Beruflich Dokumente
Kultur Dokumente
Revision: 9
Date of Revision: Oct 10, 2012
ii
TABLE OF CONTENTS
DISCLAIMER ................................................................................ iv
ACKNOWLEDGEMENTS ..................................................................... v
1.
INTRODUCTION........................................................................ 1
1.1
Purpose of the Standard Operating Procedure ............................. 1
1.2
Emergency Contact Information .............................................. 1
1.3
User Fees and Instrument Booking ........................................... 2
1.4
Basic Theoretical Background ................................................. 2
1.4.1
Sample Preparation Suggestions......................................... 3
1.5
Instrumentation ................................................................. 5
1.5.1
Overview .................................................................... 5
1.5.2
Lasers and Filters .......................................................... 5
1.5.3
Objectives .................................................................. 5
1.5.4
Scan Head................................................................... 6
2.
3.
4.
5.
6.
PROTOCOL ............................................................................. 9
6.1
Start-up........................................................................... 9
6.1.1
Viewing sample with transmitted light and Kolher illumination .. 10
6.1.2
Epifluorescence Imaging ................................................ 11
6.2
Preparation for Imaging ........................................................ 1
6.3
XY Image Acquisition ........................................................... 3
6.4
Z-Series or 3-D Stack Image Acquisition ..................................... 4
6.5
T-Series Image Acquisition..................................................... 4
6.6
Image Analysis ................................................................... 5
6.7
Saving and Viewing Final Images ............................................. 9
6.8
Cleaning Objectives .......................................................... 10
6.9
Shutdown ....................................................................... 10
7.
TROUBLESHOOTING ................................................................ 10
8.
9.
iii
REFERENCES .............................................................................. 13
APPENDIX 1: FLUOROCHROME PEAK EXCITATION AND EMISSION WAVELENGTHS 14
APPENDIX 2: USER LOG .................................................................. 20
APPENDIX 3: PREVENTATIVE MAINTENANCE LOG .................................... 22
iv
DISCLAIMER
The materials contained in this document have been compiled from sources
believed to be reliable and to represent the best opinions on the subject. This
document is intended to serve only as a starting point for good practices and
does not purport to specify minimal legal standards. No warranty, guarantee,
or representation is made by Laurier as to the accuracy or sufficiency of
information contained herein, and Laurier assumes no responsibility in
connection therewith.
ACKNOWLEDGEMENTS
The following individuals of Laurier contributed to the writing, editing, and
production of this manual: Gena Braun (Instrumentation Technician); Dr. Diano
Marrone (Psychology); Stephanie Kibbee (Environmental/Occupational Health
and Safety Office).
This manual was prepared for Laurier. Any corrections, additions or comments
should be brought to the attention of the Instrumentation Technician at
519-884-0710 ext. 2361.
1. INTRODUCTION
1.1 Purpose of the Standard Operating Procedure
This standard operating procedure (SOP) is NOT a substitute for training
and/or reading the appropriate manuals before use. All principle
investigators and supervisors must document that training has been
received by students and staff who will be using the Olympus FV1000
confocal laser scanning microscope.
A list of authorized users will be maintained by the Instrumentation Technician.
This SOP is intended to promote consistent and safe use of the Olympus
FV1000. This SOP covers the potential hazards, personal protection
requirements, spill and accident procedures, waste disposal considerations,
and instrument operation for the Olympus FV1000.
Immediately notify the Laser Safety Officer at extension 3108 for all laserrelated injuries and near-misses. If the contacts listed above cannot be
reached, or in the case of a more general emergency, call 9-911 from any
campus phone and/or Special Constable Service at ext. 3333.
http://www.wlu.ca/documents/42977/Laser_Safety_Manual_2010.pdf
Hourly: $ 20
Annually: $ 1500
(Internal users do not pay a training fee)
External users:
Figure 1-1: Basic illustration of confocal imaging; only the signal that is
confocal with the emission pinhole is detected, reducing out-of focus
background signal and improving the image significantly.
(i.e. pH) and has a relatively broad emission spectrum, so it is not ideal for
studies that require dual or triple labeling. Appendix 1 lists several dyes and
their excitation/emission wavelengths.
