Sie sind auf Seite 1von 64

Chandler

ChanScope and Dew Point Tester

User Manual

Process Instruments
PN 13-2270 REV L

150 Freeport Road


Pittsburgh, PA 15238

Offices
USA - HEADQUARTERS
AMETEK Process Instruments
150 Freeport Road
Pittsburgh, PA 15238, USA
Phone: 412-828-9040
Fax: 412-826-0399
USA - Delaware
AMETEK Process Instruments
455 Corporate Boulevard
Newark, Delaware 19702, USA
Phone: 302-456-4400 (Main)

800-537-6044 (Service)

800-222-6789 (Ordering)
Fax: 302-456-4444
USA - Texas
AMETEK Process Instruments
9750 Whithorn Drive
Houston, Texas 77095, USA
Phone: 281-463-2820
Fax: 281-463-2701
USA - Oklahoma
Chandler Engineering
2001 North Indianwood Avenue
Broken Arrow, OK 74012, USA
Phone: 918-250-7200
Fax: 918-459-0165
CANADA
AMETEK Process Instruments
2876 Sunridge Way N.E.
Calgary, Alberta
Canada T1Y 7H9
Phone: 403-235-8400
Fax: 403-248-3550

GERMANY
AMETEK GmbH
Postfach 2165
D-40644 Meerbusch

OR
Rudolf-Diesel Strasse 16
D-40670 Meerbusch
Germany
Phone: 49-21-59-9136-0
Fax: 49-21-59-9136-39
SINGAPORE
AMETEK Singapore PVT. Ltd.
10 Ang Mo Kio Street 65
#05-12 Techpoint
Singapore 569059
Republic of Singapore
Phone: 65-484-2388
Fax: 65-481-6588
FRANCE
AMETEK - APIF
Rond Point de lpine des champs
Buroplus Bat D
78990 Elancourt, France
Phone: 33 1 30 68 89 20
Fax: 33 1 30 38 89 29
CHINA
AMETEK Process Instruments
Room 406, Metro Tower
No. 30 Tian Yao Qiao Road
Shanghai, 200030
Phone: 8621 6426 7049
Fax: 8621 6426 7054

2005 AMETEK
This manual is a guide for the use of the AMETEK ChanScope and Dew Point Tester models. Data herein has been verified and validated and is
believed adequate for the intended use of this software. If the software or procedures are used for purposes over and above the capabilities specified
herein, confirmation of their validity and suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no obligation or
liability. This publication is not a license to operate under, or a recommendation to infringe upon, any process patents.

ii | AMETEK ChanScope and Dew Point Tester

Table of Contents
CHAPTER 1 Overview
Dew Point Tester.................................................................................................. 1-1

Description.......................................................................................... 1-1
Bureau of Mines-Type Dew Point Tester (Standard).................... 1-1
ChanScope.................................................................................... 1-1

Theory of Operation............................................................................................ 1-2


Additional Equipment.......................................................................................... 1-3

Chiller................................................................................................. 1-3
Illuminated Magnifier......................................................................... 1-3
Tripod.................................................................................................. 1-3
Propane Cylinder................................................................................ 1-3
CO2 Bottle........................................................................................... 1-3

Part Numbering Codes......................................................................................... 1-4


Technical Support................................................................................................ 1-5
CHAPTER 2 Specifications
Standard Bureau of Mines-type Dew Point Tester.............................................. 2-1

Accuracy............................................................................................. 2-1
Dew Point Temperature Range........................................................... 2-1
Operating Pressure.............................................................................. 2-1
Connections......................................................................................... 2-1
Weight and Dimensions...................................................................... 2-1

ChanScope Digital Dew Point Tester.................................................................. 2-2

Accuracy............................................................................................. 2-2
Dew Point Temperature Range........................................................... 2-2
Operating Pressure.............................................................................. 2-2
Connections......................................................................................... 2-2
Power.................................................................................................. 2-2
Weight and Dimensions...................................................................... 2-2

CHAPTER 3 Installation and Set Up


Inspect Shipping Contents................................................................................... 3-1
Safety Requirements............................................................................................ 3-2
Mechanical Installation........................................................................................ 3-3

Tools/Equipment Required................................................................. 3-3


Tubing and Fittings............................................................................. 3-3
Installing the Dew Point Tester........................................................... 3-4
Connecting the Refrigerant................................................................. 3-4
Installing the Glycol Filter Assembly................................................. 3-5
Selecting Temperature Display (ChanScope only)............................. 3-5

Start Up................................................................................................................ 3-6

| iii

CHAPTER 4 Operation
Using Glycol Filters............................................................................................. 4-1

Determining Equilibrium.................................................................... 4-1

Test Precautions................................................................................................... 4-2


Test Procedure..................................................................................................... 4-4
Confirmation of Dew Point.................................................................................. 4-5

Interpretation of Dew Point Temperature........................................... 4-5

Differentiating between Water and Hydrocarbon Dew Points............................ 4-6


Special Test Procedure for Propane..................................................................... 4-7
CHAPTER 5 Maintenance and Troubleshooting
Maintenance......................................................................................................... 5-1

Tools Required.................................................................................... 5-1


Replacing the Glycol Filter................................................................. 5-2
Cleaning the Window and Mirror....................................................... 5-2
Recharging Batteries (ChanScope Dew Point Tester)........................ 5-2
Bypassing the RTD............................................................................. 5-3
Calibration........................................................................................... 5-4
Selecting Temperature Display (ChanScope only)............................. 5-4

Troubleshooting Guide........................................................................................ 5-5


Replacement Parts............................................................................................... 5-9
ChanScope Digital Dew Point Tester................................................................ 5-11

Models 13-175, 13-200, 13-210 (10,000 Series SN)........................ 5-11

ChanScope II Digital Dew Point Tester............................................................. 5-12

Models 13-1175, 13-1200, 13-1210 (20,000 Series SN).................. 5-12

CHAPTER 6 Drawings
DWG 13-0052 - Glass Window Assy (sour gas)................................................. 6-1
DWG 13-0054 - Glycol Filter Assy..................................................................... 6-2
DWG 13-0072 - Immersion Cooling Rod Assy.................................................. 6-3
DWG 13-0075 - Model 13-0075 Dew Point Tester............................................. 6-4
DWG 13-0090 - Chillers, Regular and Quick..................................................... 6-5
DWG 13-0095 - Saturated Moisture Content Air or Nitrogen (F)..................... 6-6
DWG 13-0095 - Saturated Moisture Content Air or Nitrogen (C).................... 6-7
DWG 13-0097 - Saturated Moisture Content Natural Gas (F).......................... 6-8
DWG 13-0097 - Saturated Moisture Content Natural Gas (C).......................... 6-9
Table - Equilibrium Water Vapor Contents........................................................ 6-10
DWG 13-0100 - Bureau of Mines Dew Point Tester......................................... 6-15
DWG 13-0125 - Conversion Chart, Dew Point to Moisture Content................ 6-16
DWG 13-0150 - Installation, Dew Point Tester................................................ 6-17
DWG 13-0250 - Dew Point Indicator Assy....................................................... 6-18
DWG 13-0303 - Illuminated Magnifier Assy (Model 13-075).......................... 6-19
DWG 13-0304 - Illuminated Magnifier (Models 13-100, 13-110).................... 6-20
DWG 13-0305 - Illuminator Assy..................................................................... 6-21
DWG 13-1250 - ChanScope II.......................................................................... 6-22

iv | AMETEK ChanScope and Dew Point Tester

Safety Notes
warnings, cautions, and notes contained in this manual emphasize critical instructions as follows:
An operating procedure which, if not strictly observed, may result in personal injury or environmental contamination.

An operating procedure which, if not strictly observed, may result in damage to the equipment.

