Beruflich Dokumente
Kultur Dokumente
2080A102/B101/M101/M102
SmartLab
Horizontal Sample Mount X-Ray Diffractometer
for Thin Film Analysis
Instruction Manual
Rigaku Corporation
NOTICE
1.
This manual described in it may not be disclosed to a third party or copied, in whole or in
part, without the written consent of Rigaku.
2.
As a rule, one set of the instruction manual has to be purchased for each product.
3.
If there are any missing or incorrectly collated pages in the delivered instruction manual,
contact the nearest Rigaku branch office or sales office for instruction manual replacement.
4.
In no event will Rigaku be responsible for the results of the use of this manual.
5.
The contents of this manual are subject to change without prior notice.
Safety Precaution - 1
ME11507B
Safety Precaution - 2
ME11507B
Safety Precaution - 3
3.1 Notes on Radiation Enclosure and X-ray Shutter Safety Release Operation
(1) Rigakus radiation enclosure is designed to eliminate the possibility of X-ray exposure outside the
radiation enclosure during normal use of the X-ray apparatus. When you use the radiation enclosure in
conjunction with the safety mechanism during the operation of Rigakus standard X-ray diffractometer,
you can use X-rays only when the radiation enclosure door is closed. Even if you carelessly open the
door, the safety mechanism actuates to automatically stop X-ray generation.
(2) In the safety release state (in which the DOOR OPEN switch is depressed to blink the DOOR-OPEN or
OPEN SAFETY lamp), however, the safety mechanism is disabled so that X-ray generation does not
stop when you open the door.
(3) The X-ray shutter safety mechanism works so that the X-ray shutter does not open while the radiation
enclosure door is open no matter whether the automatic or manual mode is chosen. However, if the
X-ray shutter safety mechanism is placed in the release state when the radiation enclosure is in the
safety release state, the shutter can be opened and closed while the radiation enclosure door is open so
that any parts of your body inadvertently positioned in the X-ray path inside the radiation enclosure
could be exposed to X-rays. Therefore, when the equipment is used with the radiation enclosure and
X-ray shutter placed in the safety release state, the equipment must always be handled and operated
under the direct guidance and control of certified X-ray apparatus handling personnel.
(4) The key for disabling the safety mechanism should always be handled under the control of a certified
X-ray apparatus supervisor. It should normally be removed from the safety release unit and placed
under the control of a certified X-ray apparatus supervisor.
(5) If the safety mechanism is disabled, the controlled area enlarges beyond the outer surface of the X-ray
diffractometer enclosure. Therefore, when disabling the safety mechanism with the key, define the
controlled area anew (by enclosing the area with a wall or the like, drawing a white line or striped line
mark on the floor, encircling the area with a rope, setting up flags and joining them to mark the area, or
marking the area in some other conspicuous manner) and provide an adequate safeguard (e.g., protective
aprons, protective gloves, and shielding screens) to minimize the exposure of X-ray handling personnel
to radiation.
ME11507B
Safety Precaution - 4
(6) If Rigakus radiation enclosure and safety mechanism are not purchased, the automatic X-ray shutoff
feature will not be exercised in accordance with radiation enclosure door opening/closing. If the X-ray
shutter mounted coupling protector or an optical system part such as the slit box and direct beam stopper
is removed from the standard assembly or modified, the personnel outside the equipment may also be
exposed to X-rays. Never remove or modify such parts.
NOTICE :For the details of the radiation enclosure and X-ray shutter safety mechanism, refer to after
mentioned text of this manual.
NOTICE :The safety mechanism reduces the possibility of radiation exposure. However, it does not
guarantee that you are absolutely safe from radiation exposure.
ME11507B
Safety Precaution - 5
DO NOT REMOVE
SHUTTER
RADIATION HAZARD
ME11507B
Safety Precaution - 6
ME11507B
Safety Precaution - 7
ME11507B
Safety Precaution - 8
HOT
KEEP AWAY
FROM FURNACE
ME11507B
Safety Precaution - 9
9. Equipment Transfer/Relocation
While the equipment is transferred or relocated, it may be adversely affected by vibration or other impact.
When transferring or relocating the equipment, consult your nearest Rigaku office.
ME11507B
Safety Precaution - 10
ME11507B
Safety Precaution - 11
Mechanism
(1) The X-ray tube outer wall is made of heavy metal.
(2) The solenoid shutter shield section consists of a
main shutter element (heavy metal, over3.5 mm)
and an auxiliary shutter element (heavy metal,
over3.5 mm). (Except the rotary shutter)
(3) When the solenoid shutter is open, it is indicated by
the X-ray shutter red lamp.
(4) When the shutter open indicator lamp system is
faulty due to lamp burnout, the solenoid shutter
does not remain open (opens for a moment and then
closes).
(5) The main shutter element is placed under the remote
control of a computer.
Function
Complete shielding
Same as above
Warning
Fail-safe operation
(6) When the mechanical connection to the goniometer Safety assurance and
or other external X-ray utilization section is lost, the radiation exposure
auxiliary shutter element is closed by its internal
possibility minimization
spring. In this instance, the main shutter element Same as above
also closes. (Except the rotary shutter)
Radiation enclosure
(7) X-ray generation is indicated by the red warning
Warning
lamp on the radiation enclosure.
(8) X-ray generation is unachievable when the X-ray
Fail-safe operation
generation indicator lamp is burned out.
(9) The goniometer and other mechanical operation
Shielding from scattered
sections are entirely covered by the radiation
X-rays
enclosure (box with a 3.2 mm thick iron lid).
(10) The radiation enclosure side wall positioned in the Shielding from possible
direct X-ray radiation direction is provided with a direct X-rays
2.5 mm thick, large lead plate.
(11) The X-ray shutter does not open if the radiation
Fail-safe operation
enclosure door is open.
(12) X-ray generation is unachievable when the warning Fail-safe operation
lamp described in paragraph (7) above is burned
out.
SAFETY RELEASE unit (13) Key input is needed to disable the safety feature.