The appropriate mounting medium depends on the nature of the specimen
(fixed or live), and the magnification to be used. For fixed slides, the medium
should contain anti-fade reagents to minimize photobleaching. Conversely,
anti-fade reagents can be toxic to living organisms and are typically not
recommended for live specimens. The medium and anti-fade reagent must also
be compatible with the selected dye; some anti-fade compounds can cleave
the fluorescent molecules of certain dyes and cause the signal to degrade with
in a few days or less.
If you plan to image using the 60x or the 100x oil immersion lenses, your
mounting medium must have a refractive index that is similar to oil. Consider
using 50-80% glycerol or 2,2-thiodiethanol in your mounting medium to improve
the image (see Staudt et al., 2007). Again, confirm that your dyes are
compatible with all the components of your mounting medium before preparing
the specimen. All mounting media must be completely dry before the
specimen is viewed on the FV1000.
All sample preparation should be done in a separate lab, NOT in the room
with the FV1000.
Additional information on mounting media and anti-fade reagents has been
summarized by Tonny Collins (Wright Cell Imaging Facility, Toronto) at:
http://www.uhnresearch.ca/facilities/wcif/PDF/Mountants.pdf
1.5 Instrumentation
1.5.1 Overview
The FV1000 inverted CLSM can be used to examine both fixed and live
specimens. Light is produced by a several lasers, and passes through the
appropriate filters and mirrors and a pinhole before being scanned across the
image in a raster patter. High speed imaging (up to 16 frames/sec for a 256 x
256 image) and rapid spectral scanning (100 nm/sec) are carried out using
galvo scanning mirrors. Wavelengths can be resolved to 2 nm on this system,
and all filters are ion-sputter coated to provide improved transmission
efficiency and allow imaging at lower laser intensity. Fluorescence produced by
the specimen passes back through mirrors and filters and a pinhole to be
detected by the appropriate photomultiplier tube (PMT). The available lasers
and optics are described in more detail below.
1.5.3 Objectives
The terms that follow the objective magnification indicate the corrections applied to
the lens, as follows:
- Plan indicates a flat field correction, which corrects for field curvature
produced by the lens.
- S-Apo indicates apochromatic correction which is the highest degree of
correction for spherical and chromatic aberration.
- Fluor indicates fluorite aberration correction, which is multi-colour
chromatic and spherical aberration.
- NA is the numerical aperture of the lens, which determines resolution and
depth of field.
- WD is the working distance.
There are 3 air objectives, and two oil-immersion objectives available on the FV1000
confocal microscope:
- 10x (Plan S-Apo, NA 0.4, WD 3.1 mm)
- 20x (Plan S-Apo, NA 0.75, WD 0.65 mm) (DIC)
- 40x (Plan Fluor, NA 0.6, WD 2.7-4.0 mm, correction collar) (DIC)
- 60x oil (Plan Apo, NA 1.42, WD 0.15 mm) (DIC)
- 100x oil (Plan S-Apo, NA 1.4, WD 0.12 mm) (DIC)
The correction collar on the 40x objective can be adjusted to account for variation in
coverslip thickness. All of the other objectives must be used with a 0.17 mm (#1)
coverslip.
Extra care must be taken when using oil objectives:
- An oil immersion objective must be cleaned before adding any more oil to
view a new slide. If it is not cleaned, excess oil can run down the side of the
objective and into the microscope.
- DO NOT use any of the air objectives (10x, 20x, or 40x) immediately after
using an oil immersion (60x or 100x) objective. The 20x and 40x objectives in
particular have a very small working distance, and can easily come into contact
with any oil left on the slide. The slide must be cleaned before switching to the
air objectives.
2. POTENTIAL HAZARDS
The FV1000 is a class 3B laser system and uses a variety of high power
lasers which present laser radiation and heat hazards. The laser radiation
from this instrument can cause serious eye damage if direct or reflected
laser light enters you eyes. Follow all warning labels on the equipment.
Never look directly at or touch a slide while scanning.
Never switch objectives while scanning.
The FV1000 operates under high voltage. Never tamper with any of the
connections or electrical cables, and contact the Instrumentation Technician
immediately if any of the cables appear damaged.
Do not expose your hand or finger to the laser beam output from the objective
mount hole, objective tip or condenser lens, or your skin may be damaged.
Never attempt to output the laser beam outside the system by inserting a
mirror or a similar object in the light path. The laser beam may enter your
eyes, and this is extremely hazardous.
Make sure that the slide is laying flat on the stage. If the specimen inclines, the
laser beam may reflect into your eyes, which is extremely hazardous.