Important information that should not be overlooked.

NOTE

Electrical Safety
Always shut down power source(s) before performing maintenance or troubleshooting. Only a qualified
electrician should make electrical connections and ground checks.
Any use of the equipment in a manner not specified by the manufacturer may impair the safety protection
originally provided by the equipment.

Grounding
Instrument grounding is mandatory. Performance specifications and safety protection are void if instrument is operated from an improperly grounded power source.
Verify ground continuity of all equipment before applying power.

| v

Warning Labels
These symbols may appear on the instrument in order to alert you of existing conditions.
Protective Conductor Terminal
(BORNIER DE LECRAN DE PROTECTION)
Schutzerde
Caution - Risk of electric shock
(ATTENTION-RISQUE DE DCHARGE LECTRIQUE)
Achtung - Hochspannung Lebensgefahr
Caution - (Refer to accompanying documents)
(ATTENTION-SE RFERER AUX DOCUMENTS JOINTS)
Achtung (Beachten Sie beiliegende Dokumente)
CAUTION - Hot Surface
(ATTENTION-SURFACE CHAUDE)
Achtung - Heie Oberflche

Environmental Information (WEEE)


This AMETEK product contains materials that can be reclaimed and recycled. In some cases the product
may contain materials known to be hazardous to the environment or human health. In order to prevent
the release of harmful substances into the environment and to conserve our natural resources, AMETEK
recommends that you arrange to recycle this product when it reaches its end of life.
Waste Electrical and Electronic Equipment (WEEE) should never be disposed of in a municipal waste
system (residential trash). The Wheelie Bin marking on this product is a reminder to dispose of the product properly after it has completed its useful life and been removed from service. Metals, plastics and
other components are recyclable and you can do your part by one of the following these steps:

When the equipment is ready to be disposed of, take it to your local or regional waste collection administration for recycling.

In some cases, your end-of-life product may be traded in for credit towards
the purchase of new AMETEK instruments. Contact your dealer to see if this
program is available in your area.

If you need further assistance in recycling your AMETEK product, contact


our office listed in the front of the instruction manual.

vi | AMETEK ChanScope and Dew Point Tester

WARRANTY AND CLAIMS


We warrant that any equipment of our own manufacture or manufactured for us pursuant to our specifications which
shall not be, at the time of shipment thereof by or for us, free from defects in material or workmanship under normal
use and service will be repaired or replaced (at our option) by us free of charge, provided that written notice of such
defect is received by us within twelve (12) months from date of shipment of portable analyzers or within eighteen (18)
months from date of shipment or twelve (12) months from date of installation of permanent equipment, whichever
period is shorter. All equipment requiring repair or replacement under the warranty shall be returned to us at our factory,
or at such other location as we may designate, transportation prepaid. Such returned equipment shall be examined by
us and if it is found to be defective as a result of defective materials or workmanship, it shall be repaired or replaced
as aforesaid. Our obligation does not include the cost of furnishing any labor in connection with the installation of
such repaired or replaced equipment or parts thereof, nor does it include the responsibility or cost of transportation. In
addition, instead of repairing or replacing the equipment returned to us as aforesaid, we may, at our option, take back
the defective equipment, and refund in full settlement the purchase price thereof paid by Buyer.
Process photometric analyzers, process moisture analyzers, and sampling systems are warranted to perform the
intended measurement, only in the event that the customer has supplied, and AMETEK has accepted, valid sample
stream composition data, process conditions, and electrical area classification prior to order acknowledgment. The
photometric light sources are warranted for ninety (90) days from date of shipment. Resale items warranty is limited to
the transferable portion of the original equipment manufacturers warranty to AMETEK. If you are returning equipment
from outside the United States, a statement should appear on the documentation accompanying the equipment being
returned declaring that the goods being returned for repair are American goods, the name of the firm who purchased
the goods, and the shipment date.
The warranty shall not apply to any equipment (or part thereof) which has been tampered with or altered after leaving our control or which has been replaced by anyone except us, or which has been subject to misuse, neglect, abuse
or improper use. Misuse or abuse of the equipment, or any part thereof, shall be construed to include, but shall not
be limited to, damage by negligence, accident, fire or force of the elements. Improper use or misapplications shall be
construed to include improper or inadequate protection against shock, vibration, high or low temperature, overpressure,
excess voltage and the like, or operating the equipment with or in a corrosive, explosive or combustible medium, unless the equipment is specifically designed for such service, or exposure to any other service or environment of greater
severity than that for which the equipment was designed.
The warranty does not apply to used or secondhand equipment nor extend to anyone other than the original purchaser
from us.
THIS WARRANTY IS GIVEN AND ACCEPTED IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESS
OR IMPLIED, INCLUDING WITHOUT LIMITATION AND WARRANTIES OF FITNESS OR OF MERCHANTABILITY OTHER THAN AS EXPRESSLY SET FORTH HEREIN, AND OF ALL OTHER OBLIGATIONS OR LIABILITIES ON OUR PART. IN NO EVENT SHALL WE BE LIABLE UNDER THIS WARRANTY OR ANY OTHER
PROVISION OF THIS AGREEMENT FOR ANY ANTICIPATED OR LOST PROFITS, INCIDENTAL DAMAGES,
CONSEQUENTIAL DAMAGES, TIME CHANGES OR ANY OTHER LOSSES INCURRED BY THE ORIGINAL
PURCHASER OR ANY THIRD PARTY IN CONNECTION WITH THE PURCHASE, INSTALLATION, REPAIR OR
OPERATION OF EQUIPMENT, OR ANY PART THEREOF COVERED BY THIS WARRANTY OR OTHERWISE.
WE MAKE NO WARRANTY, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF FITNESS OR OF MERCHANTABILITY, AS TO ANY OTHER MANUFACTURERS EQUIPMENT,
WHETHER SOLD SEPARATELY OR IN CONJUNCTION WITH EQUIPMENT OF OUR MANUFACTURE. WE
DO NOT AUTHORIZE ANY REPRESENTATIVE OR OTHER PERSON TO ASSUME FOR US ANY LIABILITY
IN CONNECTION WITH EQUIPMENT, OR ANY PART THEREOF, COVERED BY THIS WARRANTY.

| vii

This page intentionally left blank.

viii | AMETEK ChanScope and Dew Point Tester

OVERVIEW

Dew Point Tester


Description
Bureau of Mines-Type Dew Point Tester (Standard)
When a gas-water vapor is cooled with no liquid water present, the absolute humidity of the vapor content remains constant but the relative humidity increases
until it reaches 100% saturation, after which moisture begins to condense. The
temperature at which condensation is first detected is known as the dew point.
The Bureau of Mines-type dew point tester duplicates the conditions required
by the definition of dew point. However, the measurement method is modified
so that it is suitable for measuring dew point at high pressures. This is done by
enclosing a mirror upon which moisture or hydrocarbons condense (when chilled
by a refrigerant) in a pressure chamber. A window in the chamber allows viewing
of moisture condensation.

ChanScope
The digital ChanScope provides state-of-the-art temperature measurement of the
dew point of pressurized gas. The ChanScope retains all of the proven reliability and quality of the original Bureau of Mines-type dew point tester and gives
instantaneous feedback using an LEDreadout during a test.
Both the standard (Bureau of Mines type) dew point tester and the ChanScope
include a sample pressure gauge and tripod socket for ease of use in the field.

NOTE

All AMETEK Chandler dew point testers meet ASTM D1142-GPA


2140 standards.

