Beruflich Dokumente
Kultur Dokumente
BME (418)
Spring 2011
Table of Contents
Exp#
Title
Page
##
Introduction
Differential Amplifiers
Pre-Lab #1
5
12
Optoelectronic Components
Pre-Lab #2
15
19
21
28
7
8
Pre-Lab #4
The Electrocardiograph
Recording
(ECG)
Pre-Lab #5
I. Analog to Pulse Shaping.
II.Visual and Sound Pulse Indicators.
Pre-Lab #6
Rate Meters
Pre-Lab #7
I. Pulse Rate digital meters.
II. Pulse rate by photopllethy- smography.
Pre-Lab #8
30
35
37
43
45
52
54
57
59
68
Temperature Measurements
Pre-Lab #9
70
75
10
Respiratory Rate
Pre-Lab #10
77
80
11
82
12
87
##
References
95
Introduction
The course contains two pre wired panels with a variety of cables and accessories.
It provides training in basic monitoring circuitry, such as ECG, EEG, EMG, Pulse
rate, GSR, and temperature monitors.
The first part of this lab explains the electronic circuit of the Electrocardiograph
(ECG) Instrument in addition to EMG and EEG.
The electrocardiograph recorder is an instrument, which can record the low-level
voltages produced by the heart. Recorded from a patient's limbs and chest, these
voltages produce a tracing called an electrocardiogram.
Since magnetic power fields create common-mode signals, which interfere with
the desired signal, special amplifier designs are needed.
Figure#1 show the block diagram of the experimental circuit on panel SIP3851.This circuit will be more fully described in the experiments 1, 2, 3, 4,5,6,7, and 8
in addition to exp# 14 and 15.
The second part, explains the electronic circuit of Pulse Rate Digital meter,
Temperature Digital Meter and Galvanic Skin Resistance. Figure #2 shows the
block diagram of the experimental circuit on panel SIP385-2. This circuit will be
more fully described in the experiments 9,10,11,12 and 13
Of key importance is the ratio of [differential signal gain (Adiff)] to the [common
mode gain (Acm)]. This relationship is called common mode rejection ratio
(CMRR) and is expressed by equation
CMRR 20 log
Adiff
Acom
In determining the CMRR, the two signals are adjusted at the input to produce the
same output voltages (Voutcm= Voutdiff). Therefore, CMRR can be determined
from the input voltages as shown below:
CMRR
Vin cm
Vin diff
Good amplifiers have CMRR values, which range from +60dB to +100dB.
The adverse effect of electrode contact resistance on signal input is an additional
consideration. Losses incurred at the lead contacts with the skin decrease the
available input signal to each amplifier. High input impedance of the amplifier
minimizes the signal loss.
Since the LF347 has an input impedance of over 1012 ohms, only a very small
portion of the input signal voltage would be lost. Since the LF347 is a quad as
shown in Figure#3, three of the amplifiers (A2, A3 and A4)can be used for
recording the ECG, EMG, or EEG signal voltage. Two amplifiers (A2 and A3) are
used for the differential input, and one for a single-ended output amplifier. A
typical circuit arrangement is shown in section B in the insertion panel. Amplifier
A1 is used in section A as an inverter.
We use the function generator instead of an ECG simulator (or human) in the first
exp's to perform the experimental procedure. Therefore, section A is used to
develop two signals out of phase from a single-ended function generator and the
difference between them is in mV (to simulate the low-level voltages produced by
the heart (QRS wave).
The following two equations are for the input voltages to the instrumentation
amplifier (section B) when the mode switch in section A is set to the DIFF and
COMM respectively:
DIFF :
Tp 2 Tp1
Tp 3 Tp1
R 24
Tp1
R 24 R 26 1000
Tp 1
1000
R 22
Tp1
COMM :
Tp 2 Tp 3 Tp 1
The gain of the amplifier A4 is determined by the ratio of the R11/R8 (100 K /
8.2K) resistors (A = 12.2). The gain of the two amplifiers (A2 and A3) is
determined from
A diff 2
R6
29 . 4
R5
The gain of the input stage can be controlled by varying the resistor (R5) between
pins 6 and 13. As the resistor is made smaller, the gain is increased. This resistor
should not be zero since the circuit might oscillate under high gain conditions.
In the circuit shown in the panel, resistors of 1% tolerance (or less) should be
used, since both halves of the circuit must match. The value of potentiometer R10
should be adjusted to obtain the best balance of the two signals input to A4. This
balance achieves the highest possible CMRR.
Objectives:
1.Test and evaluate differential amplifiers as use din biomedical instrument
2. Determine the gain and common-mode rejection ratio of differential amplifiers.
Materials Required:
1.
2.
3.
4.
5.
Experimental Procedure:
1. Balancing the Differential Amplifier (A4).
Input: Using the function generator feed a 10 Hz sin wave and 1 Vp-p into Tp1
(Section A).
Output: Is taken from Tp6 using the digital scope.
a. Set the mode switch in section A to the COMM position.
b. Adjust R10 (section B) as necessary for minimum sin wave output at Tp6.
TP6 = ..
V TP 6
V TP 3
e. Using the digital scope, connect Tp4 to channel 1 and Tp5 to channel 2 and draw
the wave shapes of the signals.
Channel 1 ..
Channel 2 ..
Adiff
V TP 6
V TP 3 V TP 2
Tp6 = ..
Adiff = ..
Input: Using the function generator feed a 10 Hz sin wave and 5 Vp-p into Tp1
(Section A).
Output: Is taken from Tp6 using the digital scope.
a. Set the mode switch in section A to the COMM position.
b. Measure the p-p signal voltages at:
Tp2
Tp3
Tp6.
c. Calculate the common mode gain (Acm).
Acomm
V TP 6
V TP 2 V TP 3
10
Input: Using the function generator feed a 2 Vp-p and sin wave into Tp1.
Output: Is taken from Tp6 using the digital Scope.
V (cutoff) = ..
11
Faculty of Engineering
Name:
Number:
12
I. Objectives:
Switch
DIFF
Tp2
Tp3
13
COMM
III. Calculate the differential gain between Tp6 and (Tp3 Tp2)
14
Theory:
In medical instrumentation, patient safety is a major consideration. When taking
an ECG, the right leg may be common to earth ground on some medical recorders.
If a line voltage come in contact with a hand, arm or part of the body, a current
flow through the body takes place. Less than 1mA of current through the heart is
sufficient to cause fibrillation. If, however, the patient is totally isolated from
ground, the potential shock hazard is greatly reduced
Optical coupling
The optocoupler is often used to isolate two circuits, which normally share a
common ground to reduce potential equipment shock hazards to a patient.
The Optocoupler (Section C) contains a light source (LED) and a sensor. The
sensor can be a photodiode, phototransistor, or photo Darlington. In this form,
only light waves representing the signal.
