Beruflich Dokumente
Kultur Dokumente
Measure
VOCs
ppb, ppm, %
in air, water,
soil, waste
Solvents
1,3 BD
ETO
VCM
THC
S cpds
P cpds
H 2S
NH3 PH3
AsH3
BTEX
othes:
Call us
Technologies
Photoionization, w/l P hotoionization, Flame Ionization, Thermal
Conductivity, Far UV absorbance, Flame P hotometer (S, P)
Applications
Portable GCs
Model 121
Leak Detection VOC s
Simple Apps (1-4 cpds)
PID Only 1 detector
Model 122
Air/water/IH surveys VOC s
Apps. 1-10 cpds)
PID only (1 detector)
Model 312
Ambient air, water, soil, QC
Apps 1 to >30 cpds
PID, FID, TCD, FPD, FUV
( i or 2 detectors)
Embedded PC/WOS 7.0/PW
UPGRADE YOUR LAB WITH ONE OF OUR EXCELLENT PORTABLE GCS
CHROMATOGRAMS
PI D
PID
The PID provides a response to a wide range of organic and
some inorganic compounds at part per billion (ppb) levels.
The HNU PID consists of an ultraviolet lamp and an ion
chamber. The detector measures the concentration of
gases present in a sample using photoionization.
Photoionization occurs when a molecule absorbs a photon
(light energy) of sufficient energy, creating a positive ion
and an electron as shown below:
Schematic of PID
FID
The flame ionization prosess occursonly for
hydrocarbons when a carbon- carbon bond is broken
via a thermal process in the flame. This results in
the formation of carbon ions which are collected in
the H2/air flame in an ionization chamber. The
response (current) is proportional to the
concentration.
TCD
Measures the difference between the thermal
transfer characteristics of the gas and a
reference cell. The sample and reference filaments
are two legs of a Wheatstone Bridge.A constant
current is applied with a resultant rise/decrease in
the filament temperature as the sample passes.
This difference in resistance is proportional to the
concentration.
FPD
The sample is burned in a hydrogen rich
flame which excites sulfur or phosphorus and
produces chemiluminescence emission of a blue (S394 nm) or green (P- 524 nm) photon. This emission
is detected by a PMTwith an interference filter.
FUV
The Far Ultraviolet Detector (FUV) provides a nearly
universal response (except for the noble gases) to
organic and inorganic compounds at low part per million
(ppm) levels. The internal volume of the FUV is only 40 mL
making it an ideal choice for use with capillary columns.
The
FUVAD consists of a 10.2 eV ultraviolet lamp, an
absorption cell, and a photodiode.
FEATURES/RANGE
Model 54-00 FID
Measures: selective for hydrocarbons
Concentration: sub ppm to %
Destructive detector: mass flow
the output of which is directly
proportional to the ratio of the
compounds carbon mass to the total
compound mass. Range: 106
EPA Methods, OSHA Methods
Quality control
Specific for S or P
Range S: > 103
Range P: > 105
ppb to ppm
GC312
GC 312
GC121
Hand held field portable GC for1-4 commpounds.
Measures peak height (PH) of component within a
retention time window and auto zero.. Data sample
storage and up to 5 chromatograms.Bluetooth
Weight: belt mounted tank 3.2 pounds and 10 liters of
1-4 compounds
carrier gas
GC122
Hand carry this portable GC for 1-10 compounds. PH
measurement and autozero. Data storage 100+
data plus chromatograms. Carrier gas tank
enclosed. Bluetooth
Common Parameters
Columns: 15, 30 M Capillary columns (3 D), 2-3 M
micropack columns. columns are easily changeable
Prgrammable parameters: Temperature 30-60oC,
Retention time, window, analysis time, response
factor, library
1-10 compounds
Dimensions:10" W x 4.75 H
x10.75 L; Weight 5.2#
GC312
GC122
GC121
Processor
Display
PC with windows 7
7 color touch Screen
Detectors
U Processor
LCD 4 line x 16 character
backlit
PID only
U Processor
LCD 2 line x 16 character
backlit
PID only
Weight
Dimensions
Battery operation
Battery type
Temperature control
Dual ovens
Data Processing
Columns
Number of components
Availability
10.5 L x 3 W x 4.5 H
8-10 hours
NiMH
30-75 oC Isothermal
No
Concentration PH only,
display on remote PC in
Excel, transfer by
Bluetooth