Beruflich Dokumente
Kultur Dokumente
Notiuni Curs I
Semnale si sisteme
Reprezentarea imaginii
Caracteristicile si factorii calitativi ai imaginilor
medicale
University Politehnica of Bucharest
Cursul II
Notiuni Curs I
Semnale si sisteme
Reprezentarea imaginii
Caracteristicile si factorii calitativi ai imaginilor
medicale
University Politehnica of Bucharest
Some modalities are very good at depicting anatomical structure (bone, different tissue types,
boundary between different organs)
X-ray, X-ray CT
MRI
Some modalities do not depict anatomical structures well, but reflect the functional
status (blood flow, oxygenation, etc.)
Ultrasound
PET, functional MRI
Boundaries between the two classes are blurring as the imaging resolution continues to improve
Spectrul electromagnetic
The electromagnetic spectrum comparing the size of objects that can be studied with various
techniques.
Lesson II
Notiuni Curs I
Semnale si sisteme
Reprezentarea imaginii
Caracteristicile si factorii calitativi ai imaginilor
medicale
University Politehnica of Bucharest
10
Semnale si sisteme
The object being imaged is an input signal
Typically a 3D signal
The imaging system is a transformation of the input signal to an
output signal
The data measured is an output signal
A 2D signal (an image, e.g. an X-ray) or a series of 2D signals
(e.g. measured projections from a CT scan), or 4D data (a series of 3D
volume in time)
Image reconstruction
An inverse process: from the measured output signal -> desired
images of the object (a series of 2D slices)
INPUT SIGNAL -> SYSTEM or PROCESS -> OUTPUT SIGNAL
Exemple de semnale
Transformarile semnalelor
Frecventa spatiala
Fourier: Exemple
Fourier: Exemple
Fourier - Aplicatii
A chest radiograph is illustrated in (a) with its 2-D Fourier spectrum in (b). The spatial
frequency data show a broad range of values with significant vertical and horizontal
features associated with the vertebral column and ribs, respectively.
Fourier - Aplicatii
A potential use of the FT and its inverse is the removal
of unwanted or corrupt data from a digital image and
this process is illustrated in the final figure below.
An extreme example of a corrupt image may be
generated by adding together the two images.
The Fourier spectrum in panel (b) portrays the
frequency characteristics of the summed image.
The undesirable features attributable to the sinusoidal
pattern may be removed by editing of the data in the
frequency domain as in panel (c), before the inverse FT
is performed to recover an image largely free of
artifact, as in panel (d).
Lesson III
Notiuni Curs I
Semnale si sisteme
Reprezentarea imaginii
Caracteristicile si factorii calitativi ai imaginilor
medicale
University Politehnica of Bucharest
30
Reprezentarea imaginii
semnal bidimensional in spatiul continuu= definit ca o
functie de 2 variabile in spatiu continuu: f(x,y)
f(x,y) = interpretata ca o functie de luminanta variabila in
spatiul bidimensional (x,y)-> r(x,y)
=> d.p.d.v. al notatiei f(x,y) = functie definita in domeniu
spatial
Teoretic -> modelul imaginii intr-un spatiu continuu
nelimitat-> un plan bidimensional, dar abordarea folosita in
practica -> dimensiunea imaginii este intotdeauna limitata
la un spatiu finit (dreptunghi)
University Politehnica of Bucharest
31
Grayscale image
Surface representation
32
f ( x, y ) = [ fR ( x, y ), fG ( x, y ), fB ( x, y )]T
University Politehnica of Bucharest
33
34
Spatial sampling
ideal spatial sampling =
the analogue image
a 2-D sampling signal s(x,y) (formed
of an infinite number of periodically repeated Dirac impulses)
s ( x, y ) =
( x kx, y iy )
i = k =
the sampling
f s ( x, y ) = f ( x, y ) s ( x, y )
35
Spatial sampling
only those discrete values of the analogue image are
needed that correspond to the nodes of the sampling grid:
f = f (kx, iy )
The information carried
by the intermediate image
values is lost
so that the image matrix
= the complete image
content
University Politehnica of Bucharest
36
Sampling
Sampling = choosing which
points you want to have
represent a given image;
analog image -> sampling = a
mapping from a continuum
of points in space to a
discrete set
digital image -> sampling = a
mapping from one discrete
set of points to another
(smaller) set
University Politehnica of Bucharest
37
a pixel is a point
It is NOT a box, disc or teeny wee light
It has no dimension
It occupies no area
It can have a coordinate
More than a point, it is a SAMPLE
38
http://www.cs.brown.edu/exploratories/freeS
oftware/repository/edu/brown/cs/exploratori
es/applets/sampling/introduction_to_samplin
g_guide.html
University Politehnica
of Bucharest
39
40
41
42
Digitization
the representation of an object, image, sound, document or
a signal (usually an analog signal) by a discrete set of its points or samples
The result:
digital representation or digital image, for the object,
digital form for the signal
To obtain numbers tractable by computers, the sample values obtained by
sampling are consequently digitized by an analogue to- digital (A/D)
converter.