Overview | 1-1

Theory of Operation
Water contamination in gases can cause corrosion and other equipment problems,
especially in regulation and metering stations. The determination of hydrocarbon
contamination in pipeline gas is necessary to meet hydrocarbon dew point specifications in natural gas transportation contracts. Gases most frequently tested for
water content include natural gas, sour gas, heat treating furnace gas, and industrial gases such as nitrogen, oxygen, hydrogen, and other bottled gases. The most
accurate and widely used instrument available for determining water content is
the Bureau of Mines-type chilled mirror dew point tester. When this instrument is
used properly, measured dew point accuracy is 0.2F.
Each dew point tester incorporates a pressure-tight chamber to contain the gas or
vapor test sample and valves for controlling gas flow. There is a clear plastic window at one end of the chamber to observe the interior and the mirror. The mirror
itself is highly polished stainless steel and is attached at its center to a copper
thermowell. A chiller attached to the thermowell controls expansions of refrigerant and the temperature of the mirror. The greater the flow rate of refrigerant, the
lower the temperature. The mirrors temperature is indicated by a thermometer
and uses a bulb that is located close to the back surface of the mirror. An electronic Resistance Temperature Detection (RTD) is used on some model dew point
testers.

1-2 | AMETEK ChanScope and Dew Point Tester

Additional Equipment
Chiller
There are three types of chiller devices available for use with the standard dew
point tester or the ChanScope.

Standard Chiller
Quick Chiller
Liquid Nitrogen Chiller

The Quick Chiller and the Liquid Nitrogen Chiller allow much faster dew point
determination but there is an increased risk of error and temperature overshoot if
components are not allowed to reach thermal equilibrium.

The Quick Chiller can be used for dew point measurements down to -20F
(-30C) with propane, and down to -90F (-70C) with liquid CO2 refrigerant.

The Nitrogen Chiller is a dipstick type and can be used with liquid N2 for
dew points down to and below -200F (-130C). The Nitrogen Chiller can
also be used with dry ice/acetone for dew points down to -100F (-75C).

Certified Thermometers
Certified thermometers that have been tested for accuracy are also available.
They are supplied with calibration traceable to NIST.

Illuminated Magnifier
A four-power, prefocused magnifier is available as an option, and can be attached
to the instrument to illuminate and magnify the mirror. The battery-illuminated
magnifier ensures optimum accuracy of dew point determinations regardless of
field lighting. These illuminated magnifiers are available for the following dew
point tester models: Model 13-075 (large window, 1000 PSIG) and Models 13100 (small window, 3000 PSIG), and 13-110 (small window, 5000 PSIG). The
magnifier is a simple design and is easy to operate. It is held on the dew point
tester using an O-ring.

Overview | 1-3

Glycol Filter
Glycol vapor contained in a gas that has passed through a glycol dehydrator will
condense on the testers mirror complicating observation of water dew point. A
glycol filter assembly can be installed in the tester to remove the glycol vapor
before the contaminant enters the instrument. The filter is especially useful when
the glycol condensation temperature is above the water dew point.

Tripod
The tripod is lightweight, rigid aluminum. It has two section legs that are extendable to 54 1/2 for support in the field.

Propane Cylinder
The propane cylinder is a size 20 DOT-approved cylinder with a delivery control
valve and a 1/4 male NPT thread. The capacity of the cylinder is 3.75 gallons
(14.2 liters) of propane. The cylinder is shipped empty and can be filled at a local
gas supplier.

CO2 Bottle
The CO2 bottle is a size 20 DOT-approved cylinder with a dip tube, delivery control valve, and a 1/4 male NPT thread. The capacity of the cylinder is 5 lb (2.3
kg) of CO2. The cylinder can be purchased empty, or full of welding grade liquid
CO2.

1-4 | AMETEK ChanScope and Dew Point Tester

Part Numbering Codes


Part Numbering Code for Standard Dew Point Tester

13-

Pressure Range
075
0-1000 PSIG
100
0-3000 PSIG
110
0-5000 PSIG

Model B
Model A2
Model A2

Chiller Device
C - Standard Chiller
Q - Quick Chiller
L - Liquid Nitrogen Chiller

Application
N - Normal Service
R - Off-Shore Service
S - Sour Gas Service
X - Cleaned for O2 Service

Standard Thermometers
025 -30 to 110F
027 -90 to +40F
038 -100 to +10C
060 -35 to +50C
096 -200 to 0F
099 -0 to +220F
Certified Thermometers
146 -90 to +40F
148 -200 to 0F
149 -30 to 110F
152 -100 to +10C
157 -35 to +50C

13-

100-

025

13-100-C-N-025: Typical Part Number

Part Numbering Code for Digital ChanScope Dew Point Tester

13-

Pressure Range
1175 0-1000 PSIG
1200 0-3000 PSIG
1210 0-5000 PSIG

All dew point testers come with


spare O-rings, ten feet of stainless
steel sample tubing with stainless
fittings, and ten feet of stainless
tubing with stainless fittings for
refrigerant gas, carrying case, and
manual.

Chiller Device
C - Standard Chiller
Q - Quick Chiller
L - Liquid Nitrogen Chiller

Application
N - Normal Service
R - Off-Shore Service
S - Sour Gas Service
X - Cleaned for O2 Service

Battery Charger Power


1 - 115 VAC 50/60 Hz
2 - 230 VAC 50/60 Hz

13-

1200-

13-1200-C-N-1: Typical Part Number

Overview | 1-5

Technical Support
AMETEK is committed to providing you with the best technical support in the
industry. If you need service or application assistance, please call Chandler at
918-250-7200, or call your local AMETEK/Chandler representative.
If you need to return equipment, you will be asked to provide the following information before obtaining a Return Material Authorization (RMA) number.

Billing and shipping address

Model number

Serial number

Purchase order number

Telephone number

NOTE

Before returning material, you must get an RMA number from the
factory.

1-6 | AMETEK ChanScope and Dew Point Tester

SPECIFICATIONS

Standard Bureau of Mines-type Dew Point Tester


Accuracy:
0.5F ( 0.25C)

Dew Point Temperature Range:


Dependent on which thermometer is chosen.

Operating Pressure:
Model 13-075 Standard Tester: 0 to 1000 PSIG (6.9 mPa)
Model 13-100 Standard Tester: 0 to 3000 PSIG (20.7 mPa)
Model 13-110 Standard Tester: 0 to 5000 PSIG (34.6 mPa)

Connections:
Sample inlet and outlet: 1/8 NPT
Refrigerant:

1/8 tube

Weight and Dimensions


Weight:

25 lb (11 kg)

Carrying Case:

14 x 18 x 7 (35.5 cm x 20 cm x 17.8 cm)

Specifications | 2-1

ChanScope Digital Dew Point Tester


Accuracy:
0.2C ( 0.4F) at 40C (104F) down to -90C (-130F)

Dew Point Temperature Range:


-129C (-200F) to ambient

Ambient Operating Temperature Range:


-20C (-4F) to 40C (104F)

Operating Pressure:
Model 13-1175 ChanScope Tester: 0 to 1000 PSIG (6.9 mPa)
Model 13-1200 ChanScope Tester: 0 to 3000 PSIG (20.7 mPa)
Model 13-1210 ChanScope Tester: 0 to 5000 PSIG (34.6 mPa)

Connections:
Sample inlet and outlet: 1/8 NPT
Refrigerant:

1/8 tube

Power
Battery-powered (NiMH), 100-240 VAC, 50/60 Hz Smart Charger with three
interchangeable blades

Weight and Dimensions


Weight:

25 lb (11 kg)

Carrying Case:

14 x 18 x 7 (35.5 cm x 20 cm x 17.8 cm)

2-2 | AMETEK ChanScope and Dew Point Tester

INSTALLATION AND SET UP

This chapter shows you how to set up your dew point tester. It includes the following:

Inspecting Shipping Contents

Safety Requirements

Mechanical Installation
- Equipment Required
- Tubing and Fittings
- Refrigerant Connections
- Glycol Filter Assembly
- Temperature Display Units (ChanScope only)

Start-Up
- Observing the dew point temperature

Inspect Shipping Contents


The dew point tester and its accessories were inspected before being shipped
from the factory. Remove any packing material from the units. Check for damage and verify all parts on the packing list were shipped. If any damage is found,
notify AMETEK as soon as possible and notify the shipper.

