Beruflich Dokumente
Kultur Dokumente
User Manual
Process Instruments
PN 13-2270 REV L
Offices
USA - HEADQUARTERS
AMETEK Process Instruments
150 Freeport Road
Pittsburgh, PA 15238, USA
Phone: 412-828-9040
Fax: 412-826-0399
USA - Delaware
AMETEK Process Instruments
455 Corporate Boulevard
Newark, Delaware 19702, USA
Phone: 302-456-4400 (Main)
800-537-6044 (Service)
800-222-6789 (Ordering)
Fax: 302-456-4444
USA - Texas
AMETEK Process Instruments
9750 Whithorn Drive
Houston, Texas 77095, USA
Phone: 281-463-2820
Fax: 281-463-2701
USA - Oklahoma
Chandler Engineering
2001 North Indianwood Avenue
Broken Arrow, OK 74012, USA
Phone: 918-250-7200
Fax: 918-459-0165
CANADA
AMETEK Process Instruments
2876 Sunridge Way N.E.
Calgary, Alberta
Canada T1Y 7H9
Phone: 403-235-8400
Fax: 403-248-3550
GERMANY
AMETEK GmbH
Postfach 2165
D-40644 Meerbusch
OR
Rudolf-Diesel Strasse 16
D-40670 Meerbusch
Germany
Phone: 49-21-59-9136-0
Fax: 49-21-59-9136-39
SINGAPORE
AMETEK Singapore PVT. Ltd.
10 Ang Mo Kio Street 65
#05-12 Techpoint
Singapore 569059
Republic of Singapore
Phone: 65-484-2388
Fax: 65-481-6588
FRANCE
AMETEK - APIF
Rond Point de lpine des champs
Buroplus Bat D
78990 Elancourt, France
Phone: 33 1 30 68 89 20
Fax: 33 1 30 38 89 29
CHINA
AMETEK Process Instruments
Room 406, Metro Tower
No. 30 Tian Yao Qiao Road
Shanghai, 200030
Phone: 8621 6426 7049
Fax: 8621 6426 7054
2005 AMETEK
This manual is a guide for the use of the AMETEK ChanScope and Dew Point Tester models. Data herein has been verified and validated and is
believed adequate for the intended use of this software. If the software or procedures are used for purposes over and above the capabilities specified
herein, confirmation of their validity and suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no obligation or
liability. This publication is not a license to operate under, or a recommendation to infringe upon, any process patents.
Table of Contents
CHAPTER 1 Overview
Dew Point Tester.................................................................................................. 1-1
Description.......................................................................................... 1-1
Bureau of Mines-Type Dew Point Tester (Standard).................... 1-1
ChanScope.................................................................................... 1-1
Chiller................................................................................................. 1-3
Illuminated Magnifier......................................................................... 1-3
Tripod.................................................................................................. 1-3
Propane Cylinder................................................................................ 1-3
CO2 Bottle........................................................................................... 1-3
Accuracy............................................................................................. 2-1
Dew Point Temperature Range........................................................... 2-1
Operating Pressure.............................................................................. 2-1
Connections......................................................................................... 2-1
Weight and Dimensions...................................................................... 2-1
Accuracy............................................................................................. 2-2
Dew Point Temperature Range........................................................... 2-2
Operating Pressure.............................................................................. 2-2
Connections......................................................................................... 2-2
Power.................................................................................................. 2-2
Weight and Dimensions...................................................................... 2-2
| iii
CHAPTER 4 Operation
Using Glycol Filters............................................................................................. 4-1
CHAPTER 6 Drawings
DWG 13-0052 - Glass Window Assy (sour gas)................................................. 6-1
DWG 13-0054 - Glycol Filter Assy..................................................................... 6-2
DWG 13-0072 - Immersion Cooling Rod Assy.................................................. 6-3
DWG 13-0075 - Model 13-0075 Dew Point Tester............................................. 6-4
DWG 13-0090 - Chillers, Regular and Quick..................................................... 6-5
DWG 13-0095 - Saturated Moisture Content Air or Nitrogen (F)..................... 6-6
DWG 13-0095 - Saturated Moisture Content Air or Nitrogen (C).................... 6-7
DWG 13-0097 - Saturated Moisture Content Natural Gas (F).......................... 6-8
DWG 13-0097 - Saturated Moisture Content Natural Gas (C).......................... 6-9
Table - Equilibrium Water Vapor Contents........................................................ 6-10
DWG 13-0100 - Bureau of Mines Dew Point Tester......................................... 6-15
DWG 13-0125 - Conversion Chart, Dew Point to Moisture Content................ 6-16
DWG 13-0150 - Installation, Dew Point Tester................................................ 6-17
DWG 13-0250 - Dew Point Indicator Assy....................................................... 6-18
DWG 13-0303 - Illuminated Magnifier Assy (Model 13-075).......................... 6-19
DWG 13-0304 - Illuminated Magnifier (Models 13-100, 13-110).................... 6-20
DWG 13-0305 - Illuminator Assy..................................................................... 6-21
DWG 13-1250 - ChanScope II.......................................................................... 6-22
Safety Notes
warnings, cautions, and notes contained in this manual emphasize critical instructions as follows:
An operating procedure which, if not strictly observed, may result in personal injury or environmental contamination.
