Beruflich Dokumente
Kultur Dokumente
IM/8037
Issue 9
Sodium Monitor
8037
ABB
The Company
EN ISO 9001:2000
We are an established world force in the design and manufacture of instrumentation for
industrial process control, flow measurement, gas and liquid analysis and environmental
applications.
The UKAS Calibration Laboratory No. 0255 is just one of the ten flow calibration plants operated
by the Company and is indicative of our dedication to quality and accuracy.
Stonehouse, U.K.
0255
Electrical Safety
This equipment complies with the requirements of CEI/IEC 61010-1:2001-2 'Safety Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use'. If the equipment is used in a manner NOT specified by the Company, the protection
provided by the equipment may be impaired.
Symbols
One or more of the following symbols may appear on the equipment labelling:
Warning Refer to the manual for instructions
Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual for
any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of the
Technical Publications Department.
Sodium Monitor
8037
Contents
Contents
1
Introduction ..................................................................... 2
Maintenance ................................................................. 29
8.1 Chemical Solutions ................................................ 29
8.1.1 Reagent Solutions ...................................... 29
8.1.2 Standard Solutions ..................................... 29
8.1.3 Etch Solution .............................................. 30
8.1.4 Salt Bridge Solution .................................... 30
8.2 Scheduled Servicing .............................................. 30
8.2.1 Weekly ....................................................... 30
8.2.2 Monthly ...................................................... 30
8.2.3 12-Monthly ................................................. 30
8.3 Shut-Down Procedures ......................................... 31
8.3.1 Storage of Electrodes ................................. 31
8.4 Unscheduled Servicing .......................................... 31
8.4.1 Diagnostic Messages ................................. 31
8.4.2 Calibration pass and fail limits ..................... 31
8.4.3 Replacement of Plastic Tubing ................... 31
Setting Up ...................................................................... 13
4.1 Setting Up the Sensor Panel .................................. 13
4.1.1 Cable Attached Electrodes ......................... 13
4.1.2 Cable Detached Electrodes ........................ 13
4.2 Sensor Panel .......................................................... 13
4.3 Fitting the reservoir (option) .................................... 14
4.4 Principle of Operation ............................................. 15
IM/8037
Issue 9
10 Specification ................................................................. 35
Appendix A Optional 2nd Retransmission Output ....... 36
A.1 Setting Up ............................................................. 36
Sodium Monitor
8037
1 Introduction
1 Introduction
This manual describes the installation and operation of the 8037
Sodium Monitor. Fig. 1.1 shows a schematic of the system.
REAGENT
SOLUTION
STANDARD
SOLUTION
1
(LOW)
sive
Toxic
Flam
Sodium 8037
IM/8037
Issue 9
Sodium Monitor
8037
2 Mechanical Installation
2 Mechanical Installation
2.1.1 Transmitters
Caution. Mount the instruments in a clean, vibration-free
area, avoiding direct radiant heat, sunlight and drafts.
Avoid sites containing chlorination equipment.
REAGENT
SOLUTION
STANDARD
SOLUTION
1
(LOW)
sive
Toxic
Flam
Sodium 8037
Maximum Distance
10 metres
55 C
Max.
20 C
Min.
IP66
C Within Environmental Limits
Fig. 2.1 Siting Requirements Transmitter/Sensor Panel
IM/8037
Issue 9
Sodium Monitor
8037
2 Mechanical Installation
2 Drill suitable
holes
Transmitter Unit
1
Mark fixing centres
(see Fig. 2.3)
250mm required
for standard
solution container.
REAGENT
SOLUTION
STANDARD
SOLUTION
1
(LOW)
sive
Toxic
Reagent
Container
Flam
Sodium 8037
Standard
Solution
Container
Fix instrument to
wall using
suitable fixings
Dimensions in mm (in)
Fixing Centres
160 (6.3)
68
(2.68)
42
(1.65)
3 Secure plates
250 214
(9.84) (8.43)
232
(9.13)
Allowance for
Cable Bends
200 (7.9)
61 (23/8) o.d.
Vertical Post
IM/8037
Issue 9
Sodium Monitor
8037
2 Mechanical Installation
Dimensions in mm (in)
96 (3.78)
12 (0.47)
191 (7.52)
96
(3.78)
92 +0.8
(3.62 +0.03
0 )
0
Panel Cut-out
92 +0.8
0
(3.62 +0.03
0 )
IM/8037
Issue 9
Sodium Monitor
8037
2 Mechanical Installation
REAGENT
SOLUTION
STANDARD
SOLUTION
1
(LOW)
sive
Toxic
able
Flam
Sodium 8037
690
2.4.2 Drains
The drains from the sensor unit consist of two stub pipe
connections suitable for 10 mm (3/8 in.) bore plastic or rubber
tubing. Alkaline effluent from the sensor panel flow cell appears
at one connection and uncontaminated sample drains from the
other.
