Features
Precise bi-directional and highly dynamic flow measurement with the non-intrusive clamp-on technology
High precision at fast and slow flow rates, high temperature and zero point stability
Portable, easy-to-use flow transmitter with 2 flow channels, multiple inputs/outputs, an integrated data logger
with a serial interface
Water and dust-tight (IP65); resistant against oil, many
liquids and dirt
Li-Ion battery provides up to 14 hours of measurement
operation
Automatic loading of calibration data and transducer
detection for a fast and easy set-up (less than 5 min),
providing precise and long-term stable results
User-friendly design
Transducers available for a wide range of inner pipe
diameters ( 7...1600 mm ) and fluid temperatures
( -40...+200 C)
Probe for wall thickness measurement available
Robust, water-tight (IP67) transport case with comprehensive accessories
QuickFix for fast mounting of the flow transmitter in difficult conditions
Applications
Designed for industrial use in harsh environments, in gas
processing and natural gas extraction, chemical industry
and in the petroleum industry. Practical applications:
Measurement on natural gas pipelines and in natural
gas storage installations
Measurement of synthesized gas and injection gas
Measurement for the gas supply industry
Supervision of permanently installed meters, service
and maintenance
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
FLUXUS G601
Technical Specification
Table of Contents
Function ........................................................................................................................................................... 3
Measurement Principle ..................................................................................................................................... 3
Calculation of Volumetric Flow Rate ................................................................................................................. 3
Number of Sound Paths.................................................................................................................................... 4
Typical Measurement Setup ............................................................................................................................. 5
Standard Volumetric Flow Rate ........................................................................................................................ 5
Flow Transmitter ............................................................................................................................................. 6
Technical Data .................................................................................................................................................. 6
Dimensions ....................................................................................................................................................... 8
Standard Scope of Supply ................................................................................................................................ 9
Connection of Adapters................................................................................................................................... 10
Example for the Equipment of a Transport Case ............................................................................................ 11
Transducers................................................................................................................................................... 12
Transducer Selection ...................................................................................................................................... 12
Transducer Order Code .................................................................................................................................. 15
Technical Data ................................................................................................................................................ 16
Transducer Mounting Fixture ...................................................................................................................... 21
Coupling Materials for Transducers............................................................................................................ 23
Damping Mats (optional) .............................................................................................................................. 24
Connection Systems..................................................................................................................................... 25
Transducer Cable............................................................................................................................................ 25
Clamp-on Temperature Probe (optional) .................................................................................................... 26
Wall Thickness Measurement (optional)..................................................................................................... 27
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
Function
Measurement Principle
In order to measure the flow of a medium in a pipe, ultrasonic signals are used, employing the transit time difference principle. Ultrasonic signals are emitted by a transducer installed on the pipe and received by a second transducer. These signals are emitted alternately in the flow direction and against it.
As the medium in which the signals propagate is flowing, the transit time of the ultrasonic signals in the flow
direction is shorter than against the flow direction.
The transit time difference, t, is measured and allows the flowmeter to determine the average flow velocity
along the propagation path of the ultrasonic signals. A flow profile correction is then performed in order to obtain the area averaged flow velocity, which is proportional to the volumetric flow rate.
Two integrated microprocessors control the entire measuring process. This allows the flowmeter to remove
disturbance signals, and to check each received ultrasonic wave for its validity which reduces noise.
t
Path of the ultrasonic signal
t
1
t
2
V
kRe
A
ka
t
tfl
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
FLUXUS G601
Technical Specification
a - transducer distance
a<0
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
damping mat
transducers
U L T R A S O N IC
F L O W M E T E R
C H A N N E L A
C H A N N E L B
Q
N E X T
Q -
E N T E R
M U X
1
0
O N
5
Q
6
2
O F F
L IG H T
3 x O F F
B R K
D IS P
B A T T E R Y
Q +
E N T E R
D IS P
M O D E
O N
transmitter
outputs
temperature probe
inputs
VN
V
pN
p
TN
T
K
The operational pressure p and the operational temperature T of the medium will be entered directly as fixed
values into the transmitter.
or:
If inputs are installed (optional), pressure and temperature can be measured by the customer and fed in the
transmitter.