An operating procedure which, if not strictly observed, may result in damage to the equipment.
NOTE
Electrical Safety
Always shut down power source(s) before performing maintenance or troubleshooting. Only a qualified
electrician should make electrical connections and ground checks.
Any use of the equipment in a manner not specified by the manufacturer may impair the safety protection
originally provided by the equipment.
Grounding
Instrument grounding is mandatory. Performance specifications and safety protection are void if instrument is operated from an improperly grounded power source.
Verify ground continuity of all equipment before applying power.
| v
Warning Labels
These symbols may appear on the instrument in order to alert you of existing conditions.
Protective Conductor Terminal
(BORNIER DE LECRAN DE PROTECTION)
Schutzerde
Caution - Risk of electric shock
(ATTENTION-RISQUE DE DCHARGE LECTRIQUE)
Achtung - Hochspannung Lebensgefahr
Caution - (Refer to accompanying documents)
(ATTENTION-SE RFERER AUX DOCUMENTS JOINTS)
Achtung (Beachten Sie beiliegende Dokumente)
CAUTION - Hot Surface
(ATTENTION-SURFACE CHAUDE)
Achtung - Heie Oberflche
When the equipment is ready to be disposed of, take it to your local or regional waste collection administration for recycling.
In some cases, your end-of-life product may be traded in for credit towards
the purchase of new AMETEK instruments. Contact your dealer to see if this
program is available in your area.
| vii
OVERVIEW
ChanScope
The digital ChanScope provides state-of-the-art temperature measurement of the
dew point of pressurized gas. The ChanScope retains all of the proven reliability and quality of the original Bureau of Mines-type dew point tester and gives
instantaneous feedback using an LEDreadout during a test.
Both the standard (Bureau of Mines type) dew point tester and the ChanScope
include a sample pressure gauge and tripod socket for ease of use in the field.
NOTE
Overview | 1-1
Theory of Operation
Water contamination in gases can cause corrosion and other equipment problems,
especially in regulation and metering stations. The determination of hydrocarbon
contamination in pipeline gas is necessary to meet hydrocarbon dew point specifications in natural gas transportation contracts. Gases most frequently tested for
water content include natural gas, sour gas, heat treating furnace gas, and industrial gases such as nitrogen, oxygen, hydrogen, and other bottled gases. The most
accurate and widely used instrument available for determining water content is
the Bureau of Mines-type chilled mirror dew point tester. When this instrument is
used properly, measured dew point accuracy is 0.2F.
Each dew point tester incorporates a pressure-tight chamber to contain the gas or
vapor test sample and valves for controlling gas flow. There is a clear plastic window at one end of the chamber to observe the interior and the mirror. The mirror
itself is highly polished stainless steel and is attached at its center to a copper
thermowell. A chiller attached to the thermowell controls expansions of refrigerant and the temperature of the mirror. The greater the flow rate of refrigerant, the
lower the temperature. The mirrors temperature is indicated by a thermometer
and uses a bulb that is located close to the back surface of the mirror. An electronic Resistance Temperature Detection (RTD) is used on some model dew point
testers.