All equipment is calibrated and tested prior to shipment.

NOTE

Installation and Set Up | 3-1

Safety Requirements

Vent the sample gas to a safe area to prevent the possibility of creating an
explosive atmosphere in an enclosed area.

Always wear safety glasses when working with refrigerant and high pressure
gases.

Do not exceed the pressure rating of the instrument. Refer to Chapter 2


Specifications for instrument operating pressures.
If operating pressure is exceeded, component blow out could
result in serious injury to the operator.

Clean the instrument prior to using it with oxygen.

For models equipped with ChanScope digital dew point indicators:

Do not expose the battery or internal electronics to temperatures higher than


the range -20C to 70C (-4F to 148F). Battery packs should be recycled at
an appropriate facility for Ni-MH batteries. Use only Chandler part #13-1274
batteries as replacements. Using other battery packs will cause damage to the
electronics and void your warranty.
Use extreme caution if you are using the instrument with oxygen or
any other explosive gas reactive to oil. The entire instrument and all
sample tubing must be thoroughly cleaned.

NOTE

No utilities are required for the standard dew point tester. The ChanScope requires 100-265 VAC, 50/60 Hz for the battery charger.

3-2 | AMETEK ChanScope and Dew Point Tester

Mechanical Installation
Tools/Equipment Required

7/16, 1/2 open-ended wrenches, 10 adjustable wrench

Isopropyl alcohol (99.5% pure)

Cotton balls or swabs

Air duster

Teflon tape

Refrigerant: propane, liquid CO2 or liquid nitrogen


- Liquid CO2 requires using a siphon dip tube cylinder so that low-specified temperatures can be achieved.
-

Liquid propane requires inverting the propane cylinder.

Do not use regulators on CO2 or propane cylinder tanks.

Tubing and Fittings


The instrument comes with 1/8 metal tubing and fittings for making the connection to the gas pipeline or tank from which you will be obtaining the sample. A
nylon hose should not be used for the sample lines (unless it is lined with Teflon
or equal) because unlined hose may add or remove water from the gas causing an
error in the dew point determination. Nylon hose can be used for the refrigerant
connection. However, stainless steel tubing, 1/8 in diameter, is recommended if
the tubing length is greater than ten feet.

Installation and Set Up | 3-3

Installing the Dew Point Tester


An adjustable tripod is available for mounting the tester when used in the field.
Observe the following precautions when setting up the instrument and doing testing in the field.

Take a representative sample. We recommend that you use a sample probe


inserted into the center 1/3 of the pipeline. The probe should be inserted in a
horizontal, straight run of pipe. It should also be located at least ten pipe diameters upstream and downstream from any turbulence causing devices such
as an elbow, valve, orifice meter, etc.

Ensure that the temperature of the sampling line and the dew point tester
are above the dew point temperature of the gas in the pipeline. Heat-traced
sample tubing may be required in cold ambient conditions.

Ensure that the sampling line does not contain any condensed liquids such as
water, hydrocarbons, glycol, etc.

There must be no appreciable reduction in pressure between the gas source


and the dew point tester.

Connecting the Refrigerant


Once the dew point tester sample connection to the gas supply is properly set up,
you must connect the refrigerant valve to a cylinder of propane, liquid CO2, or
other liquefied gas that will be used as a refrigerant.

Inversion Cooling Rod

Flask

Clamp
Holder

Chain
Clamp

Figure 3-1.
Liquid nitrogen
refrigerant connections.

3-4 | AMETEK ChanScope and Dew Point Tester

Open the valve on the cylinder and invert it so that


the liquid refrigerant is available at the chiller. If you
are using a siphon dip tube, you do not have to invert
the cylinder.

Propane is a satisfactory refrigerant for dew point


temperatures down to about -20F (-28C).

Liquid carbon dioxide (CO2) may be used for


temperatures down to -80F (-62C).

For lower dew points you must use liquid nitrogen (Figure 3-1).

Installing the Glycol Filter Assembly

The glycol filter assembly is attached in-line to the inlet of the dew point tester
(Figure 3-2). Once installed, the filter assembly will not interfere with the testers
attachment to the tripod.

Selecting Temperature Display (ChanScope only)

O-Ring
Seal

Glycol Filter
Cartridge

Filter Housing

Cartridge

O-Ring Seal
Install
this end to
dew point
tester

Figure 3-2.
Glycol filter
installation.

Remove the rear cover of the ChanScope enclosure to change the temperature
display units. Selections are degrees Celsius (C) or degrees Fahrenheit (F).
There is a circuit assembly inside the enclosure. Set the miniature switch setting
#1 to ON for Fahrenheit or to OFF for Celsius. Switches 2, 3 and 4 are not used.

Installation and Set Up | 3-5

Start Up
1. Connect the sample tubing from the sample source to the dew point tester
sample inlet valve.
2. Turn the sample inlet valve until it is fully open.
3. Partially open the sample outlet valve just enough to allow flow through the
tester body but not enough to cause a drop in the sample pressure.
4. Allow the sample to flow through the tester until sufficient time has passed to
allow system equilibrium. This should be approximately five to ten minutes.
5. Connect the refrigerant tubing from the refrigerant cylinder to the refrigerant
inlet connection valve on the chiller device.
6. Slowly open the refrigerant valve to start the flow of refrigerant through
the chiller device. Once the temperature starts to drop, close the refrigerant
valve.
7. Slowly open and close the refrigerant valve so that the temperature of the
sample gas drops slowly.
8. Observe the mirror and thermometer/digital readout. At the same time, make
note of the temperature at which a spot of moisture first appears at the center
of the mirror.
9. Allow the mirror to warm up and make note of the temperature at which the
moisture spot disappears.
10. Repeat Steps 7 and 8 until the temperatures observed are within 2F (1C).
11. Record the average or mean temperature that you observed as the dew point
temperature.

3-6 | AMETEK ChanScope and Dew Point Tester

OPERATION

To prevent serious injury or incorrect measurements, adhere to all


test procedures when setting up the tester and performing tests in the
field.

Using Glycol Filters


When using the standard dew point tester, there may be times when glycol vapors
are present as you are measuring the dew point in the gas. These vapors can
lead to difficulty in determining the water dew point because the glycol vapors
also condense on the mirror, making observation of the water dew point difficult
or uncertain. A glycol filter can be used with the dew point tester under these
circumstances to remove the glycol vapor to make it easier to determine the true
water dew point.
The filtering material in the glycol filter absorbs both water vapor and glycol vapor. However, though it quickly becomes saturated with water vapor, it continues
to absorb glycol vapor for a somewhat longer time. It is during this latter period
that water dew points can be determined satisfactorily since once the filter is
water saturated, no more water vapor is removed from the gas and the dew points
will be the same if read without the filter.

Determining Equilibrium
When the sample gas is introduced at the side of the pressure chamber, it flows
through the glycol filter before entering the tester. The filter material should not
cause a drop in pressure at normal flow rates. However, you must allow sufficient
time for the filter material to attain equilibrium with the sample gas pressure. The
time required to reach equilibrium depends on the concentration of water and
glycol in the gas and the density of the filter material in the cartridge. Equilibrium is determined by running preliminary dew point tests. Two consecutive test
results without glycol masking of the mirror indicate that the tester is operating
normally.