Do not bend or pull any of the laser fiber cables. If the laser fiber cable is
damaged, the laser light may leak outside it and create a hazardous situation.
Should such an event occur, immediately turn off the laser power and contact
the Instrumentation Technician.
The air outlet of the laser cooling fan blows out warm air and the mercury lamp
housing gets hot. Do not place a flammable or non-heat-resistant object near
these items. The cleaning fluids (mixture of alcohol or ether) used to clean the
optics are highly flammable. Take special care in handling.
The FV1000 rests on a floating table. The table dampens any vibrations to the
system, and is connected to a cylinder of compressed air. Do not adjust the
regulator on the compressed air cylinder or attempt to change an empty
cylinder; the cylinder is maintained and replaced by the Instrumentation
Technician.
See the WLU Laboratory Health and Safety Manual for additional
information on personal protective equipment:
http://www.wlu.ca/documents/23120/Laboratory_Health_%26_
Safety_Manual__Feb_2007_Final.pdf.
4. ACCIDENT PROCEDURES
All incidents must be reported to the Instrumentation Technician and if
applicable, a students supervisor. All accidents, incidents and near misses
must be reported to the Environmental/Occupational Health and Safety (EOHS)
Office via the WLU Employee Accident/Incident/Occupational Disease Report
form (www.wlu.ca/eohs/forms). To meet regulatory requirements, these forms
must be submitted to EOHS within 24 hours of occurrence, with the exception
of critical injuries, which must be reported immediately to the EOHS Office by
telephone. Critical injuries include any of the following; place life in jeopardy,
produce unconsciousness, result in substantial loss of blood, involve fracture of
a leg or arm but not a finger or toe, involve amputation of a leg, arm, hand or
foot, but not a finger or toe, consist of burns to a major portion of the body, or
cause the loss of sight in an eye. Additional details regarding incident reporting
can be found in the WLU Accident Incident Procedure (www.wlu.ca/eohs).
The WLU Laboratory Health and Safety Manual provides detailed instructions
for dealing with major and minor spills. Before using ANY hazardous materials,
make sure you understand the proper clean-up procedure. All sample
preparation and disposal should be done in a separate lab, NOT in the room
with the FV1000.
6. PROTOCOL
Anyone using the confocal microscope must receive hands on training. This
document is a summary of the procedure and is only intended to help you
remember the various steps.
6.1 Start-up
You only need to turn on the components that you plan to use. Turning lasers
and the mercury lamp on and off and running them affects their lifespan, so if
you are sure you wont be using a given lamp or laser, dont turn it on. You
must turn on all the lasers you might potentially need at the start, because
you should not turn lasers on once the computer is on. The 405+635 diode laser
power supply also runs the laser combiner, so this component must always be
turned on, regardless of the wavelengths you are interested in viewing. If
you just need the computer to look at data or transfer it to a CD/DVD, turn
only the computer and monitors on.
If you will be using the microscope to collect z-stacks or time series
images, it is recommended to turn the system on, including the temperature
controller, up to two hours before imaging. This will allow all of the
components to reach thermal equilibrium and reduces z-drift during imaging.
ALL mounting media or other substances used to secure coverslips (i.e. nail
polish) must be COMPLETELY DRY before the slide is viewed on the confocal.
Mounting media may otherwise leak onto the objective and is very difficult to
remove.
1. TURN ON THE FAN.
2. Make sure the window shade is down all the way.
3. Sign-in in the user log book and fill in mercury lamp hours at start, and
lasers to be used.
4. Turn on the microscope components according to the numbers on labels:
1. U-RFL-T: Mercury burner power supply required for epifluorescence
observation.
2. Power bar. This will turn on the following:
a. FV10-MCPSU: Power supply for the diode lasers (405 and 643 nm).
b. Prior ProScan II: Stage controller.
c. IX2-UCB: Microscope controller power supply (on the windowsill).
d. Scan head power supply.
3. Only turn on if required for your dye: Melles Griot argon (458, 488, and
515 nm ) laser power supply. Turn on the power switch and then the
key.
5. Turn on the computer and monitors.
6. Log on to the computer:
a. Login: Administrator
b. Password: fluoview.
10
7. Double click on the Olympus FV10-ASW 1.7 software icon and log on using
your username and password.