The gas compressibility coefficient K of the gas is entered in the transmitter:
as fixed value or
as approximation according to e.g. AGA8 or GERG
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
FLUXUS G601
Technical Specification
Flow Transmitter
Technical Data
FLUXUS
design
measurement
measurement principle
flow velocity
repeatability
medium
temperature compensation
accuracy
volumetric flow rate
flow transmitter
power supply
battery
power consumption
number of flow measuring
channels
signal attenuation
measuring cycle (1 channel)
response time
housing material
degree of protection according to IEC/EN 60529
dimensions
weight
fixation
ambient temperature
display
menu language
measuring functions
physical quantities
totalizer
calculation functions
diagnostic functions
data logger
loggable values
capacity
G601
portable
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS
communication
interface
serial data kit
software (all Windows
versions)
FLUXUS G601
G601
RS232/USB
- FluxData: download of measurement data, graphical presentation,
conversion to other formats (e.g. for Excel)
- FluxKoef: creating medium data sets
able
adapter
transport case
dimensions
outputs
number
accessories
range
accuracy
active output
passive output
range
open collector
optorelay
binary output as alarm output
- functions
limit, change of flow direction or error
binary output as pulse output
- pulse value
0.01...1000 units
- pulse width
1...1000 ms
inputs
The inputs are galvanically isolated from the transmitter.
number
see standard scope of supply on page 9, max. 4
accessories
input adapter (if number of inputs > 2)
temperature input
type
Pt100/Pt1000
connection
4-wire
range
-150...+560 C
resolution
0.01 K
accuracy
0.01 % of reading 0.03 K
current input
accuracy
0.1 % of reading 10 A
passive input
Ri = 50 , Pi < 0.3 W
- range
-20...+20 mA
voltage input
range
0...1 V
accuracy
0.1 % of reading 1 mV
internal resistance
Ri = 1 M
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
FLUXUS G601
Technical Specification
Dimensions
59
FLUXUS G601
C H A N N E L B
Q
N E X T
O N
Q -
E N T E R
M U X
D IS P
5
Q
M O D E
3 x O F F
B R K
E N T E R
D IS P
O F F
L IG H T
Q +
213
U L T R A S O N IC F L O W M E T E R
C H A N N E L A
B A T T E R Y
O N
226
in mm
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
G601
Extended Standard
flow measurement on gas
application
calculation of standard
volumetric flow rate
transducer frequency
outputs
passive current output
binary output
frequency output
inputs
temperature input
passive current input
voltage input
accessories
transport case
power supply unit,
mains cable
battery
output adapter
input adapter
adapter for voltage and
current inputs
QuickFix pipe
mounting fixture for
transmitter
serial data kit
measuring tape
wall thickness probe
user manual,
Quick Start Guide
connector board at the
upper side of the
transmitter
G601
Multifunctional
G601 CA-Energy
flow measurement on
compressed air, industrial
gases and liquids
2
2
-
2
1
1
2
2
1
2
2
-
2
-
1
2
1
2
2
-
x
x
x
x
x
x
x
x
x
-
x
2
-
x
x
2
3
x
2
2
x
x
x
x
x
x
x
x
x
x
x
x
x
x
C H A
C H B
-
C O M M
P 4
P 3
P 2
D C
P 1
- IN
C H
T 1 /T 3
A
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
T 2 /T 4
C H
C O M M
P 4
P 3
P 2
+
O u tp u t
D C
P 1
- IN
C H
T 1 /T 3
P 3 ...P 8
A
T 2 /T 4
C H
C O M M
P 2
D C
P 1
- IN
C H
T 1 /T 3
A
In p u t
O u tp u t
T 2 /T 4
C H B
C O M M
P 4
P 3
P 2
+
+
O u tp u t
D C - IN
P 1
+
In p u t
O u tp u t
In p u t
FLUXUS G601
Technical Specification
O u tp u t
P 6
P 7
P 8
Connection of Adapters
P 4
P 5
input adapter
P 3
+
output adapter
transducers
measuring channel A
transducers
measuring channel B
C H
P 3 ...P 8
A
T 1 /T 3
T 2 /T 4
C H
C O M M
P 2
D C
P 1
- IN
+
O u tp u t
outputs
In p u t
inputs
RS232
10
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
coupling compound
transducers
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
transmitter
11
FLUXUS G601
Technical Specification
Transducers
Transducer Selection
Step 1a
Select a Lamb wave transducer:
transducer order code
GLG
11
GLH
GLK
GLM
23
15
9
5
GLP
1...3
0.5...1
GLQ
10
15
20
25
pipe wall thickness [mm]
Step 1b
If the pipe wall thickness is not in the range of the Lamb wave transducers, select a shear wave transducer:
transducer order code
14
GSG
GSK
GSM
2.5
GSP
GSQ
1.5
1
23
1.5
5
10
15
20
25
pipe wall thickness [mm]
recommended
possible
Step 2
inner pipe diameter d dependent on the flow velocity v of the medium in the pipe
The transducers are selected from the characteristics (see next page). Lamb wave transducers are selected
from the left column, shear wave transducers from the right column.
