Site Survey
1. Instructions
The Field Engineer or Sales Representative must schedule a preliminary Site Survey visit in
accordance with local branch operating procedures.
Conduct the survey in cooperation with the customers biomedical or maintenance engineer. You
must define acceptance or rejection of each requirement listed in the Site Survey form by
checking Accepted or Rejected in the Preliminary Visit section.
Describe the reason for the rejection of any parameter, and inform the customer of all
unacceptable parameters and the reasons for rejection.
If any parameters are marked Rejected during a preliminary site visit, a final visit must be made
to ensure that all unacceptable conditions have been satisfied. Time must be allowed for
corrective action to be taken before the final visit. If, after the final site survey visit, there are still
unacceptable conditions, notify the customer and Service Manager.
The Service Manager should notify the customer that specific system installation requirements
have not been met and that delivery of the system may be delayed if all the site problems are not
resolved before the scheduled installation date.
2. Distribution
Forward a completed copy of the site survey forms to:
a. Customer (Laboratory Supervisor)
b. Randox Office
3. Customer Details
Institution:
Address
City:
County:
Post Code:
Country:
Telephone:
Lab Director:
Primary Contact:
Telephone:
4. Installation Route
Analyser dimensions are as follows:
Width
970 mm
Depth
582 mm
Height
690 mm
Weight
150 kg
Provide a brief description of the route the analyzer will take from the delivery area to the
proposed installation site. Particular attention should be made to doorways, staircases, etc.
5. Power Requirements
The Rx Imola requires a 3-pin (or 2-pin depending on site location) power outlet with a ground
(A-Type, S-Type or O-Type) within the reach of the analysers accessory power cable.
Specifications
Power
900 VA
Input Voltages
Yes
No
The system should not be installed near sources of high frequency noise (i.e. centrifuges or
discharge units).
Is the proposed installation site free of such noise ()?
Accepted
Yes
No
Rejected
6. UPS
The UPS should be capable of supplying 1230 watts for the RX Imola / PC system.
Furthermore, the required maximum power requirements should take into account the
laboratory water system. The UPS needs to maintain this supply for as long as the power out
situation lasts or as determined by the customer.
Maximum Power Consumption (Watts)
Rx Imola
900
164
26
Accepted
Rejected
7. Water Requirements
The analyzer needs a purified water generator. The water supply must be within acceptable
reach of the analyser.
Specifications
Pressure
Capacity
18 L
Connection Tubing
Inside diameter: 12
Outside diameter: 18
Does the purified water generator meet the required specifications ()?
Accepted
mm
Yes
mm
No
Rejected
8. Drainage Requirements
The drainage system must capable of removing at least 18 L of waste per hour.
The drainage outlet should be positioned at least 500 mm lower than the proposed analyzer
position.
Analyser
500 mm
Waste Outlet
Accepted
Yes
No
Rejected
9. Environmental Requirements
The analyser installation site must satisfy the following environmental conditions:
()
Task
No direct sunlight
Dust free environment
No vibration or excessive sound levels
A floor and supporting worktop capable of holding the analyser weight (~160 kg)
Placement on a horizontal surface
No variation in voltage supply (less than 10 %)
Avoid placement under fluorescent lighting
Accepted
Yes
No
Rejected
()
Task
Indoor temperature: 15 30C
Variation: > 2C per hour
Good air ventilation
Indoor humidity: 45 85% (without condensation)
No exposure to air conditioning or heat exchangers
Accepted
Yes
No
Rejected
11. Space
Refer to the diagram below for required installation space. This is essential for maintenance and for
exhaust ventilation. (Unit: mm)
150+
PC
300+
Analyser
300+
1340
300+
690
600
970
500+
Front
2470
Accepted
Yes
No
Rejected
Baud Rate
Data Bit
Stop Bit
Parity Bit
Parity: None/Even/Odd
Protocol (E1381)
Normal/No Handshake
Check the proposed installation position of the analyser within the laboratory and the
connection point for the Host PC.
Is there sufficient RS232 cable length to allow connection?
Are all communication conditions satisfied ()?
Accepted
Yes
Yes
No
No
Rejected
Use the space below to sketch a proposed location of the system in the laboratory, with
particular reference to:
a.
b.
c.
d.
Signature: _______________________________
Final Visit:
Date:
Signature: _______________________________
Signature: _______________________________
Final Visit:
Date:
Signature: _______________________________
Yes
No
10