8. Wait for the software and microscope to initialize.
9. Click on Device, then Microscope Controller.
11
10. Focus the condensor knob (#5 in Fugre 6-1) so that the edges of the
diaphragm AND the speciman appear very sharp (do not use the fine/course
focus for the objectives).
11. If nessicary, center the condensor using the centering screws (it helps to
open the field diaphragm so that the edges of the diaphragm almost fill the
field of view, #3 in Figure 6-1).
12. When the condensor is focused and centered, open the field diaphragm so
that it completely fills the eye pieces (#1 in Figure 6-1).
13. If you need DIC images only, proceed to 6.2, otherwise proceed to section
6.1.2 for epifluorescence imaging.
Figure 6-1: Inverted microscope components (live cell chamber not shown)
12
Figure 6-2: Software windows and basic description of controls for the FV-ASW software
d. Size (#3, Figure 6-2): intially set at 512 x 512 pixels, and adjusted
later to optimize resolution.
e. Area (#4, Figure 6-2): leave at 1:1 initally (no zoom).
f. Laser (#5, Figure 6-2): lasers are selected automatically for
fluorescence according to the dye that you select. Do not adjust
the laser powers at this point.
g. Microscope (#6, Figure 6-2): make sure the correct objective is
selected from the drop down menu.
6. In the Image Acquisition Control window:
a. If you are using more than one dye, check the Sequential box to
minimize bleedthrough.
i. Line sequential scan minimizines the time difference
between images for each channel by scanning only a line in
each channel at a time. This is typically the best choice.
ii. Frame sequential scan takes one full image on a given
channel before scanning the next channel.
b. Click on XY repeat to initiate sample scanning (#10, Figure 6-2).
Alternatively, the Focus x2 or Focus x4 buttons can be used to
obtain a very rough scan that skips lines, doubles or quadrupoles
the scanning speed, and minimizes photobleaching. Remember
that while the image is being scanned, the specimen is prone to
photobleaching avoid scanning for any longer than neccesary.
c. The focus between what is seen with epi (though the oculars) and
what is seen on the screen is slightly different. You wil need to
move the objective up VERY CAREFULLY (while in FINE FOCUS,
turn the knob very slowely counter clockwise) to bring the image
into focus. Only a very small amount of adjustment should be
required. If the image does not come in to focus easily, switch
back to a lower magnification and/or switch back to epiflourescence mode to focus again.
d. Adjust the brightness of the image:
i. If the signal is low in all channels, slow the scan speed
down while scanning until signal can be observed in all
channels.
ii. Press Ctrl+H to show the high and low limits of detection. If
the signal is too high and saturating the detector, it will
show up as red. Any blue on the image indicates zero signal
detection.
iii. Adjust HV on each of the channels (#12, Figure 6-2)
independantly until just a few pixels of red can be seen
(avoid setting HV higher than 700 as this will increase
background instrument noise). (Note: The HV for DIC in
the TD1 channel is usually much lower try starting around
200). If the HV is quite high and the signal is still too dim,
try adjusting the C.A. (confocal aperture); an increased CA
will increase brightness, but also decrease resolution, and
increase cross-section thickness. The scan rate can also be
6. If you are satisfied with the image, proceed to section 6.6 for image
analysis.
Single/Panel/Tile
Zoom, 1:1, Fit to
Window
Slider
Active Overlay
Numbered buttons
down the side of the
window (1, 2, etc)
Pencil button
Animation arrows
3-D button
Series Analysis
Line Series
Colocalization
Function
Use the LUT to change the gamma, intensity, contrast,
and pseudocolour of any channel to enhance dim images
or highlight certain features with a different colour.
Change the view of the sample, or look at more than one
image at once. Channels can be viewed separately or
overlaid using the Tile button.
These buttons adjust the size of the image on the screen.
Use to scan through the frames collected for this
speciman (primarily for Z- or T-series images).
Add text to the image describing Z position, time, or
wavelength.
Use to select the desired channels displayed in the image.
Allows access to region of interest (ROI) buttons and other
drawing tools (i.e. text, scale bar, grid). After a ROI is
selected, imaging can be conducted on that area
specifically to avoid unnecessary bleaching of entire
speciman in light path.
Used to play an animation or scroll through Z- or T-series.
Allows viewing of a Z-series from any angle, selection of
cross sections, rotation along a given axis and creation of
animation (using the More arrow).
