Sie sind auf Seite 1von 4

SHAKER

OPERATING MANUAL
A)

SHAKING SPEED
1. Press SHAKER/SET key
2. Press UP/DOWN key
3. Press ENTER BUTTON
- Finish to set value
4. Press START/STOP key
- Shaking starts in clockwise rotation
* Recall function of RPM value
- Previous RPM value starts if new RPM is set again

B)

TIMER PROGRAMMING
- Programming time cannot be done during shaking.
So, stop shaking by pressing STARTS/STOP key and start
programming.
Input total run time
1. Press SHAKER/SET key twice
- This section is for setting total run time
- run.H shown on MODE display
2. Press UP/DOWN Button to set Time and then press ENTER
key to store the value (Max Input Time: 1-999 hours)
- After setting the value, TIMER LED is on
3. Press SHAKER/SET key
- This section is for setting minute and seconds.
- run.H shown on MODE display
4. Press UP/DOWN key to set minute and seconds and then press
ENTER key to store the value.
(Max Input Time: 10 seconds ~ 59 minutes 59 seconds)

ROLL SHAKING
MANUAL OPERASI
1. Switch ON main switch
2. The device can be turned with the flip
switch A on the side of the unit
3. Adjust the speed
4. Set the operation mode
5. Set the timer and temperature

HOT PLATE UGO

OPERATION MANUAL
1. Switch ON
2. The graphic display light and shows rats, then the display
shows key function:
F1 (Opr)

F2 (Func) F3 (Mem) F4 (Esc)

Press F1 F1 (Chr)

F2 (Plt)

Press F2 F2 (Cal)

F1 (Set)

Press F1 (for temperature)


F1

F2

F3 OK

F4 (Esc)

3. Switch OFF
4. Press F4 until the display shows rats.

PLETHYSMOMETER - OPERATION MANUAL


SET-UP

1. Fasten the cell to a laboratory upright by a right-angle clamp and insert the water
reservoir on the same or a
separate stand.
2. The cell is fitted with a slide valve provided with two side pipe connections. Branch
them by the tubes
stretches provided to the water reservoir and to a sink.
3. Keep the valve wings horizontal (SHUT) and fill the reservoir with a solution
prepared in a separated vessel
according to the following directions. Add to distilled water.
A)

2-3 ml /1 wetting compound (provided with standard package) which minimizes


drop and meniscus build-up leading to more accurate measures.
B)
About 6 milimoles of any univalent alkali metal chloride (NaCl, LiCl, KCL, RbCl,
CsCL). In practice 0.4-0.5 g/l of common easily available NaCl (0.3-0.4 g for the
0.7 litre reservoir)
- Conductivity attained by this molarity provides optimum transducer
performances.
- Moreover, as the transducer works satisfactorily up to 12-15 mMoles per litre,
any solution alteration
brought about by rat dropping, in most cases increasing conductivity, does
not impair its operation.
- Stir !
THE SOLUTION MUST ALWAYS BE FRESH, SEE FAST CALIBRATION *
1. Combine a gentle push and rotation of a transducer head to make sure it is properly
fitted to the smaller
tube and check that its cable is clamped.
2. Fill the cell by turning the valve winged cap approx. 45C leftwards, according to the
side to which the
reservoir is linked.
3. Reposition the valve cap horizontally to stop the flow and turn it otherwise to drain.
4. Fill the drain a couple of times to practice and to purge the system from air bubbles.
5. Adjust the water level in the marked range, close to the upper red line.
6. Connect the transducer cable and the hold pedal to the Plethysmometer, WATER
CELL AND PEDAL
SWITCH connectors respectively.
SWITCHING ON
1. Switch ON. The graphic display lights and shows UGO BASILE aft.3 seconds, the
display shows the
configuration of the previous experiment. The operator can maintain it, if
convenient, without the need of
fresh setting.
2. Warm up 2-3 minutes.
CALIBRATION
1. When the 1740 is put in operation for the first time, it requires to be calibrated with
a volume probe.
2. Use the calibration probes provided with the standard package:- 0.5 or 1.0 ml probe for 7156 cell
- 1.0 or 2.0 ml probe for 7152 cell
- 2.0 or 4.0 ml probe for 7157 and 7159 cell
3. Calibration is also necessary if the instrument has beeb OFF duty for a long time
after having changed the
solution, or replaced the cell, the transducer, etc.
4. To calibrate, depress F1 (CAL) key, from the main menu. The display zeroed
5. F1 (UP ARROW) key if depressed more times, enables the operator to select the
volume of the calibration
probe in the sequence loop 0.5, 1.0, 2.0, 4.0, 8.0 ml
6. Select the calibration probe and depress the F2 to confirm the choice.

NOTE: BEFORE MAKING A SECOND TEST, DO NOT FORGET TO ZERO THE 7140
AND TO DRY THE PROBE TAKES OUT SOME DROPS.
* FAST CALIBRATION:

Before each working session, it is advisable to check the calibration: simply zero
the display and dip the suitable calibration probe into the cell, then read the value
shown by the instrument. If not successful, never use a solution older than 2-3
days. For more information please refer manual book or senior operator.

Das könnte Ihnen auch gefallen