TSG-T-E-TR-TS2110

TSG, ISAC

Issue No. A

Test Report on
Thermal Stability Measurement by DSC

Page 1 of 2
Revision. 00
Date: 01-07-2013

EXPERIMENTAL HEAT TRANSFER SECTION


THERMAL TESTING DIVISION, THERMAL SYSTEMS GROUP
ISRO SATELLITE CENTRE, BANGALORE 560017
TEST REPORT
Date: 15/05/2014
1

Test Report No.

TR-0515/14 (A)

Requisition No.

TSG/TTD/EHTS-0410/14 (C)

Property Measured

Thermal Stability

Sample

1. PE (Polyethylene) Film
2. PP-PE-PP (Polypropylene-Polyethylene- Polypropylene)
Film

Temperature Range

00C to 2000C

Facility Used

TAS-DSC Q100

Heating / Cooling rate

50C/min

Uncertainty in Measured Data

Within 5%

Environment

Normal Room Pressure with inert gas purge


Base Purge Nitrogen 300ml/min.
Sample Purge Helium 25ml/min.

10

Calibration

DSC was Calibrated with SRMs Prior to Unknown Sample


Measurements as per ASTM standard test methods.

11

Sample Application

TDP

12

Results

Enclosed

13

Expected Value as per I/O

NA

14

Literature Value

15

Observations/comments

Melting point for PE=1350C; PP=1650C(supplied by indenter)


Two different virgin test samples were subjected to DSC
studies in the temperature range of 00C to 2000C. In the first
sample (PE) melting point is observed at 136.80C against the
literature value 1350C and in the second sample (PP-PE-PP)
observed two melting points at 131.20C & 158.30C
respectively against the literature value of 1350C & 1650C
respectively. 1st melting point corresponds to melting of PE
and 2nd melting point corresponds to melting of PP.
Typical thermal curves of DSC are enclosed.

Note:-No Part of this Test Report may be Reproduced, Stored or published in any Form or by any
Means With out the Prior Written Permission from the Experimental Heat Transfer Section,
Thermal Testing Division, ISRO Satellite Centre, Bangalore 560017.

TSG-T-E-TR-TS2110

Issue No. A

Page 2 of 2

Test Report on
Thermal Stability Measurement by DSC

TSG, ISAC

Revision. 00
Date: 01-07-2013

EXPERIMENTAL HEAT TRANSFER SECTION


THERMAL TESTING DIVISION, THERMAL SYSTEMS GROUP
ISRO SATELLITE CENTRE, BANGALORE 560017
TEST REPORT
Date: 15/05/2014
Sample: PE-Celgard2730
Size: 3.0200 mg
Method: DSC
Comment: Thermal Stability -1st sample 1st run

File: C:...\April-14\PE-Celgard2730-110414.001
Operator: DRK
Run Date: 11-Apr-2014 09:36
Instrument: DSC Q100 V9.6 Build 290

DSC

0
l

129.08C
252.8J/g

-1

Pan: Aluminum
Gas1: Helium 25.0 ml/min
Method 1: Equilibrate at 0 C
Method 2: Ramp 5.00 C/min to 200.00 C

Heat Flow (mW)

-2

l
l
l
l
l

-3

l
l
l
l

-4
l

136.83C

-5

-6

50

100

Exo Up

150

200
Universal V4.5A TA Instruments

Temperature (C)

Sample: PP-PE-PP-Celgard232x
Size: 3.0500 mg
Method: DSC
Comment: Thermal Stability -1st sample 1st run

File: C:...\PP-PE-PP-Celgard232x-110414.001
Operator: DRK
Run Date: 11-Apr-2014 13:35
Instrument: DSC Q100 V9.6 Build 290

DSC

Heat Flow (mW)

-1

Pan: Aluminum
Gas1: Helium 25.0 ml/min
Method 1: Equilibrate at 0 C
Method 2: Ramp 5.00 C/min to 200.00 C

128.02C
33.60J/g
l

155.51C
34.62J/g
l
l

l ll

-2

l
l

l
l

l
l
l
l

158.33C

-3
l
l

131.19C
-4
Exo Up

50

100

Temperature (C)

150

200
Universal V4.5A TA Instrume

Note:-No Part of this Test Report may be Reproduced, Stored or published in any Form or by any Means Without the
Prior Written Permission from the Experimental Heat Transfer Section, Thermal Testing Division, ISRO Satellite
Centre, Bangalore 560017.

TSG-T-E-TR-TS2120

TSG, ISAC

Issue No. A

Test Report on
Thermal Stability Measurement by TGA

Page 1 of 2
Revision. 00
Date: 01-07-2013

EXPERIMENTAL HEAT TRANSFER SECTION


THERMAL TESTING DIVISION, THERMAL SYSTEMS GROUP
ISRO SATELLITE CENTRE, BANGALORE 560017
TEST REPORT
Date: 15/05/2014
1

Test Report No.