Indication of the
When the safety feature is disabled, the SAFETY possibility of radiation
RELEASE unit red LED blinks and the alarm
exposure
buzzer intermittently beeps, indicating that you can
manually open and close the X-ray shutter.
(14) When you press the DOOR OPEN button, its
Shielding from X-rays in
signal is processed by the computer, and the yellow the event of inadvertent
lamp blinks at the operation panel to indicate that door opening
the door can be opened.
NOTICE :The safe mechanisms remarkably reduce the possibility of radiation exposure. However,
they do not guarantee that you are completely safe from radiation exposure in all sorts of
operations.
ME11507B
Copyrights
1.
Duplication or reproduction of this manual in whole or in part is strictly prohibited, whether via
hardcopy or electronically. If copies must be made, you must obtain our written approval before
doing so for each specific case.
2.
No part of this manual may be disclosed to third parties. If the contents of this manual must be
disclosed to a third party, you must obtain our written approval before doing so for each specific
case.
3.
No part of this manual may be cited without our permission. No part of this manual may be
translated or disclosed to a third party without permission. If you must cite or translate any part of
this manual, you must obtain our written approval before doing so for each specific case.
4.
In general, unless special agreements are reached, each product unit is provided with a copy of
this manual.
5.
Liability
1.
Rigaku shall not be held liable for any accidents caused by or resulting from any of the following.
* Use of the product for a purpose other than the purpose intended
* End of product life
* Unauthorized modifications
* Inadequate maintenance by the user
* Natural phenomena, armed conflicts, civil disturbances
* Use or action in breach of instructions given in this manual
* Installation conditions failing to meet the recommended ambient parameters
* Consumables
2.
Rigaku Corporation cannot be responsible for the results of using this manual to operate the
product or the effects of the results of such operation.
Relocation
Please contact Rigaku before attempting to move the product from the originally installed location.
Contents
Contents
1.
Overview.................................................................................................................................................... 1
2.
2.2
2.3
2.4
2.5
2.6
2.7
2.8
2.9
3.
4.
5.
2.11
2.12
2.13
Main unit........................................................................................................................................... 7
3.2
Goniometer ....................................................................................................................................... 9
3.3
3.4
3.5
3.6
3.7
Software Configuration.......................................................................................................................... 15
4.1
4.2
4.3
4.4
5.2
5.3
5.4
5.5
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
Contents
6.
7.
6.2
6.3
Optics ....................................................................................................................................................... 23
7.1
7.2
7.3
7.2.1
7.2.2
7.2.3
7.4
7.5
7.6
7.5.1
7.5.2
Attenuator ....................................................................................................................................... 50
7.8.1
8.
7.8
7.6.1
7.7
Detector ................................................................................................................................................... 53
8.1
Counter adaptor............................................................................................................................... 53
8.2
Scintillation counter........................................................................................................................ 55
8.3
9.
Attachments ............................................................................................................................................ 61
10.
ii
10.1.2
10.2
10.3
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
Contents
11.
Sample holders...................................................................................................................................... 69
11.1
11.2
11.3
12.
Accessories............................................................................................................................................. 71
13.
14.
15.
16.
13.1
13.2
Bulk samples................................................................................................................................. 74
13.3
13.4
Measurements ....................................................................................................................................... 75
14.1
14.2
14.3
Periodic Maintenance........................................................................................................................... 77
15.1
Optics maintenance....................................................................................................................... 77
15.2
Cooling water................................................................................................................................ 78
15.3
15.4
Target............................................................................................................................................ 78
15.5
Filament ........................................................................................................................................ 78
15.6
15.7
Troubleshooting.................................................................................................................................... 79
Appendix
iii
1.
Overview
This instrument is a thin film analysis system equipped with a high-accuracy theta-theta goniometer
featuring a horizontal sample mount. By changing slits (selection slits), the operator can use the
para-focusing optics to measure polycrystalline samples, the parallel beam optics incorporating a
multilayer mirror suitable for high-precision measurement to measure polycrystalline and thin film
samples, or the SAXS optics to measure nano-scale samples. A high-resolution optics system provided
with a 2- or 4-bounce monochromator on the incident side and a 2-bounce analyzer on the receiving
side can be easily installed simply by exchanging units. Adding an in-plane arm allows various
in-plane measurements. SmartLab is capable of performing a broad range of measurements required for
thin film analysis.
Soller slit
Para-focusin
g method
Soller slit
Soller slit
SmartLab
9-kW system
Parallel beam
Ni filter
2-bounce
monochromator
+
Soller slit
Soller slit
PSA
Soller slit
PSA
Soller slit
2-bounce analyzer
2-bounce analyzer
Micro area
In-plane
optics
Soller slit
Soller slit
Soller slit
PSC
PSA
Parallel beam
CBO unit
Fig. 1.1
Phase ID
analysis/
quantitative
analysis
Soller slit
Soller slit
2-bounce
monochromator
+
PSC
Diffracted beam
monochromator
Soller slit
Soller slit
4-bounce
monochromator
SAXS
Soller slit
Micro region
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
Characteristic x-rays
X-Rays are generated by accelerating electrons to very high speeds in a vacuum and directing
them against the anode (target). The x-ray spectra generated by electrons colliding against the
target can be divided into two categories: a continuous spectrum indicating continuous x-rays
(white x-rays) and a discrete spectrum for characteristic x-rays.
4000
Intensity
Characteristic
x-rays
X
2000
Continuous
x-rays
0
0
0.5
Wavelength
1 ()
1.5
Figure
The wavelengths of characteristic x-rays depend on the type of target used. Typical x-ray
diffractometry uses K x-rays generated by several types of metal targets, as shown in the
following table. K x-rays contain K1 x-rays and K2 x-rays whose wavelengths are quite
close. Although this does not pose serious problems for ordinary measurement of powder
samples for phase indentification (ID) analysis, optimal results can be achieved by using only
K1 x-rays in certain cases when making measurements for crystal structure analysis with
powder samples or when performing precise measurements of thin film samples. In recent
years, it has become possible to use just K1 x-rays by employing an incident optical system
comprised of a multilayer mirror and a Ge or Si monochromator crystal.
