Operating Documentation
GE MEDICAL SYSTEMS
REV 3
2154943
WARNING
AVERTISSEMENT
WARNUNG
AVISO
GE MEDICAL SYSTEMS
REV 3
2154943
ATENO
AVVERTENZA
GE MEDICAL SYSTEMS
REV 3
2154943
GE MEDICAL SYSTEMS
REV 3
2154943
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation damage in shipment
written on all copies of the freight or express bill before delivery is accepted or signed for by a General Electric
representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier
immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for
inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within
this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (414) 8273449 / 8*2853449 immediately after damage is found. At this
time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item damaged and
extent of damage.
Complete instructions regarding claim procedure are found in Section S of the Policy & Procedure Bulletins.
CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.
Devices not to be used near this equipment:
Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmitter,
radiocontrolled toy, etc.
Keep power to these devices turned off when near this equipment.
Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be around
this equipment to fully comply with the above regulation.
GE MEDICAL SYSTEMS
REV 5
2154943
REVISION HISTORY
REV
0
1
2
3
4
5
DATE
. . . . . . Apr. 1, 1996
. . . . . May. 20, 1996
. . . . . . Jan. 1, 1997
. . . . . Jun. 18, 1998
. . . . . June 14, 2000
. . . . . Apr. 20, 2001
Initial Release
Periodic changes
Periodic changes
Periodic changes
The use of cleaning Pad is prohibited.
Added KYOUEI Brush.
REV
Title page . . . . . . . . . . 5
Warning Label(a to d) 3
A .................. 5
B . . . . . . . . . . . . . . blank
i to ii . . . . . . . . . . . . . . . 5
PAGE
REV
Section 1
11 to 16 . . . . . . . . . . 5
Section 2
21 to 25 . . . . . . . . . . 5
26 . . . . . . . . . . . . blank
PAGE
REV
PAGE
REV
Section 3
31 to 313 . . . . . . . . . 5
314 . . . . . . . . . . . blank
Section 5
51 to 519 . . . . . . . . . 4
520 . . . . . . . . . . . blank
Section 4
41 to 456 . . . . . . . . . 5
Section 6
61 to 612 . . . . . . . . . 4
PAGE
REV
Tab APPENDIX
A1 to A4 . . . . . . . . . 3
Blank/Rear Cover . . .
GE MEDICAL SYSTEMS
blank
2154943
GE MEDICAL SYSTEMS
REV 5
2154943
TABLE OF CONTENTS
SECTION
PAGE
11
Distinction the Gantry covers among EMC100 type, EMC80 type and NO EMC type
Distinction the Ring between Panel type and Stainless type . . . . . . . . . . . . . . . . . . . . .
Distinction the Brush Assy between ETC/KYOUEI type and B.T.B type . . . . . . . . . . .
Distinction the Gantry Interface board between DTR board and DTR2 board . . . . . .
Distinction the Gantry Interface board among
STRR(early) board, STRR(late) board and STRR2 . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12
13
14
15
21
21
22
16
CAUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
MAINTENANCE PROHIBIT ITEMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
21
23
SECTION 3 MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
31
31
32
33
31
32
34
34
36
37
39
312
313
SECTION 4 REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
41
34
35
36
37
41
42
43
44
45
46
47
41
41
44
45
46
410
414
414
425
430
448
454
456
GE MEDICAL SYSTEMS
REV 5
2154943
SECTION
PAGE
51
51
53
53
58
512
512
514
515
517
518
519
61
53
54
55
56
57
61
62
63
64
ERROR CODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TROUBLE SHOOTING FLOW CHART USING ERROR CODE . . . . . . . . . . . . . . . . .
TROUBLE SHOOTING FLOW CHART USING HEARING SENSE . . . . . . . . . . . . . .
SLIP RING AGING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ii
61
62
63
64
GE MEDICAL SYSTEMS
REV 5
2154943
Table 11
Distinction
System
PANEL
TYPE
EMC 80 TYPE
PANEL
TYPE
STANDARD
PANEL
TYPE
Signal Brush
(*3)
Interface BD
(Rotational)
(*5)
Slip Ring
Type
B.T.B
DTRE
ETC / KYOUEI
DTRE
B.T.B
DTRE
ETC / KYOUEI
DTRE
B.T.B
DTR2
ETC / KYOUEI
DTR2
NO EMC TYPE
STAINLESS
B.T.B
DTR
DTR2
EXTRA (*6)
NO EMC TYPE
PANEL
TYPE
Interface BD
(Stationary)
(*4)
ETC / KYOUEI
DTR
ETC / KYOUEI
DTR2
C
D
E
F
STRR (early)
STRR (late)
STRR2
STRR (early)
STRR (late)
STRR2
11
GE MEDICAL SYSTEMS
REV 5
1-1
2154943
DISTINCTION THE GANTRY COVERS AMONG EMC 100 TYPE, EMC 80 TYPE AND NO EMC TYPE
2.
3.
Gantry Covers
EMC 100
TYPE
EMC 80 TYPE
NO EMC
TYPE
FRAME GND
BRUSH
12
GE MEDICAL SYSTEMS
REV 5
1-2
2154943
2.
3.
Check the signal slip ring at the Gantry rear side (see illustration 12).
Illustration 12
PANEL TYPE
RESIN COLOR : CREAM COLOR
STAINLESS TYPE RESIN COLOR : BLACK COLOR
13
GE MEDICAL SYSTEMS
REV 5
1-3
2154943
DISTINCTION THE BRUSH ASSY BETWEEN ETC/KYOUEI TYPE AND B.T.B TYPE
2.
3.
Remove only the Slip Ring cover where is installed upper the Signal Brush assy. The location is as follows ( illustration 13).
Illustration 13
4.
Check the MANUFACTURED DATE PRINT or the joint solder between the tip and plate on the Signal Brush assy
(see illustration 14).
Illustration 14
WHITE
PAINT
EXIST
ETC/KYOUEI
TYPE
JOINT SOLDER
NOEXIST
FLAT
BALL
FLAT
BALL
B.T.B TYPE
14
GE MEDICAL SYSTEMS
REV 5
1-4
2154943
DISTINCTION THE GANTRY INTERFACE BOARD BETWEEN DTR BOARD AND DTR2 BOARD
The distinction between DTR board and DTR2 board is required to remove the DAS top cover.
1.
Remove the DAS top cover. Refer to the each system Component Replacement manual.
2.
ADC
ASSY
OFST
COMP
ASSY
CIF
ASSY
DAS REAR
VIEW
15
GE MEDICAL SYSTEMS
REV 5
2154943
1-5 DISTINCTION THE GANTRY INTERFACE BOARD AMONG STRR(EARLY) BOARD AND STRR(LATE)
BOARD AND STRR2
1.
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
2.
3.
4.
5.
Set the Gantry to 1 oclock position and lower the table to the lowest position.
6.
7.
Remove the four nuts and the back side cover (see illustration 16).
8.
Check the Parts # Silk Print which is printed either P9183YD or 2109655 or P9183YN (see illustration 16).
Illustration 16
FOUR NUTS
BACK COVER
INTERFACE BOAD
PARTS NO.
P9183YD
2109665
STRR2
P9183YN
16
GE MEDICAL SYSTEMS
REV 5
2154943
2-1
CAUTION
WARNING!
ELECTROCUTION!! THERE IS A LIVE CURRENT RUNNING THROUGH THE
GANTRY SLIP RING POWER BRUSHES, DO NOT SERVICE THE GANTRY SLIP
RINGS WITHOUT TURNING OFF ROTATE, XG(SAFETY LOOP),
SLIP RING115V BREAKERS.
Rotate
XG
Table/Tilt
(Safety Loop)
Control
Slip Ring
115V
(DAS/
DET)
CAUTION
Exercise caution when working with slip ring components to avoid contact, inhalation, and
ingestion of slip ring brush powder.
Use Gloves : Large size :46194427P400, XLarge size :46194427P401
Do NOT remove gloves near an exposed slip ring. The powder inside the glove can
contaminate the ring.
Use a HEPA Vacuum Cleaner : 46297933P1 to remove brush black powder (dust).
21
GE MEDICAL SYSTEMS
REV 5
2-1
2154943
CAUTION (continued)
CAUTION
There are small projections at the thin ground ring (see illustration) and at the left edge of the
slip ring panel. Take care not to be hurt during the slip ring cleaning.
PROJECTION
PROJECTION
GROUND RING
CAUTION
Use the Gloves not to be hurt, when you are performing the maintenance on the Slip Ring and
Brush.
22
GE MEDICAL SYSTEMS
REV 5
2-2
2154943
DO NOT INSTALL THE ETC/KYOUEI BRUSH TO THE STAINLESS SIGNAL SLIP RING. OTHERWISE, YOU MUST REPLACE THE SLIP RING
ASSY.
DO NOT WIPE THE ETC/KYOUEI BRUSH, B.T.B BRUSH USING LIQUID CLEANER.
23
GE MEDICAL SYSTEMS
REV 5
2-2
2154943
SIGNAL BRUSH
ASSY
DO NOT
LOOSEN THESE
SCREWS
Note
Do NOT loosen any screws except of Allow Label
screws.
Note
When the slip ring $ brush alignment is NOT correctly adjusted, these screws are must be
loosened.
Note
Loosen these screws, only when you replace STAINLESS RING with PANEL TYPE RING.
24
GE MEDICAL SYSTEMS
REV 5
2-2
2154943
DO NOT
LOOSEN THESE
SCREWS
Note
Do NOT loosen any screws except of Allow Label
screws.
Note
When the slip ring $ brush alignment is NOT correctly aligned, these screws must be
loosened.
Note
Loosen these screws, only when you replace STAINLESS RING with PANEL TYPE RING.
25
GE MEDICAL SYSTEMS
blank
2154943
26
GE MEDICAL SYSTEMS
REV 5
2154943
SECTION 3 MAINTENANCE
3-1
Table 31
Gantry
Revolutions
Brush Type
Reference
Section A B C D E F G H I J K L
Check Items
FG Brush assy
Cleaning
3-3
Cleaning
3-4
Wear Check
5-1
Cleaning
3-5
Cleaning
3-6
Wear Check
5-4
Cleaning
3-7
Wear Check
5-5
D D D D D D D D D D D D
D D
Every 500K
(or 6 Months)
3-2
Every 900K
(or 12 Months)
4-7
D D
Every 1000K
(or 12 Months)
3-2
31
D D D
D D
GE MEDICAL SYSTEMS
REV 5
3-2
2154943
NOTICE
Do NOT remove the Power brush Assy and Signal brush main support.