Additional Equipment
Chiller
There are three types of chiller devices available for use with the standard dew
point tester or the ChanScope.
Standard Chiller
Quick Chiller
Liquid Nitrogen Chiller
The Quick Chiller and the Liquid Nitrogen Chiller allow much faster dew point
determination but there is an increased risk of error and temperature overshoot if
components are not allowed to reach thermal equilibrium.
The Quick Chiller can be used for dew point measurements down to -20F
(-30C) with propane, and down to -90F (-70C) with liquid CO2 refrigerant.
The Nitrogen Chiller is a dipstick type and can be used with liquid N2 for
dew points down to and below -200F (-130C). The Nitrogen Chiller can
also be used with dry ice/acetone for dew points down to -100F (-75C).
Certified Thermometers
Certified thermometers that have been tested for accuracy are also available.
They are supplied with calibration traceable to NIST.
Illuminated Magnifier
A four-power, prefocused magnifier is available as an option, and can be attached
to the instrument to illuminate and magnify the mirror. The battery-illuminated
magnifier ensures optimum accuracy of dew point determinations regardless of
field lighting. These illuminated magnifiers are available for the following dew
point tester models: Model 13-075 (large window, 1000 PSIG) and Models 13100 (small window, 3000 PSIG), and 13-110 (small window, 5000 PSIG). The
magnifier is a simple design and is easy to operate. It is held on the dew point
tester using an O-ring.
Overview | 1-3
Glycol Filter
Glycol vapor contained in a gas that has passed through a glycol dehydrator will
condense on the testers mirror complicating observation of water dew point. A
glycol filter assembly can be installed in the tester to remove the glycol vapor
before the contaminant enters the instrument. The filter is especially useful when
the glycol condensation temperature is above the water dew point.
Tripod
The tripod is lightweight, rigid aluminum. It has two section legs that are extendable to 54 1/2 for support in the field.
Propane Cylinder
The propane cylinder is a size 20 DOT-approved cylinder with a delivery control
valve and a 1/4 male NPT thread. The capacity of the cylinder is 3.75 gallons
(14.2 liters) of propane. The cylinder is shipped empty and can be filled at a local
gas supplier.
CO2 Bottle
The CO2 bottle is a size 20 DOT-approved cylinder with a dip tube, delivery control valve, and a 1/4 male NPT thread. The capacity of the cylinder is 5 lb (2.3
kg) of CO2. The cylinder can be purchased empty, or full of welding grade liquid
CO2.
13-
Pressure Range
075
0-1000 PSIG
100
0-3000 PSIG
110
0-5000 PSIG
Model B
Model A2
Model A2
Chiller Device
C - Standard Chiller
Q - Quick Chiller
L - Liquid Nitrogen Chiller
Application
N - Normal Service
R - Off-Shore Service
S - Sour Gas Service
X - Cleaned for O2 Service
Standard Thermometers
025 -30 to 110F
027 -90 to +40F
038 -100 to +10C
060 -35 to +50C
096 -200 to 0F
099 -0 to +220F
Certified Thermometers
146 -90 to +40F
148 -200 to 0F
149 -30 to 110F
152 -100 to +10C
157 -35 to +50C
13-
100-
025
13-
Pressure Range
1175 0-1000 PSIG
1200 0-3000 PSIG
1210 0-5000 PSIG
Chiller Device
C - Standard Chiller
Q - Quick Chiller
L - Liquid Nitrogen Chiller
Application
N - Normal Service
R - Off-Shore Service
S - Sour Gas Service
X - Cleaned for O2 Service
13-
1200-
Overview | 1-5
Technical Support
AMETEK is committed to providing you with the best technical support in the
industry. If you need service or application assistance, please call Chandler at
918-250-7200, or call your local AMETEK/Chandler representative.
If you need to return equipment, you will be asked to provide the following information before obtaining a Return Material Authorization (RMA) number.
Model number
Serial number
Telephone number
NOTE
Before returning material, you must get an RMA number from the
factory.