Table
Target
Element
Atomic
number
Cr
24
Fe
26
Co
27
Cu
29
Mo
42
Ag
47
W
74
Wavelength ()
K1
2.294
1.940
1.793
1.544
0.7135
0.5638
0.2138
2.290
1.936
1.789
1.541
0.7093
0.5594
0.2090
2.085
1.757
1.621
1.392
0.6323
0.4970
0.1844
2.1
2.
Product Features
2.1
2.2
2.3
X-ray generator
Even with a horizontal sample mount goniometer with a moving x-ray source, the product can
incorporate a state of the art high-intensity 9 kW rotating anode x-ray generator. When combined
with a multilayer mirror, this x-ray generator produces a high intensity x-ray beam (approx. 6 to 7
times the intensity provided by a sealed-tube system) equal to an 18-kW rotating anode x-ray
generator, while reducing power consumption by 50% due to the high-brightness focal spot. The
system also offers lower operating costs compared to earlier models. A 3 kW sealed tube x-ray
generator can also be used with the SmartLab system. The specifications of each generator are
summarized in the following table.
Table 2.1
2.4
Generator specifications
Max. load
Max. voltage
Max. current
Target metal
9 kW
45 kV
200 mA
Cu
3 kW
60 kV
50 mA
Cu
2.
Product Features
2.5
2.6
2.7
1.
2.
3.
2-bounce analyzer
Reflectivity measurements requiring high resolution, RSM measurements, and rocking curve
measurements, etc.
1.
2.
Standard incident optics unit to select Soller slit, 2-bounce monochromator (with Soller slit), or
4-bounce monochromator
3.
Standard incident slit box equipped with variable slit and length-limiting slit
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
2.8
2.8
2.9
1.
Standard receiving slit box # 1 for installation of a K x-ray filter for measurements using
para-focusing optics
2.
3.
Standard receiving optics unit # 2 for installation of parallel slit such as Soller slits
4.
5.
Type of analyzer
The control software checks whether the necessary optical devices are installed and displays a
message for the optical configurations suitable for the users intended application. The measurement
data file stores parameters for the optical devices during measurement to improve data
reproducibility and traceability.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
2.
Product Features
2.11
Control software
The software used to control the instrument also guides the user through required measurement
procedures and condition-setting processes, in addition to providing conventional instrument control
functions. Optical device configurations, optics alignment methods, sample alignment methods, and
measurement conditions specific to various measurement needs are grouped in units called Package
Measurements. By selecting an appropriate Package Measurement for the analysis purpose, the user
is guided through the procedures of optics alignment, sample alignment, and data measurement. The
program provides the optimal alignment and measurement conditions for the desired analysis.
Each Package Measurement was prepared by specialists with expertise in the specific field of
measurement methodology. Even users with limited experience with x-ray diffraction or x-ray
reflectivity measurements can perform measurements in the same way a specialist would. The
software also allows customization of alignment and measurement conditions for special
measurement needs. Manual control is also possible. The software is designed to meet a wide range
of user needs.
2.12
2.13
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
3.
Fig. 3.1
3.1
1. Rotary pump
2. Control PC
Main unit
4
3
Fig. 3.2
1. Main panel
2. Operating panel
3. Door
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
3.
9
5
10
7, 8
6
Fig. 3.3
The power supply and the control cable for the rotary
pump connect here.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
3.2
3.2
Goniometer
Goniometer
4
5
7
Fig. 3.4
Goniometer
1. Theta_s arm
2. X-ray tube
3. Incident optics
4. Theta_d arm
5. Receiving optics
6. Detector
X-ray detector.
7. Sample attachment
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
3.
Theta_s arm
Selection slit
PB, BB, SA, etc.
Theta_d arm
K filter
Receiving slit box # 2 (see section 7.7)
Receiving optical device adaptor/analyzer
PSA
PSA 0.114 deg., Ge(220)x2, etc.
Sample attachment
Sample plate
Height reference sample plate, 4-inch sample
plate, etc.
10
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
3.3
3.3
X-ray tube
X-ray tube
4
6
2
1
5
Fig. 3.5
2. Rotary target
3. Shutter
5. Ion gauge
1
2
Fig. 3.6
2.
X-ray tube
3. Shutter
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
11
3.
3.4
Theta_s arm
2
1
Fig. 3.7
12
Theta_s arm
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
3.5
3.5
Theta_d arm
Theta_d arm
3
6
Fig. 3.8
1
Theta_d arm
6. Attenuator
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
13
3.
3.6
Fig. 3.9
3.7
1. Counter adaptor
2. Base
2
3
Fig. 3.10
14
1. Attachment base
2. Attachment
3. Sample spacer
4. Sample plate
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
4.1
4.
Control software
Software Configuration
4.1
Control software
SmartLab Guidance
4.2
4.3
3D Explore (option)
4.4
JADE
NANO-Solver
analysis.
processing
QA-Tex
Visual RIETAN
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
15
Minimum
excitation
voltage
(kV)
Cu
8.86
Co
7.71
35 to 50
25 to 35
Fe
7.10
35 to 45
25 to 35
Cr
5.98
30 to 40
20 to 30
Optimum voltage
(equivalent load) (kV)
Intensity
P/B ratio at
at
maximum
maximum
40 to 55
25 to 35
5.1
5.
5.1
Fig. 5.1
Main panel
The red lamp goes on when the instrument is starting up or when an error occurs.
For detailed information, refer to section 16 in this manual.
5.2
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
17
5.
5.3
Fig. 5.2
Tip:
For more information on the startup procedure, refer to the SmartLab Guidance
Package Measurement Manual (ME13405A).
CAUTION: When not using the water circulation pump, make sure the
cooling water is flowing properly. If the cooling water is not
flowing or the amount of water is incorrect, an alarm will be
issued and remain active for three minutes after the startup
procedure is initiated. Adjust the water volume to shut off
the alarm within three minutes, then repeat the startup
procedure.
18
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
5.4
5.4
Fig. 5.3
Tip:
When not using the instrument, turn off the x-ray generation without shutting off the exhaust control.