440
325
25
25
200
optional
environmental
cover
25
160
IM/8037
Issue 9
Sodium Monitor
8037
3 Electrical Connections
3 Electrical Connections
Warning. Before making any connections, ensure that the power supply, any high voltage-operated control circuits, and high
common mode voltage are switched off.
Although certain instruments are fitted with internal fuse protection, a suitably rated external protection device, e.g. fuze or
miniature circuit breaker (MCB), must also be fitted by the installer.
Transmitter
Screen
Sodium Electrode
Core
Reference Electrode
Black
Green
With or without
reservoir
connection
Red
Blue
Earth connection to
entrainment tube mounting
stud on panel
Green/Yellow
Sodium Electrode
Red
Transmitter
Core
Black
Black
Reference Electrode
Green
Red
Pt1000 Temperature
Compensator
Blue
Green/Yellow
Earth Connection to
entrainment tube
mounting stud on panel
IM/8037
Issue 9
Sodium Monitor
8037
3 Electrical Connections
Remove
protection
cover
1
slide
down
3
Slacken
captive
screws
2
Pull out
slightly. . .
Earth Studs
. . and
2 .slide
off
1
Remove nuts and
protection cover
Mains Cover
Remove
mains cover
IM/8037
Issue 9
Sodium Monitor
8037
3 Electrical Connections
NC C NO
NC C NO
Relay Contacts
Relay Contacts
Diode
R
External L
AC supply
Load
A AC Applications
C
External +
DC supply
Load
B DC Applications
IM/8037
Issue 9
Sodium Monitor
8037
3 Electrical Connections
Note. When 'attached' type electrodes are used (see Section 4.1.1, page 13) it is advisable to make their electrical connections
at the time of installing the electrodes into the liquid handling panel see Section 4.1, page 13.
1
1
Channel 2
Channel 1
Sodium/Ref Electrodes
1
2 Coax core
(Sodium Electrode)
3
4 Black (Ref. Electrode.)
+
4
Channel 1
Channel 2
PT1000
Temperature
Compensator
5 Red
6 Green
7 Blue
Retrans.
Serial
(If fitted)
Relay 1
Relay 2
Power
Supply
Serial
Retransmission
RS422/
RS485
Relay 1
Relay 2
Power
Supply
Output
Relays
Mains Supply
1
2
3
4
5
6
Rx+
Rx
Tx+
Tx
0V
Retrans.
Output
1
2
3
4
5
6
NC
C
NO
NC
C
NO
Relay 1
N Neutral
L Line
Earth
Relay 2
Earth Stud (on case)
see Fig. 3.1
NC = Normally Closed
C
Common
NO = Normally Open
10
IM/8037
Issue 9
Sodium Monitor
8037
3 Electrical Connections
Note. When 'attached' type electrodes are used (see Section 4.1.1, page 13) it is advisable to make their electrical connections
at the time of installing the electrodes into the liquid handling panel see Section 4.1, page 13.
Earth Stud
1
2
3
4
5
6
7
8
9
E
N
L
1
2
3
4
5
6
7
8
9
10
11
12
TBB
TBA
Retransmission Output
Relay 1
Relay 2
Mains Supply
Normally Closed
Common
Normally Open
Normally Closed
Common
Normally Open
Neutral
Live
1
2
3
4
5
6
7
8
9
E
N
L
1
2
3
4
5
6
7
8
9
10
11
12
0V
Rx
RS422/RS485
Rx+
Serial Interface (If fitted)
Tx
Tx+
Blue
Pt1000
Temperature
Green
compensator
Red
Reference Electrode (Black)
Coax core
Sodium Electrode
Screen
Earth
IM/8037
Issue 9
11
Sodium Monitor
8037
3 Electrical Connections
Remove cap
and screw
230
Slacken
captive
screws and
remove
protection
cover
230
Remove front
panel screws
Slide instrument
out of case
Undo captive
screw
230
230
12
IM/8037
Issue 9
Sodium Monitor
8037
4 Setting Up
4 Setting Up
1. Cable attached.
2. Cable detached (see below).
4.1.1 Cable Attached Electrodes
1. Unpack the sodium electrode and carefully remove the
rubber teat. Unscrew the cylindrical plastic holder from the
flow cell see Fig. 4.1, and slide the electrode fully into the
holder. Carefully screw the holder, with electrode, into the
appropriate bore of the flow cell so that the electrode bulb
passes through the 'O' ring.