SPECIFICATIONS
Operating Pressure:
Model 13-075 Standard Tester: 0 to 1000 PSIG (6.9 mPa)
Model 13-100 Standard Tester: 0 to 3000 PSIG (20.7 mPa)
Model 13-110 Standard Tester: 0 to 5000 PSIG (34.6 mPa)
Connections:
Sample inlet and outlet: 1/8 NPT
Refrigerant:
1/8 tube
25 lb (11 kg)
Carrying Case:
Specifications | 2-1
Operating Pressure:
Model 13-1175 ChanScope Tester: 0 to 1000 PSIG (6.9 mPa)
Model 13-1200 ChanScope Tester: 0 to 3000 PSIG (20.7 mPa)
Model 13-1210 ChanScope Tester: 0 to 5000 PSIG (34.6 mPa)
Connections:
Sample inlet and outlet: 1/8 NPT
Refrigerant:
1/8 tube
Power
Battery-powered (NiMH), 100-240 VAC, 50/60 Hz Smart Charger with three
interchangeable blades
25 lb (11 kg)
Carrying Case:
This chapter shows you how to set up your dew point tester. It includes the following:
Safety Requirements
Mechanical Installation
- Equipment Required
- Tubing and Fittings
- Refrigerant Connections
- Glycol Filter Assembly
- Temperature Display Units (ChanScope only)
Start-Up
- Observing the dew point temperature
NOTE
Safety Requirements
Vent the sample gas to a safe area to prevent the possibility of creating an
explosive atmosphere in an enclosed area.
Always wear safety glasses when working with refrigerant and high pressure
gases.
NOTE
No utilities are required for the standard dew point tester. The ChanScope requires 100-265 VAC, 50/60 Hz for the battery charger.
Mechanical Installation
Tools/Equipment Required
Air duster
Teflon tape
Ensure that the temperature of the sampling line and the dew point tester
are above the dew point temperature of the gas in the pipeline. Heat-traced
sample tubing may be required in cold ambient conditions.
Ensure that the sampling line does not contain any condensed liquids such as
water, hydrocarbons, glycol, etc.
Flask
Clamp
Holder
Chain
Clamp
Figure 3-1.
Liquid nitrogen
refrigerant connections.
For lower dew points you must use liquid nitrogen (Figure 3-1).
The glycol filter assembly is attached in-line to the inlet of the dew point tester
(Figure 3-2). Once installed, the filter assembly will not interfere with the testers
attachment to the tripod.
O-Ring
Seal
Glycol Filter
Cartridge
Filter Housing
Cartridge
O-Ring Seal
Install
this end to
dew point
tester
Figure 3-2.
Glycol filter
installation.
Remove the rear cover of the ChanScope enclosure to change the temperature
display units. Selections are degrees Celsius (C) or degrees Fahrenheit (F).
There is a circuit assembly inside the enclosure. Set the miniature switch setting
#1 to ON for Fahrenheit or to OFF for Celsius. Switches 2, 3 and 4 are not used.
Start Up
1. Connect the sample tubing from the sample source to the dew point tester
sample inlet valve.
2. Turn the sample inlet valve until it is fully open.
3. Partially open the sample outlet valve just enough to allow flow through the
tester body but not enough to cause a drop in the sample pressure.
4. Allow the sample to flow through the tester until sufficient time has passed to
allow system equilibrium. This should be approximately five to ten minutes.
5. Connect the refrigerant tubing from the refrigerant cylinder to the refrigerant
inlet connection valve on the chiller device.
6. Slowly open the refrigerant valve to start the flow of refrigerant through
the chiller device. Once the temperature starts to drop, close the refrigerant
valve.
7. Slowly open and close the refrigerant valve so that the temperature of the
sample gas drops slowly.
8. Observe the mirror and thermometer/digital readout. At the same time, make
note of the temperature at which a spot of moisture first appears at the center
of the mirror.
9. Allow the mirror to warm up and make note of the temperature at which the
moisture spot disappears.
10. Repeat Steps 7 and 8 until the temperatures observed are within 2F (1C).
11. Record the average or mean temperature that you observed as the dew point
temperature.
