Beruflich Dokumente
Kultur Dokumente
OPERATING MANUAL
A)
SHAKING SPEED
1. Press SHAKER/SET key
2. Press UP/DOWN key
3. Press ENTER BUTTON
- Finish to set value
4. Press START/STOP key
- Shaking starts in clockwise rotation
* Recall function of RPM value
- Previous RPM value starts if new RPM is set again
B)
TIMER PROGRAMMING
- Programming time cannot be done during shaking.
So, stop shaking by pressing STARTS/STOP key and start
programming.
Input total run time
1. Press SHAKER/SET key twice
- This section is for setting total run time
- run.H shown on MODE display
2. Press UP/DOWN Button to set Time and then press ENTER
key to store the value (Max Input Time: 1-999 hours)
- After setting the value, TIMER LED is on
3. Press SHAKER/SET key
- This section is for setting minute and seconds.
- run.H shown on MODE display
4. Press UP/DOWN key to set minute and seconds and then press
ENTER key to store the value.
(Max Input Time: 10 seconds ~ 59 minutes 59 seconds)
ROLL SHAKING
MANUAL OPERASI
1. Switch ON main switch
2. The device can be turned with the flip
switch A on the side of the unit
3. Adjust the speed
4. Set the operation mode
5. Set the timer and temperature
OPERATION MANUAL
1. Switch ON
2. The graphic display light and shows rats, then the display
shows key function:
F1 (Opr)
Press F1 F1 (Chr)
F2 (Plt)
Press F2 F2 (Cal)
F1 (Set)
F2
F3 OK
F4 (Esc)
3. Switch OFF
4. Press F4 until the display shows rats.
1. Fasten the cell to a laboratory upright by a right-angle clamp and insert the water
reservoir on the same or a
separate stand.
2. The cell is fitted with a slide valve provided with two side pipe connections. Branch
them by the tubes
stretches provided to the water reservoir and to a sink.
3. Keep the valve wings horizontal (SHUT) and fill the reservoir with a solution
prepared in a separated vessel
according to the following directions. Add to distilled water.
A)
NOTE: BEFORE MAKING A SECOND TEST, DO NOT FORGET TO ZERO THE 7140
AND TO DRY THE PROBE TAKES OUT SOME DROPS.
* FAST CALIBRATION:
Before each working session, it is advisable to check the calibration: simply zero
the display and dip the suitable calibration probe into the cell, then read the value
shown by the instrument. If not successful, never use a solution older than 2-3
days. For more information please refer manual book or senior operator.