15
The high gain capability of the Darlington reduces the need for another stage of
amplification.
The 5Kohms potentiometer (R15 located in Section B) adjusts the signal level to
the optocoupler stage; the light level in the LED is set by the DC current flow
through transistor Q1. The output signal from amplifier A4 (Section B) amplitude
modulates the light intensity of the LED in the optocoupler. The Darlington
transistor sensor also receives light changes, which are further amplified by the
high gain cascaded transistor circuit.
The important parameters for most optocouplers are their transfer efficiency,
measured in terms of the current transfer ratio (CTR), the output transistor's
maximum collector-emitter voltage rating VCE(max), the input LED's maximum
current rating IF (max), and the optocoupler's bandwidth. CTR is the ratio between
a current change in the output detector and the current change in the input LED
that produces it. Typical values for CTR range from 10% to 50% for optocouplers
with a phototransistor and up to 2000% for those with a Darlington transistor pair
in the output.
The gain of optocoupler, which uses a Darlington sensor, ranges from 5 to over
75.
*****************************************************************
Objectives:
Optocoupler parameters measurement;
1. The current transfer ratio (CTR)
2. The frequency response
Material Required:
1. SIP 385-1 Kit
2. Digital storage Oscilloscope
3. Function Generator
4. DMM
16
Experimental Procedure:
The Current transfer ratio CTR:
1. Consider the opto-coupler circuit section C, Adjust RI7 for a voltage drop of
0.11Vdc across R18.
2. Compute the current in R18 (LED forward current If1)
3. Measure the voltage drop across R19 and Compute the current in R19 (IC1:
Collector Current)
4. Calculate CTR.
CTR
I c1
I f1
I c
I f
Frequency response:
1. Set the source switch to DIFF.
2. Set the sensitivity control to its max.
3. Open switches Al, A2, A3 and A4.
4. Feed a 30Hz sine wave into the input section A to adjust the voltage at TP6 to
2Vp-p.
7. Close A4, which places C9 in parallel with C10, repeat step 6. Flower=..
17
18
Faculty of Engineering
Name:
Number:
19
I. Objectives:
II. Explain the main purpose and the operation principle of the opto-coupler shown
below:
.
.
.
.
III. Define the CTR of the opto-coupler. If the voltage across 220 = 0.11V, and the
voltage across 470 = 1V, calculate the CTR.
.
.
.
20
Theory:
Often, real world signals of interest are mixed with undesirable noise signals
(power line interference in ECG signals etc.). Circuits such as filters are used to
attenuate the amplitudes of the signals, which are not desirable. Depending on the
frequencies that are desirable, filters can be low-pass, high-pass, band-pass or
notch filters. The principle of action of ideal filters is shown in Figure#1:
Circuits made with real-world components cannot achieve the sharp cut-off
characteristics of the ideal filters shown above, but with some degree of
approximation, we can get fairly close. Filters are essential in circuits such as
ECG monitors where a considerable degree of interference is picked up because
of the surrounding electrical equipment as well as the movements of the patient.
Another area in which filters could be used is in hearing aids.
The frequency at which the output voltage (Vout) equals 0.707 x Vin is referred to
as the high- or low- frequency roll-off point. This point is also defined as the
frequency at which the output voltage has dropped by 3 dB (decibels). This point
can be described as follows:
Vout 0.707Vin
21
G 20 log
Vout
dB
Vin
The difference in frequencies at the -3 dB roll-off points at the high and low
ends of a response curve is called the bandwidth of a filter. The term bandwidth
usually applies to band-pass and band- rejection (notch) filters.
A band-pass filter passes all signals that fall within a band defined by a lower
und an upper frequency limit. It attenuates all other frequencies outside of this
specified band.
The bandwidth (BW) is defined as the difference between the upper cut-off
frequency and the lower cut-off frequency as shown in Equation
BW f u f L
The frequency about which the pass band is centered is called the center
frequency. It is defined as the geometric mean of the two cut-off frequencies as
shown in Equation:
fo
f L fu
fo
BW
22
DF
1
Q
This damping factor is important in defining the kind of response in the pass
band and the shape of the roll-off curve.
The instrumentation amplifier shown in Section B of the panel is also a bandpass filter, the high frequency roll off is caused by the following capacitors:
1. Feed-Back capacitors connected form an output to an input: C39, C3 and C6 .
2. Stray Capacitors connected form an output or an input to ground, C5 and C7.
C4 and C5 are controlled by switch A1 and A2 respectively, for example when
switch A1 is closed (ON), C4 is connected to Tp6 through R13.and when it is
open (off), C4 in not connected to Tp6.
The low frequency roll off is caused by the coupling capacitors connected in
series with the input or with the output: C1, C2, C9 and C10, also C9 and C10 are
controlled by switch A4.
Switch A3 is used to control the amplitude of the signal at TP6.1 as follows:
Tp 6 R15
R15 R13 R14
23
Notch Filter
The notch filter is also known as a band-stop, band- reject, or band-elimination
filter. In effect, a notch filter performs in exactly the opposite way from a bandpass filter. The notch filter rejects or attenuates all frequencies inside its
response curve between the 3 dB points and passes all other frequencies. The
bandwidth is defined as the difference in the frequencies at which the response
is 3 dB down. These filters are often used to reduce 50 or 60 Hz signals in
sensitive medical instruments such as EEGs and ECGs.
24
Notice that low-pass, band-pass and high pass outputs are all available from the
same circuit.
Section F of the panel shows a circuit for an active band-pass filter, which is tunable from
8-14 Hz; this filter is used in the EEG instrument (alpha waves).
*********************************************************
Objectives:
At the end of this experiment, students will:
1. Understand the characteristics of active and passive filters.
2. Understand and study the function and use of active notch filter
3. Measure the frequency response of the state variable and notch filters
(Bandwidth and gain).
4. Evaluate the operating characteristics of a state variable active filter.
Material Required:
1. SIP 385-1 Kit
2. Digital storage Oscilloscope
3. Function Generator
4. DMM
Experimental Procedure:
25
Input: Using the function generator feed a 2 Vp-p and sin wave into Tp1.
Output: Is taken from Tp6.1 when both switches A1 and A2 are closed
d. Measure the high frequency roll off when both switches A1&A2 are open.
Input: Using the function generator feed a 50 Hz (Fo) sin wave and 4 Vp-p into
Tp12.
Output: Is taken from Tp13 using the digital Scope.
26
b. Adjust the potentiometers R31 and R32 for minimum sin wave output;
a. Record the output voltage at 20Hz (below the notch frequency) and 150Hz
(above the notch frequency);
V fo
V20 Hz
);
Steps:
1. Calculate the output voltages at -3dB points;
VC = VD= 0.707* Vout at 20Hz=
2. Beginning at 20Hz, slowly increase the generator's frequency until the output
voltage decreases to VC,
Record this frequency as fC = ..