Red
Black
Delivery tube
Sodium
electrode
Reference electrode
(With or without
reservoir connection)
Plastic
holder
'O' Ring
0211 164
IM/8037
Issue 9
13
Sodium Monitor
8037
4 Setting Up
Filler Plug
Filler Cap
Clip
Push the reservoir
onto the clip.
4
Reservoir
Reference
Electrode
'O' Ring
14
IM/8037
Issue 9
Sodium Monitor
8037
4 Setting Up
Standard
Solution
Container
Reagent
Solution
Container
KCl Reservoir
(option)
Constant
Head Unit
Output to
8037 Transmitter
Entrainment
'T' Piece
3-Way
Changeover
Valve
Sample
Inlet
Flowcell
Sodium
Electrode
Reference
Electrode
Pt1000
Contaminated
Drain
Uncontaminated
Drain
IM/8037
Issue 9
15
Sodium Monitor
8037
4 Setting Up
Earth Cable
Reagent Container
REAGENT
SOLUTION
Toxic
Flam
Sodium 8037
Reference
Electrode
Standard
Solution
Inlet
To Constant
Head Unit
Temperature Compensator
3-way
Changeover
Valve
Sample Inlet
Drain
16
IM/8037
Issue 9
Sodium Monitor
8037
5.1 Displays
The display comprises a 5-digit, 7-segment digital upper display
line and a 16-character dot-matrix lower display line. The upper
display line shows numerical values of sodium ion
concentration, temperature, alarm set points or programmable
parameters. The lower display line shows the associated units
or programming information.
Advance to
next page
Page 1
Parameter 1
Parameter 2
Parameter 3
Parameter 4
Page 2
Parameter 1
Parameter 2
Parameter 3
or
For majority
of parameters
99.5
Sodium
Page X
Parameter 1
Advance to
next parameter
g/kg
Parameter 2
Parameter 3
Parameter 4
Alarm
LEDs
Parameter Value
Adjust
Membrane Switches
or
New value is
automatically stored
Select
or
New value is
automatically stored
IM/8037
Issue 9
17
Start
18
100
g/l
Set Std.2
ug/l
10. 0
Set Std.1
1. 0 0
-----
g/l
A2 Setpoint
g/l
10. 0
A1 Setpoint
15
00000
Alter Cal. Code
00000
Temp Comp.Factor
0. 0 020
Recovery (min)
15
-----
15
(%)
mg/l
g/l
00000
Alter Sec. Code
Test Retrans
RTX. Span
RTX. Zero
1. 0 0
Sample Recovery
%Slope
millivolt offset
Calibrating Std.2
50. 0
(1K5)
(1K0)
100
Temp Span
Temp Zero
mV Span (+1000)
10. 0
Log
YES
4-20
Output Type
Hold Outputs
RTX. Type
0. 1 00
Sample Recovery
1. 0 0
millivolt offset
Calibrating Std.1
-240
Connect Std.2
-----
Calibrating Std.1
Connect Std.1
88888
Lin
mV Zero (-1000)
88888
YES
-----
Calibrate
-----
g/l
NO
88888
A2 Setpoint
Low
-----
NO
A2 Action
-----
Electrical Cal.
00000
-300
(s)
10. 0
Set Defaults
Filter Time
-----
g/l
-----
A1 Setpoint
30
Low
FACTORY SETTINGS
-----
Electrical Calibration
Page
88888
Secure Parameters
Set Up
Language Page
A1 Action
-----
g/l
----Units
----SET UP OUTPUTS
Set Up
Outputs Page
----SET UP PARAMETERS
Set Up
Parameters Page
-----
Connect Std.1
100 -----
g/l
Set Std.1
10. 0
English
-----
SECURITY CODE
00000
Access to
Secure Parameters
1. 0 0 ----- -300
millivolt offset
-240
%Slope
Temperature (C)
----- ----250
.
00000
g/l
-300
millivolts
Sodium
SENSOR CAL.
One Point
Calibration Page
-----
Two Point
Calibration Page
9. 9 5
Operation Page
Sodium Monitor
8037
6 Startup And Operation
Ensure all electrical connections have been made and switch on the power supply. If the instrument is being commissioned for the
first time, calibration and programming of parameters is required.
