Microtech Calibration Lab, Plot No. 143, Indira Housing Society, Kabnoor,
Ichalkaranji, Kolhapur, Maharashtra
Mechanical Calibration
Issue Date
03.03.2014
Certificate Number
C-0229
Valid Until
02.03.2016
Last Amended on
Page
1 of 6
Quantity Measured/
Instrument
Range / Frequency
*Calibration Measurement
Capability ()
Remarks
Upto 100 mm
1.4 m
Upto 100 mm
100 mm to 200 mm
200 mm to 400 mm
400 mm to 600 mm
3.5 m
5.0 m
7.3 m
8.5 m
Upto 100 mm
4.1 m
Upto 600 mm
Upto 600 mm
15.0 m
18.5 m
Upto 600 mm
Upto 600 mm
16.0 m
19.5 m
Upto 300 mm
15.0 m
0.01mm to 10 mm
0.001mm to 1 mm
4.7 m
1.7 m
I. DIMENSION
1.
OUTSIDE
MICROMETER$
L.C. 0.001mm
2.
OUTSIDE
MICROMETER$
L.C. 0.01 mm
3.
INSIDE MICROMETER$
L.C.0.01 mm
4.
5.
VERNIER CALIPERS$
L.C. : 0.01/ 0.02 mm
HEIGHT GAUGE$
L.C. : 0.01 / 0.02 mm
6.
DEPTH VERNIER$
7.
PLUNGER DIAL$
Laboratory
Microtech Calibration Lab, Plot No. 143, Indira Housing Society, Kabnoor,
Ichalkaranji, Kolhapur, Maharashtra
Mechanical Calibration
Issue Date
03.03.2014
Certificate Number
C-0229
Valid Until
02.03.2016
Last Amended on
Page
2 of 6
Quantity Measured/
Instrument
8.
9.
10.
Range / Frequency
*Calibration Measurement
Capability ()
Remarks
Upto 1mm
3.8 m
L.C. 0.001 mm
Upto 1mm
1.6 m
2.5 m
Upto 50 mm
58.0 m
Upto 150 mm
10.4 m
BORE GAUGE$
(Transmission Accuracy
only)
PISTOL CALIPER$
L.C. 0.1mm
11.
V BLOCK$
(Parallelism &
Squareness
Symetricity)
12.
MEASURING SCALE $
L.C. 1.0 mm
Upto 1000 mm
144 m
MEASURING TAPE$
L.C. 1.0 mm
Upto 5000mm
250.0 m
13.
Laboratory
Microtech Calibration Lab, Plot No. 143, Indira Housing Society, Kabnoor,
Ichalkaranji, Kolhapur, Maharashtra
Mechanical Calibration
Issue Date
03.03.2014
Certificate Number
C-0229
Valid Until
02.03.2016
Last Amended on
Page
3 of 6
Quantity Measured/
Instrument
Range / Frequency
*Calibration Measurement
Capability ()
Remarks
DEGREE PROTECTOR$
L.C.1
0 to 90 to 0
40 min of arc
BEVEL PROTECTOR$
L.C. 5
0 to 90 to 0
5 min of arc
COMBINATION SET$
L.C.1
0 to 90 to 0
40 min of arc
17.
FEELER GAUGE$
Upto 5.0mm
1.2 m
18.
CYLINDRICAL
SETTING MASTER$
Upto 100mm
100 mm to 300mm
1.7 m
3.3 m
19.
SETTING PIECE /
LENGTH BAR$
Upto 100mm
100 mm to 300mm
1.7 m
3.3 m
14.
15.
16.
Laboratory
Microtech Calibration Lab, Plot No. 143, Indira Housing Society, Kabnoor,
Ichalkaranji, Kolhapur, Maharashtra
Mechanical Calibration
Issue Date
03.03.2014
Certificate Number
C-0229
Valid Until
02.03.2016
Last Amended on
Page
4 of 6
Quantity Measured/
Instrument
Range / Frequency
*Calibration Measurement
Capability ()
Remarks
4mm to 300mm
4.2 m
Using Length
Measuring Machine
Upto 200mm
4.9 m
20.
21.
22.
COMPARATOR STAND
FLATNESS OF PLATE$
200mmx200mm
3.5 m
23.
Upto 100mm
100mm to 300mm
1.7 m
3.3 m
24.
SNAP GAUGE$
2mm to 300mm
4.2 m
25.
THREAD PLUG
GAUGE $
(Effective Dia Only)
2mm to 100 mm
3.0 m
26.
THREAD PLUG
GAUGE$
(Effective Dia Only)
100mm to 300mm
4.1 m
27.
2mm to 100 mm
4.5 m
Laboratory
Microtech Calibration Lab, Plot No. 143, Indira Housing Society, Kabnoor,
Ichalkaranji, Kolhapur, Maharashtra
Mechanical Calibration
Issue Date
03.03.2014
Certificate Number
C-0229
Valid Until
02.03.2016
Last Amended on
Page
5 of 6
Quantity Measured/
Instrument
Range / Frequency
*Calibration Measurement
Capability ()
Remarks
2mm to 100 mm
2.0 m
100mm to 300mm
4.2 m
11.8 m
28.
29.
THREAD RING
GAUGE$
30.
BENCH CENTRE
(Co-axillity &
Parallelism)
Upto 500mm
31.
SURFACE PLATE
6.0
L+ W
125
Where L&W is in mm
32.
PROFILE PROJECTOR
L.C.0.001mm
Linear 0 to 100
Angular 0 to 360
Magnification 10x20x
8.8
3.0 min
1.0 %
0.35% FSD
Using Electronic
Pressure Calibrator
As per DKD R-6-1
By Comparison Method
II. PRESSURE
1.
PRESSURE GAUGE#
(Analogue/ Digital)
Laboratory
Microtech Calibration Lab, Plot No. 143, Indira Housing Society, Kabnoor,
Ichalkaranji, Kolhapur, Maharashtra
Mechanical Calibration
Issue Date
03.03.2014
Certificate Number
C-0229
Valid Until
02.03.2016
Last Amended on
Page
6 of 6
Quantity Measured/
Instrument
Range / Frequency
*Calibration Measurement
Capability ()
Remarks
The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing actual
environmental conditions and master equipment used.