This continues to supply power to the controller so that the x-ray generator can be started (e.g., for
aging) from the PC.
To completely shut down the power supply when not using the instrument for extended periods,
power off the instrument.
Tip:
CAUTION: When not using the water circulation pump, wait three
minutes following shutdown (i.e., three minutes after the red
lamp located at the upper section of the instrument goes
out), then halt the supply of cooling water. (Halting the
cooling water supply in less than three minutes may result in
instrument damage.)
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
19
5.
5.5
Fig. 5.4
Main panel
This procedure completely shuts down the instrument internal power supply.
Only the LINE lamp will remain lit. This lamp indicates that power is being
supplied to the main unit.
20
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
6.1
6.
6.1
Fig. 6.1
Operating panel
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
21
6.
6.2
6.3
Fig. 6.2
22
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.1
7.
Optics
This section introduces the names of parts and describes how they are installed.
7.1
Fig. 7.1
CBO unit
(1) Attach the CBO unit to the shutter. Press the CBO unit against the surface in the back and
secure the unit in place with the Allen wrench provided.
Fig. 7.2
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
23
7.
Optics
(2) Insert the cable into the CBO connector of the incident connector box.
Fig. 7.3
Cable connection
(3) Insert a selection slit into the slit box of the CBO unit.
Fig. 7.4
24
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.1
Table 7.1
Tip:
Selection slits
Name of optics
Abbreviation
For para-focusing
BB
PB
For small-angle
scattering
measurements (option)
SA
MA
For small-angle
scattering
high-intensity
measurements (option)
SA01
Illustration
The type of selection slit installed can be identified from the PC.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
25
7.
Optics
7.2
7.2.1
Fig. 7.5
(1)
Install the incident parallel slit adaptor to the CBO unit by sliding it down from the top
section of the CBO unit. Secure it in place with the Allen wrench provided.
Fig. 7.6
26
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.2
(2)
Connect the cable to the MONO/ADPT connector on the incident connector box.
Fig. 7.7
(3)
Cable connection
Install the parallel slit on the adaptor and secure it in place with the Allen wrench provided.
Fig. 7.8
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
27
7.
Optics
Table 7.2
Type
Open
Divergence
limit angle
Abbreviation
None
Soller slit
Soller slit
2.5
1.0
Illustration
(option)
(option)
In-plane PSC
(parallel slit collimator)
deg
(option)
In-plane PSC
0.5
(option)
In-plane PSC
(option)
Tip:
28
0.15
The type of installed incident parallel slit can be identified from the PC.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.2
7.2.2
Fig. 7.9
(1)
2-Bounce monochromator
Attach the 2-bounce monochromator to the CBO unit by sliding it down from the top
section of the CBO unit. Secure it in place with the Allen wrench provided.
Fig. 7.10
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
29
7.
Optics
(2)
Connect the cable to the MONO/ADPT connector on the incident connector box.
Fig. 7.11
(3)
Install a parallel slit on the 2-bounce monochromator. Secure it in place with the Allen
wrench provided.
Fig. 7.12
30
Cable connection
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.2
Table 7.3
2-bounce monochromators
Diffraction
Crystal
plane
Ge
Abbreviation
220
Ge (220) x 2
400
Ge (400) x 2
Illustration
(option)
Ge
(option)
Tip:
Table 7.4
Type
Divergence
Abbreviation
Open
None
5.0
2.5
0.5
Illustration
Soller slit
(option)
Soller slit
(option)
In-plane PSC
(option)
Tip:
deg
The type of parallel slit installed on the adaptor can be identified from the PC.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
31
7.
Optics
7.2.3
Fig. 7.13
(1)
Attach the 4-bounce monochromator to the CBO unit by sliding it down from the top
section of the CBO unit. Secure it in place with the Allen wrench provided.
Fig. 7.14
32
4-bounce monochromator
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.2
(2)
Connect the cable to the MONO/ADPT connector on the incident connector box.
Fig. 7.15
Table 7.5
Crystal
Ge
4-bounce monochromators
Diffraction
plane
Cable connection
Abbreviation
220
Ge (220) x 4
440
Ge (440) x 4
Illustration
(option)
Ge
(option)
Tip:
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
33
7.
Optics
7.3
This variable slit box is installed on the Zs axis on the theta_s arm.
7.3.1
Fig. 7.16
(1)
Attach the standard incident slit box to the Zs axis. Secure it in place with the Allen wrench
provided.
Fig. 7.17
34
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.3
(2)
Connect the cable to the SLIT connector on the incident connector box.
Fig. 7.18
(3)
Cable connection
Insert a length-limiting slit to control beam length along the longitudinal axis.
Fig. 7.19
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
35
7.
Optics
Table 7.6
Tip:
36
Length-limiting slits
Length of limiting
slit
Abbreviation
15 mm
15
10 mm
10
5 mm
2 mm
0.5 mm for
micro area optics
(option)
0.5
Illustration
The type of inserted length-limiting slit can be identified from the PC.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.4
7.4
7.4.1
Fig. 7.20
(1)
Attach the standard receiving slit box # 1 to the rail of the theta_d arm. Secure it in place
with the Allen wrench provided.
Fig. 7.21
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
37
7.
Optics
(2)
Connect the cable to the SLIT1 connector on the receiving connector box.
Fig. 7.22
(3)
Fig. 7.23
Tip:
38
Cable connection
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.4
Fig. 7.24
Tip:
(4)
The installation position can be read from the scale on the theta_d arm.
Fig. 7.25
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
39
7.
Optics
7.5
7.5.1
Fig. 7.26
(1)
Attach the receiving optical device adaptor to the rail of the theta_d arm. Secure it in place
with the Allen wrench provided.
Fig. 7.27
40
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.5
(2)
Connect the cable to the ADPT1 connector on the receiving connector box.
Fig. 7.28
(3)
Cable connection
Attach a parallel slit analyzer to the adaptor. Secure it in place with the Allen wrench
provided.
Fig. 7.29
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
41
7.