Special Equipment
D HEPA Vacuum Cleaner : 46297933P1
D HEPA Filters for Vacuum Cleaner : 46297948P1
D Kim Wipe (2254011)
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1
Switch OFF the Rotate, XG and Slip Ring switches at the rear Gantry base.
Remove the signal brush assy. (Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT)
Remove ALL of the slip ling covers, if the system has the slip ring cover at the Gantry rear side.
(Refer to Section 4-1 SLIP RING COVER REMOVAL)
Using the Vacuum Cleaner, clean the followings until they do NOT have black powder.
D Inner side of the cone portion of the front Gantry cover
D Inner side of the cone portion of the rear Gantry cover
D Gantry rear lower side (especially, lower ring cover stays and lower cable duct cover)
D All of the inner side of the slip ring covers (removed in step 5)
D Power rings
FRONT AND REAR GANTRY COVER
POWER RINGS
INNER SIDE OF
CONE
32
ALL OF INNER SIDE
OF RING COVERS
GE MEDICAL SYSTEMS
REV 5
3-2
7
2154943
33
GE MEDICAL SYSTEMS
REV 5
3-3
2154943
3-3-1
Special Equipment
D Alcohol (2254012)
D Kim Wipe (2254011)
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
NOTICE
Do NOT be performed on the stainless type Slip Ring.
1
Switch OFF the Rotate, XG and Slip Ring switches at the rear Gantry base.
Remove only the slip ring cover which covers the Signal Brush Assy, if the system has the slip ring cover at the
Gantry rear side. (Refer to Section 4-1 SLIP RING COVER REMOVAL)
Remove the signal brush assy. (Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT)
Switch ON the Rotate , Table/Tilt and Control switches and turn OFF the XG and Slip Ring switches
at the rear Gantry base.
CAUTION
TAKE CARE NOT TO BE HURT DURING THE SLIP RING CLEANING.
There are small projections at the thin ground ring (see illustration) and at the left edge of the
slip ring panel.
PROJECTION
PROJECTION
GROUND RING
34
GE MEDICAL SYSTEMS
REV 5
3-3-1
7
2154943
4
TGP BOARD
SYS
MNL
5
ABT
FIX
SPD2
SPD0
FAST
SLOW
CONT
90_
Wipe off the signal slip ring surface using Kim Wipe with alcohol by rotating the Gantry at slowest speed
(see illustration 33).
Illustration 33
Wipe off the signal slip ring surface using dry Kim Wipe by rotating the Gantry at slowest speed.
(For removing the remaining alcohol.) (see illustration 33).
10
Make sure that all slip ring surface is dry. Otherwise, there are some liquid on the slip ring surface and the the
contact between the slip ring and the brush assy will be bad.
35
GE MEDICAL SYSTEMS
REV 5
3-3-2
2154943
Signal Slip Ring Cleaning (Stainless Type)
Special Equipment
D HEPA Vacuum Cleaner : 46297933P1
D HEPA Filters for Vacuum Cleaner : 46297948P1
D Kim Wipe (2254011)
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1
Switch OFF the Rotate, XG and Slip Ring switches at the rear Gantry base.
Remove the signal brush assy. (Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT)
Switch ON the Rotate , Table/Tilt and Control switches and turn OFF the XG and Slip Ring switches
at the rear Gantry base.
Clean the signal slip ring surface using the Vacuum Cleaner by rotating the Gantry at slowest speed
(see illustration 33).
36
GE MEDICAL SYSTEMS
REV 5
3-4
2154943
Special Equipment
D Sand Paper : U0042BF (For B.T.B Type Signal Brush Only)
D Kim Wipe : 2254011 (For ETC/KYOUEI Type Signal Brush Only)
D Oscilloscope
D 1.5 V battery (Type D)
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Remove only the slip ring cover which covers the Signal Brush Assy, if the system has the slip ring cover at the
Gantry rear side. (Refer to Section 4-1 SLIP RING COVER REMOVAL)
CAUTION
Do NOT touch the slip ring surfaces and signal brushes !
Remove the SIgnal Brush Assy. (Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT)
CAUTION
Do NOT touch the slip ring surfaces and signal brushes !
NOTICE
Do NOT bend the spring plate of the brush tip.
37
GE MEDICAL SYSTEMS
REV 5
3-4
6
2154943
CARBON
TRACK
BLACK POWDER
NOTICE
Do NOT file off, rubbing the Signal Brush surface.
8
Verify that all brush tips have no carbon tracks, black powder then install the brush Assy.
38
GE MEDICAL SYSTEMS
REV 5
3-5
2154943
Special Equipment
D HEPA Vacuum Cleaner : 46297933P1
D HEPA Filters for Vacuum Cleaner : 46297948P1
D Resurfacer : P9183SU
D Flexible Abrasive : P9183SV
D Kim Wipe (2254011)
D HEPA Vacuum Cleaner : 46297933P1
D HEPA Filters for Vacuum Cleaner : 46297948P1
D Gloves : Large size :46194427P400, XLarge size :46194427P401
D Alcohol (2254012)
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
CAUTION
Possible irreparable damage!
The inner signal slip rings are very sensitive, never touch or apply any abrasive, or other
cleaning material, to the signal slip rings. See Illustration 35 .
Perform the power slip ring cleaning procedure very carefully, otherwise irreparable damage
will result.
CAUTION
Handle the power brushes carefully since they are fragile (made of graphite).
Keep grease or dust away from the brushes as these materials will considerably deteriorate
the life and performance of the brushes.
39
GE MEDICAL SYSTEMS
REV 5
3-5
2154943
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Remove only the slip ring cover which covers the Signal Brush Assy, if the system has the slip ring cover at the
Gantry rear side. (Refer to Section 4-1 SLIP RING COVER REMOVAL)
Switch ON the service switch on the SUB board near the TGP board.
Rotate the gantry by hand until the effected sections of the power slip rings, the sections with black projections,
reach the power brush position, and use the Resurfacer (P9183SU) and the Flexible Abrasive (P9183SV) to
scrape clean the rings, in the following way (see illustration 35):
a. Remove the Signal Brush Assy. (Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT)
b. Scrape off black projections from the rings using the resurfacer, until the ring surface is leveled; checking this
by touching your finger on the surface.
c. Clean the scraped surfaces using the abrasive to remove filings/powder.
d. The flexible abrasive will soon turn black with use. Cut away the blackened surface to expose a fresh, new
surface and continue scraping clean the power slip rings.
e. Clean up the dust using Vacuum Cleaner.
f. Wipe the power slip ring using Kim Wipe with alcohol.
g. Clean the power slip ring using the dry Kim Wipe.
Illustration 35
115 V
CLEANING POINT
310
GND
550 V
GE MEDICAL SYSTEMS
REV 5
3-5
2154943
Switch OFF the service switch on the SUB board near the TGP board.
Switch ON the Rotate , Table/Tilt and Control switches and turn OFF the XG and Slip Ring switches
at the rear Gantry base.
10
Clean the power slip ring surface using Vacuum Cleaner during the Gantry rotation (see illustration 35).
311
GE MEDICAL SYSTEMS
REV 5
3-6
2154943
Special Equipment
D Kim Wipe (2254011)
D HEPA Vacuum Cleaner : 46297933P1
D HEPA Filters for Vacuum Cleaner : 46297948P1
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1
Remove the Power Brush Assy. (Refer to Section 4-3 POWER BRUSH REPLACEMENT)
Use dry Kim Wipe / Vacuum cleaner to clean the filings/powder from the ring side of the power holder base and
power brush guides (see illustration 36).
NOTICE
Do NOT touch the slip ring surfaces
Illustration 36
PLASTIC COVER
312
GE MEDICAL SYSTEMS
REV 5
3-7
2154943
Special Equipment
D Kim Wipe (2254011)
D Sand Paper : U0042BF
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1
Remove all FG brushes from the Gantry. Refer to FG Brush Assy Replacement
Check visually if any carbon track, black powder are attached to brushes.
Illustration 37
FG Brush Surface
CARBON
TRACK
BLACK POWDER
Remove any carbon tracks and black powder from the brush using a dry Kim Wipe or Sand Paper : U0042BF.
NOTICE
Do NOT bend the spring plate of the brush tip.
4
Verify that all brush tips have no carbon tracks, black powder then install the brush Assy.
313
GE MEDICAL SYSTEMS
blank
2154943
314
GE MEDICAL SYSTEMS
REV 5
2154943
SECTION 4 REPLACEMENT
4-1
4-1-1
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
NUTS
4 NUTS
SUB COVER
TWO NUTS
(LOOSEN, DO NOT REMOVE)
41
TWO NUTS
(LOOSEN, DO NOT REMOVE)
CT other Systems
GE MEDICAL SYSTEMS
REV 5
4-1-1
2154943
Remove the right side ring cover from the rear view, 14 nuts.
Remove the left side ring cover from the rear view, 10 nuts.
Remove the lower slip ring cover with rear cover holder, 8 nuts and 3 Allen screws.
Illustration 42
Ring Cover
SIGNAL
BRUSH COVER
POWER
BRUSH COVER
42
GE MEDICAL SYSTEMS
REV 5
4-1-1
2154943
NOTICE
When the right and left side covers are installed, they are set over the tabs of the upper and
lower covers (see illustration below).
Illustration 43
UPPER COVER
SIDE COVER
LEFT SIDE
COVER
RIGHT
SIDE COVER
LOWER COVER
NOTICE
When all slip ring covers are installed, verify if there is no strange noise during the Gantry
rotation by hand.
Illustration 44
OUTER COVER
INNER COVER
NO INTERFERENCE NOISE
43
GE MEDICAL SYSTEMS
REV 5
4-1-2
2154943
EMC 80 Type Slip Ring Cover Removal
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Seven Nuts
Eight Nuts and Two Screws
Ten Nuts
Ten Nuts
Eight Nuts
Eight Nuts and Three Allen Screws
NOTICE
When the right and left side covers are installed, they are set over the tabs of the upper and
lower covers (see illustration 43).