Displays the intensity profile for a selected ROI. This
profile can be used to determine the size of a given
structure, for example.
Measure the size, area, etc., of selected ROI. Also
accessible from the Analysis Menu.
Displays an intensity profile (intensity vs. frequency) for
the selected ROI. Also accessible from the Analysis Menu.
Displays intensity vales for a given ROI and each channel
for each frame in a series (i.e. can use it to determine
which frame best displays a given feature/ROI). Also
accessible from the Analysis Menu.
Illustrates the change in intensity of a line as a series (Z
or T) progresses.
Graph displays extent of colocalization, and statistics
page calculations Pearson's coefficient, overlap and
colocalization indexes.
Function
Adjust digital zoom.
Change the text and line formatting for labeling ROIs, and
view information on ROIs selected.
If accessible, the acquisition conditions for the image can
be read.
Allows stepping through Z, time, or animation series
frames.
Determines the organization of the tiles from a set of
images.
Click to display the vertical or horizontal intensity
distribution along a line.
Function
Display image information in Data Manager Window.
Change how files are displayed in the Explorer window.
Filter Setting
and
Filter
Threshold
Image Calculation
Correcting Pixel Gaps
Correcting Z Gaps
Ratio/Concentration
Edit Experiment
Function
Mathematical filters can be used to improve image clarity
or emphasize certain characteristics. The filters available
are: sharpen, average, DIC, sobel, median, shading,
lapacian.
Sharpen: highlights the edges of images, improves blurred
images, but also increases noise.
Sobel: Emphasizes contours.
Lapacian: emphasizes intensity changes.
Press Preview at bottom of Filter window first, then
select various filters to observe result. Select the Single
or Series button, if doing a single image, enter the desired
frame in the Image Position box. If you want to save the
filtered image, press New Image.
This feature is used to make the image binary (two
colour). The C and Z indicate the channel and frame to
display. The thresholds can be adjusted by clicking on the
channel in the table below threshold, and manually
entering new values, or by moving the bars on the graph
Used to subtract/add/divide images or constants.
This window can be used to correct for image shift
between channels. Image position can be used to select
the desire frame. The white box in the Preview area can
be dragged to view other areas of the image.
To correct for Z position shift between channels. Moving
the yellow bar changes the Z-slice image.
The Ratio tab is used to look at changes in intensity
between two channels. Use the drop down menu to select
the desired operation, and the equation for that operation
will be displayed in the Equation box. Equations include
baseline/background subtraction, ratio of channels, FRET
image comparison (photo bleached image required).
The Concentration tab is used to analyze a change in ion
intensity over time. This feature requires knowledge of
intensity values with and without ion, and the dissociation
constant (in nM). See the Help menu for an explanation
of F values.
Combine single channels, append or extract series, from
two different images to create a new image or series
Additional Tools:
Virtual channel: Useful if fluorescent signals from two reagents overlap.
Delay image acquisition: After a set time has passed, acquisition can be started
or terminated using XYt imaging.
Image noise reduction: Can be done using Kalman filtering (takes several scans
and averages them).
Live plot window: plots change in intensity in ROI over time. First outline ROI in
Live View window by clicking on the pen tool. Then select Live Plot from Live
menu.
Live Tiling: allows viewing of the specimen as time elapses while the image is
being acquired (button in Live View window has the image of a wrench).
10
6.9 Shutdown
1. Switch back to the 10x objective, then manually lower the objective by
turning the knob clockwise.
2. Remove your slide.
3. If you used the oil objectives, make sure the oil has been removed as per
section 6.8.
4. Make sure you have saved your data and then exit the Fluoview software.
Close the FL (mercury lamp) shutter as recommended by the Caution
message.
5. If the argon laser (#3) it turned on, turn its key to the upright position,
but DO NOT TURN OFF THE POWER SWITCH YET. This allows the fan to
continue to blow until the laser has cooled down.
6. Turn off power bar (#2)
7. Turn off the Mercury power supply (#1).
8. When the fan has stopped running on the argon laser, the power can be
turned off (#3).
9. Transfer your data to a CD or memory stick if desired (recommended).
10. Shutdown the computer.
11. Cover the microscope. Make sure the cover DOES NOT touch the mercury
lamp, which is hot and may melt the cover.
12. Clean up anything you have left in the room.
13. Fill in the rest of the log book.
7. TROUBLESHOOTING
For troubleshooting tips, see the Help menu in the Fluoview software.
