Beruflich Dokumente
Kultur Dokumente
Microscopy
Lecture 1
magnification.
1. Introduction
Resources
There are many books on microscopy, and it is worth browsing the
library. The course synopsis also lists many books. Much of the course
There exist a very large number of such techniques, and we cannot hope
Microscopy is a fast developing field and there are also excellent on-line
http://www.matter.org.uk/tem
http://www.scribd.com/doc/27761327/Microscopy-With-Light-andElectrons
find it helpful to know that by the end of the course I would expect you to
be able to:
Use words and sketch diagrams to explain the principle behind all
the microscopy techniques outlined in this course including optical
microscopy, transmission electron microscopy, scanning
electron microscopy, scanning probe microscopy, atom probe,
These lecture notes are largely based on the course developed over
many years by Prof M L Jenkins and Prof Peter Nellist (University of
Oxford) and I am extremely grateful to them for making these available to
me. Many additional sources have been used to compile these notes and
I would like to thank colleagues including Dr Rik Brydson, Dr. Andrew
Scott and Dr Peter Evennett (University of Leeds) from whom I have
received data.
2. Background
The introduction of the microscope as a tool for the biologist brought
examine?
3. Types of microscopy
In microscopy, we need to generate magnification so that we can see the
microstructure, and we need to detect a particular type of signal.
By the end of this course you should be able to assign all the
Ions produced in the high field near the specimen tip are accelerated
towards the screen and imaged.
the Time Of Flight of the ions: the time between the pulse and the
The stresses in the tip produced by the field may affect it (rearrange
position and the order of arrival of the ions on the position sensitive
It can tell you where atoms are but not what they are.
the (X,Y) position of the ion impact on the detector: the X-Y
sensitive detector
The atom probe (also referred to as position sensitive atom probe or
PoSAP) can position single atoms and has ~60% detection efficiency.
The resolution of the technique is sub-nanometre laterally and single
atom layer in depth. A typical specimen is needle shaped with a radius of
50nm. The volume analyzed is typically 20 nm by 20 nm by 50 nm (Fig
1.6).
The main limitation of the technique is the requirement for a very sharp
needle sample (preferably conducting). It has the advantage of being
able to detect even low mass elements including hydrides.
The atom probe microscope uses the same projection principle as the
FIM but can provide three dimensional information of the sample
structure and composition as shown in Fig. 1.5. A voltage pulse (or laser)
is used to stimulate atoms being emitted from the surface, i.e. deliberate
use is made of field-evaporation. The detector allows us to
simultaneously measure:
More details in :
Setna, Hyde, Cerezo, Smith and Chrisholm. Position sensitive atom probe study of the decomposition
of a Cu 2.6at% C alloy. Applied surface science 67 (1993) 368-379
http://www.sciencedirect.com/science/article/pii/016943329390340H
Cerezo et al. Materials Characterisation 25 :143 -156 (1990)
http://www.sciencedirect.com/science/article/pii/104458039090026G
For prospects of using this techniques for biomaterials see the book Nanosystem characterization
tools in the life sciences By Challa S. S. R. Kumar