NOTICE
When all slip ring covers are installed, verify if there is no strange noise during the Gantry
rotation by hand (see illustration 44).
44
GE MEDICAL SYSTEMS
REV 5
4-1-3
2154943
NO EMC Type Slip Ring Cover Removal
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
POWER
BRUSHES
SIGNAL
BRUSHES
NOTICE
When all slip ring covers are installed, verify if there is no strange noise during the Gantry
rotation by hand (see illustration 44).
45
GE MEDICAL SYSTEMS
REV 5
4-2
2154943
There are three(3) types of the Signal Brush (B.T.B, ETC and KYOUEI).
B.T.B (2143818)
ETC (2148394)
KYOUEI (2274988)
NOTICE
Do NOT install the ETC and KYOUEI brushes to the Stainless Slip Ring. Otherwise, you must
replace the Ring.
Special Equipment
D Kim Wipe (2254011)
D Alcohol (2254012)
D Sand Paper : U0042BF (For B.T.B Type Signal Brush Only)
D Oscilloscope
D 1.5 V battery (Type D)
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Remove ALL of the slip ling covers, if the system has the slip ring cover at the Gantry rear side.
(Refer to Section 4-1 SLIP RING COVER REMOVAL)
NOTICE
Do NOT touch the slip ring surfaces !
46
GE MEDICAL SYSTEMS
REV 5
4-2
5
2154943
SLIP RING
STRS
BOARD
SLIP RING
Remove the four screws B and then remove the two signal brush assys (see illustration 46).
Illustration 46
FIBER OPTIC
CONNECTORS
20 SCREWS
SCREW B
47
CONNECTORS
GE MEDICAL SYSTEMS
REV 5
4-2
2154943
Wipe to clean the filings/powder on the STRS board using dry Kim Wipe.
10
Verify that the SHEET BRUSH SUPPORT is not touching the signal ring surface. If it is touching, remove the
brush assembly, correct it bending the SHEET BRUSH SUPPORT by hand, then install again.
NOTICE
For a new brush tip, it is OK that its end will be contacted with ring surface.
Do NOT bend the tip support spring.
Illustration 47
GOOD
NO GOOD
SIGNAL RING
11
12
4
TGP BOARD
SYS
ABT
MNL
FIX
48
SPD2
SPD0
FAST
CONT
SLOW
90_
GE MEDICAL SYSTEMS
REV 5
4-2
2154943
13
14
15
16
17
49
GE MEDICAL SYSTEMS
REV 5
4-3
2154943
Special Equipment
D Kim Wipe (2254011)
D HEPA Vacuum Cleaner : 46297933P1
D HEPA Filters for Vacuum Cleaner : 46297948P1
D Tie Wraps
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Remove ALL of the slip ling covers, if the system has the slip ring cover at the Gantry rear side.
(Refer to Section 4-1 SLIP RING COVER REMOVAL)
NOTICE
Do NOT touch the slip ring surfaces !
5
PLASTIC COVER
410
GE MEDICAL SYSTEMS
REV 5
4-3
2154943
POWER BARS
ALLEN SCREW
ALLOW LABEL
XCONDENSER
(YELLOW)
FOR EMC100 ONLY
A SCREWS
WHITE
BLACK
GREY
YCONDENSERS
(BLACK)
ALLOW LABEL
TOROIDAL CORE
TIE WRAP
411
GE MEDICAL SYSTEMS
REV 5
4-3
2154943
Dislodge (pullout) the power holder from its side base by pulling it free of the two support pins.
10
Unscrew the two power guide screws from each brush guide (10) on the back of the power holder assembly, and
then pull the power brush guides out from the power holder base.
11
Unscrew the power bar screws and carefully pull the power bars free from the power holder base.
Illustration 411
POWER BRUSH
412
GE MEDICAL SYSTEMS
REV 5
4-3
2154943
12
Unscrew the brushes, one at a time, remove them and replace them with the new brushes. Verify that the new
brush is the same type as the one it is replacing.
13
Exchange the screws labelled M3 in illustration 412 from the old brushes to the new brushes, one at a time.
14
Return the power bars to their original positions with the brushes facing downward into the holes of the power
holder base.
15
Insert the center shaft of the springpin assemblies shown in illustration 412 into the hole on the top of the two
power brushes, and gently slide the power brush guides into their original positions.
16
17
18
19
POWER BAR
CHECK LINE
413
POWER BRUSH
GE MEDICAL SYSTEMS
REV 5
4-4
2154943
4-4-1
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80)
NOTICE
Do not reinstall the OLD brush (ETC, KYOUEI or B.T.B brush), which was attached in the
Gantry. Use the NEW signal brush which is included in the SIGNAL SLIP RING parts.
Special Equipment
D Alcohol (2254012)
D HEPA Vacuum Cleaner : 46297933P1
D HEPA Filters for Vacuum Cleaner : 46297948P1
D Kim Wipe (2254011)
D Tie Wraps
D Felt tip marker
D Functional Check/Adjustment Manual (CDROM)
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Remove ALL of the slip ling covers, if the system has the slip ring cover at the Gantry rear side.
(Refer to Section 4-1 SLIP RING COVER REMOVAL)
NOTICE
Do NOT touch the slip ring surfaces !
414
GE MEDICAL SYSTEMS
REV 5
4-4-1
5
2154943
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80) (continued)
Verify that the Gantry is balanced for every position of the Gantry.
If it is not, perform Rotation Balance Adjustment of the Table/Gantry tab of the Functional Check/Adjustment
manual.
CAUTION
If the Gantry is NOT balanced for every position of the Gantry, the Gantry may rotate by itself
during replacing the Slip Ring assy.
Remove the Signal Brush Assy. (Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT)
Remove the signal brush support from the Gantry (see illustration 413).
Illustration 413
SIGNAL BRUSH
SUPPORT
ALLEN
SCREWS
415
GE MEDICAL SYSTEMS
REV 5
4-4-1
9
2154943
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80) (continued)
Remove the Power Brush Assy from the Gantry (see illustration 414).
a. Cut the upper two tie wraps securing the power bar cables.
b. Remove the two Allen screws.
c. Dislodge (pullout) the power holder brush assembly and mounting plate together from the support bracket by
pulling it free of the two support pins.
Illustration 414
x2
CUT
DO NOT CUT
416
GE MEDICAL SYSTEMS
REV 5
4-4-1
2154943
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80) (continued)
10
11
Switch ON the service switch on the SUB board near the TGP board.
12
Remove the eight inner slip ring covers. Each cover has two nuts and two Allen screws (see illustration 415).
Illustration 415
TWO(2) ALLEN
SCREWS
TWO(2) NUTS
INNER RING COVER 8
417
GE MEDICAL SYSTEMS
REV 5
4-4-1
13
2154943
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80) (continued)
NOTICE
Each stud bolt has two(2) washers to keep clearance between upper cover and board.
Therefore, be careful not to lose them.
f. Remove the two(2) nuts and lower cover from the Gantry.
Illustration 416
LOWER COVER
DTRE BOARD
NUT
UPPER COVER
J3
JI
SCREW
MOUNTING BOLT
WASHERS
418
J2
GE MEDICAL SYSTEMS
REV 5
4-4-1
14
2154943
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80) (continued)
Remove the Ground screws, support nut and ground plates (1 and 2) from the back(inside) of the signal brushes
(see illustration 417). Performing other places by rotating the Gantry. (By hand).
Illustration 417
Ground Cables
GROUND PLATE 1
GROUND PLATE 2
Note
Place the cables on the inside of the Gantry so that they do not catch during Gantry Rotation.
419
GE MEDICAL SYSTEMS
REV 5
4-4-1
2154943
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80) (continued)
15
Disconnect all signal ring connectors :10 Pin Connector x 2 + 4 Pin Connector x 3 as shown in illustration 418.
16
NOTICE
Irreparable Damage.
The studs and screws projecting from the back of the slip ring can easily scratch the power
slip ring surfaces during installation or removal. Handle with care.
17
Unscrew eight(8) Philip screws which hold the signal slip rings in place.
18
19
20
Clean up any wear powder. Refer to Section 3-2 SCATTERED POWDER CLEANING.
Illustration 418
4 PIN
CONNECTOR
10 PIN
CONNECTOR
PHILIP SCREWS
420
GE MEDICAL SYSTEMS
REV 5
4-4-1
21
2154943
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80) (continued)
Verify that the new signal slip ring pattern and the support metal ring is isolated (RMW) before installing to the
Gantry. However, the GND line only is 0W (see illustration 419).
Illustration 419
RMW
GND LIME
0W
421
GE MEDICAL SYSTEMS
REV 5
4-4-1
2154943
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80) (continued)
22
23
Tighten the eight Philip screws in crossangle order as shown in illustration 421.
Illustration 420
STUD HOLES
Illustration 421
4
3
1
6
5
4
2
3
7
5
2
422
GE MEDICAL SYSTEMS
REV 5
4-4-1
2154943
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80) (continued)
24
Connect all signal ring connectors :10 Pin Connector x 2 + 4 Pin Connector x 3 as shown in illustration 418.
25
Tighten the rear slip ring ground nuts as follows (see illustration 417) :
a. Tighten the rear slip ring ground screws securely, total 2 x 8 =16.
b. Tighten the rear slip ring support nuts, total 4 x 8 = 32 (14 kgf.cm).
c. Tighten the nuts again (14 kgf.cm); mark each nut after it is tightened.
Depending on the curve of the rings once may not be enough.
d. Verify that all nuts are marked as tightened, total 32.
26
27
Verify that there is NO the filings/powder from the surface of power slip rings and Gantry rear lower side.
POWER RINGS
28
Clean the signal slip ring surface using Kim wipe with alcohol.
29
Wipe off the signal slip ring surface using dry Kim Wipe.
30
Install the NEW Signal Brush Assy which is included in the SIGNAL SLIP RING parts.
(Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT)
31
Install the Power Brush Assy to the Gantry (see step 9).
423
GE MEDICAL SYSTEMS
REV 5
4-4-1
2154943
Signal Slip Ring Replacement (Panel Type EMC 100, EMC 80) (continued)
32
33
Switch OFF the service switch on the SUB board near the TGP board.
34
Rotate the Gantry using the TGP service switches (MNL/FIX/SPD2/FAST/CONT) for 2,000 rotations at fastest
speed (see illustration 48).
