostan Division, Conoco Inc. 1000 South Pine, Ponca City, OK 74602-1267.
Thermal-Lube Inc., 255 avenue Labrosse, Pointe-Claire, QC H9R 1A3.
--`,,`,```,,,``,`,``,,`,,,```,,`-`-`,,`,,`,`,,`---
3
Licensee=Chevron HIST ASTM 9 loc Richmond, CA/1000001103
Not for Resale, 06/16/2008 10:53:41 MDT
D 7418 07
tion) provides a simple means by which to check cell loading.
The integrated absorbance values used to ascertain that the cell
is fully loaded shall be determined by individual laboratories.
(1)
D 7418 07
cell with a check fluid, and record its absorption spectrum. The
noise level is calculated by measuring the maximum peak-topeak noise in milliabsorbance units (mAU) over the frequency
range of 2000-1900 cm-1. The software of most commercial
spectrometers includes a noise calculation that can be used to
measure peak-to-peak noise over a selected frequency range.
The peak-to-peak noise level can be improved by increasing
the number of scans, whereby the noise level is decreased by a
factor of N1/2 when the number of scans is increased by a factor
of N. Various fringe reduction techniques outlined in Appendix
X1 can be used to help reduce sample spectral noise caused
by fringing. Acceptable sample spectral noise levels shall be
determined during a round robin and will be available on or
before July 1, 2009.
13. Keywords
13.1 condition monitoring; differential trend analysis; Fourier transform infrared; FT-IR; hydrocarbon based lubricants;
infrared; IR; lubricants; oils; petroleum lubricants
--`,,`,```,,,``,`,``,,`,,,```,,`-`-`,,`,,`,`,,`---
APPENDIXES
(Nonmandatory Information)
X1. FRINGING EFFECTS IN INFRARED LIQUID TRANSMISSION SAMPLE CELLS AND FRINGE REDUCTION METHODS
5
Licensee=Chevron HIST ASTM 9 loc Richmond, CA/1000001103
Not for Resale, 06/16/2008 10:53:41 MDT
--`,,`,```,,,``,`,``,,`,,,```,,`-`-`,,`,,`,`,,`---
D 7418 07
FIG. X1.1 Spectrum of a Mineral Oil Recorded Using an Air and an Empty Cell Background Before (A) and After (B) Fringe Reduction
by Tilting the Cell Holder at Brewsters Angle
NOTE X2.1When cell windows have been wedged to reduce interference fringes (Appendix X1), interference fringes may not be observed in
the absorption spectrum of the empty cell. In this instance, the cell
6
Licensee=Chevron HIST ASTM 9 loc Richmond, CA/1000001103
Not for Resale, 06/16/2008 10:53:41 MDT
D 7418 07
FIG. X2.1 Spectrum of an Empty Cell Used for Calculation of Cell Pathlength
fluid. Plot the absorbance values obtained from the peak height
measurements versus their respective pathlength values for
each of the assembled cells. Carry out a linear regression to
obtain an equation which has the following form:
(X2.1)
where:
E = cell pathlength in mm, and
N = number of fringes between n1 (cm-1) and n2 (cm-1).
y 5 mx1 b
where:
m = slope,
b = intercept,
x = absorbance values, and
y = pathlength values.
(X2.2)
7
Licensee=Chevron HIST ASTM 9 loc Richmond, CA/1000001103
Not for Resale, 06/16/2008 10:53:41 MDT
--`,,`,```,,,``,`,``,,`,,,```,,`-`-`,,`,,`,`,,`---
D 7418 07
REFERENCES
Improving Predictive Maintenance Programs, Proceedings of the
1992 Joint Oil Analysis Program International Condition Monitoring
Conference, JOAP-TSC, 1992, pp. 233-254.
(6) Powell, J., and Compton, D., Automated FTIR Spectrometry for
Monitoring Hydrocarbon-Based Engine Oils, Lubrication Engineering, 49, 3, March 1993, pp. 233-239.
(7) Toms, A., and Powell, J., Molecular Analysis of Lubricants by FT-IR
Spectrometry, P/PM Technology, 10, 4, August 1997, pp. 58-64.
(8) Lukas, M., and Anderson, D., Laboratory Used Oil Analysis Methods, CRC/STLE Tribology Data Handbook, Booser, E., ed., CRC
Press, 1997.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).
--`,,`,```,,,``,`,``,,`,,,```,,`-`-`,,`,,`,`,,`---
8
Licensee=Chevron HIST ASTM 9 loc Richmond, CA/1000001103
Not for Resale, 06/16/2008 10:53:41 MDT