Operation | 4-1

Test Precautions
These precautions must be observed to prevent serious injury and to ensure accurate measurements.
1. Vent the sample gas to a safe area to prevent the possibility of creating an
explosive atmosphere in an enclosed area.
2. Always wear safety goggles when working with refrigerants and high -pressure gases.
3. Do not exceed the pressure rating of the tester.
-

Models 13-100/13-1200 tester maximum working pressure is 3000 PSIG


(20.7 mPa).

Models 13-075/13-1175 tester working pressure is limited to 1000 PSIG


(6.9 mPa).

Models 13-110/13-1210 tester working pressure is 5000 PSIG (34.5 mPa).


Working pressure must not be exceeded as this can result in component blow out causing serious injury to the operator.

4. Clean the tester properly before using oxygen.


5. Take a representative sample using a sample probe inserted into the center
1/3 of the pipeline.
6. Ensure that the sampling line does not contain condensed water, hydrocarbons, glycol, etc.
7. Ensure that the temperature of the sampling line and the dew point tester are
above the dew point temperature of the sample gas. When the atmospheric
temperature is below the dew point temperature of the tester or the gas in the
pipeline or tank, the tester and sample line should be heated with a flexible
heat tape to avoid condensation between the gas source and the instrument.
8. Ensure that there is no appreciable reduction in pressure between the gas
source and the dew point tester.
9. If you have installed a filter to remove glycol from the gas, the material in the
filter absorbs both glycol and water. The filter usually becomes saturated with
water within five minutes but continues to absorb glycol. Only after the filter
is saturated with water can you make accurate dew point readings.

4-2 | AMETEK ChanScope and Dew Point Tester

Use extreme caution if you are using the tester on oxygen or other
explosive gases reactive to oil. The entire instrument, including the
sample line, must be thoroughly cleaned. Testers that have been
ordered cleaned for oxygen from the factory will have an oxygen
pressure gauge installed and will be tagged with the date they were
cleaned.

Operation | 4-3

Test Procedure
Follow this procedure in the ordered sequence to ensure accurate readings.
1. Open the sample inlet valve and purge the tester by flowing the test gas
through the sample chamber to expel all the air.
2. Regulate the flow of the test sample by partially closing the outlet valve and
then adjust to ensure that the sample flows continuously at full sample pressure through the chamber.
3. Adjust the outlet valve so that the gas sample can just be detected on your
fingertip. An aspirator bulb or vacuum pump can be used to bring atmospheric or low-pressure samples past the mirror.

A low flow of approximately 5 SCFH is recommended.

NOTE

4. ChanScope

For models with a digital dew point indicator, verify that the decimal point at
the far right side of the display is not blinking.

A blinking decimal point indicates that the tester is in pause or hold mode.
When in this mode, the displayed temperature is frozen and will not
change until you press the HOLD button. The HOLD button is a pushbutton or mementary switch; press HOLD at any time to ENTER or EXIT this
mode.

5. Open the refrigerant valve periodically for a few seconds; allow sufficient
time between valve openings to lower the mirror temperature. You can make
a quick observation to determine an approximate dew point. However, for the
final determination, you should lower the temperature at a rate of no more
than one or two degrees per minute.
6. Observe the mirror and thermometer/digital readout at the same time and
note the temperature at which a spot of moisture first appears at the center of
the mirror.
7. Allow the mirror to warm up and note the temperature at which the moisture
spot disappears.
8. Repeat Steps 5 and 6 until the temperatures observed are within 2F (1C).
9. Record the mean or average temperature observed as the dew point.

4-4 | AMETEK ChanScope and Dew Point Tester

Confirmation of Dew Point


In extreme cases, water dew point determination can be difficult, even after the
gas has passed through a glycol filter. An alternate method for determining if the
dew point is water is to operate at a different pressure.
Example:
1. A dew point determination is done in the normal manner at 1000 PSIG and a
dew point of 30F is obtained. Plotting on the Saturated Moisture Content
of Natural Gas as a Function of Temperature and Pressure chart (Drawing
13-0097) obtains a water vapor content of 6.3 lb/million cubic feet.
2. Pressure in the tester is reduced to 300 PSIG by partially closing the inlet
valve and adjusting the outlet valve to obtain the approximate same flow rate.
3. A dew point determination is now run at this reduced pressure.

If a dew point of 9F is obtained, you can be sure that the original determination at 1000 PSIG was correct since the water content remained the same (6.3
lb/MMscf) with the change is pressure and dew point. The water dew point is
lowered as the pressure is lowered, but the water content of the gas remains
the same. Had the water content been different at the second pressure and
dew point, then one of the dew points might have been a hydrocarbon, glycol
or alcohol dew point.

Interpretation of Dew Point Temperature


Dew points can be converted to water content using Drawings 13-0095, 13-0097,
and 13-0125 found in this manual. The ASTM tables are also included.

Operation | 4-5

Differentiating between Water and Hydrocarbon Dew Points


Sometimes, the hydrocarbon dew point of the gas will be the same or higher
temperature than the water dew point. In this case, the two dew points must be
differentiated.

The hydrocarbon wets the mirror and the film expands in an iridescent ring to
cover the mirrors surface, or appears as a waterfall of droplets that appear to
be pouring from the mirrors center.

The water condensate remains concentrated in the center of the mirror.

You can also verify the substance on the mirror by determining the
dew point at a lower pressure.

NOTE

4-6 | AMETEK ChanScope and Dew Point Tester

Special Test Procedure for Propane


Dew point tests of commercial propane are usually made at atmospheric pressure instead of at full sample flow as with other gases. Therefore, when testing
propane, the procedure for opening and closing the valves should be reversed to
ensure atmospheric pressure inside the tester.
Commercial propane is usually considered dry if the dew point is -15F or
lower when measured at atmospheric pressure.

NOTE

For detailed information refer to GPA Publication 2140-80 Liquid Petroleum Gas Specification and Test Methods, 1980 revision,
Tulsa, Oklahoma.

Operation | 4-7

This page intentionally left blank.

4-8 | AMETEK ChanScope and Dew Point Tester

Maintenance and Troubleshooting

Maintenance
Tools Required
Standard Dew Point Tester








7/16, 1/2, 1/4 open-end wrenches


10 adjustable wrench
1/8 Allen-head driver for loosening and tightening the window nut
Slotted screwdriver
Bench vise
Isopropyl alcohol (99.5%)
Air duster
Teflon tape
Cotton balls or swabs

ChanScope Dew Point Tester


Same as above plus:

Phillips screwdriver
1/8 hex wrench

Maintenance and Troubleshooting | 5-1

Replacing the Glycol Filter


If a glycol filter it being used, it will eventually become saturated with glycol and
will need to be replaced.
1. Disconnect the sample tubing from the glycol filter housing.
2. Open the filter housing and remove the old filter cartridge. Remove the Oring from the old filter cartridge and place it on the new filter cartridge.
3. Place the new filter cartridge in the filter housing and reassemble.
4. Reconnect the sample inlet tubing.

Cleaning the Window and Mirror


Both the window and mirror of the tester are easily removed for cleaning.

Use a cotton ball, cotton swab, or soft cloth to clean the window and mirror
surfaces. Be careful to avoid scratching the surfaces.

If a hydrocarbon or other contaminant remains after cleansing, use a soft


cloth with soap and water followed by a rinse using isopropyl alcohol.

To clean the high-pressure window, remove the screw and pull the unit out of
the window nut. Be careful to avoid scratching the window.

A scratched window should be replaced. A fracture could develop


from the scratch when the tester is subjected to high pressure.

NOTE

Recharging Batteries (ChanScope Dew Point Tester)


If the LED display on the ChanScope dew point tester goes blank, it is because
the battery pack needs to be recharged. The battery pack is a nickel metal hydride (NiMH) 1600 mAh and will last approximately 12 hours if fully charged. It
requires approximately 3 hours to fully recharge.