35
Perform the Signal Brush Contact Surface Cleaning (refer to Section 3-4 SIGNAL BRUSH CONTACT SURFACE
CLEANING).
36
37
Check the contact status of signal ring and the signal brush.
(Refer to Section 5-2-1 Signal Ring Brush Contact Check (PANEL TYPE))
38
39
424
GE MEDICAL SYSTEMS
REV 5
4-4-2
2154943
Signal Slip Ring Replacement (Panel Type No EMC)
NOTICE
Do not reinstall the OLD brush (ETC, KYOUEI or B.T.B brush), which was attached in the
Gantry. Use the NEW signal brush which is included in the SIGNAL SLIP RING parts.
Special Equipment
D Alcohol (2254012)
D HEPA Vacuum Cleaner : 46297933P1
D HEPA Filters for Vacuum Cleaner : 46297948P1
D Kim Wipe (2254011)
D Tie Wraps
D Felt tip marker
D Functional Check/Adjustment Manual (CDROM)
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Remove ALL of the slip ling covers, if the system has the slip ring cover at the Gantry rear side.
(Refer to Section 4-1 SLIP RING COVER REMOVAL)
NOTICE
Do NOT touch the slip ring surfaces !
5
Verify that the Gantry is balanced for every position of the Gantry.
If it is not, perform Rotation Balance Adjustment of the Table/Gantry tab of the Functional Check/Adjustment
manual.
CAUTION
If the Gantry is NOT balanced for every position of the Gantry, the Gantry may rotate by itself
during replacing the Slip Ring assy.
425
GE MEDICAL SYSTEMS
REV 5
4-4-2
2154943
Remove the Signal Brush Assy. (Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT)
Remove the signal brush support from the Gantry (see illustration 413).
Remove the Power Brush Assy from the Gantry (see illustration 414).
a. Cut the upper two tie wraps securing the power bar cables.
b. Remove the two Allen screws.
c. Dislodge (pullout) the power holder brush assembly and mounting plate together from the support bracket by
pulling it free of the two support pins.
10
11
Switch ON the service switch on the SUB board near the TGP board.
12
Rotate the Gantry by hand and mark the eight(8) inner slip ring covers (see illustration 422).
This will allow proper installation later since not all the covers have plastic edge guards
Illustration 422
HOME
PLASTIC
EDGE GUARDS
426
GE MEDICAL SYSTEMS
REV 5
4-4-2
13
2154943
Remove the eight(8) inner slip ring covers. Each cover has four(4) nuts (see illustration 423).
Illustration 423
BACK COVER
FOUR NUTS
14
Disconnect the ground cable nuts 1 and screws 2 , and support nuts 3 from the back (inside) of the signal
brushes (see illustration 424). Performing other places by rotating the Gantry. (By hand)
Illustration 424
Ground Cables
NUT 16 (7mm)
FLANGE NUT 16
TORQUE : 14 kg.cm
3
SUPPORT NUT
16
TORQUE : 14 kg.cm
Note
Place the cables on the inside of the Gantry so that they do not catch during Gantry Rotation.
427
GE MEDICAL SYSTEMS
REV 5
4-4-2
2154943
15
Disconnect all signal ring connectors :10 Pin Connector x 2 + 4 Pin Connector x 3 as shown in illustration 418.
16
NOTICE
Irreparable Damage.
The studs and screws projecting from the back of the slip ring can easily scratch the power
slip ring surfaces during installation or removal. Handle with care.
17
Unscrew eight(8) Philip screws which hold the signal slip rings in place.
18
19
20
Clean up any wear powder. Refer to Section 3-2 SCATTERED POWDER CLEANING.
21
Verify that the new signal slip ring pattern and the support metal ring is isolated (RMW) before installing to the
Gantry. However, the GND line only is 0W (see illustration 419).
22
23
Tighten the eight Philip screws in crossangle order as shown in illustration 421.
24
Connect all signal ring connectors :10 Pin Connector x 2 + 4 Pin Connector x 3 as shown in illustration 418.
25
Tighten the rear slip ring ground nuts as follows (see illustration 417) :
a. Tighten the rear slip ring ground nuts 1 and screws 2 and support ring nuts 3
b. Tighten the nuts again (14 kgf.cm); mark each nut after it is tightened.
Depending on the curve of the rings once may not be enough.
c. Verify that all nuts are marked as tightened, total 32.
26
Install the eight(8) inner slip ring covers (Perform steps 13 to 11 in their reverse order).
27
Verify that there is NO the filings/powder from the surface of power slip rings and Gantry rear lower side.
28
Clean the signal slip ring surface using Kim Wipe with alcohol.
29
Wipe off the signal slip ring surface using dry Kim Wipe.
30
Install the NEW Signal Brush Assy which is included in the SIGNAL SLIP RING parts.
(Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT)
31
Install the Power Brush Assy to the Gantry (see step 9).
428
GE MEDICAL SYSTEMS
REV 5
4-4-2
2154943
32
33
Switch OFF the service switch on the SUB board near the TGP board.
34
Rotate the Gantry using the TGP service switches (MNL/FIX/SPD2/FAST/CONT) for 2,000 rotations at fastest
speed (see illustration 48).
35
36
37
Check the contact status of signal ring and the signal brush.
(Refer to Section 5-2-1 Signal Ring Brush Contact Check (PANEL TYPE))
38
39
429
GE MEDICAL SYSTEMS
REV 5
4-4-3
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type)
NOTICE
The SIGNAL SLIP RING REPLACEMENT (STAINLESS ' PANEL TYPE) is required the special
service engineer who has experience this replacement. Call A.S.C, E.S.C or O.L.C. before this
work.
NOTICE
Do not reinstall the OLD brush (B.T.B brush), which was attached in the Gantry.
Use the NEW signal brush which is included in the Slip Ring Kit : 2147710.
Special Equipment
D Dial Gauge with stand
D Torque wrench
D Lock tight
D Tie Wraps
D Kim Wipe (2254011)
D Alcohol (2254012)
D Functional Check/Adjustment Manual (CDROM)
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
D Slip Ring Kit : 2147710. The detail is as follows :
Item
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Part Name
SLIP RING ASSY
SR SHIM (0.1mm)
SR SHIM (0.2mm)
SR SHIM (0.4mm)
POS PIN1 (Small)
POS PIN2 (Large)
P9223QK JIG ASSY
POWER HOLDER JIG
RATING PLAGE
PLATE ASSY
SCREW
SCREW
SIGNAL BRUSH
Maintenance Manual
Part Number
2149878
P9183JD
P9183JE
P9183JF
P9183LP
P9183LN
2113741
P9183QN
2111567
2113740
N9410ES
N9410JS
2274988
2154943
Qty.
1
10
10
10
2
2
1
1
1
1
1
1
1
1
430
Description
For Ring Frame Adjustment
For Ring Frame Adjustment
For Ring Frame Adjustment
For Brush Holder Alignment
For Brush Holder Alignment
For Signal Brush Position Alignment
For Power Brush Holder Position Alignment
For Reference only, No Locator Card
For 25MHz DAS only (Discard this for 40MHz DAS)
For 25MHz DAS only (Discard this for 40MHz DAS)
For 25MHz DAS only (Discard this for 40MHz DAS)
GE MEDICAL SYSTEMS
REV 5
4-4-3
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Remove ALL of the slip ling covers, if the system has the slip ring cover at the Gantry rear side.
(Refer to Section 4-1 SLIP RING COVER REMOVAL)
NOTICE
Do NOT touch the slip ring surfaces !
5
Verify that the Gantry is balanced for every position of the Gantry.
If it is not, perform Rotation Balance Adjustment of the Table/Gantry tab of the Functional Check/Adjustment
manual.
CAUTION
If the Gantry is NOT balanced for every position of the Gantry, the Gantry may rotate by itself
during replacing the Slip Ring assy.
Remove the Signal Brush Assy. (Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT)
Remove the signal brush support from the Gantry (see illustration 413).
Remove the Power Brush Assy from the Gantry (see illustration 414).
a. Cut the upper two tie wraps securing the power bar cables.
b. Remove the two Allen screws.
c. Dislodge (pullout) the power holder brush assembly and mounting plate together from the support bracket by
pulling it free of the two support pins.
10
11
Switch ON the service switch on the SUB board near the TGP board.
12
Rotate the Gantry by hand and mark the eight(8) inner slip ring covers (see illustration 422).
This will allow proper installation later since not all the covers have plastic edge guards.
13
Remove the eight(8) inner slip ring covers. Each cover has four(4) nuts (see illustration 423).
431
GE MEDICAL SYSTEMS
REV 5
4-4-3
14
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
Disconnect the power cables from the back (inside) of the power slip rings (see illustration 425).
a. Disconnect the 550 VDC and 550 VRTN power cables (x2 locations).
b. Disconnect the 115V power cables, one blue and one brown (x2 locations).
15
Place the power cables on the inside of the Gantry and fix the cables to the cable stay bracket using tie wraps
so that they do not catch during Gantry Rotation.
Illustration 425
DOUBLE
NUTS
BROWN 115V
(OUTSIDE)
16
Set the Gantry to 7 oclock position and lower the table to the lowest position.
17
Disconnect the ground cable attached to the back (inside) of the power slip rings (see illustration 426).
Illustration 426
STRR2
STRR (EARLY TYPE)
STRR (LATE TYPE)
GROUND CABLE
DOUBLE NUTS
FRONT VIEW
432
GE MEDICAL SYSTEMS
REV 5
4-4-3
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
18
Place the ground cable on the inside of the Gantry and fix the cables to the cable stay bracket using tie wraps
so that it does not catch during Gantry Rotation (see illustration 426).
19
Remove either STRR(early type) or STRR(late type) or STRR2 (see illustration 427).
a. Disconnect all connector from the board.
b. Disconnect all nuts from the board.
c. Use TieWraps to secure the cables/connectors disconnected from the board to the rotating Gantry frame
(see illustration 428).
d. Remove the board.
Note
Do not dispose either STRR(early type) or STRR(late type) or STRR2
Illustration 427
STRR Assy
CN1
CN2
CN2
CNF
CN1 CN3
CN3
Illustration 428
STRR2
12 NUTS
Cable TieWrapping
433
CN2
CNS
CN2
GE MEDICAL SYSTEMS
REV 5
4-4-3
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
20
21
Switch OFF the service switch on the SUB board near the TGP board.