8. PREVENTATIVE MAINTENANCE
Users are not to perform maintenance. Unless noted otherwise, these
procedures are carried out by the Instrumentation Technician.
8.1 Monthly
-
If the system has not been used for a month or more, it will be run for ~ 8
hours to ensure long-term laser stability
Objective lenses will be checked and cleaned if required
Check the 5% CO2 and humidity canister for the Weather Station chamber
Check the CO2 for the anti-vibration table
11
12
REFERENCES
Laboratory Health and Safety Manual. 2007. Wilfrid Laurier University
Environmental/Occupational Health and Safety Office.
Olympus website: http://www.olympusfluoview.com.
Pawley, JB (Ed.). 2006. Handbook of Biological Confocal Microscopy, 3rd
Edition. Springer Science + Business Media: Singapore.
Staudt, T, Lang, MC, Medda, R, Engelhardt, J, & Hell, S.W. 2007. 2,2'thiodiethanol: a new water soluble mounting medium for high resolution
optical microscopy. Microsc Res Tech 70: 1-9.
Texas A&M University. Microscopy and Imaging Center. Olympus FV1000.
http://microscopy.tamu.edu/instruments/light-microscopy/olympusfv1000-confocal-microscope.html. Accessed May 26, 2008.
13
14
Excitation Wavelength
Emission Wavelength
Acid Fuchsin
540
630
Acridine Orange
(Bound to DNA)
502
526
455-600
560-680
Acridine Yellow
470
550
Acriflavin
Acridine Red
436
520
AFA (Acriflavin
Feulgen SITSA)
355-425
460
Alizarin Complexon
530-560
580
Alizarin Red
530-560
580
Allophycocyanin
650
661
ACMA
430
474
AMCA-S, AMC
345
445
Aminoactinomycin D
555
655
7-Aminoactinomycin D-AAD
546
647
Aminocoumarin
350
445
Anthroyl Stearate
361-381
446
Astrazon Brilliant
Red 4G
500
585
Astrazon Orange R
470
540
Astrazon Red 6B
520
595
Astrazon Yellow
7 GLL
450
480
Atabrine
436
490
Auramine
460
550
Aurophosphine
450-490
515
Aurophosphine G
450
580
BAO 9-(Bisaminophenyloxadiazole)
365
395
BCECF
505
530
Berberine Sulphate
430
550
Bisbenzamide
360
600-610
BOBO-1, BO-PRO-1
462
481
390
470
Blancophor SV
370
435
15
503
512
Bodipy TMR
542
574
Bodipy TR
589
617
BOPRO 1
462
481
Brilliant Sulphoflavin FF
430
520
Calcein
494
517
Calcien Blue
370
435
Calcium Green
505
532
Calcium Orange
549
576
Calcofluor RW Solution
370
440
Calcofluor White
440
500-520
380
475
365
435
548(low pH)
576(high pH)
587(low pH)
635(high pH)
6-Carboxyrhodamine 6G
525
555
Cascade Blue
400
425
Catecholamine
410
470
450-490
515
504(low pH)
514(high pH)
587(low pH)
540(high pH)
Coriphosphine O
460
575
Coumarin-Phalloidin
387
470
CY3.18
554
568
CY5.18
649
666
CY7
710
805
340
525
340-380
430
340
578
DAPI
350
470
DiA
456
590
280
460
Chinacrine
CL-NERF
Di-8-ANEPPS
488
605
310-370
520
DiI [DiIC18(3)]
549
565
DiO [DiOC18(3)]
484
501
Diphenyl Brilliant
Flavine 7GFF
430
520
DM-NERF
497(low pH)
510(high pH)
527(low pH)
536(high pH)
Dopamine
340
490-520
16
345
530
Eosin
525
545
Erythrosin ITC
530
558
Ethidium Bromide
510
595
Euchrysin
430
540
FIF (Formaldehyde
Induced Fluorescence)
405
435
375-530
612
Fluorescein
494
518
490
525
Fluo 3
485
503
FM1-43
479
Flazo Orange
