Sie sind auf Seite 1von 530

LONG TERM STORAGE INSTRUCTIONS

The analyzer system itself is intended to be installed


indoors, on the inside of the shelter.
Equipment installed on the outside of the analyzer
shelter is rated minimum Nema 4 and able to
withstand the elements that will be encountered at the
end users final installation site. Consideration for long
term storage of the analyzer system as an entirety,
should include ensuring that the door and all other
entries to the system are secured shut, and all
electrical and gas entries are sealed to prevent the
ingress of moisture or any other foreign objects or
matter that could be detrimental to the analyzer and
all other equipment housed inside the system.
For long term storage, the Analyzer Systems should
be kept in a sealed container, preferably indoors, in
an area free of moisture and elements such as rain,
wind, dust, excessively frigid temperatures, and direct
sunlight.

Servomex 2500 Series


Process Gas Analysers
Installation Manual
Ref:02500/005C/2
Order as part 02500005C

WARNING, CAUTIONS AND NOTES


This publication includes WARNINGS, CAUTIONS and NOTES which provide,
where appropriate, information relating to the following:

WARNINGS: Hazards which will result in personal injury or death.

CAUTIONS: Hazards which will result in equipment or property damage.

NOTES: Alert the user to pertinent facts and conditions.

WARNING

LETHAL VOLTAGES: THE ELECTRICAL POWER USED IN THIS


EQUIPMENT IS AT A VOLTAGE HIGH ENOUGH TO ENDANGER LIFE.

BEFORE CARRYING OUT MAINTENANCE OR REPAIR THE EQUIPMENT


MUST BE DISCONNECTED FROM THE ELECTRICAL SUPPLY. TESTS
MUST BE MADE TO ENSURE THAT DISCONNECTION IS COMPLETE.

IF FOR ANY REASON THE POWER SUPPLY CANNOT BE


DISCONNECTED, FUNCTIONAL TESTING, MAINTENANCE AND REPAIR
OF THE ELECTRICAL UNITS IS ONLY TO BE UNDERTAKEN AS A LAST
RESORT AND MUST BE CARRIED OUT BY PERSONS FULLY AWARE
OF THE DANGER INVOLVED.

NOTE
This instruction manual for the 2500 Series range of analysers covers installation,
routine maintenance and fault diagnosis procedures, and it should be thoroughly
read and retained by the personnel responsible.

ii

Table of Contents
SECTION 1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
1.1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
1.2
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
1.3
Basic Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3
1.4
Options and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
1.4.1
Heated Sample Cell Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
1.4.2
Heated Sample Cell Steam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
1.4.3
Optional Output Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
1.4.4
Compensation Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
1.4.5
Application Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
SECTION 2 INSTALLATION GENERAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.2
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.2.1
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.2.2
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.2.3
Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
2.3
Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
2.4
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3
2.4.1
Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3
2.4.2
Mounting Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4
2.5
Instructions Specific to Hazardous Area Installations . . . . . . . . . . . . . . . . . . . 2.7
2.6
ATEX Label Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.8
SECTION 3 INSTALLATION ELECTRICAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
3.1
Electrical Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
3.2
Signal Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
3.2.1
Analogue Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7
3.2.2
Analogue Output Link Selections on Sensor Interface
Board (SIB) PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.9
3.2.3
Relay Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.10
3.2.4
Relay Output Link Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.10
3.3
Control Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.11
3.3.1
Range Change Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.12
3.3.2
Autocalibration Initiation Input . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.12
3.3.3
Password Keyswitch Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12
3.3.4
Sample Flow Sensor Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.13
3.3.5
Solenoid Valve Drives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.13
3.3.6
Externally Powered Solenoid Valve Connections . . . . . . . . . . . . .3.14
3.3.7
Internally Powered Solenoid Valve Connections . . . . . . . . . . . . . . 3.15
3.4
Digital Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.16
3.4.1
RS-232 Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.16
3.5
Pressure Transducer Connections (If Supplied) . . . . . . . . . . . . . . . . . . . . . 3.23

iii

SECTION 4 INSTALLATION GAS CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . 4.1


4.1
Purge Connections (EU1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
4.1.1
General Purge Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
4.2
End Boss Purge Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
4.2.1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
4.2.2
End Boss Scrubbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
4.2.3
End Boss Vents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
4.2.4
End Boss Purge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
4.3
Steam Heated Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6
4.4
Process Stream Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6
4.4.1
Basic Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6
4.4.2
High Integrity Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6
4.5
Power-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6
SECTION 5 FAULT DIAGNOSIS AND CELL MAINTENANCE . . . . . . . . . . . . . . 5.1
5.1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
5.2
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
5.3
General Fault Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
5.4
Serious Fault (Shutdown) Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5
5.5
Cell Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5
5.6
Routine Leak Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.9
SECTION 6 SPARES LISTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
6.1
Spares List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
6.1.1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
6.1.2
Sample Cell Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2
6.1.3
Source Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
6.1.4
Detector Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
6.1.5
Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
6.1.6
Scrubbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
6.2
Recommended Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.5
SECTION 7 INSTRUMENT SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
7.1
Generic 2500 Series Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
7.1.1
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
7.1.2
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
7.1.3
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
7.1.4
Performance Characteristics (each component) . . . . . . . . . . . . . . . 7.2
7.1.5
Performance Characteristics EMC . . . . . . . . . . . . . . . . . . . . . . . . 7.2
7.1.6
Sample Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
7.1.7
Sample and Ambient Performance . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
7.1.8
Solenoid Valve Relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
7.1.9
Analogue Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
7.1.10
Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
7.1.11
Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
7.1.12
Ingress Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
7.1.13
Optical Purge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
7.1.14
485 Purge Controller (2500 EU1 Only) . . . . . . . . . . . . . . . . . . . . . . 7.5
7.1.15
Alternative Purge Controller (2500 EU1 Only) . . . . . . . . . . . . . . . . 7.5

iv

SECTION 8 CE MARKING AND OTHER SAFETY APPROVALS . . . . . . . . . . . . 8.1


8.1
EMC Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
8.2
Low Voltage Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
8.3
Pressure Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
8.4
ATEX Directive and other non-European Hazardous Area approvals . . . . . . . 8.2

List of Figures
Figure 1.1
Figure 1.2
Figure 2.1
Figure 2.2
Figure 3.1a
Figure 3.1b
Figure 3.2
Figure 3.3
Figure 3.4
Figure 3.5
Figure 3.6
Figure 4.1
Figure 4.2
Figure 5.1

2500 Overall View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2


2500 Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
2500 Positioning Restrictions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5
2500 Mounting Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.6
2500 AC Power and Valve Connections (Purged Filter) . . . . . . . . . . . 3.3
2500 AC Power and Valve Connections (Non-Purged Filter) . . . . . . . 3.4
2500 Series Terminal Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5
Signal Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
SIB and Optional Output PCB's . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7
Autocalibration Valve Configuration . . . . . . . . . . . . . . . . . . . . . . . . . .3.16
Typical Pressure Transducer Assembly . . . . . . . . . . . . . . . . . . . . . . . 3.24
Purge Connection Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
Cell End Boss Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5
General Cell Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.6

List of Tables
Table 3.1
Table 3.2
Table 3.3
Table 3.4
Table 3.5
Table 3.6
Table 3.7
Table 3.8
Table 3.9
Table 3.10
Table 5.1
Table 5.2
Table 5.3
Table 6.1
Table 6.2
Table 6.3
Table 6.4
Table 6.5
Table 6.6
Table 6.7
Table 7.1
Table 7.2
Table 7.3
Table 7.4
Table 7.5
Table 8.1

vi

Analogue Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8


Relay Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.10
External Control Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11
Externally Powered Solenoid Valve Connections . . . . . . . . . . . . . . 3.14
Internally Powered Solenoid Valve Connections . . . . . . . . . . . . . .3.15
Serial Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.17
Fault Number Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.21
ASCII Characters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.22
RS232 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.22
Pressure Transducer Connections . . . . . . . . . . . . . . . . . . . . . . . . . 3.23
Diagnostic Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
General Fault Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
Serious Fault Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5
General Spares List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
Sample Cell Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2
Source Unit Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
Detector Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
Window Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
Scrubber Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
Recommended Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.5
General Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . 7.1
Performance Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.2
Sampling Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
Sample and Ambient Influence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
Serial Output Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
02500 Safety Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.2

SECTION 1
1.1

INTRODUCTION

Introduction

This manual contains information regarding installation and hardware configuration of


the Servomex 2500 Series analysers (2500GEN/EU1/EU2/FM2, 2510GEN/EU1/EU2/
FM2/HTV, 2520GEN/EU2/FM2, 2550GEN/EU1/EU2/FM2 variants).
GEN General Purpose/Safe area use.
HTV High Temperature Variant.
EU2 Certified for UK/European Zone 2 areas.
EU1 Certified for UK/European Zone 1 areas.
FM2 Certified for US Class 1 Div.2 areas.
Note that the EU1, EU2 and FM2 versions can also be used where flammable samples
are to be monitored in otherwise safe areas.
The Servomex 2500 Series has been designed and manufactured using Servomex
quality systems, which have been certified to ISO9001.
A separate Quickstart manual is also supplied with the 2500, 2520 and 2550 analysers,
reference part number 02500/003C. This details software configuration and operation
of the analyser. An alternative Quickstart manual is supplied with the 2510 analysers,
reference part number 02510/003C. Extra copies of either may be ordered from
Servomex.
Some variants are supplied with a Safety Certificate Manual, reference part number
02500/008C that details the hazardous area approvals and declarations applicable.
Details of the hardware and instructions for servicing, by qualified personnel only, are
presented in the 2500 Service Manual. This may be ordered from Servomex, reference
part number 02500/002C.
Technical assistance and spare parts are available from Servomex outlets (or their
local agents) listed on the back cover.
Use this manual for:
Installation

To take commissioning to the point where the analyser is


powered and operational. The installer is advised to read
this manual completely before commencing installation.

Use the Quickstart manual for:


Configuration

How to set up the clock, passwords, alarm levels, analogue


outputs, relays and other parameters.

Calibration

How to use the manual and automatic calibration/checking


facilities.

Review

How to review configuration and set-up information.

1.1

1.2

General Description

The Servomex 2500 Series is a single beam, multi wavelength process analyser
suitable for monitoring up to three components in a gas sample stream (2550). It is
supplied configured to the customer's precise analytical requirements for a stated
analysis in a specific background stream.
The general layout of the 2500 Series is shown in Figure 1.1. It has been designed for
modern industrial environments with the emphasis on rugged construction, reliable
performance, simple operation and easy servicing. The analyser is controlled using an
on-board microprocessor, which gives flexibility to the user, enabling him/her to
configure the operation of the 2500 to suit the particular process environment. The
2500 is operated via a very simple control panel, which is mounted on the analyser
itself.
The gas or liquid sample to be analysed must be passed continuously through the
analyser's sample cell. The 2500 is designed for continuous 24 hour/day operation and
should not normally be switched off. Versions of the 2500 analyser are suitable for use
in safe and hazardous areas. They are also designed to I.P. 65. (without electrical cell
heater, I.P. 50 with cell heater).

Figure 1.1

2500 Overall View

Note that the EU1, EU2 and FM2 versions can also be used when flammable samples
are to be monitored in otherwise safe areas (see Safety Certificate Manual).

1.2

WARNING
The EU1 versions are intended for use with a suitably certified Purge Control
System (PCS). Do not install a EU1 version in a hazardous area without a
purge controller.
The analyser's software is menu driven and has been designed to be as intuitive as
possible, enabling the user to fully operate and control the 2500 with the minimum of
familiarisation.
NOTE
Full technical specifications for 2500 Series Analysers are presented in section 7 of this
manual.

1.3

Basic Construction

The basic construction of the 2500 is shown in Figure 1.2. It comprises 2 cast end
assemblies (with hinged opening/removable covers), connected by a rigid mounting
beam, or chassis. The sample cell is mounted between the 2 end assemblies and is
removable for cleaning. This configuration results in a single optical axis, i.e. straight
through beam design, with no alignment adjustments necessary. The source end (right
side) contains the soft U.V. or infrared source, which generates a broad beam of
energy across the appropriate spectrum. This is mounted on the chopper box
assembly, which contains the interference filters, mounted on a rotating chopper wheel.
These filters select the appropriate wavelengths for the analysis for which the 2500 is
configured. In most applications, there is one measurement wavelength and one
reference wavelength for each measured component.
The infrared or visible beam, alternating in wavelength, is then focused through the
sample cell, where energy at the measurement wavelength is selectively absorbed by
the components being analysed. The remaining energy falls on the detector in the
detector end.
The sample cell is a simple thick metal construction (316 stainless steel as standard)
fitted with optical windows (each 6 mm thick) to allow passage of the beam. It is
supported between the 2 ends of the 2500 by short cylindrical bosses, which may be
scrubbed or purged, depending on application. The whole cell assembly is easily
removed by loosening the two bosses and lifting it away, as described in Section 5.

1.3

Figure 1.2

2500 Construction

The detector end (left side) contains the detector assembly and also the main
electronic circuit boards. The display and control panel is also mounted in the hinged
lid of this end.
1.4

Options and Accessories

The 2500 supplied may have been fitted with customer specified options and/or
accessories. The full list of available items with brief descriptions is given here.
WARNING
The surface temperature of heated cells can reach or exceed 100C
The following symbol is used on the heated cell:

CAUTION, HOT SURFACE

1.4.1

Heated Sample Cell Electrical

Where fitted, this is powered and controlled by the 2500 itself and is suitable for gas
samples. It may be used in both general purpose and some hazardous areas. The
normal temperature setting for common gas applications is 60C, although it may be
supplied set to any temperature from 30 180C for HTV variants or from 30 130C
for other versions.
1.4

1.4.2

Heated Sample Cell Steam

Where fitted, this is a steam heating coil which must be supplied with low pressure
steam by the user and vented to atmosphere, to maintain a cell temperature of
approximately 100C. This allows cell heating in hazardous areas up to Zone 1/Div. I
classification, and is normally used with the EU1 version only.
1.4.3

Optional Output Cards

As standard, all 2500 Analysers have 2 analogue mA (current) outputs and 3 relay
contact pairs. The analogue outputs are user configured as described in the Quickstart
manual. The relay contacts are user-assigned to any combination of concentration
alarms, Fault alarm, or calibration in progress, as similarly described in the Quickstart
manual. These standard outputs originate from the Sensor Interface Board (SIB PCB)
of the 2500.
An additional output PCB is available as option:
The 02000/916 PCB provides an extra 2 x (mA) plus 2 x (relay) outputs
This additional output PCB can be fitted into the normally vacant slot 2 in the 2500.
These extra outputs are configured along with the standard outputs in the usual way
(See Quickstart manual).
NOTE
Any or all of these analogue outputs may be configured as a voltage output by hardware
link selection, as an alternative to current output.

1.4.4

Compensation Options

The output of any infrared analyser is affected by changes in sample pressure and
sample temperature, unless compensation is applied. Alternatively, these parameters
can be kept constant by a suitable sample conditioning system. Ambient (barometric)
pressure will also affect sample pressure in the analyser if a gas sample is being simply
vented to atmosphere. The following compensation options may be fitted to a 2500.

Sample Pressure Compensation

A solid state pressure transducer is fitted and plumbed into one sample cell connection.
A continuous pressure measurement over the range of the standard transducer, 0 -345
kPa (0 50 psia), is made by the 2500. During calibration in the factory, an empirical
sample pressure/span concentration relationship is determined and stored in memory
to calibrate the compensation. Suitable for gas sample measurements only.
This option is detailed in Section 3.5

1.5

.
WARNING
Standard transducers are not suitable for oxygen service.

Sample Temperature Compensation

A type 'K' thermocouple in stainless steel sheath is supplied, and a continuous


temperature measurement is made by the 2500 over the range -50 to 200C (-58 to
392F). During calibration in the factory, an empirical sample temperature/span
concentration relationship is determined and stored in memory to calibrate the
compensation. Suitable for gas or liquid samples, and for safe and hazardous area
installations. Not available with heated cell.
1.4.5

Application Options

A range of standard and special application options is available with the 2500 and it will
have been supplied configured with one particular option to make the customerspecified measurements. Full details of the application configuration (i.e. optical
materials, operating wavelengths, cell path length, wetted materials etc.) will be found
on the Product Specification document in the front of this manual.

1.6

SECTION 2
2.1

INSTALLATION GENERAL

Introduction

This section provides all the information required to install the 2500 analyser. The
installer is advised to read this section through completely before commencing the
installation.
2.2

Preparation

2.2.1

Tools

Installation of the 2500 requires the use of standard hand tools only, and an Allen key
provided with the instrument to open the end case covers. Preparation of suitable
mounting holes in support brackets or panels may require appropriate power tools.
2.2.2

Safety Precautions

The electrical power used in this equipment is at a voltage high enough to endanger
life therefore normal electrical safety precautions must be observed. Where necessary
a warning is placed prior to the procedure it refers to as shown in the following example:
WARNING
Always ensure that the electrical mains power supply is disconnected before
commencing work.

WARNING
Electrostatic hazard: Clean display and keypad using a clean cloth moistened
with water.
2.2.3

Unpacking and Inspection


WARNING

2500 analysers weigh at least 25 kg (55 lb) and care must be taken when
handling. Use 2 people if necessary and only lift using the yellow handles
provided, not the sample cell.
Lift in the approved manner and remove the 2500 analyser from its packing and inspect
for any damage incurred during transit. If any damage has occurred inform Servomex
or their agents immediately. Retain packing, in case it is required to return the analyser
to the supplier.

2.1

After the initial visual inspection, carry out the following checks. Beware gas strut
action of the hinged covers.

Check that the serial number of the 2500 (located inside source end and also
marked on the external rating label) corresponds to the serial number given in
the Manufacturing Data Sheet found at the front of this manual.

Check that the specification sheet details are in order and agree with the
purchase requirements. Pay particular attention to the Manufacturing Data
Sheet supplied, and any Special Instrument Process Sheets attached.

Check that the accessories are present and undamaged. Standard 2500
accessories provided are: Fuse kit, Allen key, Plastic plugs (refer to section
2.4.2) and Ferrites (refer to section 3.2).

2.3

Electrical Connections
CAUTION

Specific voltage rating information is given on the label located on the outside of the
2500 Analyser and copied within the enclosure. Ensure that the analyser is correctly
configured for the supply voltage.
The 2500 Analyser has no integral switch for disconnecting the electrical supply.
The installer must include a means of isolating the electrical power by means of a
switch or circuit breaker located close to the 2500 Analyser. It must be marked as
the disconnecting device for the analyser and be easily accessible
The electrical supply to the 2500 Analyser MUST be protected by a suitable fuse or
over current protection device rated at no greater than 6 Amps.
Where the main electrical supply is not referenced to ground, an isolating transformer
must be fitted and its secondary winding suitably earthed. The following mains power
cable requirements must be met:

Three core cable with Line/Live (L), Neutral (N), Earth/Ground (E) conductors.

The cable voltage rating should conform to the supply voltage. The type of
cable used should conform to current local regulations, and be suitable for the
installation environment. For use in areas exposed to weather or mechanical
stress, armoured or sheathed cable is recommended.

All electrical cables enter the body of the 2500 Analyser via user-supplied cable
glands. These cable glands must be air tight and/or suitably approved if the
units are to be purged and/or mounted in hazardous areas.

The electrical supply connection terminals are suitable the following cable:
Flexible conductors

0.5 to 1.5mm2 (20 to 16 AWG)

Solid conductors

0.5 to 2.5mm2 (20 to 14 AWG)

Cables should be suitable for temperatures of at least 75C.

2.2

2.4

Installation
WARNINGS

The 2500 Series analysers are not suitable for use in hazardous areas
unless the correct certification labels have been affixed by Servomex
and a suitable purge control system is installed (EU1 versions).

The installer must be satisfied that the 2500 analyser installation


conforms to the relevant safety requirements, national electrical code
and any other local regulations, and that the installation is safe for any
extremes of conditions which may be experienced in the operating
environment of the analyser.

If the equipment is used in a manner not specified by the manufacturer


the protection provided by the equipment might be impaired.

Many process samples, gas streams and test samples are toxic,
asphyxiant, corrosive, flammable or a combination of some or all of
these. It is the responsibility of the installer to ensure that adequate
precautions are taken during installation of the 2500 analyser,
connection of samples, and during any other checks or tests. All
sample line connections must be checked for no leaks.

2.4.1

Positioning

Sensible positioning and an accurate and secure installation will minimise


maintenance, instrument breakdowns and will provide reliable service.
The location should be reasonably vibration free, subject to minimal fluctuations in
ambient temperature and provide reasonable access to the instrument and the
availability of any required services.
If the 2500 is to be mounted outside it must be suitably protected from all extremes of
climatic conditions by a suitable insulated enclosure, paying particular attention to the
ambient temperature range and the rate of temperature change specifications.
EU2 versions must be protected against the entry of solid particles or liquids by a
suitable cover or enclosure where necessary.
EU1 versions must be used with either a Servomex 485 Purge Control Unit with the
associated Servomex 486 Relay Unit, or another suitable Purge Control System, which
must meet the necessary pressurisation requirements, and is supplied with the
necessary pressurisation and control components.
FM2 versions fitted with electically heated sample cells must be installed in locations
which provide adequate protection against the entry of solid foreign bodies and liquids
capable of impairing safety.

2.3

WARNING
EU1 Zone 1 versions must be used with either the Servomex 485 Purge
Control Unit with the associated Servomex 486 Relay Unit, or a third-party
purge controller, which meets the necessary pressurisation and certification
requirements as detailed in the Safety Certificate Manual.
If the 2500 has been supplied with a Servomex 485 Purge Unit with the associated
Servomex 486 Relay Unit, installation information is provided in a separate manual
part no. 00485001C, supplied with the analyser system.
If a gas sample conditioning system is to be connected it must be designed to provide
a filtered, clean, non-condensing sample for the analyser with no risk of a possible
carry- over of condensate into the sample cell.
If a liquid sample conditioning system is to be connected it must be designed to provide
a filtered, single-phase sample for the analyser.
2.4.2

Mounting Details

The following paragraphs provide mounting information for the 2500 Analyser. Use the
yellow lifting handles to manoeuvre the 2500 into position, using two people where
necessary.
The analyser may be mounted vertically or horizontally, ideally at chest height for best
visibility and ease of access, but with some restrictions (refer to Fig 2.1).

CAUTION

For Liquids the analyser should be mounted horizontally with the cell
inlet port at the bottom.

For Gases the analyser should be mounted horizontally with the cell
inlet port on top, especially if there is risk of particulates or
condensates in the sample.

The 2500 may be mounted vertically for gas samples, provided the cell
outlet is at the bottom. This is only suitable if there is no risk of
particulates or condensates in the gas sample.

2.4

When mounting the analyser ensure that the panel or brackets employed are adequate
to take the weight and there is a minimum clearance above the casing of 500mm (20")
to provide space to open the end covers.
Fig 2.2 provides mounting details and a table of mounting hole spacings for the various
cell lengths. Be sure to identify the correct hole spacing.
When the 2500 is to be vertically mounted the display may be rotated 90 to bring it into
a legible position. This is done by removing the internal metal cover and ribbon cable
clamp, and removing the retaining bracket. The display unit can then be withdrawn a
little and rotated. Reassemble the disturbed components in reverse order, carefully
folding the ribbon cable to prevent any stress on it.

Figure 2.1

2500 Positioning Restrictions

2.5

Figure 2.2

2500 Mounting Details

Once mounted, the two yellow lifting handles and their spacers MUST be removed by
unscrewing the securing screws. Press the black pop-in plugs supplied into the vacant
screw holes and retain the handles for possible future use.

2.6

2.5

Instructions Specific to Hazardous Area Installations


(Reference European ATEX Directive 94/9/EC, Annex ll, 1.0.6.)

The following instructions apply to equipment covered by certificate numbers: Sira


02ATEX1395X and Sira 03ATEX2236X
2.5.1

The equipment may be located where flammable gases and vapours of


groups llA, llB and llC may be present. The equipment is only certified for use
in ambient temperatures in the range -20C to +55C for type 'p' approval and
-5C to +55 C for type 'n' approval, and should not be used outside these
ranges.
(Note performance specification is 0C to +55C)

2.5.2

The equipment has not been assessed as a safety-related device (as


referred to by Directive 94/9/EC, Annex ll, clause 1.5).

2.5.3

Installation of this equipment shall be carried out by suitably trained


personnel in accordance with the applicable code of practice (EN 60079-14
within Europe).

2.5.4

Repair of this equipment shall be carried out by the manufacturer or in


accordance with the applicable code of practice (IEC 60079-19).

2.5.5

If the equipment is likely to come into contact with aggressive substances,


then it is the responsibility of the user to take suitable precautions that prevent
it from being adversely affected, thus ensuring that the type of protection is
not compromised.
Aggressive substances

e.g. acid liquids or gases that may attack metals,


or solvents that may affect polymeric materials.

Suitable Precautions

e.g. regular checks as part of the routine


inspections or establishing from the material's
data sheet that it is resistant to specific
chemicals.

2.7

2.6

ATEX Label Information

2.6.1

EU1 T5 (Unheated Cell)

2.6.2

EU1 T4 (Heated Cell)

2.8

2.6.3

EU2 T3

2.6.4

EU2 T4

2.9

2.10

SECTION 3

INSTALLATION ELECTRICAL
WARNINGS

Lethal voltage: mains AC power supplies are potentially lethal. The


installer must ensure that the power supply has been isolated before
commencing installation.

The installer must be satisfied that the 2500 analyser installation


conforms to the relevant safety requirements, national electrical code
and any other local regulations, and that the installation is safe for any
extremes of conditions which may be experienced in the operating
environment of the analyser.

WARNING
The following symbols are used on the warning label fitted on the inside of the
Source End Enclosure Cover.

CAUTION, ELECTRIC SHOCK HAZARD

CAUTION, CONSULT MANUAL

3.1

Electrical Power Connections

For convenience the gas struts may be removed during installation by carefully levering
them off. Ensure that the cable used is suited to the installation environment, and is the
correct rating. Ensure that the fuse is suited to current rating of the analyser and that
all glands and cable connections are secured firmly.
The mains power connection to this instrument is made via an electrical filter unit
attached to the outside of the case. The power cable into this unit does not need to be
screened.

3.1

WARNING (EU1 VERSION)


AC Power for Zone 1 (EU1) versions may only be connected via a suitably
certified Purge Control System. Refer to the PCS manual for details. The
mains power connection to the instrument is made via the purged electrical
filter unit attached to the outside of the case.
Refer to Figure 3.1 a) or b) for illustration.
To connect the 2500 analyser to the mains supply:
1.

Isolate the mains AC supply at source.

2.

Remove filter unit cover.

3.

Fit suitable cable gland to the mains filter unit.

4.

Insert cable through gland and secure.

5.

Wire to terminals.

6.

The mains AC supply is connected internally to TB30. The voltage selection is


made by wire link on TB31 as illustrated in Figure 3.1 a) or b).

7.

Connect the 2 external earthing studs to earth/ground. The RF earth provided


must be locally earthed using as suitable solid conductor connected to a local
physical earth to ensure optimum RFI protection. The external system earth must
be connected to the mounting panel / cubicle / frame / local Intrinsically Safe
earth, as appropriate.
DO NOT APPLY POWER YET.
NOTE

When installed in hazardous locations, appropriate wiring practice must be used


(e.g. within Europe installation must be in accordance with IEC 60079-14)

3.2

Figure 3.1a

2500 AC Power and Valve Connections (Purged Filter)


3.3

Figure 3.1b
3.4

2500 AC Power and Valve Connections (Non-Purged Filter)

Figure 3.2

2500 Series Terminal Locations


3.5

3.2

Signal Connections
NOTE

All signal cables must have a braided overall screen or armour. The screen must be
terminated at the point of entry to the case. This will be by using a gland, which
makes a connection between the cable screen and the case. Beware of ground/
earth loops if the screens are also connected at the user end.
Where two cables are passed through a single entry, a metal gland specifically
intended for two separate cables shall be used.
To minimise the effects of interference from the RF fields, a Steward type 28B0562200 or equivalent ferrite sleeve (Servomex Part No. 2824-0017) shall be placed
over each mA output cable pair. Refer to Figure 3.3.

Figure 3.3

3.6

Signal Cable Connections

3.2.1

Analogue Outputs

Figure 3.4

SIB and Optional Output PCB's

Each analogue output is supplied set for current output and may be configured by the
user to be 0-20mA or 4-20mA, and be assigned a particular range of operation, in
software. This is described in the Quickstart manual. In addition, a "Range 2" setting
for each analogue output can also be configured in software. Maximum impedance is
1k ohm for current output. If desired, each analogue output may be changed to voltage
output of either 0-10V or 2-10V, minimum impedance 1M ohms. This is done by
soldering links into position "LINK 10" and "LINK 11" on the relevant SIB (Sensor
Interface Board) PCB or Option PCB for the first and second analogue output
respectively. See Section 3.2.2. In every case, fit a suitable gland, insert, secure and
strip a suitable cable pair before connecting the cores to the appropriate terminals.

3.7

WARNING (EU1 VERSION)


Where the 2500 is in a Zone 1 area, a suitable Relay Box should be used to isolate
the outputs. This Relay Box should be switched from the Purge Control System.

Each analogue output (and each relay output) has an individual identity in software.
This identity is a number, which relates to the "slot" in which the PCB (SIB or Option
PCB) containing the actual output (or relay) physically resides.
For example, the first analogue output is from the standard SIB PCB which resides in
"slot 1" and therefore its software identity is "1.1" (i.e. first slot first output channel). The
second analogue output is "1.2", and the first relay output is "1.3", etc.
The following tables show the physical locations for connections to the analogue
outputs.
Table 3.1 Analogue Output Connections
Output

Software ID

Analogue 1

1.1

TB 24-4 (+) and TB 24-3 (-)

Analogue 2

1.2

TB 24-2 (+) and TB 24-1 (-)

Analogue 3*

2.1

TB 22-1 (+) and TB 22-2 (-)

Analogue 4*

2.2

TB 22-3 (+) and TB 22-4 (-)

*option PCB
Terminal locations are as illustrated in Figure 3.2.

3.8

Terminal Connections

3.2.2

Analogue Output Link Selections on Sensor Interface Board (SIB) PCB

After connecting up the required analogue outputs, the user should make the following
hardware link selections on the SIB PCB and (where fitted) Option PCB(s). Refer to
Figure 3.3.4 for general schematic of both these types of PCB. Carefully remove the
PCB for link selection and be sure to replace it in the correct slot and in the correct
orientation.

Current/Voltage
Each output may be individually selected as current (0/4-20mA) or voltage
(0/2-10V). Factory setting is CURRENT.
Analogue o/p No.1
Analogue o/p No.2

Current:

Link 10 OPEN

Voltage:

Link 10 CLOSED (solder link)

Current:

Link 11 OPEN

Voltage:

Link 11 CLOSED (solder link)

Action on System Failure


To enhance fail-safe operation, each output may be selected to drive HIGH
(to 20.5mA/10.2V) or LOW (to 0 mA/0V.) if the 2500 detects a system failure
(Shutdown state, i.e. serious fault and measurement invalid). Factory setting is
HIGH.
Analogue o/p No.1

HIGH on Shutdown: SW1/3 OFF


LOW on Shutdown: SW1/3 ON

Analogue o/p No.2

HIGH on Shutdown: SW1/2 OFF


LOW on Shutdown: SW1/2 ON

Delay on Fail Safe Action


The detection of system failure (Shutdown, i.e. serious fault/measurement
invalid) is practically instantaneous and therefore a sufficiently large, rapid,
mains transient or brown-out may also cause the fail-safe action described
above to occur immediately. In installations where this would be undesirable,
the fail-safe action may be delayed for 5 seconds to overcome such possible
transient effects.
Factory setting is NO DELAY.
Both Analogue o/p

DELAY on fail-safe:

SW1/1 OFF

NO DELAY on fail-safe:

SW1/1 ON

3.9

3.2.3

Relay Outputs
NOTE

Each relay output is supplied set for NC (i.e. Normally Closed under "Safe"
condition) operation, and may be assigned to a particular alarm or signalling
function in software. This is described in the Quickstart manual.
User selection of NC or NO operation is described in Section 3.2.4. As with the
analogue outputs, each relay has its own software identity relating to the slot in which
the PCB, on which it is physically located, resides. The following tables show the
physical locations for connections to the relay outputs. All relay outputs are rated
240VAC/1.0A or 30VDC/1.0A. In each case, fit suitable glands and cable and connect
to the relevant terminals. Select the correct procedure according to the architecture.
Table 3.2 Relay Output Connections
Output

Software ID

Terminal Connections

Relay 1

1.3

TB 23-1 and TB 23-2

Relay 2

1.4

TB 23-3 and TB 23-4

Relay 3

1.5

TB 23-5 and TB 23-6

Relay 4*

2.3

TB 22-5 and TB 22-6

Relay 5*

2.4

TB 22-7 and TB 23-7

*option PCB
Terminal locations are as illustrated in Figure 3.2.
3.2.4

Relay Output Link Selection

After connecting up the required relay output, the NC/NO selections may be altered on
the SIB PCB and (where fitted) output Option PCB(s). Refer to Figure 3.4 for general
schematic and carefully remove and replace the PCB as before, if a change is to be
made.
Normally Closed/Normally Open (NC/NO)
The factory setting of all relay output wire links is soldered in the NC position. This
means that the contacts will open to signal an alarm. The contacts will also be open if
there is no power applied to the 2500, i.e. they are fail-safe. If, however, it is required
to change to NO operation (i.e. close to signal an alarm) this can be done by cutting
and reordering the wire links as listed below.

3.10

WARNING
Setting the relays to Normally Open (N.O.) operation will mean that there is no
fail-safe action since any loss of cable continuity will prevent an alarm being
signalled.

NOTE
Regardless of N.O./ N.C. setting, the relays will always be in "alarm" state wherever
there is no power applied to the 2500, or when they are unassigned.

Relay o/p No. 1

N.C. Link 3 A

N.O. Link 3 B

Relay o/p No. 2

N.C. Link 4 A

N.O. Link 4 B

Relay o/p No. 3

N.C. Link 5 A

N.O. Link 5 B

3.3

Control Connections
NOTE

All control cables must have a braided overall screen or armour. The screen must be
terminated at the point of entry to the case. This will be by using a gland, which
makes a connection between the cable screen and the case. Beware of ground/earth
loops if the screens are also connected at the user end.
Where two cables are passed through a single entry, a metal gland specifically
intended for two separate cables shall be used.
As standard the following connections are provided on the 2500, and these are made
to the 2500 Analyser PCB's utilising a suitable gland in an adjacent hole. Refer to
Figure 3.2 and Table 3.3.
Table 3.3 External Control Connections
PCB

Function

Connections

Transmitter PCB
02500904A

Range Change

TB26 1
TB25 3

Transmitter PCB
02500904A

Autocalibration Initialise

TB26 2
TB25 3

Transmitter PCB
02500904A

Password Keyswitch

TB26 3
TB25 3

Power Control PCB


02500911A

Flow Failure

TB36 1
TB36 2
3.11

3.3.1

Range Change Input

Every analogue output can be independently scaled in software to represent some or


all of the calibrated range of the instrument, as described in the Quickstart manual.
However, under some process conditions (e.g. plant start-up) it may be desirable to
have different output ranges on demand. When the Range Change Input is used all
analogue outputs will change to their user-designated "Range 2" settings. Note that
"Range 2" settings can be identical to "Range 1" settings if desired.
NOTE
"Range 2" is selected as long as contacts remain closed.
Range 2 settings selected by Contact closure between:
TB26 1 and TB25 3

3.3.2

Autocalibration Initiation Input

Autocalibration/Autocheck may be triggered by the user from the control panel or by


using the 2500's own real time clock system. Alternatively, it may be triggered at will
using a manual remote switch or by a host device, closing a contact.
Autocalibration initialised by contact closure between:
TB26 2 and TB25 3
NOTE

Contacts must be closed for at least 2 seconds but no more than 59 seconds.

Remote initiation starts the "one cycle" routine using all the user-set
parameters. It will not repeat unless contact closure is made again.

User is advised to set the Autocalibration period in software to zero


(i.e. internal timer disabled) when using remote Autocalibration initiation,
to prevent duplication of Autocalibrations.

3.3.3

Password Keyswitch Input

As an alternative to the use of passwords via the control panel, a simple Keyswitch can
be connected which enables access to both Operator and Supervisor levels without
password demands. The user must first set SW1/8 to ON on the Microprocessor PCB,
to enable this feature. See the Quickstart Manual for more details.
Keyswitch setting selected by Contact closure between: TB26 3 and TB25 3

3.12

.
WARNING
If the Keyswitch is left "ON" (contacts closed), access is continuously
available to all normally protected areas of the software.
3.3.4

Sample Flow Sensor Input

If the sampling system supplying the process sample to the 2500 is fitted with a loss of
flow sensor, this can be connected to the 2500 to enable the "loss of sample flow"
diagnostic to operate, which in turn triggers the analyser's Fault alarm. See Figure 3.1.
2500 Power Control PCB 02500911A
Loss of flow diagnostic triggered by contact closure between: TB36 1 and TB36 2.
NOTE
In common with all fault diagnostics this alarm will only be cleared by restoration of
flow/opening of contacts.

WARNING
Where the 2500 analyser is in a hazardous area, suitable protection concepts
must be used to connect the loss of flow sensor to the 2500 analyser.
3.3.5

Solenoid Valve Drives

Autocalibration routines are provided as standard on the 2500, and if they are to be
used, suitable 3-way solenoid valves can be plumbed into the sampling system as
shown in Fig 3.5. Two or three valves may be used and all should be located
conveniently close to the 2500. Each valve is switched via the 2500 Power Control PCB
relays as illustrated. The valves are selected for operation by the 2500 as shown in
Table 3.4 and Table 3.5. Connect the valves to the appropriate terminals using suitable
screened cables and glands, and attach screens to the nearest earth stud with the
minimum length of screen. See Figure 3.1.
The Autocalibration relay contacts are rated at 240VAC/1.0A, 30VDC/1.0A. The
contacts are fitted with an R-C snubber (100R + 47nF) to provide transient protection
when used with AC solenoids. 220/240Vac solenoid valves with a power rating less
than about 6VA are not recommended because the 'leakage' current through the
snubber may prevent low power solenoids from de-energising correctly.
Solenoid valves may use an external power source, or the internal 24VDC/12VA power
source may be used. When using the internal power source, ensure that the 12VA
rating is not exceeded. DC solenoid valves must have built-in suppression diodes fitted
across the coil to protect the relay contacts.

3.13

WARNING
If installation is in a hazardous area, use suitable certified solenoid valves.

NOTE
Valve (SV3) reverts to OFF in Shutdown (Serious Fault) condition. The use of the
optional 3rd "Sample/Inert valve will prevent sample entering the 2500 sample cell
while the cell or chopper box under temperature diagnostic is activated, i.e. heated
cell is not fully warmed up. This can be used whether or not Autocalibration is used
and will be appropriate in some applications: e.g.

if there is a possibility of condensation forming in the sample cell during warm


up, then nitrogen or another purge gas must be used to flush the cell during
it's warming up period.

if a hazardous sample is being monitored, the cell should be purged in the


event of a serious fault.

3.3.6

Externally Powered Solenoid Valve Connections

Connect via solenoid valve relays as detailed. Ensure that the relay rating is not
exceeded.
Table 3.4 Externally Powered Solenoid Valve Connections
Valve State (ON=Current Flow Through
Solenoid)

3.14

Span/Zero
Valve SV1
TB34 1
TB34 2

Cal/Sample
Valve SV2
TB34 3
TB34 4

Sample/Inert
Valve SV3
TB34 5
TB34 6

Cell or Chopper Box


Under Temperature

either

either

OFF(Inert)

NORMAL STATE OFF

OFF (Zero)

ON(Sample)

ON(Sample)

Zero Sample Required

OFF (Zero)

ON(Cal.)

ON(Sample)

Span Sample Required

ON(Span)

ON(Cal.)

ON(Sample)

3.3.7

Internally Powered Solenoid Valve Connections

Connect to solenoid valve relays SV1, SV2 and SV3 as detailed below. Take care to
ensure that internal 12VA (total) rating of the 24VDC supply is not exceeded.
Table 3.5 Internally Powered Solenoid Valve Connections
Valve State (ON=Current Flow Through
Solenoid)
Span/Zero
Valve SV1
TB34 1 (-)
TB36 6 (+)

Cal/Sample
Valve SV2
TB34 3 (-)
TB36 5 (+)

Sample/Inert
Valve SV3
TB34 5 (-)
TB36 5(+)

Cell or Chopper Box


Under Temperature

either

either

OFF(Inert)

NORMAL STATE OFF

OFF (Zero)

ON(Sample)

ON(Sample)

Zero Sample Required

OFF (Zero)

ON(Cal.)

ON(Sample)

Span Sample Required

ON(Span)

ON(Cal.)

ON(Sample)

NOTE
To ensure internally powered solenoid valve function, fit links between:
TB34 2 and TB36 4
TB34 4 and TB36 3
TB34 6 and TB36 3

CAUTION
TB36 terminals 5 and 6 (+24V) are always live and should not be grounded.