22
Unscrew the slip ring frame support screws, sixteen, using a wrench as follows: (see illustration 429).
(A strong torque may be required since these screws are installed using lock tight.)
a. Remove all the screws except the top and bottom screws.
b. Loosen the top and bottom screws.
c. Remove the top and bottom screws while performing step 23
d. Place the sixteen(16) support screws in a safe place, they will be used again.
Illustration 429
16 SCREWS (8x2)
(TORQUE 58.5kgDcm)
434
GE MEDICAL SYSTEMS
REV 5
4-4-3
23
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
Pull the slip ring frame assembly out from the lower alignment pin while holding the top pin, two people.
Illustration 430
PUSH
PERSON A
PULL
PULL
PERSON B
24
Support the slip ring frame assembly as shown in Illustration below, two people.
25
Remove the slip ring frame assembly, pull it free from the top alignment pin.
CAUTION
The slip ring frame assembly is heavy (40kg), be very careful.
Illustration 431
PERSON A
PERSON B
DO NOT TOUCH
RING SURFACES
435
GE MEDICAL SYSTEMS
REV 5
4-4-3
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
26
Clean up any wear powder. Refer to Section 3-2 SCATTERED POWDER CLEANING.
27
Verify that the new signal slip ring pattern and the support metal ring is isolated (RMW) before installing to the
Gantry. However, the GND line only is 0W (see illustration 419).
28
Hold the new slip ring frame with the number 1 on the back facing upward.
29
Install the new slip ring frame using the positioning pins as shown in illustration below
30
Screw the slip ring frame support screws with lock tight loosely into place, sixteen. After applying lock tight, do
not take a break until all screws are firmly tightened.
31
Tighten the eight Philip screws in crossangle order as shown in illustration 421.
32
Connect the 550 VDC and 550 VRTN power cables attached to the back (inside) of the power slip rings (see
illustration 425).
33
Connect the 115V power cables, one blue and one brown (x2 locations) (see illustration 425).
34
Connect the ground cable attached to the back (inside) of the power slip rings (see illustration 426).
Illustration 432
NUMBER 1 :
TOP PIN
1
8
2
7
3
6
4
5
NUMBER 5 :
BOTTOM PIN
436
GE MEDICAL SYSTEMS
REV 5
4-4-3
35
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
For STRR : P9183YD (early type) or STRR : 2109665 (late type) (see illustration 434)
a. Remove the connector and terminal strip labelled A .
b. Remove the support plate (see illustration 433).
Illustration 433
SUPPORT PLATE
STRR Board
DUMMY
JUMPER
CONNECTOR
CN2
WIRES (2)
G, H
7mm
CONNECTORS
CN1, CN3
CONNECTOR
DUMMY
JUMPER
437
GE MEDICAL SYSTEMS
REV 5
4-4-3
36
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
CONNECTOR
CN2
CONNECTOR
TERMINAL
Z Y H G
DUMMY
JUMPER
WIRES (4)
STEEL
PLATE
37
Cut the ID tag off (it is no longer connected to the STRR or STRR2 board) of connector CN2 and tie wrap the
cable to the holes on the inside frame as shown in illustration below.
Illustration 436
GROUND CABLE
438
GE MEDICAL SYSTEMS
REV 5
4-4-3
38
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
39
40
41
Set the gauge pin on top of the inner most signal brush. Do not touch the signal rings with the gauge pin (see
illustration 437).
42
Switch ON the service switch on the SUB board near the TGP board.
43
Rotate the Gantry by hand one complete turn and measure the difference between the highest and lowest slip
ring height.
Illustration 437
GAUGE PIN
SIGNAL BRUSH
439
GE MEDICAL SYSTEMS
REV 5
4-4-3
44
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
If the total difference in height is more than 0.5 mm (+/ 0.25 mm) then proceed as follows:
a. Rotate the Gantry by hand one complete turn and locate the lowest point (farthest inward) of the slip ring.
b. Mark the lowest point with a piece of tape or chalk.
c. Rotate the Gantry until the lowest point is at the bottom (6 oclock) position.
d. Select a shim to increase the height the appropriate amount.
e. Loosen the outside and inside shim screws and insert the shim in the closest joint (see illustration 438).
f. Repeat until the total height difference is 0.5 mm or less.
45
Stick some cellophane tape on the tip of the gauge pin so that it will slip easily on the ring surface and not damage
the ring surface.
Illustration 438
SHIM
LOOSEN
SCREWS
(2)
DO NOT LOOSEN
LOOSEN
SCREWS
(2)
440
GE MEDICAL SYSTEMS
REV 5
4-4-3
46
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
Measure the total height difference for the power rings as follows (see illustration 439) :
a. Set the gauge pin on top of the outer most power ring.
b. Rotate the Gantry and measure the difference between the highest and lowest slip ring height.
c. Mark the lowest point with a piece of tape or chalk.
d. Repeat step a. to c. for each power ring.
Illustration 439
441
GE MEDICAL SYSTEMS
REV 5
4-4-3
47
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
If the total difference in height for one ring is more than 0.5 mm (+/ 0.25 mm) then proceed as follows (see illustration 440) :
a. Estimate the average lowest point for the power rings based on all the readings in step 46
b. Rotate the Gantry until the lowest point is at 9 oclock position form the Gantry back side.
c. Select a shim to increase the height the appropriate amount.
d. Loosen the outside and inside shim screws and insert the shim in the closest joint.
48
Repeat step 46 and 47 until the total height difference is 0.5 mm or less for all power rings.
Illustration 440
LOOSEN
SCREWS
(2)
SHIM
DO NOT LOOSEN
LOOSEN
SCREWS
(2)
442
GE MEDICAL SYSTEMS
REV 5
4-4-3
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
49
Measure the total height difference for each signal slip ring (refer to steps 41 and 43)
50
51
52
Measure the total height difference for each power slip ring (refer to step 46).
53
54
55
Repeat the above procedure until the total height difference is 0.5 mm or less for all slip rings.
56
Remove the dial gauge from the signal brush main support.
57
58
Rotate the Gantry, so that the lowest point measured in step 50 is at 3 oclock position from the Gantry back side.
59
Insert the two POS PIN2s (Large) into the signal brush main support (see illustration 441).
60
Align the closest pair of outer slip ring frame positioning holes with the adjustment screws. If the adjustment
screws do not fit into the positioning holes, slightly loosen the three screws (see illustration 441).
61
Screw the two POS PIN2s (Large) tightly by hand (see illustration 441).
Illustration 441
POS PIN2
(Large)
SIGNAL
BRUSH
SUPPORT
POS PIN2
(Large)
THREE
SCREWS
POS PIN1
(Small)
443
SIGNAL
BRUSH
SUPPORT
POS PIN 1
(Small)
GE MEDICAL SYSTEMS
REV 5
4-4-3
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
62
Switch OFF the service switch on the SUB board near the TGP board.
63
Adjust the signal brush main support as follows (see illustration 442) :
a. Screw the two POS PIN1s (Small) tightly by hand (see illustration 441).
b. Loosen the front and rear screws.
Illustration 442
REAR SUPPORT
SCREWS (2)
FRONT SUPPORT
SCREWS (2)
REAR SUPPORT
SCREWS (2)
c. Person A : Push the signal brush main support by hand so that two POS PIN1s (Small) and two POS PIN2s
(Large) are the status as shown in illustration 441.
Person B : Tighten, bit by bit, the front, rear and support screws.
d. Rotate 1/6 rotation(CCW) the two POS PIN1s (Small) and the two POS PIN2s (Large) (see illustration 441).
e. Verify that two POS PIN1s (Small) and two POS PIN2s (Large) are still touching to the slip ring panel. If not,
repeat from step b. to d.
f. Remove the two POS PIN1s (Small) and the two POS PIN2s (Large).
444
GE MEDICAL SYSTEMS
REV 5
4-4-3
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
64
Switch ON the service switch on the SUB board near the TGP board.
65
66
Insert the two POS PIN2s (Large) into the signal brush main support (see illustration 441).
67
Align the closest pair of outer slip ring frame positioning holes with the adjustment screws (see illustration 441).
68
69
Adjust the alignment tool until its corners are perfectly seated in the alignment holes.
70
71
Looking from the side, as in illustration, verify that the adjusting plate is comfortably in place. If it is twisted or
misaligned, then loosen the screw R and properly align it and the ring alignment tool again.
72
73
74
Remove the slip ring alignment tool from the signal brush main support.
75
SCREW R
UPPER PLATE
SCREW B
BOTTOM PLATE
445
GE MEDICAL SYSTEMS
REV 5
4-4-3
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
76
77
Perform this adjustment only if the Gantry serial number listed below, otherwise proceed to the next section:
P9183AF 111919YM7 or later
P9183AG 112234YM0 or later
P9183AH 121609YM2 or later
P9183AJ
121708YM2 or later
ProSeed
NB3038 or later
ProSeed Accell NA3041 or later
78
Rotate the Gantry, so that the lowest point measured in step 53 is at 9 oclock position from the Gantry back side.
79
80
Loosen the power brush holder, three screws accessed from the side.
81
82
Align the closest pair of outer slip ring frame positioning holes with the adjustment screws.
83
Adjust the power holder until the adjustment screws fit comfortably into the slip ring frame holes.
84
Tighten the power brush holder into place, three screws accessed from side.
85
86
SLIP
RING
FRAME
SIGNAL
BRUSH
SUPPORT
POS PIN 1
(Small)
POWER BRUSH
HOLDER SCREW 3
446
GE MEDICAL SYSTEMS
REV 5
4-4-3
2154943
Signal Slip Ring Replacement (from Stainless Type to Panel Type) (continued)
87
Install the inner slip ring covers (Perform steps 13 to 11 in their reverse order).
88
Verify that there is NO the filings/powder from the power rings and from the base frame surfaces located below
the slip ring housing. Refer to Section 3-2 SCATTERED POWDER CLEANING.
89
Clean the slip ring surface using Kim Wipe with alcohol.
90
Wipe off the signal slip ring surface using dry Kim Wipe .
91
Install the NEW Signal Brush Assy which is included in the Slip Ring Kit : 2147710.
Refer to
92
Install the Power Brush Assy to the Gantry (see step 9).