TR-0515/14 (B)

Requisition No.

TSG/TTD/EHTS-0410/14 (B)

Property Measured

Thermal Stability

Sample

1. PE (Polyethylene) Film
2. PP-PE-PP (Polypropylene-Polyethylene- Polypropylene)
Film

Temperature Range

RT to 2000C

Facility Used

TAS-TGA Q500

Heating / Cooling rate

50C/min

Uncertainty in Measured Data

Within 5%

Environment

Normal Room Pressure with inert gas purge


Furnace Purge Nitrogen 40ml/min.
Sample Purge Nitrogen 60ml/min.

10

Calibration

TGA was calibrated for temperature with manufacturer


recommended standard procedures prior to unknown sample
measurements.

11

Sample Application

TDP

12

Results

Thermal curves enclosed

13

Expected Value as per I/O

Not Known

14

Literature Value

15

Observations/comments

NA
Two different virgin test samples were subjected to TGA
studies in the temperature range of RT to 2000C from the
material as received from the indenter for its thermal stability.
In the 1st sample (PE) weight loss of about 0.22% and 0.32%
was observed in the temperature range of RT to 1400C and
1400C to 2000C respectively. The left over residue was in the
form of lump.
In the 2nd sample (PP-PE-PP) weight loss of about 0.11% and
0.09% was observed in the temperature range of RT to 1400C
and 1400C to 1700C respectively. The left over residue was in
the form of lump.
Typical thermal curves of TGA are enclosed.

Note:-No Part of this Test Report may be Reproduced, Stored or published in any Form or by any Means With out the
Prior Written Permission from the Experimental Heat Transfer Section, Thermal Testing Division, ISRO Satellite
Centre, Bangalore 560017.

TSG-T-E-TR-TS2120

Issue No. A

Page 2 of 2

Test Report on
Thermal Stability Measurement by TGA

TSG, ISAC

Revision. 00
Date: 01-07-2013

EXPERIMENTAL HEAT TRANSFER SECTION


THERMAL TESTING DIVISION, THERMAL SYSTEMS GROUP
ISRO SATELLITE CENTRE, BANGALORE 560017
TEST REPORT
Date: 15/05/2014
Sample: PE-Celgard2730
Size: 2.1250 mg
Method: TGA standard
Comment: Thermal Stability Measurement-1st sample

File: C:...\April-14\PE-Celgard2730-110414.001
Operator: DRK
Run Date: 11-Apr-2014 13:14
Instrument: TGA Q500 V6.5 Build 196

TGA

102

0.020

Pan: Platinum
Balance Gas: Nitrogen 40.0 ml/min
Sample Gas: Nitrogen 60.0 ml/min
Method 1: Ramp 5.00 C/min to 200.00 C

0.015

101
p

[ l ] Weight (%)

p
p

p
p
l

100

p
p

l
p

l
p

0.005

0.2248%
(0.004776mg)

0.000
0.3176%
(0.006749mg)

-0.005

Residue:
99.48% Lump
(2.114mg)

99

[ p ] Deriv. Weight (%/C)

0.010

-0.010

-0.015

98

50

100

150

200

Universal V4.5A TA Instruments

Temperature (C)
Sample: PP-PE-PP-Celgard232x
Size: 1.7030 mg
Method: TGA standard
Comment: Thermal Stability Measurement-1st sample

-0.020

File: C:...\PP-PE-PP-Celgard232x-110414.001
Operator: DRK
Run Date: 11-Apr-2014 14:33
Instrument: TGA Q500 V6.5 Build 196

TGA

102

0.05
Pan: Platinum
Balance Gas: Nitrogen 40.0 ml/min
Sample Gas: Nitrogen 60.0 ml/min
Method 1: Ramp 5.00 C/min to 200.00 C

0.04
0.03

[ l ] Weight (%)

0.02
p

p
l

100

p
l

p
l

l
p

p
l

p
l

0.08640%
(0.001471mg)
p

p
l
l
0.1115%
(0.001899mg)

p
l
p

0.01
p

0.00
l

l
Residue:
99.65% Lump
(1.697mg)

-0.01

[ p ] Deriv. Weight (%/C)

101

-0.02

99
-0.03
-0.04
98

50

100

Temperature (C)

150

200

-0.05

Universal V4.5A TA Instruments

Note:-No Part of this Test Report may be Reproduced, Stored or published in any Form or by any Means With out the
Prior Written Permission from the Experimental Heat Transfer Section, Thermal Testing Division, ISRO Satellite
Centre, Bangalore 560017.

Das könnte Ihnen auch gefallen