The finite code: only limited-precision numbers => the measured values
rounded to the nearest of the available levels ( 256, 4096, or 65,536 levels
corresponding to 8-, 12-, or 16-bit fixed point)
University Politehnica of Bucharest
43
Digitization
As any natural image has limited dimensions, the result of
digitization is a finite set of numbers representing the sample
values picture elements, pixels*, that may be arranged into a
matrix.
Contemporary systems use matrices of sizes approximately
in the range between 0.3 and (rather rarely) 16 Mpixels (Mpx),
corresponding to square matrices of 512 512 to 4096 4096;
however, many systems use the rectangular (nonsquare) format.
This image size, together with the bit depth, determines the memory
requirements (unless data compression is used): e.g., a 5-Mpx 16-bit
gray-scale image requires 10 Mbytes of memory space, and a color
image three times as much.University Politehnica of Bucharest
44
Cursul IV
Cuprins
Caracteristicile si factorii calitativi in imaginile
medicale:
calitatea imaginii,
contrastul imaginii,
Sensibilitate la contrast,
Functia de transfer a modulatiei
Contrastul local
Factorul incetosare - blur si vizibilitatea detaliului,
zgomot, artefact, distorsiune,
compromisuri,
Caracteristicile tesutului
48
The objective = make an object within the patient's body visible to the observer.
The visibility: depends on the characteristics of the imaging system and the
manner in which it is operated.
most variables are adjustable physical quantities associated with the imaging
process: kilovoltage (radiography), gain (sonography), and echo time (TE) (MRI)
The values selected -> determine the quality of the image and the visibility of
University Politehnica of Bucharest
49
specific body features.
IMAGE QUALITY
50
IMAGE QUALITY
The quality of a medical image: the imaging method, the
characteristics of the equipment, and the imaging variables selected
by the operator
Image quality = a composite of at least five factors: contrast, blur,
noise, artifacts, and distortion.
The human body contains many structures and objects that are
simultaneously imaged by most imaging methods.
We often consider a single object in relation to its immediate
background
The visibility of an object is determined by this relationship rather
than by the overall characteristics of the total image.
University Politehnica of Bucharest
51
Image Contrast
Contrast means difference.
52
Image Contrast
The physical contrast of an object must represent a difference in
one or more tissue characteristics.
For example, in radiography, objects can be imaged relative to their
surrounding tissue if there is an adequate difference in either
density or atomic number and if the object is sufficiently thick.
When a value is assigned to contrast, it refers to the difference
between two specific points or areas in an image.
In most cases we are interested in the contrast between a specific
structure or object in the image and the area around it or its
background.
The degree of physical object contrast required for an object to be
visible in an image depends on the imaging method and the
characteristics of the imaging system.
Image Contrast
The primary characteristic of an
imaging system that establishes the
relationship between image contrast
and object contrast is its contrast
sensitivity.
Consider the situation shown below.
The circular objects are the same size
but are filled with different
concentrations of iodine contrast
medium.
That is, they have different levels of
object contrast. When the imaging
system has a relatively low contrast
sensitivity, only objects with a high
Increasing Contrast Sensitivity Increases Image
concentration of iodine (ie, high
Contrast and the Visibility of Objects in the Body
object contrast) will be visible in the
image.