93
94
Switch OFF the service switch on the SUB board near the TGP board.
95
Rotate the Gantry using the TGP service switches (MNL/FIX/SPD2/FAST/CONT) for 2,000 rotations at fastest
speed (see illustration 48).
96
97
98
Check the contact status of signal ring and the signal brush.
(Refer to Section 5-2-1 Signal Ring Brush Contact Check (PANEL TYPE))
99
447
GE MEDICAL SYSTEMS
REV 5
4-5
2154943
NOTICE
Do NOT wipe the Power Ring.
Special Equipment
D Torque Wrench
D Kim Wipe (2254011)
D HEPA Vacuum Cleaner : 46297933P1
D HEPA Filters for Vacuum Cleaner : 46297948P1
D Tie wraps
D Felt tip marker
D Functional Check/Adjustment Manual (CDROM)
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Remove ALL of the slip ling covers, if the system has the slip ring cover at the Gantry rear side.
(Refer to Section 4-1 SLIP RING COVER REMOVAL)
NOTICE
Do NOT touch the slip ring surfaces !
5
Verify that the Gantry is balanced for every position of the Gantry.
If it is not, perform Rotation Balance Adjustment of the Table/Gantry tab of the Functional Check/Adjustment
manual.
CAUTION
If the Gantry is NOT balanced for every position of the Gantry, the Gantry may rotate by itself
during replacing the Slip Ring assy.
448
GE MEDICAL SYSTEMS
REV 5
4-5
2154943
Remove the Signal Brush Assy. Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT.
Remove the signal brush support from the Gantry (see illustration 413).
Remove the Power Brush Assy from the Gantry (see illustration 414).
a. Cut the upper two tie wraps securing the power bar cables.
b. Remove the two Allen screws.
c. Dislodge (pullout) the power holder brush assembly and mounting plate together from the support bracket by
pulling it free of the two support pins.
Switch ON the service switch on the SUB board near the TGP board.
10
Remove the eight(8) inner slip ring covers. Each cover has two(2) nuts and two(2) Allen screws (see illustration
445). There are two kinds of inner cover.
Illustration 445
NO EMC
TWO(2) ALLEN
SCREWS
FOUR NUTS
INNER RING
COVER 8
449
TWO(2)
NUTS
GE MEDICAL SYSTEMS
REV 5
2154943
4-5
11
12
Disconnect the power cables attached to the back (inside) stud of each 115V power slip ring.
Note
Place the cables on the inside of the Gantry so that they do not catch during Gantry Rotation.
13
Remove the nuts which hold the 115V slip rings in place. Manually rotate the Gantry to access the nuts (2 slip
rings X 16 nuts, 7mm).
CAUTION
Irreparable Damage.
The studs and screws projecting from the back of the slip ring can easily scratch other slip
ring surfaces during installation or removal. Handle with care.
Illustration 446
TORQUE : 24 kgf.cm
450
BROWN
(OUTSIDE)
GE MEDICAL SYSTEMS
REV 5
2154943
4-5
14
15
16
Insert the new slip rings into position; the power studs, slightly larger, should be at the 2 and 10 oclock
positions.
Illustration 447
451
GE MEDICAL SYSTEMS
REV 5
4-5
17
2154943
Tighten nut
. Then rotate the Gantry 8 nuts and tighten, 4 nuts and tighten, and 8 nuts and tighten.
Next rotate 10 nuts and repeat for 8nuts, 4 nuts, 8 nuts, etc.
(START, 8, 4, 8, 10, 8, 4, 8, 13, 8, 4, 8, 10, 8, 4, 8)
18
Tighten the slip ring nuts in crossangle order a second time (20kgf.cm). Depending on the curve of the rings
once may not be enough. Mark each nut after it is tightened.
19
Tighten again if necessary. Verify that all nuts are marked as tightened.
Illustration 448
16
8
12
5
FRONT VIEW FROM TABLE
Rotate
8 nuts
START
10 nuts
Rotate
4 nuts
8 nuts
5
13 nuts
9
10 nuts
13
8 nuts
6
8 nuts
8 nuts
10
14
4 nuts
4 nuts
14
Rotate
8 nuts
3
13
11
15
8 nuts
8 nuts
8 nuts
11
10
7
12
16
452
15
GE MEDICAL SYSTEMS
REV 5
4-5
20
2154943
21
22
Switch ON the service switch on the SUB board near the TGP board.
23
Verify that no abnormal noise sounds during the Gantry rotation by hand.
24
When the Gantry rotates manually by hand, verify that the gantry should NOT be in contact with the covers.
25
Switch ON the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
26
Switch OFF the service switch on the SUB board near the TGP board.
27
28
29
30
31
32
453
GE MEDICAL SYSTEMS
REV 5
4-6
2154943
Special Equipment
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Remove ALL of the slip ling covers. (Refer to Section 4-1 SLIP RING COVER REMOVAL)
FG Brush Replacement
POSITIONING HOLE
PIN
DO NOT LOOSEN
LOOSEN NUT
(TORQUE:25kgDcm)
454
GE MEDICAL SYSTEMS
REV 5
4-6
2154943
CIRCLE HOLE
BRUSH TIP
455
GE MEDICAL SYSTEMS
REV 5
4-7
2154943
Special Equipment
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1
Switch OFF the Rotate, XG and Slip Ring115V switches at the rear Gantry base.
Remove ALL of the slip ling covers. (Refer to Section 4-1 SLIP RING COVER REMOVAL)
Remove the CONTACT BRUSH ASSY from the Gantry back side and install the new one. (There are four(4)
CONTACT BRUSH ASSYs at the the Gantry back side.)
DO NOT LOOSEN
456
GE MEDICAL SYSTEMS
REV 4
2154943
Special Equipment
D Kim Wipe (2254011)
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1.
Switch OFF the Rotate, XG and Slip Ring switches at the rear Gantry base.
2.
3.
4.
Remove only the slip ring cover which covers the Signal Brush Assy, if the system has the slip ring cover at the
Gantry rear side. (Refer to Section 4-1 SLIP RING COVER REMOVAL)
NOTICE
Do NOT touch the slip ring surfaces !
5.
Remove the Signal Brush Assy. Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT.
6.
Visually inspect the upper and lower signal brushes for wear. There is a check line on each brush (see illustration
51). If any brush is worn below this line, the Signal Brush Assy must be replaced in 100,000 rotations.
NOTICE
Do NOT touch the signal brush surface!
Illustration 51
CHECK LINE
NEW
OUT OF SPECIFICATION
51
GE MEDICAL SYSTEMS
REV 4
5-1
7.
2154943
Illustration 52
SLIP RING
CLEAN A.
STRS BOARD
CLEAN A.
STRS BOARD
SIGNAL BRUSH ASSY
52
GE MEDICAL SYSTEMS
REV 4
5-2
5-2-1
2154943
Special Equipment
D Oscilloscope
D 1.5 V battery (Type D)
Equivalent Circuit
120
W
STRS BOARD
6VDC
(EXTERNAL BATTERY SOURCE)
53
OSCILLOSCOPE
GE MEDICAL SYSTEMS
REV 4
5-2-1
2154943
1.
Switch OFF the Rotate, XG and Slip Ring switches at the rear Gantry base.
2.
3.
4.
Remove only the slip ring cover which covers the Signal Brush Assy, if the system has the slip ring cover at the
Gantry rear side. (Refer to Section 4-1 SLIP RING COVER REMOVAL)
NOTICE
Do NOT touch the slip ring surfaces !
5.
6.
Set the Gantry to 9 oclock position and lower the table to the lowest position.
7.
8.
Remove the three socket bolts and slip ring connector cover.
Illustration 53
REAR VIEW
CONNECTOR COVER
54
GE MEDICAL SYSTEMS
REV 4
5-2-1
9.
2154943
Exchange the following system cable connectors (4 pin, 10 pin) with the jumper connectors for shorting (4 pin,
10 pin) at the gantry rear base (see illustration 54). However, cut TieWraps for these cables.
10. Tie the cables/connectors, which were disconnected in previous step, using TieWraps in order to fix them during
the Gantry rotation.
Illustration 54
Swapping Connectors
REAR VIEW
SWAP
CONNECTORS
BROWN COLOR
CONNECTOR
BLACK COLOR
CONNECTOR
TIEWRAPS
CONNECTOR COVER
55
GE MEDICAL SYSTEMS
REV 4
5-2-1
2154943
Signal Ring Brush Contact Check (PANEL TYPE) (continued)
11. Disconnect CN1A, CN1B, CN2, CN3 from the STRS board.
12. Remove the jumper pin JP1 from the STRS board.
13. Connect the Oscilloscope probes and power supply (1.5 V battery (Type D)
STRS Board as shown for CASE 1 in illustration 55.
Note
The Oscilloscope mode setting is as follows. But the example uses SONY TEKTRONIX 2230.
Vertical axis
:1.0 V/ div. (DC range)
Horizontal axis
:0.2 2.0 m sec/ div.
Menu setting
Display
Vectors : ON
Smooth : OFF
Advanced Function ACQ_MODE_SETUP_TREE
: ACC PEAK
REPETITIVE
FAST RECORD
: ACC PEAK
Illustration 55
6 VDC POWER
6 VDC POWER
TPI
TPJ1
TPJ2
CASE 1
OSCILLOSCOPE
TEST JUMPER
(REMOVED DURING TEST)
OSCILLOSCOPE
JP 1
CN1A
CN1B
6 VDC POWER
6 VDC POWER
TPA
TPC
TPE
TPG
CN2
CASE 2
OSCILLOSCOPE
OSCILLOSCOPE
STRS
56
GE MEDICAL SYSTEMS
REV 4
5-2-1
2154943
SWITCHING ORDER
4
TGP BOARD
SYS
ABT
MNL
FIX
SPD2
SPD0
FAST
CONT
SLOW
90_
SWITCHING ORDER
1
TGP BOARD
SYS
ABT
MNL
FIX
SPD2
SPD0
FAST
CONT
SLOW
90_
17. Disconnect the Oscilloscope probes and power supply (1.5 V battery (Type D)
the STRS Board as shown in illustration 55.
18. Connect the Oscilloscope probes and power supply (1.5 V battery (Type D)
STRS Board as shown for CASE 2 in illustration 55 (TPA).