OPERATION
Determining Equilibrium
When the sample gas is introduced at the side of the pressure chamber, it flows
through the glycol filter before entering the tester. The filter material should not
cause a drop in pressure at normal flow rates. However, you must allow sufficient
time for the filter material to attain equilibrium with the sample gas pressure. The
time required to reach equilibrium depends on the concentration of water and
glycol in the gas and the density of the filter material in the cartridge. Equilibrium is determined by running preliminary dew point tests. Two consecutive test
results without glycol masking of the mirror indicate that the tester is operating
normally.
Operation | 4-1
Test Precautions
These precautions must be observed to prevent serious injury and to ensure accurate measurements.
1. Vent the sample gas to a safe area to prevent the possibility of creating an
explosive atmosphere in an enclosed area.
2. Always wear safety goggles when working with refrigerants and high -pressure gases.
3. Do not exceed the pressure rating of the tester.
-
Use extreme caution if you are using the tester on oxygen or other
explosive gases reactive to oil. The entire instrument, including the
sample line, must be thoroughly cleaned. Testers that have been
ordered cleaned for oxygen from the factory will have an oxygen
pressure gauge installed and will be tagged with the date they were
cleaned.
Operation | 4-3
Test Procedure
Follow this procedure in the ordered sequence to ensure accurate readings.
1. Open the sample inlet valve and purge the tester by flowing the test gas
through the sample chamber to expel all the air.
2. Regulate the flow of the test sample by partially closing the outlet valve and
then adjust to ensure that the sample flows continuously at full sample pressure through the chamber.
3. Adjust the outlet valve so that the gas sample can just be detected on your
fingertip. An aspirator bulb or vacuum pump can be used to bring atmospheric or low-pressure samples past the mirror.
NOTE
4. ChanScope
For models with a digital dew point indicator, verify that the decimal point at
the far right side of the display is not blinking.
A blinking decimal point indicates that the tester is in pause or hold mode.
When in this mode, the displayed temperature is frozen and will not
change until you press the HOLD button. The HOLD button is a pushbutton or mementary switch; press HOLD at any time to ENTER or EXIT this
mode.
5. Open the refrigerant valve periodically for a few seconds; allow sufficient
time between valve openings to lower the mirror temperature. You can make
a quick observation to determine an approximate dew point. However, for the
final determination, you should lower the temperature at a rate of no more
than one or two degrees per minute.
6. Observe the mirror and thermometer/digital readout at the same time and
note the temperature at which a spot of moisture first appears at the center of
the mirror.
7. Allow the mirror to warm up and note the temperature at which the moisture
spot disappears.
8. Repeat Steps 5 and 6 until the temperatures observed are within 2F (1C).
9. Record the mean or average temperature observed as the dew point.
If a dew point of 9F is obtained, you can be sure that the original determination at 1000 PSIG was correct since the water content remained the same (6.3
lb/MMscf) with the change is pressure and dew point. The water dew point is
lowered as the pressure is lowered, but the water content of the gas remains
the same. Had the water content been different at the second pressure and
dew point, then one of the dew points might have been a hydrocarbon, glycol
or alcohol dew point.
Operation | 4-5
The hydrocarbon wets the mirror and the film expands in an iridescent ring to
cover the mirrors surface, or appears as a waterfall of droplets that appear to
be pouring from the mirrors center.
You can also verify the substance on the mirror by determining the
dew point at a lower pressure.
NOTE
NOTE
For detailed information refer to GPA Publication 2140-80 Liquid Petroleum Gas Specification and Test Methods, 1980 revision,
Tulsa, Oklahoma.
Operation | 4-7
Maintenance
Tools Required
Standard Dew Point Tester
Phillips screwdriver
1/8 hex wrench
Use a cotton ball, cotton swab, or soft cloth to clean the window and mirror
surfaces. Be careful to avoid scratching the surfaces.
To clean the high-pressure window, remove the screw and pull the unit out of
the window nut. Be careful to avoid scratching the window.
NOTE
Plug the battery charger into the wall receptacle from 100-265 VAC, 50/60
Hz using the appropriate supplied blade adapter, and plug the other end into
the jack on the ChanScope.