3. Beginning at 150 Hz, slowly decrease the generator's frequency until the output
voltage decreases to VD,
Record this frequency as fD = ..
g. Compute the width of the notch at its -3dB points,
f = fD fC=
h. Compute the voltage attenuation in dB both above and below the notch,
27
Faculty of Engineering
Name:
Number:
28
I. Objectives:
II. Explain how Switch A (A1, A2, and A4) affects the frequency response of the
differential amplifier located in Section B of the panel.
A1:..
A2:..
A4:..
III. If Tp1= 2Vp-p, Calculate the maximum diff voltage and the cut-off voltage
at Tp6.1.
..
..
IV. What is the main function of the notch filter located in section (E) in the panel?
..
..
29
Noise is all the unwanted electronic signals coming in that you dont want to
measure. This one we will leave to you: you have to determine how noisy your
circuit is, and justify the method you use to measure noise. Remember from class
that we characterize the noise characteristic of an amplifier by the Noise Figure.
How can you measure the noise figure of your circuit?
Noise in amplifiers can be caused both by the shot noise within active components
and by the resistors connected to active components. Limiting the frequency
response of the amplifier can reduce the noise levels in amplifiers. In a biomedical
system, the input amplifier section is responsible for most of this noise. External
and internal sources, as well as passive and active components, all contribute to
this noise. Shot noise, intrinsic to solid-state device junctions, is possibly the
greatest consideration. Patient electrodes also generate noise.
At very low levels, the wiring of printed circuit boards and cables introduces noise
through ground loops. Taking a voltage measurement between two distant points
on the ground plane reveals the existence of microvolt variations. These, in turn,
feed input signals to the amplifiers.
Power supplies, when their output impedance is not low enough (under a few
ohms), are another source of noise. The noise may actually be a collection of
pulses, spikes, and 50/60 or 120 Hz voltages. These voltages can usually be
isolated by inserting RC filters between the power supply lines and the IC
circuitry.
30
An amplifier with a gain of 100 might have a noise output of 400 n V when its
total input noise is 4 n V. In order to measure such low-level signals, a high- gain
oscilloscope with low noise is required. Typical oscilloscopes will measure
voltage levels as low as +10 mV /cm. In addition, they themselves generate input
noise in the microvolt range.
The most important noise sources in the panel and how to minimize it are
summarized in the Table below.
# Noise sources
1 60 or 50 Hz power
interference
2 Circuit Noise (IC's)
3 Other
biopotentials,
when
measuring ECG, everything else
creates
noise
like
other
biopotentials (EEG, EMG or
EOG)
4 Motion artifacts (man-made
noise)
5 Electrode Noise
High quality electrode and good
contact
31
Signal
Frequency
Range (Hz)
Amplitude (mV)
ECG
0.3-50
EMG
3-2000
EEG
0.3-30
0.002- 0.05
Table#2
We can control and change the frequency range in the panel (according to
the bio signal we measure) using the control switches: A1, A2 and A4.
************************************************************
Objectives
Upon completion of this study, which includes theory, laboratory testing, and
evaluation, the student will be able to:
1. Measuring the signal to noise ration in a biomedical instrumentation amplifier
2. Measuring the BW and the effect of noise on the bandwidth, the equivalent
input noise
3. Select ICs for low noise applications.
Material Required:
1. SIP 385-1 Kit
2. Digital storage Oscilloscope
3. Function Generator
4. DMM.
5. jumper leads.
32
Experimental Procedures:
Steps:
1. Measuring the total gain of the differential amplifier Adiff.
Input: Using the function generator feed a 1Vp-p ,10Hz and sin wave into Tp1.
Adiff = .
2. Compute the Nin levels for each of the Nout levels determined in step a using the
following equation
Input Noise
Output Noise
Gain
33
Input: Using the function generator feed a 1mVp-p differential signal at Tp2 and
Tp3, 10Hz and sin wave into Tp1.
With Both A1 and A2 closed, Measure:
1. The signal voltage (S):.
2. The Nout voltage (N):.
3. Calculate the S / N.
Input: Using the function generator feed a 1mVp-p differential signal at Tp2 and
Tp3, and sin wave into Tp1.
Output: From Tp15 using the digital scope.
*Close A1, A3, B1 and B3.
*Open A2, A4, B2 and B4.
1. Measure and record the upper cutoff frequency (upper 3db response):
a. Maximum Vout =
b. Cutoff voltage (Vcutoff ) =
c. FH =
34
Faculty of Engineering
Name:
Number:
35
I. Objectives:
II. Explain the main noise sources in the panel and how to minimize it?
...
...
...
...
III. The noise output No at Tp6.1 =30mV, calculate the noise the input noise (Nin)
...
...
...
36
Theory:
The P wave is generated when the SA node develops its initial potential. Signal
transit through the atria to the AV node is indicated by the P-Q interval. The R
wave, often called the QRS complex, is generated as this potential is conducted
through the Bundles of His, the Purkinje system, and the ventricles.
Repolarization, in preparation for the next stimulus, generates the T wave.
Deviations in amplitude, timing, and polarity of the various EKG waves all
indicate conduction abnormalities.
37
There are many different configurations to place the ECG electrodes on a patient.
Three common ones are listed below in Table 1
Lead 1
Lead 2
Lead 3
+ LA RA
+ LL RA
+ LL LA
LA
Left Arm
LL
Left Leg
RA
Right Arm
RL
Right Leg
Table# 2. Short hand notations used in ECG descriptions.
An ECG Amplifier
The circuit diagram of sections B show the ECG amplifier. Bio-potential signals
are very weak signals. Even the strongest ECG signal has a magnitude of less than
10 mV. Furthermore, ECG signals have very low drive, i.e. source has very high
output impedance. Therefore, an ECG amplifier is usually required to have the
following properties:
38
Input leakage current is defined as the current an amplifier sends to the unit
(human body in our case) connected to its input terminals.
A high CMMR is essential since the capacitive coupling from the external
electrical sources such as power lines would create a strong common mode signal
in comparison to the differential ECG signal. A high CMMR would mean that the
AD is much larger than AC, and the differential amplification of ECG signal in the
order of 1 milli-Volt would be possible in the presence of common 50/60 Hz
signal coupled from the power mains, which would be in the order of tens of
Volts. Figure#2 below shows the typical situation in electrical bio-potential
amplification.
*************************************************************
Objectives
1. Learn the sources of bio-potentials in the human body and understand the
techniques used for measurement of these electrical potentials.