Optics
Table 7.7
Aperture angle
Length
Abbreviation
Open
45 mm
PSA open
1.0
(option)
45 mm
0.5
45 mm
0.114
(option)
90 mm
PSA 0.114
Open
90 mm
Illustration
deg
PSA open
Tip: The type of parallel slit analyzer installed on the adaptor can be identified from
the PC.
42
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.5
7.5.2
Fig. 7.30
(1)
2-bounce analyzer
Attach the 2-bounce analyzer to the theta_d arm. Secure it in place with the Allen wrench
provided.
Fig. 7.31
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
43
7.
Optics
(2)
Connect the cable to the ADPT1 connector on the receiving connector box.
Fig. 7.32
Cable connection
Table 7.8
Crystal
Ge
Diffraction
plane
Crystal types
Abbreviation
220
Ge (220) x 2
400
Ge (400) x 2
Illustration
(option)
Ge
(option)
Tip: The type of crystal currently installed can be identified from the PC.
44
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.6
7.6
7.6.1
Fig. 7.33
(1)
Attach the receiving parallel slit adaptor to the rail of the theta_d arm. Secure it in place
with the Allen wrench provided.
Fig. 7.34
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
45
7.
Optics
(2)
Connect the cable to the ADPT2 connector on the receiving connector box.
Fig. 7.35
(3)
Attach a parallel slit to the adaptor. Secure it in place with the Allen wrench provided.
Fig. 7.36
46
Cable connection
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.6
Table 7.9
Type
Aperture
angle
Abbreviation
Soller slit
5.0
Soller slit
(option)
2.5
In-plane PSA
(option)
1.0
In-plane PSA
(use of PSA 0.5 deg.)
0.5
In-plane PSA
(use of PSA 0.114
deg.)
0.114
Tip:
Illustration
The type of parallel slit currently installed on the adaptor can be identified from
the PC.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
47
7.
Optics
7.7
7.7.1
Fig. 7.37
(1)
Attach the standard receiving slit box # 2 to the rail of the theta_d arm. Secure it in place
with the Allen wrench provided.
Fig. 7.38
48
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.7
(2)
Connect the cable to the SLIT2 connector on the receiving connector box.
Fig. 7.39
(3)
Cable connection
The installation position can be read from the scale on the theta_d arm.
Tip:
When using the para-focusing method, set the installation position to 300 mm.
Fig. 7.40
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
49
7.
Optics
7.8
Attenuator
The attenuator adjusts the intensity of the detected x-rays.
7.8.1
Standard attenuator
The system uses several types of attenuators on a rotating disc. Attenuators for optics alignment
are also built-in.
The type of attenuator can be switched from the PC. Attenuators can be switched automatically
for various measurements based on detected intensity.
Fig. 7.41
(1)
Attach the standard attenuator to the rail of the theta_d arm. Secure it in place with the
Allen wrench provided.
Fig. 7.42
50
Standard attenuator
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
7.8
(2)
Attenuator
Connect the cable to the ATT connector on the receiving connector box.
Fig. 7.43
Tip:
Cable connection
The attenuator types used during actual measurements are as follows: Open,
1/70, 1/1000, 1/10000. The attenuator types used during alignments are as
follows: 9 kWBB, 9 kWPB, 3 kWBB, 3 kWPB for alignment.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
51
K filter
The characteristic x-rays used for x-ray diffractometry generally contain K and K x-rays. A
substance that passes K x-rays while absorbing most K x-rays to monochromatize the beam
is called a K filter. The mass absorption coefficient (hereafter referred to as absorption
coefficient) of an element becomes smaller as wavelengths become shorter, and vice versa.
However, when x-rays with energy greater than the binding energy between the nucleus and
electrons of an element, strike the element, the absorption coefficient suddenly increases due to
the photoelectric effect. This discontinuous point in the absorption coefficient is called an
absorption edge.
LI absorption
edge
LII absorption edge
LIII absorption
edge
K absorption
edge
Wavelength ()
Figure
For efficient absorption of K x-rays only, select an element with its absorption edge
wavelength located between the K x-ray and K x-ray wavelengths. Those elements will
have an atomic number one or two less than that of the target element. The following table
shows the target elements commonly used for x-ray diffractometry and their corresponding
filters.
Table
Target
Wavelength ()
K1
K1
K filters
Filter
Material
Wavelength of
absorption edge
()
Thickness
(mm)
When I K I K = 1 / 100
Thickness
K1
transmissivity
(g/cm2)
1
Cr
2.290
2.085
2.269
0.011
0.007
63
Fe
1.936
1.757
Mn
1.896
0.011
0.008
62
Co
1.789
1.621
Fe
1.743
0.012
0.009
61
Cu
1.541
1.392
Ni
1.488
0.015
0.013
55
Mo
0.7093
0.6323
Zr
0.689
0.081
0.053
43
Ag
0.5594
0.4970
Rh
0.534
0.062
0.077
41
8.1
8.
Counter adaptor
Detector
8.1
Counter adaptor
This adaptor is used to install the detector (scintillation counter).
Fig. 8.1
Counter adaptor
(1) Attach the counter adaptor to the rail of the theta_d arm. Secure it in place with the Allen
wrench provided.
Fig. 8.2
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
53
8.
Detector
(2) Connect the cable to the COUNTER connector on the receiving connector box.
Fig. 8.3
54
Cable connection
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
8.2
8.2
Scintillation counter
Scintillation counter
The scintillation counter is attached to the counter adaptor. It detects x-rays received by the
receiving optics.
Determine whether the scintillation counter is installed from the PC.
Fig. 8.4
Scintillation counter
(1) When detecting x-rays from the receiving optics directly (in absence of the diffracted beam
monochromator), install the scintillation counter to the base with the Allen wrench provided.
Fig. 8.5
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
55
8.
Detector
(2) Attach the base to the counter adaptor. Secure it in place with the Allen wrench provided.
Fig. 8.6
(3) The installation position can be read from the scale on the theta_d arm.