Plug the battery charger into the wall receptacle from 100-265 VAC, 50/60
Hz using the appropriate supplied blade adapter, and plug the other end into
the jack on the ChanScope.

The charger is not for outdoor use.

NOTE

5-2 | AMETEK ChanScope and Dew Point Tester

Charger Functionality

Charging begins when the battery pack is connected to the charger. The LED
is orange before the fast charge starts, and then the LED changes to red.
When the cells are fully charged and the voltage drops (because of the -dV
signal from the cells), the charger will go into a top-off charge mode before
switching to trickle charge mode.

During the top-off charge session, the LED will be green with short, intermittent orange flashes. When the top-off charge is complete, the charger will go
into trickle charge mode, changing the LED to green. The charge current
is now reduced to a safe level, permitting the charger to stay connected to
a NiCd battery pack without damaging the cells. Depending on the specific
manufacturer of NiMH cells, trickle charging in excess of 24 hours may not
be permitted.

If the safety timer disconnects before -dV, the top-off charge will not be
engaged. The charger will then go directly to trickle charge mode and the
LED will be green. If the battery voltage is far below normal, the charger will
stop the fast-charge current and switch to trickle charge mode. The LED
will then indicate error by flickering green and red.

If the mains is turned off, the charger will reset and start a new charge cycle
when the mains is turned on again.

NOTE

You can operate the tester during charging. The tester cannot be
operated without a battery.

Turn the tester off when not in use to conserve energy.

Use a ChanScope II battery charger only (PN 13-2019). Other chargers will cause internal damage.

NOTE

Recycle the battery pack at an appropriate facility for NiMH batteries. Visit
www.rbrc.org or call 1-800-8-BATTERY for locations.

Maintenance and Troubleshooting | 5-3

Bypassing the RTD


You can bypass the RTD at any time by operating the tester with the thermometer
instead.

Loosen the gland nut on the RTD and gently pull the RTD out. Replace with
the appropriate thermometer.

Turn the power switch on to illuminate the mirror and proceed as usual.

Remember that the LED displays now indicates ambient temperature


when operating in this mode.

NOTE

Replacing the batteries and changing the temperature display unit


selection are the only operations allowed that require removing the
back cover of the dew point indicator. All optical and electronic
adjustments are made only at the factory. Failure to observe this may
void the warranty.

5-4 | AMETEK ChanScope and Dew Point Tester

Calibration
Except for the thermometer, calibration is not possible or necessary on the Standard Dew Point Testers (Models 13-075, 13-100, and 13-110. The thermometer
can be removed and checked for accuracy at any reputable calibration service
equipped to do so.
Calibration of the ChanScope RTD and the electronics is done at the factory.
Return the tester after obtaining an RMA.

Selecting Temperature Display (ChanScope only)


Remove the rear cover of the ChanScope enclosure to change the temperature
display units. Selections are degrees Celsius (C) or degrees Fahrenheit (F).
There is a circuit assembly inside the enclosure. Set the miniature switch setting
#1 to ON for Fahrenheit or to OFF for Celsius. Switches 2, 3 and 4 are not used.

Maintenance and Troubleshooting | 5-5

TROUBLESHOOTING THE CHANSCOPE

PROBLEM

CAUSE

SOLUTION

Intermittent display

RTD, battery
connections, or switch

Check RTD cable, battery


connections, and switch

Display is segmented, or no
display.

Low batteries, voltage


less than minimum

Check charger output, replace


battery pack; recharge.

With Chanscope OFF, charger turns OFF prematurely.


Battery not fully charged.

Battery may have


developed a memory.

Remove charger from outlet.


Turn on Chanscope and discharge
battery completely.
Plug charger into outlet to charge.

Excessive glare or light on


mirror.

Lamp too deep in


holder.

Remove cover, adjust lamp depth.

RTD not making proper


contact.

Check for corrosion in well and


depth of RTD in well to bottom. Use
thrmometer to verify. Check RTD
at 32F (0C) in insulated ice water
bath, stirring the crushed ice.

Calibration.

Contact factory for calibration.

Device is in HOLD
mode.

Press HOLD button on the front


panel to resume normal operation.

Temperature not reading


correctly.

Temperature not changing,


but decimal point is flashing
in lower right portion of
display.

5-6 | AMETEK ChanScope and Dew Point Tester

TROUBLESHOOTING DEW POINT READINGS


PROBLEM

CAUSE

SOLUTION

Haze on entire monitor


obscures dew point.

Possible glycol
contamination.

Install glycol filter.

Condensed water in sample


line.

Temperature of sample
line and tester below
the temperature of
sample gas. Not using
a probe to sample gas,
slugs of water entering
the sample line from
the pipe wall.

Heat trace sample line. Purge line


with dry nitrogen. Avoid excessive
fittings, gauges, etc., in sample line
that may be difficult to dry down
after condensation. Use sample
probe.

Low dew point using glycol


filter.

Moisture equilibrium
in glycol filter not yet
reached.

Glycol filter requires time to saturate


with water. Allow to saturate for 5 to
10 minutes.

CO2 and propane refrigerant


not chilling to specified
temperature.

Using gas instead of


liquid. Using a regulator
on the cylinder.

CO2 cylinder must be a siphon dip


tube cylinder. Propane cylinders
must be inverted. Do not use a
regulator.

Unable to distinguish water


dew point from hydrocarbon
dew point.

Both dew points can


occur at a lower,
same, or higher
temperature depending
on hydrocarbon
components present
and the amount of
water.

Hydrocarbon dew point wets


the mirror with an iridescent ring
followed by condensing droplets.
This is your dew point. Viewing the
hydrocarbon dew point at an angle
to the mirror can be helpful. The
water condensate begins at the
mirrors center and moves outward
across the mirror.

Maintenance and Troubleshooting | 5-7

REPLACEMENT PARTS
MODELS 13-075, 13-100, 13-110, 13-210, 13-1175, 13-1200 and 13-1210
PART #
13-0005
13-0006
13-0007
13-0009
13-0018
13-0020
13-0021
13-0022
13-0024
13-0025
13-0027
13-0038
13-0046
13-0047
13-0048
13-0052
13-0053
13-0054
13-0060
13-0064

DESCRIPTION
Window, Models 13-100 & 13-200 and 13-1200
Window retaining nut, Model 13-100 & 13-110
Mirror assembly, regular chiller
Mirror insulator, Models 13-153 & 13-154 kits
Chiller body, regular
Thermoguard assembly
Lock nut, thermoguard
Bulb shield, thermoguard
Shield, thermoguard
Thermometer, range -30 to +110F
Thermometer, range -90 to +40F
Thermometer, range -100 to +10C
Retaining nut, quick chiller
Block, with tube, quick chiller
Quick chiller assembly
Glass window assembly, Models 13-100, 13-110, 13-200, 13-210, 13-1200,
13-1210 (sour gas and 5,000 PSIG)
Cooling rod, nitrogen chiller
Glycol filter assembly
Thermometer, range -35 to +50C
Window, glass, Models 13-100, 13-110, 13-200, 13-210, 13-1200, 13-1210
(sour gas and 5,000 PSIG)

13-0065

Adapter, glass window, Models 13-100, 13-200, 13-210, 13-1200, 13-1210


(sour gas and 5,000 PSIG)

13-0068
13-0071
13-0072
13-0077
13-0078
13-0079
13-0084
13-0085
13-0086
13-0101M
13-0102
13-0111
13-0116
13-0117
13-0119
13-0122