2. Record the ECG from the simulator.
3. record the ECG for human subject
4. Study the effects of the artifices.
39
Material Required:
1. Insertion Panel.
2. Patient Lead Cables.
3. ECG Electrode (4).
4. PKg.of alcohol-treated Gauze Pads.
5. Lead Selector Box.
6. ECG Simulator.
7. Scope and DMM.
Experimental Procedures:
1. On the ECG simulator, set the rate to 70bpm. Select lead II on the lead
selector.
2. Display the output (TP15) by the oscilloscope
40
Portion
P
P-Q
QRS
S
ST
T
Two adjacent
R waves
Amplitude (V)
Width (sec)
--------
----------------Table#3
Questions
1. What does a double beat mean?
41
42
Faculty of Engineering
Name:
Number:
43
I. Objectives:
III. If the number of beats per minute = 70BPM, calculate the period between two
adjacent R-waves in seconds.
.
.
IV. What are the most important properties of the ECG amplifier?
44
EXP#6:
I. Analog to Pulse Shaping
II. Visual and Sound Pulse Indicators
Theory:
I. Analog to Pulse Shaping:
The system of counting events such as pulse rate, respiratory rate, etc. requires that
an analog signal be converted into pulsed events and displayed on a meter or
recorder. In other words, analog signals must be timed and counted, events must
be marked and or pulse periods must be indicated. To accomplish this, a portion of
an analog signal is used to trigger a pulse-forming circuit. The amplitude, polarity,
or rise time of a specific signal voltage is chosen as the trigger to activate an event
marker.
Figure # 1 below shows the sequence that leads to a pulse count including the
changes in the wave shape of the signal1.
45
2. Polarity correction Circuit (Section I): Clip off a portion of the signal to use its
rise or fall time(R wave) to trigger the pulse stretcher circuit. In other words, this
circuit removes P and T waves from the ECG complex to produce the trigger
voltage(R wave) as shown in Figure # 1.
Section (I) consists of two circuits: Balanced Splitter (transistor Q2) and Fullwave rectifier (D7- D10).
This polarity correcting circuit also makes use of the R wave as a trigger
regardless of its polarity, it produces a pulse when it receives either a positive or a
negative2 going input signal, and lead reversal is then no longer a problem. If the
R wave would appear inverted at the pulse stretcher input, and if there were no
polarity correcting circuit, the patient leads would have to be reversed.
The rectangular pulse must have constant amplitude and pulse width, it is obtained
from a Schmitt trigger NAND gate connected as monostable oscillator. This
circuit produces fixed amplitude and pulse width pulses.
46
4. For monitoring, average the pulse count with a low pass filter amplifier
(Integrator) and apply the averaged voltage to a DC meter. This will be discussed
in details in Exp#7.
47
Most monitoring equipment incorporates in its design both a visual and a sound
indicators. Which are used to verify that either an event has just taken place or is
currently in progress.
The visual indicator (LED, D18) is controlled by the pulse produced by the
Schmitt trigger (pin 3 of 74LS132). The pulse width is approximately 0.1 second.
Shorter pulse duration cause a chirpy sound, 0.1 second is a reasonable time when
both visual and audio indicators are involved. Each time the Schmitt trigger is
pulsed, the light flashes.
The other two Schmitt Trigger NAND gates of the IC74LS132 shown in Figure#3
are used as a pulse tone generator as shown in Figure#4 below.
The first NAND gate act as an oscillator that produces pulses within the normal
human audio range (20_20K) Hz. The frequency of the audio pulses (Tp19)
generated by the oscillator is controlled by R57 and C33 using the following
equation:
0.65
R57 C 33
The second nand gate inserts the audio pulses at Tp19 with the pulses at Pin 3. The
wave shape at Tp20 is shown in Figure# 1.
48
The IC74LS132, however, is not a power device and an audio amplifier would be
needed as a follow up for driving a speaker (audio indicator).this circuit (audio
amplifier) is already wired in the master builder and power supply base.
Audio pulses can be formed by using a 555 timer connected as a stable as shown
in Figure #5 below,
1.44
( R1 2 R2 ) C
********************************************************
Objectives:
1. Utilize a Schmitt trigger circuit for controlling other circuits.
2. Study an indicator circuits, which utilizes a pulsed LED and a loudspeaker.
Material Required:
1. SIP 385-1 Kit
2. Digital storage Oscilloscope
3. Function Generator
4. DMM.
5. Jumper leads
49
Experimental Procedures:
Input: Using the function generator feed a 0.5Vp-p, 1Hz and square wave into
Tp10 (section D).
1. Observe and record the wave shapes at the following test points:
a. Tp11:
c. Output of the pulse stretcher (pin 3 of 74LS132 IC); what are the width and the
period of the pulses?
*Width = ..
*Period =
d. Tp19
2. Compare the pulse width at pin 3 with the length of the total oscillation period,
Are they the same?
50
3. Calculations:
3. If you vary the function generator repetition rate, does the frequency of the
following pulses vary?
* Transistor Q3?
51
Faculty of Engineering
Exp.6
I. Analog to Pulse shaping
II. Visual and Sound Pulse indicators
Pre-lab
Name:
Number:
52
I. Objectives:
Pulse Amplifier:..
Pulse Stretcher:
III. The IC 74LS132 located in (Section K) is a quad Schmitt-trigger, What is the main
function of each one (a, b, c, and d)?
a:
b:
c:
d:..
53
Theory:
The rate metering circuit converts varying pulses into stable pulses whose
amplitude and width are fixed and its frequency is variable as explained in the
previous experiment. Rate meters operate by averaging or by measuring the time
between pulses. The averaging method (see section L in the insertion panel)
changes the varying frequency of the signal into an average DC level; this DC
voltage can derive an electromagnetic meter.
The output pulses from the pulse stretcher circuit with constant amplitude and
width are averaged per time using the double integrator circuit shown in section L
of the panel, the wave shape of the output signal at Tp22 is triangular with a
negative DC level.
The double integrator requires 6 to 8 pulses (beats or R waves) for the meter to
reach its averaging level, while the single integrator(one capacitor in the feedback) requires 15 to 18 pulses.
The averaging method of rate counting is one of two techniques used. The second
method is pulse-to-pulse counting method: the period between pulses is counted
with a frequency meter. This method will be explained in details in exp#8.
************************************************************
Objectives
1. Study how the rate of a periodic signal can be counted
2. Study the double integrator circuit.
54
Material Required:
1. SIP 385-1 Kit
2. Digital storage Oscilloscope
3. Function Generator
4. DMM.
5. jumper leads
Experimental Procedures:
Input: Using the function generator feed a 1Hz and positive pulses into Tp18.1
(section J).
1. First start with no input pulses. What is the DC voltage output of the integrator
(Tp22)? ..
The meter reads
2. With 1 Hz input pulses, adjust R65 for suitable BPM. Measure the frequency
and period of pulses at Tp22.