Lamb wave transducers: If the values d and v are not in the range, the diagonal arrangement with 1 sound
path may be used, i.e. the same characteristics can be used with doubling the inner pipe diameter. If the values are still not in the range, shear waves transducers regarding the pipe wall thickness have to be selected
in step 1b.
12
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
Dmm
1200
GLG
Dmm
1200
1000
1000
800
800
600
600
400
400
200
200
10
15
20
25
30
35
vms
Dmm
1200
GSG
10
15
20
25
30
35
vms
GLH
1000
800
600
400
200
0
10
15
20
25
30
35
vms
Dmm
1200
GLK
1000
1000
800
800
600
600
400
400
10
15
20
25
30
35
vms
Dmm
120
100
100
80
80
60
60
40
40
20
15
20
25
30
35
vms
GSM
20
5
10
15
20
25
30
35
vms
Dmm
10
15
20
25
30
35
vms
Dmm
GLP
140
120
100
100
80
80
60
60
40
40
20
GSP
140
120
20
10
15
20
25
30
35
vms
Dmm
10
15
20
25
30
35
vms
Dmm
GLQ
140
120
100
100
80
80
60
60
40
40
20
GSQ
140
120
10
140
120
Dmm
GLM
140
GSK
200
200
Dmm
1200
20
5
10
15
20
25
30
35
vms
10
15
20
25
30
35
vms
inner pipe diameter and max. flow velocity for a typical application with natural gas, nitrogen, oxygen in reflection arrangement with
2 sound paths (Lamb wave transducers)/1 sound path (shear wave transducers)
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
13
FLUXUS G601
Technical Specification
Step 3
min. medium pressure
Lamb wave transducer
transducer
medium pressure1 [bar]
order code
metal pipe
plastic pipe
min.
min. extended
min.
GLG
15
10
1
GLH
15
10
GSK
30
20
GLK
GSM
30
20
GSP
30
20
GSQ
30
20
GLM
GLP
GLQ
1
depending on application, typical absolute value for natural gas, nitrogen, compressed air
Example
step
1
2
mm
mm
m/s
bar
12
GLG or GLH
800
15
GLG
12
GLG or GLH
600
15
GLG or GLH
17
GLG
17
GLG or GLH
influence of acoustic
noise is reduced with
increased transducer
frequency, thus
recommended: GLH
12
GLG or GLH
800
30
values not in the range
of the characteristics,
but by using diagonal
arrangement with 1
sound path, the inner
pipe diameter in the
characteristics is doubled: GLG
17
GLG
30
GS
300
15
GSK
35
GSK
Step 4
for the characters 4...11 of the transducer order code (ambient temperature, explosion protection, connection
system, extension cable) see page 15
Step 5
for the technical data of the selected transducer see page 16 et seqq.
14
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
5, 6
7, 8
9...11
no. of character
extension cable
connection
system
explosion
protection
ambient
temperature
transducer
frequency
transducer
description
GL
GS
0.2 MHz
0.5 MHz
1 MHz
2 MHz
4 MHz
N
000
example
GL
NN
NL
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
15
FLUXUS G601
Technical Specification
Technical Data
Shear Wave Transducers
GDG1NZ7
GSG-NNNNL
MHz 0.2
GDK1NZ7
GSK-NNNNL
0.5
medium pressure1
min. extended
min.
bar
bar
metal pipe: 20
metal pipe: 30
plastic pipe: 1
metal pipe: 20
metal pipe: 30
plastic pipe: 1
250
380
810
1100
70
80
500
720
14
-
5
-
1699
5
1699
5
mm
mm
mm
129.5
51
67
126.5
51
67.5
contact surface
degree of protection
according to IEC/
EN 60529
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing
technical type
order code
transducer frequency
ambient temperature
min.