3.15

Figure 3.5
3.4

Autocalibration Valve Configuration

Digital Connections
NOTE

All digital cables must have a braided overall screen or armour. The screen must be
terminated at the point of entry to the case. This will be by using a gland, which
makes a connection between the cable screen and the case. Beware of ground/earth
loops if the screens are also connected at the user end.
Where two cables are passed through a single entry, a metal gland specifically
intended for two separate cables shall be used.
Digital connections in RS/EIA-232 format are provided for attaching a local portable
printer to receive set-up information directly down loaded from the 2500, or to report
measurement and analyser status information to a DCS or datalogger.
3.4.1

RS-232 Connection

See the Quickstart manual for configuration of the RS-232 port and operation of this
function. See Figure 3.2 and Table 3.9 for connections. The format of the serial data
stream is as follows:

3.16

Table 3.6 Serial Data Format


Item

Size

Description

<CR>

1 byte

Carriage return character (ASCII code 13).

date;

8 bytes

dd/mm/yy or mm/dd/yy depending on user


settings.

time;

8 bytes

hh:mm:ss

number of
components

1 byte

1 byte Number of components fitted


(range 1-3)

<CR><LF>

2 bytes

Carriage return and Line Feed characters


(ASCII codes 13 and 10 respectively).

component 1 formula; Max 6 bytes

Chemical formula defined for component 1.

component 1
concentration;

Max 5 bytes

Concentration for component 1 as defined


for measure display.

component 1 units;

Max 3 bytes

Units defined for component 1.

component 1 alarm
status;

4 bytes

One byte for each alarm. Set to alarm


number (1,2,3 or 4) when alarm is raised,
<space> when alarm not raised

component 1
Autocalibration
status;

1 byte or
15 bytes

Indicates Autocalibration phase.


0 = Not in Autocalibration
1 = In span preflush
2 = In zero cal
3 = zero corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
zero before calibration and the zero after
calibration respectively.
4 = In span cal
5 = span corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
span before calibration and the span after
calibration respectively.
6 = In post flush

<CR><LF>

2 bytes

Carriage return and Line Feed characters


(ASCII codes 13 and 10 respectively).

Note: The data for components 2 and 3 will only be present on a 2550
instrument.
component 2 formula; Max 6 bytes

Chemical formula defined for component 2.


(if fitted)
3.17

Table 3.6 Serial Data Format


component 2
concentration;

Max5 bytes

Concentration for component 2 as defined


for measure display. (if fitted)

component 2 units;

Max 3 bytes

Units defined for component 2. (if fitted)

component 2 alarm
status;

4 bytes

One byte for each alarm. Set to alarm


number (1,2,3 or 4) when alarm is raised,
<space> when alarm not raised. (if fitted)

component 2
Autocalibration
status;

1 byte or
15 bytes

Indicates Autocalibration phase.


0 = Not in Autocalibration l
1 = In span preflush
2 = In zero cal
3 = zero corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
zero before calibration and the zero after
calibration respectively.
4 = In span cal
5 = span corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
span before calibration and the span after
calibration respectively.
6 = In post flush

<CR><LF>

2 bytes

Carriage return and Line Feed characters


(ASCII codes 13 and 10 respectively).

component 3 formula; Max 6 bytes

Chemical formula defined for component 3.


(if fitted)

component 3
concentration;

Max 5 bytes

Concentration for component 3 as defined


for measure display. (if fitted)

component 3 units;

Max 3 bytes

Units defined for component 3. (if fitted)

component 3 alarm
status;

4 bytes

One byte for each alarm. Set to alarm


number (1,2,3 or 4) when alarm is raised,
<space> when alarm not raised. (if fitted)

3.18

Table 3.6 Serial Data Format


component 3
Autocalibration
status;

1 byte or
15 bytes

Indicates Autocalibration phase.


0 = Not in Autocalibration
1 = In span preflush
2 = In zero cal
3 = zero corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
zero before calibration and the zero after
calibration respectively.
4 = In span cal
5 = span corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
span before calibration and the span after
calibration respectively.
6 = In post flush

<CR><LF>

2 bytes

Carriage return and Line Feed characters


(ASCII codes 13 and 10 respectively).

Note: End of 2550 specific data


fault status;

1 byte

fault numbers;

Set to 'F' when fault is active, <space> when


no fault active
List of fault numbers of active faults,
separated by ','. Fault numbers are defined
in table 3.7.

<CR><LF>

2 bytes

Carriage return and Line Feed characters


(ASCII codes 13 and 10 respectively).

Chopper
temperature; C;

Max 4 bytes

Chopper temperature as displayed in


diagnostic display.

Compensation
temperature; C;

Max 4 bytes

Compensation (detector) temperature as


displayed in diagnostic display.

cell/sample
temperature; C;

Max 5 bytes

Cell/sample temperature as displayed in


diagnostic display.

source voltage; V;

Max 5 bytes

Source voltage as displayed in diagnostic


display.

Intensity 1;Intensity 2; Max 6*6


Intensity 3;Intensity 4; bytes
Intensity 5;Intensity 6;

Intensity channel values diagnostic values


representing up to six input positions used
for measure and reference signals.

<CR><LF>

Carriage return and Line Feed characters


(ASCII codes 13 and 10 respectively).

2 bytes

3.19

Table 3.6 Serial Data Format


checksum;

4 bytes

ASCII representation of 16 bit checksum


(modulo 65536), calculated by adding all
data preceding the checksum together.

<CR><LF>

2 bytes

Carriage return and Line Feed characters


(ASCII codes 13 and 10 respectively).

<CR><LF>

2 bytes

Carriage return and Line Feed characters


(ASCII codes 13 and 10 respectively).

Serial port parameters are as follows:


Baud rate: 2400
Parity: Even
Stop Bits: 1
Data Bits: 8
Handshake: hardware
Example Data:
Note: Checksums have not been calculated
Normal Operation Example Output
<CR>09/01/99;13:42:30;2;<CR><LF>CO;0.5;%; 2 4;0;<CR><LF>CO2;3.8;%; ;0;<CR><LF> ;;<CR><LF>50.0;51.5;183.2;5.0;
123456;521452;319322;4235421;478525;368965;<CR><LF>????;<CR><LF><CR><LF>

3.20

Table 3.7
Fault Number
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38

Fault Number Reference


Display Message
OPTICAL BENCH POWER FAILURE
SOURCE VOLTAGE HIGH
SOURCE VOLTAGE LOW
INFRARED SOURCE FAILURE
CHOPPER MOTOR OUT OF LOCK
SAMPLE FLOW FAILURE
CHOPPER TEMP HIGH
CHOPPER TEMP LOW
CHOPPER TEMP SENSOR FAILURE
CELL TEMP HIGH
CELL TEMP LOW
CELL TEMP SENSOR FAILURE
SAMPLE TEMP SENSOR FAILURE
PRESSURE SENSOR FAILURE
BAD REFERENCE VOLTAGE
COMPENSATION T SENSOR FAILURE
DETECTOR SIGNAL HIGH
DETECTOR SIGNAL LOW
Reserved
Reserved
Reserved
AUTO CAL BAD PREFLUSH
ZERO OUT OF TOLERANCE
SPAN OUT OF TOLERANCE
Reserved
Reserved
Reserved
Reserved
Reserved
PASSWORD VIOLATION
SYSTEM CLOCK INOPERATIVE
SYSTEM ADC INOPERATIVE
DEFAULT SYSTEM DATA CORRUPTION
CALIBRATION DATA CORRUPTION
UTILITY DATA CORRUPTION
ALARM/RELAY DATA CORRUPTION
ANALOGUE ASSIGN DATA CORRUPTION
BAD SPAN SPAN IGNORED

3.21

Some characters defined by the 2500 are non-standard characters, i.e. not part of the
ASCII character set. These either occupy the ASCII codes above 127 or redefine an
ASCII code as another character. When these characters are used, for example in the
component formula, they will be substituted with a valid ASCII character prior to output
on the RS232. The non-standard characters and their substitute values are as follows:
Table 3.8 ASCII Characters
ASCII Code

Character

Substitute Character

126

<space>

127

<space>

128

129

130

131

132

133

134

135

136

137

138

<space>

139

<space>

140

<space>

141

??

<space>

142

<space>

143

<space>

Table 3.9 RS232 Connections


2500 Terminals

Connector

External RS232 Terminals

0V (Ref.)(0VR)

TB20 1

Signal GND/0V

Clear to Send (CTS)

TB20 3

Data Terminal Ready (DTR)

Transmit (Tx)

TB20 4

Receive (Rx)

3.22

WARNING
Do not use an uncertified dcs, datalogger or printer in a hazardous area.

3.5

Pressure Transducer Connections (If Supplied)

The pressure transducer cabling is terminated to the source end right hand side (RHS)
of the analyser.
The pressure transducer is supplied pre-wired to the analyser. The cabling is
connected to PCB 02500911A TB35 as shown in Table 3.10
.
Table 3.10
PCB 02500911A

Pressure Transducer Connections

Interconnecting Cable

Pressure Transducer

TB35 5 (signal +)

Identified '1'

Terminal 1 (+)

TB35 1 (signal -)

Identified '2'

Terminal 2 (-)

The general layout of the pressure transducer option is illustrated in Figure 3.6.

3.23

Figure 3.6

3.24

Typical Pressure Transducer Assembly

SECTION 4
4.1

INSTALLATION GAS CONNECTIONS

Purge Connections (EU1)

The next phase of the installation is to connect up any safety purge system required for
safe operation in a hazardous area.
4.1.1

General Purge Connections


WARNING (EU1 VERSION)

Where appropriate, the user must refer to the detailed instructions provided
in the manual supplied with the Servomex 485/486 purge units or other Purge
Control System, and use them in conjunction with the following instructions.

Figure 4.1

Purge Connection Schematic

4.1

The general schematic for a hazardous area purge system is shown in Figure 4.1.
WARNINGS (FIGURE 4.1)
1.

If this component is mounted in a hazardous area it should be protected


by one of the norms listed in EN 50014.

2.

Cable glands used in EEx 'd' apparatus shall be approved Ex 'd'


components, which maintain the degree of protection necessary.

3.

All cable glands for use in electrical cable entries shall be of a type, which
seals and clamps the cable.

NOTES (FIGURE 4.1)


1.

The PCS shall include an automatic device, which operates when the
overpressure falls below the minimum prescribed value. This device operates
in turn a pressure switch, which can be used for alarm and/or isolation
purposes.

2.

Where the PCS is used to protect the 2500 bench unit only, the protective gas
supply shall be connected from the PCS to the 2500. The minimum level of
overpressure shall be monitored at or close to the outlet aperture of the 2500.
The maximum level of overpressure shall be monitored at or close to the inlet
and the PCS shall incorporate a limiting device to ensure that the maximum
overpressure specified in normal service (80 mBar) is not exceeded.

3.

A protective device shall be fitted to the purge exhaust aperture to prevent


sparks or incandescent particles from being ejected from the pressurised
enclosure.

4.

The PCS shall require to be put through a purge cycle after a pressure loss
before power can be restored to the instrument.

5.

The PCS shall contain a pressure sensing element which is factory set, to a
level not less than the specified minimum (50 Pa, 0.5 Mb) and shall measure
pressure at the point in the system where the lowest pressure occurs.

6.

The PCS shall supply a flow rate of protective gas, not less than 10 SLPM, and
shall incorporate an automatic device to monitor flow during the purge time
specified below.

7.

The PCS shall maintain the minimum specified overpressure during the purge
phase and afterwards.

8.

The PCS shall incorporate a purge timer, which shall be set to not less than
16 minutes. The purge timer shall be set to zero if either the minimum
overpressure or minimum purge flow is not maintained.

9.

The PCS shall incorporate a device to ensure that power is not connected to
the apparatus being protected until the specified purge time has elapsed.

4.2

4.2

End Boss Purge Connection

Depending on application, fittings for 1/8" o.d. tube may be provided to purge the end
bosses on either side of the sample cell and these should now be connected. Refer to
the following application notes and Figure 4.2 where necessary.
4.2.1

Introduction

The 2500 measuring cell is mounted on two short tubes or "end bosses" which
preserve the integrity of the optical system, while providing thermal isolation. The
measuring cell and the analyser enclosure are both independently sealed using
windows or lenses and elastomeric seals. The end bosses form the interface between
the two, and the implications of a cell window leak must be considered when
flammable, corrosive or toxic gases are measured.
Under ideal circumstances, the gas in the end bosses should be totally transparent to
the wavelengths of infrared energy used for the measurement. However, simply filling
the bosses with nitrogen and sealing them is not always satisfactory, because the
elastomeric seals are permeable and atmospheric gases such as carbon dioxide or
water vapour can leak in. Specifically, a problem will arise when the gas being
measured is also carbon dioxide or water vapour, or when the measurement is cross
sensitive to either of those gases: the effect of a gradual ingress of the atmospheric
gases will be seen as drift in the measurement. A further problem arises with low-level
measurements of carbon monoxide. In this case it is out-gassing of carbon monoxide
from the seals themselves which can cause drift.
4.2.2

End Boss Scrubbers

Problems, relating to cross interference within the end bosses, can usually be resolved
by fitting chemical "scrubbers" to the end bosses. These are small metal housings
packed with a chemically active material, which absorbs or reacts with the problem gas.
A number of materials are available which absorb water vapour (molecular sieve),
carbon dioxide (soda lime) and carbon monoxide (Hopcalite).
When a 2500 is specified, a suitable scrubber material is stipulated for the end bosses,
where necessary, and the analyser will be delivered with these scrubbers fitted.

NOTE
If the analyser is dismantled for any reason, the scrubbers must not be left exposed
to normal atmosphere for more than a few minutes or they will be exhausted.
Similarly, after long-term use they may need to be replaced, and this must always
be with the correct type.

4.3

4.2.3

End Boss Vents

When a flammable, severely corrosive or toxic gas or liquid is present in the


sample cell, either as the sample or a background gas, additional issues arise.
The construction of the cell window seal is such that any leak of the seal or failure of
the window will result in the sample gas entering the end boss, but not being released
to the surrounding atmosphere or directly into the analyser enclosure. This is a
deliberate safety feature of the design. However it is necessary that steps are taken to
deal with the leak so that no permanent hazard arises, and permanent sealing of the
end bosses and the use of scrubbers is not generally acceptable. Instead, the end
bosses will be supplied with adaptors for connecting vent or purge lines.
If the measurement is not affected by atmospheric gases, then a single vent connection
of at least 4mm I.D. should be made to each end boss, which will vent to atmosphere
in a safe location where release of the sample will not cause a hazard. If the sample
pressure exceeds 50kPa (7 psig), the vent should be 8mm I.D., and the sample cell
outlet connection should be fitted with a non return valve, and the sample cell inlet
connection should be fitted with a restrictor which will limit the total flow to less than
10 litres/min in the event of total window failure.
4.2.4

End Boss Purge

If the measurement is affected by atmospheric gases, then in addition to the vent


described above, the second connection to each end boss should be connected to a
supply of dry (1) nitrogen (2) regulated to a flow of 100ml/min using a suitable flow
meter. Ensure that the end boss purge is always vented to atmospheric pressure and
the pressure within the end boss dead volume space does not exceed 2.5psig.
Suitable precautions should be made to ensure that, in the unlikely event of a
catastrophic failure of the cell window /sealing, sample gas is prevented from
contaminating the purge gas supply.
(1)

Dry means a dew point of below -30 deg C.

(2)

Dry CO2-free air may also be used in most applications except low-level CO2.

Hazardous Area Certification flammable samples, additional note


The analysers are certified for use in either Zone 2 / Div II or Zone 1. In all cases this
certification does not permit the analyser (or any part of its enclosure) to be exposed to
a permanently flammable atmosphere. Sealing the end-bosses with scrubbers when
measuring a flammable sample would mean that should there be a cell window leak,
the part of the enclosure that the end boss seals against would become exposed to a
permanently flammable atmosphere. Venting or purging the end bosses resolves this
issue.

4.4

NOTE
Vent lines, if required, should be connected to end boss bottom tappings. Purge gas
should be connected to top tappings when required. When not required, top tappings
should be plugged.
IF FITTED, COMMENCE END BOSS PURGE FROM THIS POINT ONWARDS.

WARNING
Where the sample is corrosive or toxic, the vent lines from the end bosses
must take away any potential leakage to a safe disposal point.

Figure 4.2

Cell End Boss Connection

4.5

4.3

Steam Heated Cell

Some 2500's are fitted with steam heated cells, and at this stage suitable low pressure
steam, typically at 14-21kPa, 2-3 psig venting to atmosphere after the cell, should be
connected ready for use, but do not pass steam into the heating coil yet. Fittings for
1/8" o.d. tube are normally provided.
4.4

Process Stream Connection

4.4.1

Basic Operation

The process stream connections to and from the 2500 sample cell inlet and outlet
should now be made. The inlet and outlet are 1/4" o.d. stubs. Connections will normally
be from a local sample conditioning system. Ensure that any pressure compensation
transducer is correctly fitted. Ensure that all connections are made securely with the
correct compression fittings, of suitable materials. Gas samples should enter the cell
from above. (Refer to Figure 2.1).
Do NOT pass process samples into the 2500 yet.

4.4.2

High Integrity Operation

This is the recommended mode of operation and where appropriate, the process
sample can be routed via the "Sample/Inert" solenoid valve, controlled by the 2500,
which will only permit sample to enter the sample cell when it is fully warmed up and
no serious faults are present. See Section 3.3.5.
Do NOT pass process samples into the 2500 yet.
WARNING
In either case for toxic/flammable samples ensure that the sample inlet line
has a suitable restrictor inserted to limit sample flow in case of catastrophic
failure of sample tubing/cell connections.
4.5

Power-Up

1.

Confirm that all electrical and plumbing connections have been correctly made as
described in the preceding sections.

2.

Ensure purge gas is applied to the Purge Controller (Where Fitted).

3.

Apply power to the 2500 (via the Purge Control System for EU1 Versions).

4.

Apply steam to any steam-heated cells fitted.

5.

The 2500 will show the start-up and Identity display as follows:
Self-checks (Fault and Alarm LEDs tested)
System O.K.
Analyser identity/Software revision number.
Serial number and order number.
Calibration (measurand, range, units).
Commence measurement.

4.6

6.

The normal measurement display will then appear and the 2500 will be
operational.

NOTES

The measurement will not be valid at this point since process sample is not
yet flowing.

If a heated cell is fitted, it will not yet be up to temperature and the intermittent
"warming up" message will be displayed. Depending on set point and/or
environment this may take several hours to reach temperature and the
message to cease.

The 2500 chopper box will also take up to 2 hours to reach specified
temperature and the "warming up" message will also be displayed until this is
accomplished.

WARNING
Check for no leaks of sample at full operating pressure and temperature
before proceeding. This is especially important for toxic/flammable samples.
While the analyser is warming up, the software configuration to suit the user's particular
requirements may now be accomplished. Proceed to the Quickstart Manual for details
of configuration. Note that during the warming up period, all outputs and alarms are live
and fully operational. All diagnostics are also operational, EXCEPT cell temperature
and chopper box temperature. These last two diagnostics only become operational
once the temperature set points have been reached or after 2 hours (for chopper box)
or 10 hours (for cell) have elapsed from power-up.
At this point process sample may be turned on and allowed to flow through the
2500 sampling system.

NOTE
Where the 3rd (sample/inert) solenoid valve is in use (see Section 3.3.5) the inert
medium will automatically be supplied to the analyser until the cell reaches correct
temperature.

4.7

4.8

SECTION 5
5.1

FAULT DIAGNOSIS AND CELL MAINTENANCE

Introduction

A fault will be signalled wherever any parameter measured by the 2500's diagnostics
becomes out of tolerance. Any Fault condition will be indicated by the Fault LED on the
control panel being illuminated.
The Fault signal can be assigned to a Relay, as described in the Quickstart Manual.
Faults are categorised as either:
1.

General Faults parameter out of tolerance, intervention is required as


soon as possible. The analyser is still operational and measuring.

2.

Serious Faults (Shutdown) performance seriously degraded,


measurement withdrawn and analogue outputs driven HIGH (or low, as
configured in Section 3.2.2).

NOTE
For detailed instructions on all Service and Maintenance operations for suitably
skilled personnel, please refer to the 2500 Series Service Manual part number
02500/002C.
5.2

Diagnostics

The diagnostics listed in Table 5.1 can be interrogated via the DIAGNOSTICS function
in DISPLAY menu. See Quickstart Manual.

5.1

Table 5.1 Diagnostic Displays


Display

Description

CHOPPER TEMP C

Chopper box temperature

COMPENSATION C

Detector temperature

CELL TEMP C

Cell temperature (if enabled)

SAMPLE TEMP C

Sample Temperature *

[Measurement 1] ABS au

Absolute absorbance units (M1)

[Measurement 2] ABS au

Absolute absorbance units (M2) *

[Measurement 3] ABS au

Absolute absorbance units (M3) *

SAMPLE PRESSURE (units)

Sample pressure *

SOURCE VOLTS V

Source voltage

MOTOR DRIVE %
SET POINT %

Motor drive power and set point

DETECTOR SIG M1

Detector signal level (M1) measurement

DETECTOR SIG M2

Detector signal level (M2) measurement *

DETECTOR SIG M3

Detector signal level (M3) measurement *

DETECTOR SIG R1

Detector signal (R1) reference

DETECTOR SIG R2

Detector signal (R2) reference *

DETECTOR SIG R3

Detector signal (R3) reference *

TEMP CORR M1

Temperature Correction (M1)

TEMP CORR M2

Temperature Correction (M2) *

TEMP CORR M3

Temperature Correction (M3) *

TEMP CORR R1

Temperature Correction (R1)

TEMP CORR R2

Temperature Correction (R2) *

TEMP CORR R3

Temperature Correction (R3) *

* (IF FITTED).

5.2

5.3

General Fault Conditions

When the Fault LED is illuminated, the DISPLAY FAULTS function should be used to
determine the nature of the fault. The diagnostics (Section 5.2) may assist in defining
the cause, together with the comments given in Table 5.2. Action to cure the fault
should be taken as soon as possible.
Table 5.2 General Fault Messages
Display Message

History Log
Entry

Description/comments

SOURCE VOLTAGE
HIGH

SOURCE HIGH

Possible Power Control PCB fault

SOURCE VOLTAGE
LOW

SOURCE LOW

Possible Power Control PCB fault

SAMPLE FLOW
FAILURE

FLOW FAIL

User Sample Flow Alarm active

CHOPPER TEMP HIGH

CH TEMP HIGH

Ambient temp. too high/control


problem

CHOPPER TEMP LOW

CH TEMP LOW

Ambient temp. too low/control


problem

CHOPPER TEMP
SENSOR FAILURE

CH TEMP FAIL

Chopper sensor failed/


disconnected/outside range.

CELL TEMP HIGH

CELL T HIGH

Possible Power Control PCB fault

CELL TEMP LOW

CELL T LOW

Possible Power Control PCB fault

CELL TEMP SENSOR


FAILURE

CELL T FAIL

Cell sensor failed/disconnected/


outside range

SAMPLE TEMP
SENSOR FAILURE

SAMPLE T FAIL

Sample sensor failed/disconnected/


outside range

PRESSURE SENSOR
FAILURE

NO PRESSURE

Sample sensor failed/disconnected/


outside range

BAD REFERENCE
VOLTAGE

REF OLT FAIL

Possible Transmitter PCB fault

DETECTOR SIGNAL
LOW

DET SIG LO

Probable cell contamination clean


cell

COMPENSATION T
SENSOR FAILURE

COMP T FAIL

Detector sensor failed/disconnected

ZERO OUT OF
TOLERANCE

ZERO OUT TOL

Failed Autocalibration bad zero


sample use One Cycle or Manual
Zero

5.3

Table 5.2 General Fault Messages


SPAN OUT OF
TOLERANCE

SPAN OUT TOL

Failed Autocalibration bad span


sample use One Cycle or Manual
Span

AUTOCAL BAD
PREFLUSH

BAD PREFLUSH

Failed Autocalibration bad


preflush use One Cycle

PASSWORD
VIOLATION

BAD PASSWORD

3 consecutive Password failures

SYSTEM CLOCK
INOPERATIVE

BAD CLOCK

Reset time or replace Clock chip

DEFAULT SYSTEM
DATA CORRUPTION

BAD SYS DATA

Factory calibration corrupt use


Associate

SYSTEM ADC
INOPERATIVE

BAD ADC

Possible SIB PCB fault

CALIBRATION DATA
CORRUPTION

BAD CAL DATA

User calibration corrupt repeat


calibration

UTILITY DATA
CORRUPTION

BAD UTI DATA

Utility file corrupt renew all entries

ALARM/RELAY DATA
CORRUPTION

BAD RLY DATA

Alarms/Relays file corrupt-renew all


entries

ANALOGUE ASSIGN
DATA CORRUPTION

BAD ANL DATA

Analogue output file corrupt-renew


all entries

BAD SPAN SPAN


IGNORED

BAD SPAN

Coarse span adjustment limit


exceeded Check span sample

NOTE
DETECTOR SIGNAL LOW may briefly occur at power-up.

5.4

5.4

Serious Fault (Shutdown) Conditions

If the Fault LED is illuminated and the Analogue Outputs have been driven HIGH (or
LOW, depending on selection), a Serious Fault exists and the Measurement Display
will show "Measurement Invalid".
Immediate action should be taken to rectify the fault, and qualified service personnel
should be called.
The SERIOUS FAULT messages listed in Table 5.3 will result in a Shutdown.
Table 5.3 Serious Fault Displays
Display Message

History Log
Entry

Description/comments

INFRARED SOURCE
FAILURE

SOURCE FAIL

Source failure

CHOPPER MOTOR
OUT OF LOCK

OUT OF LOCK

Motor or timing failure

DETECTOR SIGNAL
HIGH

DET SIG HI

Possible loss of IR filter, loss of


liquid sample.

NOTE

Detector Signal High may also occur and cause temporary shutdown on
liquid analysers if there is no liquid sample in the cell, e.g. at initial power-up.
Check for loss of flow if sensor fitted. Try Manual Zero on good zero sample if
sample flowing.

The user interface remains operational during Shutdown and can be used to
diagnose the likely cause of the fault via the Fault Display and Fault History.

5.5

Cell Maintenance

As a guide the cell should be removed for cleaning when the Reference detector
signals (Section 5.2) fall below 50% of their original values. In aggressive sample
conditions, the loss of signal may be due to general corrosion or pitting of the cell
windows. In this case the cell windows may have to be replaced. Otherwise, the cell
and windows can be dismantled and carefully cleaned and all items (except O-rings)
re-used. See Figure 5.1 for general construction of the cell.

5.5

Figure 5.1

General Cell Construction

CAUTION
Use only Servomex supplied spare parts. The use of inferior replacement
components may degrade the performance and safety of the instrument.

5.6

WARNING

Lethal voltages: mains AC power supplies are potentially lethal. The


maintainer must ensure that all mains power supplies are disconnected
from the supply before maintenance work is started.

Purge the measuring cell and associated pipework with an inert gas
before work is started.

It is the users responsibility to ensure that work is carried out in an


adequately ventilated space taking due account of risks associated
with release of flammable or toxic samples. Attention is also drawn to
the risks associated with the release of asphyxiants.

To remove the cell assembly from the 2500 proceed as follows:


Flush the cell thoroughly with Nitrogen and disconnect the process connections. Where
fitted, remove cell heater connection plug from the socket at the source end and allow
sufficient time for it to cool (this may take some hours). Remove any purge connections
from the end bosses if fitted. Slacken 3 x M6 socket screws securing the cell bosses at
either end of the sample cell.
Carefully disconnect the cell thermocouple from the cell body. Support the cell
assembly and pull the bosses inward from the ends of the 2500 sufficient to enable the
cell to be removed and taken to a workshop for maintenance. If necessary, slacken the
screws holding the source end casting onto the mounting chassis to permit easy
removal.
CAUTION
Do not pull or strain heater connection leads at any stage.
If the scrubbers (if fitted) are not to be removed, leave the bosses on the analyser and
blank the open ends of the bosses using the 2 blanking plugs (02500460), supplied
with the Special Tools Kit S2500979, to preserve the existing scrubbers. Otherwise
take the entire cell and boss assembly to a workshop.
Working on a clean bench in a suitable environment, dismantle the cell. Pull the end
bosses off the cell body. Note that these items are one-piece stainless steel. If the
bosses are fitted with absorbent scrubbers, these should be removed and replaced
(A variety of types may be fitted see spares list for appropriate part numbers).
WARNING
Special materials are used for the windows. Some of these are toxic
(i.e. germanium, barium fluoride, zinc selenide, etc.) these substances must
be handled in accordance with the relevant local standards.

5.7

The windows are usually ground from optical grade single crystal material and as such
may be EXTREMELY EXPENSIVE. The windows may have special anti-reflective
coatings and they may easily be damaged by brittle fracture. The standard material
chosen by Servomex is Calcium Fluoride. This material is non-toxic and reasonably
robust. However, some applications may require materials such as Germanium,
Barium Fluoride, Zinc Selenide, etc. which are toxic and must be handled with extreme
care. See the Manufacturing Data Sheet at the front of this manual for details of the
windows fitted.
Windows should be kept optically clean, i.e. free from dirt, scratches, grime, grease,
fingerprints etc. When handling windows it is recommended that cotton gloves or finger
cots be worn.

WARNING ZINC SELENIDE


If windows of zinc selenide are fitted and have been damaged by acidic attack
(e.g. acidic condensate or organic acids have entered the cell) it will be likely
that some of the zinc selenide will be reduced to selenium metal. This is
highly toxic and on no account should attacked zinc selenide material be
handled with bare fingers. Such material should be bagged and disposed of
according to local regulations.
Cleaning of the cell windows should be carried out using cotton buds or lens tissue,
detergent and/or a suitable solvent such as Isopropanol. Avoid abrasive cleaning
compounds, as these will damage the optical surface.

WARNING
The use of some solvents (iso-propyl alcohol, alcohol etc.) constitutes a
hazard. These substances must be handled in accordance with the relevant
local standards.
To remove the cell windows proceed as follows:
Ensure that the end bosses have been removed from the cell body. Starting at one end
first, unscrew and remove one locking ring with a cell key (supplied in Special Tools Kit
Servomex p/n S2500979). Remove the 5 wavy washers, spacer, and 'O' ring being
careful not to damage them or the window. To remove the window, seal the outlet and
with the cell in a vertical position, connect a hand operated aspirator (Servomex p/n
2387-0514) to the inlet. Apply a gentle pressure to lift the window up squarely and
remove it. Finally remove the PTFE gasket. Dismantle the other end of the cell in a
similar manner.
The 'O' ring spacer is threaded internally. The Special Tools Kit Servomex p/n
S2500979 contains an extractor, which screws into the spacer to aid removal.
The windows may now be carefully cleaned. The cell body may also be cleaned as
necessary.

5.8

Ensure all items are thoroughly clean and dry before reassembly, which is the reverse
procedure but with the following points noted:

The windows should be 'floated' into place squarely using a hand aspirator or
equivalent, reversing the removal procedure.

Use new 'O' rings and other cell sealing kit items making sure that all dust, hairs
etc, are rigorously excluded from them. Ensure the window surfaces are also
free from solvent residues or finger grease.

The locking ring has a coated thread to prevent seizure PTFE tape is not
required. The locking ring should be tightened down in stages, pausing for a
moment between stages until the metal faces of the cell body and locking ring
meet when fully tightened, with no gap between them. Where Teflon 'O' rings
are being used, this will require significant torque, with longer pauses. DO NOT
use the cell inlet/outlet stubs as levers.

After re-assembly, perform a leak check on the cell using a manometer before installing
it in the 2500. Do not submerge the cell or fill it with water.
Re-installation of the complete cell assembly onto the 2500 is the reverse of the
dismantling process with the added requirement that, unless fitted with a purge, the end
bosses should be briefly flushed with clean dry nitrogen after the cell has been
mounted and then new absorbent scrubbers should be quickly screwed into them. Be
sure to fully tighten the screws holding the source end casting onto the mounting
chassis if they have been slackened.
Finally, connect up the cell thermocouple to the cell body, reconnect the cell heater
plug (where fitted) and connect up any purge connections and the process
connections. Return the analyser to service as described in Section 4.5.
Reset zero if necessary on a true zero sample.
5.6

Routine Leak Checks


WARNING

It is the users responsibility to keep all sample connections to the 2500


sample cell, purge connections (if fitted) to the end bosses and steam coil
connections completely free from leaks. All plumbing connections should
therefore be regularly inspected for signs of leakage and urgent corrective
action taken if any are found. This is especially important for process
samples, which are flammable, toxic, liquid, or any combination of these.

5.9

5.10

SECTION 6
6.1

SPARES LISTS

Spares List

The following spares list applies to all versions of the 2500 covered by this manual.
The 2500's Model and Serial Number MUST be advised when ordering spares.
NOTE
For ATEX versions EU1 and EU2 only those boards listed may be used as spares
6.1.1

General
Table 6.1 General Spares List
Description

Part Number

Power Supply

S2000925

Display PCB

02501903

Display/Keypad Assembly

S2501999

Keypad Assembly

02501998

Sensor Interface PCB

02000934

Option PCB

02000916

Transmitter PCB

S2500904A

Motor Connection PCB

S2500906

Power Control PCB

S2500911A

Interconnecting Ribbon Cable Assembly 1-256mm path

S2500928A

Interconnecting Ribbon Cable Assembly 512mm path

S2500928B

Interconnecting Ribbon Cable Assembly 1000mm path

S2500928C

Display Ribbon Cable Assembly

S2500929

Interconnecting Cableform 1-256mm path

S2500944A

Interconnecting Cableform 512mm path

S2500944B

Interconnecting Cableform 1000mm path

S2500944C

Fixings Kit (All fasteners)

S2500974

Seals Kit (All Seals excluding Cell Seals)

S2500975

Chopper Box Insulation

S2500977

Transformer, 115/230 volts

00608333

Transformer, 100/200 volts

00608334
6.1

Table 6.1 General Spares List

6.1.2

Special Tools Kit

S2500979

Enclosure Earth Studs

S2500984

Pressure Transducer

S2500985

Chopper Motor

S2500986

Chopper Box Seals Kit

S2500988

Gas Strut Kit

S2500989

Motor Connection PCB

S2500906

Fuse Kit

02501996

Multi component Adaptor PCB (2550 only)

S2550901

10 x Gas Filter Correlation Adaptor PCB (2510 only)

S2510901A

40 x Gas Filter Correlation Adaptor PCB (2510 only)

S2510901B

Processor PCB + Firmware (2510 only)

S2510965A

Processor PCB + Firmware (2500, 2520, 2550)

S2500965A

Thermal Fuse (98C)

2536-0329

Sample Cell Spares


Table 6.2 Sample Cell Spares

6.2

Description

Part Number

Thermocouple (300mm Sheath)

02500502

Cell Servicing Kit

S2500981

Cell Seals Kit (Viton + PTFE)

S2500982A

Cell Seals Kit (Chemraz)

S2500982B

Special Tools Kit

S2500979

PTFE Tape

1835-3026

Hand Aspirator

2387-0514

6.1.3

Source Units

(Refer to Final Specification Sheet)

Table 6.3 Source Unit Spares

6.1.4

Description

Part Number

Long Wave, CaF2

S2500931C

Long Wave, BaF2

S2500931B

Long Wave, ZnSe

S2500931A

Short Wave, CaF2

S2500943A

Short Wave, Glass

S2500943B

Soft UV CaF2 (2520 Only)

S2520933

Detector Units

(Refer to Final Specification Sheet)

Table 6.4 Detector Spares


Description

Part Number

CaF2

S2501905A

CVDFET

S2501905C

KRS-5

S2501905B

Photodiode (2520 Only)

S2600905

6.3

6.1.5

Windows

(Refer to Final Specification Sheet)

Table 6.5 Window Spares

6.1.6

Description

Part Number

CaF2

S2500980A

BaF2

S2500980B

ZnSe

S2500980C

IR Quartz

S2500980D

Glass

S2500980E

Germanium

S2500980F

Sapphire

S2500980G

Scrubbers

(Refer to Final Specification Sheet)

Table 6.6 Scrubber Spares

6.4

Detector / Cell Boss Scrubber (H2O) Marked 'W'

S1200921

Detector / Cell Boss Scrubber (H2O/CO2) Marked 'CW'

S1200922

Detector / Cell Boss Scrubber (H2O/CO2/CO) Marked 'H'

S1200924

Scrubber Sachet Kit, H2O Mix

1723-8010

Scrubber Sachet Kit, H2O/CO2 Mix

S2000511

Scrubber Sachet Kit, H2O/CO2/CO Mix

S2000512

6.2

Recommended Spares

Quantity for 2 years.

Table 6.7 Recommended Spares


Item

Part Number

1-3 Analysers

4-9 Analysers

Fuse Kit

02501996

Cell Servicing Kit

S2500981

PTFE Tape

1835-3026

Chopper Motor

S2500986

Cell sealing Kit

Refer to application

Hand Aspirator

2387-0514

Windows

Refer to application

Chopper Box Scrubber

Refer to application

Detector & Cell Boss


Scrubber

Refer to application

Source

Refer to application

Detector

Refer to application

Microprocessor PCB

Refer to application

Sensor Interface PCB

02000934

Power Supply

S2000925

NOTE
Scrubbers should be used as soon as received, or otherwise they must be stored in
their original unopened packaging in an airtight container, or preferably in a
desiccator. Scrubbers should be used on a "first in, first out" basis.

6.5

6.6

SECTION 7

INSTRUMENT SPECIFICATIONS

7.1

Generic 2500 Series Performance

7.1.1

Environmental Specifications
Table 7.1 General Environmental Specifications
Operating Temperature

GEN, EU1, EU2, FM2, HTV 0-55C (32-131F)

Operating Humidity

0-95% RH, non-condensing

Storage Temperature

-40C(-104F) to +70C (158F)

Storage Humidity

0-95% RH, non-condensing

Altitude

Up to 3000 M

Installation Category II
Pollution Degree II

Over-voltage impulse to withstand up to 2500V


in accordance with IEC 60664-1

Rate of Ambient Temperature


Change

Less than 25C/h, (45F/h) Change

Warm-up Time

Typically 2-10h, depending on application and


environment

7.1.2

Dimensions

Length:

Max 1618mm, (63.7") min 615mm (24.2")

Height:

241mm (9.5")(allow 500 mm (20")


to open end covers)

Width:

248mm (9.8")

Weight:

from 25 kg (55 lb)

7.1.3

Power Supply

115/230V AC 15%, 50/60Hz or 100/200V AC 15%, 50/60Hz.


120 VA without optional electrically heated cell.
300 VA with optional electrically heated cell.
Power Supply
Voltage Effects:

Less than 1% fsd for 15% change.

Power Supply
Frequency Effects:

Less than 1% fsd for 47 to 53 Hz or 57 to 63 Hz.

7.1

7.1.4

Performance Characteristics (each component)


Table 7.2 Performance Characteristics
Intrinsic Error*

Less than 1% fsd

Repeatability

0.5% fsd

Output Fluctuation (Noise)

Less than 1% fsd (peak-peak) at minimum T90

Linearity Error

Typically less than 1% fsd

Short Term Zero Drift:

Less than 1% fsd per week

Response Time (T90)

User adjustable from 11s minimum (electronic


only, excludes sampling)

Obscuration

Less than 3% fsd zero error for 50% broadband


obscuration of cell windows

Influence Errors

Dependant on application (see specification


details)

* error when used under reference conditions.

2510 HCL measurement less than 2% fsd (peak-peak) at minimum T90.

7.1.5

Performance Characteristics EMC

Configuration of equipment:
Fitted with an unscreened power cable. Also included were two separate mA
output cables, each individually screened. Each signal pair was fitted with
ferrite bead (Servomex part No. 2824-0017). The screens and drain wires were
terminated at the glands.
Results:
Met the requirements of EN 61326, Table 4, Class B (Equipment for use in
domestic establishments) in respect of conducted and radiated emissions.
Met the requirements of EN 61326, Table A1 (Equipment for use in industrial
establishments) in respect of electrostatic discharges, fast transients/bursts,
surges, conducted RF disturbances, voltage dips and interrupts.
An additional error of 3% FS over the intrinsic error given in Table 7.2 may be
observed at some frequencies under the influence of radiated RF fields
specified for industrial environments.
Connection of additional or other signal/control cables and pressure
compensation could lead to a deterioration in performance at some RF
frequencies.

7.2

7.1.6

Sample Specification
Table 7.3 Sampling Specification
Sample Temperature

0-180C/32-356F (see sampling influences)

Sample Pressure

0-150 psig/0-10 barg (normal) (For special high


pressure operation consult Servomex)

Sample Flow (typical)

GAS 0.2-5.0 litres/min


LIQUIDS 0.3-1.0 litres/min
(Constant flow rate preferred)

Sample Wetted Materials

Refer to Final Specification Sheet

7.1.7

Sample and Ambient Performance


Table 7.4 Sample and Ambient Influence

Sample Pressure
Stability

Output may change up to 1.6% fsd for a 1% change in


sample (Gas Analysers) pressure. With optional
empirical sample pressure compensation, effect is
reduced to 1% fsd or less for a 20% change in sample
pressure for most typical samples.

Sample Temperature
Stability

Output may change up to 0.3% fsd for gases for a 1C


change in sample temperature. With optional empirical
sample temperature compensation, effect is reduced to
1% fsd or less for a 20C change in sample
temperature for most typical samples.

Sample Flow Stability

Depends on cell volume and vent conditions but for


unheated cells the primary effect is less than 1% fsd
change in output for flow rate changes of 10%.

Ambient Temperature
Influence

Less than 1% fsd zero drift due to rate of ambient


temperature change of 25C/hr (45F/hr) over a
maximum of 25C (45F) change.