Frequency (Hz) =
Period (ms) =
55
4. Observe and record the wave shape at Tp22 and measure the DC voltage output.
5. Draw on the same graph the wave shape of the signals at the input of
integrator/metering circuit and at Tp22.
Questions:
1. In any system of counting events (such as heart rate) analog signals must be
converted into pulses, the sequence leading to a pulse count by the
averaging method is as follows:
d. Average the pulse count with a low pass filter and apply the averaged
voltage to a DC meter, What is the main function of :
56
Faculty of Engineering
Name:
Number:
57
I. Objectives:
58
EXP#8:
I. Pulse Rate Digital Meters
II. Pulse Rate by Photoplethysmography
Theory:
I. Pulse Rate Digital Meters:
The pulse rate (F) is obtained from the pulse period (T) as
60
T
Where:
F= pulse rate in beats per minute (bpm), and
T= time in seconds.
General block diagram of the Pulse Rate digital meter shown in sections M
OP in the panel using the function generator is shown below:
59
Most frequency meters cannot register frequency rates below 10 Hz. To measure
low frequencies, a phase-locked loop IC that can provide frequency multiplication
is used as a frequency synthesizer. The PLL, performing as a synthesizer, tracks
the incoming signal with respect to time. Figure #1 shows the block diagram of a
synthesizer used in the panel, which incorporates a frequency divider.
The VCO oscillator can operate at 60, 600 or 6000 Hz, or at much higher
frequencies. For the panel, the VCO operates at 600Hz.
The oscillator frequency determines the divider requirements. The divider can be
60
formed using one, two or three IC's, depending on the availability of components.
For the panel, we have three dividers providing divide by 10, 6 and 10 respectively
as shown in Figure# 1.
Pulses applied to pin 14 of phase comparator (Tp4) are compared with the output
of the VCO after it has been divided by 600 (Tp11). Any difference in frequency
or phase between inputs 14 and 3 of the phase comparator appears as an error
voltage at pin 13. After filtering by R44, R45 and C12, this error voltage appears
at the input of the VCO (pin 9). It is of a proper magnitude and polarity to drive
the VCO toward that frequency which will produce phase and frequency
coherence at the phase comparator input.
The VCO frequency range is determined by R47, R48 and C 13. For the
component values shown, this particular VCO ranges from 60 Hz to 2,200 Hz per
second.
The filter component values were selected to provide a good capture range as well
as good damping characteristics. The circuit, as shown, will capture and lock onto
any frequency within the range of the VCO in under 5 seconds. If a pulse train
having a rate of 1.25 Hz per second is applied to the input of the phase
comparator, the output frequency of the VCO will adjust to 750 Hz (600*1.25) per
second within 5 seconds after the input signal has been applied.
************************************************************
Objectives
1. Describe how a phase locked loop (PLL) circuit can be used for frequency
multiplication.
2. Evaluate the operation of a PLL based pulse rate counter.
3. Evaluate the operation of a digital counter and LED seven-segment display.
61
Material Required:
1. SIP 385-2 Kit
2. Digital storage Oscilloscope
3. Function Generator
4. DMM.
Experimental Procedures:
1. Consider the pulse rate metering circuit consists of section M,O and P in
SIP 385-2 kit
2. Using function generator, feed 1Hz, 1vp-p and sin wave into Tp1
3. Set the two Rate/Temp switches to rate position. In addition, increase the
sensitivity until the digital meter shows a steady pulse rate.
4. Display the output at TP4 by using oscilloscope. Plot it
6. Record the pulse period and then compute the pulse rate from the period
12. If the frequency at TP11= 1.25Hz, calculate the frequencies at Tp8, Tp9,
Tp10 and Tp21 in Hz.
13. With no input, the frequency meter reads 4 or 5 BPM, Explain?
62
Theory:
The photoplethysmograph measures transmittance of light through a capillary bed
in order to determine pulse rate. A light source (e.g. LED) transmits light through
a capillary bed (e.g fingertip or ear lobe) and photo-detectors (e.g. phototransistor)
are placed appropriately to measure the reflected and/or transmitted light. It can
also be used to measure blood flow rate, though this measurement is extremely
sensitive to motion artifact.
With each beat of the heart, arterial blood pressure rises (systole period) and the
extremities increase: (slightly) in physical size. In addition, increased oxygenation
decreases the optical density of surface tissue.
During the heart's period of relaxation (diastole period), the pressure falls; density
in- creases, and the extremities decrease in physical size. Since these cyclical
changes follow the cardiac cycle, they can be used to determine the peripheral
pulse rate. Measuring pulse rate by counting periodic fluctuations in some physical
parameter is called plethysmography.
63
1. When cold air or water contacts the finger's tissue, the vessels contract, blood
flow is decreased, and sensitivity is diminished. Therefore, fingers should always
be kept warm.
2. If the probe holding the optical devices presses too firmly, the vessels are
constricted, blood circulation is reduced, and sensitivity falls off.
3. The physical placement of the LED and sensor is very important. Light should
not spill over and bypass the finger
The next figure shows a sketch of the pulse wave shape as recorded on the strip
chart recorder. The systolic peak is used to trigger a comparator or Schmitt trigger
circuit for rate counting.
When pressure falls as a result of the closing of the heart's aortic valve, a dip
(called a dicrotic notch) takes place.
64
Objectives:
Material Required:
1. SIP 385-2 Kit
2. Digital storage Oscilloscope
3. Function Generator
4. DMM.
5. Photoplethysmograp sensor.
Experimental Procedure:
1. Consider the circuit in section M (SIP385-2)
2. Attach the light sensor to the right forefinger, taking care not to make too tight
a fit. The light source and cell should be facing the palm side of the finger. The
transducer should fit firmly, but should not constrict blood flow. Connect the
cable to the panel by inserting the plug into the pulse rate jack. Relax your arm
on your lap, on the arm rest of a chair. The forearm should not be above the
level of the heart.
3. Set the two toggle switches (Sections 0 and P) to the rate position. Connect an
65
oscilloscope to the output jack (Section M) and record the wave shape.
Increase the sensitivity until the rate meter shows a steady pulse rate. Record
your pulse rate.
4. Confirm that the pulse rate stops when the blood flow stops. Occlude (stop by
closing) the blood flow by tightly squeezing the lower half of the finger being
tested between the sensor and the knuckle. After a few moments, release the
occlusion. What happens to the pulse rate during and just after occlusion?
5. Stand in place with the pulse sensor attached to the finger. Your arm should be
extended down by your side for at least one minute. Record your pulse rate: _
bpm. Adjust the sensitivity control as necessary. You will compare this normal
beat with your pulse rate after exercising.