If the imaging system has a high
contrast sensitivity, the lower-contrast
University Politehnica of Bucharest
55
objects will also be visible.
Contrast Sensitivity
contrast sensitivity = a characteristic of the imaging method and the
variables of the particular imaging system
It relates to the system's ability to translate physical object contrast
into image contrast.
It is difficult to compare the contrast sensitivity of various imaging
methods because many are based on different tissue characteristics.
certain methods do have higher contrast sensitivity than others.
Exemplu
(CT) generally has a higher contrast sensitivity than conventional
radiography
This is demonstrated by the ability of CT to image soft tissue objects
(masses) that cannot be imaged with radiography.
University Politehnica of Bucharest
56
Contrast Sensitivity
Consider the image:
Here is a series of objects with
different degrees of physical
contrast. They could be vessels
filled with different concentrations
of contrast medium. The highest
concentration (and contrast) is at
the bottom. Now imagine a
curtain coming down from the top
and covering some of the objects
so that they are no longer visible.
Contrast sensitivity is the
characteristic of the imaging
system that raises and lowers the
curtain. Increasing sensitivity
raises the curtain and allows us to
see more objects in the body. A
Effect of Contrast Sensitivity on Object Visibility
system with low contrast
sensitivity allows us to visualize
only objects with relatively high
University Politehnica of Bucharest inherent physical contrast.
57
Cursul V
63
64
Noise
65
Noise
In the image below we find our
familiar array of body objects
arranged according to physical
contrast and size.
We now add a third factor, noise,
which will affect the boundary
between visible and invisible
objects.
The general effect of increasing
image noise is to lower the curtain
and reduce object visibility.
67
Cursul 6
Artifacts, Distortion
most imaging methods can create image features that do not
represent a body structure or object
an artifact does not significantly affect object visibility and diagnostic
accuracy.
But artifacts can obscure a part of an image or may be interpreted as
an anatomical feature.
Distortion: A medical image should not only make internal body
objects visible, but should give an accurate impression of their size,
shape, and relative positions.
University Politehnica of Bucharest
70
Artifacts
Artifacts: image features that do not correspond to a real object, and are not due
to noise
artefactul de miscare: blurring or streaks due to patient motion
artefactul in forma de stea star artifact: in CT, due to presence of metallic
material in a patient
artefactul dat de cresterea in intensitate a fascicolului: broad dark bands or
streaks, due to significant beam attenuation caused by certain materials
artefactul in forma de inel: because detectors are out of calibration
Artifacts, Distortion
Acuratetea
Acuratetea cantitativa
Acuratetea diagnosticului
Sensibilitate
Specificitate
a+b
c+d
Total
a+c
b+d
a+b+c+d
Sensibilitatea =
a
100
a+c
Boala vizat
Total
Rezultatul
testului
screening
Total
Specificit atea =
d
100
b+d
Prezent
Absent
Pozitiv
a+b
Negativ
c+d
a+c
b+d
a+b+c+d
Compromises
why we do not adjust each imaging procedure to yield maximum
visibility?
the variables that affect image quality also affect factors such as
radiation exposure to the patient and imaging time.
In general, an imaging procedure should be set up to produce
adequate image quality and visibility without excessive patient
exposure or imaging time.
84
85
Examples
In projection imaging (radiography and fluoroscopy), images are
formed by projecting an x-ray beam through the patient's body
and casting shadows onto an appropriate receptor that converts
the invisible x-ray image into a visible light image.
The primary advantage: is that a large volume of the patient's
body can be viewed with one image.
A disadvantage is that structures and objects are often
superimposed so that the image of one might interfere with the
visibility of another.
86
Examples
Tomographic imaging, sonography, single photon emission
tomography (SPECT), positron emission tomography (PET), and
MRI produces images of selected planes or slices of tissue in the
patient's body.
The general advantage of a tomographic image is the increased
visibility of objects within the imaged plane.
One factor that contributes to this is the absence of overlying
objects.
The major disadvantage is that only a small slice of a patient's
body can be visualized with one image.