NOTE
Charger Functionality
Charging begins when the battery pack is connected to the charger. The LED
is orange before the fast charge starts, and then the LED changes to red.
When the cells are fully charged and the voltage drops (because of the -dV
signal from the cells), the charger will go into a top-off charge mode before
switching to trickle charge mode.
During the top-off charge session, the LED will be green with short, intermittent orange flashes. When the top-off charge is complete, the charger will go
into trickle charge mode, changing the LED to green. The charge current
is now reduced to a safe level, permitting the charger to stay connected to
a NiCd battery pack without damaging the cells. Depending on the specific
manufacturer of NiMH cells, trickle charging in excess of 24 hours may not
be permitted.
If the safety timer disconnects before -dV, the top-off charge will not be
engaged. The charger will then go directly to trickle charge mode and the
LED will be green. If the battery voltage is far below normal, the charger will
stop the fast-charge current and switch to trickle charge mode. The LED
will then indicate error by flickering green and red.
If the mains is turned off, the charger will reset and start a new charge cycle
when the mains is turned on again.
NOTE
You can operate the tester during charging. The tester cannot be
operated without a battery.
Use a ChanScope II battery charger only (PN 13-2019). Other chargers will cause internal damage.
NOTE
Recycle the battery pack at an appropriate facility for NiMH batteries. Visit
www.rbrc.org or call 1-800-8-BATTERY for locations.
Loosen the gland nut on the RTD and gently pull the RTD out. Replace with
the appropriate thermometer.
Turn the power switch on to illuminate the mirror and proceed as usual.
NOTE
Calibration
Except for the thermometer, calibration is not possible or necessary on the Standard Dew Point Testers (Models 13-075, 13-100, and 13-110. The thermometer
can be removed and checked for accuracy at any reputable calibration service
equipped to do so.
Calibration of the ChanScope RTD and the electronics is done at the factory.
Return the tester after obtaining an RMA.
PROBLEM
CAUSE
SOLUTION
Intermittent display
RTD, battery
connections, or switch
Display is segmented, or no
display.
Calibration.
Device is in HOLD
mode.
CAUSE
SOLUTION
Possible glycol
contamination.
Temperature of sample
line and tester below
the temperature of
sample gas. Not using
a probe to sample gas,
slugs of water entering
the sample line from
the pipe wall.
Moisture equilibrium
in glycol filter not yet
reached.
REPLACEMENT PARTS
MODELS 13-075, 13-100, 13-110, 13-210, 13-1175, 13-1200 and 13-1210
PART #
13-0005
13-0006
13-0007
13-0009
13-0018
13-0020
13-0021
13-0022
13-0024
13-0025
13-0027
13-0038
13-0046
13-0047
13-0048
13-0052
13-0053
13-0054
13-0060
13-0064
DESCRIPTION
Window, Models 13-100 & 13-200 and 13-1200
Window retaining nut, Model 13-100 & 13-110
Mirror assembly, regular chiller
Mirror insulator, Models 13-153 & 13-154 kits
Chiller body, regular
Thermoguard assembly
Lock nut, thermoguard
Bulb shield, thermoguard
Shield, thermoguard
Thermometer, range -30 to +110F
Thermometer, range -90 to +40F
Thermometer, range -100 to +10C
Retaining nut, quick chiller
Block, with tube, quick chiller
Quick chiller assembly
Glass window assembly, Models 13-100, 13-110, 13-200, 13-210, 13-1200,
13-1210 (sour gas and 5,000 PSIG)