IM/8037
Issue 9
Sodium Monitor
8037
Either single or two point calibrations can be carried out depending on the operating conditions. The Slope of a sodium electrode is
relatively stable, therefore frequent two point calibration may be unnecessary. It is suggested that single point calibration should be
done weekly and two point calibration carried out monthly, but a suitable schedule must be determined to suit the operating
conditions.
Before starting an automatic calibration sequence, rinse the standard solution containers with high purity water and fill with fresh
standard solutions. If a single point calibration is to be done only one solution (STD1) is required. If a two point calibration is to be
done, the lower value solution should be used first.
Note. A flashing display indicates an out of range condition.
9.95
Sodium
g/l
300
millivolts
250
Temperature
%Slope
(C)
100
0
The % Slope is displayed numerically and in bars (see calibration pass and fail limits).
Note. The % Slope is shown on the upper display.
The millivolt offset is displayed. This represents the difference in millivolt output of the electrode
pair compared with the ideal output.
millivolt offset
1.00
10.0
A1 Setpoint
A2 Setpoint
g/l
g/l
-----
Sensor Cal.
IM/8037
Issue 9
19
Sodium Monitor
8037
-----
00000
Enter the required calibration password between 00000 and 19999. If an incorrect value is
entered, access to the sensor calibration is prevented and the display returns to the top of the
sensor calibration page.
SENSOR CAL.
-----
10.0
Set Std.1
Set the calibration standard solution value between 0.10 g kg-1 and 10.0 mg kg-1.
Note. See Section 8.1.2 regarding the minimum concentration for static sodium solutions.
g/l
Set the calibration standard solution value between 0.50 g kg-1 and 10.0 mg kg-1.
100
Set Std.2
Note. Standard 2 is always greater than Standard 1, and must be at least five times the value
of Standard 1.
g/l
Connect the Standard 1 solution using the quick release coupling, and operate the 3-way
changeover valve to allow the solution to pass through the flowcell.
-----
Connect Std. 1
300
During calibration (15 minutes), the millivolt output from the electrode pair is displayed.
Calibrating Std.1
-----
Disconnect the quick release coupling from the Standard 1 bottle and connect the Standard 2
solution in the same manner as Standard 1.
Connect Std. 2
240
During calibration the millivolt output from the electrode pair is displayed.
Calibrating Std.2
Millivolt offset
20
IM/8037
Issue 9
Sodium Monitor
8037
100
After 15 minutes, Millivolt Offset is displayed for a few seconds, followed by the %Slope
display.
100
The % Slope is calculated and is used to indicate the quality of the electrode pair.
Millivolt offset
%Slope
9.95
Sample Recovery
Return to top
of Operation
Page
IM/8037
Issue 9
00000
The display remains showing %Slope until the scroll button is pressed, when the concentration
is displayed during the Sample Recovery period (usually 30 minutes). At the same time as
operating the scroll button, the 3-way change-over valve should be returned to sample.
At the end of this period, the display automatically returns to the Operating Page.
Advance to Access to Secure Parameters page see Section 7.1, page 23.
SECURITY CODE
21
Sodium Monitor
8037
----SENSOR CAL.
00000
Cal. User Code
-----
10.0
Set Std. 1
g/l
Set the calibration standard solution value between 0.10 g kg-1 and 10.0 mg kg-1.
Notes.
If the required value cannot be set, select 2 Point Calibration and check that Std. 2 is not
less than minimum see Section 6.3.1, page 20.
See Section 8.1.2 regarding the minimum concentration for static sodium solution.
-----
Connect the Standard 1 solution, using the quick release coupling, Operate the 3-way
changeover valve to change from sample input to standard solution input.
Connect Std. 1
-300
During calibration (15 minutes), the millivolt output from the electrode pair is displayed.
Calibrating Std.1
millivolt offset
1.00
The millivolt offset is recalculated. This represents the difference in millivolt output of the
electrode pair compared with the ideal output.
The display remains showing millivolt offset until the scroll button is pressed, when the
concentration is displayed during the Sample Recovery period (usually 30 minutes). At the
same time as operating the scroll button, the 3-way changeover valve should be returned to
the sample. At the end of the recovery period, the display automatically returns to Operating
Page.
Sample recovery time does not start until this parameter is selected.
Sample Recovery
00000
SECURITY CODE
22
Advance to Access to Secure Parameters page see Section 7.1, page 23.
Note. When the %Slope has been established by a two point calibration, a one point
calibration can be carried out on a static sodium standard, as described above, or on the
sample, if the sodium concentration of the latter is known.