Regular chiller assembly


Extension, tripod adapter, nitrogen chiller
Liquid nitrogen chiller assembly
Nut, window (Model 13-075)
Nut, mirror, retaining
Window, plastic, Models 13-075, 13-175, & 13-1175
Adapter, gauge
Body, Models 13-075, 13-175, & 13-1175
Adapter, tripod
Body, Model 13-100, 13-110, 13-200, 13-210, 13-1200, 13-1210
Mirror, swivel
Body assembly, magnifier
Window, magnifier, eye guard
Lens retainer, Model 13-075
Lens, magnifier, objective
Eye guard, plastic, magnifier

5-8 | AMETEK ChanScope and Dew Point Tester

PART #

DESCRIPTION

13-0124
13-0146
13-0147
13-0148
13-0149

Lens retainer, Model 13-100, 13-110 magnifier


Thermometer, certified range -90 to +40F
Expansion tube only, quick chiller
Thermometer, certified range -200 to +0F
Thermometer, certified range -30 to +110F
Replacement Seal Kit for 13-100 includes:
(6)
P-0063
O-ring
(1)
P-0065
O-ring
(3)
P-0067
O-ring
(1)
P-0071
O-ring
13-0153
(1)
P-0221
O-ring
(1)
13-0009 Insulator (2 parts)
13-0154
Same as 13-153, plus P-70 plug O-ring used for Model 13-075
13-0157
Thermometer, certified range -5 to +50C
13-0158
Thermometer, certified range -130 to -20C
13-0161
Extension, tripod adapter
13-0210
L-bracket, battery holder
13-0277
CO2 bottle, 20 cu ft, with siphon tube, nut and tailpiece (full)
13-0303
Illuminated magnifier assembly, Model 13-075
13-0304
Illuminated magnifier assembly, Model 13-100 & 13-110
13-2277
CO2 bottle, 20 cu.ft., with siphen tube, nut and tailpiece (empty)
36-0028
Hose, refrigeration, 1/4 flexible, 1250 PSIG x 10 feet
36-0045
Tube, assembly, 1/8 x 10 feet SST, brass fittings, welded
36-0050
Refrigerant tube assembly, nylon (sour gas service)
36-0087
Tube, assembly, 1/8 x 10 feet, SST, with brass fittings
36-0000-10
Tubing, coiled, SST, 1/8 x 10 feet
187-20403
Connector, 1/4 MP x 1/8 T, SST
188-13044
Connector, 1/8 MP x 1/8 T, SST
C03322
Connector, 1/8 FP x 1/8 T, SST
C05013
Lubricant
C08161
Valve, inlet and outlet
C08435
Cartridge, mega-glysorb filter (package of 10)
C08517
Connector, 1/4 FP x 1/8 T, SST
C08884
O-ring for gauge adapter (sour gas service)
C08885
O-ring for window (sour gas service)
C08886
O-ring, mirror (sour gas service)
C09490
Battery holder, magnifier
C09509
Valve handle
C09549
Batteries, AAA
C09550
Light bulb, magnifier
C09608
Pressure gauge, oxygen service, 1000 PSIG
C10499*
Gauge, digital pressure, 1000 PSI
C10500*
Gauge, digital pressure, 3000 PSI
C10501*
Gauge, digital pressure, 5000 PSI
* Consult factory for optional measurement units.

Maintenance and Troubleshooting | 5-9

P-0018
P-0020
P-0063
P-0067
P-0070

Clamp, magnifier
Clamp holder, nitrogen chiller
O-ring, gauge adapter, quick chiller
O-ring, window mirror (included in 13-153 & 13-154 kits)
O-ring, 13-0303 magnifier (included in 13-154 kit)

P-0071

O-ring, window for Model 13-075, 13-0304 magnifier (included in


13-153 and 13-154 kits)

P-0182
P-0301
P-0535
P-0536
P-0583
P-0591
P-0594
P-0711
P-0720
P-0986
P-0987
P-0988
P-1037
P-1360
P-1367
P-1392
P-1818
P-2300
P-2346

Hose nipple, chillers


Valve, chiller
Tripod
20# propane cylinder with valve & 1/4 MNPT connector
Pressure gauge, brass, 0-15 PSIG, Model 13-075
Pressure gauge, brass, 0-1000 PSIG, Model 13-075
Pressure gauge, brass, 0-3000 PSIG, Model 13-100, 13-110
Quick disconnect coupling, chiller
Pressure gauge, brass, 0-200 PSIG, Model 13-075
Flask, N2, chiller
Chain clampo, nitrogen chiller
Plastic handle for carrying case
Pressure gauge, brass, oxygen service, 3000 PSIG
Nut, brass (for CO2 bottle)
Tailpiece, brass, 1/4 MP (for CO2 bottle)
Pressure gauge, 0-3000 PSIG, Model 13-100 sour gas service
Pressure gauge, brass, 0-5000 PSIG, Model 13-110 & 13-210
Valve, SST, 6000 PSI (sour gas service)
Pressure gauge, 0-5000 PSIG, Model 13-0100 & 13-210, SST

5-10 | AMETEK ChanScope and Dew Point Tester

ChanScope Digital Dew Point Tester


MODELS 13-175, 13-200, 13-210 (10,000 Series SN)

PART #
13-0204
13-0205
13-0210
13-0214
13-0215
13-0217
13-0219*
13-0220
13-0227
13-0230
13-0234
13-0236*
13-0241
13-0250
13-0253
13-0254
13-1238
187-20535
C08846
P-0221
P-1044
P-2770
P-2820
P-2822***
P-2826
P-2828
P-2830
P-3113
P-3115
P-3116
*

DESCRIPTION
Window nut, Model 13-175
Window nut, Model 13-0200, 13-0210
L-bracket, battery
Optical assembly
Beam splitter assembly
Prism assembly
Logic board assembly
LED board assembly
Eyepiece assembly
Gland nut
Switch
RTD assembly with strain relief (spring-loaded)
Battery holder, two cell
Dewpoint indicator (ChanScope) assembly
Wiring harness assembly
Cable assembly
Black case with foam inserts
Adapter, 220 VAC to 115 VAC, 50/60 Hz, use with P-3116 charger
Kit, 230 VAC plug adapter
O-ring, RTD gland nut
Lamp, screw base
Lamp, bayonet base
O-ring, tube support
Eye Cup
Screw, 2-56 x .250, rnd Head, slot
Screw, 6-32 x .250, rnd head, Phillips
Screw, 6-32 x 1.0, skt head, SST
Battery, C cell, NiCad recharge
Holder, (1) C cell, #2173
Battery charger, NiCad

RTD and circuit assembly should be purchased together and be factory-calibrated.


Consult the factory for price and availability.

*** OBSOLETE. Replaced by C11468 Eye Cup and 13-0227 Eyepiece Assembly.
Consult the factory for price and availability.

Maintenance and Troubleshooting | 5-11

ChanScope II Digital Dew Point Tester


MODELS 13-1175, 13-1200, 13-1210 (20,000 Series SN)

PART #
13-0204
13-0205
13-0215
13-0227
13-0236*
13-0307*
13-1214
13-1217
13-1238
13-1250
13-1252
13-1274
13-1276
13-1277
13-2019**
C09867
C11003
C11468
C11921
C11922
C11923
C12024
P-2820
P-2821
P-2826
P-2828
H-6033
P-2836
*

DESCRIPTION
Window nut, Model 13-1175
Window nut, Model 13-1200, 13-1210
Beam splitter assembly
Eyepiece assembly
RTD assembly with strain relief (spring-loaded)
Circuit assembly
Optical assembly
Prism assembly
Case with foam inserts
ChanScope II assembly
ChanScope LED/lamp wiring harness assembly
Battery pack, rechargeable
Switch/battery/harness assembly
Assembly, battery charger connector wiring harness
Battery charger, NiMH, 1600 mAh, 100-265 VAC, 50/60 Hz, with 3 blades
Pushbutton switch
Charger connection cap assembly
Eye cup
Adapter, charger, US blade
Adapter, charger, Euro blade
Adapter, charger, UK blade
Charger, Z22 PDA, 110-240V
O-ring, tube support
Rocker switch
Screw, 2-56 x .250 rnd head, slotted
Screw, 6-32 x .250, rnd head, Phillips
Screw, 6-32 x 1.125, skt head, SST (4 required for eyepiece)
Strain relief, RTD, cable

RTD and circuit assembly should be purchased together and be factory-calibrated.