Cooling rod, nitrogen chiller
Glycol filter assembly
Thermometer, range -35 to +50C
Window, glass, Models 13-100, 13-110, 13-200, 13-210, 13-1200, 13-1210
(sour gas and 5,000 PSIG)
13-0065
13-0068
13-0071
13-0072
13-0077
13-0078
13-0079
13-0084
13-0085
13-0086
13-0101M
13-0102
13-0111
13-0116
13-0117
13-0119
13-0122
PART #
DESCRIPTION
13-0124
13-0146
13-0147
13-0148
13-0149
P-0018
P-0020
P-0063
P-0067
P-0070
Clamp, magnifier
Clamp holder, nitrogen chiller
O-ring, gauge adapter, quick chiller
O-ring, window mirror (included in 13-153 & 13-154 kits)
O-ring, 13-0303 magnifier (included in 13-154 kit)
P-0071
P-0182
P-0301
P-0535
P-0536
P-0583
P-0591
P-0594
P-0711
P-0720
P-0986
P-0987
P-0988
P-1037
P-1360
P-1367
P-1392
P-1818
P-2300
P-2346
PART #
13-0204
13-0205
13-0210
13-0214
13-0215
13-0217
13-0219*
13-0220
13-0227
13-0230
13-0234
13-0236*
13-0241
13-0250
13-0253
13-0254
13-1238
187-20535
C08846
P-0221
P-1044
P-2770
P-2820
P-2822***
P-2826
P-2828
P-2830
P-3113
P-3115
P-3116
*
DESCRIPTION
Window nut, Model 13-175
Window nut, Model 13-0200, 13-0210
L-bracket, battery
Optical assembly
Beam splitter assembly
Prism assembly
Logic board assembly
LED board assembly
Eyepiece assembly
Gland nut
Switch
RTD assembly with strain relief (spring-loaded)
Battery holder, two cell
Dewpoint indicator (ChanScope) assembly
Wiring harness assembly
Cable assembly
Black case with foam inserts
Adapter, 220 VAC to 115 VAC, 50/60 Hz, use with P-3116 charger
Kit, 230 VAC plug adapter
O-ring, RTD gland nut
Lamp, screw base
Lamp, bayonet base
O-ring, tube support
Eye Cup
Screw, 2-56 x .250, rnd Head, slot
Screw, 6-32 x .250, rnd head, Phillips
Screw, 6-32 x 1.0, skt head, SST
Battery, C cell, NiCad recharge
Holder, (1) C cell, #2173
Battery charger, NiCad
*** OBSOLETE. Replaced by C11468 Eye Cup and 13-0227 Eyepiece Assembly.
Consult the factory for price and availability.
PART #
13-0204
13-0205
13-0215
13-0227
13-0236*
13-0307*
13-1214
13-1217
13-1238
13-1250
13-1252
13-1274
13-1276
13-1277
13-2019**
C09867
C11003
C11468
C11921
C11922
C11923
C12024
P-2820
P-2821
P-2826
P-2828
H-6033
P-2836
*
DESCRIPTION
Window nut, Model 13-1175
Window nut, Model 13-1200, 13-1210
Beam splitter assembly
Eyepiece assembly
RTD assembly with strain relief (spring-loaded)
Circuit assembly
Optical assembly
Prism assembly
Case with foam inserts
ChanScope II assembly
ChanScope LED/lamp wiring harness assembly
Battery pack, rechargeable
Switch/battery/harness assembly
Assembly, battery charger connector wiring harness
Battery charger, NiMH, 1600 mAh, 100-265 VAC, 50/60 Hz, with 3 blades
Pushbutton switch
Charger connection cap assembly
Eye cup
Adapter, charger, US blade
Adapter, charger, Euro blade
Adapter, charger, UK blade
Charger, Z22 PDA, 110-240V
O-ring, tube support
Rocker switch
Screw, 2-56 x .250 rnd head, slotted
Screw, 6-32 x .250, rnd head, Phillips
Screw, 6-32 x 1.125, skt head, SST (4 required for eyepiece)
Strain relief, RTD, cable
DRAWINGS
Drawings | 6-1
10
Install
this end to
dew point
tester
11
ITEM
1
2
PART #
36-0000-10
C03322
Figure 6-2.
Glycol filter assembly.
DWG 13-0054.
4
5
6
7
8
9
10
11
P-2577
C08435
P-3178
C04364
C08434
C08435
DESCRIPTION
1/8 SST Tubing
Tube Connector
Valve - supplied with
dew point tester
Bushing
O-Ring Seal
Filter Cartridge
Elbow
Adapter
Filter Housing
O-Ring Seal
Cartridge
Dew point tester and tripod are not included in 13-0072 immersion
cooling rod assembly.
NOTE
6
7
ITEM
1
2
3
Figure 6-3.