IM/8037
Issue 9
Sodium Monitor
8037
00000
SECURITY CODE
-----
Security Code
Enter the required code number, between 00000 and19999, to gain access to the secure
parameters. If an incorrect value is entered, access to subsequent programming pages is
prevented and the display reverts to the Operating Page.
Advance to Select Language Page, see Section 7.2, page 23.
English
-----
Use the buttons to select the required language (English, French, German or Spanish).
English
-----
SET UP PARAMETERS
IM/8037
Issue 9
23
Sodium Monitor
8037
----SET UP PARAMETERS
----Units
Units
The measured units can be displayed in g kg-1, g l-1 or ppb.
g/kg
g/l
ppb
-----
Temperature Units
-----
This is used to prevent short term variations in reading. A setting of 100 seconds will usually be
found to give acceptably stable readings. The range is 10 to 500.
-----
If 'YES' is selected, default values of zero millivolt offset and 100% Slope are inserted in the
software.
Temp. Units
Filter Time
Set Defaults
(C)
(s)
NO
-----
SET UP OUTPUTS
24
IM/8037
Issue 9
Sodium Monitor
8037
-----
or
SET UP OUTPUTS
press to advance to Set Up Outputs Page, see Section 7.4, page 25.
----A1 Action
Alarm 1 Action
For 'Fail-safe' alarm operation the relay's alarm state must be the same as the power-down
state, i.e. the relay is de-energized.
High
Low
The set point band is defined as the actual value of the set point plus or minus the hysteresis
value. The hysteresis value is 1% of the full span value displayed in the Set Up Parameter
Page see Section 7.3, page 24. Alarm action occurs if the input value is above or below the
set point band. If the input moves within the set point band the last alarm action is maintained.
Off
1.00
A1 Setpoint
The Alarm 1 set point can be set to any value between 0.10 g kg-1 and 10.0 mg kg-1.
g/kg
Alarm 2 Action
----A2 Action
High
Low
Off
10.0
A2 Setpoint
The Alarm 2 set point can be set to any value between 0.10 g kg-1 and 10.0 mg kg-1.
g/kg
-----
-----
The current output and the relays can be held to prevent inadvertent operation during a sensor
calibration. Select 'YES' to hold outputs, otherwise select 'NO'.
RTX. Type
Hold Outputs
4-20
YES
----Ouput Type
IM/8037
Issue 9
Log
25
Sodium Monitor
8037
Lin
----Output Type
The current output can be set to Logarithmic or Linear. To select Logarithmic select 'Log'.
Otherwise select 'Lin'.
Log
Lin
0.100
RTX. Zero
g/kg
10.0
RTX. Span
mg/kg
50.0
Test Retrans
The zero current output can be set to any value between 0.10 g kg-1 and 100 g kg-1. This is
only available for Logarithmic output.
Note. For Linear output, the zero value is always 0 g kg-1.
The span current output can be set to any value between 10.0 g kg-1 and 10.0 mg kg-1
(logarithmic or linear).
Note. For Logarithmic output, the minimum span is two decades, and the maximum is four
decades.
The instrument can transmit a signal of 0, 25, 50, 75 or 100% of the current output range.
(%)
00000
00000
Return to
top of page
-----
FACTORY SETTINGS
26
IM/8037
Issue 9
Sodium Monitor
8037
-----
Parameters in these pages are factory set and should not normally require adjustment. They
can only be set up if the necessary equipment is available.
00000
-----
FACTORY SETTINGS
Electrical Cal.
-----
NO
Calibrate
Select 'YES' to access the electrical calibration sequence. Select 'NO' to advance to Adjust
current output zero.
YES
88888
Set the millivolt source to -1000 mV and allow the instrument display to stabilize.
mV Zero (-1000)
88888
Set the millivolt source to +1000 mV and allow the instrument display to stabilize.
mV Span (+1000)
88888
Set the temperature simulator resistance box to 1000 ohms, and allow the instrument to
stabilize.
88888
Set the temperature simulator resistance box to 1500 ohms, and allow the instrument to
stabilize.
-----
-----
IM/8037
Issue 9
27
Sodium Monitor
8037
-----
15
15
30
Recovery Time.
Usually set to 15 minutes, but can be set between 1 and 30 minutes. (1 minute is used for
electrical simulation of calibration).
Usually set to 15 minutes, but can be set between 1 and 30 minutes (1 minute is used
electrical simulation of calibration).
Time after returning to sample before the alarms are reactivated. Usually set to 30 minutes, but
adjustable between 1 and 120 minutes.
