Beruflich Dokumente
Kultur Dokumente
resulting chromatograms will be evaluated to determine how well the analyte peaks are
resolved.
Reservoirs for EACH HPLC machine
o 1 L of acetonitrile
o 1 L of 0.5% (V/V) acetic acid in Milli-Q water (add 5 mL of HPLC acetic acid to 1 L
bottle and adjust to the volume with milli-Q water)
Samples to run for mobile phase optimization:
a. Highest concentration mixed standard.
b. Pure solutions of each of the three compounds
HPLC parameters:
o Flow rate: 1.5 mL/min
o Detector wavelength: 274 nm (and 300 nm)
o Run time: 20 minutes
o Column temperature: 25oC
o Injection volume: 20 L
Mobile phase composition to run
100% acetic acid
95/5 acetic acid/acetonitrile
93/7 acetic acid/acetonitril
92/8 acetic acid/acetonitrile
90/10 acetic acid/acetonitrile
88/12 acetic acid/acetonitr
83/17 acetic acid/acetonitrile
85/15 acetic acid/acetonitrile
80/20 acetic acid/acetonitr
Follow the standard operating procedure instructions for the HPLC, using the parameters
listed above. Be sure to enter your assigned mobile phase percentages into the pump
parameters field.
After collecting the HPLC data at your assigned mobile phase composition, determine the
retention factor (k) for all three compounds and also calculate the selectivity factor
() and resolution (R) values between all pairs of compounds.
Record data in an excel file.
Prepare a single graph of k vs % acetonitrile for the five components. Use this graph to
determine the % acetonitrile which gives the greatest difference in capacity factor
between the peaks. This should be your optimum setting for later analysis. Set the pump
to this setting and repeat the two samples to confirm that you are getting a good
separation.
Prepare a graph of log (k) vs 1/T with temperature in Kelvin for each of the five
components. The log (k) should be a linear function of 1/T. You can do least squares trend
lines to determine the optimal temperature for separation.
Once the optimum setting is determined, you will run the calibration standards and
samples described in the lab manual using this combination. Set up a sequence file with
all of your calibration standards and chocolate samples and run it overnight.
SPE procedure
a. Conditioning SPE tube in vacuum (5 mmHg)
Activation: 1 mL methanol
Equilibration: 1 mL of Milli-Q water
Reversed phase type silicas and nonpolar adsorption media are usually conditioned with a
water-miscible organic solvent such as methanol, followed by water or an aqueous buffer.
Methanol wets the surface of the sorbent and penetrates bonded alkyl phases, allowing
water to wet the silica surface efficiently.
b. Sample loading step - make sure you added the internal standard to the
chocolate before this step!
0.5 mL of sample
Accurately transfer the sample to the tube or reservoir, using a volumetric pipette or
micropipette. The sample must be in a form that is compatible with SPE
c. Wash step
1 mL of milli-Q water
If compounds of interest are retained on the packing, wash off unwanted un-retained
materials using the same solution in which the sample was dissolved, or another solution
that will not remove the desired compounds.
d. Elution step place clean sample collection tube under each extraction column
before starting elution!
2.5 mL of ethanol
Rinse the packing with a small volume of a solution that removes compounds of interest
(typically 200 L to 2.5 mL depending on the tube size), but retain any impurities not
removed in the washing step.
Two small aliquots generally elute compounds of interest more efficiently than one larger
aliquot. Recovery of analytes is best when each aliquot remains in contact with the tube
packing for 20 seconds to 1 minute. Slow or drop wise flow rates in this step are beneficial.
Collect elutes and further prepare as appropriate.
e. Reconstitution for each sample
All samples need to be evaporated to dryness using the solvent evaporator system. If your
samples are not completely dry at the end of lab, tightly cap and place them in the
refrigerator and continue drying next week.
The dried chocolate extraction residues should be reconstituted in 2.00 mL of Milli-Q water
and filtered using 0.22 m filters into autosampler vials. When sample has been filtered,
run HPLC right away.