6. Turn the sensitivity control to zero after noting the initial setting. Run in place
for one or two minutes so that your heart establishes a rapid beat. Stop running
and increase the sensitivity control to the original setting. Record your new
pulse rate. Is the pulse rate faster?
7. After your pulse has returned to its normal rate, breathe heavily for a while and
observe how heavy breathing affects your pulse rate.
.
8. Hold your breath for at least 20-30 seconds and note the effects on your pulse
rate.
a. Stand in position with your hand by your side. Record your pulse rate.
Slowly raise your transducer-attached arm, outstretched, until it is
b. level with your heart. Record your pulse rate and note any required changes
necessary in the sensitivity control setting.
c. Raise the arm fully above your head and hold it in position while recording your
pulse rate. If the pulse disappears, slowly lower the arm and determine in what
position the pulse reappears. Describe the effects.
9. Attach the probe to a finger of the right hand. Record the pulse rate. When
recorded, dip the left hand in cold water. Dry the hand and transfer the probe to
the left hand; then record the pulse on a cold finger. Dip the right hand in very
warm water, dry it, and then record pulse rate.
66
10. What do you think happens to a person's pulse rate and blood circulation?
when the body has been exposed to extremely cold temperature?
11. What effect do you think smoking has on the pulse rate? (Ask a doctor to
describe the effects of smoking on blood circulation and breathing.)
12. Set the Sensitivity Control to Maximum (fully CW). Feed a function generator
(25Hz, sin wave and minimum output of the generator you have) into TPl.
Connect the oscilloscope to TP3. Record the voltage level on
TP3..
13. Find the voltage level on TP 1. Determine the gain of the circuit
.
67
Faculty of Engineering
Exp.8
I.
Name:
Number:
68
I. Objectives:
II. For the pulse rate digital meter block diagram shown in the manual, what is the main
purpose of the comparator and P.L.L?
.
.
.
III. For the P.L.L block diagram shown in the manual, if the frequency of the signal at
Tp4= 1 Hz, calculate the frequencies at:
Tp8=.
Tp9=
Tp10=
Tp11=
IV. Explain how the photoplethysmograph measures number of pulses per time?
..
..
V. What are the main components of the photoplethysmograph monitor?
1.
2.
3.
69
Theory:
In this Laboratory, the surface body temperature is going to be measured at
different locations on the body.
Body temperature can either be measured mechanically, with a mercury
thermometer, or electrically, with a thermistor, diode probe, or a thermocouple.
The mercury thermometer is slow, although accurate. Among these electronic
devices, the thermistor probe has proven to be the most popular.
RTH Ro
1 1
B
T To
In all cases, accuracy, resolution, linearity and temperature range are important
design considerations. In biomedical measurements, a resolution of 0.1C would
be desirable, although a 0.5C resolution is quite adequate.
70
I Is
VD
nVT
Where
I: diode current.
IS: Reverse bias saturation current.
VD: Voltage across the diode.
VT: Thermal voltage=
KT
q
Silicon diodes, such as the Fairchild FDH600, can be used as temperature sensors
by virtue of the fact that the forward voltage across the diode is nearly a linear
function of temperature as shown in the above equation. Figure # 1 shows a
plotting of the diodes voltage versus temperature in degrees centigrade.
Amplifiers E1 and E2 provide the necessary gain for converting a silicon diode
71
The diode sensor is more sensitive than a thermocouple but is somewhat less than
a thermistor. Thermistor and diode probes are utilized in biomedical
instrumentation.
************************************************************
Objectives
1. Describe how temperature varies over the surface of the body.
2. Study the characteristics of diode sensor that used in temperature
measurements.
3. Evaluate the performance of typical circuit used in temperature measurement.
Material Required:
1. SIP 385-2 Kit
2. Digital storage Oscilloscope
3. Function Generator
4. DMM.
5. Diode Probe and sheath.
Experimental Procedure:
I. Circuit Calibration:
a) Place the C/F toggle switch in the C position.
b) Place the rate/Temperature switches in the temperature position.
72
c) Insert the plug of the temp probe into the jack in section N.
d) Adjust R30 until the digital display reads the same as Room Temperature.
e) Place the C/F switch in the F position and adjust R89 to obtain the
equivalent Fahrenheit temperature. F = 9/5 C + 32.
f) R88 is calibrated to read 37C when the probe is not inserted, remove the
probe and adjust R88 for a 37C reading.
1. Insert the sensor into experimental panel and set the Rate/Temp switches to
temp.
2. Cover the probe with a plastic sheaths, place the covered probe under your
tongue. Record your oral temperature.
..
Temp( Co)
Site
Forehead-frontal
Neck, side
Right forearm
Right forefinger
Table#1
73
Temp in C
VTP5
V TP6
Freq at TP7
Room
Temp..
Oral
Temp..
Between thumb
and
fore
finger.
Table#2
74
Faculty of Engineering
Name:
Number:
75
I. Objectives:
II. What are the main differences between diode probe and thermistor probe?
1. ..
2. ..
III. What is the main function of LM566 used in the block diagram of the temperature
digital meter?
..
..
IV. Calculate the frequency of the signal at Tp7 if the voltage at pin 5 of LM566 = 3volts
and the C/F toggle switch in the C position (return to LM566 datasheet).
..
..
76
Theory:
The respiratory system is responsible for bringing oxygen into the body and
carbon dioxide from the body. Different parameters can be measured to show the
performance of the human respiratory system. In evaluating the lung function,
three main factors are considered:
A small diode temperature sensor is inserted into one nostril. The sensor tracks the
temperature difference between inhalation and exhalation. This change in
temperature is amplified and made to control a comparator or Schmitt trigger
circuit. The resultant pulse rate can then be counted by either the meter averaging
method or pulse-by-pulse methods. .
77
A temperature sensitive diode is held under one nostril. The short time constant3 of
the sensor enables the element to accurately track temperature fluctuations. The
mV signal is amplified and used to fire a comparator (D4). The comparators pulse
rate is converted into a frequency by the P.L.L. The frequency change feeds the
frequency counting circuit, which has been previously described.
Most of the circuits used here have been previously encountered. In the laboratory,
an overall system is evaluated. The temperature sensor on one panel (SIP 385-2) is
connected to the meter rate section on the second panel (SIP 385-1).
********************************************************
Objectives:
1. Describe the basic functioning of the respiratory system.
2. Measure the respiratory rate.
3. Evaluate a respiratory recording instrument.
Material Required
1.
2.
3.
4.
5.
6.
Experimental Procedure
Respiratory rate measurements will require the use of both panels. The meter rate
circuit is requiring since the breathing pattern is too slow (12-20 BPM) for period
recording by the digital display.
Time constant of a medical sensor is the time required for the sensor to reach 63.2 % of its final response.
78
Before taking any measurements, check that the two insertion panels are work
well, in other words, the digital meter must read approximately your respiratory
rate and the visual alarm LED(Section j)must be ON for each temperature
change.