Recovery (min)
0. 0 020
Temp.Comp Factor
00000
Return to
top of page
-----
Sodium
28
g/l
IM/8037
Issue 9
Sodium Monitor
8037
8 Maintenance
8.1 Chemical Solutions
The reagents and calibration solutions detailed in this section
are required to keep the monitor operating. Solutions should be
stored in plastic bottles and where possible, should be freshly
made.
8.1.1 Reagent Solutions
Two alternative reagent solutions may be used, depending on
the required lower limit of measurement. Concentrated
ammonia solution, which provides adjustment of sample pH to
10.7 is suitable for measurements of sodium ion to
approximately 0.5 g kg-1. At concentrations below this,
hydrogen ion interference becomes significant and a reagent of
50% diethylamine solution should be used. This adjusts the
sample pH between 11.2 and 11.5 and enables measurements
to be made down to 0.1 g kg-1.
1. Concentrated ammonia solution one litre.
Warning. This reagent should only be handled under
a fume hood. It causes burns and is irritating to the
eyes, respiratory system and skin. Wear rubber gloves
and eye protection. In warm weather pressure
increases in the bulk container of ammonia and the
cap must be released with care.
A 35% w/v solution (s.g. 0.88) is recommended, but lower
concentrations, to a minimum of 30% w/v (s.g. 0.89), can
be used see Section 8.1.1, page 29 which relates
sodium concentration and pH.
2. Diethylamine Solution (50%)
Warning. Diethylamine is an extremely inflammable
and irritating colorless liquid with a strong smell of
ammonia. It should be handled with care at all times.
The following points should also be noted:
Avoid breathing vapor and avoid contact with
skin and eyes.
Work under a fume hood, wearing rubber
gloves and eye protection.
In the event of a fire, extinguish with water spray,
foam, dry powder or carbon dioxide.
If a spillage occurs, shut off all possible sources
of ignition, and instruct others to keep at a safe
distance. Mop up spillage with plenty of water,
diluting greatly. Ventilate the area well to
evaporate any remaining liquid and dispel vapor.
8 Maintenance
8.1.2 Standard Solutions
The following instructions refer to the preparation of 100 g l-1
and 1 mg l-1 sodium, LOW and HIGH standard solutions
respectively, but any concentrations can be prepared within the
measuring range selected by appropriate dilution of the stock
solution. For all practical purposes, 'g l-1' can be considered to
be equal to 'g kg-1' ('ppb') and 'mg l-1' equal to 'mg kg-1'
('ppm').
Stock solution (1000 mg l-1 sodium ions)
1. Dissolve 2.543 (0.001) g of analytical reagent grade
sodium chloride in approximately 100 ml high purity water.
Transfer this solution to a one litre volumetric flask and
make up to the one litre mark with more high purity water
to give a stock solution of 1000 mg l-1 sodium ions. Store
in a plastic container.
2. Pipette 10 ml of this solution to a one litre volumetric flask.
Make up to the one litre mark with high purity water to give
a solution of 10 mg l-1 sodium ions.
3. Pipette 10 ml of the 10 mg l-1 solution into a one litre
volumetric flask and make up to the one litre mark with
high purity water to give the LOW standard solution of
100 g l-1 sodium ions. Transfer this solution to the bottle
labelled STANDARD SOLUTION 1 (LOW).
4. Transfer 100 ml of the 10 mg l-1 solution to a one litre
volumetric flask and make up to the mark with high purity
water to give the HIGH standard solution of 1 mg l-1
sodium ions. Transfer this solution to the bottle labelled
STANDARD SOLUTION 2 (HIGH).
a. It is not advisable to prepare static sodium solutions
of less than 50 g l-1 because low concentration
solutions rapidly become contaminated and change
in concentration.
b. Although the HIGH and LOW standard solutions are
typically one decade apart in sodium concentration,
any concentration difference can be used within the
constraints of i) above and the requirement that the
HIGH solution should be at least five times the
concentration of the LOW solution. It is necessary to
have a significant change in electrode output to
achieve an accurate calibration.
IM/8037
Issue 9
29
Sodium Monitor
8037
8 Maintenance
30
Temperature
IM/8037
Issue 9
Sodium Monitor
8037
8 Maintenance
Condition
16 Character
Display
Flashes Sodium
Value
Flashes Sodium
Value
Sample Temperature
> 55 C
Flashes 'Hot'
Sample Temperature
< 5 C
Flashes 'Cold'
Solution is cold
IM/8037
Issue 9
31
Sodium Monitor
8037
8 Maintenance
Percent Slope
Displayed Lights
Action
110%
5 lights flashing
Calibration unacceptable.