Fig. 8.7
56
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
8.3
8.3
Fig. 8.8
(1) Install the scintillation counter on the diffracted beam monochromator unit. Secure it in place
with the Allen wrench provided.
Fig. 8.9
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
57
8.
Detector
(2) Install the unit on the counter adaptor. Secure it in place with the tool provided.
Fig. 8.10
(3) The unit installation position can be read from the scale on the theta_d arm.
Fig. 8.11
58
Installation position
351.5 mm
359.8 mm
364.7 mm
376.1 mm
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
8.3
Fig. 8.12
CAUTION: Use the RSm only when using the monochromator in Bent
mode for the para-focusing method. The RSm is not used
when using the monochromator in Flat mode.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
59
Table
Type of
crystal
Resolution
(FWHM)
Ge(220)x2
crystal
0.008<
Ge(400)x2
crystal
0.0022
(when sample d
spacing is
1.41A)
Ge(220)x4
crystals
0.003<
Ge(440)x4
crystals
0.0015<
Intensity
[unspecified unit]
[]
Slit
collimation
Ge(220)2
Ge(220)2 crystal
Ge(220)4
Ge(220)4
crystal
Zoom
view
Cu K
W-L
61.00
63.00
69.00
65.00
69.10
69.20
67.00
69.00
2/ [deg]
Figure
9.
Attachments
You can switch between attachments with various adjust axes for different measurement purposes.
Pin
Fig. 9.1
(4) Make sure the right and left clasps are open.
(5) Install the attachment by aligning the groove of the attachment with the pin on the front.
(6) Align the blades on both sides.
(7) Close the clasps and engage the front hook.
Fig. 9.2
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
61
9.
Attachments
Table 9.1
Name
Types of attachments
Description
Standard attachment
Illustration
RxRy attachment
(option)
XY-20 attachment
(option)
Tip:
62
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
10.1
10.
Sample plates
Select a sample plate based on the sample to be measured.
10.1
Fig. 10.1
CAUTION: Placing a sample larger than the sample plate may damage
the sample during measurement.
Magnets supplied with the product are generally used to secure the sample in place. Refer to 13.
Examples of sample mounting.
Fig. 10.2
Tip: The sample holding magnets can be used for samples measuring up to
approximately 3 mm in thickness.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
63
10.
Sample plates
The Z axis is used to align the sample surface with the center of the x-ray beam. Since the thickness
range for which the Z axis can be adjusted is 3 mm, you must select the appropriate sample spacer
based on the thickness of the sample to be measured.
Table 10.1
Sample spacers
Sample thickness
Abbreviation
0-3 mm
0-3 mm
3-6 mm
3-6 mm
Illustration
(option)
6-9 mm
6-9 mm
(option)
9-12 mm
9-12 mm
(option)
0-3 mm (blue
0-3 mm (*1)
stamped mark)
12-15 mm
12-15 mm
(option)
3-6 mm (blue
3-6 mm (*1)
stamped mark)
15 mm-18 mm
15-18 mm
(option)
6-9 mm (blue
6-9 mm (*1)
stamped mark)
*1: Sample thickness for using XY-4 attachment. Refer to the abbreviations in blue.
64
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
10.1
Table 10.2
Name
Illustration
Attachment
Sample spacer
0-3 mm
3-6 mm
6-9 mm
Standard attachment
9-12 mm
RxRy attachment
12-15 mm
XY-20 mm
attachment
Sample plate
Measurable
sample thickness
0-3 mm
3-6 mm
plate(*1)
6-9 mm
9-12 mm
plate(*1)
12-15 mm
18-21 mm
21-24 mm
(blue stamped
0-3 mm
3-6 mm
6-9 mm
mark)(*2)
4-inch area mapping sample plate (for 9-12 mm)
9-12 mm
12-15 mm
*1: Both X and Y must be 0 mm when the sample is placed on the XY-20 mm attachment.
*2: When the XY-4 attachment is used, the thickness indicated by the blue stamp mark on the
sample spacer is valid.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
65
10.
Sample plates
10.1.1
Fig. 10.3
To remove the sample spacer from the attachment, apply pressure to the bar indicated by the
arrow in Fig. 10.4, then turn the sample spacer counterclockwise.
Fig. 10.4
66
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
10.1
10.1.2
Fig. 10.5
Tip: Refer to 10.1.1 Installing and removing sample spacers for information on
direct mounting (removing) a wafer sample plate to (from) an attachment.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
67
10.
Sample plates
10.2
Fig. 10.6
10.3
Fig. 10.7
Tip:
Although the height reference sample plate and transmission SAXS sample plate
can be used on top of other attachments, it is recommended for maximum
precision that they be used attached to the standard attachment. If using the
height reference sample plate or transmission SAXS sample plate with an
attachment other than the standard attachment, you must set each attachment axis
to 0. Optics alignments, however, must be done with the height reference sample
plate attached to the standard attachment. Install/remove sample plates in the
same way as wafer sample plates.
68
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
11.1
11.
Sample holders
Use a sample holder when measuring samples that require preparation, such as powder and bulk
samples.
Tip:
11.1
Fig. 11.1
11.2
Fig. 11.2
11.3
Fig. 11.3
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
69
dI D = 1 exp 2x
Gx = 0
sin
dI
0 D
Substitute the values for angle of the incident x-ray beam and x-ray absorption coefficient
(/cm) in the following formula to obtain x cm, the effective thickness.
x=
where,
K x = ln (1 G x )
50
75
90
95
99
99.5
0.69
1.39
2.30
3.00
4.61
6.91
The following table shows the results of calculations of the sample thicknesses x cm that give
a diffracted x-ray usage rate of 90% when Kx = 2.30 and = 90.