Consult the factory for price and availability.

** Replaces 13-1275 charger, 187-20535, 220 VAC to 115 VAC adapter,


and C08846 plug adapter kit.

5-12 | AMETEK ChanScope and Dew Point Tester

DRAWINGS

1. (13-0052) Glass Window Assembly for sour gas/5000 PSIG DPT



includes:
- (13-0064) Glass Window
- (13-0065) Adapter
2. Use perfluoroelastomer O-rings for corrosive gases.
Figure 6-1.
Glass window
assembly for sour
gas/5000 PSIG DPT
DWG 13-0052.

3. Lubricate O-ring (C08885) with silicon O-ring lubricant (C05013).

Drawings | 6-1

10
Install
this end to
dew point
tester

11
ITEM
1
2

PART #
36-0000-10
C03322

Figure 6-2.
Glycol filter assembly.
DWG 13-0054.

6-2 | AMETEK ChanScope and Dew Point Tester

4
5
6
7
8
9
10
11

P-2577
C08435
P-3178
C04364
C08434
C08435

DESCRIPTION
1/8 SST Tubing
Tube Connector
Valve - supplied with
dew point tester
Bushing
O-Ring Seal
Filter Cartridge
Elbow
Adapter
Filter Housing
O-Ring Seal
Cartridge

Dew point tester and tripod are not included in 13-0072 immersion
cooling rod assembly.

NOTE

6
7

Approximate Dew Points obtainable with refrigerants:


Dry Ice in solvent:
-60F
Liquid nitrogen: -200F

ITEM
1
2
3

Figure 6-3.
Immersion cooling rod
assembly.
DWG 13-0072.

4
5
6
7

PART #

DESCRIPTION

P-0020
Clamp Holder
13-0071
Extension
Model 13-0100 or Model 13-0075 with
13-0007 Regular Mirror
13-0053
Immersion Cooling Rod
P-0986
Flask
P-0987
Chain Clamp
P-0535
Tripod

Drawings | 6-3

14 3/4 approx

13
4
14

12

15
16

11
10

Thermometer
specify temperature range

17

10 1/2
approx

18

1/8 NPT

19

1/8 NPT
Gas Outlet

23

ITEM

Figure 6-4.
Dew Point Tester,
Model 13-0075 with
quick chiller assembly.
DWG 13-0075.

PART #

DESCRIPTION

C03322

Connector

C08161

Valve

P-1190

Elbow

Refrigerant Outlet

P-0301

Chiller valve

13-0086

Adapter, Tripod

C08161

Valve

13-0079

Window

13-0077

Nut

10

13-0085

Body

11

P-0071

Extension

12

13-0084

Adapter

13

P-0591

Pressure Gauge

14

P-0063

O-Ring, Gauge Adapter

15

P-0067

O-Ring, Window Mirror

16

13-0078

Nut

17

13-0048

Quick Chiller Assembly

18

13-0046

Retaining Nut

19

13-0021

Lock Nut, Thermoguard

20

13-0020

Thermoguard Assembly

21

13-0024

Shield

22

13-0023

Adjusting Knob

23

Refrigerant Inlet 1/8 FNPT

6-4 | AMETEK ChanScope and Dew Point Tester

21
20

22

Figure 6-5.
Chillers, regular and
quick.
DWG 13-0090.

Drawings | 6-5

Figure 6-6.
Chart.

6-6 | AMETEK ChanScope and Dew Point Tester

Figure 6-7.
Chart.

Drawings | 6-7

Figure 6-8.
Chart.

6-8 | AMETEK ChanScope and Dew Point Tester

Figure 6-9.
Chart.

Drawings | 6-9

Figure 6-10.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 1.

6-10 | AMETEK ChanScope and Dew Point Tester

Figure 6-11.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 2.

Drawings | 6-11

Figure 6-12.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 3.

6-12 | AMETEK ChanScope and Dew Point Tester

Figure 6-13.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 4.

Drawings | 6-13

Figure 6-14.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 5.

6-14 | AMETEK ChanScope and Dew Point Tester

Figure 6-15.
Bureau of Mines
dew point tester with
regular chiller.
DWG 13-0100.

Drawings | 6-15

Figure 6-16.
Dew point to moisture
content conversion
chart.

6-16 | AMETEK ChanScope and Dew Point Tester

Figure 6-17.
Installation of dew
point tester.
DWG 13-0150.

Drawings | 6-17

Figure 6-17.
Dew point indicator
assembly.
DWG 13-2016.

6-18 | AMETEK ChanScope and Dew Point Tester

13-0116

13-0122

13-0111

13-0119

13-0117

P-0070

13-0305

H-8014

PART #

Illuminator Assembly

O-Ring

Lens Retainer

Body Assembly
1/4-20 x .25 LG, Allen
HD set screw
Lens

Viewer

Window Eyeguard

#8-32 x .5 LG, BHP

DESCRIPTION

4. Rotate 13-0305 illuminator counterclockwise to turn it on.

3. Screw 13-0305 illuminator assembly into 13-0117 lens retainer.

2. Lubricate the magnifier O-ring and push the magnifier inside of the round
window nut.

1. Remove 13-0102 observation mirror from the dew point tester body.

INSTALLATION INSTRUCTIONS

ITEM

Figure 6-18.
Illuminated magnifier
assembly, Model 13-075.
DWG 13-0303.

Drawings | 6-19

INSTALLATION INSTRUCTIONS
1. Remove 13-0102 observation mirror from the dew point tester body.
2. Lubricate the magnifier O-ring and push the magnifier over the outside of the
round window nut.
3. Screw 13-0305 illuminator assembly into 13-0124 lens retainer.
4. Rotate 13-0305 illuminator counterclockwise to turn it on.

Used with Models 13-100


& 13-110 Dew Point Tester,
with 13-0006 round window
nut.

ITEM

Figure 6-19.
Illuminated magnifier
assembly, Models 13100 and 13-110.
DWG 13-0304.

6-20 | AMETEK ChanScope and Dew Point Tester

1
2
3
4
5

PART #
13-0305
P-0071
13-0124
H-8014
13-0122

6
7
8
9

13-0116
13-0111
13-0119

DESCRIPTION
Illuminator
O-Ring
Lens Retainer
#8-32 x .5 LG, BHP
Viewer
1/4-20 x .25 LG, Allen
HD set screw
Window Eyeguard
Body Assembly
Lens

ITEM

PART #

DESCRIPTION

C09490

Flashlight, Maglight Model M3A016

13-0301

Adapter, flashlight x 3/8-32

13-0302

Light pipe, 0.25

C09549

Batteries, AAA

C09550

Light bulb (replacement only), LM-3A-001

Figure 6-20.
Illuminator Assembly.
DWG 13-0305.

Drawings | 6-21

Figure 6-21.
ChanScope II.
DWG 13-2017.

6-22 | AMETEK ChanScope and Dew Point Tester

13-2019 Battery Charger


C11922 Charger Adapter (Euro blades)
C11923 Charger Adapter (UK blades)

Not Shown:

Das könnte Ihnen auch gefallen