19. Repeat step 15. to 16. for TPA and then for the remaining signal brushes by connecting the test lines for each
respective signal brush (TPC, TPE, and TPG).
57
GE MEDICAL SYSTEMS
REV 4
5-2-2
2154943
Signal Ring Brush Contact Check (STAINLESS TYPE)
Special Equipment
D Oscilloscope
D 1.5 V battery (Type D)
Equivalent Circuit
120
W
OSCILLOSCOPE
STRS BOARD
6VDC
(EXTERNAL BATTERY SOURCE)
58
GE MEDICAL SYSTEMS
REV 4
5-2-2
2154943
Signal Ring Brush Contact Check (STAINLESS TYPE) (continued)
1.
Switch OFF the Rotate, XG and Slip Ring switches at the rear Gantry base.
2.
3.
4.
5.
Set the Gantry to 7 oclock position and lower the table to the lowest position.
6.
7.
Remove the four nuts and the back side cover which the STRR or STRR2 board exists on.
Illustration 56
BACK COVER
FOUR NUTS
STRR/STRR
2
STRR(2109665)
STRR2(P9183YN)
STRR(P9183YD)
CN1
CN2
CN2
CNF
CN1
CN3
59
CN2
CNS
CN3
CN1
GE MEDICAL SYSTEMS
REV 4
5-2-2
Perform the following job according to STRR board type (see illustration 56):
For STRR board (P9183YD)
a. Disconnect CN1, CN2, CN3 and black/white cable from the STRR board.
b. Remove ten nuts from the STRR board.
c. Use TieWraps to secure the cables/connectors disconnected from the STRR board to the rotating Gantry
frame.
d. Attach five short cables to the A stud $ B stud, C stud $ D stud, E stud$ F stud, G stud $ H stud, I stud$ J
stud where the STRR board was installed (see illustration 57).
Note
The short cable is included as part of P9180GM which is shipped with the CT system.
Illustration 57
TO GANTRY CENTER
SHORT CABLE
8.
2154943
510
GE MEDICAL SYSTEMS
REV 4
5-2-2
9.
2154943
Disconnect CN1A, CN1B, CN2, CN3 from the STRS board (see illustration 55).
10. Remove the jumper pin JP1 from the STRS board (see illustration 55).
11. Connect the Oscilloscope probes and power supply (1.5 V battery (Type D)
STRS Board as shown for CASE 1 (see illustration 55).
Note
The Oscilloscope mode setting is as follows. But the example uses SONY TEKTRONIX 2230.
Vertical axis
:1.0 V/ div. (DC range)
Horizontal axis
:0.2 2.0 m sec/ div.
Menu setting
Display
Vectors : ON
Smooth : OFF
Advanced Function ACQ_MODE_SETUP_TREE
REPETITIVE
: ACC PEAK
FAST RECORD
: ACC PEAK
12. Switch ON the Rotate switch at the rear Gantry base.
13. Rotate the Gantry using the TGP service switches (MNL, FIX, SPD2, FAST, CONT), and then either turn OFF
the breaker on the Servo Amp. or turn ON the Service Switch and while the Gantry is rotating by its own inertia,
continuously verify that there are no small drops or fluctuations in voltage. There should be no voltage drop greater than 0.5 volts.
14. Change only two TGP service switches (MNLOFF, CONTHOME).
15. Disconnect the Oscilloscope probes and power supply (1.5 V battery (Type D)
the STRS Board (see illustration 55).
16. Perform the following job according to STRR board type (see illustration 55):
For STRR board (P9183YD)
a. Connect the lines as shown in CASE 2 of illustration 55 (TPA), and attach the jumper across the respective
studs at the STRR board location (for A & B slip rings).
For STRR board (2109665) and STRR2 board (P9183YN)
a. Connect the lines as shown in CASE 2 of illustration 55 (TPA).
17. Repeat from step 13. to 14. for TPA and then for the remaining signal brushes by connecting the test lines for
each respective signal brush (TPC, TPE, and TPG) (see illustration 55).
511
GE MEDICAL SYSTEMS
REV 4
5-3
2154943
5-3-1
Special Equipment
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1.
2.
Switch OFF the Rotate, XG and Slip Ring switches at the rear Gantry base.
3.
4.
5.
Remove the slip ring cover, if the system has the slip ring cover at the Gantry rear side. (Refer to Section 4-1
SLIP RING COVER REMOVAL)
6.
Remove or disconnect the two inner connectors located at the back side of the slip ring (see illustration 58).
Illustration 58
Swapping Connectors
TIEWRAPS
JUMPER
SYSTEM
SYSTEM
JUMPER
BLACK COLOR
CONNECTOR
CONNECTOR COVER
512
BROWN COLOR
CONNECTOR
GE MEDICAL SYSTEMS
REV 4
5-3-1
7.
2154943
Signal Brush Short Circuit Check (PANEL TYPE) (continued)
Remove the two cable connectors, two fiber optic connectors and 20 screws on the STRS board, then remove
the STRS board (see illustration 59). This isolates the slip rings from the external connections.
Illustration 59
FIBER OPTIC
CONNECTORS
20 SCREWS
CONNECTORS
NOTICE
The resistance measurement must be performed at the brush support. Never place the multimeter probe onto the slip ring surface. It can permanently damage the slip ring.
8.
Measure the resistance between the signal brushes at the brush support. Perform the measurement with all possible combinations (see illustration 510).
9.
The measurement must be 40W between DAS signal brushes and infinite (open) for the other combinations (see
illustration 510).
Illustration 510
BRUSH
SUPPORT
513
GE MEDICAL SYSTEMS
REV 4
5-3-2
2154943
Signal Brush Short Circuit Check (STAINLESS TYPE)
Special Equipment
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1.
Set the Gantry to 1 oclock position and lower the table to the lowest position.
2.
Switch OFF the Rotate, XG and Slip Ring switches at the rear Gantry base.
3.
4.
5.
Remove the four nuts and the back side cover which STRR or STRR2 board exists on (see illustration 56).
6.
Perform the following job according to STRR board type (see illustration 56):
For STRR board (P9183YD)
a. DIsconnect all connectors (CN1, CN2, CN3) and WHITE color cable, BLACK color cable from STRR.
For STRR board (2109665)
a. DIsconnect all connectors (CN1, CN2, CN3, CNS) and WHITE color cable, BLACK color cable from STRR.
For STRR board (P9183YN)
a. DIsconnect all connectors (CN1, CN2, CNS) from STRR2.
7.
Remove the two cable connectors, two fiber optic connectors and 20 screws on the STRS board, then remove
the STRS board (see illustration 59). This isolates the slip rings from the external connections.
NOTICE
The resistance measurement must be performed at the brush support. Never place the multimeter probe onto the slip ring surface. It can permanently damage the slip ring.
8.
Measure the resistance between the signal brushes at the brush support. Perform the measurement with all possible combinations (see illustration 510).
9.
The measurement must be 40W between DAS signal brushes and infinite (open) for the other combinations (see
illustration 510).
514
GE MEDICAL SYSTEMS
REV 4
5-4
2154943
Special Equipment
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1.
Switch OFF the Rotate, XG and Slip Ring switches at the rear Gantry base.
2.
3.
4.
Remove only the slip ring cover which covers the Power Brush Assy, if the system has the slip ring cover at the
Gantry rear side. (Refer to Section 4-1 SLIP RING COVER REMOVAL)
NOTICE
Do NOT touch the slip ring surfaces !
5.
Illustration 511
PLASTIC COVER
515
GE MEDICAL SYSTEMS
REV 4
5-4
6.
2154943
Illustration 512
POWER BAR
CHECK LINE
POWER BRUSH
If any brush has reached the power bar surface, the Power Brush must be replaced in 100,000 rotations.
(Refer to Section 4-3 POWER BRUSH REPLACEMENT)
516
GE MEDICAL SYSTEMS
REV 4
5-5
2154943
Special Equipment
D Kim Wipe (2254011)
D Gloves : Large size :46194427P400, XLarge size :46194427P401
Note
Use the Gloves during working in order to avoid the slip ring surface contamination and the brush
tip contamination by body oil.
1.
Remove all FG brushes from the Gantry. Refer to Section 4-6 FG BRUSH ASSY REPLACEMENT (EMC 100).
2.
Visually inspect the upper and lower FG brushes for wear. There is a check line on each brush (see illustration
513). If any brush is worn below this line, the FG brush must be replaced in 100,000 rotations.
NOTICE
Do NOT touch the FG brush surface!
Illustration 513
CHECK LINE
NEW
OUT OF SPECIFICATION
517
GE MEDICAL SYSTEMS
REV 4
5-6
2154943
1.
2.
3.
CALIBRATION
.
.
Select No. = $$ CR
4.
5.
6.
:
:
:
:
:
7.
8.
9.
Set the display to normal status. (Press the [Cancel] key a a few times.)
518
GE MEDICAL SYSTEMS
REV 4
5-7
1.
2154943
TEST SCAN
Execute the following scan only.
Table 51
Test Scan
System
Sytec SRi
Sytec Synergy
Legato
ProSpeed VX
ProSpeed
ProSpeed Plus
ProSeed Accell
ProSeed Accell EI
ProSeed
ProSeed EF
ProSpeed S
ProSpeed
ProSpeed S Fast
ProSpeed SX
ProSpeed SX Power
Lemage
Lemage SX
Lemage SX/E
Lemage Supreme
ProSpeed Advantage
ProSpeed SA
PTM
FOV
(cm)
KV
TIME
(sec)
mA
FOCUS
THK
(mm)
42cm P.P
42
120
1.5
200
35cm P.P
50
120
350
42cm P.P
50
120
1.5
42cm P.P
50
120
42cm P.P
50
42cm P.P
50
SPECIFICATION
XC
XO
10
M10
x13
M103
Large
10
M10
x9
M103
200
10
M10
x15.5
M103
1.5
250
10
M10
x16
M103
120
1.5
160
10
M10
x15.5
M103
120
250
Large
10
M10
x15.2
Visually check each image for rings, bands, streaks, shadings or artifacts near the image center. And verify that
there are no error message.
Position a 2 cm diameter ROI at the locations listed below and record the results on the Data/Work sheets.
Xc Mean CT number in the center ROI.
Xo Mean CT number 10 cm off center ROI.
sC Standard deviation of the center ROI.
s0 Standard deviation of the 10 cm off center ROI.