7.1.8

Solenoid Valve Relays

3 x Solenoid valve relays with volt free contacts rated 240VAC/1.0A, 30VDC/1.0A.
1 x 24VDC/12VA auxiliary supply for solenoid valves

7.3

7.1.9

Analogue Outputs

2 x mA outputs, plus 3 x relay outputs. Optionally additional 2 x mA, plus 2 x relay


outputs.
All analogue mA outputs are assignable and selected as 0-20mA or 4-20mA (max.
impedance 1KOhm) or alternatively 0-10V or 2-10V (min. impedance 1MOhm).
Dual ranges on every Analogue output. Optional additional mA outputs should be
specified for 3 component analysers.
All relay outputs are assignable contact pairs selected by links as NC or NO, and are
rated 240VAC/1.0A, 30VDC/1.0A.
7.1.10

Serial Output

Single ASCII data logging output port (RS232).


Table 7.5 Serial Output Specification

7.1.11

Baud Rate

2400

Parity

Even

Stop Bits

Data Bits

Handshake

Hardware

Inputs

Terminals for external switch contacts are provided for:

Range change

Autocalibration/Autocheck initiate

Key switch (password)

Flow Sensor

7.1.12

Ingress Protection

Designed to IP65 (without electrical cell heater, IP50 with electrical cell heater).

7.1.13

Optical Purge

Where used, end bosses require total 200cm3/min clean, dry purge gas (see Section 4.2)
delivered at 1-3 psig.

7.4

7.1.14

485 Purge Controller (2500 EU1 Only)

Requires maximum of 40 l/min clean dry Nitrogen or clean dry Air at 3-7 barg,
depending on application.

7.1.15

Alternative Purge Controller (2500 EU1 Only)

Minimum flow rate = 10 SLPM for 11 minutes.


Minimum overpressure = 0.5 mBar.
Maximum overpressure = 80 mBar.
Refer to Certification Manual (Certificate SIRA 02ATEX1395X) for definitive
requirements and operational parameters.

7.5

7.6

SECTION 8 CE MARKING AND OTHER SAFETY APPROVALS


All 2500, 2510 and 2550 Analysers are CE marked and comply with all appropriate
European Community Directives.
8.1

EMC Directive

Electromagnetic Compatibility Directive 89/336/EEC (as amended by 92/31/EEC and


93/68/EEC) by the application of the following standard:
EN 61326: 1997 +A1: 1998 +A2: 2001 Calling:
Emissions; Table 4, Class B. Equipment used in establishments directly
connected to a low voltage supply which supplies buildings for domestic
purposes;
Immunity: Table A1 Industrial Environments.

8.2

Low Voltage Directive


NOTE

Assessment under the Low Voltage Directive is not required for CE marking if the
equipment is approved under the ATEX Directive for use in potentially explosive
atmospheres
Low Voltage Directive 73/23/EEC and the CE Marking Directive 93/68/EEC by the
application of the following standards:
EN 61010-1 (equivalent to IEC 61010-1)
Safety requirements for electrical equipment for measurement, control and
laboratory use general requirements:
2500, 2510 and 2550 Analysers are rated in accordance with:
IEC 60664 for Installation Category II, which is characterised as being local level (i.e.
not distribution level), appliances and portable equipment with over-voltage impulse
withstand up to 2500V.
8.3

Pressure Directive

All the relevant design aspects of the 2500 series analysers have been assessed in
respect of the requirements of the Pressure Equipment Directive (97/23/EC). These
products are excluded from the scope of this directive when installed and operated
within their published specifications.

8.1

8.4

ATEX Directive and other non-European Hazardous Area approvals

ATEX Directive 94/9/EEC


NOTE
Depending on the variant ordered an additional Safety Certificate Manual may be
supplied. This contains all the relevant safety certificates and declarations for the
EU1, EU2 and FM2 variants.

WARNING
The installer and/or the user of the analyser must be satisfied that the
installation complies with any "Special Conditions for Safe Use" or
"Schedules of Limitations" contained in the Certificates included in the
separate certification manual.
Table 8.1 02500 Safety Approvals
Variant
GEN/HTV

EU1

EU2

Certification
Only suitable for installation in a non-hazardous area.
The 2500, 2510 and 2550 EU1 version is ATEX CAT 2
equipment certified to Type 'p' standard EN 50016 and is
suitable for use in European and UK Zone 1 areas. The
analyser is marked EEx p ia [ia] IIC.
The EU2 version is ATEX CAT 3 equipment certified to
Type 'n' standard EN 50021, and is suitable for use in
Zone 2 areas. The analyser is marked
EEx nCL IIC T3 for cell temperatures up to 130C and
EEx nCL IIC T4 for cell temperatures up to 80C.
A special condition of this certification is that the analyser
must be protected against the entry of solid particles or
liquids by a suitable cover or enclosure where necessary.
The 02500 Series FM2 version is approved for operation
in US Class 1 Div. II location, without safety purging.

FM2

8.2

02500 Series FM2 units fitted with electrically heated


sample cells must be installed in locations which provide
adequate protection against the entry of solid foreign
bodies and liquids capable of impairing safety.

Servomex 2500 Series


Process Gas Analyser
Service Manual
This Service Manual applies to the following
Analysers: 2500, 2510, 02520 and 2550
Ref : 02500 / 002C / 1
Order as part No. 02500 002C

WARNINGS, CAUTIONS AND NOTES


This publication includes WARNINGS, CAUTIONS AND NOTES which provide,
where appropriate, information relating to the following:
WARNINGS

Hazards which will result in personal injury or death.

CAUTIONS

Hazards which will result in equipment or property damage.

NOTES :

Alert the user to pertinent facts and conditions.

WARNING
1.

LETHAL VOLTAGES: THE ELECTRICAL POWER USED IN THIS EQUIPMENT IS AT A


VOLTAGE HIGH ENOUGH TO ENDANGER LIFE.

2.

BEFORE CARRYING OUT SERVICING OR REPAIR THE EQUIPMENT MUST BE


DISCONNECTED FROM THE ELECTRICAL SUPPLY. TESTS MUST BE MADE TO
ENSURE THAT DISCONNECTION IS COMPLETE.

3.

IF FOR ANY REASON THE POWER SUPPLY CANNOT BE DISCONNECTED,


FUNCTIONAL TESTING, MAINTENANCE AND REPAIR OF THE ELECTRICAL UNITS
IS ONLY TO BE UNDERTAKEN AS A LAST RESORT AND MUST BE CARRIED OUT
BY PERSONS FULLY AWARE OF THE DANGER INVOLVED.

4.

THIS MANUAL COVERS SERVICING OF ANALYSERS WHICH HAVE CERTIFICATION


FOR USE IN HAZARDOUS AREAS. THE ENGINEER MUST ENSURE THAT ALL
INSTRUCTIONS REFERRING TO SERVICING OF COMPONENTS AFFECTING
CERTIFICATION ARE RIGOROUSLY ADHERED TO.

NOTE
This service manual is for use with the 2500 Single Beam Infrared, 2510 Gas Filter Infrared, 2520
Soft UV Analyser and 2550 Multicomponent Infrared analysers.
All analysers in the 2500 Series are built using, wherever possible, standard components. The main
exterior differences: The 2510 analyser has additional sample cell insulation and a larger metal cell
cover. On the inside: The 2510 and 2550 analysers utilise an additional printed circuit board
mounted on the Transmitter PCB to adjust the signal gain. The 2510 detector housing is plated steel
whereas others are aluminium. 2520 analysers utilise different source and detector assemblies
from the other analysers.
Refer to Section 9 - Parts Lists, for a detailed breakdown of the analyser components.
This manual covers disassembly procedures for the equipment, and brief technical descriptions of
component parts of the equipment. It should be thoroughly read and retained by the Service
Engineer.
I

NOTES

II

CONTENTS

Section
1.

Introduction

2.

Description

3.

Fault Diagnosis

4.

Cell Maintenance

5.

Disassembly - Major Components

6.

Chopper Box

7.

Reassembly - Leak Testing and Flushing

8.

Printed Circuit Boards

9.

Parts Lists

10.

Engineering Drawings

III

NOTES

IV

SECTION 1: INTRODUCTION
LIST OF CONTENTS
Section

Page

1.1

INTRODUCTION...........................................................................................................................1.3

1.2

HAZARDOUS AREAS..................................................................................................................1.4

1.3

USING THIS MANUAL .................................................................................................................1.4

1.4

ROUTINE LEAK CHECKS...........................................................................................................1.4

1. 1

NOTES

1. 2

1.1

INTRODUCTION
This 2500C Series Service Manual is intended for use by Servomex trained service personnel.
The manual contains technical descriptions, disassembly and assembly instructions, fault
diagnosis charts, repair procedures, tools and test equipment lists as well as necessary drawings
and illustrations.
It is recommended that during servicing, service personnel have access to the relevant 2500C
Infrared Analyser User Manual, and in addition they should be familiar with general operating
principles of the equipment, display messages and the use of the keypad.
Before undertaking service or repair of the analyser it is recommended that reference is made to
the appropriate section of this manual for details of procedures that may be relevant. The analyser
is a sophisticated device containing sensitive electronic and optical components and extreme care
should be taken when handling these components. Contaminants such as dirt/fingerprints left
on optical assemblies will cause degraded performance.
Printed circuit boards (PCB=s) containing CMOS devices can be damaged by static electricity even
when the power is off. For this reason avoid touching any circuitry on the PCB=s. To discharge any
static that you may hold, touch an unpainted part of analyser (chassis earth point).
Before beginning work, confirm with the customer the location of the analyser and the equipment's
configuration (precise analytical set-up for a stated analysis in a specific background stream).
The versions covered by this manual are:
02500C/ 02510C/ 02520C/ 02550C GEN
02500C/ 02510C/ 02520C/ 02550C EU2
02500C/ 02510C/ 02520C/ 02550C FM2
02500C/ 02510C/ 02550C EU1

General Purpose/Safe Area.


Certified for UK/European Zone 2 areas.
Certified for US Class 1 Div. II areas.
Certified for UK/European Zone 1 areas.

The following abbreviations are used throughout this manual:


PCB
fsd
hex
hex.skt
wrt
DVM
htr
VA
VD
VP
VI(A/B)

:
:
:
:
:
:
:
:
:
:
:

Printed Circuit Board


Full Scale Deflection (full span)
Hexagonal (e.g. hexagonal Allen key)
Hexagonal Socket Head
With Respect To/compared to
Digital Volt Meter (suitable for mains AC and lower voltages)
Heater
Voltage Supply (analogue circuits)
Voltage Supply (digital circuits)
Voltage Supply (power circuits)
Voltage Supplies (isolated)

Where items are referenced with a part No. this No. is a Servomex part No. unless otherwise
specified.

1. 3

1.2

HAZARDOUS AREAS
WARNING
BEFORE STARTING ANY WORK WHATSOEVER, ASCERTAIN WHERE THE
EQUIPMENT IS INSTALLED. IF INSTALLED IN A HAZARDOUS AREA
OBTAIN THE NECESSARY PROTECTIVE CLOTHING REQUIRED SUITABLE
FOR THAT PARTICULAR AREA AND NOTE ANY RESTRICTIONS ON USE
OF TOOLS OR EQUIPMENT.

NOTE: It is possible that the EU2 and FM2 versions are being used when flammable samples are
to be monitored in otherwise safe areas.
1.3

USING THIS MANUAL


The layout of this manual is in a sectional (modular) format but with intersectional referencing
where applicable.
Key sections are:

Section 3: Fault Diagnosis


Section 4: Cell Maintenance
Section 5: Disassembly
Section 6: Chopper Box

However, non-Servomex service engineers should read "Section: 2 Description" first to become
familiar with 2500 series construction.
Once problems have been diagnosed, the service engineer should move on to the relevant
sections. Any spares that may be required are listed in Section 9.
1.4

ROUTINE LEAK CHECKS


It is the user's responsibility to keep all sample connections to the analyser's sample cell,
purge connections (if fitted) to the end bosses and steam coil connections completely free
from leaks. All plumbing connections should therefore be regularly inspected for signs of
leakage and urgent corrective action taken if any are found. This is especially important for
process samples which are flammable, toxic, liquid, or any combination of these.
It is recommended that the following parts be leak checked annually and replaced if
necessary:End boss scrubbers,
Purge port seals,
Chopper box blanking plug seal,
(Refer to section 9 for component part Nos.)

1. 4

SECTION 2: DESCRIPTION
LIST OF CONTENTS
Section

Page

2.1

GENERAL .....................................................................................................................................2.5

2.2

DESCRIPTION ..............................................................................................................................2.5
2.2.1
2.2.2
2.2.3
2.2.4
2.2.5
2.2.6
2.2.7
2.2.8
2.2.9
2.2.10

Introduction ......................................................................................................................2.5
Source-End Assembly.....................................................................................................2.5
Power Control PCB..........................................................................................................2.6
Source Assembly.............................................................................................................2.6
Sample Cell Assemblies..................................................................................................2.6
Detector-End Assembly...................................................................................................2.7
Mains Transformer...........................................................................................................2.7
Transmitter PCB ..............................................................................................................2.7
Detector Box Assembly ...................................................................................................2.8
Keypad and Display PCB Assembly...............................................................................2.8

LIST OF ILLUSTRATIONS
Figure
2.1
2.2

Page
Overall View..................................................................................................................................2.3
Keypad and Display ....................................................................................................................2.4

2. 1

NOTES

2. 2

Figure 2.1 Overall View


2. 3

16 Character
2 line VF Display
Hard key to select
Measurement Display
Hard key to select
Menu Display
Fault warning LED

Alarm warning LED

Up/Down/Left/Right
Cursor keys to select menu
options and increase/decrease
numerical values

Key to enter data or accept highlighted option within Menu

Key to escape or return to previous level within Menu

Key will also return display to Top Level Menu from lower menu levels

Fig 2.2 Keypad & Display


2. 4

2.1

GENERAL
This section provides information on what the analyser contains and how the analyser is
assembled. It would be helpful for the service engineer to read completely through this section
before continuing.
The 2500 series process analysers are single beam, dual or multi-wavelength infrared process
analysers, suitable for monitoring single components, multi-components or a component group in a
gas or liquid sample.
Samples to be analysed are normally passed continuously through the analyser's sample cell for
24 hours a day and the analyser is not normally switched off.
General layout of the equipment is shown in Fig 2.1. The analyser is microprocessor controlled
which provides flexibility in configuring the analyser to suit various process environments.
Operation is via a control panel mounted on the analyser.

2.2

DESCRIPTION

2.2.1

Introduction
The analyser (Fig.2.1) comprises two end assemblies, source and detector, connected by a
mounting chassis. The sample cell is mounted between the two end assemblies.
The source-end assembly houses the source (which generates the broad beam of energy across
the complete spectrum), Power Control PCB and the chopper box assembly.
The sample cell is secured between the two end assemblies and can be removed for servicing.
The detector-end assembly houses the detector box assembly, main electronic printed circuit
boards (PCB) and keypad/display panel assembly.

2.2.2

Source - End Assembly


The source-end assembly (02501/940) houses the chopper box assembly, source assembly and
Power Control PCB. In addition the source-end assembly forms the mount for the cell
thermocouple connector, pressure transducer connector, external cable earth and the sample cell.
The left hand side panel contains entry glands for electrical cables.

2. 5

2.2.3

Power Control PCB


The Power Control PCB (02500/911A) is mounted horizontally. At one end it locates in 2 brackets
attached to the rear wall and is secured to the floor of the unit via spacers using 3 M4 x 16 hex.skt
captive screws. The board provides the mount for 2 ribbon cable plugs PL1, PL2, and 2 cable
connectors PL3 and PL4. PL1 and PL2 provide ribbon cable connections to the Transmitter PCB
and detector box; PL3 and PL4 allow cable connections to the mains transformer and source
assembly.

2.2.4

Source Assembly
Three types of source assembly can be fitted; a sealed Long Wave, Short Wave or Soft UV type.
They are secured to the chopper box lid with 3 hex.skt M3 x 12 screws.
The sealed long wave source assembly is conical in shape with a tapered end, having a flanged
rim containing fixing holes for connection to the chopper box lid. The other end provides electrical
connections to Power Control PCB PL4 via a 4-way cable and plug. Three different types of
lenses can be fitted dependent on requirements. Component parts are sealed with adhesive
potting during manufacture.
The short wave source assembly ( 02500/943) consists of an M29 Quartz Tungsten Halogen lamp
rated 6V, 10W, but under-run on the 2500C at around 4.6V. It is mounted on a PCB via a terminal
block. The PCB / lamp is fitted into a machined block which is mounted to the chopper box. Early
versions of the PCB / lamp assembly mounted to the chopper box using a bracket. There are 2
types of SW source arrangement, one with a CaF2 lens, one with a crown glass lens, but the actual
source bulb is the same in both cases.
The Soft UV Sealed Source assembly consists of a Gas Filled Tungsten Lamp, fitted into a special
housing. The lamp is normally run at 4.4V. Component parts are sealed with adhesive potting
during manufacture.

2.2.5

Sample Cell Assemblies


The cell assemblies have a range of different path lengths. Servomex Drawing No. 02500/834
shows path lengths of 1 to 64mm and Servomex Drawing No.02500/835 shows path lengths of
128 to 1000mm. Sample cells are of thick stainless steel, monel, hastelloy or titanium construction
fitted with 6mm thick windows. Eight different types of windows can be fitted depending on
customer requirements. Window types are:
CaF2 (Calcium Fluoride)
BaF2 (Barium Fluoride)
ZnSe (Zinc Selenide)
Infrasil Quartz
Crown Glass
Germanium
UV Quartz
Sapphire
A thermocouple connector and 2 sample fluid process connections are fitted to each cell.

2. 6

Cells are mounted between the source and detector ends using flanged cylindrical cell bosses and
secured in position using 3 M6 x 12 hex.skt screws. These bosses may be fitted with screw-in
chemical scrubbers (absorbers) or purge fittings.
Heated versions of these cells may be fitted - Electrically Heated and Steam Heated.
2.2.6

Detector - End Assembly


The detector-end assembly (02501/941) contains a mains transformer, Transmitter PCB,
Microprocessor PCB, Sensor Head Interface Board (SIB PCB), plug-in Power Supply, Optional
PCB=s (if fitted), Display PCB, keypad and detector box assembly.
Externally, the detector-end rear panel forms the mount for the sample cell and the mounting
chassis. Internally, the rear panel supports the detector box assembly and 2 brackets which
provides support for the Transmitter PCB. Right hand side panel contains entry glands which
allow electrical cables to be inserted for connections that may be required. When not in use they
are plugged.

2.2.7

Mains Transformer
The mains transformer is mounted in the base of the unit, enclosed within a metal frame, beneath
the Transmitter PCB. The transformer is secured in position by 2 bolts that pass through the base
of the unit.

2.2.8

Transmitter PCB
The Transmitter PCB (02500/904A) is mounted horizontally above the mains transformer and at
one end locates in 2 brackets mounted on the rear panel and is secured to the floor of the unit via
spacers using 2 M4 x 16 hex.skt screws. The PCB provides the mount for three 96-way
connectors SK1, SK3, SK5 and two 48-way connectors SK2, SK4; the Microprocessor, SIB, Option
and Power Supply PCB=s respectively are vertically located. The PCB=s are enclosed by a three
sided metal cage; the Power Supply PCB provides the fourth side, and is thermally connected to
the detector end enclosure. The PSU is fixed to the heatsink using 2 M4 x 12 screws. The cage
sides have PCB guide runners mounted which provides alignment as the PCB=s are being fitted. A
securable hinged lid completes the cage. There are also three ribbon plugs PL1, 2 and 3 which
provide ribbon cable connections to Power Control PCB, Display PCB and detector box.

2. 7

2.2.9

Detector Box Assembly


The detector box assembly is mounted on the inside rear panel of the detector-end. The detector
box comprises a detector block, Detector PCB assembly, 02501/905 for 2500, 2510 and 2550
analysers or 02600/905 for 2520 analyser) scrubber assembly (if fitted), adaptor and dowel pin (if
fitted), detector boss and lens assembly. The detector block, cylindrical in shape, locates in the
aperture provided in the rear panel of the box and is secured using 3 hex.skt M4 x 10 screws. The
Detector PCB assembly is mounted on the face of the detector block, positioned over an `O' ring
and secured with 3 hex.skt M3 x 6 screws. A 16-way ribbon cable provides interconnections to the
Transmitter PCB. An earth protection lead is routed from the PCB to the RFI earth terminal in the
detector end assembly. The scrubber assembly (if fitted), with bonded seal, passes through an
aperture in the `top' of the box and screwed into the detector block. A lid is fitted to the box and
secured in place with 4 hex.skt M4 screws.
The 2510 detector assembly is similar in all above respects except that the detector box is
manufactured from plated steel and the lid is secured by two M3 screws mounted on the top edge.
Note: The scrubber and leak test port are not fitted to purge certified analysers, these are replaced
by a purge adaptor and flexible purge duct fitted between the detector box and the detector end
enclosure.

2.2.10 Keypad and Display Assembly


The Keypad and Display Assembly is mounted in the hinged lid and enclosed with a vented metal
cover. The assembly comprises of a mounting plate fitted with a transparent window, keypad and
Display PCB Assembly.
The window is sealed to the mounting plate with an >O= ring and retained by means of a bracket
and six M4 x 6mm hex skt screws. The Display PCB Assembly is mounted on the plate using four
M4 x 12mm hex skt screws.
The display, which is mounted on the Display PCB Assembly, is a 2-line 16 character vacuum
flourescent display which ensures good visibility under all conditions.
The keypad (see Figure 2.2) comprises of eight single function >hard= keys, four of which are
identified as Measure, Menu, Exit and Enter, and four cursor direction keys. Operation of the keys
requires only a light pressure and they give tactile feedback when pressed.
The keypad is bonded to the mounting plate and connected to the Display PCB Assembly via a 10way connector. The connecting ribbon is encapsulated as it passes through the mounting plate.
Interconnections between this assembly and the Transmitter board are via a 50-way ribbon cable

2. 8

SECTION 3: FAULT DIAGNOSIS


LIST OF CONTENTS
Section

Page

3.1

INTRODUCTION...........................................................................................................................3.4

3.2

GENERAL FAULTS......................................................................................................................3.4

3.3

SERIOUS FAULTS .......................................................................................................................3.8

3.4

DIAGNOSTICS..............................................................................................................................3.8

LIST OF ILLUSTRATIONS
Figure

Page

3.1

2500, 2520 and 2550 Menu Map..................................................................................................3.2

3.2

2510 Menu Map.............................................................................................................................3.3

3. 1

Fig 3.1 2500, 2520 and 2550 Menu Map


3. 2

Fig 3.2 2510 Menu Map


3. 3

3.1

INTRODUCTION
This section is basically an expanded version of sections found in the 2500 User Manual. The
following charts should be used by the service engineer to check the more easily remedied faults
by undertaking the actions suggested. The service engineer should initially decide the nature of the
fault and then use the charts to arrive at the correct diagnostic solution. Additional information can
be obtained with reference to the relevant section of this manual.
Using the charts
The first column, Fault (Display Message), indicates the type of fault and how it is displayed. The
second column provides a diagnosis of the fault and column three sets out remedial actions
including reference to other sections of this manual and User Manual from where additional
information can be obtained.

3.2

GENERAL FAULTS
General faults category covers parameters out of tolerance with the analyser remaining operational
and measuring.
WARNING
TB32/1 SUPPLIES 110V AC NOMINAL
SEVERE SHOCK HAZARD

3. 4

General Faults (cont)


Fault
(Display Message)

Diagnosis

Action

Source voltage high

Source voltage high

1 Check fuses F30 and F32.


2 Check switch SW3 and SW4
settings. Refer to Section 8.1.10.1 and
8.10
3 Change Power Control PCB.

Source voltage low

Source voltage low

1 Check fuses F30 and F32.


2 Check switch SW3 and SW4
settings. Refer to Section 8.1.10.1 and
8.10
3 Change Power Control PCB.

Sample flow failure

User sample flow alarm


active

Check short circuit between TB36


pins 1 and 2. Refer to Section
8.1.10.2 and 8.10

Chopper temp high

Ambient too high or other


fault

1 Check fuse F32


2 Check PL2/PL3 for positive
connection
3 Refer to Section 8.1.10.3
4 Change Power Control PCB.

Chopper temp low

Ambient temp too low or


other fault

1 Check fuse F32


2 Check PL2/PL3 for positive
connection
3 Refer to Section 8.1.10.3
4 Change Power Control PCB.

Chopper temp
sensor failure

Failed, disconnected, or
out of range

1 Check sensor connection


2 Check sensor or change if necessary

Cell temp high

Cell over temperature

1 Check fuse F31


2 Check cell heater connections at
TB32 and TB33
3 Refer to Section 8.1.10.4
4 Change Power Control PCB

Cell temp low

Cell under temperature

1 Check fuse F31


2 Check cell heater connections at
TB32 and TB33
3 Refer to Section 8.1.10.4
4 Change Power Control PCB

Cell temp sensor


failure

Failed, disconnected or
out of range

1 Check sensor connection


2 Check sensor; change if necessary

3. 5

General Faults (cont)


Fault
(Display Message)

Diagnosis

Action

Sample temp
sensor failure

Failed, disconnected or
out of range

1 Check sensor connection


2 Check sensor; change if necessary

Pressure sensor
failure

Failed, disconnected or
out of range

1 Check sensor connection


2 Check sensor; change if necessary

Bad reference
voltage

This is a check on the


4mA level in the 4 - 20
mA current loop from
the Transmitter PCB to
the SIB PCB

1 Change Power Supply PCB


2 Check wiring of the DIN Controller
(if used)
3 Change Transmitter PCB
4 Change SIB PC

Detector signal low

Likely cell contamination


or severe source
degradation

1 Check signal at TP4 wrt TP2 on the


Transmitter PCB. Refer to Fig. 8.6(b).
Refer to section 8.1.10.5
2 Clean cell; refer to Section 4.1

Detector temp
sensor failure

Failed, disconnected or
out of range

1 Check sensor connections


2 Check sensor, or change if necessary

Source temp high

Possible Power Control


PCB failure or strong
reference interference

1 Check source voltage


2 Check SW3 and SW4 settings refer to
Section 8.1.10.5 and 8.10
3 Change Power Control PCB

Source temp low

Possible Power Control


PCB failure or strong
reference interference
or severe source
degradation

1 Check source voltage


2 Check SW3 and SW4 settings refer to
Section 8.1.10.5 and 8.10
3 Change Power Control PCB

Zero out of tolerance

Failed Autocal,
Bad zero sample

Use One Cycle or Manual Zero;


refer to 2500C User Manual

Span out of tolerance

Failed Autocal,
Bad Span sample

Use One Cycle or Manual Span;


refer to 2500C User Manual

Autocal bad preflush

Failed Autocal,
bad preflush

1 Use One Cycle


2 Check solenoids and span gas
3 Refer to User Manual

3 consecutive password
failures

WARNING Unauthorised personnel


attempting access

Microprocessor PCB
clock IC faulty

1 Change Microprocessor PCB


2 Refer to Section 8.5.3

Password violation
System clock
inoperative

3. 6

General Faults (cont)


Fault
(Display Message)

Diagnosis

Action

System AD
inoperative

Possible SIB PCB fault

Change SIB PCB

Default system
data corruption

Factory calibration
corrupt

Use ASSOCIATE function to


reconstruct default data. Refer to
Section 8.2.3 and User Manual

Calibrate data
corruption

User calibration corrupt

1 Re-calibrate; refer to User Manual


2 If corruption repeated after entry
change Microprocessor PCB

Utility data corruption

Utility file corrupt

1 Renew all entries; refer to user


manual
2 If corruption repeated after entry
change Microprocessor PCB

Alarm/relay data
corruption

Alarm relays corrupt

1 Renew all entries; refer to user


manual
2 If corruption repeated after entry
change Microprocessor PCB

Analogue assign data


corruption

Analogue output file


corrupt

1 Renew all entries; refer to user


manual
2 If corruption repeated after entry
change Microprocessor PCB

3. 7

3.3

SERIOUS FAULTS (SHUTDOWN)


If any of the following messages are displayed it indicates serious degraded performance.
Measurement is withdrawn and analogue output driven high or low depending on configuration.
The service engineer should initially decide the nature of the fault.
Fault
(Display Message)

Diagnosis

Action

Infrared Source
Failure

Source driver failure (low)

Refer to sections 8.1.9, 8.1.10.1, 8.1.10.6


& 8.10

Chopper Motor Out of


Lock

Failure of motor, timing or


control circuit

Refer to sections 8.1.9 & 8.1.10.7

Detector Signal High

Loss of IR filter from


chopper wheel or loss of
liquid sample or high
absorbance gas
background from cell

Refer to section 8.1.10.8

3.4

DIAGNOSTICS
The diagnostic reference chart below can be interrogated via the SETUP/DISPLAY/DIAGNOSTICS function of the 2500 menu map. Refer to Fig.3.1 of this section and
Section 4.6 of the User Manual. The chart provides the service engineer with information
regarding how and where the measurements are taken.
Using the chart
The first column shows display messages, the second column describes function and the third
column provides the location of measurement being made and relevant comments.

3. 8

Diagnostics (cont)
Display

Description

Operation

Chopper
Temp C

Chopper box
temperature

Normal working temperature


65EC approx. Measured by the AD590
transducer device mounted on the
Motor Connection PCB.

Detector Temp C

Detector temperature, as
a compensating
temperature

Normal reading approx 10EC above


external local ambient.
Measured by AD590 on the Detector
PCB. Used for ambient compensation.

Cell Temp C

Cell temperature

Measured by type K thermocouple in


cell thermowell. Connected to the Power
Control PCB.

Sample Temp C

Sample temperature
(if option fitted)

Measured by type K thermocouple via


fitting on cell port. Connected to the
Power Control PCB.

Ambient
Pressure (units)

Ambient pressure
(if option fitted)

Measured by transducer mounted on


Power Control PCB. Used for sample
barometric pressure compensation.

Sample
Pressure

Sample pressure
(if option fitted)

Measured by transducer via fitment on


cell port and connected to Power
Control PCB.

Source Volts V

Source voltage

Factory set on Power Control PCB


according to source type. Should not be
adjusted. Refer to Section 8.1.10.1.

Motor Drive %

Motor drive power.

Motor drive will vary around a


percentage level under normal
conditions.

Set Point %

Set point

Set point is a factory set benchmark.

Detector Sig M

Detector signal level


(M = measurement)

Signal measurements in absolute units.


Will decrease with increasing
measurement component.

Detector Sig
R1 or R

Detector signal
(R1 = reference 1)

R1 signal intensity in absolute units.


Should remain roughly constant.

3. 9

NOTES

3. 10

SECTION 4: CELL MAINTENANCE


LIST OF CONTENTS
Section

Page

4.1

CLEANING THE SAMPLE CELL ................................................................................................4.4

4.2

CELL PRESSURE AND TEMPERATURE TRANSDUCER FAULTS.......................................4.8

4.3

ELECTRICALLY HEATED CELL FAULTS ................................................................................4.8

LIST OF FIGURES
Figure
4.1

Page
General Cell Construction.............................................................................................................4.3

4. 1

NOTES

4. 2

Fig.4.1 General Cell Construction


4. 3

4.1

CLEANING THE SAMPLE CELL


This section is basically as that found in the 2500 User Manual but with Service Manual references
incorporated.
The DETECTOR SIGNAL LOW fault is a primary indication of a badly contaminated sample cell.
However, before proceeding to dismantle and clean the cell, it should be confirmed that this is the
case, by looking at the detector signals levels in Section 3.4, Diagnostics. Typical reference
detector signals under normal working conditions can be anywhere in the range 500,000 to
900,000 depending on application.
As a guide the cell should be removed for cleaning when the reference detector signals fall below
50% of their original values. In aggressive sample conditions, the loss of signal may be due to
general corrosion or pitting of the cell windows. In this case the cell windows may have to be
replaced. Otherwise, the cell and windows can be dismantled and carefully cleaned and all items
(except O rings) re-used. See Fig 4.1 for general construction of the cell.
WARNING
LETHAL VOLTAGES: MAINS AC POWER SUPPLIES ARE LETHAL.
THE SERVICE ENGINEER MUST ENSURE THAT ALL MAINS POWER SUPPLIES
ARE DISCONNECTED FROM THE SUPPLY BEFORE WORK IS STARTED.
PURGE THE MEASURING CELL AND ASSOCIATED PIPEWORK WITH AN INERT
GAS BEFORE WORK IS STARTED.
IT IS THE SERVICE ENGINEERS RESPONSIBILITY TO ENSURE THAT WORK IS
CARRIED OUT IN AN ADEQUATELY VENTILATED SPACE - TAKING DUE
ACCOUNT OF RISKS ASSOCIATED WITH RELEASE OF FLAMMABLE OR TOXIC
SAMPLES. ATTENTION IS ALSO DRAWN TO THE RISKS ASSOCIATED WITH THE
RELEASE OF ASPHYXIANTS.
BURN HAZARD: SOME 2500 SERIES CELLS MAY BE HEATED TO 180EC AND
WILL TAKE SEVERAL HOURS TO COOL AFTER BEING SWITCHED OFF. IN
PARTICULAR, THE CELL INLET/OUTLET CONNECTIONS AND THE CELL END
BOSSES SHOULD NOT BE TOUCHED WHILE HOT.

CAUTION
The Service Engineer is reminded to use only Servomex supplied spare parts. The use
of inferior replacement components may degrade the performance and safety of the
instrument.
4. 4

To remove the cell assembly from the analyser proceed as follows with reference to Fig. 4.1 and
Fig. 5.1:
1

Isolate the process sample and flush the cell thoroughly with nitrogen, clean dry air, or
some appropriate inert medium.

Disconnect the process connections.

If fitted, disconnect the cell heater connector from the source end.
Do not strain heater connection leads.
WARNING
Very hot cells may take some hours to cool.

Allow enough time for the cell to cool before continuing.

If fitted, remove end boss purge connections.

Using a 5mm hex tool, slacken the three M6 screws securing the cell bosses at either end
of the sample cell.

If necessary, using a 5mm hex tool, slacken the four M6 screws connecting the mounting
chassis at the source end.

Carefully disconnect the thermocouple from the cell body and ensure it is protected from
damage.

Remove the cell complete with end bosses, end boss scrubbers (if fitted) are not to be
removed. The entire assembly can now be taken to the workshop. Caution, the end
bosses will need to be manually retained on the cell.
Alternatively, remove the end bosses from the cell and re-assemble to the enclosures. Fit
the blanking plugs, supplied with maintenance kit part No. 02500944, to preserve existing
scrubbers.
Working on a clean bench in a suitable environment, dismantle the cell. Pull the end
bosses off the cell body. Note that these items are one piece stainless steel. If the bosses
are fitted with absorbent scrubbers, these should be removed and replaced (Refer to
Section 9.2.).
WARNING
SPECIAL MATERIALS ARE USED FOR THE WINDOWS. SOME OF THESE
ARE TOXIC (I.E. GERMANIUM, BARIUM FLUORIDE, ZINC SELENIDE, ETC.) THESE
SUBSTANCES MUST BE HANDLED IN ACCORDANCE WITH THE RELEVANT LOCAL
STANDARDS.

4. 5

The windows are usually ground from optical single crystal material and as such may be
EXTREMELY EXPENSIVE. The windows may have special anti-reflective coatings and they may
easily be damaged by brittle fracture. The standard material chosen by Servomex is calcium
fluoride. This material is non-toxic and reasonably robust. However, some applications may
require materials such as germanium, barium fluoride, zinc selenide, etc. which are toxic and must
be handled with extreme care. See the specification sheets at the front of the User Manual for
details of the windows fitted.
Windows should be kept optically clean, i.e. free from dirt, scratches, grime, grease, fingerprints
etc. When handling windows it is recommended that cotton gloves or finger cots be worn.
WARNING - ZINC SELENIDE
If windows of Zinc Selenide are fitted and have been damaged by acidic
attack (e.g. acidic condensate or organic acids have entered the cell) it will
be likely that some of the Zinc Selenide will be reduced to Selenium metal.
This is HIGHLY TOXIC and on no account should attacked Zinc Selenide
material be handled with bare fingers. Such material should be bagged and
disposed of according to local regulations.
Cleaning of the cell windows should be carried out using cotton buds or lens tissue, detergent
and/or a suitable solvent such as Isopropanol or Arklone/Freon TF (1,1,2-Trichloro-1,2,2,
-Trifluoroethane). Avoid abrasive cleaning compounds as these will damage the optical surface.
WARNING
THE USE OF SOME SOLVENTS (ISO-PROPYL ALCOHOL, ALCOHOL ETC.)
CONSTITUTES A HAZARD. THESE SUBSTANCES MUST BE HANDLED IN
ACCORDANCE WITH THE RELEVANT LOCAL STANDARDS.
To remove the cell windows proceed as follows:

4. 6

Check that end bosses have been removed from the cell body.

Unscrew and remove, the locking ring from one end, with the special key supplied in
Special Tools kit (S2500979).

Remove the 5 wavy washers, spacer and 'O' ring, avoiding damage to the window. Seal
the cell outlet and stand the cell in a vertical position.

Connect a hand operated aspirator ( 2387-0514) to the inlet.

Apply gentle pressure to lift the window up squarely. Remove the window.

Remove the PTFE seat gasket.

The windows may now be carefully cleaned. The cell body may also be cleaned as
necessary.

Repeat instructions 1 to 7 for the other end of the cell.


Ensure all items are thoroughly clean and dry before reassembly, which is the reverse
procedure but with the following points noted:

The windows should be floated into place squarely using a hand aspirator or equivalent,
reversing the removal procedure.

Use new O rings and other cell sealing kit items making sure that all dust, hairs etc, are
rigorously excluded from them. Ensure the window surfaces are also free from solvent
residues or finger grease.

The latest design of the locking ring has a coated thread to prevent seizure - PTFE tape is
not required. FOR NON-COATED ITEMS ONLY - the threads should be covered with only
2 turns of PTFE tape (1835-3026) and the ring tightened down in stages, pausing for a
minute between stages until the metal faces of the cell body and locking ring meet when
fully tightened, with no gap between them. Where Teflon 'O' rings are being used, this will
require significant torque, with longer pauses. DO NOT use the cell inlet/outlet stubs as
levers.

After re-assembly, perform the leak check on the cell using a manometer, before installing it in the
analyser. Refer to Section 1.4 (routine leak checks) of this manual.
Do not submerge the cell or fill it with water.
Re-installation of the complete cell assembly onto the analyser is the reverse of the dismantling
process with the added requirement that, unless fitted with a purge, the end bosses should be
briefly flushed with clean dry nitrogen after the cell has been mounted and then new absorbent
scrubbers should be quickly screwed into them (refer to Section 7). Be sure to fully tighten the
screws holding the source end onto the mounting chassis if they have been slackened.
Finally, connect up the cell thermocouple to the cell body, reconnect the cell heater plug (where
fitted) and connect up any purge connections and the process connections. Return the analyser to
service as described in the 2500 Series Installation Manual. Reset zero if necessary on a true zero
sample.

4. 7

4.2

CELL PRESSURE AND TEMPERATURE TRANSDUCER FAULTS


Any fault diagnosed as a transducer failure should first be investigated by checking that the
transducer has not become simply disconnected or that the prevailing pressure or temperature
conditions are outside the working range. Otherwise, the transducer itself should be exchanged for
a known functioning unit to confirm whether or not the transducer has failed.

4.3

ELECTRICALLY HEATED CELL FAULTS


The standard electrically heated cell is factory set to operate at a fixed temperature for the
specified application in the range 30-130EC, and must not be adjusted in the field. Any suspected
fault should be investigated at the heater connections on the Power Control PCB first (see Section
8.1.8) by checking availability of power and continuity of heater windings. Also check fuse F31 on
the Power Control PCB.
All standard heated cells are protected by a pair of sacrificial thermal fuses. The temperature (T)
rating of the fuses depends on the T-rating of the analyser. Analysers rated T4 must not be
operated at cell temperatures above 80EC while 130EC is the limit for analysers rated T3. If these
limits are exceeded for a length of time, the fuses may fail and in this case the cell may need repair
by Servomex. Continuity of the heater winding will confirm this fault.
Heated cells, protected by thermal fuses, are fitted to 02500C GEN, 02500C EU2 and
02500C FM2 analysers
Power is normally supplied to the heater winding direct from the Power Control PCB at either 30,
60 or 110V, depending on the length of sample cell. Refer to Section 8.1.8.
Special heated sample cells operating from 130 to 180EC may be fitted to 02500 GEN analysers.
These are fitted with painted aluminium covers, and do not have thermal fuses. They are normally
used for high temperature emissions measurements.