6. Exercise by running for one minute, and then record your respiratory rate.
7. With each breath, how many mV changes appear at TP5?
79
Faculty of Engineering
Name:
Number:
80
I. Objectives:
II. Explain how the temperature sensor (diode probe) measures the number of breaths
per minute?
.
.
III. Explain the operating principle of the respiratory rate block diagram shown in the
manual.
.
.
.
.
81
The skin exhibits a resistance to the flow of an applied external current. This
resistance, which is normally in the range of 100,000 to 1,000,000 ohms,
decreases in response during periods of emotional stress. Resistance changes are
particularly notice- able on the palms of the hands and soles of the feet.
In addition, the surface of the skin exhibits an electrical potential. This potential,
which may range as high as 50 m V, is also influenced by emotional states.
Both the Galvanic Skin Resistance (GSR) and the Galvanic Skin Potential (GSP)
can be used to indicate the level of emotional response to a stimulus. Fluctuations
in these skin measurements are referred to as psycho-galvanic reflexes.
The GSR is obtained by passing a small constant current (less than 20 A) through
skin tissue. Electrode sites are chosen on areas of high concentrations of sweat
glands. When sweat glands are stimulated as a result of some stress they secrete a
conducting fluid. This increased fluid content lowers the skin's resistance.
Since the GSR changes by less than 1/2 of 1percent of its initial value, a high gain
amplifier is necessary where G = 1000 or more. In addition, a low-pass input filter
(0.3-25 Hz) is needed to remove extraneous bio-potentials and other artifacts. As
with the ECG measurements, the test results are adversely affected by a high
electrode contact resistance.
82
The GSR monitor uses a DC current source passed through a changing skin tissue
to generate a signal voltage. Changes in skin resistance are reflected as changes in
signal voltage. The signal is amplified and then fed to a light display and tone
generator. The greater the evoked response, the higher pitched is the tone and the
greater the number of lights which turn on.
83
As each LED lights on, the respective voltage to the connecting resistor goes high.
The LEDs turn on sequentially, placing additional resistors in parallel. The voltage
applied to pin5 steadily decreases. When the tenth LED turns on, all resistors R71
to R80 are in parallel and the tone is at its highest pitch.
2 V VC
Frequency at pin4 (566) =
RCV
************************************************************
Objectives:
1. Measuring the skin resistance
2. Describe the influence of emotional or physical stimuli
3. Learn the design of skin resistance monitot
Material Required:
1. SIP 385-2 Kit
2. Digital storage Oscilloscope
3. Function Generator
4. DMM.
5. Finger electrodes with cables.
Experimental Procedure:
1. Hold one electrode in each hand, gripping the electrode between the thumb and
forefinger. Read the dermal resistance on the ohmic scale using DMM.
2. Using your own saliva, wet the surface of the fingers which are in contact with
the electrodes. Again measure your skin resistance. Does the resistance
increase or decrease?
84
3. Attach the two ECG electrodes to the palm side of two fingers. Insert the GSR
plug into the GSR input (section Q). Increase the control sensitivity to 2 or 4
divisions. Press the CAL (calibration) pushbutton and observe if the full light
scale is reached. With each press of the test button, the top light should be
reached. If it does not, increase the sensitivity setting.
4. The subject should be seated and relaxed, with eyes closed. The arm and hand
and not the electrode should rest on the arm of the chair or on the lap. There
should be no strain on the hand. Allow the subject to remain relaxed for at least
one to two minutes. There should be a minimum of room noise. The quiet
condition is part of the subject's conditioning. If responses are observed, they
are being caused by the subject's cognitive process.
5. Stand behind the subject and lightly touch the subject's hair. Say nothing. Note
the light bar response and rotate the sensitivity control as necessary to obtain a
near full-scale deflection. Continue touching the subject in different places so
that a response is evoked. Say nothing. Try clapping your hands near the
subject's ear and observe the response that follows. Try touching the nose with
a feather.
6. Approximately, how long is the time delay between the stimulus and the
response?
7. When good responses are obtained, stimulate the cognitive process by asking
stressful questions. The subject's eyes should remain closed.
8. Good responses, consistently obtained, indicate that the subject is conditioned
and the eyes can now be opened. Ask the subject to look at the light bar and
continue your questioning, touching, etc. in order to evoke a response.
9. Have the subject count backwards, in odd numbers, while the eyes are closed.
Frustration can cause an emotional response. Is a response obtained?
10. In addition to the light bar as an indicator of the GSR response, the glide-tone
is available. Adjust the tone volume (R67, Section R) to a comfortable level.
With each rapid response, a sliding tone pulse is produced. If a tone is not
produced, increase the sensitivity setting.
11. The subject should try to prevent the tone change from occurring. Again, the
stress created in preventing the tone causes the response to happen. Before
removing the electrodes, return the sensitivity setting to zero. Repeat the
testing procedures on several subjects. Always readjust the sensitivity control
85
12. Place a DMM between TP 13 and TP 12. Measure the voltage present (V1).
Press the "Cal" switch (V2) and measure the difference in voltage (V1-V2).
V1>V2, Why?
13. Determine the total constant current being supplied by measuring the voltage
drop across R49 (39 K),Ic.
14. From Ic, calculate the voltage drop across R50 (470 )
15. Measure the DC voltage required at TP15 to obtain a full light bar swing.
*High
frequency
(all
=..
LED's
are
18. Can you replace the 566 with a 555timer (a stable)? And why?
86
ON)
EXP#12:
I. Electromyograms (EMG)
II. Electroencephalograms (EEG)
I. Electromyograms (EMG)
Theory:
It is possible to record the action of skeletal muscle in the body using either strain
gage sensors monitoring the displacements and forces produced by the muscle or
biopotential electrodes sensing electrical activation. Direct force measurements
require intimate contact between the muscle and strain gage. For body surface
recordings, this presents a problem. However, the electrical activity of skeletal
muscles can be recorded by applying electrodes to the skin above the muscle in
question. The pattern produced by the combined action potentials of many motor
units is called an electromyogram as shown in Figure#1.
87
Objectives:
1. Describe the action potentials of muscles as produced by the stimulation of
nerves
2. record the EMG potential on the biceps and triceps muscles
3. record the EMG potential on the forehead
Material Required:
1.
2.
3.
4.
5.
6.
7.
8.
Experimental Procedure:
1. Consider the EMG recording system which consists of section A, B, E and G
in SIP385-1 kit.
2. Close switches A3, B1 and B3.
3. Initially attach the RA and LL electrode leads to electrodes located on the inner
forearm. One electrode should be near the wrist; the other about 8 inches
distant on the same forearm. The electrode marked RL should be centered in
the area between the other two electrodes, not touching either one. This is the
common electrode, and it remains in this location during all tests.
