Digital Engler Viscometer

Standards: ASTM D1665,


IP212, DIN 51560
Datasheet

S.D.M. Product number: 140


Web site: www.sdmtorino.com

Email: info@sdmtorino.com

1.Use of the instrument

This instrument is used for determination of specific viscosity of tars and their fluid products. This is an empirical flow test.
2.Instrument Description

Main unit: Structure made in stainless steel, with front opened jacket, liquid bath, stainless steel heater, stirring device, closing-ball ended rod, calibrated brass test cup with stainless steel orifice, PTFE tipped closing rod, cooling coil used for
external cooling device (closed circuit), insulated double wall, level indicator and control with heat breaking circuit and warning lamp and digital temperature reader and controller with PID algorithm and class A PT100 probe with overtemperature
alarm.

3.Technical Specifications:

Supply: 220Vac, 50/60Hz

Bath capacity: 7 liters

Setpoint stability: 0,1C

Power: 800W

Working Temperature: ambient to 100C(140F)

External dimensions of the instrument: 40 x 29 x 64 cm.

Weight: 12 Kg.

Accessories:
10-0261
10-0265
10-0267
10-0332

FILTER FUNNEL With stainless steel wire mesh


TESTING FLASK 50 ml, pack of 3 pcs Graduate and calibrated at 20C
KOHLRAUSCH FLASK 200 ml, pack of 3 pcs Graduation mark and calibrated at 20C
DIGITAL STOPWATCH 7 digit LCD, max.10 hours, 1/100 sec, digit h=8 mm

Thermometers
T-AS23C
T-AS24C
T-AS25C
T-IP76C

THERMOMETER
THERMOMETER
THERMOMETER
THERMOMETER

ASTM 23C
ASTM 24C
ASTM 25C
IP 76C

Consumables for 2 years


10-0265
10-0267

TESTING FLASK 50 ml ,pack of 3 pcs x1


KOHLRAUSCH FLASK 200 ml, pack of 3 pcs x1

SPARE PARTS
15-0270
15-0271
15-0165/W

14-0001
11-0008
15-0003/120
16-0005
15-0015
15-0004
12-0001

ENGLER CUP
ENGLER LEVEL-CONTROL
ORIFICE WRENCH
PROBE PT100A
HEATER 800W
LEVEL SWITCH
DIGITAL THERMOREGULATOR
STATIC RELAY
BIPOLAR GREEN SWITCH
MOTOR STIRRER

Note 1: The instrument is available in 2 variants: with one or two places.


Tip: Its recommended that before begining the test to check the liquid level, use Down arrow on the temperature controller to change the test temperature setpoint and press P to confirm it, wait for the controller to
reach the setpoint temperature and that the temperature reading is stable; after this the intrument is ready
to perform the viscosity test.
Aditional informations:
All variants, on the temperature controller have an over temperature alarm and maximum and minimum
limit of temperature which means that the user can use the temperature setpoint within that range.
In adition when a intstrument leaves S.D.M., the temperature controller its already calibrated and PID
parameters are tuned for the standard temperature range for that instrument.
Its also recommended to make a calibration once a year to adjust the controllers linearization.
Note 2. To calibrate the linearization of the temperature controller please read the Technical manual or
refer to your provider.