4. 8

SECTION 5: DISASSEMBLY - MAJOR COMPONENTS


LIST OF CONTENTS
Section

Page

5.1

INTRODUCTION........................................................................................................................ 5.11

5.2

PREPARATION ......................................................................................................................... 5.11


5.2.1
5.2.2
5.2.3
5.2.4

5.3

Tools and Equipment.................................................................................................... 5.11


Safety Precautions........................................................................................................ 5.11
Analyser Location and Inspection................................................................................ 5.12
Electrical Mains AC Supply Disconnection.................................................................. 5.12

DISASSEMBLY.......................................................................................................................... 5.12
5.3.1
5.3.2
5.3.3
5.3.3.1
5.3.3.2
5.3.4
5.3.4.1
5.3.4.2
5.3.4.3
5.3.4.4
5.3.4.5
5.3.4.6
5.3.4.7

2500 Analyser ................................................................................................. 5.13


Sample Cell Assembly .................................................................................... 5.13
Source End Assembly..................................................................................... 5.13
Source Assembly ............................................................................................ 5.13
Power Control PCB......................................................................................... 5.14
Detector End Assembly .................................................................................. 5.15
Power Supply .................................................................................................. 5.15
Option PCB (if fitted), SIB PCB, Processor PCB ........................................... 5.15
Transmitter PCB.............................................................................................. 5.16
Transmitter PCB Cage.................................................................................... 5.16
Mains Transformer.......................................................................................... 5.16
Detector Box.................................................................................................... 5.17
Keypad and Display PCB Assembly .............................................................. 5.18

5. 1

LIST OF ILLUSTRATIONS
Figure

Page

5.1(a) Source End - Exploded View ........................................................................................................5.4


5.1(b) Source End - Exploded View ........................................................................................................5.5
5.2(a) Detector End - Exploded View......................................................................................................5.6
5.2(b) Detector End - Exploded View......................................................................................................5.7
5.3

Chopper Box - Exploded View......................................................................................................5.8

5.4

Detector Box - Exploded View ......................................................................................................5.9

5.5

AC Voltage Connections............................................................................................................ 5.10

5. 2

NOTES

5. 3

Figure 5.1(a) Source End - Exploded View


5. 4

Figure 5.1(b) Source End - Exploded View


5. 5

Figure 5.2(a) Detector End - Exploded View


5. 6

Figure 5.2(b) Detector End - Exploded View


5. 7

Figure 5.3 Chopper Box - Exploded View


5. 8

Figure 5.4 Detector Box - Exploded View


5. 9

Figure 5.5 AC Voltage Connections


5. 10

WARNING
These service and disassembly instructions are for use by trained and
experienced service engineers. To avoid injury and electric shock do not
perform any action not contained in the instructions, unless suitably
qualified.
5.1

INTRODUCTION
This section provides disassembly instructions for the 2500 series analysers. Instructions are
given in a logical sequence beginning with the removal of the sample cell followed by the source
end assembly and then the detector end assembly with direction for each subsequent item given
under its own heading. Reassembly is the reverse of disassembly except where specific cautions
are noted.
To familiarise themselves with disassembly procedure, Service Engineers should read through this
section before commencing.

5.2

PREPARATION

5.2.1

Tools and Equipment


Tools
Servicing of the analyser requires standard hand tools only: a 5mm Hex tool to open end case
covers has been supplied with the analyser.
Equipment
DVM with 4.5 digit display
Cotton gloves or finger cots
Cotton buds or lens tissue
Clean dry Nitrogen at low pressure

5.2.2

Safety Precautions
The electrical supply to this equipment is at a high enough voltage to endanger life therefore
normal electrical safety precautions must be observed. Where necessary a warning is placed prior
to the procedure it refers to as shown in the following example:
WARNING
ALWAYS ENSURE THAT THE ELECTRICAL MAINS POWER SUPPLY IS
DISCONNECTED BEFORE COMMENCING WORK.

5. 11

5.2.3

Analyser Location and Inspection


Depending on location it may be preferable or necessary to move the analyser to a more suitable
area so that servicing may be carried out safely. If so, obtain help to move the analyser. Locate
and re-fit the yellow transit handles originally supplied attached to the analyser. Remove black
pop-in plugs to do so.
Inspect the analyser. Check serial number details against service requirements.

5.2.4

Electrical Mains AC Supply Disconnection


WARNING
LETHAL VOLTAGE: MAINS AC POWER SUPPLIES ARE POTENTIALLY
LETHAL. THE SERVICE ENGINEER MUST
ENSURE THAT THE POWER SUPPLY HAS
BEEN ISOLATED BEFORE COMMENCING
SERVICING OR DISASSEMBLY.
Locate and switch off the isolation switch and disconnect the mains supply to the analyser at AC
source. Ensure that the process stream is isolated where necessary.

5.3

DISASSEMBLY
WARNING

5. 12

THE SERVICE ENGINEER MUST BE SATISFIED THAT THE 2500 SERIES


ANALYSER INSTALLATION AFTER REASSEMBLY CONFORMS TO
THE RELEVANT SAFETY REQUIREMENTS, NATIONAL ELECTRICAL
CODE AND ANY OTHER LOCAL REGULATIONS, AND THAT THE
INSTALLATION IS SAFE FOR ANY EXTREMES OF CONDITIONS
WHICH MAY BE EXPERIENCED IN THE OPERATING ENVIRONMENT
OF THE ANALYSER.

THE SERVICE ENGINEER MUST ENSURE THAT ANY CERTIFICATION


LABELS AFFIXED ON INSTALLATION REMAIN SO AFFIXED.

MANY PROCESS SAMPLES, GAS OR LIQUID STREAMS AND TEST


SAMPLES ARE TOXIC, ASPHYXIANT, CORROSIVE, FLAMMABLE OR
A COMBINATION OF SOME OR ALL OF THESE. IT IS THE
RESPONSIBILITY OF THE SERVICE ENGINEER TO ENSURE THAT
ADEQUATE PRECAUTIONS ARE TAKEN DURING DISASSEMBLY,
ASSEMBLY, CHECKS OR TESTS AND SUBSEQUENT REINSTALLATION OF THE 2500 SERIES ANALYSER AND CONNECTION
OF SAMPLES. ALL SAMPLE LINE CONNECTIONS MUST BE
CHECKED FOR NO LEAKS.

5.3.1

2500, 2510, 2520, 2550 Analysers


To disconnect the analyser from the AC mains supply:

5.3.2

Ensure the mains AC supply is isolated at source

Remove 4 screws retaining mains filter unit cover

Using a suitable size slotted screwdriver, loosen the screws on the terminal block inside
the mains filter unit, remove and insulate the mains supply leads.

Sample Cell Assembly


For instructions to remove the sample cell and cell windows please refer to Section 4: Cell
Maintenance.

5.3.3

Source-End Assembly

5.3.3.1 Source Assembly


Refer to Fig. 5.3
Standard sources are long-life sealed units and serviced by replacement if necessary.
Note that special short wave sources No. 02500/943 are serviced by simple bulb replacement and
recalibration. Do not touch with bare fingers.
Removal of Long wave and Soft UV sources:
1

Unplug the electrical cable from PL4 on the Power Control PCB (item 6b Fig. 5.1a).

Unscrew and remove the 3 hex.skt screws (item 4). Note: On long wave source, it may be
helpful to remove the leak test plug (items 9 and 10) to enable a full turn with the hex key.

Place assembly and screws on one side for re-connection which is the reverse of
disconnection. Ensure the leak test plug is reinstalled.

Note:

The Chopper Box must be flushed with Nitrogen after source reinstallation. See Section
7. The analyser should be checked for span and zero after returning to operation.

5. 13

Removal of Short Wave source:


1

Remove the faulty source bulb by releasing its legs from the screwed terminal block which
holds it.

Using cotton gloves, or holding it with tissue paper, take the new source out of its packet
and mount it in exactly the same position as the old source.

Note: DO NOT touch the new halogen source bulb with bare fingers or irreparable
degradation will occur!

Power up the analyser. Observe the illumination of the new source bulb to confirm
satisfactory replacement. Allow the unit to fully warm up.

Inspect the received detector signal levels (which should be typically 250,000-500,000) If
the highest signal (Meas. 1 or Ref. 2) is not around 500,000 try to increase this signal by
making fine adjustments to the orientation of the source. Do this by loosening the screws
which hold the mounting bracket to the source body and by sliding the mounting bracket
vertically and/or horizontally to gain maximum signal. Make small adjustments and allow
several seconds for the diagnostics to respond after each movement. Lock the screws at
position of maximum signals.

If this maximum is still significantly below 500,000 very carefully adjust the detector gain
potentiometer beneath the sealing label on top of the detector housing (item 20 in Fig.
5.2(b)) in the detector end of the analyser, to achieve the correct value. Conversely, if the
signal is too high, reduce it with the detector gain pot. Make sure that the correct zero
sample is flowing through the analyser sample cell all the time. Note that significant
adjustments of this potentiometer will significantly affect the calibration of the analyser.

The analyser should be checked for span and zero after returning to operation.

5.3.3.2 Power Control PCB


Refer to Fig. 5.1(a & b).
To remove the Power Control PCB (item 6):

5. 14

Disconnect all wires, except wire links, from terminal blocks. Label or record all locations
to ensure reconnections are correct.

Unplug all cable connectors. With the exception of PL1, press in side latches to release.
For PL1 press down side latches to release.

Unscrew the 3 captive screws (item 6c) spaced along the front edge of the PCB securing
it to the floor of the end-assembly.

Insert a finger of one hand under the corner of the PCB in the cut-out provided and with
the other hand, grasp TB34 and together pull the PCB out.

5.3.4

On reinstallation, ensure the back of the PCB properly engages in the support brackets in
the back of the end-assembly.

Detector End Assembly

5.3.4.1 Power Supply


Refer to Fig. 5.2(a).
To remove the Power Supply (item 33):
1

At the metal cage (item 30), loosen the M3 hex.skt screws (item 30b) each side and open
the lid.

Remove the 2 hex.skt screws (item 33a) securing the Power Supply to the heatsink
(item 33b)
Grasp the Power Supply firmly, pull upwards to unplug it.

5.3.4.2 Option (if fitted), Sensor Interface Board (SIB) and Processor PCB=s
Caution
Although the SIB and Option PCB locations are identified on the Transmitter
PCB they are fitted with identical edge connectors and can be transposed. No
damage will be done to the PCB=s electrically if they are but the analyser will
not function correctly (identified by a fault message being displayed).
Refer to Fig. 5.2(a).
To remove the PCB=s (Items 31,32 and Option PCB):
1

At the metal cage (item 30), loosen the hex.skt screws (item 30b), each side and open the
lid.

Grasp the PCB firmly with both hands, pull upwards to unplug from the connector.

Before reassembly ensure that connector pins are not damaged in any way.

5. 15

5.3.4.3 Transmitter PCB


Refer to Fig. 5.2(a).
To remove the PCB (item 28):
1

Carry out actions 5.3.4.1 and 5.3.4.2 if required.

Disconnect all wires. Label or record all locations to ensure reconnections are correct.

Unplug all cable connectors. To release PL1 and PL2 press side latches down and out.
For PL3 press latches in to release.

Unscrew the 2 captive screws (item 28a), one on each front corner of the PCB, securing it
to floor of the end-assembly.

Grasp the metal cage and pull the PCB upwards and out from support brackets.

On reinstallation, ensure the back of the PCB properly engages in the support brackets in
the back of the end-assembly.

5.3.4.4 Transmitter PCB Cage


Refer to Fig. 5.2(a).
To remove the cage (item 30):
1

Carry out action 5.3.4.3.

Turn the PCB assembly upside down to rest on the cage.

Using a slotted screwdriver remove the 4 screws (item 30a) securing the cage to the PCB.

Separate the cage and PCB.

5.3.4.5 Mains Transformer


Refer to Fig. 5.2(b)
To remove the mains transformer (item 21):

5. 16

Carry out actions 5.3.4.1 to 5.3.4.3 inclusive.

Unscrew the 2 nuts and washers (item 26a) securing the transformer cover (item 26), and
take the cover off.

Separate the plug and socket connection fitted to the bracket (items 22 - 24) mounted
alongside the transformer.

Disconnect the earth tag from the earth stud.

Unscrew the two M5 hex.skt screws (item 21a) from the underside of the case which
secure the transformer. Remove the transformer.

5.3.4.6 Detector Box


Refer to Fig 5.4.
To remove the detector box:
1

Carry out actions 5.3.4.1 to 5.3.4.3 inclusive, referring to Fig. 5.2.

Unscrew the 4 screws (item 4, Fig. 5.4) securing the lid (item 3) using a 3mm hex tool,
and remove the lid (2510 analyser type secured with 2 screws located on top edge).

Unscrew the 3 screws (item 6) securing the Detector PCB assembly (item 5) to the
detector mounting block (item 12) and take the PCB from the box.

Remove the leak test port adaptor and dowel pin (items 7 and 8) and remove absorbent
scrubber (items 9 and 10).
Note: The scrubber and leak test port are not fitted to purge certified analysers, these are
replaced by a purge adaptor and flexible purge duct fitted between the detector box and
the detector end enclosure.
Loosen a purge duct clamp to allow the detector box to be removed.

Loosen the screws (item 2) securing the detector block assembly and box to the detector
mounting boss (item 15).

Holding the assembly, fully unscrew the already loosened screws. The assembly and box
can now be separated and both items removed.

After reassembly, the detector block must be flushed and sealed with Nitrogen. Refer to
Section 7. If the detector is replaced, the analyser should be checked for span and zero
after returning to operation.

5. 17

5.3.4.7 Keypad and Display PCB Assembly


CAUTION
The display panel mounted on the PCB is made of glass and therefore fragile.
Handle with care during disassembly.

Refer to Fig. 5.2 (a and b)


To remove the Display PCB and Keypad (items 17 and 18):

5. 18

At the detector end cover unscrew the 4 screws securing the cover and allow the cover to
rise. Beware of gas strut action (item 5).

To provide a comfortable working position, disconnect one end of the gas strut. To do this,
insert the blade of a screwdriver or similar between the strut and the side of the chassis
and ease the strut from the locating stud. This will allow the cover to be fully opened.
Carefully rest the opened cover on the source end assembly or sample cell.

Unscrew the 4 hex.skt screws (item 16a) securing the vented metal cover (item 16) to the
detector - end cover and 2 screws securing the ferrite clamp (items 17c and 17d).
Remove the ferrite clamp and screws. Take care not to drop the ferrite clamp as it may be
damaged.

Carefully unplug the ribbon cable (item 17b) from the Display PCB.

At the Display PCB, unscrew the 4 hex.skt screws (item 17a) securing the Display PCB to
the Keypad (item 18). Carefully lift the Display PCB and disconnect the keypad ribbon
connector from the back of the PCB. When refitting the Display, remember to connect the
keypad ribbon cable before securing the Display PCB. Note that it is much easier to refit
the Display PCB to the Keypad before securing the Keypad into the cover. Also note the
earth wire on one of the Display PCB securing screws.

Unscrew the 4 hex.skt screws (item 18a) securing the Keypad to the cover and remove
the Keypad. Note the rubber O-ring gasket in the groove on the front of the Keypad. When
refitting the Keypad, ensure the O-ring gasket is correctly located in the groove and tighten
the 4 hex skt screws equally to ensure the Keypad seats correctly in the cover.

Carefully reconnect the ribbon cable connector to the Display PCB, arrange the ribbon
cable and refit the ferrite clamp.

Refit the vented metal cover. Refit the gas strut.

SECTION 6: CHOPPER BOX


LIST OF CONTENTS
Section

Page

6.1

INTRODUCTION...........................................................................................................................6.2

6.2

DESCRIPTION ..............................................................................................................................6.2

6.3

DISASSEMBLY.............................................................................................................................6.3

6. 1

NOTES

6. 2

6.1

INTRODUCTION
The chopper box No. 02500/920 is mounted in the source end and contains the solid-state
interference filters located on a rotating chopper wheel. Appropriate wavelengths configured for
analysis are selected by these filters. The infrared source assembly is mounted on the box lid.

6.2

DESCRIPTION
The chopper box assembly comprises: machined chopper box; chopper box lid; motor mounting
plate, filter mount and cut off filter, motor and filter disc assembly; Motor Connection PCB; lens
assembly; scrubber sachet port. Two heater resistors are mounted in open ended recesses in the
top and bottom of the box.
The motor assembly (with ribbon cable)(S2500986) locates in an open ended elongated hole on
the mounting plate and is secured in position with 2 M3 x 6 screws. The ribbon cable passes
through an opening in the side panel and connects to the Motor Connection PCB.
The Motor Connection PCB (S2500906) is fitted to the right hand side of the box, sealed via an `O'
ring, effectively forming a side panel, with 3 M4 x 6 pan head screws. Locating dowels are
provided for correct alignment.
The mounting plate (02500410) also forms the mount for the cut-off filter and filter retaining plate.
The cut-off filter, often basically triangular in shape, is first placed over a recessed machined
aperture. The filter retaining plate (Servomex Part No. 01201414) is then placed over it and
secured with 2 pan head M3 x 5 screws. The plate, complete with the aforementioned items, is
fitted to the box with 4 M4 x 16 hex.skt captive screws.
The filter disc assembly is positioned on the motor drive spindle and secured with 3 hex.skt M3 x 5
screws. The measure and reference interference filters are mounted on the outer side of the disc,
facing away from the motor, with wavelengths engraved next to each one.
The lens assembly, comprising mounted lens, sleeve, spacer and `O' rings, is fitted to the rear of
the box and held in position with 2 M4 x 16 hex.skt screws.
A scrubber sachet (if fitted) is located in the port on the left side of the box, retained and sealed
with an >O= ring with an M25 brass blanking plug. The plug can be removed using a 10mm hex
key. Note: Purge certified units are not fitted with either a scrubber sachet or blanking plug, the
blanking plug being replaced by a purge adaptor and flexible purge duct.
The chopper box lid (02500/407) provides a circular recessed mount for the source assembly and
a leak test port with mating plug and adaptor. The lid is secured to the box using 4 hex.skt M4 x 16
captive screws.
The resistors (2666 - 2255) are 10 ohm 15 watt devices and are secured directly to the box each
with 2 pan head M2 x 6 screws. The resistors act as heating elements for the box. Temperature of
the box is sensed by a thermistor mounted under one of the resistors. Resistors and thermistor
are connected via the Motor Connection PCB.

6. 3

6.3

DISASSEMBLY
Refer to Fig. 5.3.
1

Disconnect source cable from PL4 on the Power Control PCB (see Section 5.3.3.1).

At the lid (item 2), unscrew the 4 captive screws (items 5 and 6) using 3mm hex tool and
remove the lid, with the source attached. Store carefully. On no account touch a short
wave source lamp with bare fingers.

Using the 2.5mm hex tool remove the 3 screws (item 18) securing the filter disc assembly
(item 17) from the motor drive spindle. Carefully hold the filter disc lightly at the edges to
prevent rotation while doing this. Do not touch the filters.

Align the cut-out section of the filter disc with the opto-coupler (item 35) and carefully pull
off the disc without touching the filters. Put in a safe place and protect filters from damage
by wrapping with lens tissue or soft clean cloth. Note correct orientation (engraved
markings and filters usually on the front).

Carefully unplug the motor assembly ribbon cable connector from the Motor Connection
PCB (item 25).

At the mounting plate (item 15) unscrew the 4 captive screws (items 14 and 6) and
remove plate and motor assembly (item 16). Do not touch the fixed cut-off (bandpass)
filter.

Unscrew the 2 pan head screws (item 36) retaining the motor to the mounting plate.

Remove the motor, by gently pushing it backwards out of the mounting hole and then
sideways via the slot.

On reinstallation, ensure the filter disc is correctly orientated, i.e. measure and reference
infrared filters facing away from the motor. (NB some wheels have filters on both sides,
with bandpass filters towards the motor).

10

The completely reassembled chopper box must be flushed and sealed with clean dry
nitrogen after reassembly. Refer to Section 7.
Note: Purge certified analysers have a vented chopper box assembly, therefore flushing
and leak testing is unnecessary.

Note : If any changes to the filters have been necessary, a complete recalibration must be
conducted, - Consult Servomex.

6. 4

SECTION 7: REASSEMBLY - LEAK TESTING AND FLUSHING


LIST OF CONTENTS
Section

Page

7.1

INTRODUCTION...........................................................................................................................7.3

7.2

LEAK TESTING ............................................................................................................................7.3

7.3

FLUSHING ....................................................................................................................................7.3

7.4

SCRUBBER REPLACEMENT ....................................................................................................7.3

7.5

VOIDS ............................................................................................................................................7.3
7.5.1
7.5.2
7.5.3

Cell Assembly .................................................................................................................7.3


Detector Assembly ..........................................................................................................7.4
Chopper Box assembly ...................................................................................................7.4

7. 1

NOTES

7. 2

7.1

INTRODUCTION
The analyser contains a number of sealed voids (dead spaces) which are in the optical path and
these are the chopper box, detector box and 2 cell end bosses.
If any of these voids are opened during servicing or repair they must be leak tested and
flushed and sealed with clean dry nitrogen after reassembly and have their scrubbers
replaced, if fitted.

7.2

LEAK TESTING
A leak test port is fitted to the chopper box lid and to the detector block, with a 4mm stainless
dowel which may be removed by pressing on the red release collet and pulling on the dowel pin
simultaneously. A 4mm OD tube may be fitted to the leak test port; the tube is automatically sealed
via an >O= ring contained in the leak test port. To refit the dowel pin, press into position using
gentle pressure. Ensure that voids are not contaminated during leak testing.
Procedure
Leak tests should be carried out in a stable temperature environment otherwise erroneous results
may occur.
Connect a manometer to the void and pressurise to 6 feet of water gauge (3 psig/20kPa). Allow
approximately 1 minute for the manometer to stabilise and note the manometer level. After
approximately 30 minutes there should be no detectable pressure drop. If available , the use of
Helium leak test equipment should be considered as an alternative.
Note: Purge certified analysers have vented chopper box and detector assemblies, therefore leak
testing of these items is unnecessary.

7.3

FLUSHING
Pass clean dry nitrogen through the void at a flow rate in excess of 1l/min. Supply pressure not to
exceed 5 psi. Seal void immediately after flushing so as to avoid the ingress of air.

7.4

SCRUBBER REPLACEMENT
It is recommended that when a void is opened the scrubber (if fitted), is always replaced. Spare
scrubbers supplied on delivery of the analyser should only be used if they are still sealed in their
original packaging and stored in an airtight container. Scrubbers have a G1/8 (1/8" BSPP) thread
and are sealed with a Dowty bonded seal.
Note: Purge certified analysers are not fitted with scrubbers.

7.5

VOIDS

7.5.1

Cell Assembly
There are cell end bosses at each end of the cell. These may be fitted with two scrubbers. One of
these may be removed to facilitate leak testing and flushing. On instruments where the end
bosses are constantly purged no testing is necessary.
7. 3

On some instruments there is an additional window fitted in the casting insert which divides the
end boss cavities in two. The >cell side= end boss void may be leak tested and flushed via the cell
end boss scrubber ports. No facility is provided for leak testing the >end-box side= void although
this should still be flushed at time of assembly.
7.5.2

Detector Assembly
This is normally fitted with a scrubber and a leak test port. Remove the dowel from the leak test
port and fit a 4mm OD tube to facilitate leak testing. Loosen the scrubber 2 turns and introduce
nitrogen via the leak test port to flush. Tighten the scrubber, remove the 4mm OD tube and refit the
dowel to the leak test port.
Note: Not necessary on purge certified analysers.

7.5.3

Chopper Box
This is fitted with a leak test port in the centre of the lid. Remove the dowel and fit a 4mm OD tube
to facilitate leak testing. Loosen the brass scrubber port 2 turns and flush the box via the tube.
Tighten the scrubber port and refit the dowel pin to the leak test port.
Note: Not necessary on purge certified analysers.

7. 4

SECTION 8: PRINTED CIRCUIT BOARDS


LIST OF CONTENTS
Section
8.1

Page
POWER CONTROL PCB .............................................................................................................8.6
8.1.1 General.............................................................................................................................8.6
8.1.2 Mains Input Voltages .......................................................................................................8.6
8.1.3 Solenoid Valve Driver ......................................................................................................8.7
8.1.4 Source Voltage Control ...................................................................................................8.7
8.1.5 Chopper Box Heater Control...........................................................................................8.8
8.1.6 Motor Control ...................................................................................................................8.9
8.1.7 Pressure Sensor Amplifier...............................................................................................8.9
8.1.8 Cell Temperature Control............................................................................................. 8.10
8.1.9 Diagnostics.................................................................................................................... 8.11
8.1.10 Fault Displays ............................................................................................................... 8.12
8.1.10.1 Source Voltage High/Low............................................................................................. 8.12
8.1.10.2 Sample Flow Failure ..................................................................................................... 8.13
8.1.10.3 Chopper Temp High/Low ............................................................................................. 8.13
8.1.10.4 Cell Temp High/Low ..................................................................................................... 8.13
8.1.10.5 Source Temp High/Low................................................................................................ 8.13
8.1.10.6 Infrared Source Failure................................................................................................. 8.14
8.1.10.7 Chopper Motor out of Lock........................................................................................... 8.14
8.1.10.8 Detector Signal High .................................................................................................... 8.14

8.2

TRANSMITTER PCB ................................................................................................................. 8.15


8.2.1
8.2.2
8.2.3
8.2.4
8.2.5

Three Channel Synchronous Detector ........................................................................ 8.17


Analogue/Digital Multiplexer and I/O (Input/Output).................................................... 8.17
Serial EEPROM ............................................................................................................ 8.18
Autocal Latches ............................................................................................................ 8.18
Switch Settings.............................................................................................................. 8.18

8.3

SENSOR INTERFACE PCB...................................................................................................... 8.19

8.4

OPTION PCB=s .......................................................................................................................... 8.20

8.5

MICROPROCESSOR PCB ....................................................................................................... 8.21


8.5.1
8.5.2
8.5.3

Watchdog Circuitry ....................................................................................................... 8.21


Memory Test ................................................................................................................. 8.22
Diagnostics.................................................................................................................... 8.22
.....continued overleaf

8. 1

LIST OF CONTENTS (continued)


Section
8.6

Page
DISPLAY PCB............................................................................................................................ 8.22
8.6.1

Diagnostics.................................................................................................................... 8.23

8.7

POWER SUPPLY....................................................................................................................... 8.23

8.8

DETECTOR PCB ....................................................................................................................... 8.24


8.8.1 Diagnostics .................................................................................................................... 8.24
8.8.1.1 Detector Setup for 2500 Series Analyser..................................................................... 8.25
8.8.1.2 Detector Setup for 2510 Series Analyser..................................................................... 8.25
8.8.1.3 Detector Setup for 2550 Series Analyser..................................................................... 8.25

8.9

MOTOR CONNECTION PCB.................................................................................................... 8.26

LIST OF ILLUSTRATIONS
Figure

Page

8.1

2500C PCB Locations...................................................................................................................8.3

8.2

Option Board Layout .....................................................................................................................8.4

8.3

Microprocessor PCB Layout.........................................................................................................8.5

8.4

Diagnostic Waveforms ............................................................................................................... 8.16

8. 2

Note:

Power Supply Part Number 02000925


Optional Output PCB Part number 02000916
Sensor Interface PCB part number 02000934

Figure 8.1 2500C PCB Locations


8. 3

Figure 8.2 Option Board Layouts

8. 4

Figure 8.3 Microprocessor PCB Layout


8. 5

8.1

POWER CONTROL PCB

8.1.1

General
The Power Control PCB (02500/911A) provides power distribution, control signals and
interconnections to most of the analyser's circuitry. The PCB also provides most of the fuse
protection for the analyser. Fuses values and circuitry protection are listed below.
Fuse

Value

F30
F31
F32
F33
F34
F35
F36
F37
F38

1A(T)
2A(F)
2A(F)
1A(T)
2.5A(T)HBC
2.5A(T)HBC
3.15A(T)HBC
1A(T)
800mA(F)

T = anti-surge

F = quick blow

Protection
20V AC supply to Source Power Supply
20V AC supply to Cell Heater Control
20V AC supply to Chopper Box Heater Control
20V AC supply to Solenoid Valve Drivers
230V AC from TB31 to Transformer Primary
115V AC from TB31 to Transformer Primary
AC mains input to mains filter
AC mains input to DC power supply
Source Supply
HBC (or H) = high break capacity

The PCB is located in the source-end assembly. For removal instructions refer to Section 5.
8.1.2

Mains Input Voltages


The incoming mains AC voltages are routed to the primary winding of the mains transformer and
the Power Supply PCB via terminal blocks (TB) TB30, TB31, mains filter (X1) and PL3. Analyser
working voltages are selected at TB31 using wire links. Voltages requirements and their wire links
are as follows:
115V nominal TB31 pin 1(com) to TB31 pin 2
230V nominal TB31 pin 1(com) to TB31 pin 3
Filtered mains voltages are taken via PL3 to the Power Supply PCB fitted to the Transmitter PCB.
The transformer secondary provides operating voltages (wrt 0V) of 110V, 60V, 30V, for the sample
cell heater, operating voltages of 20V for solenoid valve drivers, source power supply and chopper
box heater control.

8. 6

8.1.3

Solenoid Valve Driver


The solenoid valve driver circuitry provides three volt free contacts rated 240V AC/1.0A,
30VDC/1.0A. These are designated SV1, SV2 and SV3.
An auxiliary circuit provides +24VDC for internal solenoid valve driver, maximum power available
is 12W. The out put is routed to TB36-5&6(+24VDC), 0V connection being made to TB36-3&4.
An AC voltage of 20V is input via PL3 and fed, via a 1 amp fuse F33 to full wave bridge rectifier
BR2 for conversion to an unstabilised, smoothed +24V DC.
Solenoid valve driver input voltages and signals are listed below.
Circuit Reference

PL3 pin

PL1 pin

20V AC
0V AC
SV3
SV2
SV1

15
14
-

30
31
32

Valve driver terminals are listed below. Note that these are volt free contacts.

8.1.4

Circuit Reference

TB34 pin

SV1
SV2
SV3

1,2
3,4
5,6

Source Voltage Control - Infrared Analysers


The source voltage control circuitry provides operating voltages and signals for the IR source
assembly.
20V AC is input via PL3 and full wave rectified and smoothed by BR1 and capacitor C1 to +24V
DC providing the operating voltage for the switched mode power supply IC3. The line input is fuse
protected by F30 (1 amp). Further DC regulated DC voltages of +12V and +5V are required which
are provided by on-board regulators, IC18 and IC1.
The IR source assembly voltages are monitored continuously with information being fed back to
on-board control circuitry and via PL1 to the Microprocessor PCB. A green coloured light emitting
diode (LED) D6, monitoring the SOURCE FAIL signal line, will be lit under normal operating
conditions (see paragraph 8.1.9). Test points, TP1 and TP2, are available for external monitoring
of the +/- ve sense lines.
Source voltage control input signals are listed below.
8. 7

Circuit Reference

PL1 pin

PL3 pin

20V AC(from : PS PCB)


0V AC (from : PS PCB)
Sync. pulse
Source fail

27
26

9
8
-

Source voltage control output signals are listed below.

8.1.5

Circuit Reference

PL1 pin

PL4 pin

Source voltage
Source voltage error
-ve supply
-ve sense
+ve sense
+ve supply

7
8
-

1
2
3
4

Chopper Box Heater Control


The chopper box heater control circuitry provides the necessary voltages and signal lines to
maintain the chopper box assembly at a steady temperature. Signals from the chopper box
thermistor are fed via PL2 to a Wheatstone bridge arrangement. Changes in voltage across the
bridge are sensed by differential amplifiers which in turn control 20V AC voltage to the heaters 1
and 2. Test points TP4 and TP5 are provided which enable monitoring of differential amplifier
signal levels.
Input signals are listed below.

8. 8

Circuit Reference

PL1 pin

PL2 pin

0VA
-15VA
+15VA
Chop.box temp.+ve
Chop.box temp.-ve
Sync.pulse
20V AC c/box htr
Chop.box therm.
0V AC c/box htr
Chop.box therm.

13,14
15,16
17,18
1
2
13
18
20

12
11
-

PL3 pin

Signal outputs are listed below.

8.1.6

Circuit Reference

PL1 pin

PL2 pin

Sync.pulse
20V AC switched(c/box htr)
0V AC (c/box htr)

27
-

17,19
15,16

Motor Control
The motor control circuitry maintains the chopper box motor at a constant 11.25 Hz (675 rpm) and
provides visual indications if faults occur.
LEDs D5 and D7, normally lit, will extinguish when motor operation is impaired in any way. D5
monitors the +24VP DC rail and if extinguished, this will indicate the +ve supply to the board is
missing. The +24VP, input via PL1 pins 37 and 38 is used to provide a stabilised +12VP supply
(IC26) for the control circuitry. If D7 is extinguished it will indicate that the motor has stopped.
Motor control input signals are listed below.
Circuit Reference

PL1 pin

PL2 pin

0VP
+24VP
Hall I/P (b)
Hall I/P (c)
Hall I/P (a)
Motor phase (c)
Motor phase (b)
Motor phase (c)
0VD
+5VD

33,34,35,36
37,38
-

3
4
5
6
7
8
9
10
12
14

Motor control output signals are listed below.

8.1.7

Circuit Reference

PL1 pin

Motor current
Motor lock

6
25

Pressure Sensor Amplifier


The pressure sensor amplifier circuitry, when utilised, simply amplifies signals from the pressure
transducer via TB35 pins 1 and 2. the pressure signal is output at PL1 pin 27.

8. 9

8.1.8

Cell Temperature Control


The cell temperature control circuitry receives voltage signals from the cell thermocouple via PL5.
The signals are then amplified, and in conjunction with switch SW2, activate the TRIACs to allow
the correct AC heater voltage to reach the sample cell. The circuitry generates its own +5V
reference voltage, derived from the +15V rail, required for the differential amplifier. A test point TP3
wrt 0V is provided to monitor the CELL TEMP. voltage level.
Cell temperature control input signals are listed below.
Circuit Reference

PL5 pin

T.C. +ve
T.C. -ve

1
2

Cell temperature control output signals are listed below.


Circuit Reference

PL1 pin

TB33 pin

Cell sample temp.


Cell sample temp.S.P.
Cell T.C. fail
Cell heater (A)
Cell heater (B)

9
10
11
-

1
4

Switch settings for SW2 showing the respective cell temperatures are listed below.

8. 10

SW2 Setting ............Cell Temp. EC

SW2 Setting .......... Cell Temp. EC

0 .................................................. 30

8.................................................110

1 .................................................. 40

9.................................................120

2 .................................................. 50

A ................................................130

3 .................................................. 60

B ................................................140

4 .................................................. 70

C ................................................150

5 .................................................. 80

D ................................................160

6 .................................................. 90

E ................................................170

7 ................................................ 100

F.................................................180

8.1.9 Diagnostics
The Power Control PCB has 3 diagnostic LED's, D5, D6 and D7. They should be used to identify
the nature of any fault. All circuits operate fully independently with the exception that any motor
failure will also shut down the source voltage to 0.4V.
+24v
Supply D5

Source
Voltage D6

Motor
Lock D7

Diagnosis

On

On

On

Normal operation

On

Off

Off

PL2 disconnected;
Chopper motor failure;
Motor control failure.

On

On

Off

As above but with source also


disconnected/open circuit

On

Off

On

Source driver failure;


fuses blown (check F30,34,35,38)

Off

Off

Off

No mains power;
fuses blown (Check F36,37);
PL1 Disconnected; PS PCB (+24v)
failure

Off

On

Off

PL1 disconnected; PS (+24v) failure;


Fuse F37 blown and source also open
circuit.

8. 11

8.1.10 Fault Displays


8.1.10.1 Source Voltage High/Low
Check source voltage between TP2 wrt TP1 (earth) and check settings of switches SW3 and SW4.
Note: switch settings have been factory set and must not be altered.
SW3 Settings
Source Voltage (Volts)

Rotary
Switch SW3
position

SW4-1OFF & SW4-2ON

0
1
2
3
4
5
6
7
8
9
A
B
C
D
E
F

0.75
0.80
0.85
0.90
0.95
1.00
1.05
1.10
1.15/0.75
1.20/0.80
1.25/0.85
1.30/0.90
1.35/0.95
1.40/1.00
1.45/1.05
1.50/1.10

SW4-1OFF & SW4-2OFF SW4-1ON & SW4-2OFF


1.5
1.6
1.7
1.8
1.9
2.0
2.1
2.2
2.3/1.5
2.4/1.6
2.5/1.7
2.6/1.8
2.7/1.9
2.8/2.0
2.9/2.1
3.0/2.2

3.0
3.2
3.4
3.6
3.8
4.0
4.2
4.4
4.6/3.0
4.8/3.2
5.0/3.4
5.2/3.6
5.4/3.8
5.6/4.0
5.8/4.2
6.0/4.4

Note: Dual voltages are shown in some cases these are selected by SW4-3.
SW4 Settings
Switch position
3 ON
3 OFF

Function
Disable 2.7V limit
Enable 2.7V limit (switch positions 0 to 7 repeated

For standard long wavelength IR sources the voltage is typically set from 2.0V to 2.2V. For short
wavelength sources in the NIR region (2500 liquid applications, etc.) the Quartz Halogen source is
set to 2.6V. For short wavelength sources in the visible and Soft UV versions it is set to 4.4V.
If the source voltage supplied is not according to the switch settings, and no other diagnostics are
indicated, a control chip failure is diagnosed and the Power Control PCB must be replaced. Note
that a motor failure shuts down the source voltage to 0.4V.
8. 12

8.1.10.2 Sample Flow Failure


Check the sample flow and flow switch operation where fitted. This diagnostic will be triggered by
continuity between TB36/1 and TB36/2.
8.1.10.3 Chopper Temp High/Low
Check chopper box temperature at TP4 using a digital voltmeter (DVM) (0.10V per EC -20 to
+80EC) wrt TP5. Standard chopper box set point is 65EC although steady operation in the range
64-68EC is acceptable. Check ambient temperature and ensure analyser has had sufficient time
to warm up if recently switched on (requires up to 2 hours). Ambient temperature outside the
specified range (refer to instrument specification in appropriate User Manual) may result in these
faults.
8.1.10.4 Cell Temp High/Low
Check cell temperature at TP3 (1mV per EC) wrt TP5, and compare with cell temperature
diagnostic display and the factory setpoint (see specification sheet in the User Manual). Check
ambient temperature and ensure that analyser has sufficient time to warm up if recently switched
on. Determine whether overheating or excessive cooling may be due to the sample stream.
WARNING
TB32/1 supplies 110V AC nominal
Severe shock hazard
If no discontinuity/supply failure is found, a control circuit failure is indicated and the Power Control
PCB must be replaced.
8.1.10.5 Source Temp High/Low
Source temp low would normally occur as a consequence of an infrared source failure fault. The
source unit should be checked for resistance of the element (measured between pins 1 & 4 of the
source pins) and should be replaced if it is greater than 3 ohms, and the instrument recalibrated.
The only circumstances where source temp high or source temp low faults would otherwise arise
would be on 2500 SBTW (triple wavelength) analysers where severe source ageing has occurred
or strong interferences are present on one or both of the reference wavelengths. The latter is the
more likely, and steps should be taken to identify whether process composition changes or other
application problems are the cause, before a change of Power Control PCB is attempted.

8. 13

8.1.10.6 Infrared Source (Driver) Failure


An actual infrared source drive circuit failure will be confirmed by D5 and D7 being on, and D6
being off, on the Power Control PCB. Otherwise, if D7 is off, the fault could be related to the
chopper motor being out of lock since source power is interlocked to motor lock (source drive
reduces to 0.4V). Refer to next section (8.1.10.8).
8.1.10.7 Chopper Motor out of Lock
A chopper motor out of lock fault indicates that the motor is not rotating at the correct speed (which
is 675 revolutions per minute). Under these conditions the circuit may reduce the source voltage to
prevent possible damage to some types of IR filter as there is a risk that these might otherwise
remain stationary in the IR beam. Under these conditions Infrared Source Failure and Detector
Signal Low faults may also be displayed.
Check whether Power Control PCB D7 is off or on.
If D7 is on (motor out of lock) :
Check interconnections to DIN Controller (if fitted)
Change SIB PCB
Change Power Control PCB
If D7 is off (motor out of lock) :
Check whether D5 (+24V supply) is off or on.
If D5 is off,
check PL1 connections are good and check other items as
noted in section 8.1.9
If D5 is on,
check PL2 connections are good;
check and replace chopper motor;
change Power Control PCB
8.1.10.8 Detector Signal High
Excessive detector signal may be due to loss of a liquid sample, loss of a normal background gas
sample which has a high absorbance, or possible loss of an IR filter from the chopper wheel.
Check signals at Transmitter PCB, refer to Section 8.2.

8. 14

8.2

TRANSMITTER PCB
The Transmitter PCB (02500/904A) provides the interface between the Power Control PCB and
most of the other PCBs used by the analyser (see Fig 8.1). It provides three test points and a 0V
reference point along its front edge.
Functions carried out by the Transmitter PCB are three channel synchronous detection, analogue
and digital multiplexing, input/outputs and storage of calibration data in EEPROM devices.
Note: the term transmitter is generic and is merely descriptive of how voltages and signals are
progressed for onward transmission although there is a printer RS232C port routed from the
Microprocessor PCB.

