Sie sind auf Seite 1von 4

Laboratory

Gatrad Cal-test Laboratories and Research Pvt. Ltd., 248, GIDC


Industrial Area, Telephone Exchange Road, Odhav, Ahmedabad, Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Thermal Calibration

Issue Date

27.01.2014

Certificate Number

C-0337

Valid Until

26.01.2016

Last Amended on

Page

1 of 4

Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

-80C to -40C
-40C to 100C
100C to 600C

0.12C
0.04C
0.14C

Using SPRT sensor with


Chub-E4 DRO of Hart
Scientific and
Temperature Baths.
Model : 9173 metrology
well calibrator & Model:
7380 Hart scientific
compact bath
by Comparison Method

600 C to 1200 C

2C

Using Standard S type


thermocouple with
Chub-E4 DRO of Hart
Scientific and
Temperature Baths.
by Comparison Method

-80C to -40C
-40C to 400C
400C to 600C

0.12C
0.04C
0.14C

Using SPRT sensor with


Chub-E4 DRO of Hart
Scientific and
Temperature Baths
Model : 9173 metrology
well calibrator &
Model: 7380 Hart
scientific compact bath
by Comparison Method

I. TEMPERATURE
1.

2.

Temperature Indicator /
Gauge/ Controller /
Recorder/ Transmitter
with Sensor$

RTD Sensor$

Laboratory

Gatrad Cal-test Laboratories and Research Pvt. Ltd., 248, GIDC


Industrial Area, Telephone Exchange Road, Odhav, Ahmedabad, Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Thermal Calibration

Issue Date

27.01.2014

Certificate Number

C-0337

Valid Until

26.01.2016

Last Amended on

Page

2 of 4

Quantity Measured/
Instrument

3.

Thermocouple Sensor$

4.

Liquid in Glass
thermometer$

5.

Temperature Indicator of
Baths/ Oven / Dry Block$

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

-80C to -40C
-40C to 100C
100C to 600C

0.12C
0.04C
0.14C

Using SPRT sensor with


Chub-E4 DRO of Hart
Scientific and
Temperature Baths
Model: 9173 metrology
well calibrator &
Model: 7380 Hart
scientific compact bath
by Comparison Method

600 C to 1200 C

2 C

Using Standards type


thermocouple with
Chub-E4 DRO of Hart
Scientific and
Temperature Baths.
by Comparison Method

-80C to -40C
-40C to 100C

0.12C
0.04C

Using SPRT sensor with


Chub-E4 DRO of Hart
Scientific and Liquid
Temperature Baths
Model : 9173 metrology
well calibrator
by Comparison Method

-80C to 0C
0C to 600C
600C to 1200C

0.13C
0.06C
2 C

Using SPRT sensor with


Chub-E4 DRO of Hart
By using Standard S
type thermocouple with
Chub-E4 DRO of Hart
Scientific
by Comparison Method

Laboratory

Gatrad Cal-test Laboratories and Research Pvt. Ltd., 248, GIDC


Industrial Area, Telephone Exchange Road, Odhav, Ahmedabad, Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Thermal Calibration

Issue Date

27.01.2014

Certificate Number

C-0337

Valid Until

26.01.2016

Last Amended on

Page

3 of 4

Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

6.

Temperature Indicator /
Thermo-Humidity meter/
Recorder / Logger with
inbuilt Sensor$

-5C to 50C

1.0C

Using ThermoHumidity meter


Rotronic Hgroflex4 and
Incubator
by Comparison Method

7.

Temperature Indicator /
Controller / Recorder /
Transmitter with Sensors/
RTDs / Thermocouples
sensors

-25 C to 150 C
150 C to 600 C

0.2C
0.6C

Using PRT sensor with


Indicator and
Temperature Baths

600 C to 1200 C

2.0 C

Using S-type
thermocouple with
Indicator and
Temperature Baths

-60 C to 600 C

0.5C

Using PRT sensor with


Indicator

600 C to 1200 C

2.0 C

Using S-type
thermocouple with
Indicator

8.

Temperature Indicators
of Baths / Furnace /
Incubators / Ovens /
Coolers /Freezers
(Single Point)
II. HUMIDITY
1.

Humidity Indicators/
Thermo-Hygrometer$

15% to 95% RH @25

2.1% RH

Using Humidity
Generator along with
Humidity meter

2.

Humidity Indicators/
Thermo-Hygrometer
With Sensor$

35% RH
50% RH
75% RH
95% RH

0.50% RH
0.90% RH
0.90% RH
1.20% RH

Using Humidity
Standard Solutions
by Comparison Method

Laboratory

Gatrad Cal-test Laboratories and Research Pvt. Ltd., 248, GIDC


Industrial Area, Telephone Exchange Road, Odhav, Ahmedabad, Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Thermal Calibration

Issue Date

27.01.2014

Certificate Number

C-0337

Valid Until

26.01.2016

Last Amended on

Page

4 of 4

Quantity Measured/
Instrument

3.

Humidity Indicators of
Humidity Chambers/
Generators/ Calibrators
Single Point Calibration

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

15% RH to 95% RH @25C

2.8%RH

Using Rotronic
Humidity Meter

* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%


$

Only in Permanent Laboratory


Only for Site Calibration
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing actual
environmental conditions and master equipment used.