C
max.
C
temperature
compensation
-40
+130
x
-40
+130
x
depending on application, typical absolute value for natural gas, nitrogen, compressed air
16
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
GDP1NZ7
GSP-NNNNL
2
GDQ1NZ7
GSQ-NNNNL
4
medium pressure1
min. extended
min.
bar
bar
metal pipe: 20
metal pipe: 30
plastic pipe: 1
metal pipe: 20
metal pipe: 30
plastic pipe: 1
metal pipe: 20
metal pipe: 30
plastic pipe: 1
30
40
80
120
15
20
40
60
6
10
20
30
2.5
-
1.5
-
1
-
1699
4
1699
4
1699
3
mm
mm
mm
60
30
33.5
60
30
33.5
42.5
18
21.5
ambient temperature
min.
C
max.
C
temperature
compensation
remark
-40
+130
x
contact surface
degree of protection
according to IEC/
EN 60529
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing
-40
+130
x
technical type
order code
transducer frequency
-40
+130
x
on request
depending on application, typical absolute value for natural gas, nitrogen, compressed air
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
17
FLUXUS G601
Technical Specification
GDP1EZ7
GSP-ENNNL
2
GDQ1EZ7
GSQ-ENNNL
4
medium pressure1
min. extended
min.
bar
bar
metal pipe: 20
metal pipe: 30
plastic pipe: 1
metal pipe: 20
metal pipe: 30
plastic pipe: 1
metal pipe: 20
metal pipe: 30
plastic pipe: 1
30
40
80
120
15
20
40
60
6
10
20
30
2.5
-
1.5
-
1
-
1699
4
1699
4
1699
3
mm
mm
mm
60
30
33.5
60
30
33.5
42.5
18
21.5
contact surface
degree of protection
according to IEC/
EN 60529
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing
technical type
order code
transducer frequency
ambient temperature
min.
C
max.
C
temperature
compensation
remark
-30
+200
x
-30
+200
x
-30
+200
x
on request
depending on application, typical absolute value for natural gas, nitrogen, compressed air
18
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
GRK1NC3
GLK-NNNNL
0.5
medium pressure1
min. extended
bar
metal pipe: 10
metal pipe: 10
min.
bar
metal pipe: 15
plastic pipe: 1
metal pipe: 15
plastic pipe: 1
metal pipe:
10 (d > 120 mm)
5 (d < 120 mm)
metal pipe:
15 (d > 120 mm)
10 (d < 120 mm)
plastic pipe: 1
190
220
900
1600
120
140
600
1000
60
80
300
500
11
23
7
15
4
9
1699
5
1699
5
1699
5
mm
mm
mm
128.5
51
67.5
128.5
51
67.5
128.5
51
67.5
contact surface
degree of protection
according to IEC/
EN 60529
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing
ambient temperature
min.
C
max.
C
temperature
compensation
-40
+170
x
-40
+170
x
GRG1NC3
GLG-NNNNL
MHz 0.2
technical type
order code
transducer frequency
-40
+170
x
depending on application, typical absolute value for natural gas, nitrogen, compressed air
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
19
FLUXUS G601
Technical Specification
GRP1NC3
GLP-NNNNL
2
GRQ1NC3
GLQ-NNNNL
4
medium pressure1
min. extended
min.
bar
bar
metal pipe:
10 (d > 60 mm)
5 (d < 60 mm)
plastic pipe: 1
metal pipe:
10 (d > 35 mm)
5 (d < 35 mm)
plastic pipe: 1
metal pipe:
10 (d > 15 mm)
5 (d < 15 mm)
plastic pipe: 1
30
40
90
150
15
20
50
70
7
10
22
35
2
5
1
3
0.5
1
1699
4
1699
4
1699
3
mm
mm
mm
74
32
40.5
74
32
40.5
42
22
25.5
ambient temperature
min.
C
max.