8. 15

Figure 8.4 Diagnostic Waveforms


8. 16

8.2.1 Three Channel Synchronous Detector


The three channel synchronous detector (TCSD) circuitry receives pulses from the detector box,
via connector PL3. The output from the detector is a composite signal comprising a stream of
pulses corresponding to the transmission of energy through the sample at each filter wavelength.
It is changes in the magnitude of these pulses with change in sample composition which enable
measurements to be made. Pulse amplitudes are determined by sampling the pulse stream. For
optimum accuracy, two samples are taken for each pulse; one, just before the pulse (baseline) and
one during the pulse (peak).These samples are fed into a differential amplifier and filter circuit. The
subsequent output represents the pulse height and hence the transmission of the sample at the
wavelength concerned. A standard 2500 has three such circuits (each capturing two samples)
and hence output signals from up to three filters. Using a dual beam oscilloscope and with
reference to Fig.8.4, monitor the signal at TP2 wrt 0V. On the transmitter PCB.
To aid sampling, the periphery of the filter disc assembly is slotted, with the slots being detected by
a photo-interrupter (part of the Motor Connection PCB Fig.5.3 item 35). The three filter disc
assembly has thirteen slots, twelve of which are used to control the six samples, (one to start each
sample, one to end it) and a thirteenth, wider slot, which is detected by a discriminator circuit and
serves to identify the first filter. The pulse stream from the opto-interruptor may be viewed, using
the oscilloscope on TP3 wrt 0V and the output from the discriminator may be viewed on TP1 wrt
0V. The two or three signals, monitored at TP2, feed into two or three inputs of the analogue
multiplexer. The multiplexer has a total of six inputs for transmission signals, and the other three
connect to plug PL5 together with the detector signal pulse stream, the discriminator output and
the timing slot pulse stream. This permits connection of optional additional circuits (on 2510 and
2550 models) to allow up to six filter pulses to be sampled.
8.2.2 Analogue/Digital Multiplexer and I/O (Input/Outputs)
The analogue multiplexer provides sixteen inputs, six of which are used for the incoming filter disc
signals. Remaining inputs are used for sample pressure and temperature signals and various
analogue diagnostics such as source voltage and motor current. The output from the analogue
multiplexer is passed via a variable gain circuit to a current output circuit which has a range of 4 20 mA.
The digital multiplexer has eight inputs and receives digital diagnostic data such as motor lock, cell
thermocouple fail, etc. Output from the digital multiplexer is fed to a CMOS driver and filter circuit.
Both analogue and digital multiplexers are controlled by four common control lines (although only
three are connected to the digital multiplexer) which are fed via resistor and zener diode filters
under the control of the Microprocessor PCB via the SIB PCB and this allows the Microprocessor
PCB to read any digital or analogue signals.
In addition to the four control lines, three further digital lines, buffered in a similar way to the control
lines are provided to control the gain of the analogue output stage, referred to earlier.

8. 17

8.2.3 Serial EEPROM


The non-volatile serial EEPROM, (Electronically Erasable Programmable Read Only Memory)
forming part of the Microprocessor PCB memory, contains factory set information relating to the
calibration, such as calibration data points, span and zero constants, temperature and pressure
compensation factors and filter temperature coefficients. The stored data is unique to each
instrument. An exact backup copy is stored in a second EEPROM fitted to the Transmitter PCB. If
data in the microprocessor EEPROM becomes corrupted, data can be downloaded to the
Microprocessor PCB by using the ASSOCIATE command, available from the keypad.
8.2.4 Autocal Latches
There are four latches on the Transmitter PCB. The control lines are each connected to the input
of one latch. The latches are clocked by a dedicated signal via the SIB PCB. The outputs from the
latches are fed to drivers for solenoid valves, mounted on the Power Control PCB, used by
Autocal. This enables the Microprocessor PCB to energise any combination of valves as required.
If failure or disconnection of the Microprocessor occurs while Autocal is running it could leave
calibration gases permanently connected so an automatic time out is provided. This circuit clears
the latches after 4 seconds if activity on the control line ceases.
8.2.5

Switch Settings
If the Transmitter PCB is replaced ensure that PCB switch SW2 settings are set according to the
analyser model:
For models 2500 / 2520 and 2510 set SW2 1 & 2 to OFF
For model 2550 set SW2 1 & 2 to ON

8. 18

Transmitter input signals:


Circuit Reference

PL1 pin

Chopper box temp.


Pressure
Motor current
Source voltage
Source voltage error
Cell sample voltage
C/S temp. S.P.
Cell T.C. fail
Chopper temp. S.P.
Motor lock
Source fail
Sync.pulse
Flow fail
Solenoid valve driver SV3
Solenoid valve driver SV2
Solenoid valve driver SV1

1
5
6
7
8
9
10
11
12
25
26
27
28
30
31
32

Transmitter output signals:

8.3

Circuit Reference

PL3 pin

Detector shield
Detector signal
Detector temp.

1,2,15,16
3,4,
10

SENSOR HEAD INTERFACE BOARD (SIB PCB)


The Sensor Head Interface board (SIB) (02000/934) connects to the Transmitter PCB (sockets
SK2 and SK3) via PL1 and PL2. The SIB is powered by +5VD,+/- 15VA and +24VP received from
the PS PCB. The +5VD and +/- 15VA provides circuitry requirements and +24VP is used to
provide +/-15VIA and +/- 15VIB via DC/DC converters for 2 isolated analogue measurement signal
outputs (mA or V) and also to energise 3 relay contact pair outputs (NO or NC). A simplified board
layout is shown in Fig.8.2.

8. 19

Input voltages are listed below.


Circuit Reference

PL1 row b pin

PL2 pin

+5VD
0VD
+15VA
0VA
-15VA
+24VP
0VP

13,14,15,16
1,2,3,4
-

1C
1B
1A
3A,3B,3C
2A,2B,2C

Connections to analogue signal outputs and relays are made at TB24 and TB23 mounted on the
Transmitter PCB.
Analogue signal output connections are listed below.
Circuit Reference

TB24 pin

1.1+
1.11.2+
1.2-

4
3
2
1

Relay output connections are listed below.


Circuit Reference

TB23 pin

1.3+
1.31.4+
1.41.5+
1.5-

1
2
3
4
5
6

In addition the SIB performs A/D conversion, drives control lines and reads the digital output line
from the Transmitter PCB.
8.4

OPTION PCB=s
When the Option PCB (02000/916) is fitted, it is located in sockets SK4 and SK5. Option PCB=s are
depopulated SIB=s. The 02000/916 retains sufficient circuitry to drive 2 analogue and 2 relay
outputs. Board layouts are shown in Fig.8.2.

8. 20

8.5

MICROPROCESSOR PCB
The microprocessor PCB (02000/912) is plugged into socket SK1 of the Transmitter PCB and is
powered by +5VD.
The PCB carries out all of the measurement and diagnostic functions of the analyser. The PCB
contains all the necessary integrated circuits to support operation of the 68HC001 microprocessor,
around which the Microprocessor PCB is based. Components include RAM, ROM (Flash
EEPROM, for firmware storage) FRAM (for non-volatile storage of user settings etc) and the
required microprocessor housekeeping logic such as an address decoder, and data acknowledge
(DTACK) generator.
Three LEDs DS01, DS02 and DS03 that form part of the circuitry are used and provide a visual
indication of the state of some operational aspects of the PCB. DS01 (green) is hardware driven
and indicates that the microprocessor is running. DS02 and DS03 (red) indicate the results of
memory tests.
The PCB, in addition to controlling analyser operation and storing factory set data, drives the
following peripheral functions:
Real time clock and backup power source
Keypad interface circuit
Front panel FAULT and ALARM LED drivers
Bus Interface to Display, SIB and Option PCB=s
RS232 printer port
RS485 digital link port
Keyswitch input
The PCB is not a field serviceable item and if faulty should be replaced with a known working PCB.
If a replacement PCB is to be fitted, ensure that link P002 is not fitted and switch (S001) settings
are identical to the faulty (original) PCB. SW1/S001, contact 1 to ON (all other contacts are OFF)
is used to configure the PCB for printer use via the RS232 port. Refer to the board layout shown
Fig.8.3.

8.5.1 Watchdog Circuitry


Functions of RAM, ROM and FRAM devices are monitored by a watchdog circuit. The watchdog
circuit has the following functions:
1

It ensures that the microprocessor will be prevented from running if the input from the PS
PCB indicates that the supply voltage is too low for safe operation.

It monitors a timer (on the PCB) which must be re-initialised every 500ms, (approx) by the
software to prevent time out when it will generate a momentary hardware reset. This will
ensure that if the software should crash and consequently fail to re-initialise the timer, the
software will be re-started within 500ms.

8. 21

In the event of a hardware fault which stops the microprocessor, the watchdog circuit will
repeatedly pulse the RESET line and cause LED DS01 to flash. See Section 8.5.3.

8.5.2 Memory Test


The memory test is a software check on RAM, ROM and FRAM and is carried out whenever the
microprocessor is reset, eg. on power-up, and after watchdog time out. Initially LEDs DS02 and
DS03 will be lit. If the RAM test is satisfactory DS02 will be extinguished, and if the ROM test is OK,
DS03 will also be extinguished. During normal operation both LEDs will be unlit.
8.5.3 Diagnostics
The conditions of LEDs DS01,DS02 and DS03, shown below, can be used to determine possible
problems related to the Microprocessor PCB.
RUN
DS01

RAM fail
DS02

ROM fail
DS03

Diagnosis

On

Off

Off

Normal operation

Flashing
Flashing
On
On

On
Off
On
Off

Off
On
Off
On

Probably a Microprocessor
PCB fault caused by failure
of all or one of the RAM,
ROM and FRAM devices.
Blank display screen.

Off

On

On

Probably an external
RESET caused by a PS. PCB
fault or low mains voltage.

Loss of the peripheral functions, listed in Section 8.5, without total loss of analyser operation, could
be caused by failure of control circuits on the Microprocessor PCB, and should be checked by
PCB replacement.
8.6

DISPLAY PCB
The Display PCB (02501/903) is connected to the Transmitter PCB PL2 by ribbon cable which
carries power and high speed address and data bus to the PCB. Voltage requirements by the
PCB are +5VD only.
The display, mounted on the reverse side of the PCB, features a 2 x 16 character vacuum
fluorescent display (VFD) module. Each character is made up from a 5 x 7 array of dots. The VFD
module internally generates the required higher working voltage from the 5V supply.
The PCB is directly coupled to the microprocessor bus, and operates by copying bit patterns from
a dedicated RAM out to the display drivers. The characters are multiplexed to reduce the number
of drivers required, and the whole display is scanned approximately every 20ms. The bit patterns
are created in the RAM by the microprocessor, so the software can construct any character that is

8. 22

required, including the subscript numbers used in chemical formulae, which are not normally seen
on dot matrix displays.
The circuit has a master blanking line which when high, blanks the whole display irrespective of
the RAM contents. This is driven from the Microprocessor PCB, and defaults high when the
microprocessor is first powered up, or is reset. The line can only be taken low by the software,
and this is only occurs when the display RAM has been loaded with a valid bit pattern.
8.6.1 Diagnostics
In the event of an erroneous display, the fault may be on the Display PCB or the Microprocessor
PCB, and it can be impossible to tell without substituting PCB's because their functions are so
closely interrelated.

8.7

Fault

Diagnosis

Display characters lit but message garbled

Microprocessor PCB fault

Whole characters and dots from characters


missing

Display PCB fault

Display not lit after initialisation routines

Display module failure; cracked


display glass

POWER SUPPLY
Regulated DC voltages, derived from the AC mains supply, are provided by a switched mode
power supply (02000925) which plugs into socket SK6. Voltages are routed to plug-in PCBs via
on-board connections and to Power Control and Display PCBs via PL1 and PL2 respectively. The
power supply has internal in-line fuse protection and the inrush current is thermally limited. The
mains input to the power supply, which is filtered, can vary between 85V and 265V without
affecting the power supply output. The LINE/LIVE input is protected by a 1A anti-surge fuse F37.
Output voltages are:
+5V (0.2V) at 1A
15V (0.75V) at 200mA
+24V (1.92V) at 1A

8. 23

PS PCB input and output pin connections are listed below.


Mains input:
Circuit Reference

PL4 pin

SK6 pin

Mains (L)
Mains (N)
Mains (E)

2
3
1

28A,32A
26C,30C
20A,22C,24A

Output voltages are listed below.

8.8

Circuit Reference

PL1 pin

PL3 pin

SK6 pin

+5VD
0VD
+15VA
0VA
-15VA
+24VP
0VP

23,24
19,20,21,22
7
8
9
10
11

7
5,6
8,9
13,14
11,12

2A,2B,2C
4A,4B,4C
17,18
13,14
2,15,16
37,38
33,34,35,36

DETECTOR PCB
The pyroelectric detector used on 2500, 2510 and 2550 analysers consists of a lithium tantalate
pyroelectric element, which is sensitive to any changes in incident radiation. The detector used on
a 2520 analyser is a UV photodiode type.
The detector response to the chopped light from the source is a near triangle wave which is AC
coupled into the first stage amplifier (IC1). This is a non inverting amplifier with a band pass gain,
matched to the chopping frequency and a gain which is variable between 48 and 1000 by adjusting
RV1. The setting of RV1 is fundamental to the calibration of the analyser and should NOT be
adjusted in the field.
The amplified signal is fed into a differentiating amplifier (IC2) with a gain of 1. The differential of a
triangle is a square wave and the result is a "squared" up signal.
Changes in the sample absorption result in changes to the square peaks proportional to
1/absorbance.

8.8.1

Diagnostics
Setup varies depending upon analyser type, but should always be performed using a
ZERO sample.

8. 24

8.8.1.1 Detector Setup for 2500 Series Analyser


Use an oscilloscope to observe the detector waveform on TP2 of the 02500/904 Transmitter PCB
using TP4 as the earth. Trigger the oscilloscope from the falling edge of the signal on TP1 of the
same PCB.
Adjust RV1 on the 02500/905 Detector pre-amp PCB (visible through the top of the detector
assembly) so that the largest of the peaks is set to 3V0 0V2
Note: If the noise level on the output of the calibrated analyser is greater than 1% f.s.d. (on ZERO
sample) then it is permissible to set the largest peak to a maximum of approx. 4V6 to improve the
signal to noise ratio.
Peak amplitude should be measured from the flat region of the baseline immediately before the
rising edge of the peak, to the flat region at the top of the peak.
Check that the waveform is steady, and that the value of the DETECTOR SIG R, in DIAGNOSTICS
should be greater than 160000.
8.8.1.2 Detector Setup for 2510 Series Analyser
Use a D.V.M. and connect the -ve lead to TP1 (0V) on the 02510/901 Adapter PCB and the D.V.M.
+ve lead to TP3 on the same PCB. Adjust RV1 on the 02500/905 Detector pre-amp PCB (visible
through the top of the detector screening box) until the D.V.M. reading (VN2) is 5V 0V1
Connect the D.V.M. +ve lead to TP4 on the 02510/901 Adapter PCB - on the same PCB use SW1
as a coarse adjustment and RV1 as a fine adjustment to obtain a D.V.M. reading (VDiff) of 0V5
0V1
Connect the D.V.M +ve lead to TP5 on the 02510/901 Adapter PCB and check that the D.V.M.
reading (VGas) is within 0V1 of the value of VN2.
8.8.1.3 Detector Setup for 2550 Series Analyser
Use an oscilloscope to observe the detector waveform on TP2 of the 02500/904 Transmitter PCB,
using TP4 as the earth. Trigger the oscilloscope from the falling edge of the signal on TP1 of the
same PCB.
Up to six peaks will be observed dependant upon the application.
Adjust RV1 on the 02500/905 Detector pre-amp PCB (visible through the top of the detector
assembly) so that the largest peak is optimised at 3V0 0V2.
Peak amplitude should be measured from the flat region of the baseline immediately before the
rising edge of the peak, to the flat region at the top of the peak.
Check that the waveform is steady, and that the value of the DETECTOR SIG R1 in
DIAGNOSTICS should be greater than 160000.
8. 25

8.9

MOTOR CONNECTION PCB


The Motor Connection PCB (02500/906) forms part of the chopper box and provides
interconnections for the chopper motor drive voltages and monitors chopper box temperature, all
under supervision of the Power Control PCB. A thermal fuse mounted on the board will fail if the
chopper box temperature exceeds 95 degrees Celsius. Normal operating temperature is approx.
65EC.
The opto-interrupter (IC1), in conjunction with the filter disc assembly provides the pulse stream
data which is fed to the analogue multiplexer on the Transmitter PCB. The periphery of the filter
disc assembly is slotted, with the slots being detected by the photo-interrupter. The three filter disc
assembly has thirteen slots, twelve of which are used to control the six samples, (one to start each
sample, one to end it) and a thirteenth, wider slot, which is detected by a discriminator circuit and
serves to identify the first filter. Pulse stream data can be monitored with an oscilloscope, at TP3
(wrt 0V) on the Transmitter PCB.

8. 26

SECTION 9: PARTS LISTS


LIST OF CONTENTS
Section

Page

9.1

INTRODUCTION...........................................................................................................................9.3

9.2

PARTS LISTS................................................................................................................................9.3
9.2.1
9.2.2
9.2.3
9.2.4

Source End Assembly .....................................................................................................9.3


Chopper Box ....................................................................................................................9.4
Detector End Assembly...................................................................................................9.5
Detector Box ....................................................................................................................9.7

9. 1

NOTES

9. 2

9.1

INTRODUCTION
This section lists parts numbers and names for use with disassembly of the analyser.
Spares list and recommended spares requirements can be found in the Section 7 of the 2500 User
Manual, including all window and window sealing spares.
Due to the large variation in sample cell types, other cell part numbers are not given here. Please
contact Servomex or their agents with details of cell type if replacement parts are required.

9.2

PARTS LISTS

9.2.1

Source End Assembly


Refer to Fig. 5.1a and 5.1b.
Item No.

Part No.

5
6
8
10
12*
12*
12*
13
14
15*
15*
15*

S2500989
S2500911A
02500502
02500463
S1200921
S1200922
S1200924
02500471
refer to 9.2.2
S2500981
S2500982A
S2500982B

Parts Name and Description


Gas strut kit
Power Control PCB
Cell thermocouple (300mm long sheath)
End boss
Scrubber - H2O Mix.
Scrubber - H2O/CO2 Mix
Scrubber - H2O/CO2/CO Mix
Casting insert
Source
Cell Servicing Kit (includes S2500982A)
Cell Seals Kit - Viton & PTFE seals
Cell Seals Kit - Chemraz seals

*Application dependent, refer to Analyser specification.

9. 3

9.2.2

Chopper Box
Refer to Fig. 5.3.
Item No.
1
2
3

5
6
7
9
10
12
13
14
15
16
17
19
20
21
22
23

24
25
27
28
29

30
31
32
33
34
38
Note:

9. 4

Parts No.
02500466
02520401
02500407
02520402
S2500931A
S2500931B
S2500931C
S2500943A
S2500943B
S2520933
2259-2543
3941-0555
2323-6297
2475-0941
2112-3014
01201414
Consult Servomex
2259-2543
02500410
02520403
S2500986
Consult Servomex
2356-1083
2321-4059
02500451
2323-6266
1723-8010
S2000511
S2000512
2666-2255
S2500906
2323-6398
2536-0329
S2500933A
S2500933B
S2500933C
S2500933D
S2500933E
2323-6297
2323-6273
02520922A
02520406
2259-1780
02510401
02500468

Parts Name and Description


Chopper box, machined
Chopper box, 2520 analyser
Lid, chopper box, machined
Lid, chopper box, 2520 analyser
Source assembly
Zinc selenide
Barium fluoride
Calcium fluoride
Source (Short wave) CaF2
Source (Short wave) Glass
Source Soft UV CaF2
Screw, M4 x 16
Captive screw retainer
>O= ring, Viton
Adaptor, 5mm male x 4 O/D
Dowel, 4 O/D x 28mm long
Mount, filter
Cut-off filter
Screw, M4 x 16
Mounting plate
Mounting plate, 2520 analyser
Motor assembly
Filter wheel assembly
Plug, brass 3" BSPP (if fitted)
Sealing washer 3" BSP (if fitted)
Blanking plug, M25
>O= ring, Viton
Scrubber sachet (if fitted)
H20 Mix.
H20/CO2 Mix.
H2O/CO2/CO Mix.
Resistor, 10 ohm, 15W
Motor Connection PCB assembly
>O= ring, Viton
Thermal Fuse 98C
Lens assembly
Zinc selenide
Barium fluoride
Calcium fluoride
Crown glass
Germanium
>O= ring, Viton
>O= ring, Viton
Front lens assembly, CaF2, 2520 analyser
Source insulator, 2520 analyser
Screw, M3 x 30
Spacer, Motor Connection PCB (2510 & 2520)
Insulator mounting

>O= rings shown above are only available in Seals Kit Part No. S2500975.
Fixings are only available in Fixings Kit Part No. S2500974.

9.2.3

Detector End Assembly


Refer to Fig. 5.2a and 5.2b.
Item No.

Part No.

Part Name and Description

10
16
17
18
19
21

02500463
02501405
02501903
02501998
S2501999
00608333
00608334
02500418
02500449
S2500904A
02500493
02500415
02500416
* See Note 2
02000934
02000916
S2000925

Cell end boss


Display PCB cover
Display PCB
Keypad
Keypad and Display
Transformer (115/230V)
Transformer (100/200V)
Bracket
Transformer cover
Transmitter PCB
Cage Assembly
Side
Side
Cover
Processor PCB
SIB PCB
Option PCB
Power Supply

34

S2510901A
S2510901B
S2550901

Gas Filter Adaptor PCB (2510 x10 Gain)


Gas Filter Adaptor PCB (2510 x40 Gain)
Multicomponent Adaptor PCB (2550)

35
36

00022919
S2501352

Earth Strap Kit


Window

23
26
28
30

31
32
33

* Note:
1) Adaptor PCB item 34 is fitted to 2510 and 2550 analysers only.
2) Order spare Processor PCB or Firmware according to instrument type (refer to table below).
Instrument

Spare Processor PCB

Spare Firmware

02500C

S2500965A

S2550665FA

02510C

S2510965A

S2510665FA

02520C

S2500965A

S2550665FA

02550C

S2500965A

S2550665FA

9. 5

9.2.4

Detector Box
Refer to Fig. 5.4.
Item No.
1/3

Part No.
02500413
02510402, 02510403
02600413

7
8
9

10
11
12
13
14

15

S2501905A
S2501905B
S2501905C
S2600905
2475-0941
2112-3014
S1200921
S1200922
S1200924
2321-4035
2323-6134
02500412
2323-6297
02500933A
02500933B
02500933C
02500933D
02500933E
02500411

Parts Name and Description


Box detector, screening 2500, 2550
2510 analyser
2520 analyser
Detector PCB Assembly :
CaF2 window
KRS - 5 window
CVD-FET /CaF2 /SW
UV Photo diode
Adaptor,5mm male x 4 O/D (if fitted)
Dowel pin 4 O/D x 28mm long (if fitted)
Scrubber Assembly (if fitted)
H2O Mix
H2O/CO2 Mix.
H2O/CO2/CO Mix.
Bonded seal
>O= ring Viton
Block, detector mounting 2500, 2550
>O= ring, Viton
Lens assembly
Zinc selenide
Barium fluoride
Calcium fluoride
Crown glass
Germanium
Boss, detector mounting 2500, 2550

Note that >O= rings shown above are only available in Seals Kit S2500975.
Note: Purge certified analysers do not have adaptor and dowel (items 7 & 8) or scrubber (item 9)
fitted.

9. 6

SECTION 10 : ENGINEERING DRAWINGS

Drawing Number
02501/171
02500/834
02500/835

Description
Power Distribution Circuit Diagram
Cell Assemblies 1 to 64mm Path Lengths
Cell Assemblies 128 to 1000mm Path Lengths

The engineering drawings in this manual are intended for guidance to component location only.
For component part numbers refer to the relevant section of this manual.
Servicing of faulty circuit boards is carried out on a PCB replacement basis.

10. 1

NOTES

10. 2

1701 SOUTH SUTRO TERRACE


CARSON CITY, NV. 89706
TELEPHONE (800) 993-9309
(775) 883-2500
FAX: 775-883-6388

MODEL 3040
SINGLE CHANNEL
SAMPLE COOLER

See us on the WEB at http://www.universalanalyzers.com


e-mail address: sales@universalanalyzers.com
MAN3040 Rev F

1701 SOUTH SUTRO TERRACE


CARSON CITY, NV. 89706
TELEPHONE (800) 993-9309
(775) 883-2500
FAX: 775-883-6388

LIMITED WARRANTY
ALL PRODUCTS MANUFACTURED BY UNIVERSAL ANALYZERS INC. ARE
WARRANTED TO BE FREE OF MANUFACTURING DEFECTS FOR A PERIOD OF ONE
YEAR FROM THE DATE OF RECEIPT AT THE CUSTOMERS RECEIVING AREA AND
FOR AN ADDITIONAL PERIOD OF UP TO 90 DAYS IF THE PRODUCT IS PLACED IN
SERVICE AFTER BEING IN STORAGE. THIS WARRANTY COVERS MATERIALS AND
LABOR TO RESTORE ANY PRODUCTS TO ORIGINAL FACTORY SPECIFICATIONS IF
A DEFECT IS FOUND WITHIN THE WARRANTY PERIOD.
THE DEFECTIVE PRODUCT SHOULD BE SENT, FREIGHT PREPAID, TO THE
FACTORY IN CARSON CITY, NEVADA. REPAIRS WILL BE PERFORMED AT THE
FACTORY AND RETURNED, PREPAID, BY THE SAME SHIPPING METHOD USED TO
SEND THE PRODUCT TO THE FACTORY.
THIS WARRANTY DOES NOT APPLY WHERE THE EQUIPMENT HAS SUSTAINED
DAMAGE DUE TO NEGLECT, MODIFICATION, CORROSION, OR OTHER REASON
BEYOND THE SCOPE OF THE NORMAL DEFINITION OF MANUFACTURING
DEFECT.
FURTHER, THIS WARRANTY IS LIMITED TO REPLACING THE DEFECTIVE
COMPONENTS AND RETURNING THE EQUIPMENT MANUFACTURED BY
UNIVERSAL ANALYZERS INC. TO THE CUSTOMER IN WORKING CONDITION. ANY
OTHER CLAIMS ARE OUTSIDE THE SCOPE OF THIS WARRANTY. NO WARRANTIES
ARE MADE AS TO THE SUITABILITY OF THE USE OF THE EQUIPMENT IN ANY
PARTICULAR APPLICATION OR LOCATION. THE SUITABILITY OF THE USE OF THE
EQUIPMENT IS THE RESPONSIBILITY OF THE CUSTOMER AND THE INSTALLING
CONTRACTOR.

Universal Analyzers Inc.

1701 SOUTH SUTRO TERRACE


TELEPHONE: (775) 883-2500

CARSON CITY, NV 89706


FAX: (775) 883-6388

UNIVERSAL ANALYZERS MODEL 3040 SAMPLE COOLER


SPECIFICATIONS
SAMPLE FLOW RATE:

0 to 6 L/M (at STP)

MAXIMUM INLET TEMPERATURE:


STAINLESS STEEL HEAT EXCHANGER:
KYNAR/GLASS HEAT EXCHANGER:

700o F. (371o C.)


280o F. (138o C.)

MAXIMUM INLET GAS DEW POINT:

180o F. (82o C.)

MAXIMUM INLET WATER CONCENTRATION:

50%*

MINIMUM AMBIENT TEMPERATURE:

32o F. (0o C.)

MAXIMUM AMBIENT TEMPERATURE:

105o F. (41o C.)*

MAXIMUM COOLING POWER, (SECOND STAGE):

126 BTU'S/Hr. (120 kJ/Hr.)

OUTLET SAMPLE DEW POINT:

41o F. (5o C.), adjustable

MAXIMUM INPUT POWER:

350 WATTS

VOLTAGE:

95-125 VAC, 50/60 Hz or


190-250 VAC, 50/60 Hz

ELECTRICAL CLASSIFICATION:

NEMA 1
Class I, Grps. A, B, C, & D,
Division 2 (FM Approved)

DIMENSIONS:

18" x 7 1/2" x 11", HWD

WEIGHT:

33 LB's., (15 KG)

SOLUBLE GAS REMOVAL RATES:

NO
NO2
SO2
CO
CO2

*at reduced flow rates. See capacity chart.

3040 Manual Rev A

0% LOSS
<10% LOSS
< 2% LOSS
0% LOSS
< 2% LOSS

MODEL 3040 SAMPLE COOLER


OPERATING INSTRUCTIONS
APPLICATION
In order to analyze combustion products or incinerator effluents utilizing a direct
extractive sampling technique, it is important to remove the water vapor without
removing the water soluble fraction(s) from the gas sample. The heat exchangers
(impingers) used in the Universal Analyzers gas sample coolers are designed to minimize
the gas/condensate area and time of contact to reduce to a minimum, the amount of mass
transfer of those water soluble components from the gas phase into the liquid phase. The
result is a dry gas sample which has the same composition on a dry basis before and after
passing through the chiller.
A gas sample is usually taken from a stack with a probe extending into the center of the
stack mounted onto or adjacent to a heated filter. The heated filter is maintained at a
temperature above the dew point of the stack gas, usually 300o to 400o F. in order to
avoid cementing the filtered particulates to the filter medium with condensate. A means
is usually provided to automatically blow the particulates trapped by the filter, back into
the stack on a periodic basis.
The stack gas sample is clean but "wet" after passing through the filter assembly. The
moisture in the gas sample comes from the fuel as a product of combustion, from the
humidity in the air which supports the combustion and from the water content which was
trapped in the fuel. This latter source of water in the sample can be from burning moist
coal, wet garbage, or from water injected into the fire box. Water from all of these
sources will remove the water soluble gasses from the sample stream if allowed to
condense in the sample line prior to the controlled separation within the Universal
Analyzers' heat exchangers in the sample cooler.
In order to maintain the temperature of the gas sample above the dew point as it is
transported to the gas sample cooler, a heat traced sample line is usually employed. The
heat traced sample line can be very short if the gas sample cooler is located close to the
heated stack filter or it could be several hundred feet long if the gas sample cooler is
located in the analyzer shack some distance from the stack.
The Universal Analyzers gas sample cooler contains the special impinger type heat
exchanger(s). These are mounted within heat transfer blocks which are cooled by
thermoelectric elements utilizing the "Peltier Effect" discovered in France over half a
century ago. Where high water contents are encountered, it is efficient to remove the
condensate in two stages, one at the temperature of the air in the vicinity of the
"Precooler" and then by passing the sample into a heat exchanger cooled by the
thermoelectric elements. The precooler can remove water which will condense at the
temperature of the environment. In high water content samples, this could be as much as
80% of the water in the sample.

The thermoelectrically cooled stage is temperature controlled at a factory setting of 5o C.


This temperature can be adjusted with a set point potentiometer easily as described
below.
The gas sample conditioning system should contain additional components to insure that
a clean, dry sample is presented to the analyzer panel for minimum analyzer maintenance.
A sensor should be provided to sense the presence of condensate, should any exist in the
tubing following the chiller. This is the optional WCO (Water Carry-Over) sensor which
can be provided with a Universal Analyzers chiller. A visible coalescing filter which
collects particulates on the outside of the cylindrical filter, surrounded by a transparent
bowl will allow the operator to inspect the condition of the heated stack filter. The WCO
sensor can be provided with the sensing elements in the bottom of the filter bowl to
provide an early warning if the coalescing filter removes liquid from the sample stream.
This combination filter/moisture sensor is called the "WCOF".
The location of the sample pump (usually an oil-less diaphragm pump) within the
sampling system is the subject of much debate. If the pump is located upstream from the
chiller, it should have a heated head to avoid the presence of a two phase mixture which
shortens the life of a diaphragm and causes maintenance problems. It can be shown that
passing the gas sample through the chiller under a slight positive pressure will result in a
lower gas dew point than if the gas were at a slight vacuum. A common location for the
sample pump is in the sample line after the chiller. This allows the sample pump to
handle a cool, dry sample with much reduce maintenance problems. Some design
engineers have taken the approach to use a sample cooler having two chilled heat
exchangers. The pump is placed in the sample line between the two heat exchangers.
The first chilled heat exchanger takes enough of the water vapor from the sample to
protect the pump. The second heat exchanger is under a slight pressure and the dew point
is at it's minimum because it is on the discharge side of the pump. The sample pump
location within the sample system is a matter of choice and good engineering.
The condensate must be removed from the heat exchanger(s) using one of the following
methods:
1. A continuously running peristaltic tubing pump can be used
with the heat exchanger either under pressure or vacuum. This is an easy
solution which lends itself to leak testing because of the positive
displacement nature of the peristaltic pump. It is, however, a device which
requires periodic maintenance to replace the tubing.
2. Condensate can be removed from the heat exchanger using an
eductor with a flow limiter to draw some of the hot, wet sample through
the heat exchanger along with the condensate. This method utilizes the
heat exchanger as a bypass condenser and serves the purpose of reducing
the time lag in the sample line without loading the chiller with the water
vapor in the bypassed sample.

3. A float drain trap can be used to collect the condensate running


out of the bottom of the heat exchanger(s) if the heat exchangers are at a
slight positive pressure with respect to the atmosphere. The float drain
trap functions like a steam trap. The float rises to discharge the
condensate when there is enough to lift the float.
4. A small drain tank can be provided with a dual level,
conductivity type level control to collect the condensate. When the tank if
full, the level controller will start a pump to remove the condensate until it
reaches to the lower electrode. When contact is lost, the pump stops and
the condensate is again allowed to fill the tank. Typically, a peristaltic
pump is used to remove the condensate. The advantage is that the pump
only runs occasionally and bypasses no gas sample. The peristaltic pump
requires less maintenance because it does not run continuously.
Finally, a means of controlling the flow of the sample to the analyzers must be
considered. This can be as simple as providing a flow meter with a flow control needle
valve to pinch off the sample flow causing the sample pump to pump higher on the pump
curve. One option which is used is to provide an adjustable back-pressure regulator
between the inlet and outlet of the sample pump to allow a portion of the gas pumped to
be recirculated back to the inlet if discharge pressure exceeds the back-pressure control
point. Some analyzers have their own sample pump which may be sufficient to supply
the analyzer but insufficient to pull the sample through the chiller, sample line, and
heated stack filter. These can be supplied by piping the sample from the external sample
pump into the run of a tee with a flow meter which registers the flow of excess sample
from the branch of the tee to atmosphere. The internal analyzer sample pump can then
withdraw sample from the opposite run of the tee which is essentially at atmospheric
pressure and unaffected by pressure changes within the sample line due to changes in
filter pressure drop or sample pump efficiency.
DESCRIPTION
The key to the success of the Universal Analyzers Sample Cooler being able to condense
the water from a wet gas sample with a minimal loss of the water soluble gas fraction, is
due to the design of the heat exchanger. Please refer to Drawings P0146 thru P0149. The
separation occurs in a classical impinger which has a highly polished cylindrical surface
cooled to the desired dew point temperature. The hot wet sample is brought to the bottom
of the cylinder through an insulated tube and allowed to rise through a narrow annular
area at a relatively high Reynolds number to insure the entire sample is influenced by the
cold surface. The condensate falls down the cold polished surface in the form of a sheet
(as opposed to droplets or the bubbling of the gas sample through the condensate) which
minimizes the surface area in contact with the gas sample.
The temperature of the cylindrical condensation surface of the heat exchanger is
maintained through intimate contact with an aluminum heat transfer block. In the Model

3040 Sample Cooler, the heat transfer block is cooled by the use of two Thermoelectric
Elements at a controlled temperature. The factory temperature setting is 5o C. The
temperature sensor is an AD592 semiconductor device. The controller is a proportional
controller with a proportional band of about 1o C. The set point is field adjustable by
opening the door on the front of the chiller, moving a jumper (J3) with needle nose pliers,
and adjusting a trim-pot to a new temperature value shown on the display. The jumper
needs to be placed in the original position to, again, indicate the temperature of the heat
transfer block.
The heat which is removed from the gas sample (and that which is created by the
Thermoelectric Elements) is discharged by a heat sink which is cooled with a centrifugal
drum blower. The turbulence and velocity provided by the high static pressure capability
of the blower allows for efficient cooling of the heat sinks. The heat sinks are fabricated
from solid blocks of aluminum which eliminates the epoxy joints in more conventional
heat sink designs which are barriers to heat conduction. The result is a heat removal
system with superior performance under all conditions.
Universal Analyzers Thermoelectric Sample Coolers have a digital display as a front
panel indication of the of the operating temperature, (o C.) of the heat exchangers. In
addition, there are two green LED lamps to indicate the status of the cooler. The
"COOL" lamp lights when the operating temperature is between 0o C and 10o. If the
operating temperature is above or below this range, the light will be out and the alarm
relay will be in the alarm position. The "DRY" lamp indicates that there is no water in
contact with the moisture sensor listed as the "WCO" or "WCOF" options. If the "DRY"
lamp is out, the alarm relay will be in the alarm position.
The alarm relay is a single relay activated by either of the above alarm conditions. The
alarm relay is energized in the "safe" condition and will relax into the de-energized, alarm
position if a fault is detected. If a broken thermocouple occurs, the result will be that the
temperature indication is out of range and the alarm relay will indicate a fault.
A thermal switch is mounted on the heat sink to serve as a safety device to turn off power
to the thermoelectric elements if the heat sink approaches 160o F. This protects the
thermoelectric elements from melting in the event high temperature conditions are
encountered for any reason.
The 3000 Series thermoelectric sample coolers are designed to incorporate a peristaltic
pump for condensate removal within the enclosure to reduce the total volume required for
the system installation. Alternatively, it also has provisions for incorporating an eductor,
air filter, and dual heat sample pump with the second head used for generating the
necessary compressed air for the eductor.
Electrical connections are made to external wiring which will enter conduit fittings on the
bottom of the chiller. Internal conduit tubes direct the wires as they are pushed into the

chiller into the vicinity of the terminal strips.


accessible from the front of the chiller.

All of the terminal strips are easily

If it is desired to connect conduit to the chiller, a conduit coupling can be screwed onto
the 1/2" male conduit threads which extend from the bottom.
"PEM Nuts" are provided on the right hand side to mount accessories such as a sample
pump, visible filter, condensate alarm sensor, and flow meter. These options can be user
or factory installed.
The chiller can be changed from 115 VAC to 230 VAC operation with the movement of a
circuit card on the power supply. Accessories such as the sample pump or peristaltic
pump must be rated for the correct supplied line voltage.
INSTALLATION INSTRUCTIONS
Thermoelectric Sample Coolers should be installed away from heat sources in a well
ventilated area of an instrument rack or enclosure. Completely enclosing any instrument
which consumes as much power as a thermoelectric chiller will cause the temperature of
the interior of an unventilated enclosure to rise to too great a level for the sample cooler
to perform reliably. Where it is necessary for the enclosure to be tightly gasketed, a water
cooled thermoelectric chiller should be used. Contact the factory for recommendations.
Air purging an enclosure does not provide enough ventilation to remove the heat which
will be generated internally by the sample chiller. On the other hand, there are air
conditioners and vortex coolers which can provide the necessary cooling for enclosing
thermoelectric chillers.
The 3000 series sample cooler has a flange on the top and bottom to allow it to be wall
mounted or mounted to rails in an instrument rack. Accessories mounted on the side can
be supported by the mounting flange on the cooler.
A 3/8" tubing fitting is provided at the top of the heat exchanger as the sample inlet. The
dry sample outlet from the cooler is the 1/4" Kynar tubing fitting coming out of the top of
the heat exchanger at an angle. The gas sample should be further conditioned with an
additional filter as a safety measure before it enters the analyzer(s). The WCOF visible
filter which incorporates a safety moisture sensor is designed for this purpose. A similar
filter without the moisture sensor is also available from Universal Analyzers. A separate
moisture sensor, the WCO, can also be provided where the user prefers that form factor.
Contact the factory for drawings.
A sample pump is normally required to pull the sample through the sample cooler and to
force it through the visible filter. The moisture sensor, if not a part of the visible filter
should be located ahead of the visible filter.

There must be a method to remove the condensate from the heat exchanger. Several
methods are discussed above. A drain line from the peristaltic pump, eductor, or drain
pot must be run to sewer, a container, or to the ground outside the instrument enclosure to
avoid collecting water (condensate) on the floor.
If an eductor is utilized to remove the condensate, the outlet tube length should be no
longer than two feet in order to keep too much back pressure from the outlet of the
eductor. The outlet tube can be placed in a larger pipe to channel the condensate to a
drain.
The electrical power, about 3 amps at 115 VAC or 1.5 amps at 230 VAC 50/60 HZ
should be supplied. Installation shall be in accord with the manufacture's instructions and
the National Electrical Code (ANSI/NFPA 70). Tampering and replacement with
nonfactory components may adversely affect the safe use of the system. For the 115
VAC case, a power cord is supplied. It can be replaced with conduit wiring easily. See
the electrical installation Drawing P0093. The alarm relay has two sets of "Form C"
contacts brought out to the center terminal strip, TB2. One set are totally dry contacts.
The other set are MOV protected and designed to interrupt power to the sample pump
when in the alarm condition. When the pump is powered through the connector on the
power supply board, the MOV protected contacts are hot, with line power supplied.
The moisture sensor can be connected to the "Viking" connector mounted on the exterior
of the cabinet below the heat exchangers or, alternatively, connected to TB3 on the power
supply board.
START UP PROCEDURE
Apply power to the sample cooler. The indicated temperature will start to drop
immediately. It should be below the over-temperature alarm point in approximately four
minutes and the "COOL" green LED lamp should light. When the temperature reaches
the control set point, the rate at which the temperature drops will be reduced. The
temperature will stabilize within 1o C. of the control set point.
Start the sample gas flow by turning on the sample pump and adjusting the flow control
valve to the desired flow rate. Water should be observed to be removed from the bottom
of each heat exchanger when steady state conditions are established.
If an eductor is utilized to remove the condensate, a strong flow of air should be felt to be
flowing from the eductor outlet tube. The outlet tube length should be no longer than two
feet. The outlet tube can be placed in a larger pipe to channel the condensate to a drain.
The green "DRY" LED light will be on. If an optional moisture is installed, the light will
go out and the alarm relay de-energized when the sensor detects condensate. A test can
be performed by putting water on the moisture sensor. If the sample pump is powered

through the alarm relay contacts, it will cease running. When the moisture sensor is
wiped dry, the sample pump will start.
Turn on the analyzer(s) and initiate the calibration cycle.
TROUBLE SHOOTING
The presence of water in liquid form after the sample cooler is an indication of a fault in
the system. Reasons for the presence of condensate in the system after the sample cooler
could be one or more of the following:
1. Overloading of the cooling capacity of the cooler due to too
much water vapor in the sample or due to too great a sample flow rate.
2. The condensate removal equipment (peristaltic pump, eductor,
or drain pot) may be faulty. The heat exchanger(s) may be full of
condensate.
3. An air leak may be in the condensate removal system allowing
air to enter and blow the condensate back into the heat exchanger. (This
assumes the heat exchanger is under a slight vacuum.)
4. The temperature of the air passing through the cooler to cool the
heat sink is too high. This could be due to placement of the cooler in a
tightly sealed box.
5. The sample cooler could have failed.
Dimensional drawings, installation drawings, and schematics are included as part of this
manual. If additional information is required, telephone assistance can be obtained by
calling (775) 883-2500 or FAX request to (775) 883-6388.

D
NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

7-1/2"
5-1/2"

FM

WET SAMPLE GAS


INLET
3/8" TUBE FITTING

APPROVED

11-1/4" 11-3/4"

DRY SAMPLE GAS


OUTLET
1/4" TUBE FITTING

HOLES FOR
1/4" BOLTS
(4 TYP)
(BOLTS SUPPLIED
BY OTHERS)

THERMOELECTRICALLY
COOLED STAGE
EX
HA
US
T

AIR

FL
OW

SOLID STATE
RELAY
DISPLAY
CIRCUIT BOARD

HIGH PRESSURE
BLOWER/S
HEAT SINKS

17"
EX
HA
US
T

18"

EX
HA
US
T

EX
HA
US
T

AIR

AIR

FL
OW

FL
OW

FL
OW

5,
46
5,
57

POWER SUPPLY
CIRCUIT BOARD

AIR

3
1/2" CONDUIT PORT
(2 TYP)

OPTIONAL
SINGLE HEAD
PERISTALTIC PUMP

CONDENSATE DRAIN
LINE

USER ACCESSABLE
TERMINAL BLOCKS

B 04/16/98
A 1/25/94
REV

SINGLE IMPINGER
THERMOELECTRIC
SAMPLE COOLER

FONT/PATENT #/NEW HEAD/RELAY EAM TB


LARGE TEXT/PAT PEND ADDED HWM TB

DATE

MODEL 3040 SAMPLE COOLER


OUTLINE
PART NO.

FOR

UNIVERSAL ANALYZERS INC.

1701 S. Sutro Terrace


Carson City, Nevada 89706 USA

DRAWING NO
HARVEY MITCHELL
P0082
TED BARBEN
DATE
SCALE
SIZE
12/16/93
1:1
C

DRAWN BY
APVD BY

DWN APVD

DESCRIPTION
REVISIONS

SHEET
1 OF 1

13 5/8"
5 1/2"
3 13/16"

WET GAS
SAMPLE INLET
3/8" TUBING
FITTING

SAMPLE COOLER

DRY SAMPLE GAS


TO ANALYZER

18 13/16"
18"

17"
SINGLE HEAD
ADI MINI
DIA-VAC
PUMP

8
,57
65
5,4

PERISTALTIC
PUMP HEADS

B
5,465,57

WATER
CARRY OVER
SENSOR
CONNECTOR

FLOW METER
VISIBLE FILTER
WITH WATER
CARRY OVER
SENSOR

FRONT VIEW
DESCRIPTION

REV DATE

DWN APVD

REVISIONS

3000 SERIES SAMPLE COOLER


ACCESSORY OUTLINE FOR
SYSTEMS WITH PERISTALTIC PUMP

FOR
INSTRUMENT

PART NO.
NOT ISSUED

ISOMETRIC VIEW

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWN BY
APVD BY

DATE
07/07/1998

DRAWING NO

E. Musselman
T. Barben II

P0294
SCALE
1:1

SIZE
E

SHEET
1 OF 1

24V
ADJ

+ 4

+ 3

+ 2

SOLID STATE RELAY


MODEL Z240D10

+ 1

+3

TB3

W
C
O

RED
BLACK

POWER SUPPLY
BOARD

PELTIER CONNECTIONS
RED
BLACK
RED
BLACK
RED
BLACK
RED
BLACK

RED

1
2

MOISTURE
SENSOR
CONNECTOR

-4

- CAUTION TERMINALS
1,2, & 3 ARE
HOT WHEN
SAMPLE
PUMP IS
INSTALLED
TB2

PELTIER
ELEMENTS

A MOV PROT.
L N.C. 1
2
A
R N.O. 3
M
R
E
L
A
Y

#1

N.C.

DRAIN

5
N.O.

#4

BLOWER

6 AMP FUSE

MODELS
3040
3050
3080

JUMPER CARD
115V
230V

12 AMP FUSE
TB1

P HOT L1
O
W COM L2
E
R

SAMPLE

DRY CIRC.

1
2
3

MODEL
3080
BLOWER
BLACK
BLUE
RED
GREEN

BLACK

#2

WHITE

G
B R
BL LAC EEN
AC K
K

#3
AD592

BL
AC
K

IN
DRA
D
RE

DR
AIN
RE
D

K
AC
BL ACK
BL

THERMAL
SAFETY
SWITCH

MOISTURE SENSOR
CONNECTOR
R

PERISTALTIC
DRAIN PUMP

3
FM
APPROVED

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

DISPLAY BOARD

2
(JUMPER)
TEMPERATURE
OR
SET POINT

TEMP
SET

10/24/03

REV

DATE

SSR CONNECTIONS TO 1-BLK / 4-RED


DESCRIPTION
REVISIONS

EM

PART NO.

FOR

UNIVERSAL ANALYZERS INC.

1701 S. Sutro Terrace


Carson City, Nevada 89706 USA

CALIBRATION ADJUST
(FACTORY SET)

DRAWN BY

DRAWING NO

APVD BY

P0093

HARVEY MITCHELL
TED BARBEN
DATE
SCALE
3/17/94
NONE

TEMPERATURE
SENSOR

MODEL 3040 3050 3080


INTERNAL 115VAC WIRING
DIAGRAM

POWER
SUPPLY

SET POINT
ADJUST

TB

DWN APVD

SIZE

SHEET
1 OF 2

24V
ADJ

+ 4

+ 3

+ 2

SOLID STATE RELAY


MODEL Z240D10

+ 1

+3

TB3

W
C
O

RED
BLACK

POWER SUPPLY
BOARD

PELTIER CONNECTIONS
RED
BLACK
RED
BLACK
RED
BLACK
RED
BLACK

RED

1
2

MOISTURE
SENSOR
CONNECTOR

-4

- CAUTION TERMINALS
1,2, & 3 ARE
HOT WHEN
SAMPLE
PUMP IS
INSTALLED
TB2

PELTIER
ELEMENTS

A MOV PROT.
L N.C. 1
2
A
R N.O. 3
M
R
E
L
A
Y

#1

N.C.

DRAIN

5
N.O.

#4

BLOWER

6 AMP FUSE

MODELS
3040
3050
3080

JUMPER CARD
115V
230V

12 AMP FUSE
TB1

P HOT L1
O
W COM L2
E
R

SAMPLE

DRY CIRC.

1
2
3

MODEL
3080
BLOWER
WHITE
BLUE
BLACK
RED
GREEN

#2

WHITE

G
B R
BL LAC EEN
AC K
K

#3
AD592

BL
AC
K

IN
DRA
D
RE

DR
AIN
RE
D

K
AC
BL ACK
BL

THERMAL
SAFETY
SWITCH

MOISTURE SENSOR
CONNECTOR
R

PERISTALTIC
DRAIN PUMP

3
FM
APPROVED

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

DISPLAY BOARD

2
(JUMPER)
TEMPERATURE
OR
SET POINT

TEMP
SET

10/24/03

REV

DATE

SSR CONNECTIONS TO 1-BLK / 4-RED


DESCRIPTION
REVISIONS

EM

PART NO.

FOR

UNIVERSAL ANALYZERS INC.

1701 S. Sutro Terrace


Carson City, Nevada 89706 USA

CALIBRATION ADJUST
(FACTORY SET)

HARVEY MITCHELL
APVD BY
TED BARBEN
DATE
SCALE
3/17/94
NONE
DRAWN BY

TEMPERATURE
SENSOR

MODEL 3040 3050 3080


INTERNAL 230VAC WIRING
DIAGRAM

POWER
SUPPLY

SET POINT
ADJUST

TB

DWN APVD

DRAWING NO

P0093
SIZE

SHEET
2 OF 2

TO
DISPLAY
BOARD

POWER SUPPLY
BOARD

24V
ADJ

+ 4

+ 3

+ 2

+ 1

TB3

W
C
O

RED

1
2

MOISTURE
SENSOR
CONNECTOR

-CAUTIONTERMINALS
1,2 & 3 ARE
HOT WHEN
SAMPLE
PUMP IS
INSTALLED
TB2

R
E
L
A
Y

DRY CIRC.

N.C.

4
5

N.O.

BLACK
WHITE
GREEN
DRY CIRCUIT
TO ALARM
(CONTACT CLOSES
WHEN IN ALARM)

6 AMP FUSE

SAMPLE
DRAIN
BLOWER
JUMPER CARD
115V
230V

12 AMP FUSE
TB1

P HOT L1
O
W COM L2
E
R

SAMPLE PUMP

A MOV PROT.
L N.C. 1
2
A
R N.O. 3
M

1
2

WATER CARRY OVER


FILTER

L 1(HOT)
L 2(COMMON)

WATER CARRY OVER SENSOR


WITH FLOW THRU ADAPTER

TERMINAL BLOCK
WIRED

A 01/05/00 ADD PAGES 2 & 3 TO DRAWING PACKAGE


REV DATE
DESCRIPTION

EAM JK
DWN APVD

REVISIONS

MOISTURE SENSOR
CONNECTOR

OR
EXTERNAL WIRING DIAGRAM
FOR MODELS 3040 3050 3080

OR

PART NO.
NOT ISSUED

FOR
INSTRUMENT

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
H. MITCHELL
P0094
T. BARBEN II
DATE
SCALE
SIZE
03/17/1994
1:1
D

DRAWN BY
APVD BY

SHEET
1 OF 3

TO
DISPLAY
BOARD

POWER SUPPLY
BOARD

24V
ADJ

+ 4

+ 3

+ 2

+ 1

TB3

W
C
O

RED

1
2

MOISTURE
SENSOR
CONNECTOR

-CAUTIONTERMINALS
1,2 & 3 ARE
HOT WHEN
SAMPLE
PUMP IS
INSTALLED
TB2

A MOV PROT.
L N.C. 1
2
A
R N.O. 3
M
R
E
L
A
Y

DRY CIRC.

N.C.

4
5

N.O.

SAMPLE PUMP
SAMPLE

DRY CIRCUIT TO ALARM


(CONTACT CLOSES
WHEN IN ALARM)

DRAIN

BLOWER

6 AMP FUSE

JUMPER CARD
115V
230V

12 AMP FUSE
TB1

P HOT L1
O
W COM L2
E
R

1
2

HOT

WATER CARRY OVER


FILTER

COMMON

WATER CARRY OVER SENSOR


WITH FLOW THRU ADAPTER

SOLID STATE RELAY


MAX 240VAC 25 AMP
1
240V

TERMINAL BLOCK
WIRED
A 01/05/00 SHOW HIGH CURRENT PUMP & SOLID STATE RELAY EAM JK
REV DATE
DWN APVD
DESCRIPTION
REVISIONS

25A

115 / 230 VAC


3+
4

BLACK

MOISTURE SENSOR
CONNECTOR

HOT (CUSTOMER SUPPLIED VOLTAGE)

OR

EXTERNAL WIRING DIAGRAM


FOR 115VAC HIGH CURRENT
SAMPLE PUMP APPLICATIONS

OR

PART NO.
FOR
NOT ISSUED MODELS 3040 3050 3080

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
H. MITCHELL
P0094
T. BARBEN II
DATE
SCALE
SIZE
03/17/1994
1:1
D

DRAWN BY
APVD BY

SHEET
2 OF 3

TO
DISPLAY
BOARD

POWER SUPPLY
BOARD

24V
ADJ

+ 4

+ 3

+ 2

+ 1

TB3

W
C
O

RED

1
2

MOISTURE
SENSOR
CONNECTOR

-CAUTIONTERMINALS
1,2 & 3 ARE
HOT WHEN
SAMPLE
PUMP IS
INSTALLED
TB2

A MOV PROT.
L N.C. 1
2
A
R N.O. 3
M
R
E
L
A
Y

DRY CIRC.

N.C.

4
5

N.O.

SAMPLE PUMP
SAMPLE

DRY CIRCUIT TO ALARM


(CONTACT CLOSES
WHEN IN ALARM)

DRAIN

BLOWER

6 AMP FUSE

JUMPER CARD
115V
230V

12 AMP FUSE
TB1

P HOT L1
O
W COM L2
E
R

1
2

HOT
COMMON

WATER CARRY OVER


FILTER

WATER CARRY OVER SENSOR


WITH FLOW THRU ADAPTER

SOLID STATE RELAY


MAX 240VAC 25 AMP
1
240V

TERMINAL BLOCK
WIRED
2

25A

115 / 230 VAC


3+
4

BLACK

L1 (CUSTOMER
SUPPLIED
VOLTAGE)

SOLID STATE RELAY


MAX 240VAC 25 AMP

1
240V

MOISTURE SENSOR
CONNECTOR

OR

EXTERNAL WIRING DIAGRAM


FOR 230VAC HIGH CURRENT
SAMPLE PUMP APPLICATIONS

OR

PART NO.
FOR
NOT ISSUED MODELS 3040 3050 3080

UNIVERSAL ANALYZERS INC.


2

25A

115 / 230 VAC


3+
4

BLACK

A 01/05/00 SHOW HIGH CURRENT PUMP & SOLID STATE RELAY EAM JK
REV DATE
DWN APVD
DESCRIPTION
REVISIONS

1701 South Sutro Terrace


Carson City, Nevada 89706 USA

L2 (CUSTOMER SUPPLIED VOLTAGE)

DRAWING NO
H. MITCHELL
P0094
T. BARBEN II
DATE
SCALE
SIZE
03/17/1994
1:1
D

DRAWN BY
APVD BY

SHEET
3 OF 3

1/4" NPT
.400

D
.300

1-1/8"

.400

MOUNTING HOLE
FOR 10-32 SCREW
(2 TYP)
FLOW

.950

3"

1/4" NPT
3

1.500
24" CABLE
B 03/10/03 Revised P/N for Item 1 & 2
REV DATE
DESCRIPTION
REVISIONS

PART NO.
6020-0001

PART NO.
5101-0002
5101-0001
5205-0007

FOR
W.C.O.

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
H. MITCHELL
P0006
TED BARBEN II
DATE
SCALE
SIZE
B
10/08/1991
1:1

DRAWN BY
APVD BY

MODEL W.C.O. WATER


CARRY OVER MOISTURE
SENSOR WITH FLOW THRU
ADAPTER

TABLE A
ITEM NAME
1 HOLDER FLOW THRU BODY
2 KYNAR MOISTURE SENSOR ASSY
3 SENSOR LOCK ASSY KIT

RPH JK
DWN APVD

SHEET
1 OF 1

HOLES FOR #10 PANHEAD


MOUNTING SCREWS
(SUPPLIED BY OTHERS)

1/2"

3"

2"

1.65"

EXPLODED FRONT VIEW

3"
2"

2
COMPLETE
FILTER P/N

1"

5/8"
1

1 1/4"

8"

1
FILTER
INLET PORT
1/4" NPT

1
FILTER
OUTLET PORT
1/4" NPT

4980-0039
4980-0058
4980-0005
4980-0009
4980-0059

FILTER PRESSURE DROP RATINGS


ELEMENT
INCHES OF WATER
P/N
5 LPM 10 LPM 20 LPM
SINTERED S.S. 4980-0023
10
14
24
4980-0022
PAPER
8
12
20
4980-0006
TEFLON
4
6
10
4980-0007
CERAMIC
4
6
10
COATED CERAMIC 4980-0018
6
8
12

1 FOR COALESCING (COATED CERAMIC WCOF)


P/N 4980-0059 SAMPLE INLET & OUTLET PORTS
ARE REVERSED.

7
1/8"

08/24/04

REV

DATE

8
7
6
5

2
1
1
1

4
3
2
1
ITEM

MIDDLE
PIN

Revised with New Bowl Design & Ext. Cable


DESCRIPTION
REVISIONS
-

1
1
1
1
QTY DWG NO.

4704-0010
3907-0052
4904-0006
5205-0038

RPH

TB

VIEW "A" - 2x SCALE


VERIFY BEFORE CONNECTING THAT
MIDDLE PIN IS SHORTER THAN
OUTSIDE PIN. IF NOT, CUT MIDDLE PIN
AS SHOWN

10-32X3/8"PH
14" EXT. CABLE
O-RING 2-030
BOWL ASSEMBLY

5110-0012
SEE TABLE
5008-0004
5110-0011
PART NO.

A
LEFT SIDE

RETAINER
FILTER ELEMNT
BRACKET
KYNAR HEAD
DESCRIPTION

FOR
INSTRUMENT

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWN BY

DRAWING NO

APVD BY

P0018

HARVEY MITCHELL
TED BARBEN II
DATE
SCALE
04/14/92
1:1

RIGHT SIDE

PART NO.
SEE TABLE

DWN APVD

VISIBLE GAS FILTER WITH MOISTURE SENSOR


(W.C.O.F.)

ITEM #3 TYPE

SIZE

SHEET
1 OF 1

D
2-5/8"
INSULATED SECTION

TOP VIEW

CHILLED SURFACE

0.25" INSIDE DIAMETER

THERMOCOUPLE
(TYPE "K")
(TYPE "J")
(TYPE "T")

"L"
(SEE TABLE)

CONDENSATE DRAIN
3/8" NPT
(FEMALE)

ANNULAR SPACE .060" WIDE


(GAS TO BE DRIED RISES IN THIS SPACE)

SIDE VIEW

DRY GAS OUTLET


1/8" NPT
(FEMALE)

B
SAMPLE GAS INLET
1/4" NPT
(FEMALE)

1). SEE SHEET 2 FOR ASSEMBLY INSTRUCTIONS.

APPROX 316 S.S.


LENGTH EXCHANGER
"L"
P/N
5200-SK50
5"
5200-SJ50
5200-ST50

10"

5200-SK10
5200-SJ10
5200-ST10

HASTELLOY C276
EXCHANGER
P/N
5200-CK50
5200-CJ50
5200-CT50

ALLOY 20
EXCHANGER
P/N
5200-AK50
5200-AJ50
5200-AT50

5200-CK10
5200-CJ10
5200-CT10

5200-AK10
5200-AJ10
5200-AT10

TITANIUM
UNIVERSAL
EXCHANGER ANALYZERS
P/N
SERIES
5200-TK50
400
5200-TJ50
500
5200-TT50
800
5200-TK10
1000
5200-TJ10
1100
5200-TT10
3000

01/23/97

REV

DATE

SHOW NEW DESIGN W/THERMOCOUPLE BRACKET

HEAT EXCHANGER OUTLINE


PART NO.
SEE TABLE

1701 South Sutro Terrace


Carson City, Nevada 89706 USA

EV MUSSELMAN

HARVEY MITCHELL
DATE
SCALE
12/11/95
1:1

FOR
SEE TABLE

UNIVERSAL ANALYZERS INC.

APVD BY

HWM

DWN APVD

METALLIC TEMPERATURE SENSING

DRAWN BY

EAM

DESCRIPTION
REVISIONS

DRAWING NO

P0146
SIZE

SHEET
1 OF 2

EXPLODED METALLIC HEAT EXCHANGER


THERMOCOUPLE
ASSY.
(TYPE "K")
(TYPE "J")
(TYPE "T")

2-021 VITON
O-RING
P/N 4904-0013
(SEE NOTES)

METALLIC
OUTER TUBE ASSEMBLY

S.S. COMPRESSION FITTING


W/TEFLON FERRULES
P/N 4954-0095

1/16" TEFLON
THERMOCOUPLE SLEEVE

METALLIC
INNER TUBE ASSEMBLY

6-32 X 1/2"
PAN HEAD SCREW

"L"
LENGTH
5"
5"
5"
5"
5"
5"
5"
5"
5"
5"
5"
5"
10"
10"
10"
10"
10"
10"
10"
10"
10"
10"
10"
10"

MATERIAL
DESC.
316 S.S.
"
"
ALLOY 20
"
"
HASTELLOY C276
"
"
TITANIUM
316 S.S.
"
"
ALLOY 20
"
"
HASTELLOY C276
"
"
TITANIUM
"
"

COMPLETE ASSY
P/N
5200-SK50
5200-SJ50
5200-ST50
5200-AK50
5200-AJ50
5200-AT50
5200-CK50
5200-CJ50
5200-CT50
5200-TK50
5200-TJ50
5200-TT50
5200-SK10
5200-SJ10
5200-ST10
5200-AK10
5200-AJ10
5200-AT10
5200-CK10
5200-CJ10
5200-CT10
5200-TK10
5200-TJ10
5200-TT10

COMPLETE ASSEMBLY

SPARE PARTS LIST


OUTER TUBE
INNER TUBE
ASSY P/N
ASSY P/N
5201-0015
5201-0012
"
"
5201-0022
"
"
5201-0020
"
"
5201-0024
"
"
5201-0013
"
"
5201-0023
"
"
5201-0021
"
"
5201-0025
"
"

"
"
"
"
"
"
"
"
5201-0028
"
"
5201-0016
"
"
"
"
"
"
"
"
5201-0029
"
"

TYPE "K" THERMOCOUPLE


ASSEMBLY P/N
1150-0011
X
X
"
X
X
"
X
X
1150-0018
X
X
1150-0011
X
X
"
X
X
"
X
X
1150-0018
X
X

TYPE "J" THERMOCOUPLE


ASSEMBLY P/N
X
1150-0012
X
X
"
X
X
"
X
X
1150-0019
X
X
1150-0012
X
X
"
X
X
"
X
X
1150-0019
X

TYPE "T" THERMOCOUPLE


ASSEMBLY P/N
X
X
1150-0013
X
X
"
X
X
"
X
X
1150-0014
X
X
1150-0013
X
X
"
X
X
"
X
X
1150-0014

2). LIGHTLY LUBRICATE O-RING


WITH SILICONE GREASE BEFORE
ASSEMBLY.

INNER TUBE
MATERIAL IDENTIFIER

1). O-RING IS FACTORY INSTALLED


IN METALLIC OUTER TUBE ASSY.

2-56 X 1/2"
PAN HEAD SCREW

NOTES:
"L"
(SEE TABLE)

OUTER TUBE
MATERIAL IDENTIFIER

01/23/97

SHOW NEW DESIGN W/THERMOCOUPLE BRACKET

EAM HWM

REV

DATE

DESCRIPTION
REVISIONS

DWN APVD

METALLIC TEMPERATURE SENSING


HEAT EXCHANGER ASSEMBLY
PART NO.
SEE TABLE

FOR
400 500 800 1000 1100 3000 SERIES

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
EV MUSSELMAN
P0146
HARVEY MITCHELL
DATE
SCALE
SIZE
B
12/11/95
1:1

DRAWN BY
APVD BY

SHEET
2 OF 2

8
B

6
B

5
VITON O-RING
P/N 4904-0013

INSULATED
INNER TUBE
ASSY

OUTER TUBE
ASSY

.305 I.D.

SAMPLE
GAS
OUTLET
C
1/4"
TUBE

PLUG
1/16" NPT

A
TOP VIEW

ANNULAR SPACE .060" WIDE


(GAS TO BE DRIED RISES IN THIS SPACE)

SAMPLE GAS INLET


3/8" TUBE

SECTION A-A

HEAT
APPROX
LENGTH EXCHANGER OUTER TUBE MATERIAL
P/N
"L"
5"
5200-S050
316 S.S.
5"
5200-T050
TITANIUM
5"
5200-A050
ALLOY 20
5"
5200-C050
HASTELLOY C-276
5"
5200-S05T TEFLON COATED 316 S.S.
10"
5200-S010
316 S.S.
10"
5200-T010
TITANIUM
10"
5200-A010
ALLOY 20
10"
5200-C010
HASTELLOY C-276
10"
5200-S01T TEFLON COATED 316 S.S.

CONDENSATE
DRAIN 3/8"NPT

"L"
(SEE TABLE)

UNIVERSAL
ANALYZERS
SERIES
400/500
400/500
400/500
400/500
400/500
800/1000/1100/3000
800/1000/1100/3000
800/1000/1100/3000
800/1000/1100/3000
800/1000/1100/3000

02/09/99

REVISE P/N TABLE FOR TEFLON COATED

REV

DATE

DESCRIPTION
REVISIONS

1). SEE SHEET 2 FOR ASSEMBLY INSTRUCTIONS.

EAM

RD

DWN APVD

HEAT EXCHANGER METALLIC


NON-TEMPERATURE SENSING
OUTLINE

PART NO.
SEE TABLE

FOR
SEE TABLE

UNIVERSAL ANALYZERS INC.


1701 South SutroTerrace
Carson City, Nevada 897
06 USA

SECTION B-B

DRAWN BY

DRAWING NO

APVD BY

P0147

EV MUSSELMAN
H. MITCHELL
DATE
SCALE
12/11/95
NONE

SIZE

SHEET
1 OF 2

VITON O-RING
#2-021
SEE NOTES

D
OUTER TUBE ASSEMBLY METALLIC

INNER TUBE
MATERIAL
IDENTIFIER

"L"
INNER TUBE ASSEMBLY

OUTER TUBE
MATERIAL
IDENTIFIER

HEAT EXCHANGER
SEPERABLE
INNER
APPROX
P/N
TUBE ASSY
"L"
P/N
LENGTH
5"
5200-S050 5201-0015
5"
5200-T050
"
5"
5200-A050
"
5"
5200-C050
"
5"
5200-S05T 5201-0041
10"
5200-S010 5201-0016
10"
5200-T010
"
10"
5200-A010
"
10"
5200-C010
"
10"
5200-S01T 5201-0042

SPARE PARTS LIST


OUTER
VITON O-RING PLUG
TUBE ASSY #2-021 P/N 1/16"NPT
P/N
P/N
5201-0012
4904-0013 4951-0058
5201-0024
"
"
5201-0022
"
"
5201-0020
"
"
5201-0043
"
"
5201-0013
4904-0013 4951-0058
5201-0025
"
"
5201-0023
"
"
5201-0021
"
"
5201-0044
"
"

NOTES:
1). O-RING IS FACTORY INSTALLED IN
METALLIC OUTER TUBE.
2). LIGHTLY LUBRICATE O-RING WITH
SILICONE GREASE BEFORE ASSEMBLY. B

02/09/99

REV

DATE

REVISE TABLE TO INCLUDE TEFLON PARTS EAM


DESCRIPTION
REVISIONS

FOR
INSTRUMENT

UNIVERSAL ANALYZERS INC.


1701 South SutroTerrace
Carson City, Nevada 8970
6 USA

DRAWN BY

DRAWING NO

APVD BY

P0147

EV MUSSELMAN
H. MITCHELL
DATE
SCALE
12/11/95
NONE

RD

DWN APVD

HEAT EXCHANGER ASSEMBLY


NON-TEMPERATURE SENSING
METALLIC
PART NO.
SEE TABLE

SIZE

SHEET
2 OF 2

CHILLED SURFACE

INSULATED SECTION

0.300" INSIDE DIAMETER

THERMOCOUPLE
(TYPE "J")
(TYPE "K")
(TYPE "T")

TOP VIEW

ANNULAR SPACE .060" WIDE


(GAS TO BE DRIED RISES IN THIS SPACE)

2-5/8"

"L"
(SEE TABLE)

CONDENSATE DRAIN
3/8" NPT
(FEMALE)

SIDE VIEW
DRY GAS OUTLET
1/8" NPT
(FEMALE)

SAMPLE GAS INLET


1/4" NPT
(FEMALE)

1). SEE SHEET 2 FOR ASSEMBLY INSTRUCTIONS.


GLASS KYNAR
LENGTH HEAT EXCHANGER
ASSEMBLY
"L"
P/N

5"

5200-KK50
5200-KJ50
5200-KT50

10"

5200-KK10
5200-KJ10
5200-KT10

15"

5200-KK15
5200-KJ15
5200-KT15

UNIVERSAL
ANALYZERS
SAMPLE COOLER
MODELS
520
530
540
570
1040
1050
1060
1080
1090

C 10/02/01
REV DATE

Add 15" Heat Exchanger Data To Table


DESCRIPTION
REVISIONS

Glass Kynar Temperature Sensing


Heat Exchanger Outline
PART NO.
SEE TABLE

FOR
SEE TABLE

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

FGA2

DRAWN BY
APVD BY

DRAWING NO

E. Mussleman
H. Mitchell

DATE
12/11/95

EM
JK
DWN APVD

P0148
SCALE
1:1

SIZE

SHEET
1 OF 2

EXPLODED GLASS KYNAR HEAT EXCHANGER

S.S. COMPRESSION FITTING


W/TEFLON FERRULES
P/N 4954-0095

THERMOCOUPLE
ASSY.
(TYPE "K")
(TYPE "J")
(TYPE "T")

1/16" THERMOCOUPLE
TEFLON INSULATOR

#2-120 VITON
O-RING
P/N 4904-0004
(SEE NOTE 2)

6-32 X 1/2"
PAN HEAD SCREW

GLASS TUBE

#2-018 VITON
O-RING
P/N 4904-0003
(SEE NOTE 2)

KYNAR INSULATED
SECTION ASSEMBLY

GLASS KYNAR
HEAT EXCHANGER
ASSEMBLY
LENGTH "L"
P/N
5"
5200-KK50
5200-KJ50
5"
5200-KT50
5"
10"
5200-KK10
10"
5200-KJ10
10"
5200-KT10
15"
5200-KK15
15"
5200-KJ15
15"
5200-KT15

KYNAR DRAIN
FITTING
P/N 5110-2002
(SEE NOTE 1)

KYNAR HEAD
P/N 5201-0017
(SEE NOTE 1)
SPARE PARTS LIST
KYNAR
TYPE "T"
TYPE "K"
TYPE "J"
GLASS
INSULATED SECTION
THERMOCOUPLE THERMOCOUPLE THERMOCOUPLE
TUBE
ASSEMBLY
P/N
P/N
P/N
P/N
P/N
1150-0013
1150-0012
5110-2003
1150-0011
5201-0002
"
"
"
"
"
"
"
"
"
"
5110-2004
5201-0001
"
"
"
"
"
"
"
"
"
"
"
"
"
5110-2006
5201-0078
"
"
"
"
"
"
"
"
"
"
"
"
"

NOTES:
1). O-RINGS ARE FACTORY INSTALLED
IN KYNAR HEAD AND KYNAR DRAIN FITTING.
2). LIGHTLY LUBRICATE O-RINGS
WITH SILICONE GREASE BEFORE
ASSEMBLY.

C
REV

10/02/01
DATE

Add 15" Data To Parts List Table


DESCRIPTION
REVISIONS

EM
JK
DWN APVD

Glass Kynar Temperature Sensing Heat Exchanger


Spare Parts List Exploded Assembly
PART NO.
SEE TABLE

7/8" WRENCH
FLATS

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

2-56 X 1/2"
PAN HEAD SCREW

FOR
SAMPLE COOLERS

DRAWN BY

COMPLETE GLASS KYNAR


HEAT EXCHANGER ASSEMBLY

APVD BY

DATE
12/11/95

DRAWING NO

E. Musselman
H. Mitchell

P0148
SCALE
1:1

SIZE

SHEET
2 OF 2

ANNULAR SPACE .060" WIDE


(GAS TO BE DRIED RISES IN THIS SPACE)
INSULATED SECTION

CHILLED SURFACE

PLUG
1/16" NPT

0.300" INSIDE DIAMETER

TOP VIEW

"L"
(SEE TABLE)

CONDENSATE DRAIN
3/8"NPT

SIDE VIEW

DRY GAS OUTLET


1/4" TUBE

B
1). SEE SHEET 2 FOR ASSEMBLY INSTRUCTIONS.
SAMPLE GAS INLET
3/8" TUBE

LENGTH
"L"

5"

10"

15"

UNIVERSAL
QTY OF
GLASS KYNAR HEAT
ANALYZERS
EXCHANGERS PER
EXCHANGER
SAMPLE
COOLER
SAMPLE COOLER
ASSEMBLY P/N
MODELS
520
1
2
530
5200-K050
2
540
570
4
1040
1
1050
2
5200-K010
1060
2
1080
2
1090
4
FGA2
5200-K015
1

10/02/01

REV

DATE

Add 15" Data To Table

EM

DESCRIPTION
REVISIONS

HEAT EXCHANGER GLASS /


KYNAR NON-TEMPERATURE
SENSING OUTLINE
PART NO.
SEE TABLE

FOR
INSTRUMENT

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
EV MUSSELMAN
P0149
H. MITCHELL
DATE
SCALE
SIZE
C
12/11/95
1:1

DRAWN BY
APVD BY

RD

DWN APVD

SHEET
1 OF 2

PLUG
1/6"NPT

EXPLODED GLASS KYNAR HEAT EXCHANGER

#2-120 VITON
O-RING
(SEE NOTE 2)

GLASS TUBE

#2-018 VITON
O-RING
(SEE NOTE 2)

KYNAR INSULATED
SECTION ASSEMBLY

KYNAR HEAD
(SEE NOTE 1)

KYNAR DRAIN
FITTING
(SEE NOTE 1)

C
GLASS KYNAR
HEAT EXCHANGER
ASSEMBLY
LENGTH "L"
P/N
5"
5200-K050
10"
5200-K010
15"
5200-K015

KYNAR
INSULATED
SECTION ASSY
P/N
5110-2003
5110-2004
5110-2006

SPARE PARTS LIST


KYNAR
GLASS
DRAIN
TUBE
FITTING
P/N
P/N
5201-0002
5110-2002
5201-0001
"
"
5201-0078

#2-120
VITON
O-RING
P/N
4904-0004
"
"

#2-018 VITON
KYNAR HEAD
O-RING
P/N
4904-0003
"
"

1/16" NPT PLUG

P/N
5201-0017
"
"

P/N
4951-0066
"
"

7/8" WRENCH
FLATS

COMPLETE GLASS KYNAR


HEAT EXCHANGER ASSEMBLY
NOTES:

10/02/01

REV

DATE

1). O-RINGS ARE FACTORY INSTALLED


IN KYNAR HEAD AND KYNAR DRAIN
FITTINGS

EAM

DESCRIPTION
REVISIONS

PART NO.
SEE TABLE

FOR
INSTRUMENT

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
EV MUSSELMAN
P0149
H. MITCHELL
DATE
SCALE
SIZE
C
12/11/95
1:1

DRAWN BY
APVD BY

RD

DWN APVD

HEAT EXCHANGER GLASS /


KYNAR NON-TEMPERATURE
SENSING SPARE PART
INSTALLATION

2). LIGHTLY LUBRICATE O-RINGS


WITH SILICONE GREASE BEFORE
ASSEMBLY.

Add 15" Data to Table

SHEET
2 OF 2

R13
C4

U2

TB1

C8

C14

R43

U6

C17 C21
R22

R37

R38
R40

U5

U4

CR5

J1

C24

R42

R28

R11

R8
R44
CR2

C11
C7
R20

CR9
CR1
C28

U9

R30

C23

C25
W1

R24
R27
C10
C12

C20

C27

U8

C26

R31

R23
C13

R17
R19

R32

U3

CR7
CR8

U1

CR4 CR3

R33

R16
R7

J3

R10
R14

R15

C
J5

C16

C15

C19

U7

C18

R35

C6

R3

R34
C22
R25

CR6

R5

R12

C3

R45

B
NOTES: UNLESS OTHERWISE SPECIFIED.
ITEM

DWG.
SIZE

DWG NO.

REVISIONS

UNLESS OTHERWISE INDICATED:


ANGLES = $$p 2$$a
MACHINED FINISH 300/OR BETTER
CHAMFER ANGLE = $$p 3$$a
CHAMFER LENGTH = $$p .015"
DIMENSIONS ENCLOSED IN A
ARE MORE CRITICAL THAN APPLICABLE
INDUSTRY STANDARDS.
ALL DIMENSIONS $$p .005" UNLESS
OTHERWISE SPECIFIED.

APV SYM DATE

ASSY OR B/M

CHANGE

TB C 1/16
98 REMOVED R10
TB B 5/24
95 REFDES U8 TO U6
TB A 2/13
94

FINISH
PER SPEC.

REQ'D

THIRD ANGLE PROJECTIONS

SUPERSCEDES
DWG.

K2

K1

R9

DIMENSIONS IN INCHES.

J6

B
Package To Prevent Handling
And Stocking Damage

C9
R18
R21
R2
R1

Q1
R36
R39
C5
Q2

R6

C1

R26

C2

R4

PART NO.

NAME

PART NO.

NAME

3600-0012

PARTS LOCATION

FOR

MODEL 3040/50/80 COOLER/DISPLAY


SCALE

FULL

DATE

2/13/94

INTERNATIONAL

UNIVERSAL ANALYZERS INC.


SIZE

B
SHEET

DRAWING NO.

E0030

REV.

DWN. CHKD.APVD.ENGR.

MRM

TB TB TB

CONT. ON.

10

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

G
+VREF
R3
10K

+10V

R6

R7
49.9K

TEMP
SET

2K
27L2
7
U3
5 +
4
6

R9
1K
20T

8
U3

R45

R5

2.49K

1K

1
27L2

CR9

J5
1

C7 +15V
.1
MLC
R8
24.9K
R11

CR2

.47uf, POLYE
R12

C28
.1
MLC

C6
.1uf
MLC

J1
1

TP GND
J1
2

+VREF

SET

R16

10K

10K

-VREF
TEMP
CAL
+15V

R20

10K
J3
R21

J1
4

COM

1K
20T
R17
8.66K
R19

J1
3
J1
5

+15V

R23
4.99K
C12
.1uf
MLC

C10

R24
4.99K
C13
.1uf
MLC

100K

8
U4

1
U4
4 27L2
R27
10K

+10mv/ C

-VREF

R31

10K

1 MEG
3

10K

U5
4 27L2

R37

R35

15K

1K

+15V
DRY

8
U5

7
27L2

11
12
13

14
U6

10
4025

8
2
1

U6

R42

R36
24.9K

Q2

WET/DRY

J5
6

100K

U6

C20
.1uf
MLC

7 4025

J5
3

+10V
C19
.1uf
MLC

CR6
LM385 -2.5

+VREF
(+5V)
-VREF
(+2.5V)

C22
.1uf
MLC

W1

+5V

8 2

C26
15uf
TANT

C15
15uf
TANT

B
REV DATE

CHANGE

06/07
99

REV DATE

1
6
7

15uf
TANT

U9
7660S

4
5

-5V
C27
15uf
TANT

NAME:3000 SERIES LOW TEMP


SCHEMATIC
CONTROLLER/DISPLAY
FOR
PART NO. 3600-0033

UNIVERSAL DATE: 2/12/98


ANALYZERS
C E0061
INC.

RELAY
DRIVE

C25

3). ALL DIODES ARE 1N4148.


2). ALL RESISTORS ARE IN OHMS.
1). ALL RESISTORS ARE 1/4W, 1%, 50ppm/ C.
NOTES: UNLESS OTHERWISE SPECIFIED.
CR5

R40
499ohm

R39
100K

U7

C18
.1uf
MLC

2N4401

4025

5
4
3

J5
2

1 MEG
CR4
GRN

78L05

R33
2K

R41
2K

COM

J5
5

C17
.1uf
MLC

CR3
GRN

COOL

+15V

+15V
6

+15V

HOT=HIGH
COOL=LOW

DISPLAY

+15V
C16
15uf
TANT

C21
.1uf
MLC

FROZEN=HIGH
OK=LOW

U1
3 1/2
DIGIT
LCD

J5
4

TEST
C14
.01uf

R38

10

+VREF
+15V

R30

R34
10K

+VREF

26

+VREF

+VREF R32

32

TEST

21
16
22
18
19
2
3
4
5
6
7
8
9
10
11
12
13
14
15
17
20
23
24
25

BP
B3
G3
OSC2 E3
B4
OSC1 D1
CREF C1
B1
CREF A1
U2 F1
7136 G1
E1
A/Z
D2
C2
B2
BUFF
A2
F2
INT
E2
D3
F3
IN+
INPOL
REF+
A3
REFC3
TEST G2

1
G

16
12
8
28
38
39
40
1
29
32
9
3
18
19
20
21
22
23
17
14
15
24
25
26
13
10
31
2
30
11
27

Q1

-100mv/ C

+VREF

J6

39

C2
33
.47uf POLYE
C4
29
.47uf POLYE
R13
28
1.74 MEG
C8
27
.15uf POLYP
31
30
+VREF
36
35
R26
37
10K
R43
15K

R1

2N4401
R2
100K

1
38 OSC3

549K

1 MEG

+10V

+VREF

+VREF +VREF

-VREF

C9
.01uf

47pf
R4

27L2

R28
10K

.47uf
POLYE
2

1 MEG

+VREF
0 C=0V
25 C=+.25V
w.r.t. +VREF

.1uf, MLC
R25

R18

TEMP

C11

100K
C24
.1uf
MLC

TEMP
SENSOR
-

R22
4.99K

R15

+10V

APPROVED

P/S
CONTROL

3
R

40
34

24.9

1 MEG
R14
20K

FM

C1

R44
10K

CR1

C3

C5 -5V
.1uf
MLC

+VREF

+5V

TP PS

+10V

+10mv/ C

DELETE C23
CHANGE

SHEET 1 OF

DWN MRM
CHK TB
APV TB
ENG TB

CR17
CR15

R1

C33
R34
R26

T3

C4

C6

L2

RH1

J4

C2

CR6
CR5
CR8
CR7
CR3
CR1

J3

TB1

F1

TB2

K1

CR22

Q5
J6

R42
R43

TB3

C41

R41

R40

C37

C1

J2
RH7

RH6

RH5

C14

U3

CR4
CR2

C17

T2
RH4

U5

U4

C31

C21

R12
R7

R29
R28
R21

R24
C12

F2

R32
C36 C23
R19
R23

C10

C9

C43

C8

U7

R3

J1

C30

R4

R22

Q4

C35

R8

R38
R13

L1

C13

C29

R30

C11

R31
R39
R11

U6

R6
R37

Q3

R10

C19

R14

C39

R15
R36
C38 C40

C15
J5

C3

RH3

RH2

T1

C25

ITEM

DWG NO.

DWG.
SIZE

PART NO.

REVISIONS

APV SYM DATE

ASSY OR B/M

UNLESS OTHERWISE
INDICATED:
.
ANGLES =_+ 2
MACHINED FINISH 300/OR
BETTER
.
CHAMFER ANGLE_= + 3
CHAMFER LENGTH_= + .015"
DIMENSIONS ENCLOSED IN A
ARE MORE CRITICAL THAN APPLICABLE
INDUSTRY STANDARDS.
ALL DIMENSIONS_ +.005" UNLESS
OTHERWISE SPECIFIED.

NAME

3000 SERIES POWER SUPPLY


CHANGE

FINISH
PER SPEC.

REQ'D

THIRD ANGLE PROJECTIONS

SUPERSCEDES
DWG.

C5

R20

T4

C20

C22

C26

R27

R5

NOTES: UNLESS OTHERWISE SPECIFIED.

DIMENSIONS IN INCHES.

Package To Prevent Handling


And Stocking Damage

R2

R25

CR20
CR19

R16

CR13
CR12

C32
R33
R17

C7

CR10
CR9

CR21

Q1

R18

C24

5
Q2

6
CR11

7
U2

PART NO.

NAME

PARTS LOCATION

3600-0011

FOR

MODEL 3040/50/80 POWER SUPPLY


SCALE

FULL

DATE

2/23/94

INTERNATIONAL

UNIVERSAL ANALYZERS INC.


SIZE

B
SHEET

DRAWING NO.

E0034

REV.

DWN. CHKD.APVD.ENGR.

MRM

CONT. ON.

10

LINE

F1

TB1

4
RH1

L1

12 AMP
250V

NEUT/
LINE 2

C1
.22
250V

R3
422K

C3
4700pf
250V

C5
4700pf
250V

+
AC

CR21
FPIL
60

TB1

115V
230V

5
J4
6
7

RH2
130V
MOV

C4
680
200VDC

0VDC

T1

1
3
1

J1

WHT

1
3

J1

GRN

F2

6 AMP
250V

J1

BLK
DRAIN PUMP
TB2
NC1 1

J2

TB2
2

TB2
NO1 3
TB2
NC2 4

J2

TB2
NO2 6

115V

J4

230V

J4

BLK
J2

115V

J2

CR3

J4

J2
J2

SAMPLE PUMP

5
6
3

RELAY CONTACTS
REQUIRE 24VDC
200mA MAX.
RESISTIVE LOAD
IN CLASS I, GROUP A, B, C, OR D
DIVISION 2 AREAS

J3
J3

BLU

+15V

J3

10K

C37
1000pf
MLC

GRN

3+

240V

3 - 32 VDC

10K

25A

R44

MOISTURE
SENSOR

Q5
2N4401

BLK

PUMP

J2-3

WHT

PUMP

J2-1

GRN

PUMP

SOLID STATE RELAY


(SSR1)
P/N 3152-0002
PUMP CONTROL
3-32VDC CONTROL
25AMP

J6

U7
LM1830N

J5
6

12

DIV 2
SAMPLE PUMP

H
G

5/25
05
4/12
01
9/11
00

13
9

FM

11
C43
15
25V
TANT

ADD SOLID STATE


RELAY & PUMP WIRES E
REVISION PER
D
SHEET 2
ADD GROUP A&B
C
CLASSIFICATION NOTE
CHANGE

WET/DRY

APPROVED

3). ALL RESISTORS ARE IN OHMS.


2). ALL CAPACITORS ARE IN MFD.
1). ALL RESISTORS ARE 1/4W, 1%, 50ppm/oC.
NOTES: UNLESS OTHERWISE SPECIFIED.
SCHEMATIC
FOR

+VREF

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

REV DATE

10

+VREF

10

TB3
2

J2-2

J5
2

7
5

C41
.1
MLC

TB3
1

RELAY MOUNTED ON PANEL NEXT OT POWER SUPPLY BOARD


(SEE DRAWING P0083)

COM

J3

CR22
1N4148

R43
24.9K

J5
5

14

RED

R40
71.5K
OHMS

R42

+15V

BLK
J3

BLOWER

R41
75
1/2W

J5
4

WHT

J6

CR4

C8
15
25V
TANT

6
12
PL5.0-16

C
RELAY
DRIVE

+15V

U2
C7
470
50VDC

10

+15V

J5

7815

N/C

K1

-160VDC

CR2

CR1
4X 1N4005

GRN

RH6
250V
MOV
RH7
250V
MOV

TB2
ARM2 5

WHT

RH4
250V
MOV
RH5
250V
MOV

ARM1

R2
C6
100K
680
1/2W
200VDC

R1
100K
1/2W

RH3
130V
MOV

GND
TB1

+160VDC

CL30

C2
.22
250V

6/7
99
10/15
96
5/16
94

REV DATE

R40 FROM 20K TO


71.5K OHMS
C19 VALUE CHANGE
RELAY RATINGS

NAME:500

WATT POWER SUPPLY

PART NO.3600-0011

UNIVERSAL DATE: 10/4/93


ANALYZERS
C E0024
INC.
SHEET 1 OF

CHANGE

DWNMRM
CHK
APV
ENG

10

G
+15V
2

+VREF

U6
3+

1
R4

1 MEG
C35

F
0VDC
+15VDC
+VREF
+24V

C15

R5
2.15K

+24V R27
ADJ 100
1T

P/S 1
CONTROL

R12
499

R30

Q3

24.9K
R28
24.9K

J5

R29
49.9K

2N4401

.1
50V
MLC
R11
10K
C36
.1
MLC

16
R39
10K
2

R32
24.9K

C9
15
25V
TANT

1
R14
24.9K

C30
15
25V
TANT

C10
15
25V
TANT

C11
.1
MLC

C12
.1
MLC

C13
.1
MLC

1N4148
T2

U3
3525A

50KHz
+15V
11

R21
100

8
2 U4

R24
100

12

CR10

R16

1N4148

24.9

R33
49.9

11T
6
7

C29

8
1
2 U5
5 4

Q1
IRFP451R
CR11A

T3
R20

C32
2200pf
MLC

D.S.
R36

R38

54.9K
C40

.01
MLC

120
5W

C22
470pf
2KV

L2

1+

FEP30DP

8T

R18
D.S.

8T

C24
D.S.

18 uH

35T

3+
CR11B

+24V

FEP30DP

1N4148
6
7

R13

R22

CR17

R25

2.43

1N4148

24.9

R34
49.9

4420

CR19
1N5245
R26
15V
10K
CR20
1N5245
15V

Q2
IRFP451R

1-

2+

2-

C33
2200pf
MLC

C25
4700
35V

3-

4+

C26
4700
35V

4-

-160VDC

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

B
SCHEMATIC
FOR

H
G
F

5/25
05
4/12
01
9/11
00

REVISION PER
SHEET 1
C35 FROM 1000pf TO
47pf
REVISION PER
SHEET 1

REV DATE

TLC3702

15
TANT
25V

CR15

14T

11T

4). D.S. COMPONENTS R15, R18, R38, AND C24


ARE NOT PRESENT.
3). ALL RESISTORS ARE IN OHMS.
2). ALL CAPACITORS ARE IN MFD.
o
1). ALL RESISTORS ARE 1/4W, 1%, 50ppm/ C.
NOTES: UNLESS OTHERWISE SPECIFIED.

CR12
1N5245
R17
15V
10K
CR13
1N5245
15V

1
5 4
4420 50V
POLYE

+15V
14

5+
U6
6-

100K

D.S.
C39

CR9
3

APPROVED

20K
C38
0.1
MLC

+160VDC

FM

10

4.99K
C17
0.1
MLC

+VREF

C14
.1
MLC

R37
15
1/2W

R8

13

8
2.49K

4X 1N4148
CR7
CR8

R7

R15

10

Q4
2N4401

50T

CR5

R6
1K
1/2W

47pf
MLC

CR6

15

C19
330pF
100V
MLC
9
C21
C31
2200pf MLC
1
R19
50V
6
POLYE 11.5K
C23
R10
5
34K
R23
1000pf 5.11
100V
7
MLC

R31

1T

1
400VDC
POLYP

+15VDC

TLC3702

T4
C20

CHANGE

E
D
C

6/7
99
10/15
96
5/16
94

REV DATE

R40 FROM 20K TO


71.5K OHMS
C19 VALUE CHANGE
RELAY RATINGS

NAME:

500 WATT POWER SUPPLY

PART NO. 3600-0011

UNIVERSAL DATE: 10/4/93


ANALYZERS
C E0024
INC.
SHEET 2 OF

CHANGE

DWNMRM
CHK
APV
ENG

22-3/4"
CARRYING HANDLE
15-1/4"

WET GAS
SAMPLE INLET
DRY SAMPLE GAS
TO ANALYZER

3/8" TUBING
FITTING
SAMPLE COOLER

SAMPLE PUMP HEAD

DOUBLE HEAD
ADI MINI DIA-VAC
PUMP
26-1/2"
FLOW METER

VISIBLE FILTER
WITH WATER
CARRY OVER
SENSOR

COMPRESSED AIR HEAD

8
,57
65
5,4

78
5,465,5

EDUCTOR

B
AIR FILTER
WATER CARRY
OVER SENSOR
CONNECTOR
FOUR LEG STAND
(ONE LEG NOT VISIBLE IN THIS VIEW)

ISOMETRIC VIEW

FRONT VIEW

12-3/8"

04/17/98

ADD MOUNTING FEET/PATENT NUMBER

REV

DATE

DESCRIPTION
REVISIONS

EAM

3000 SERIES PORTABLE SAMPLE COOLER


ACCESSORY OUTLINE FOR SYSTEMS WITH
EDUCTOR
PART NO.

FOR

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWN BY
APVD BY

DATE
12/20/93

JK

DWN APVD

DRAWING NO
HARVEY MITCHELL
P0084
TED BARBEN
SCALE
SIZE
D
1:1

SHEET
1 OF 1

22-3/4"
CARRYING HANDLE
13-5/8"
WET GAS
SAMPLE INLET

DRY SAMPLE GAS


TO ANALYZER

3/8" TUBING
FITTING
SAMPLE COOLER

SINGLE HEAD
ADI MINI
DIA-VAC
PUMP

26-1/2"

FLOW METER

VISIBLE FILTER
WITH WATER
CARRY OVER
SENSOR

8
5,465,57

5,4

65

,57

PERISTALTIC
PUMP HEADS

WATER
CARRY OVER
SENSOR
CONNECTOR
FOUR LEG STAND
(ONE LEG NOT VISIBLE IN THIS VIEW)

ISOMETRIC VIEW

FRONT VIEW

12-3/8"

04/23/98

ADD LEG/FEET/NEW HEAD DESIGN/PAT #

REV

DATE

DESCRIPTION
REVISIONS

EAM

3000 SERIES PORTABLE SAMPLE COOLER


ACCESORY OUTLINE FOR SYSTEMS WITH
PERISTALTIC PUMP
PART NO.

FOR

UNIVERSAL ANALYZERS INC.


1701 S. Sutro Terrace
Carson City, Nevada 89706 USA

DRAWN BY
APVD BY

DATE
12/20/93

DRAWING NO
HARVEY MITCHELL
P0085
TED BARBEN
SCALE
SIZE
D
1:1

JK

DWN APVD

SHEET
1 OF 1

4
1

D
*

8
7

8
6

5,4

65

,57

7
5

ASSEMBLED FRONT VIEW

(ACCESSORIES NOT SHOWN FOR CLARITY)

VIEW AS SEEN LOOKING AT BACK OF SAMPLE COOLER


(ACCESSORIES NOT SHOWN FOR CLARITY)

BILL OF MATERIALS FOR


LEG MOUNT KIT
P/N 9515-0019
ITEM
1
2
3
4
5
6
7
8

*
*

QTY
2
1
1
2
3
4
3
3

DESCRIPTION
LEG SINGLE
LEG DUAL
PLUG ROUND PLASTIC
PLUG SQUARE PLASTIC
SCREW 1/4" x 1-1/2" S.S.
SCREW #6 x 1-1/4" S.S.
WASHER 1/4" FLAT S.S.
NUT 1/4" S.S.

A 04/28/98 ADD FOOT PADS REVISE BOM TABLE


DESCRIPTION
REV DATE

EAM JK
DWN APVD

REVISIONS

P/N
6006-0001
6006-0002
4919-0007
4919-0006
4704-0047
4704-0048
4710-0001
4711-0006

LEG MOUNTING KIT


INSTALLATION FOR 3000
SERIES SAMPLE COOLERS

PART NO.
9515-0019

FOR
3000 SERIES COOLERS

UNIVERSAL ANALYZERS INC.

*NOTE: ITEMS 3 AND 4 ARE FACTORY INSTALLED

1701 South Sutro Terrace


Carson City, Nevada 89706 USA

DRAWN BY
APVD BY

DATE
02/04/1994

DRAWING NO

H. MITCHELL
T. BARBEN II

P0089
SCALE
NTS

SIZE
E

SHEET
1 OF 1

INSTALLATION:
1). INSTALL ITEMS #1 INTO BOTTOM OF
BOTH HEAT EXCHANGERS.

2). INSTALL ITEMS #2 & #3 INTO ITEMS #1


WITH ITEM #2 IN THE FRONT HEAT
EXCHANGER.
3). INSTALL ITEMS #4 INTO TOP OF LIQUID
DRAINS TYPICAL.

FRONT

4). INSTALL ITEMS #5 ONTO ITEMS #2 & #3.

PARTS TABLE
ITEM#
DESCRIPTION
QTY
P/N
1
1/2" TUBE X 3/8"NPT CONNECTOR 2 4954-0058
2
1/2" 316 S.S. X 5" BENT TUBE
1 9011-0004
3
1/2" 316 S.S. X 3" STRAIGHT TUBE
1 9011-0004
4
1/2" TUBE X 3/4"NPT CONNECTOR 2 4954-0135
5
FLOAT DRAIN VALVE
2 4955-0001

4
3

5
DESCRIPTION

REV DATE

DWN APVD

REVISIONS

FLOAT DRAIN INSTALLATION


LAYOUT

PART NO.
9515-0100

UNIVERSAL ANALYZERS INC.

RIGHT SIDE VIEW

1701 South Sutro Terrace


Carson City, Nevada 89706 USA

DRAWN BY
APVD BY

DRAWING NO

E. Musselman
T. Barben II

DATE
09/15/1999

FOR
3020, 3050, 3080

P0360
SCALE
NTS

SIZE
E

SHEET
1 OF 1

Universal Analyzers Inc.

1701 South Sutro Terrace Carson City, NV 89706


Telephone (775) 883-2500 (800) 993-9309
Fax (775) 883-6388

SPARE PARTS RECOMMENDATIONS FOR 3000 SERIES SAMPLE COOLERS


Level A, Consumable Parts (All Models):
9216-0002
Peristaltic Pump Tubing, #15, 5 ft. length
3010-0005
6 amp Slow Blow Fuse for 3000 Series Power Supply
3010-0006
12 amp Fast Acting Fuse for 3000 Series Power Supply

2 Yr Req.
9
2
2

Level B, Basic Parts (3000SS Series):


5200-S010
Heat Exchanger/Impinger 316S.S. 10 Inch
4904-0013
O-Ring, 316S.S. Heat Exchanger Viton 2-021
8010-0001
Paste, Heat Sinking - 0.1 Ounce Container
9515-0040
Ambient Heat Exchanger Conversion Kit
(Converts the Model 3040SS to a Model 3050SS)

2
2

Level B, Basic Parts (3000PV Series):


5200-K010 Heat Exchanger/Impinger Glass/Kynar 10 Inch
5201-0001
Glass Tube, Outer Heat Exchanger Replacement 10 Inch
4904-0003
O-Ring, Glass/Kynar Heat Exchanger Viton 2-018
4904-0004
O-Ring, Glass/Kynar Heat Exchanger Viton 2-120
8010-0001
Paste, Heat Sinking - 0.1 Ounce Container
9515-0041
Ambient Heat Exchanger Conversion Kit
(Converts the Model 3040PV to a Model 3050PV)

1
2
2
2

Level C, Critical Parts (All Models):


3016-0002
Peltier Element, Pair - 15Vdc 8.5 Amp 40mm Sq.
9515-0042
Insulation Kit - Heat Transfer Block (Model 3040 & 3050)
9515-0023
Insulation Kit - Heat Transfer Block (Model 3080)
1150-0017
Temperature Sensor Assembly, AD592
3103-0006
Temperature Limit Switch, 185F
3152-0004
Solid State Relay, 3-32 VDC / 240VAC 10Amp (Blower Motor)
4800-0010
Blower & Motor Squirrel Cage 115/230Vac

1
1
1

Level D, In-Depth Parts (All Models):


3600-0012
Controller Circuit Board
3600-0011
Power Supply Board - 15Vdc 500 Watt
4958-0028
Peristaltic Pump Motor, 115Vac 6RPM
4958-0031
Peristaltic Pump Motor, 230Vac 6RPM
4958-0006
Peristaltic Pump Head, #15
Optional Parts/Accessories:
4958-0025
Sample Pump - 115Vac Mini Dia-Vac Alum/Teflon Single Head
4958-0015
Sample Pump - 230Vac Mini Dia-Vac Alum/Teflon Single Head
4958-0026
Sample Pump - 115Vac Mini Dia-Vac Alum/Teflon Dual Head
4958-0016
Sample Pump - 230Vac Mini Dia-Vac Alum/Teflon Dual Head
9515-0018
Sample Pump Rebuild Kit - Mini Dia-Vac (for Single Head)
9515-0018
Sample Pump Rebuild Kit - Mini Dia-Vac (for Dual Head)
4955-0004
Sample Pressure Control, AirTrol 0-10Psig Adjustable
4955-0091
Sample Pressure Control, Swagelok 316SS 10-225Psig Adjustable
3000 Spare Parts Rev D

3
6
2
2

Optional Parts/Accessories, Continued:


3152-0002
Solid State Relay, 3-32 VDC / 240VAC 25Amp (Sample Pump)
4980-0005
WCOF Assembly - Moisture Sensor/2m Teflon Filter
5205-0006
Bowl, WCOF Filter - Replacement with Cable & O-Ring
4904-0006
O-Ring, WCOF Bowl Viton 2-030
4980-0006* Filter Element - 2m Teflon (WCOF)
6020-0001
WCO Moisture Sensor Assembly w/ Flow Thru Adaptor
5101-0001
WCO Moisture Sensor (for Non-Filtered Flow Thru Style)
9515-0046
Thermocouple Kit, Heat Exchanger - New Jersey Type K
9515-0047
Thermocouple Kit, Heat Exchanger - New Jersey Type J
9515-0048
Thermocouple Kit, Heat Exchanger - New Jersey Type T
9515-0100
Condensate Drain Kit - Automatic (Models 3050 & 3080)
9515-0101
Condensate Drain Kit - Automatic (Models 3040)
9515-0019
Portable Leg Kit (3000 Series)
* Commissioning Spare Part

2 Yr Req.
1
2
6

UNIVERSAL ANALYZERS INC.


MOISTURE CONVERSION TABLE
DEW POINT,
DEGREES C.
-100
-50
-20
-10
-5
-4
-3
-2
1
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90

% WATER BY VOLUME % WATER BY WEIGHT


AT SATURATION
AT SATURATION
0.00000139
0.00000081
0.00388
0.00241
0.102
0.0633
0.256
0.1596
0.396
0.229
0.432
0.268
0.469
0.291
0.510
0.317
0.555
0.345
0.602
0.364
0.649
0.404
0.696
0.433
0.750
0.466
0.803
0.499
0.861
0.535
0.922
0.573
0.988
0.614
1.06
0.658
1.13
0.702
1.21
0.753
1.29
0.802
1.38
0.860
1.48
0.920
1.58
0.980
1.68
1.044
2.31
1.433
3.13
2.004
4.19
2.64
5.55
3.54
7.28
4.67
9.46
6.12
12.2
8.0
15.5
10.3
19.7
13.3
24.4
16.8
30.7
21.7
38.0
27.8
46.7
35.6
57.2
45.7
69.2
58.4

WATER VAPOR
PRESSURE, mm. Hg
0.0000099
0.0295
0.776
1.950
3.014
3.280
3.569
3.880
4.223
4.579
4.937
5.294
5.70
6.10
6.54
7.01
7.51
8.05
8.58
9.15
9.8
10.5
11.2
12.0
12.8
17.5
23.8

1701 SOUTH SUTRO TERRACE


CARSON CITY, NV. 89706
TELEPHONE (800) 993-9309
(775) 883-2500
FAX: 775-883-6388

MODEL 3050
SINGLE SAMPLE
DUAL STAGE
SAMPLE COOLER

See us on the WEB at http://www.universalanalyzers.com


e-mail address: sales@universalanalyzers.com
MAN3050 Rev D

1701 SOUTH SUTRO TERRACE


CARSON CITY, NV. 89706
TELEPHONE (800) 993-9309
(775) 883-2500
FAX: 775-883-6388

LIMITED WARRANTY
ALL PRODUCTS MANUFACTURED BY UNIVERSAL ANALYZERS INC. ARE
WARRANTED TO BE FREE OF MANUFACTURING DEFECTS FOR A PERIOD OF ONE
YEAR FROM THE DATE OF RECEIPT AT THE CUSTOMERS RECEIVING AREA AND
FOR AN ADDITIONAL PERIOD OF UP TO 90 DAYS IF THE PRODUCT IS PLACED IN
SERVICE AFTER BEING IN STORAGE. THIS WARRANTY COVERS MATERIALS AND
LABOR TO RESTORE ANY PRODUCTS TO ORIGINAL FACTORY SPECIFICATIONS IF
A DEFECT IS FOUND WITHIN THE WARRANTY PERIOD.
THE DEFECTIVE PRODUCT SHOULD BE SENT, FREIGHT PREPAID, TO THE
FACTORY IN CARSON CITY, NEVADA. REPAIRS WILL BE PERFORMED AT THE
FACTORY AND RETURNED, PREPAID, BY THE SAME SHIPPING METHOD USED TO
SEND THE PRODUCT TO THE FACTORY.
THIS WARRANTY DOES NOT APPLY WHERE THE EQUIPMENT HAS SUSTAINED
DAMAGE DUE TO NEGLECT, MODIFICATION, CORROSION, OR OTHER REASON
BEYOND THE SCOPE OF THE NORMAL DEFINITION OF MANUFACTURING
DEFECT.
FURTHER, THIS WARRANTY IS LIMITED TO REPLACING THE DEFECTIVE
COMPONENTS AND RETURNING THE EQUIPMENT MANUFACTURED BY
UNIVERSAL ANALYZERS INC. TO THE CUSTOMER IN WORKING CONDITION. ANY
OTHER CLAIMS ARE OUTSIDE THE SCOPE OF THIS WARRANTY. NO WARRANTIES
ARE MADE AS TO THE SUITABILITY OF THE USE OF THE EQUIPMENT IN ANY
PARTICULAR APPLICATION OR LOCATION. THE SUITABILITY OF THE USE OF THE
EQUIPMENT IS THE RESPONSIBILITY OF THE CUSTOMER AND THE INSTALLING
CONTRACTOR.

Universal Analyzers Inc.

1701 SOUTH SUTRO TERRACE


TELEPHONE: (775) 883-2500

CARSON CITY, NV 89706


FAX: (775) 883-6388

UNIVERSAL ANALYZERS MODEL 3050 SAMPLE COOLER


SPECIFICATIONS
SAMPLE FLOW RATE:

0 to 8 L/M (at STP)

MAXIMUM INLET TEMPERATURE:


STAINLESS STEEL HEAT EXCHANGER:
KYNAR/GLASS HEAT EXCHANGER:

700o F. (371o C.)


280o F. (138o C.)

MAXIMUM INLET GAS DEW POINT:

194o F. (90o C.)

MAXIMUM INLET WATER CONCENTRATION:

70%*

MINIMUM AMBIENT TEMPERATURE:

32o F. (0o C.)

MAXIMUM AMBIENT TEMPERATURE:

105o F. (41o C.)*

MAXIMUM COOLING POWER, (SECOND STAGE):

126 BTU'S/Hr. (120 kJ/Hr.)

SAMPLE DEW POINT LEAVING FIRST STAGE:

AMBIENT + 10o F. (6o C.)

OUTLET SAMPLE DEW POINT:

41o F. (5o C.), adjustable

MAXIMUM INPUT POWER:

350 WATTS

VOLTAGE:

95-125 VAC, 50/60 Hz or


190-250 VAC, 50/60 Hz

ELECTRICAL CLASSIFICATION:

NEMA 1
Class I, Grps. A, B, C, & D,
Division 2 (FM Approved)

DIMENSIONS:

18" x 7 1/2" x 11", HWD

WEIGHT:

33 LB's., (15 KG)

SOLUBLE GAS REMOVAL RATES:

NO
NO2
SO2
CO
CO2

*at reduced flow rates. See capacity chart.

3050 Text Rev A

0% LOSS
<10% LOSS
< 2% LOSS
0% LOSS
< 2% LOSS

MODEL 3050 SAMPLE COOLER


OPERATING INSTRUCTIONS
APPLICATION
In order to analyze combustion products or incinerator effluents utilizing a direct
extractive sampling technique, it is important to remove the water vapor without
removing the water soluble fraction(s) from the gas sample. The heat exchangers
(impingers) used in the Universal Analyzers gas sample coolers are designed to minimize
the gas/condensate area and time of contact to reduce to a minimum, the amount of mass
transfer of those water soluble components from the gas phase into the liquid phase. The
result is a dry gas sample which has the same composition on a dry basis before and after
passing through the chiller.
A gas sample is usually taken from a stack with a probe extending into the center of the
stack mounted onto or adjacent to a heated filter. The heated filter is maintained at a
temperature above the dew point of the stack gas, usually 300o to 400o F. in order to
avoid cementing the filtered particulates to the filter medium with condensate. A means
is usually provided to automatically blow the particulates trapped by the filter, back into
the stack on a periodic basis.
The stack gas sample is clean but "wet" after passing through the filter assembly. The
moisture in the gas sample comes from the fuel as a product of combustion, from the
humidity in the air which supports the combustion and from the water content which was
trapped in the fuel. This latter source of water in the sample can be from burning moist
coal, wet garbage, or from water injected into the fire box. Water from all of these
sources will remove the water soluble gasses from the sample stream if allowed to
condense in the sample line prior to the controlled separation within the Universal
Analyzers' heat exchangers in the sample cooler.
In order to maintain the temperature of the gas sample above the dew point as it is
transported to the gas sample cooler, a heat traced sample line is usually employed. The
heat traced sample line can be very short if the gas sample cooler is located close to the
heated stack filter or it could be several hundred feet long if the gas sample cooler is
located in the analyzer shack some distance from the stack.
The Universal Analyzers gas sample cooler contains the special impinger type heat
exchanger(s). These are mounted within heat transfer blocks which are cooled by
thermoelectric elements utilizing the "Peltier Effect" discovered in France over half a
century ago. Where high water contents are encountered, it is efficient to remove the
condensate in two stages, one at the temperature of the air in the vicinity of the
"Precooler" and then by passing the sample into a heat exchanger cooled by the
thermoelectric elements. The precooler can remove water which will condense at the

temperature of the environment. In high water content samples, this could be as much as
80% of the water in the sample.
The thermoelectrically cooled stage is temperature controlled at a factory setting of 5o C.
This temperature can be adjusted with a set point potentiometer easily as described
below.
The gas sample conditioning system should contain additional components to insure that
a clean, dry sample is presented to the analyzer panel for minimum analyzer maintenance.
A sensor should be provided to sense the presence of condensate, should any exist in the
tubing following the chiller. This is the optional WCO (Water Carry-Over) sensor which
can be provided with a Universal Analyzers chiller. A visible coalescing filter which
collects particulates on the outside of the cylindrical filter, surrounded by a transparent
bowl will allow the operator to inspect the condition of the heated stack filter. The WCO
sensor can be provided with the sensing elements in the bottom of the filter bowl to
provide an early warning if the coalescing filter removes liquid from the sample stream.
This combination filter/moisture sensor is called the "WCOF".
The location of the sample pump (usually an oil-less diaphragm pump) within the
sampling system is the subject of much debate. If the pump is located upstream from the
chiller, it should have a heated head to avoid the presence of a two phase mixture which
shortens the life of a diaphragm and causes maintenance problems. It can be shown that
passing the gas sample through the chiller under a slight positive pressure will result in a
lower gas dew point than if the gas were at a slight vacuum. A common location for the
sample pump is in the sample line after the chiller. This allows the sample pump to
handle a cool, dry sample with much reduce maintenance problems. Some design
engineers have taken the approach to use a sample cooler having two chilled heat
exchangers. The pump is placed in the sample line between the two heat exchangers.
The first chilled heat exchanger takes enough of the water vapor from the sample to
protect the pump. The second heat exchanger is under a slight pressure and the dew point
is at it's minimum because it is on the discharge side of the pump. The sample pump
location within the sample system is a matter of choice and good engineering.
The condensate must be removed from the heat exchanger(s) using one of the following
methods:
1. A continuously running peristaltic tubing pump can be used
with the heat exchanger either under pressure or vacuum. This is an easy
solution which lends itself to leak testing because of the positive
displacement nature of the peristaltic pump. It is, however, a device which
requires periodic maintenance to replace the tubing.
2. Condensate can be removed from the heat exchanger using an
eductor with a flow limiter to draw some of the hot, wet sample through

the heat exchanger along with the condensate. This method utilizes the
heat exchanger as a bypass condenser and serves the purpose of reducing
the time lag in the sample line without loading the chiller with the water
vapor in the bypassed sample.
3. A float drain trap can be used to collect the condensate running
out of the bottom of the heat exchanger(s) if the heat exchangers are at a
slight positive pressure with respect to the atmosphere. The float drain
trap functions like a steam trap. The float rises to discharge the
condensate when there is enough to lift the float.
4. A small drain tank can be provided with a dual level,
conductivity type level control to collect the condensate. When the tank if
full, the level controller will start a pump to remove the condensate until it
reaches to the lower electrode. When contact is lost, the pump stops and
the condensate is again allowed to fill the tank. Typically, a peristaltic
pump is used to remove the condensate. The advantage is that the pump
only runs occasionally and bypasses no gas sample. The peristaltic pump
requires less maintenance because it does not run continuously.
Finally, a means of controlling the flow of the sample to the analyzers must be
considered. This can be as simple as providing a flow meter with a flow control needle
valve to pinch off the sample flow causing the sample pump to pump higher on the pump
curve. One option which is used is to provide an adjustable back-pressure regulator
between the inlet and outlet of the sample pump to allow a portion of the gas pumped to
be recirculated back to the inlet if discharge pressure exceeds the back-pressure control
point. Some analyzers have their own sample pump which may be sufficient to supply
the analyzer but insufficient to pull the sample through the chiller, sample line, and
heated stack filter. These can be supplied by piping the sample from the external sample
pump into the run of a tee with a flow meter which registers the flow of excess sample
from the branch of the tee to atmosphere. The internal analyzer sample pump can then
withdraw sample from the opposite run of the tee which is essentially at atmospheric
pressure and unaffected by pressure changes within the sample line due to changes in
filter pressure drop or sample pump efficiency.
DESCRIPTION
The key to the success of the Universal Analyzers Sample Cooler being able to condense
the water from a wet gas sample with a minimal loss of the water soluble gas fraction, is
due to the design of the heat exchanger. Please refer to Drawings P0146 thru P0149. The
separation occurs in a classical impinger which has a highly polished cylindrical surface
cooled to the desired dew point temperature. The hot wet sample is brought to the bottom
of the cylinder through an insulated tube and allowed to rise through a narrow annular

area at a relatively high Reynolds number to insure the entire sample is influenced by the
cold surface. The condensate falls down the cold polished surface in the form of a sheet
(as opposed to droplets or the bubbling of the gas sample through the condensate) which
minimizes the surface area in contact with the gas sample.
The temperature of the cylindrical condensation surface of the heat exchanger is
maintained through intimate contact with an aluminum heat transfer block. In the Model
3050 Sample Cooler, the first of the heat transfer blocks is cooled by direct contact with
the heat sink cooled by the blower. The temperature of the first of the two heat
exchangers will be about 10o F. above the temperature of the air passing through the heat
sink. (The temperature differential depends on the amount of heat that is being extracted
from the sample, which is a function of the water content of the sample.)
The second of the heat exchangers through which the sample passes is cooled by the use
of Thermoelectric Elements at a controlled temperature. The factory temperature setting
is 5o C. The temperature sensor is an AD592 semiconductor device. The controller is a
proportional controller with a proportional band of about 1o C. The set point is field
adjustable by opening the door on the front of the chiller, moving a jumper (J3) with
needle nose pliers, and adjusting a trim-pot to a new temperature value shown on the
display. The jumper needs to be placed in the original position to, again, indicate the
temperature of the heat transfer block.
The heat which is removed from the gas sample (and that which is created by the
Thermoelectric Elements) is discharged by a heat sink which is cooled with a centrifugal
drum blower. The turbulence and velocity provided by the high static pressure capability
of the blower allows for efficient cooling of the heat sinks. The heat sinks are fabricated
from solid blocks of aluminum which eliminates the epoxy joints in more conventional
heat sink designs which are barriers to heat conduction. The result is a heat removal
system with superior performance under all conditions.
Universal Analyzers Thermoelectric Sample Coolers have a digital display as a front
panel indication of the of the operating temperature, (o C.) of the heat exchangers. In
addition, there are two green LED lamps to indicate the status of the cooler. The
"COOL" lamp lights when the operating temperature is between 0o C and 10o. If the
operating temperature is above or below this range, the light will be out and the alarm
relay will be in the alarm position. The "DRY" lamp indicates that there is no water in
contact with the moisture sensor listed as the "WCO" or "WCOF" options. If the "DRY"
lamp is out, the alarm relay will be in the alarm position.
The alarm relay is a single relay activated by either of the above alarm conditions. The
alarm relay is energized in the "safe" condition and will relax into the de-energized, alarm
position if a fault is detected. If a broken thermocouple occurs, the result will be that the
temperature indication is out of range and the alarm relay will indicate a fault.

A thermal switch is mounted on the heat sink to serve as a safety device to turn off power
to the thermoelectric elements if the heat sink approaches 160o F. This protects the
thermoelectric elements from melting in the event high temperature conditions are
encountered for any reason.
The 3000 Series thermoelectric sample coolers are designed to incorporate a peristaltic
pump for condensate removal within the enclosure to reduce the total volume required for
the system installation. Alternatively, it also has provisions for incorporating an eductor,
air filter, and dual heat sample pump with the second head used for generating the
necessary compressed air for the eductor.
Electrical connections are made to external wiring which will enter conduit fittings on the
bottom of the chiller. Internal conduit tubes direct the wires as they are pushed into the
chiller into the vicinity of the terminal strips. All of the terminal strips are easily
accessible from the front of the chiller.
If it is desired to connect conduit to the chiller, a conduit coupling can be screwed onto
the 1/2" male conduit threads which extend from the bottom.
"PEM Nuts" are provided on the right hand side to mount accessories such as a sample
pump, visible filter, condensate alarm sensor, and flow meter. These options can be user
or factory installed.
The chiller can be changed from 115 VAC to 230 VAC operation with the movement of a
circuit card on the power supply. Accessories such as the sample pump or peristaltic
pump must be rated for the correct supplied line voltage.
INSTALLATION INSTRUCTIONS
Thermoelectric Sample Coolers should be installed away from heat sources in a well
ventilated area of an instrument rack or enclosure. Completely enclosing any instrument
which consumes as much power as a thermoelectric chiller will cause the temperature of
the interior of an unventilated enclosure to rise to too great a level for the sample cooler
to perform reliably. Where it is necessary for the enclosure to be tightly gasketed, a water
cooled thermoelectric chiller should be used. Contact the factory for recommendations.
Air purging an enclosure does not provide enough ventilation to remove the heat which
will be generated internally by the sample chiller. On the other hand, there are air
conditioners and vortex coolers which can provide the necessary cooling for enclosing
thermoelectric chillers.

The 3000 series sample cooler has a flange on the top and bottom to allow it to be wall
mounted or mounted to rails in an instrument rack. Accessories mounted on the side can
be supported by the mounting flange on the cooler.
Sample tubing should be brought to the front, unchilled heat exchanger first. A 3/8"
tubing fitting is provided at the top of the first heat exchanger as the sample inlet. A 1/4"
Teflon transfer tube is provided to carry the sample from the first, ambient temperature
heat exchanger to the second, chilled stage. The dry sample outlet from the cooler is the
1/4" Kynar tubing fitting coming out of the top of the second heat exchanger at an angle.
The gas sample should be further conditioned with an additional filter as a safety measure
before it enters the analyzer(s). The WCOF visible filter which incorporates a safety
moisture sensor is designed for this purpose. A similar filter without the moisture sensor
is also available from Universal Analyzers. A separate moisture sensor, the WCO, can
also be provided where the user prefers that form factor. Contact the factory for
drawings.
A sample pump is normally required to pull the sample through the sample cooler and to
force it through the visible filter. The moisture sensor, if not a part of the visible filter
should be located ahead of the visible filter.
Some have preferred to locate the sample pump in the sample line between the two heat
exchangers. The theory is that the first heat exchanger reduces the water content to the
point that water will not condense in the sample pump. A controlled portion of the
sample can be discharged to atmosphere to reduce the sample lag in the sample line
before the final cooling/dehydration stage. The final cooling stage then only handles the
sample which is required by the analyzer(s). This location of the sample pump also
allows the final stage of cooling to be performed at a slightly higher pressure which is
theoretically more efficient.
There must be a method to remove the condensate from the heat exchanger(s). Several
methods are discussed above. A drain line from the peristaltic pump, eductor, or drain
pot must be run to sewer, a container, or to the ground outside the instrument enclosure to
avoid collecting water (condensate) on the floor.
If an eductor is utilized to remove the condensate, the outlet tube length should be no
longer than two feet in order to keep too much back pressure from the outlet of the
eductor. The outlet tube can be placed in a larger pipe to channel the condensate to a
drain.
The electrical power, about 3 amps at 115 VAC or 1.5 amps at 230 VAC 50/60 HZ
should be supplied. Installation shall be in accord with the manufacture's instructions and
the National Electrical Code (ANSI/NFPA 70). Tampering and replacement with
nonfactory components may adversely affect the safe use of the system. For the 115

VAC case, a power cord is supplied. It can be replaced with conduit wiring easily. See
the electrical installation Drawing P0093. The alarm relay has two sets of "Form C"
contacts brought out to the center terminal strip, TB2. One set are totally dry contacts.
The other set are MOV protected and designed to interrupt power to the sample pump
when in the alarm condition. When the pump is powered through the connector on the
power supply board, the MOV protected contacts are hot, with line power supplied.
The moisture sensor can be connected to the "Viking" connector mounted on the exterior
of the cabinet below the heat exchangers or, alternatively, connected to TB3 on the power
supply board.
START UP PROCEDURE
Apply power to the sample cooler. The indicated temperature will start to drop
immediately. It should be below the over-temperature alarm point in approximately four
minutes and the "COOL" green LED lamp should light. When the temperature reaches
the control set point, the rate at which the temperature drops will be reduced. The
temperature will stabilize within 1o C. of the control set point.
Start the sample gas flow by turning on the sample pump and adjusting the flow control
valve to the desired flow rate. Water should be observed to be removed from the bottom
of each heat exchanger when steady state conditions are established.
If an eductor is utilized to remove the condensate, a strong flow of air should be felt to be
flowing from the eductor outlet tube. The outlet tube length should be no longer than two
feet. The outlet tube can be placed in a larger pipe to channel the condensate to a drain.
The green "DRY" LED light will be on. If an optional moisture is installed, the light will
go out and the alarm relay de-energized when the sensor detects condensate. A test can
be performed by putting water on the moisture sensor. If the sample pump is powered
through the alarm relay contacts, it will cease running. When the moisture sensor is
wiped dry, the sample pump will start.
Turn on the analyzer(s) and initiate the calibration cycle.

TROUBLE SHOOTING
Tampering and replacement with nonfactory components may adversely affect the safe
use of the system.
The presence of water in liquid form after the sample cooler is an indication of a fault in
the system. Reasons for the presence of condensate in the system after the sample cooler
could be one or more of the following:
1. Overloading of the cooling capacity of the cooler due to too
much water vapor in the sample or due to too great a sample flow rate.
2. The condensate removal equipment (peristaltic pump, eductor,
or drain pot) may be faulty. The heat exchanger(s) may be full of
condensate.
3. An air leak may be in the condensate removal system allowing
air to enter and blow the condensate back into the heat exchanger. (This
assumes the heat exchanger is under a slight vacuum.)
4. The temperature of the air passing through the cooler to cool the
heat sink is too high. This could be due to placement of the cooler in a
tightly sealed box.
5. The sample cooler could have failed.
Dimensional drawings, installation drawings, and schematics are included as part of this
manual. If additional information is required, telephone assistance can be obtained by
calling (775) 883-2500 or FAX request to (775) 883-6388.