Immersion cooling rod
assembly.
DWG 13-0072.
4
5
6
7
PART #
DESCRIPTION
P-0020
Clamp Holder
13-0071
Extension
Model 13-0100 or Model 13-0075 with
13-0007 Regular Mirror
13-0053
Immersion Cooling Rod
P-0986
Flask
P-0987
Chain Clamp
P-0535
Tripod
Drawings | 6-3
14 3/4 approx
13
4
14
12
15
16
11
10
Thermometer
specify temperature range
17
10 1/2
approx
18
1/8 NPT
19
1/8 NPT
Gas Outlet
23
ITEM
Figure 6-4.
Dew Point Tester,
Model 13-0075 with
quick chiller assembly.
DWG 13-0075.
PART #
DESCRIPTION
C03322
Connector
C08161
Valve
P-1190
Elbow
Refrigerant Outlet
P-0301
Chiller valve
13-0086
Adapter, Tripod
C08161
Valve
13-0079
Window
13-0077
Nut
10
13-0085
Body
11
P-0071
Extension
12
13-0084
Adapter
13
P-0591
Pressure Gauge
14
P-0063
15
P-0067
16
13-0078
Nut
17
13-0048
18
13-0046
Retaining Nut
19
13-0021
20
13-0020
Thermoguard Assembly
21
13-0024
Shield
22
13-0023
Adjusting Knob
23
21
20
22
Figure 6-5.
Chillers, regular and
quick.
DWG 13-0090.
Drawings | 6-5
Figure 6-6.
Chart.
Figure 6-7.
Chart.
Drawings | 6-7
Figure 6-8.
Chart.
Figure 6-9.
Chart.
Drawings | 6-9
Figure 6-10.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 1.
Figure 6-11.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 2.
Drawings | 6-11
Figure 6-12.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 3.
Figure 6-13.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 4.
Drawings | 6-13
Figure 6-14.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 5.
Figure 6-15.
Bureau of Mines
dew point tester with
regular chiller.
DWG 13-0100.
Drawings | 6-15
Figure 6-16.
Dew point to moisture
content conversion
chart.
Figure 6-17.
Installation of dew
point tester.
DWG 13-0150.
Drawings | 6-17
Figure 6-17.
Dew point indicator
assembly.
DWG 13-2016.
13-0116
13-0122
13-0111
13-0119
13-0117
P-0070
13-0305
H-8014
PART #
Illuminator Assembly
O-Ring
Lens Retainer
Body Assembly
1/4-20 x .25 LG, Allen
HD set screw
Lens
Viewer
Window Eyeguard
DESCRIPTION
2. Lubricate the magnifier O-ring and push the magnifier inside of the round
window nut.
1. Remove 13-0102 observation mirror from the dew point tester body.
INSTALLATION INSTRUCTIONS
ITEM
Figure 6-18.
Illuminated magnifier
assembly, Model 13-075.
DWG 13-0303.
Drawings | 6-19
INSTALLATION INSTRUCTIONS
1. Remove 13-0102 observation mirror from the dew point tester body.
2. Lubricate the magnifier O-ring and push the magnifier over the outside of the
round window nut.
3. Screw 13-0305 illuminator assembly into 13-0124 lens retainer.
4. Rotate 13-0305 illuminator counterclockwise to turn it on.
ITEM
Figure 6-19.
Illuminated magnifier
assembly, Models 13100 and 13-110.
DWG 13-0304.
1
2
3
4
5
PART #
13-0305
P-0071
13-0124
H-8014
13-0122
6
7
8
9
13-0116
13-0111
13-0119
DESCRIPTION
Illuminator
O-Ring
Lens Retainer
#8-32 x .5 LG, BHP
Viewer
1/4-20 x .25 LG, Allen
HD set screw
Window Eyeguard
Body Assembly
Lens
ITEM
PART #
DESCRIPTION
C09490
13-0301
13-0302
C09549
Batteries, AAA
C09550
Figure 6-20.
Illuminator Assembly.
DWG 13-0305.
Drawings | 6-21
Figure 6-21.
ChanScope II.
DWG 13-2017.
Not Shown: