Beruflich Dokumente
Kultur Dokumente
Contents Page
Introduction
Theoretical Background
Experimental Procedure
16
Discussion
22
Error Analysis
24
25
26
Conclusion
27
References
28
29
Introduction
1 page (jia Ming)
2.Theoretical Background
2.1 Packed Bed
The packed bed essentially consists of a vertical cylinder that is packed with packing
material. In most cases, liquids flow downwards from the top of the bed due to gravitational
forces while the vapors/gases travel upwards through contacting with the wetted packing
which is present to increase the surface area for reactions to take place. Optimum recovery
of the solute is achieved when the liquids present are non-volatile and the gases do not
react with the liquids present.
Packed towers are mostly utilized in stripping/ gas-liquid absorption processes as they are
relatively cheaper as compared to other processes such as and because it is very feasible to
change the geometrical parameters of the bed such as the height if the bed, the type of
packing present and such. However, the drawbacks of such processes are that although
efficient stripping occurs at low flow rates, flooding can also occur at higher flow rates and
this may hinder the performance of the bed. In addition, it is relatively difficult to allow
efficient heat exchange with the surroundings; it would not be possible to use this process if
the temperature of either the gas or liquid is too high. The dissolving capacity of the gas in
the liquid also decreases with increasing temperature.
In this particular experiment, the packing used is the Raschig rings that are small cylindrical
structures and hollow. They are inert, affordable and most importantly increase the surface
area for chemical reactions and absorption to take place. The liquid and the gas are
introduced in the counter directions to maintain the concentration gradient between both
streams to increase the driving force of mass transfer.
The stripping process in the packed beds can be improved by allowing a chemical reaction to
occur concurrently to reduce mass transfer resistance and this enhancement of the process
is often referred to as chemisorption.
For instance, in this experiment, the following reaction takes place to improve the stripping
of CO2 gas:
CO2 + 2NaOH
Na2CO3 + H2O
Therefore, there are two processes occurring concurrently. The carbon dioxide from the gas
steam is reacting with sodium hydroxide as well as being absorbed by diffusion into the
liquid stream. The latter process is taken to be instantaneous.
2.2 Different Zones in a packed column
In every packed bed in which counter current absorption is taking place together with a
chemical reaction, three zones of absorption can be identified. They are the surface reaction
zone, the interior reaction zone and the physical absorption zone.
Lin
Absorpti
on Zone
Gout
CAbsorpti
on Zone
XB4
YA4
H3
YA3
XB3
H2
YA2
H1
XB2=0
G, YA1
L, XB1=0
Where
Y: mole ratio of solute in gaseous phase
X: mole fraction of solute in liquid phase
H: height of each zone
A: Carbon Dioxide Gas
B: Sodium Hydroxide
G: Gas phase
L: Liquid phase
2.2.1 Surface Reaction Zone
This zone is present in the uppermost section identified in the column where the
concentration of liquid sodium hydroxide is the highest as compared to the concentration of
carbon dioxide gas that is the lowest. Thus, the reaction takes place at the gas-liquid
interface and as mentioned before, is taken to be instantaneous. An assumption is made
here that sodium hydroxide is non-volatile and that increasing its concentration does not
affect the rate of diffusion significantly. The rate of the reaction is governed by the mass
transfer rate of carbon dioxide from the air because the mass transfer resistance is usually
the highest in any gas phase.
Differing concentrations of
NaOH
or
Defining
or
,
G A3 dYA
G
h3
a YA 4 rAs k GA aPT
(1 YA )dYA
G Y A3
(YA3 YA4 )
ln
YA
k GA aPT YA4
YA 4
YA 3
Y
G
ln A 3
k gA aP
YA 4
Legend:
FA:
xA:
Ax:
h3 :
YA :
G:
PA: :
PT:
kGA :
a:
fractional conversion of A
cross sectional area of the reactor (m2)
height of surface reaction zone (m)
mole ratio of reactant A to inert in gas bulk
molar flow rate of inert gas (mol/(m2.s))
partial pressure of gas (N/ m2)
total pressure of gas (N/ m2)
mass
transfer
coefficient
of
A
3
with chemical reaction (mol/m s.Pa)
interfacial area per volume packing (m-1)
in
gas
Reaction locus
Gas-Liquid interface
The interior reaction zone can be found in the middle of the packed column. Since the
location of this section is lower than the section observed in the previous part, it can be
observed that the sodium hydroxide concentration decreases while the carbon dioxide
concentration increases. This change in concentrations is the reason as to why the reaction
locus shifts into the liquid film as shown above. The reaction locus will continue to shift
deeper into the liquid film until the concentration of sodium hydroxide reaches zero but will
not move into the liquid bulk. In this section, the mass transfer resistance is governed by
both the liquid and gas film resistances.
Using the Hatta equation, the overall rate equation for this section is,
PA
D C
B B
HA
D Ab
rA
1
1
o
k LA k GA H A
Using the same basis for reaction rate (interfacial mass transfer area, S),
rAs
1 dN A
S dt
YA
PT
1 YA
DB H A
XB
CT
DA b
1 X B
H
1
oA
k GA k LA
Using material balance over differential area in the interior reaction zone and integrating,
Y
G A 2 dYA
G
h2
YA 2
dYA
Y
G
ln A2
k GA aPT Y A3
DB m
XB
Y A
D Ab
D m
D mG
D mG
1 B
B X Bc B
YA3
Where
and
DA b
DA L
DA L
h2
=
height
of
interior
reaction
zone
[m]
HA
=
Henrys
law
constant*
[N.m/mol]
KGA
=
overall mass transfer coefficient (with chemical reaction)
[mol/m3s.Pa]
D A, D B =
diffusivities
of
A
and
B
respectively
[m2/s]
C, CT =
partial and total molar concentration in the liquid [mol/m3]
b
=
stoichiometric coefficient of NaOH in reaction equation
(*Note that this can only apply if solution is dilute such that Y A << 1; XB << 1, then PA = PTYA
and CB = CTXB)
YA 3
Experimental Procedures
Apparatus
2. 1 X Vertical packed bed column (Internal Diameter (ID): 0.05m, Height (H): 0.490m) along
with the following:
a. Raschig Rings as packing material (ID: 0.006m, Outer Diameter (OD): 0.008m and
Length (L): 0.009m),
b. Various valves, connecting pipes and flow meters,
c. Inlet Distributor,
d. By-passes for extraction of inlet and outlet gas samples,
e. Drainage Facility.
3. CO2 and fresh air supply system:
a. 1 X Compressed CO2 gas tank,
b. 1 X Air supply,
c. 1 X CO2 pump and air pump,
d. 2 X Flow meter with control valve,
e. Connecting pipes.
4. One large reservoir feed tank, greater than 25 L (to store NaOH) along with:
a. 1 X Fluid pump,
b. 1 X Flow meter with control valve,
c. Connecting pipes,
d. 1 X Large stirring stick,
e. 1 X Big funnel.
5. Tap Water supply system:
a. 1 X Water supply,
b. 1 X Flow meter with control valve,
c. Connecting pipes,
d. 1 X Filtration system.
6. Gas Chromatography analyser machine along with:
a. 1 X 1.0 mL Gas Syringe.
7. Acid-base Titration Apparatus:
a. 1 X 50.00 mL burette,
b. 1 X 10 mL,
c. 1 X 20 mL pipette,
d. 2 X Conical Flasks,
e. 1 X Small Funnel,
f. 2 X Beakers,
g. 1 X Retort Stand.
Experimental Procedure
10. The expertimental procedure can be summarised in Figure XXX.
10
Inconsistent
Consistent
Backward titrate to get
composition of liquid outlet
Consistent
Adjust flow rates to Run 2
values
Completed all
6 Runs
Continue with
next Run
End of experiment
11
Inconsistent
12
Water
NaOH(aq)
0.60
0.00
0.55
0.05
0.50
0.10
0.45
0.15
0.40
0.20
0.35
0.25
13
Water
NaOH(aq)
0.60
0.00
0.55
0.05
0.50
0.10
0.45
0.15
0.40
0.20
0.35
0.25
Outlet (Vol %)
Run
Sample 1
Sample 2
Average
CO2(g)
Air
CO2(g)
Air
CO2(g)
Air
CO2(g)
1
18.5
81.5
13.7
86.3
14.1
85.9
13.9
2
18.5
81.5
7.07
92.9
6.77
93.2
6.92
3
18.5
81.5
2.40
97.6
2.70
97.3
2.55
4
18.5
81.5
0.783
99.2
0.738
99.3
0.761
5
18.5
81.5
0.254
99.8
0.239
99.8
0.247
6
18.5
81.5
0.118
99.9
0.0953
99.9
0.107
Table 4.2: Composition of CO2(g) and Air measured using Gas Chromatography
Inlet (Vol %)
Run
1
2
3
4
5
6
Air
86.1
93.1
97.5
99.2
99.8
99.9
for titration
Average
0.00
1.25
1.35
1.45
1.50
1.70
Table 4.3: Volume of HCl(aq) used for Titration with liquid sample drawn from Absorption
Column
Volume of liquid sample taken from absorption column outlet for back-titration = 10mL
Volume of NaOH(aq) added for back-titration = 5mL
1.177
28.85
1.788
44.01
1000
18.02
29.41
2979.97
1.77 x 10-9
2.17 x 10-9
298
101325
1.00
55.56
0.49
0.05
0.006
0.008
0.009
[HCl(aq)]titration (mol/L)
[NaOH(aq)]stock (mol/L)
[NaOH(aq)]feed tank (mol/L)
[NaOH(aq)]titration (mol/L)
Table 4.6: Concentration of Reagents
0.1084
2.00
0.180
0.1095
15
Titration: 5mL of NaOH(aq) sample is withdrawn from the feed tank and titrated against
0.1084M of HCl.
Average volume of HCl(aq) used to fully titrate the 5mL NaOH(aq), VHCl = 8.3mL
Since the reaction occurs in a stoichiometric ratio of 1:1,
Concentration of NaOH(aq) in feed tank
=
8.3
x
0.1084
/
5
= 0.18M
Determining the inlet and outlet molar ratios
(i) Molar ratio for inlet CO2(g), YA,in
Since the inlet air remains the same for all the runs, the average inlet gas concentrations will
be used for calculations.
Inlet
molar
ratio
of
CO2(g),
YA,in = (average inlet CO2(g) conc.) / (average inlet air conc.)
=
18.5
/
81.5
= 0.227
(ii) Molar ratio for outlet CO2(g), YA,out
Sample calculation using Run 6:
Outlet
molar
ratio
of
YA,out =
(outlet
CO2
concentration)/(outlet
air
=
0.107
/
= 0.00107
The sample calculations are then repeated for Runs 1 to 5:
Run
Outlet molar ratio of CO2(g), YA,out
1
0.161
2
0.0743
3
0.0262
4
0.00767
5
0.00247
6
0.00107
Table 4.7: Outlet Molar Ratios of CO2(g)
CO2(g),
concentration)
99.9
16
Run
Inlet molar ratio of NaOH(aq), XB,in
1
0
2
0.000294
3
0.000648
4
0.00108
5
0.00162
6
0.00231
Table 4.8: Inlet Molar Ratios of NaOH(aq)
(iv) Molar ratio for outlet NaOH(aq), XB,out
Volume of outlet solution collected = 10mL
Volume of 0.1095M NaOH(aq) added
=
5mL
no. of moles of NaOH(aq) added
= 0.000548 mol
Sample Calculation using run 6:
No. of mol of unreacted NaOH(aq) =
(9.55
1.70)/1000
x
0.1084
= 0.000851 mol
Concentration of NaOH(aq) in outlet liquid = (0.000851 0.000548) / (20/1000)
= 0.0151 M
Molar flow rate of NaOH(aq) in outlet liquid
=
0.0151
M
x
0.35
L/min
= 0.00530 mol/min
Molar
ratio
of
NaOH(aq)
in
outlet
liquid,
XB,out = molar flowrate of NaOH(aq) / molar flowrate of water
=
0.00530
/
19.45
= 0.000273
Note that Run 1 is conducted without any NaOH(aq) in the inlet for pure physical absorption,
thus there will not be any NaOH(aq) present in the outlet liquid.
The sample calculations are then repeated for Runs 1 to 5:
Run
Outlet molar ratio of NaOH(aq), XB,out
1
-0.000493
2
-0.000118
3
-0.0000395
4
0.0000531
5
0.000195
6
0.000273
Table 4.9: Outlet Molar Ratios of NaOH(aq)
Negative values of XB,out indicate that all the NaOH(aq) have been reacted (amount of
NaOH(aq) left after reaction with dissolved CO2 is less than the amount added in), therefore
the negative values can all be taken to be zero.
Determining the mass transfer coefficients
(i) Overall gas mass transfer coefficient without reaction, KGAoa
To perform this calculation, only data from Run 1 is required (region of physical absorption).
17
Assume ideal gas behaviour for air at room temperature and pressure,
18
18g/mol
x
x
55560
mol/m3
CT
55560
CB,out =
=
= 15.17 mol/m3
PA,in =
= 0.227 atm
Enhancement factor,
XB,out
0.000273
CT
55560
PT
YA,in
Calculating E at the top and bottom of the column and taking the average as the overall
enhancement factor:
0.0122
Run
PA (atm)
CB,in
(mol/m3)
CB,out
(mol/m3)
Etop
Ebottom
Eaverage
0.227
-27.39
1.00
-1.17
-0.085
0.227
16.33
-6.56
2.30
0.479
1.39
0.227
36.00
-2.19
3.86
0.826
2.34
0.227
60.00
2.95
5.76
1.23
3.50
0.227
90.00
10.83
8.15
1.86
5.01
0.227
128.34
15.17
11.19
2.20
6.70
19
Run
KGAa
kGAa
kLAa (1/s)
3
(mol/m .atm.s) (mol/m3.atm.s)
1
2
3
0.375
3.80
0.0122
4
5
6
Table 4.11: Summary of Mass Transfer Coefficients Values
kLAa (1/s)
-0.00104
0.0170
0.0285
0.0427
0.0611
0.0817
KGAa
(mol/m3.atm.s)
-0.0357
0.502
0.772
1.05
1.34
1.61
According to Table XXX, the types of reaction zones present are as follows:
Run
XBC
Types of reaction zones present
1
0.000311
Physical Absorption
2
0.000376
Physical Absorption, Interior Reaction
3
0.000562
Physical Absorption, Surface Reaction
4
0.000867
Surface Reaction, Interior Reaction
5
0.00128
Surface Reaction, Interior Reaction
6
0.00182
Surface Reaction, Interior Reaction
Table 4.12: Calculations for XBC and Types of Reaction Zones Present
Heights of various reaction zones
20
YA,out = 0.0262
XB,in = 0.000648
XB,2 = 0 (physical absorption zone present)
This value of YA,2 is not possible as it has to be lower than YA,in (i.e. 0.227) after physical
absorption has taken place. This anomalous result may have occurred due to errors which
21
would be discussed below in the error analysis section. In the meantime, Run 3 will be
treated as though it does not possess a physical absorption zone in the further calculations.
Consider control area C, material balance gives:
XB,2
YA,2
XB,3 = XBC
Height of physical absorption zone,
=
=
XB,out
YA,in
Run
XBC
XB,in
XB,2
XB,3
XB,out
YA,in
YA,2
1
0.000311
0
0
0.227
-
2
0.000376
0.000294
0
0.000294
0
0.227
0.194
3
0.000562
0.000648
0
0.000562
0
0.227
0.290
22
4
0.000867
0.00108
0.0000531
0.000867
0.0000531
0.227
0.426
5
0.00128
0.00162
0.000195
0.00128
0.000195
0.227
0.583
6
0.00182
0.00231
0.000273
0.00182
0.000273
0.227
0.831
YA,3
YA,out
0.0743
0.0612
0.161
0.0743
0.0262
0.559
0.713
h1
0.49
0.173
h2
0.233
0.236
h3
0.0775
Total
0.49
0.406
0.314
Error (%)
17.1
35.9
Table 4.13: Summary of Final Experimental Values
0.0945
0.00767
1.10
0.158
0.230
0.388
20.8
0.141
0.00247
1.63
0.107
0.370
0.477
2.65
0.201
0.00107
2.32
0.0894
0.479
0.568
-15.9
Run 3 is taken to consist of the surface reaction zone only due to the unfeasible Y A,2
calculated previously above.
23
Discussion
The main aim of this experiment was to study the various reaction zones for a continuous,
counter-current gas-liquid absorption process carried out in a packed bed column. In
addition to that, the heights corresponding to the various reaction zones, as well as the mass
transfer coefficients pertaining to the entire experimental process were determined. In this
section, the assumptions made throughout the course of the experiment will first be stated,
followed by a discussion regarding the results of our experiment and whether they agree
with theoretical principles.
Assumptions
The following assumptions were made for this experiment:
1. The entire experiment was carried out under isothermal, isobaric, and continuous
steady state conditions. 20min-intervals were given after the settings were calibrated
for each run so as to ensure that the conditions were reached before the collection
of any samples.
2. Phase boundary occurs at thermodynamic equilibrium.
3. The solution in the absorption column was dilute enough (<0.05moldm-3) such that
Henrys law is valid and can be applied.
4. Reaction between the CO2(g) and NaOH(aq) was instantaneous and irreversible, such
that both the reactants cannot coexist at any point, and thus the process could be
considered as one that was mass-transfer controlled.
5. NaOH(aq) was non-volatile, no vaporisation of NaOH(aq) into the gaseous film or gas
bulk occurred.
6. The diffusion of CO2(g) from the bulk gas through the gas-liquid interface and into the
film was purely by mass transfer.
7. Run 6 had a sufficiently high NaOH(aq) concentration such that only the surface
reaction zone existed.
8. There was no other mutual exchange of materials between the gas phase and the
liquid phase except for the component (i.e. CO2(g) in this case) to be removed.
Effect of chemical reaction on the rate of absorption
Run
NaOH(aq)
flow rate Eaverage
(L/min)
kLAa (1/s)
0.00
-0.085
-0.00104
0.05
1.39
0.0170
0.10
2.34
0.0285
0.15
3.50
0.0427
0.20
5.01
0.0611
24
0.25
6.70
0.0817
25
Nonetheless, we would still expect the calculated height to be less than the actual height
since the experimental absorption column itself is not 100% efficient. Apart from that, the
complexity of the calculations might also lead to large errors of uncertainty.
26
Error Analysis
Throughout the experiments, several assumptions made coupled with experimental errors
could have lead to the aberrations that are observed in the observed data.
1. The flow rates of some of the variables in the experiment such as air, water and
sodium hydroxide had to be manually controlled by the experimenter. Constant
fluctuations were observed in the flow rates of these variables. Everytime the water
level in the packed bed had to be adjusted, it was observed that the flow rates had to
be re-adjusted to desired values. Thus everytime a manual re-adjustment had to be
made, time may be required to establish a new equilibrium reading and therefore
the concentration of carbon dioxide recorded may not have been that at equilibrium.
2. During many steps of the experiment such as reading off carbon dioxide calculations
and titration results, the steps were repeated a few times to ensure the credibility
and therefore the average values were taken. Any random errors present may be
aggravated due to this averaging and might have led to larger deviations from the
ideal values.
3. The sodium hydroxide solution in the feed tank is assumed to be perfectly mixed.
However, when the experimenter manually diluted the stock solution of sodium
hydroxide, he might not have mixed the resulting solution with the stirrer for an
adequate time period. This would mean that there might be fluctuations in the
concentration of sodium hydroxide entering the packed column at any one time.
4. When taking the sample of carbon dioxide for testing its concentration, the sample
had to be taken from the end of a tube where it is possible for the sample and air to
intermingle. This suggests that although an accurate gas chromatography is utilized
to give us the concentration, the sample itself may be contaminated in the first place.
5. It was accepted that steady state had been attained when the carbon dioxide
concentrations with an interval of 5 minutes had less than a percent difference in
their values. However, this criteria may not be representative of a steady state
having been reached.
6. The reaction between sodium hydroxide and carbon dioxide is assumed to be
instantaneous. However, in reality, this assumption may not hold.
7. The usage of Henrys constants in calculations may cause a problem if the solutions in
question are not dilute enough as Henrys Law is only applicable for dilute solutions.
27
28
29
Conclusion
1 page Jia Ming
30
References
Wolfgang Bertsch. (2000). Two-dimensional gas chromatography. concepts,
instrumentation, and applications part 2: Comprehensive two-dimensional gas
chromatography. 23(3), 167.
W.E. McCabe, J.C. Smith, and P.Harriott 2001. Unit Operations of Chemical Engineering,
McGraw Hill, New York
th
Perrys Chemical Engineers Handbook, 7 Edition. 1997 (Ed: R.H. Perry, D.W. Green, and J.O.
Maloney), McGraw-Hill, New York.
P.V.Danckwerts. (n.d.). The absorption of gases in liquids. In Retrieved from
http://pac.iupac.org/publications/pac/pdf/1965/pdf/1004x0625.pdf
Hans T. Karlsson. (2010). Absorption with chemical reaction. Unpublished raw data, Energy
and
Environment
KET010,
Retrieved
from
http://www.chemeng.lth.se/ket010/Arkiv/Absorption_Reaction.pdf
M.A. Olutoye and A. Mohammed. (2006). Modelling of a gas-absorption packed column for.
(Master's
thesis)Retrieved
from
http://www.journal.au.edu/au_techno/2006/oct06/journalTechV10N2_aticle11.pdf
31