598
Fura-2
363(low [Ca ])
335(high [Ca2+])
512(low [Ca2+])
505(high [Ca2+])
Fura Red
472(low [Ca2+])
436(high [Ca2+])
657(low [Ca2+])
637(high [Ca2+])
Genacryl Brilliant
Red B
520
590
Genacryl Brilliant
Yellow 10GF
430
485
Genacryl Pink 3G
470
583
Genacryl Yellow
5GF
430
475
Gloxalic Acid
405
460
Granular Blue
355
425
530-560
580
352
461
3-Hydroxypyrene-5,8,10-TriSulfonic Acid
403
513
7-Hydroxy-4methylcoumarin
360
455
380-415
520-530
Indo-1
350
405-482
Intrawhite Cf
Liquid
360
430
Leucophor PAF
370
430
Leucophor SF
380
465
Leucophor WS
395
465
Lissamine Rhodamine
B200 (RD200)
575
595
Lucifer Yellow CH
425
528
Lucifer Yellow VS
430
535
Haematoporphyrin
5-HydroxyTryptamine (5-HT)
2+
17
374
425
LysoSensor Green
DND-153, DND-189
442
505
384(low pH)
329(high pH)
540(low pH)
440(high pH)
LysoTracker Green
504
511
LysoTracker Yellow
534
551
LysoTracker Red
577
592
Magdala Red
524
600
Magnesium Green
506
531
Magnesium Orange
550
575
Maxilon Brilliant
Flavin 10 GFF
450
495
Maxilon Brilliant
Flavin 8 GFF
460
495
Mitotracker Green FM
490
516
Mitotracker Orange
CMTMRos
551
576
364
395
Mithramycin
450
570
NBD
465
535
LysoSensor Yellow/Blue
NBD Amine
450
530
Nile Red
515-530
525-605
Nitrobenzoxadidole
460-470
510-650
340
490-520
Noradrenaline
Nuclear Fast Red
289-530
580
Nuclear Yellow
365
495
460
510
496
524
503
522
511
530
Pararosaniline (Feulgen)
570
625
Phorwite AR Solution
360
430
Phorwite BKL
370
430
Phorwite Rev
380
430
Phorwite RPA
375
430
Phosphine 3R
465
565
Phosphine R
480-565
578
535-553
605
POPO-1, PO-PRO-1
434
456
Primuline
410
550
Procion Yellow
470
600
18
536
617
410
540
540-590
560-650
365
495
Quinacrine Mustard
423
503
R-phycoerythrin
565
575
Rhodamine 110
496
520
Rhodamine 123
511
534
Rhodamine 5 GLD
470
565
Rhodamine 6G
526
555
Pyronine B
Rhodamine B
540
625
523-557
595
Rhodamine B Extra
550
605
Rhodamine BB
540
580
Rhodamine BG
540
572
Rhodamine Green
fluorophore
502
527
Rhodamine Red
570
590
Rhodamine WT
530
555
499
525
Rose Bengal
540
550-600
Serotonin
365
520-540
520
595
500
583
530
590
Sevron Orange
440
530
Sevron Yellow L
430
490
SITS (Primuline)
395-425
450
365
460
Sodium Green
507
535
Stilbene
335
440
Snarf 1
563
639
520
595
470
570
504
523
494 6
515 7
515 7
543 13
621
634
Tetracycline
390
560
557
576
Rhodamine B 200
19
596
615
Thiazine Red R
510
580
Thioflavin S
430
550
Thioflavin TCN
350
460
Thioflavin 5
430
550
370-385
477-484
Thiozol Orange
453
480
Tinopol CBS
390
430
TOTO 1, TO-RRO-1
514
533
TOTO 3, TO-PRO-3
642
661
True Blue
365
420-430
Ultralite
656
678
Uranine B
420
520
Uvitex SFC
365
435
X-Rhodamine
580
605
Xylene Orange
546
580
XRITC
582
601
YOYO-1, YOYO-PRO-1
491
509
YOYO-3, YOYO-PRO-3
612
631
Thiolyte
20
21
NAME &
EXTENSION
SUPERVISOR
SAMPLE AND
DYES USED
MERCURY LAMP
HOURS AT
START
HE-NE
LASER
USED (Y/N)
OBJECTIVES USED
(AIR: 10, 20, 40,
OIL: 60 OR 100)
MERCURY
LAMP HOURS
AT END
TIME OUT
PROBLEMS /
COMMENTS
22
23
DATE
NAME
EXT #
TYPE OF MAINTENANCE
FREQUENCY OF MAINTENANCE
(I.E. WEEKLY)
PROBLEMS / COMMENTS