New coefficients ignored and the 8037 transmitter unit continues to use the last known good coefficients.
Check correct standard solutions have been used or entered.
93%
5 lights
86%
4 lights
79%
3 lights
Calibration accepted. New coefficients saved into non-volatile memory. Recommend electrode pair are
serviced or replaced.
<79%
2 lights flashing
Calibration unacceptable.
New coefficients ignored and the 8037 transmitter unit continues to use the last known good coefficients.
Check that correct standard solutions have been used or entered.
If solutions are satisfactory, replace the electrode pair and calibrate.
Reagent
Solution
Container
KCl Reservoir
(option)
Entrainment
'T' Piece
2.4mm i.d.
Tygon
0212 362
(140mm)
Constant
Head Unit
Uncontaminated
Drain
Flowcell
3-Way
Changeover
Valve
Sample
Inlet
Contaminated
Drain
32
IM/8037
Issue 9
Sodium Monitor
8037
9 Spares List
9 Spares List
Part no.
Description
8037 050
8037 070
8036 286
Description
No.
Required
No.
Required
1048 836
0216 558
1048 837
8037 110
1436 836
8037 115
1436 837
0216 514
1436 840
8036 286
8036 286
8036 286
8037 140
Description
1401 500
1403 500
Description
No.
Required
No.
Required
8037 150
1405 500
0214 047
1410 500
0214 059
1431 875
0214 071
1431 876
0214 526
1431 877
1431 878
8061 660
8036 216
8035 677
Entrainment/earthing tube
8037 095
8037 085
8037 086
0216 513
Description
8037 570
8037 571
2
8037 572
8037 055
4600/0295
8037 580
8037 056
4600/0405
8037 057
8037 058
8037 062
8037 063
IM/8037
Issue 9
No.
Required
33
Sodium Monitor
8037
9 Spares List
Description
No.
Required
8037 575
8037 571
8037 576
8037 577
4600/0246
4600/0285
8037 580
4600/0405
Consumables
Part no.
Description
1431 242
34
No.
Required
1
IM/8037
Issue 9
Sodium Monitor
8037
10 Specification
10 Specification
Installation Information
General
Sample Temperature:
5 to 55 C.
Display
Measured value
5-digit, 7-segment
Sample Flow:
50 ml min-1 to 500 ml min-1.
Reproducibility
5% of concentration or 0.1 g kg-1 (whichever is the greater) at
constant temperature.
Response Time
1 to 100 g kg-1 less than 4 minutes for 90% step change 100 to 1
g kg-1 less than 6 minutes for 90% step change.
Resolution
0.01g kg-1 in range 0.10 g kg-1 to 9.99 g kg-1
Sample drains:
9.5 mm i.d. flexible, atmospheric drain.
Power supply
0 to 10, 0 to 20 or 4 to 20 mA programmable.
Requirements:
115 V nom. 15 V, 50/60 Hz or 230 V nom. 30 V, 50/60 Hz
Power Consumption:
<10 VA.
Alarms
Two fail-safe, high and low concentration alarms.
Hysteresis:
1% of FSD (fixed)
Environmental Data
3 A AC 3 A DC max.
Loading:
non inductive
750 VA 30 W max.
inductive
75 VA3 W max.
Insulation, contacts to earth:
2 kV r.m.s.
Calibration:
Manual initiation of calibration sequence. Calibration frequency,
typically, weekly for 1 point, and 4 weekly for 2 point calibration.
IM/8037
Issue 9
35
Sodium Monitor
8037
1
1
7
+
Channel 1
Channel 1
Sodium/Ref Electrodes
12 - Coax core (Sodium
Electrode)
34 - Black (Ref. Electrode.)
Channel 2
Channel 2
PT1000
Temperature
Compensator
Retransmission 2
+
Retrans.
Output 2
Retrans. 2
N
4
Retrans. 1
Retransmission 1
+
Retrans.
Output 1
5 - Red
6 - Green
7 - Blue
Relay 1
Relay 2
Power
Supply
Relay 1
Relay 2
Power
Supply
Relays
Mains Supply
1
2
3
4
5
6
NC
C
NO
NC
C
NO
Relay 1
N Neutral
L Line
Earth
Relay 2
Earth Stud (on case)
see Fig. 3.1
NC = Normally Closed
C
Common
NO = Normally Open
36
IM/8037
Issue 9
Sodium Monitor
8037
Earth Stud
1
2
3
4
5
6
7
8
9
E
N
L
1
2
3
4
5
6
7
8
9
10
11
12
TBA
Retransmission
Output 1
Relay 1
Relay 2
Mains Supply
Normally Closed
Common
Normally Open
Normally Closed
Common
Normally Open
Neutral
Live
TBB
1
2
3
4
5
6
7
8
9
E
N
L
1
2
3
4
5
6
7
8
9
10
11
12
Retransmission
Output 2
+
Pt1000
Blue
Temperature
Green
compensator
Red
Reference Electrode (Black)
Coax core
Sodium Electrode
Screen
Earth
IM/8037
Issue 9
37
38
Sodium
g/kg
9. 9 5
Operation Page
SENSOR CAL.
