N INF07/123
Issue 2
1 Reagent Solutions
Warning.
Concentrated sulfuric acid must be handled with great
care at all times. In particular, ensure that, when
diluting concentrated acid, it is added to the water,
not water to the acid. Wear appropriate protective
clothing.
Concentrated ammonia solution is extremely volatile
and toxic; it must always be handled under a fume
hood. Wear appropriate protective clothing.
A volume of 2.5 l of each reagent solution maintains the monitor
in operation for a minimum of three months. The bottles are
color-coded to identify each reagent.
1st Acid 0.4M sulfuric acid (RED channel)
Place approximately 1 l high purity water in a plastic
beaker and carefully add 54.5 (0.5) ml analytical reagent
grade concentrated sulfuric acid, H2SO4, (1.84 s.g.).
Transfer the solution to a 2.5 l plastic bottle and make up
to 2.5 l with more high purity water.
Ammonium molybdate solution (VIOLET channel)
Dissolve 50 (1) g analytical reagent grade ammonium
molybdate, (NH4)6Mo7024.4H2O, in approximately 1 l high
purity water. Transfer the solution to a 2.5 l plastic bottle,
add 7.5 (0.5) ml ammonia solution, NH4OH, (0.880 s.g.)
and make up to 2.5 l with more high purity water.
2nd Acid 1.25M sulfuric acid (ORANGE channel)
Place approximately 1 l of high purity water in a
polyethylene beaker. Surround the beaker with running
cold water and add 170 (2) ml of analytical reagent grade
concentrated sulfuric acid, H2SO4 (1.84 s.g.) slowly and
carefully. Stir the solution continuously during the addition.
Add 63 (1) g of analytical reagent grade citric acid
crystals, C6H8O7.H2O, and stir to dissolve. Allow the
solution to cool to room temperature and then add 250
(5) ml of propan-2-ol (iso-propyl alcohol). Then transfer
to a 2.5 l plastic bottle. Make up to 2.5 l with more high
purity water.
Silica Analyzers
Mixing Chemical Solutions for Silica Analyzers
Note.
Store all standard solutions in tightly stoppered
polyethylene bottles.
The stock solutions remain stable for approximately
one year. Prepare standard solutions for the analyzer
immediately before they are required for use.
INF07/123
Issue 2
Silica Analyzers
Mixing Chemical Solutions for Silica Analyzers
3 Cleaning Solutions
3 Cleaning Solutions
It is important that the internal pipework of the monitor is
cleaned using the automatic cleaning schedule. Cleaning
prevents gradual fouling by molybdate precipitation in the
pipework which can introduce errors. Problems such as noise
can also be caused by contaminated pipework.
Warning. Sodium hydroxide is extremely caustic and must
be handled with great care. Wear appropriate protective
clothing.
500 ml of the cleaning solution is prepared as follows:
1. Dissolve 25 (1) g of sodium hydroxide pellets, NaOH, in
approximately 300 ml of high purity water in a plastic
bottle. Allow the solution to cool to ambient temperature.
2. Add to the solution 2.5 (0.01) g di-sodium EDTA,
C10H14O8N2Na2.2H2O and stir to dissolve.
3. Transfer the solution to a 500 ml measuring cylinder and
make to the mark with more high purity water. Mix well
and store in a tightly stoppered plastic bottle. Stored in
this bottle the solution is stable for many months.