IMAGE
+Y
Xc (0.0 cm)
519
GE MEDICAL SYSTEMS
blank
2154943
520
GE MEDICAL SYSTEMS
REV 4
2154943
6-1
ERROR CODE
When an error occurs due to the slip ring trouble, the error messages are shown on the OC screen. However, these
messages do not always imply to the slip ring trouble. there are always other possible causes (DBM, Optical Cable,
DTR etc.).
Table 61 shows the errors which only the slip ring is caused.
Table 61
Table Title
Error No.
Error Name
Error Code
Trouble Line
Error
101
Scan Error
DAS TRIG
Error
104
DAS TRIG
Error
110
1, 2
Error
112
1, 2
RX, TX,TGP
Error
117
XG Error
x 49
Error
203
DAS LINE
Error
204
DAS LINE
Error
205
DAS LINE
61
GE MEDICAL SYSTEMS
REV 4
6-2
2154943
Use this trouble shooting flowchart, when the error messages are shown on the OC screen.
Note
If you are turned around the same loop in this flow chart two times, stop anything and call A.S.C,
E.S.C or O.L.C.
Illustration 61
START
SIGNAL
BRUSH
WEAR CHECK
NG
NG
D
D
D
D
D
NG
D
D
D
D
NG
OK
SIGNAL BRUSH
RING $ BRUSH
CONTACT
CHECK
OK
SIGNAL BRUSH
SHORT SIRCUIT
CHECK
OK
TOTAL TEST
D DAS TRANSFER TEST
D TEST SCAN
OK
END
62
GE MEDICAL SYSTEMS
REV 4
6-3
2154943
Illustration 62
START
POWER BRUSH
WEAR CHECK
NG
NG
OK
OK
*1) :
END
63
GE MEDICAL SYSTEMS
REV 4
6-4
2154943
Introduction
Multiwire aging is to be performed on systems with a voltage drop due to slip ring/brush contact greater
than 0.5 Volts. This voltage drop is checked during the first three sections of this procedure; do not perform this multi
wire aging if the voltage drop is within specification.
Cautions
NOTICE
Do NOT perform the AGING for the PANEL RING !!!
CAUTION
Possible irreparable damage. The inner signal slip rings are very sensitive, never touch them.
CAUTION
Rotation Hazard.
Part of this procedure will be conducted with the Service Switch ON.
Ensure that the Gantry is balanced properly at every position
NOTICE
Excessive aging reduces the life of the signal brushes.
In any case, total aging must NOT exceed 20 hours (90000 rotations).
64
GE MEDICAL SYSTEMS
REV 4
6-4
2154943
Special Equipment
D Aging Kit : 2114540
Parts Name
Parts No.
Qty.
21145411
2118754
P9183MY
4
5
6
7
8
9
Screw
Screw
Washer
Signal Brush
Color Chart
Manual
N9506JS
N9512ZX
N9500WS
P9183FH
2120475
2119292
4
1
1
3
1
1
D Oscilloscope
D 1.5 V battery
65
GE MEDICAL SYSTEMS
REV 4
6-4
2154943
Flow Chart
The flow chart is only main items. The detail is from next subsection.
Illustration 63
SIGNAL BRUSH
RING $ BRUSH CONTACT
CHECK
OK
NG
NO
SYSTEM CAN
AFFORD TO HAVE 3 WEEKS TO SHOOT
THIS TROUBLE
YES
YES
NO
SIGNAL BRUSH
RING $ BRUSH CONTACT
CHECK
NG
OK
D
D
TOTAL TEST
DAS TRANSFER TEST
NG
TEST SCAN
OK
END
66
GE MEDICAL SYSTEMS
REV 4
6-4
2154943
Perform Signal Ring Brush Contact Check. Refer to Section 5-2 SIGNAL RING BRUSH CONTACT
CHECK.
2.
Proceed as follows:
Voltage drop < 0.5 Volts: Do not perform this procedure, and return the system to normal. Investigate other
causes.
Voltage drop > 0.5 Volts: proceed to the next subsection (Air Condition Check).
If the system can afford to have 3 weeks to shoot the trouble, perform this subsection, if not, perform next sub
section (Stainless Slip Ring Aging).
2.
File the surfaces of three signal brushes (in Aging kit) from the kit using Sand Paper.
3.
Attach these three signal brushes on the rear cover of the Gantry using Adhesive Tape. (This is to investigate
the air condition in the scan room.) (see illustration 64)
Illustration 64
ADHESIVE TAPE
4.
Three weeks later after attaching the brushes, compare the color of the brush surfaces with the Color Chart (in
Aging kit) (see illustration 65).
If the color of the brush surfaces has turned similar to color of STEP 1 or STEP 2 or STEP 3, proceed next sub
section (Stainless Slip Ring Aging). This result indicates that the air condition is no problem.
If the color of the brush surfaces has turned similar to color of STEP 4 or STEP 5 or STEP 6, this indicates that
the air contains sulfur or sulfur dioxide exceeding the acceptable level. Stop working and CALL A.S.C, E.S.C
or O.L.C.
Illustration 65
Color Chart
STEP 1
67
GE MEDICAL SYSTEMS
REV 4
6-4
2154943
Remove all slip ring covers. Refer to Section 4-1 SLIP RING COVER REMOVAL.
2.
Remove the signal brush assy. Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT.
3.
Clean up the signal slip ring. Refer to Section 3-3 SIGNAL SLIP RING CLEANING.
4.
Turn ON the Rotate , Table/Tilt and Control switches and turn OFF the XG and Slip Ring switches at
rear Gantry base.
5.
Rotate the Gantry automatically. To set at the slowest speed, set the TGP Service Switches to MNL, FIX, SPD0,
SLOW, CONT in this order. Immediately the ABT/SVE/FIX switch is set to the FIX position, the Gantry will starts
to rotate.
Illustration 66
SWITCHING ORDER
4
TGP BOARD
SYS
MNL
ABT
FIX
SPD2
SPD0
FAST
CONT
SLOW
90_
6.
Hold the Kim Wipe with alcohol onto the surface of one signal slip ring and wipe the ring using Kim Wipe.
7.
8.
Dry the surface of the slip ring totally with dry Kim Wipe.
9.
10. Set the TGP Service Switches to OFF, SVE, SPD0, SLOW, CONT
11. Turn OFF the Rotate switch at rear Gantry base.
68
GE MEDICAL SYSTEMS
REV 4
6-4
2154943
14. Adjust the ring cleaner assemblies so that they contact the signal slip ring surface at every gantry position as
follows (see illustration 67):
a. Loosen the cleaner support assy screws,
b. Rotate the gantry by hand and roughly adjust the center of the ring cleaner assemblies over the ring surfaces.
c. Tighten the cleaner support assy screws.
d. Loosen the ring cleaner assy M5 screws
e. Adjust the ring cleaner assy positions while rotating the gantry by hand.
f. Loosen the ten ring cleaner assy M3 screws(10)
one at a time.
g. Adjust each ring cleaner assy position while rotating the gantry by hand until every cleaner completely covers
each slip ring surface at every gantry position.
Illustration 67
N9512ZX
2
1
SCREW C
69
GE MEDICAL SYSTEMS
REV 4
6-4
2154943
Illustration 68
610
GE MEDICAL SYSTEMS
REV 4
6-4
2154943
611
GE MEDICAL SYSTEMS
REV 4
6-4
2154943
Install STRS board. Do NOT connect all connectors on the STRS board yet. Refer to Section 4-2 SIGNAL
BRUSH REPLACEMENT.
2.
Perform SIGNAL RING BRUSH CONTACT CHECK. Refer to Section 5-2 SIGNAL RING BRUSH CONTACT CHECK.
3.
Proceed as follows:
Voltage drop < 0.5 Volts: return the system to normal. This result is that the aging was succeeded.
Voltage drop > 0.5 Volts: call A.S.C, E.S.C or O.L.C.
4.
Connect STRS boards all connectors. Refer to Section 4-2 SIGNAL BRUSH REPLACEMENT.
5.
Install all slip ring covers. Refer to Section 4-1 SLIP RING COVER REMOVAL.
Total Test
1.
Turn OFF the Rotate and Slip Ring 115V switches at rear Gantry base.
2.
3.
Turn ON the Rotate and Slip Ring 115V switches at rear Gantry base.
4.
612
GE MEDICAL SYSTEMS
REV 3
2154943
Publication
Description
4175032
Alternating Current
3351
3351
3N
Direct Current
4175019
348
4175008
4175007
Emergency Stop
A1
GE MEDICAL SYSTEMS
REV 3
2154943
Publication
Description
Type B
4175339
4175009
Standby
Start
Table Set
Abort
Intercom
A2
GE MEDICAL SYSTEMS
REV 3
2154943
Description
Microphone (Mic)
Contrast
Brightness
_
_
A3
GE MEDICAL SYSTEMS
REV 3
2154943
CLASS 1 EQUIPMENT
Any permanently installed equipment containing operator or patient accessible surfaces must provide backup protection against electric shock,in case the BASIC INSULATION fails. In addition to BASIC INSULATION,Class1 equipment contains a direct connection to a PROTECTIVE(EARTH) CONDUCTOR which prevents shocks when a person
touches a broken piece of equipment or touches two different equipment surfaces simultaneously.
TYPE B EQUIPMENT
CLASS I, II, or III EQUIPMENT or EQUIPMENT with INTERNAL ELECTRICAL POWER SOURCES provide an adequate degree of protection against electric shock arising from (allowable) LEAKAGE CURRENTS or a breakdown
in the reliability of the protective earth connection.
ORDINARY EQUIPMENT
Enclosed EQUIPMENT without protection against the ingress of water.
OPERATION 0f EQUIPMENT
CLEANING
The ProSpeed S series system is NOT WATERPROOF. It is NOT designed to protect internal components against
the ingress of liquids.Clean external system surfaces(Gantry,table,consoles and accessories)with a soft cloth dipped
in hot water and wrung DAMP/DRY. (NOT dripping!) IF NECESSARY, use only mild (dish washing liquid) soap to remove dirt.
NOTICE
Avoid damage to equipment! Some spray and wipecleaners etch and permanently cloud
clear plastic surfaces!! Use only warm water and mild liquid soap to clean surfaces.
A4