Substance
Al2O3
Si
SiO2
CaCO3
Powder
Fe2O3
ZnO
Ag2O
(tetragonal)
PbO I(
Al
Fe
Metal
SUS(18-8)
Cu
Pb
Polyethylene
Polymer
Vinyl chloride
Single crystal Si
substance
SiO2
Density
g/cm )
3.97/2
2.42/2
2.65/2
2.71/2
5.26/2
5.60/2
7.22/2
9.53/2
2.70
7.86
7.93
8.92
11.34
0.93
1.41
2.42
2.65
Mass/cm
absorption coefficient / /g)
(cm /g)
Thickness
for 90% utilization rate (mm)
90%mm)
MoK CuK CoK CrK MoK CuK CoK CrK
3.34
31.1
48.0
97.6
1.74
0.187
0.121
0.060
6.44
60.6
93.3
189
1.48
0.157
0.102
0.050
3.71
34.5
53.1
108
2.34
0.252
0.164
0.080
8.02
70.9
107
205
1.06
0.120
0.079
0.041
27.3
216
42.7
86.5
0.160
0.020
0.103
0.051
44.7
50.7
78.2
159
0.092
0.081
0.053
0.026
24.1
204
300
548
0.132
0.016
0.011 0.0058
112
216
311
540
0.022
0.011 0.0078 0.0089
5.16
48.6
74.8
152
0.826
0.088
0.057
0.028
38.5
308
52.8
108
0.038 0.0048
0.028
0.014
37.9
278
110
106
0.038 0.0052
0.013
0.014
50.9
52.9
81.6
166
0.025
0.024
0.016 0.0078
120
232
334
579 0.0085 0.0044 0.0030 0.0018
0.59
4.0
6.1
12.5
21.0
3.09
2.02
0.989
6.73
62.0
94.2
186.2
1.21
0.132
0.086
0.044
6.44
60.6
93.3
189
0.738
0.079
0.051
0.025
3.71
34.5
53.1
108
1.17
0.126
0.082
0.040
2
12.
Accessories
Table 12.1
Name
Accessories
Illustration
Center slit
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
71
(BB)
Selection
slit (BB)
X source
X-Ray
(filament)
Sample
Optical
pathA
A (divergence beam)
(PB)
Selection
slit (PB)
X
X-Ray
source
(filament)
Sample
Optical
path B (parallel beam)
B
Fig.
13.
13.1
Wafer-shaped samples
Use a 4-inch wafer sample plate, 6-inch wafer sample plate, or 8-inch wafer sample plate to measure
wafer-shaped samples.
Fig. 13.1
Tip:
Wafer-shaped sample
Place the sample at the center of the sample plate. Use the magnets provided to hold the sample in
place. The magnets can be used with samples that are up to 3 mm thick. For information on
measuring samples thicker than 3 mm, refer to 13.2 Bulk samples.
CAUTION: Placing a sample larger than the sample plate may result in
damage to the sample during measurement.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
73
13.
13.2
Bulk samples
Use a 4-inchi wafer sample plate, 6-inch wafer sample plate, or 8-inch wafer sample plate to
measure bulk samples. Select the appropriate sample spacer and sample plate based on sample
thickness. For detailed information, refer to 10.1 Wafer sample plates and sample spacers.
Place the sample so that the sample surface to be measured is parallel to the sample plate. A piece of
clay may be used to secure the sample in place. If the bulk sample is small enough to pass through
the window on the provided aluminum sample holder, you can secure the sample in the aluminum
sample holder and use the height reference sample plate.
Fig. 13.2
13.3
Powder samples
Place powder samples in the indentation on the glass sample holder or aluminum sample holder,
then insert into the height reference sample plate.
Fig. 13.3
13.4
Powder sample
74
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
14.1
14.
Measurements
This section introduces basic operating procedures for typical measurements. For detailed information,
refer to the SmartLab Guidance Reference Manuals, Parts Manuals, and Package Measurement
Manuals.
14.1
14.2
Fig. 14.1
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
75
14.
Measurements
Fig. 14.2
(3) Set the position of each optical device. To change the optical devices of the receiving optics,
inspect the installation positions of the standard receiving slit box # 1, standard receiving slit
box # 2, and scintillation counter and enter the correct corresponding values.
14.3
Fig. 14.3
For information on measurement procedures for specific Package Measurements, refer to the
SmartLab Guidance Package Measurement Manual (ME13406A).
76
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
15.1
15.
Optics maintenance
Periodic Maintenance
Perform the following maintenance tasks at regular intervals to maintain the performance of the
instrument.
15.1
Optics maintenance
Multilayer mirror
Attenuator correction
HV/PHA adjustment
Dead-time correction
The characteristics of the detector will change over time. Replacement of the detector is required
when it reaches the end of its service life. However, you can minimize performance degradation by
making the appropriate adjustments.
We recommend performing HV/PHA adjustments and dead-time correction measurement at periodic
intervals.
To make these adjustments automatically, select Package Measurement from the Task menu, then
select Maintenance Package measurement under Utility. For detailed information on measurement
procedures, refer to the Maintenance Package measurement section of the SmartLab Guidance
Maintenance Package Measurement Manual (ME13405A).
Fig. 15.1
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
77
15.
Periodic Maintenance
15.2
Cooling water
Check the amount of cooling water when using the water circulation pump. If the water level is low,
add water. Periodic changes of cooling water are also recommended.
15.3
Sections 15.4 15.7 are maintenance required only for a 9 kW rotating anode x-ray generator.
15.4
Target
The target metal rotates at high speed to boost x-ray output. Pay close attention and note any
abnormal noise or vibrations.
When contaminant (tungsten from the filament) from the electron gun (filament) adheres to the
target metal surface, you may observe characteristic x-rays from the contamination, in addition to
characteristic x-rays from the target metal. If this happens, maintenance is required. Please contact
our service personnel.
15.5
Filament
As electrons are discharged, the filament will become thinner, eventually breaking the required
circuit. Upon breaking, the filament will release a sudden burst of electrons that can damage the
target surface. To avoid this, we recommend periodically changing the filament.
15.6
15.7
Rotary pump
Check the oil level in the vacuum pump. If the oil level is low, add oil. We also recommend periodic
oil changes.
78
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
16.
Troubleshooting
Problem
Remedy
The instrument may not be receiving electric power. Check the main
circuit breaker located on the rear panel of the instrument. Also check
the connections for the power cable and circuit breaker on the supply
side.