C
temperature
compensation
remark
-40
+170
x
contact surface
degree of protection
according to IEC/
EN 60529
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing
-40
+170
x
technical type
order code
transducer frequency
-40
+170
x
on request
depending on application, typical absolute value for natural gas, nitrogen, compressed air
20
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
7...9
no. of character
outer pipe
diameter
fixation
size
measurement
arrangement
description
transducer
transducer
mounting fixture
1, 2
VP
portable Variofix
TB
tension belts
A
all transducers
D
reflection arrangement
S
small
medium
C
chains
without fixation
055
10...550 mm
150
50...1500 mm
210
50...2100 mm
055
example
VP
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
21
FLUXUS G601
Technical Specification
tension belts TB
22
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
Technical Data
type
order code
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
ambient temperature
material
C
-30...+130
mineral grease paste
-30...+200
silicone paste
-30...+250
fluoropolymer paste
23
FLUXUS G601
Technical Specification
b
D
reflection arrangement
description
outer pipe
diameter
mm
dimensions
lxbxh
mm
transducer
techniambient
frequency
cal type temperature
G H K M P
C
remark
< 80
80
x
-
x
x
-
x D20S3
- D20S2
- D50S2
-25...+60
< 300
x A20S4
-25...+60
300
l x 100 x 0.9
x B35R2
-35...+50
transducer frequency
G, H
K, M, P
12
24
32
6
12
16
transducer frequency
G, H
m
mm
300
500
1000
24
12
32
126
K, M, P
m
6
16
63
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
Connection Systems
connection system NL
x
transmitter
transducer frequency
(3d character of transducer
order code)
N
L
1
cable length m
G, H, K
x
2
M, P
y
3
l1
25
x
2
Q
y
2
l1
25
x
2
S
l1
25
y
1
x
1
y
1
l
20
Transducer Cable
Technical Data
type
standard length
max. length
ambient temperature
sheath
material
outer diameter
cable jacket
material
outer diameter
thickness
color
shield
transducer cable
1699
see table above
2551
-
m
C
-55...+200
mm
PTFE
2.9
0.3
brown
x
TPE-O
8
PE
6
0.5
black
x
mm
mm
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
black
x
extension cable
1750
5
10
10
< 80
25
FLUXUS G601
Technical Specification
PT12N
670415-1
Pt100
connection
measuring range
accuracy T
accuracy T
kg
2x Pt100 matched
according to EN 1434-1
4-wire
-30...+250
(0.15 C + 2 . 10-3 . |T [C]|), class A
0.1 K (3 K < T < 6 K),
more corresponding to
EN 1434-1
50
aluminum
IP66
0.25
PT12F
PT12F
670415-2
670414-2
short response time
Pt100
2x Pt100 matched
according to EN 1434-1
4-wire
-50...+250
(0.15 C + 2 . 10-3 . |T [C]|), class A
0.1 K (3 K < T < 6 K),
more corresponding to
EN 1434-1
8
PEEK, stainless steel 304 (1.4301), copper
IP66
0.5
0.32
clamp-on
15
15
20
14
30
27
B
mm
mm
mm
A
extension cable
0.64
clamp-on
response time
housing
degree of protection
according to IEC/
EN 60529
weight (without
connector)
fixation
accessories
thermal conductivity
paste 200 C
thermal conductivity
foil 250 C
plastic protection plate,
insulation foam
dimensions
length l
width b
height h
dimensional drawing
PT12N
670414-1
Connection
Temperature Probe
red/blue
red
white/blue
white
Connector
pin
1
2
3, 4, 5
6
7
8
extension cable
blue
gray
red
white
2
4
5
8
1
6
3
7
Cable
type
standard length
max. length
cable jacket
26
m
m
extension cable
LIYCY 8 x 0.14 mm gray
5/10/25
200
PVC
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
Technical Specification
FLUXUS G601
Technical Data
technical type
measuring range1
resolution
accuracy
medium temperature
able
type
length
mm
mm
DWR1NZ7
DWQ1NZ7
1...250
0.01
1 % 0.1 mm
-20...+200,
short-time peak max. 500
1...200
0.01
1 % 0.1 mm
-20...+60
2616
1.5
1.5
The measuring range depends on the attenuation of the ultrasonic signal in the
pipe. For strongly attenuating plastics (e.g. PFA, PTFE, PP) the measuring range is smaller.
Cable
type
ambient temperature
cable jacket
material
outer diameter
color
shield
mm
2616
<200
FEP
5.1
black
x
DWR1NZ7
DWQ1NZ7
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25
27
FLUXUS G601
Technical Specification
FLEXIM GmbH
internet: www.flexim.com
Wolfener Str. 36
e-mail: info@flexim.com
12681 Berlin
28
Germany
TSFLUXUS_G601V1-5-1EN_Leu, 2014-02-25