-----
Two Point
One Point
Calibration Page Calibration Page
Set Up Language
Page
English
-----
SECURITY CODE
00000
Access to
Secure Parameters
----SET UP OUTPUTS
----SET UP PARAMETERS
RTX O/P
RTX Span
Log
mgl
10. 0
gl
0. 1 0
Output Type
RTX Zero
4-20
Temp Span
Temp Zero
Sodium
-----
RTX Type
-----
SET UP RETRANS 2
-----
-----
Sodium
Set Up
Outputs Page
Set Up
Parameters Page
(C)
55. 0
(C)
5. 0
Temperature
FACTORY SETTINGS
-----
Electrical Calibration
Page
Sodium Monitor
8037
Appendix A Optional 2nd Retransmission Output
Fig. A.3 Overall Programming Chart Showing Position of Second Retransmission Output Page
IM/8037
Issue 9
Sodium Monitor
8037
-----
-----
SET UP RETRANS 2
RTX Type
4-20
0-20
The retransmission 2 output can be selected from three mA current ranges to ensure
compatibility with the peripheral device connected.
0-10
----RTX O/P
SODIUM
TEMP
TEMP
SODIUM
----Output Type
Log
TEMP:
The current output can be set to Logarithmic or Linear. To set to Logarithmic select 'Log'.
Otherwise set to 'Lin'.
Lin
0.10
RTX Zero
g/l
10.0
RTX Span
mg/l
-----
The zero current output can be set to any value between 0.10 g kg-1 and 100 g kg-1. This is
only available for Logarithmic output.
Note. For Linear output the zero value is always 0 g kg-1.
The span current output can be set to any value between 10.0 g kg-1 and 10.0 mg kg-1
(Logarithmic or Linear).
Note. For Logarithmic output the minimum span is two decades, and the maximum span is
four decades.
FACTORY SETTINGS
TEMP
5.0
The zero output can be set to any value between 5C and 45C (41F to 113F). See Note
below.
55.0
The span output can be set to any value between 15C and 55C (59F to 131F). See Note
below.
Temp Zero
Temp Span
(C)
(C)
-----
FACTORY SETTINGS
Note. The minimum range between Zero Output and Span Output is 10C (18F).
IM/8037
Issue 9
39
Sodium Monitor
8037
----FACTORY SETTINGS
88888
Temp Span (1K5)
-----
Adjust RTX.Zero 2
-----
Adjust RTX.Span 2
15
40
IM/8037
Issue 9
Customer Support
Automation Systems
United Kingdom
ABB Limited
Tel: +44 (0)1453 826661
Fax: +44 (0)1453 829671
United States of America
ABB Inc.
Tel: +1 775 850 4800
Fax: +1 775 850 4808
Flexible Automation
Flow Measurement
Electromagnetic Flowmeters
Mass Flowmeters
Turbine Flowmeters
Wedge Flow Elements
Electrical Systems
Marine Equipment
Offshore Retrofit and Refurbishment
Process Analytics
Transmitters
Pressure
Temperature
Level
Interface Modules
Control Valves
Actuators
Positioners
Client Warranty
Prior to installation, the equipment referred to in this manual must
be stored in a clean, dry environment, in accordance with the
Company's published specification.
Periodic checks must be made on the equipment's condition. In
the event of a failure under warranty, the following documentation
must be provided as substantiation:
1. A listing evidencing process operation and alarm logs at time
of failure.
2. Copies of all storage, installation, operating and
maintenance records relating to the alleged faulty unit.
IM/8037
Issue 9
ABB 2007
ABB Limited
Oldends Lane, Stonehouse
Gloucestershire
GL10 3TA
UK
Tel: +44 (0)1453 826661
Fax: +44 (0)1453 829671
ABB Inc.
Analytical Instruments
9716 S. Virginia St., Ste. E
Reno, Nevada 89521
USA
Tel: +1 775 850 4800
Fax: +1 775 850 4808