The OPERATE lamp does not The controller may be in the process of starting up. Wait several
change color to green (remains minutes.
red).
If the lamp fails to turn green after several minutes, the controller may
be malfunctioning. Press the OFF button on the main panel to shut
down the instrument, then press the ON button to restart the
instrument.
An
error
occurs
during
SmartLab Guidance startup.
The
instrument
startup
procedure fails to complete
properly.
The instrument stops abruptly Make sure the OPERATE lamp on the main panel is red. The
during measurement or other instrument may have stopped due to excessive force applied to the
operation.
theta_s or theta_d arm. Press the OFF button on the main panel to stop
the instrument. Check the arms to see if any item obstructs their
operation, then restart the instrument.
The x-ray generator stops during Click XG Control in the Control menu to open the XG Control
measurement
or
other dialog box. Click Display in the Alarm section to open the Alarm
operations.
dialog box. Check for x-ray generator alarms.
A description of each alarm is displayed in the XG status description
section when the LED in the dialog box is clicked.
* If a problem other than those described above occurs or if the problem cannot be corrected by taking the
countermeasure described, please contact our service personnel.
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
79
Appendix
Appendix
The OPERATE lamp indicates the status of the SmartLab controller. The lamp is usually green, but
will change to red and begin flashing at set intervals in the event of an abnormality. To reset the
generated alarm, you must restart the instrument. After shutting down the control software (SmartLab
Guidance), turn off power to the instrument, then press the ON button to restart the instrument.
If an alarm is generated, stop, then restart the instrument. If the OPERATE lamp lights up in green in
10 to 20 seconds after the restart, the instrument may be used as usual. If the OPERATE lamp does not
turn green, flashing red instead after startup, or if the lamp turns green but changes to a flashing red
immediately during use, please contact our service personnel. Before contacting our service personnel,
please refer to Table 2 to identify the alarm being generated.
Table 1
Condition
Immediately
after Power
ON
Color
Continuously
On / flashing
Green
Flashing
250 ms
Green
Accessible
from PC
Green
100 ms
Flashing
100 ms
Continuously
On
Table 2
Controller
startup
failure
Controller
hung-up
Pattern
250 ms
10 seconds
after Power
ON
Cause of
alarm
Instrument startup
Color
Alarm generation
Continuously
On / flashing
Pattern
500 ms
Red
100 ms
Flashing
100 ms
500 ms
Red
500 ms
100 ms
Flashing
100 ms
500 ms
500 ms
Servo alarm
Red
Flashing
500 ms
Servo
overload
2000 ms
Red
Flashing
2000 ms
80
SmartLab Horizontal Sample Mount X-Ray Diffractometer for Thin Film Analysis
t
sli
on
i
t
c
el e
MA
L
g
en
th
itin
lim
gs
Standard configuration
lit
0.
)
mm
5(
SA
2,
0
,1
PB
Ge
5
,1
m
(m
S
IP
em
nc
on
ad
r
p to
Inpla
0.1 ne P
5 d SC
eg
r
4
at o
0)x
rom e(44
h
c
G
ou
4- b 4
0)x
(2 2
Inpla
n
1.0 e P
d e SC
g
BB
So
lle
rs
ilt
5.0
de
g
So
lle
rs
lit
2.5
de
g
at
on
2
tor 0)x
m a (2 2
o
r
e
h
G
oc
90
ou
2-b )x2
00
e( 4
mm
- 2 0 en t
XY chm
a
att
nt
rd t
da
an m en
t
S ch
a
att
D
RO
ad
ap
ple
s
Al
am
pl
e
old
eh
Ce
r
nt e
t
sl i
Inpla
0.1 ne P
14
de SA
g
t or
Sc
0 . 1 PS A
14
de
g
He
ig
sa h t r e
mp fer
e
le
pla nce
te
Tra
ns
s a mi s s
m p i on
le
h o SA X
ld e S
r
(3-
int
illa
tio
n
co
un
tor
ap
In-
p la
n
1 .0 e PS
de
g A
So
lle
2.5 r slit
de
g
DB
M
P
1.0 SA
de
g
6m
m
for , 69m S a
4
wa m, mple
9
fer
sa -12m spa
ce
mp
m,
r
le
(fo 12-1
r (1 5 m
8m m,
1
m21 5-18
mm m
) )) m
Inpla
0.5 ne P
de SA
g
So
ll e
5.0 r slit
de
g
ter
P
0.5 SA
de
g
Tra
ns
m
sa issi
mp on
le
pla SAX
te S
r
de ple
ow s am
p
Si ce
n
er e
f
re
am
ss r
as olde
l
G h
ce
er
lyz
Sa
RP
-4 nt
XY hme
ac
att
un
a
an
ta
be
Ku
C
ref Si w
ere
a
nc fer
es
am
ple
S
DH
rad
le
So
lle
r
op slit
en
Detector adaptor
Ry
Rx
Inpla
ne
PS
0.5
C
de
g
me
ard
o
2-b
em
nc
h
t ac
nd
Ge(400)x2
St a
So
ll
5 . 0 er s l
d e it
g
Ge(220)x2
So
lle
rs
ilt
op
en
So
ll
2.5 er sli
de t
g
un
it
PS
op A
en
PS
op A
en
Va
cu
um
8 wafer
sample plate
pa
th
Sample spacer
0-3 mm
4 wafer
sample plate
6 wafer
sample plate
Ver. 1.3
LINE OF BUSINESS
X-ray Equipment for Science and Industry
X-ray Diffractometer and Attachments
X-ray Generators
X-ray Diffraction Cameras
X-ray Spectrometers
Piezo Goniometers
X-ray Stress Analyzers
Industrial X-ray Radiographic Equipment
Vacuum Rotary Feedthrough
Thermomechanical Analyzers
Differential Scanning Calorimeters
Specifications Subject to Change without Notice
Rigaku Corporation
Tokyo Branch
Rigaku Corporation
Head Office
3-9-12, Matsubara-cho, Akishima-shi,
Tokyo 196-8666, Japan
Phone: 81-42-545-8189 Fax: 81-42-544-9223