4. Connect the Lead Selector Box plug to the jack in Section B. Set the box to
the Lead II position. The mode switch is left in the OFF position.
5. connect the oscilloscope to output of the amplifier (section G)
6. The arm with the attached electrodes should be relaxed on a table. Increase the
sensitivity gain control of the EMG amplifier (Section B) as necessary to
obtain observable output wave shapes. The scope's base line should be flat
when muscle stress does not exist.
7. Place an easily gripped object in your hand (ball of paper, rubber ball, piece of
88
wood, etc.). Start squeezing the object and observe the base line. Adjust the
sensitivity control as necessary to obtain an effective scope display. Input
signal range may vary according to the subject's strength and therefore, if the
grip is strong, you may wish to lower the sensitivity level.
8. Measure the spike amplitudes observed on the oscilloscope. The tone should
fire at random as the object is gripped and should stop when the muscle is
relaxed.
9. Check your biceps muscles. Move the electrodes RA and LL to the biceps
muscles of your right arm. Clean the areas as before. Place one electrode just
above the elbow flex and the other about 4 inches further up. While seated in a
chair, hold your right foot firmly against the floor; at the same time use your
right hand to lift your right leg. Force your leg down so that your arm must
work hard in trying to make the lift.
10. Feel your biceps muscle with your other hand. Also feel your triceps muscle.
Which muscles are contracting and which are relaxed? Observe the display on
the oscilloscope during muscle stress and note the change.
11. Check your triceps for contraction. Use the back of your hand to push your leg
down to the floor. When the pressure is increased, feel your triceps and biceps
muscles. Which are contracting and which are relaxed?
12. Move your electrodes over the triceps muscles (as you did for the biceps
89
Theory:
Low-amplitude (microvolt range) electrical potentials believed to be generated by
large numbers of nerve cells known as pyramidal cells, located in the outer layer
(cortex) of the brain, polarize and depolarize in response to various stimuli,
creating the EEG waveform. These fluctuating electrical potentials are detected by
electrodes placed on the scalp and are displayed and/or recorded on the EEG.
Measured signal (which typically ranges from 10 to 300 mV), through a series of
stages. The gain, or sensitivity, of each channel is adjustable. User-programmable
instruments require that montages be programmed at the users facility and stored
in the electroencephalographs memory until changes are needed. The EEG signal
is a composite of a range of frequencies (the 1- to 30-hertz [Hz] range usually
proves the most useful) and includes electrical noise, which is inherent in lowlevel measurements.
Typical electrode attachments sites for EEG recording are temporal, Frontal,
90
Type
Alpha
Frequency
Hz
8-14
Theta
in Amplitude in V
Conditions
30-50
Awake, resting
4-8
20-40
Stress, Frustration
Beta
14-50
Delta
1-4
Varies
Stress, Tension
Sleep,
disease
Brain
************************************************************
Objectives:
1. Describe how EEG is recorded.
2. Record EEG pattern on a subject
3. Evaluate the performance of EEG amplifiers.
Material Required:
91
1.
2.
3.
4.
5.
6.
Experimental Procedure:
1. Consider the EEG recording system which consists of section A, B, C, F and
G.
2. The first site to be observed is located on the temporal region of the head.
3. Connect the EEG electrodes cable into the lead selector box, select lead I on
the lead selector.
4. Insert the plug of the lead selector into the jack of SIP385-1.
5. Place the subject, with eyes closed, on a blanket-covered table (in a dimlylighted room, if possible). Attach the electrode to the recording sites using
electrode paste to hold the electrodes in place. Tape can be used to help hold
the electrode in place until the paste dries.
6. On the panel, select a 0.3 Hz to 50 Hz frequency response by closing switches
A1, A3, and A4 (switch A2 remains open). Open switches Bland B4 and close
switches B2 and B3. This connection inserts the notch filter. Position the
"DIFF/COMM" switch to the OFF position.
7. Connect the oscilloscope to the output of the amplifier (section G). Then
slowly rotate the Sensitivity control on the panel and observe the tracing on the
oscilloscope. Adjust the vertical sensitivity of the oscilloscope for a 1-2 cm
display.
8. Have the test subject quickly open and close his/her eyes. From the
oscilloscope display, determine if eye-blinking spikes are observed. If eyeblinking spikes are not observed, increase the sensitivity further.
9. With the eyelids closed, have the test subject shift his or her eyes rapidly from
center to right and center to left. Can you observe pattern changes associated
with the eye movement? Make a number of observations until you know what
you are observing. Observe the difference between right and left, and up and
down.
92
10. Does movement of the eyes to the right produce a positive or negative going
pulse? What does movement to the left produce? What does movement of the eyes
up and down produce?
11. The next site to be observed is located above the eyebrows, near the edge of
the hair line (Frontal region). Leads I and II will be placed high above each eye,
and the common lead is placed below the earlobe, on the neck.
12. Record an EEG tracing. Have the test subject open and close his or her eyes
and determine if eye-blinking spikes are observed. If not, readjust the Sensitivity
control on the panel until spikes are observed.
13. Position one EEG electrode on the right Frontal region and one on the right
Parietal region. Use the oscilloscope to make a recording of the waves present.
14. The alpha rhythm will be observed next. The frequency of the alpha waves is
in the range of 8-14 Hz. To select this bandpass, switch in the bandpass amplifier
by opening switches B2 and B3, and closing B1 and B4. The bandpass filter also
provides additional gain.
15. Allow the subject to relax in complete silence until he or she starts to doze. On
the oscilloscope or Compucorder observe the pattern, amplitude and types of
waves present. Count the number of waves produced per period and determine
the frequency of the waves present.
16. While the subject is lying on the table, with eyes closed and presumably
relaxed, present him or her with a problem difficult enough to cause frustration
to the test subject; i.e., the multiplication of two number by a third. Are the
observed wave patterns affected? Theta waves (4-8 Hz) increase in amplitude
with frustration. Do you observe any pulses in this frequency range?
17. Bypass the alpha filter (8-14 Hz) by closing and opening the appropriate
switches, since it also filters out the theta waves. Subject the person to various
forms of harsh stimulation. Use loud noise, physical touch, odors, bright lights,
and other activities which might cause a fright, tension, and stress in the test
subject. Observe the effects on the wave patterns that 13
93
4. Close B2 and B4, and open B1 and B3, What is the gain of alpha filter?
94
References:
1. SIP385BM
Biomedical Instrumentation
Author: Morris Tischler.
2. Electronic Design
Circuits and systems
Second edition
Author: C.J. Savant,Jr, Martin S. Roden, and Gordon L. Carpenter.
3. URL: http://www.bme.utexas.eduiugrad/UGLab/
4. URL: http://www.omgsic.com
95