2. Press: 20 ENTER
Press: PRGM
Note: Determination of
total copper needs a prior
digestion (see Digestion in
Section 2).
177
COPPER, continued
5. Press: ZERO
6. Fill another sample
The cursor will move to cell with 10 mL of
the sample.
the right, then the
display will show:
0.00 mg/L Cu
Note: If Reagent Blank
Correction is on, the
display may flash limit.
9. Within 30 minutes
after the timer beeps,
place the prepared
sample into the cell
holder. Tightly cover the
sample cell with the
instrument cap.
178
8. Press:
TIMER ENTER
A two-minute reaction
period will begin.
Note: Accuracy is not
affected by
undissolved powder.
COPPER, continued
Using AccuVac Ampuls
Method 8026
2. Press: 20 ENTER
Press: PRGM
Note: Determination of
total copper needs a prior
digestion (see Digestion in
Section 2).
6. Fill a CuVer 2
The cursor will move to Copper Reagent
AccuVac Ampul with
the right, then the
sample.
display will show:
Note: If Reagent Blank
Correction is on, the
display may flash limit.
8. Press:
5. Press: ZERO
0.00 mg/L Cu
179
TIMER ENTER
A two-minute reaction
period will begin.
COPPER, continued
180
COPPER, continued
Sampling and Storage
Collect samples in acid-cleaned glass or plastic containers.
Adjust the pH to 2 or less with nitric acid (about 2 mL per liter).
Store preserved samples up to six months at room temperature.
Before analysis, adjust the pH to 4 to 6 with 8 N potassium
hydroxide. Do not exceed pH 6, as copper may precipitate.
Correct the test result for volume additions; see Correction for
Volume Additions in Section 1 for more information. If only
dissolved copper is to be determined, filter the sample before acid
addition using the labware listed under Optional Apparatus.
Accuracy Check
Standard Additions Method
a) Fill three 25-mL graduated mixing cylinders with 25 mL of
sample.
b) Snap the neck off a Copper Voluette Ampule Standard, 75
mg/L
as Cu.
c) Use the TenSette Pipet to add 0.1, 0.2, and 0.3 mL of
standard, respectively, to the mixing cylinders. Stopper
and mix thoroughly.
d) For analysis with AccuVac Ampuls, transfer the solutions
to dry, clean 50-mL beakers to fill the ampules. For
analysis with powder pillows, transfer only 10 mL of the
solution to 10-mL sample cells.
e) Analyze each sample as described in the procedure. The
copper concentration should increase about 0.3 mg/L for
each 0.1 mL of standard added.
f) If these increases do not occur, see Standard Additions in
Section 1 for more information.
Standard Solution Method
Prepare a 1.00 mg/L copper standard by pipetting 1.00 mL of
Copper Standard Solution, 100 mg/L as Cu, into 100-mL
volumetric flask. Dilute to volume with deionized water and mix
well. Prepare this solution daily. Using this solution as the
sample, perform the copper procedure as described above.
Method Performance
Precision
In a single laboratory, using a standard solution of 2.25 mg/L Cu
181
COPPER, continued
and two representative lots of reagent with the instrument, a
single operator obtained a standard deviation of 0.02 mg/L Cu.
In a single laboratory, using a standard solution of 2.25 mg/L Cu
and two representative lots of AccuVac Ampuls with the
instrument, a single operator obtained a standard deviation of
0.02 mg/L Cu.
Estimated Detection Limit (EDL)
The EDL for program 20 (Powder Pillows and AccuVac Ampuls)
is 0.02 mg/L Cu. For more information on derivation and use of
Hachs estimated detection limit, see Section 1.
Interferences
Interfering Substances and Suggested Treatments for Powder
Pillows
Interfering
Substance
Acidity
Aluminum, Al3+
Cyanide, CN-
Hardness
Iron, Fe3+
Silver, Ag+
COPPER, continued
other complexing agents, use a Free Copper Reagent Powder
Pillow in place of the CuVer 1 pillow in Step 4. Results in Step
10 will be free copper only. Add a Hydrosulfite Reagent Powder
Pillow to the same sample and
re-read the result. This result will include the total dissolved
copper (free and complexed).
Interfering Substances and Suggested Treatments for AccuVac
Ampuls
Interfering
Substance
Acidity
Aluminum, Al3+
Cyanide, CN-
Hardness
3+
Iron, Fe
Silver, Ag+
183
COPPER, continued
REQUIRED REAGENTS & APPARATUS (Using Powder Pillows)
Description
Quantity Required
Per Test
Unit
Cat. No.
CuVer 1 Copper Reagent Powder Pillows ...............1 pillow ............ ...100/pkg.......... 21058-69
Sample Cell, 10-20-25 mL, w/cap ................................ 2......................... 6/pkg.......... 24019-06
REQUIRED REAGENTS & APPARATUS (Using AccuVac Ampuls)
CuVer 2 Copper Reagent AccuVac Ampuls ............1 ampul ................. 25/pkg.......... 25040-25
Beaker, 50 mL............................................................... 1........................... each........... 500-41H
OPTIONAL REAGENTS
Copper Standard Solution, 100 mg/L ....................................................100 mL.............. 128-42
Copper Standard Solution, Voluette Ampule, 75 mg/L Cu, 10 mL........ 16/pkg.......... 14247-10
CuVer 2 Reagent Powder Pillows ........................................................ 100/pkg.......... 21882-99
Formaldehyde, 37%, ACS .........................................................100 mL* MDB............ 2059-32
Free Copper Reagent Powder Pillows .................................................. 100/pkg.......... 21186-69
Hydrochloric Acid Solution, 6.0 N........................................................500 mL.............. 884-49
Hydrosulfite Reagent Powder Pillows.................................................. 100/pkg.......... 21188-69
Metals Drinking Water Standard, LR for Cu, Fe, Mn ...........................500 mL.......... 28337-49
Metals Drinking Water Standard, HR for Cu, Fe, Mn ...........................500 mL.......... 28336-49
Nitric Acid, ACS....................................................................................500 mL.............. 152-49
Nitric Acid Solution, 1:1 .......................................................................500 mL............ 2540-49
Potassium Chloride Solution, saturated .................................................100 mL .............. 765-42
Potassium Hydroxide Standard Solution, 8.0 N ........................ 100 mL* MDB........... 282-32H
Sodium Hydroxide Standard Solution, 5.0 N ............................ 100 mL* MDB............ 2450-32
Water, deionized............................................................................................4 L.............. 272-56
OPTIONAL APPARATUS
Description
Unit
Cat. No.
184
COPPER, continued
Pipet Tips, for 19700-01 TenSette Pipet .................................................50/pkg ..........21856-96
Pipet Tips, for 19700-01 TenSette Pipet .............................................1000/pkg ..........21856-28
Pipet, volumetric, Class A, 1.00 mL ...........................................................each ..........14515-35
Pipet Filler, safety bulb ..............................................................................each ..........14651-00