A
8

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

8
7-1/2"
5-1/2"

FM

DRY SAMPLE GAS


OUTLET
1/4" TUBE FITTING

APPROVED

11-1/4" 11-3/4"
R

WET SAMPLE GAS


INLET
3/8" TUBE FITTING

HOLES FOR
1/4" BOLTS
(4 TYP)
(BOLTS SUPPLIED
BY OTHERS)

THERMOELECTRICALLY
COOLED STAGE
EX
HA
US
TA

AMBIENT
COOLED STAGE

IR

FL

OW

SOLID STATE
RELAY
DISPLAY
CIRCUIT BOARD

HIGH PRESSURE
BLOWER/S
HEAT SINKS

17"
EX
HA
US
TA

18"

EX
HA
US
TA

EX
HA
US
TA

IR

IR

FL

FL

FL

OW

OW

OW

5,
46
5,
5

78

POWER SUPPLY
CIRCUIT BOARD

IR

3
1/2" CONDUIT PORT
(2 TYP)

OPTIONAL
DUAL HEADED
PERISTALTIC PUMP

CONDENSATE DRAIN
LINES

USER ACCESSABLE
TERMINAL BLOCKS

SINGLE SAMPLE
DUAL STAGE
SAMPLE COOLER
A

04/08/98

Font/Patent#/New Heads/Relay

REV

DATE

DESCRIPTION
REVISIONS

EM

TB

DWN APVD

MODEL 3050 SAMPLE COOLER


OUTLINE
PART NO.

FOR

UNIVERSAL ANALYZERS INC.

1701 S. Sutro Terrace


Carson City, Nevada 89706 USA

DRAWN BY

DRAWING NO

APVD BY

P0069

HARVEY MITCHELL
TED BARBEN
DATE
SCALE
09/02/93
1:1

SIZE

SHEET
1 OF 1

13 5/8"
5 1/2"
3 13/16"

WET GAS
SAMPLE INLET
3/8" TUBING
FITTING

SAMPLE COOLER

DRY SAMPLE GAS


TO ANALYZER

18 13/16"
18"

17"
SINGLE HEAD
ADI MINI
DIA-VAC
PUMP

8
,57
65
5,4

PERISTALTIC
PUMP HEADS

B
5,465,57

WATER
CARRY OVER
SENSOR
CONNECTOR

FLOW METER
VISIBLE FILTER
WITH WATER
CARRY OVER
SENSOR

FRONT VIEW
DESCRIPTION

REV DATE

DWN APVD

REVISIONS

3000 SERIES SAMPLE COOLER


ACCESSORY OUTLINE FOR
SYSTEMS WITH PERISTALTIC PUMP

FOR
INSTRUMENT

PART NO.
NOT ISSUED

ISOMETRIC VIEW

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWN BY
APVD BY

DATE
07/07/1998

DRAWING NO

E. Musselman
T. Barben II

P0294
SCALE
1:1

SIZE
E

SHEET
1 OF 1

24V
ADJ

+ 4

+ 3

+ 2

SOLID STATE RELAY


MODEL Z240D10

+ 1

+3

TB3

W
C
O

RED
BLACK

POWER SUPPLY
BOARD

PELTIER CONNECTIONS
RED
BLACK
RED
BLACK
RED
BLACK
RED
BLACK

RED

1
2

MOISTURE
SENSOR
CONNECTOR

-4

- CAUTION TERMINALS
1,2, & 3 ARE
HOT WHEN
SAMPLE
PUMP IS
INSTALLED
TB2

PELTIER
ELEMENTS

A MOV PROT.
L N.C. 1
2
A
R N.O. 3
M
R
E
L
A
Y

#1

N.C.

DRAIN

5
N.O.

#4

BLOWER

6 AMP FUSE

MODELS
3040
3050
3080

JUMPER CARD
115V
230V

12 AMP FUSE
TB1

P HOT L1
O
W COM L2
E
R

SAMPLE

DRY CIRC.

1
2
3

MODEL
3080
BLOWER
BLACK
BLUE
RED
GREEN

BLACK

#2

WHITE

G
B R
BL LAC EEN
AC K
K

#3
AD592

BL
AC
K

IN
DRA
D
RE

DR
AIN
RE
D

K
AC
BL ACK
BL

THERMAL
SAFETY
SWITCH

MOISTURE SENSOR
CONNECTOR
R

PERISTALTIC
DRAIN PUMP

3
FM
APPROVED

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

DISPLAY BOARD

2
(JUMPER)
TEMPERATURE
OR
SET POINT

TEMP
SET

10/24/03

REV

DATE

SSR CONNECTIONS TO 1-BLK / 4-RED


DESCRIPTION
REVISIONS

EM

PART NO.

FOR

UNIVERSAL ANALYZERS INC.

1701 S. Sutro Terrace


Carson City, Nevada 89706 USA

CALIBRATION ADJUST
(FACTORY SET)

DRAWN BY

DRAWING NO

APVD BY

P0093

HARVEY MITCHELL
TED BARBEN
DATE
SCALE
3/17/94
NONE

TEMPERATURE
SENSOR

MODEL 3040 3050 3080


INTERNAL 115VAC WIRING
DIAGRAM

POWER
SUPPLY

SET POINT
ADJUST

TB

DWN APVD

SIZE

SHEET
1 OF 2

24V
ADJ

+ 4

+ 3

+ 2

SOLID STATE RELAY


MODEL Z240D10

+ 1

+3

TB3

W
C
O

RED
BLACK

POWER SUPPLY
BOARD

PELTIER CONNECTIONS
RED
BLACK
RED
BLACK
RED
BLACK
RED
BLACK

RED

1
2

MOISTURE
SENSOR
CONNECTOR

-4

- CAUTION TERMINALS
1,2, & 3 ARE
HOT WHEN
SAMPLE
PUMP IS
INSTALLED
TB2

PELTIER
ELEMENTS

A MOV PROT.
L N.C. 1
2
A
R N.O. 3
M
R
E
L
A
Y

#1

N.C.

DRAIN

5
N.O.

#4

BLOWER

6 AMP FUSE

MODELS
3040
3050
3080

JUMPER CARD
115V
230V

12 AMP FUSE
TB1

P HOT L1
O
W COM L2
E
R

SAMPLE

DRY CIRC.

1
2
3

MODEL
3080
BLOWER
WHITE
BLUE
BLACK
RED
GREEN

#2

WHITE

G
B R
BL LAC EEN
AC K
K

#3
AD592

BL
AC
K

IN
DRA
D
RE

DR
AIN
RE
D

K
AC
BL ACK
BL

THERMAL
SAFETY
SWITCH

MOISTURE SENSOR
CONNECTOR
R

PERISTALTIC
DRAIN PUMP

3
FM
APPROVED

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

DISPLAY BOARD

2
(JUMPER)
TEMPERATURE
OR
SET POINT

TEMP
SET

10/24/03

REV

DATE

SSR CONNECTIONS TO 1-BLK / 4-RED


DESCRIPTION
REVISIONS

EM

PART NO.

FOR

UNIVERSAL ANALYZERS INC.

1701 S. Sutro Terrace


Carson City, Nevada 89706 USA

CALIBRATION ADJUST
(FACTORY SET)

HARVEY MITCHELL
APVD BY
TED BARBEN
DATE
SCALE
3/17/94
NONE
DRAWN BY

TEMPERATURE
SENSOR

MODEL 3040 3050 3080


INTERNAL 230VAC WIRING
DIAGRAM

POWER
SUPPLY

SET POINT
ADJUST

TB

DWN APVD

DRAWING NO

P0093
SIZE

SHEET
2 OF 2

TO
DISPLAY
BOARD

POWER SUPPLY
BOARD

24V
ADJ

+ 4

+ 3

+ 2

+ 1

TB3

W
C
O

RED

1
2

MOISTURE
SENSOR
CONNECTOR

-CAUTIONTERMINALS
1,2 & 3 ARE
HOT WHEN
SAMPLE
PUMP IS
INSTALLED
TB2

R
E
L
A
Y

DRY CIRC.

N.C.

4
5

N.O.

BLACK
WHITE
GREEN
DRY CIRCUIT
TO ALARM
(CONTACT CLOSES
WHEN IN ALARM)

6 AMP FUSE

SAMPLE
DRAIN
BLOWER
JUMPER CARD
115V
230V

12 AMP FUSE
TB1

P HOT L1
O
W COM L2
E
R

SAMPLE PUMP

A MOV PROT.
L N.C. 1
2
A
R N.O. 3
M

1
2

WATER CARRY OVER


FILTER

L 1(HOT)
L 2(COMMON)

WATER CARRY OVER SENSOR


WITH FLOW THRU ADAPTER

TERMINAL BLOCK
WIRED

A 01/05/00 ADD PAGES 2 & 3 TO DRAWING PACKAGE


REV DATE
DESCRIPTION

EAM JK
DWN APVD

REVISIONS

MOISTURE SENSOR
CONNECTOR

OR
EXTERNAL WIRING DIAGRAM
FOR MODELS 3040 3050 3080

OR

PART NO.
NOT ISSUED

FOR
INSTRUMENT

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
H. MITCHELL
P0094
T. BARBEN II
DATE
SCALE
SIZE
03/17/1994
1:1
D

DRAWN BY
APVD BY

SHEET
1 OF 3

TO
DISPLAY
BOARD

POWER SUPPLY
BOARD

24V
ADJ

+ 4

+ 3

+ 2

+ 1

TB3

W
C
O

RED

1
2

MOISTURE
SENSOR
CONNECTOR

-CAUTIONTERMINALS
1,2 & 3 ARE
HOT WHEN
SAMPLE
PUMP IS
INSTALLED
TB2

A MOV PROT.
L N.C. 1
2
A
R N.O. 3
M
R
E
L
A
Y

DRY CIRC.

N.C.

4
5

N.O.

SAMPLE PUMP
SAMPLE

DRY CIRCUIT TO ALARM


(CONTACT CLOSES
WHEN IN ALARM)

DRAIN

BLOWER

6 AMP FUSE

JUMPER CARD
115V
230V

12 AMP FUSE
TB1

P HOT L1
O
W COM L2
E
R

1
2

HOT

WATER CARRY OVER


FILTER

COMMON

WATER CARRY OVER SENSOR


WITH FLOW THRU ADAPTER

SOLID STATE RELAY


MAX 240VAC 25 AMP
1
240V

TERMINAL BLOCK
WIRED
A 01/05/00 SHOW HIGH CURRENT PUMP & SOLID STATE RELAY EAM JK
REV DATE
DWN APVD
DESCRIPTION
REVISIONS

25A

115 / 230 VAC


3+
4

BLACK

MOISTURE SENSOR
CONNECTOR

HOT (CUSTOMER SUPPLIED VOLTAGE)

OR

EXTERNAL WIRING DIAGRAM


FOR 115VAC HIGH CURRENT
SAMPLE PUMP APPLICATIONS

OR

PART NO.
FOR
NOT ISSUED MODELS 3040 3050 3080

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
H. MITCHELL
P0094
T. BARBEN II
DATE
SCALE
SIZE
03/17/1994
1:1
D

DRAWN BY
APVD BY

SHEET
2 OF 3

TO
DISPLAY
BOARD

POWER SUPPLY
BOARD

24V
ADJ

+ 4

+ 3

+ 2

+ 1

TB3

W
C
O

RED

1
2

MOISTURE
SENSOR
CONNECTOR

-CAUTIONTERMINALS
1,2 & 3 ARE
HOT WHEN
SAMPLE
PUMP IS
INSTALLED
TB2

A MOV PROT.
L N.C. 1
2
A
R N.O. 3
M
R
E
L
A
Y

DRY CIRC.

N.C.

4
5

N.O.

SAMPLE PUMP
SAMPLE

DRY CIRCUIT TO ALARM


(CONTACT CLOSES
WHEN IN ALARM)

DRAIN

BLOWER

6 AMP FUSE

JUMPER CARD
115V
230V

12 AMP FUSE
TB1

P HOT L1
O
W COM L2
E
R

1
2

HOT
COMMON

WATER CARRY OVER


FILTER

WATER CARRY OVER SENSOR


WITH FLOW THRU ADAPTER

SOLID STATE RELAY


MAX 240VAC 25 AMP
1
240V

TERMINAL BLOCK
WIRED
2

25A

115 / 230 VAC


3+
4

BLACK

L1 (CUSTOMER
SUPPLIED
VOLTAGE)

SOLID STATE RELAY


MAX 240VAC 25 AMP

1
240V

MOISTURE SENSOR
CONNECTOR

OR

EXTERNAL WIRING DIAGRAM


FOR 230VAC HIGH CURRENT
SAMPLE PUMP APPLICATIONS

OR

PART NO.
FOR
NOT ISSUED MODELS 3040 3050 3080

UNIVERSAL ANALYZERS INC.


2

25A

115 / 230 VAC


3+
4

BLACK

A 01/05/00 SHOW HIGH CURRENT PUMP & SOLID STATE RELAY EAM JK
REV DATE
DWN APVD
DESCRIPTION
REVISIONS

1701 South Sutro Terrace


Carson City, Nevada 89706 USA

L2 (CUSTOMER SUPPLIED VOLTAGE)

DRAWING NO
H. MITCHELL
P0094
T. BARBEN II
DATE
SCALE
SIZE
03/17/1994
1:1
D

DRAWN BY
APVD BY

SHEET
3 OF 3

1/4" NPT
.400

D
.300

1-1/8"

.400

MOUNTING HOLE
FOR 10-32 SCREW
(2 TYP)
FLOW

.950

3"

1/4" NPT
3

1.500
24" CABLE
B 03/10/03 Revised P/N for Item 1 & 2
REV DATE
DESCRIPTION
REVISIONS

PART NO.
6020-0001

PART NO.
5101-0002
5101-0001
5205-0007

FOR
W.C.O.

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
H. MITCHELL
P0006
TED BARBEN II
DATE
SCALE
SIZE
B
10/08/1991
1:1

DRAWN BY
APVD BY

MODEL W.C.O. WATER


CARRY OVER MOISTURE
SENSOR WITH FLOW THRU
ADAPTER

TABLE A
ITEM NAME
1 HOLDER FLOW THRU BODY
2 KYNAR MOISTURE SENSOR ASSY
3 SENSOR LOCK ASSY KIT

RPH JK
DWN APVD

SHEET
1 OF 1

HOLES FOR #10 PANHEAD


MOUNTING SCREWS
(SUPPLIED BY OTHERS)

1/2"

3"

2"

1.65"

EXPLODED FRONT VIEW

3"
2"

2
COMPLETE
FILTER P/N

1"

5/8"
1

1 1/4"

8"

1
FILTER
INLET PORT
1/4" NPT

1
FILTER
OUTLET PORT
1/4" NPT

4980-0039
4980-0058
4980-0005
4980-0009
4980-0059

FILTER PRESSURE DROP RATINGS


ELEMENT
INCHES OF WATER
P/N
5 LPM 10 LPM 20 LPM
SINTERED S.S. 4980-0023
10
14
24
4980-0022
PAPER
8
12
20
4980-0006
TEFLON
4
6
10
4980-0007
CERAMIC
4
6
10
COATED CERAMIC 4980-0018
6
8
12

1 FOR COALESCING (COATED CERAMIC WCOF)


P/N 4980-0059 SAMPLE INLET & OUTLET PORTS
ARE REVERSED.

7
1/8"

08/24/04

REV

DATE

8
7
6
5

2
1
1
1

4
3
2
1
ITEM

MIDDLE
PIN

Revised with New Bowl Design & Ext. Cable


DESCRIPTION
REVISIONS
-

1
1
1
1
QTY DWG NO.

4704-0010
3907-0052
4904-0006
5205-0038

RPH

TB

VIEW "A" - 2x SCALE


VERIFY BEFORE CONNECTING THAT
MIDDLE PIN IS SHORTER THAN
OUTSIDE PIN. IF NOT, CUT MIDDLE PIN
AS SHOWN

10-32X3/8"PH
14" EXT. CABLE
O-RING 2-030
BOWL ASSEMBLY

5110-0012
SEE TABLE
5008-0004
5110-0011
PART NO.

A
LEFT SIDE

RETAINER
FILTER ELEMNT
BRACKET
KYNAR HEAD
DESCRIPTION

FOR
INSTRUMENT

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWN BY

DRAWING NO

APVD BY

P0018

HARVEY MITCHELL
TED BARBEN II
DATE
SCALE
04/14/92
1:1

RIGHT SIDE

PART NO.
SEE TABLE

DWN APVD

VISIBLE GAS FILTER WITH MOISTURE SENSOR


(W.C.O.F.)

ITEM #3 TYPE

SIZE

SHEET
1 OF 1

D
2-5/8"
INSULATED SECTION

TOP VIEW

CHILLED SURFACE

0.25" INSIDE DIAMETER

THERMOCOUPLE
(TYPE "K")
(TYPE "J")
(TYPE "T")

"L"
(SEE TABLE)

CONDENSATE DRAIN
3/8" NPT
(FEMALE)

ANNULAR SPACE .060" WIDE


(GAS TO BE DRIED RISES IN THIS SPACE)

SIDE VIEW

DRY GAS OUTLET


1/8" NPT
(FEMALE)

B
SAMPLE GAS INLET
1/4" NPT
(FEMALE)

1). SEE SHEET 2 FOR ASSEMBLY INSTRUCTIONS.

APPROX 316 S.S.


LENGTH EXCHANGER
"L"
P/N
5200-SK50
5"
5200-SJ50
5200-ST50

10"

5200-SK10
5200-SJ10
5200-ST10

HASTELLOY C276
EXCHANGER
P/N
5200-CK50
5200-CJ50
5200-CT50

ALLOY 20
EXCHANGER
P/N
5200-AK50
5200-AJ50
5200-AT50

5200-CK10
5200-CJ10
5200-CT10

5200-AK10
5200-AJ10
5200-AT10

TITANIUM
UNIVERSAL
EXCHANGER ANALYZERS
P/N
SERIES
5200-TK50
400
5200-TJ50
500
5200-TT50
800
5200-TK10
1000
5200-TJ10
1100
5200-TT10
3000

01/23/97

REV

DATE

SHOW NEW DESIGN W/THERMOCOUPLE BRACKET

HEAT EXCHANGER OUTLINE


PART NO.
SEE TABLE

1701 South Sutro Terrace


Carson City, Nevada 89706 USA

EV MUSSELMAN

HARVEY MITCHELL
DATE
SCALE
12/11/95
1:1

FOR
SEE TABLE

UNIVERSAL ANALYZERS INC.

APVD BY

HWM

DWN APVD

METALLIC TEMPERATURE SENSING

DRAWN BY

EAM

DESCRIPTION
REVISIONS

DRAWING NO

P0146
SIZE

SHEET
1 OF 2

EXPLODED METALLIC HEAT EXCHANGER


THERMOCOUPLE
ASSY.
(TYPE "K")
(TYPE "J")
(TYPE "T")

2-021 VITON
O-RING
P/N 4904-0013
(SEE NOTES)

METALLIC
OUTER TUBE ASSEMBLY

S.S. COMPRESSION FITTING


W/TEFLON FERRULES
P/N 4954-0095

1/16" TEFLON
THERMOCOUPLE SLEEVE

METALLIC
INNER TUBE ASSEMBLY

6-32 X 1/2"
PAN HEAD SCREW

"L"
LENGTH
5"
5"
5"
5"
5"
5"
5"
5"
5"
5"
5"
5"
10"
10"
10"
10"
10"
10"
10"
10"
10"
10"
10"
10"

MATERIAL
DESC.
316 S.S.
"
"
ALLOY 20
"
"
HASTELLOY C276
"
"
TITANIUM
316 S.S.
"
"
ALLOY 20
"
"
HASTELLOY C276
"
"
TITANIUM
"
"

COMPLETE ASSY
P/N
5200-SK50
5200-SJ50
5200-ST50
5200-AK50
5200-AJ50
5200-AT50
5200-CK50
5200-CJ50
5200-CT50
5200-TK50
5200-TJ50
5200-TT50
5200-SK10
5200-SJ10
5200-ST10
5200-AK10
5200-AJ10
5200-AT10
5200-CK10
5200-CJ10
5200-CT10
5200-TK10
5200-TJ10
5200-TT10

COMPLETE ASSEMBLY

SPARE PARTS LIST


OUTER TUBE
INNER TUBE
ASSY P/N
ASSY P/N
5201-0015
5201-0012
"
"
5201-0022
"
"
5201-0020
"
"
5201-0024
"
"
5201-0013
"
"
5201-0023
"
"
5201-0021
"
"
5201-0025
"
"

"
"
"
"
"
"
"
"
5201-0028
"
"
5201-0016
"
"
"
"
"
"
"
"
5201-0029
"
"

TYPE "K" THERMOCOUPLE


ASSEMBLY P/N
1150-0011
X
X
"
X
X
"
X
X
1150-0018
X
X
1150-0011
X
X
"
X
X
"
X
X
1150-0018
X
X

TYPE "J" THERMOCOUPLE


ASSEMBLY P/N
X
1150-0012
X
X
"
X
X
"
X
X
1150-0019
X
X
1150-0012
X
X
"
X
X
"
X
X
1150-0019
X

TYPE "T" THERMOCOUPLE


ASSEMBLY P/N
X
X
1150-0013
X
X
"
X
X
"
X
X
1150-0014
X
X
1150-0013
X
X
"
X
X
"
X
X
1150-0014

2). LIGHTLY LUBRICATE O-RING


WITH SILICONE GREASE BEFORE
ASSEMBLY.

INNER TUBE
MATERIAL IDENTIFIER

1). O-RING IS FACTORY INSTALLED


IN METALLIC OUTER TUBE ASSY.

2-56 X 1/2"
PAN HEAD SCREW

NOTES:
"L"
(SEE TABLE)

OUTER TUBE
MATERIAL IDENTIFIER

01/23/97

SHOW NEW DESIGN W/THERMOCOUPLE BRACKET

EAM HWM

REV

DATE

DESCRIPTION
REVISIONS

DWN APVD

METALLIC TEMPERATURE SENSING


HEAT EXCHANGER ASSEMBLY
PART NO.
SEE TABLE

FOR
400 500 800 1000 1100 3000 SERIES

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
EV MUSSELMAN
P0146
HARVEY MITCHELL
DATE
SCALE
SIZE
B
12/11/95
1:1

DRAWN BY
APVD BY

SHEET
2 OF 2

8
B

6
B

5
VITON O-RING
P/N 4904-0013

INSULATED
INNER TUBE
ASSY

OUTER TUBE
ASSY

.305 I.D.

SAMPLE
GAS
OUTLET
C
1/4"
TUBE

PLUG
1/16" NPT

A
TOP VIEW

ANNULAR SPACE .060" WIDE


(GAS TO BE DRIED RISES IN THIS SPACE)

SAMPLE GAS INLET


3/8" TUBE

SECTION A-A

HEAT
APPROX
LENGTH EXCHANGER OUTER TUBE MATERIAL
P/N
"L"
5"
5200-S050
316 S.S.
5"
5200-T050
TITANIUM
5"
5200-A050
ALLOY 20
5"
5200-C050
HASTELLOY C-276
5"
5200-S05T TEFLON COATED 316 S.S.
10"
5200-S010
316 S.S.
10"
5200-T010
TITANIUM
10"
5200-A010
ALLOY 20
10"
5200-C010
HASTELLOY C-276
10"
5200-S01T TEFLON COATED 316 S.S.

CONDENSATE
DRAIN 3/8"NPT

"L"
(SEE TABLE)

UNIVERSAL
ANALYZERS
SERIES
400/500
400/500
400/500
400/500
400/500
800/1000/1100/3000
800/1000/1100/3000
800/1000/1100/3000
800/1000/1100/3000
800/1000/1100/3000

02/09/99

REVISE P/N TABLE FOR TEFLON COATED

REV

DATE

DESCRIPTION
REVISIONS

1). SEE SHEET 2 FOR ASSEMBLY INSTRUCTIONS.

EAM

RD

DWN APVD

HEAT EXCHANGER METALLIC


NON-TEMPERATURE SENSING
OUTLINE

PART NO.
SEE TABLE

FOR
SEE TABLE

UNIVERSAL ANALYZERS INC.


1701 South SutroTerrace
Carson City, Nevada 897
06 USA

SECTION B-B

DRAWN BY

DRAWING NO

APVD BY

P0147

EV MUSSELMAN
H. MITCHELL
DATE
SCALE
12/11/95
NONE

SIZE

SHEET
1 OF 2

VITON O-RING
#2-021
SEE NOTES

D
OUTER TUBE ASSEMBLY METALLIC

INNER TUBE
MATERIAL
IDENTIFIER

"L"
INNER TUBE ASSEMBLY

OUTER TUBE
MATERIAL
IDENTIFIER

HEAT EXCHANGER
SEPERABLE
INNER
APPROX
P/N
TUBE ASSY
"L"
P/N
LENGTH
5"
5200-S050 5201-0015
5"
5200-T050
"
5"
5200-A050
"
5"
5200-C050
"
5"
5200-S05T 5201-0041
10"
5200-S010 5201-0016
10"
5200-T010
"
10"
5200-A010
"
10"
5200-C010
"
10"
5200-S01T 5201-0042

SPARE PARTS LIST


OUTER
VITON O-RING PLUG
TUBE ASSY #2-021 P/N 1/16"NPT
P/N
P/N
5201-0012
4904-0013 4951-0058
5201-0024
"
"
5201-0022
"
"
5201-0020
"
"
5201-0043
"
"
5201-0013
4904-0013 4951-0058
5201-0025
"
"
5201-0023
"
"
5201-0021
"
"
5201-0044
"
"

NOTES:
1). O-RING IS FACTORY INSTALLED IN
METALLIC OUTER TUBE.
2). LIGHTLY LUBRICATE O-RING WITH
SILICONE GREASE BEFORE ASSEMBLY. B

02/09/99

REV

DATE

REVISE TABLE TO INCLUDE TEFLON PARTS EAM


DESCRIPTION
REVISIONS

FOR
INSTRUMENT

UNIVERSAL ANALYZERS INC.


1701 South SutroTerrace
Carson City, Nevada 8970
6 USA

DRAWN BY

DRAWING NO

APVD BY

P0147

EV MUSSELMAN
H. MITCHELL
DATE
SCALE
12/11/95
NONE

RD

DWN APVD

HEAT EXCHANGER ASSEMBLY


NON-TEMPERATURE SENSING
METALLIC
PART NO.
SEE TABLE

SIZE

SHEET
2 OF 2

CHILLED SURFACE

INSULATED SECTION

0.300" INSIDE DIAMETER

THERMOCOUPLE
(TYPE "J")
(TYPE "K")
(TYPE "T")

TOP VIEW

ANNULAR SPACE .060" WIDE


(GAS TO BE DRIED RISES IN THIS SPACE)

2-5/8"

"L"
(SEE TABLE)

CONDENSATE DRAIN
3/8" NPT
(FEMALE)

SIDE VIEW
DRY GAS OUTLET
1/8" NPT
(FEMALE)

SAMPLE GAS INLET


1/4" NPT
(FEMALE)

1). SEE SHEET 2 FOR ASSEMBLY INSTRUCTIONS.


GLASS KYNAR
LENGTH HEAT EXCHANGER
ASSEMBLY
"L"
P/N

5"

5200-KK50
5200-KJ50
5200-KT50

10"

5200-KK10
5200-KJ10
5200-KT10

15"

5200-KK15
5200-KJ15
5200-KT15

UNIVERSAL
ANALYZERS
SAMPLE COOLER
MODELS
520
530
540
570
1040
1050
1060
1080
1090

C 10/02/01
REV DATE

Add 15" Heat Exchanger Data To Table


DESCRIPTION
REVISIONS

Glass Kynar Temperature Sensing


Heat Exchanger Outline
PART NO.
SEE TABLE

FOR
SEE TABLE

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

FGA2

DRAWN BY
APVD BY

DRAWING NO

E. Mussleman
H. Mitchell

DATE
12/11/95

EM
JK
DWN APVD

P0148
SCALE
1:1

SIZE

SHEET
1 OF 2

EXPLODED GLASS KYNAR HEAT EXCHANGER

S.S. COMPRESSION FITTING


W/TEFLON FERRULES
P/N 4954-0095

THERMOCOUPLE
ASSY.
(TYPE "K")
(TYPE "J")
(TYPE "T")

1/16" THERMOCOUPLE
TEFLON INSULATOR

#2-120 VITON
O-RING
P/N 4904-0004
(SEE NOTE 2)

6-32 X 1/2"
PAN HEAD SCREW

GLASS TUBE

#2-018 VITON
O-RING
P/N 4904-0003
(SEE NOTE 2)

KYNAR INSULATED
SECTION ASSEMBLY

GLASS KYNAR
HEAT EXCHANGER
ASSEMBLY
LENGTH "L"
P/N
5"
5200-KK50
5200-KJ50
5"
5200-KT50
5"
10"
5200-KK10
10"
5200-KJ10
10"
5200-KT10
15"
5200-KK15
15"
5200-KJ15
15"
5200-KT15

KYNAR DRAIN
FITTING
P/N 5110-2002
(SEE NOTE 1)

KYNAR HEAD
P/N 5201-0017
(SEE NOTE 1)
SPARE PARTS LIST
KYNAR
TYPE "T"
TYPE "K"
TYPE "J"
GLASS
INSULATED SECTION
THERMOCOUPLE THERMOCOUPLE THERMOCOUPLE
TUBE
ASSEMBLY
P/N
P/N
P/N
P/N
P/N
1150-0013
1150-0012
5110-2003
1150-0011
5201-0002
"
"
"
"
"
"
"
"
"
"
5110-2004
5201-0001
"
"
"
"
"
"
"
"
"
"
"
"
"
5110-2006
5201-0078
"
"
"
"
"
"
"
"
"
"
"
"
"

NOTES:
1). O-RINGS ARE FACTORY INSTALLED
IN KYNAR HEAD AND KYNAR DRAIN FITTING.
2). LIGHTLY LUBRICATE O-RINGS
WITH SILICONE GREASE BEFORE
ASSEMBLY.

C
REV

10/02/01
DATE

Add 15" Data To Parts List Table


DESCRIPTION
REVISIONS

EM
JK
DWN APVD

Glass Kynar Temperature Sensing Heat Exchanger


Spare Parts List Exploded Assembly
PART NO.
SEE TABLE

7/8" WRENCH
FLATS

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

2-56 X 1/2"
PAN HEAD SCREW

FOR
SAMPLE COOLERS

DRAWN BY

COMPLETE GLASS KYNAR


HEAT EXCHANGER ASSEMBLY

APVD BY

DATE
12/11/95

DRAWING NO

E. Musselman
H. Mitchell

P0148
SCALE
1:1

SIZE

SHEET
2 OF 2

ANNULAR SPACE .060" WIDE


(GAS TO BE DRIED RISES IN THIS SPACE)
INSULATED SECTION

CHILLED SURFACE

PLUG
1/16" NPT

0.300" INSIDE DIAMETER

TOP VIEW

"L"
(SEE TABLE)

CONDENSATE DRAIN
3/8"NPT

SIDE VIEW

DRY GAS OUTLET


1/4" TUBE

B
1). SEE SHEET 2 FOR ASSEMBLY INSTRUCTIONS.
SAMPLE GAS INLET
3/8" TUBE

LENGTH
"L"

5"

10"

15"

UNIVERSAL
QTY OF
GLASS KYNAR HEAT
ANALYZERS
EXCHANGERS PER
EXCHANGER
SAMPLE
COOLER
SAMPLE COOLER
ASSEMBLY P/N
MODELS
520
1
2
530
5200-K050
2
540
570
4
1040
1
1050
2
5200-K010
1060
2
1080
2
1090
4
FGA2
5200-K015
1

10/02/01

REV

DATE

Add 15" Data To Table

EM

DESCRIPTION
REVISIONS

HEAT EXCHANGER GLASS /


KYNAR NON-TEMPERATURE
SENSING OUTLINE
PART NO.
SEE TABLE

FOR
INSTRUMENT

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
EV MUSSELMAN
P0149
H. MITCHELL
DATE
SCALE
SIZE
C
12/11/95
1:1

DRAWN BY
APVD BY

RD

DWN APVD

SHEET
1 OF 2

PLUG
1/6"NPT

EXPLODED GLASS KYNAR HEAT EXCHANGER

#2-120 VITON
O-RING
(SEE NOTE 2)

GLASS TUBE

#2-018 VITON
O-RING
(SEE NOTE 2)

KYNAR INSULATED
SECTION ASSEMBLY

KYNAR HEAD
(SEE NOTE 1)

KYNAR DRAIN
FITTING
(SEE NOTE 1)

C
GLASS KYNAR
HEAT EXCHANGER
ASSEMBLY
LENGTH "L"
P/N
5"
5200-K050
10"
5200-K010
15"
5200-K015

KYNAR
INSULATED
SECTION ASSY
P/N
5110-2003
5110-2004
5110-2006

SPARE PARTS LIST


KYNAR
GLASS
DRAIN
TUBE
FITTING
P/N
P/N
5201-0002
5110-2002
5201-0001
"
"
5201-0078

#2-120
VITON
O-RING
P/N
4904-0004
"
"

#2-018 VITON
KYNAR HEAD
O-RING
P/N
4904-0003
"
"

1/16" NPT PLUG

P/N
5201-0017
"
"

P/N
4951-0066
"
"

7/8" WRENCH
FLATS

COMPLETE GLASS KYNAR


HEAT EXCHANGER ASSEMBLY
NOTES:

10/02/01

REV

DATE

1). O-RINGS ARE FACTORY INSTALLED


IN KYNAR HEAD AND KYNAR DRAIN
FITTINGS

EAM

DESCRIPTION
REVISIONS

PART NO.
SEE TABLE

FOR
INSTRUMENT

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWING NO
EV MUSSELMAN
P0149
H. MITCHELL
DATE
SCALE
SIZE
C
12/11/95
1:1

DRAWN BY
APVD BY

RD

DWN APVD

HEAT EXCHANGER GLASS /


KYNAR NON-TEMPERATURE
SENSING SPARE PART
INSTALLATION

2). LIGHTLY LUBRICATE O-RINGS


WITH SILICONE GREASE BEFORE
ASSEMBLY.

Add 15" Data to Table

SHEET
2 OF 2

R13
C4

U2

TB1

C8

C14

R43

U6

C17 C21
R22

R37

R38
R40

U5

U4

CR5

J1

C24

R42

R28

R11

R8
R44
CR2

C11
C7
R20

CR9
CR1
C28

U9

R30

C23

C25
W1

R24
R27
C10
C12

C20

C27

U8

C26

R31

R23
C13

R17
R19

R32

U3

CR7
CR8

U1

CR4 CR3

R33

R16
R7

J3

R10
R14

R15

C
J5

C16

C15

C19

U7

C18

R35

C6

R3

R34
C22
R25

CR6

R5

R12

C3

R45

B
NOTES: UNLESS OTHERWISE SPECIFIED.
ITEM

DWG.
SIZE

DWG NO.

REVISIONS

UNLESS OTHERWISE INDICATED:


ANGLES = $$p 2$$a
MACHINED FINISH 300/OR BETTER
CHAMFER ANGLE = $$p 3$$a
CHAMFER LENGTH = $$p .015"
DIMENSIONS ENCLOSED IN A
ARE MORE CRITICAL THAN APPLICABLE
INDUSTRY STANDARDS.
ALL DIMENSIONS $$p .005" UNLESS
OTHERWISE SPECIFIED.

APV SYM DATE

ASSY OR B/M

CHANGE

TB C 1/16
98 REMOVED R10
TB B 5/24
95 REFDES U8 TO U6
TB A 2/13
94

FINISH
PER SPEC.

REQ'D

THIRD ANGLE PROJECTIONS

SUPERSCEDES
DWG.

K2

K1

R9

DIMENSIONS IN INCHES.

J6

B
Package To Prevent Handling
And Stocking Damage

C9
R18
R21
R2
R1

Q1
R36
R39
C5
Q2

R6

C1

R26

C2

R4

PART NO.

NAME

PART NO.

NAME

3600-0012

PARTS LOCATION

FOR

MODEL 3040/50/80 COOLER/DISPLAY


SCALE

FULL

DATE

2/13/94

INTERNATIONAL

UNIVERSAL ANALYZERS INC.


SIZE

B
SHEET

DRAWING NO.

E0030

REV.

DWN. CHKD.APVD.ENGR.

MRM

TB TB TB

CONT. ON.

10

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

G
+VREF
R3
10K

+10V

R6

R7
49.9K

TEMP
SET

2K
27L2
7
U3
5 +
4
6

R9
1K
20T

8
U3

R45

R5

2.49K

1K

1
27L2

CR9

J5
1

C7 +15V
.1
MLC
R8
24.9K
R11

CR2

.47uf, POLYE
R12

C28
.1
MLC

C6
.1uf
MLC

J1
1

TP GND
J1
2

+VREF

SET

R16

10K

10K

-VREF
TEMP
CAL
+15V

R20

10K
J3
R21

J1
4

COM

1K
20T
R17
8.66K
R19

J1
3
J1
5

+15V

R23
4.99K
C12
.1uf
MLC

C10

R24
4.99K
C13
.1uf
MLC

100K

8
U4

1
U4
4 27L2
R27
10K

+10mv/ C

-VREF

R31

10K

1 MEG
3

10K

U5
4 27L2

R37

R35

15K

1K

+15V
DRY

8
U5

7
27L2

11
12
13

14
U6

10
4025

8
2
1

U6

R42

R36
24.9K

Q2

WET/DRY

J5
6

100K

U6

C20
.1uf
MLC

7 4025

J5
3

+10V
C19
.1uf
MLC

CR6
LM385 -2.5

+VREF
(+5V)
-VREF
(+2.5V)

C22
.1uf
MLC

W1

+5V

8 2

C26
15uf
TANT

C15
15uf
TANT

B
REV DATE

CHANGE

06/07
99

REV DATE

1
6
7

15uf
TANT

U9
7660S

4
5

-5V
C27
15uf
TANT

NAME:3000 SERIES LOW TEMP


SCHEMATIC
CONTROLLER/DISPLAY
FOR
PART NO. 3600-0033

UNIVERSAL DATE: 2/12/98


ANALYZERS
C E0061
INC.

RELAY
DRIVE

C25

3). ALL DIODES ARE 1N4148.


2). ALL RESISTORS ARE IN OHMS.
1). ALL RESISTORS ARE 1/4W, 1%, 50ppm/ C.
NOTES: UNLESS OTHERWISE SPECIFIED.
CR5

R40
499ohm

R39
100K

U7

C18
.1uf
MLC

2N4401

4025

5
4
3

J5
2

1 MEG
CR4
GRN

78L05

R33
2K

R41
2K

COM

J5
5

C17
.1uf
MLC

CR3
GRN

COOL

+15V

+15V
6

+15V

HOT=HIGH
COOL=LOW

DISPLAY

+15V
C16
15uf
TANT

C21
.1uf
MLC

FROZEN=HIGH
OK=LOW

U1
3 1/2
DIGIT
LCD

J5
4

TEST
C14
.01uf

R38

10

+VREF
+15V

R30

R34
10K

+VREF

26

+VREF

+VREF R32

32

TEST

21
16
22
18
19
2
3
4
5
6
7
8
9
10
11
12
13
14
15
17
20
23
24
25

BP
B3
G3
OSC2 E3
B4
OSC1 D1
CREF C1
B1
CREF A1
U2 F1
7136 G1
E1
A/Z
D2
C2
B2
BUFF
A2
F2
INT
E2
D3
F3
IN+
INPOL
REF+
A3
REFC3
TEST G2

1
G

16
12
8
28
38
39
40
1
29
32
9
3
18
19
20
21
22
23
17
14
15
24
25
26
13
10
31
2
30
11
27

Q1

-100mv/ C

+VREF

J6

39

C2
33
.47uf POLYE
C4
29
.47uf POLYE
R13
28
1.74 MEG
C8
27
.15uf POLYP
31
30
+VREF
36
35
R26
37
10K
R43
15K

R1

2N4401
R2
100K

1
38 OSC3

549K

1 MEG

+10V

+VREF

+VREF +VREF

-VREF

C9
.01uf

47pf
R4

27L2

R28
10K

.47uf
POLYE
2

1 MEG

+VREF
0 C=0V
25 C=+.25V
w.r.t. +VREF

.1uf, MLC
R25

R18

TEMP

C11

100K
C24
.1uf
MLC

TEMP
SENSOR
-

R22
4.99K

R15

+10V

APPROVED

P/S
CONTROL

3
R

40
34

24.9

1 MEG
R14
20K

FM

C1

R44
10K

CR1

C3

C5 -5V
.1uf
MLC

+VREF

+5V

TP PS

+10V

+10mv/ C

DELETE C23
CHANGE

SHEET 1 OF

DWN MRM
CHK TB
APV TB
ENG TB

CR17
CR15

R1

C33
R34
R26

T3

C4

C6

L2

RH1

J4

C2

CR6
CR5
CR8
CR7
CR3
CR1

J3

TB1

F1

TB2

K1

CR22

Q5
J6

R42
R43

TB3

C41

R41

R40

C37

C1

J2
RH7

RH6

RH5

C14

U3

CR4
CR2

C17

T2
RH4

U5

U4

C31

C21

R12
R7

R29
R28
R21

R24
C12

F2

R32
C36 C23
R19
R23

C10

C9

C43

C8

U7

R3

J1

C30

R4

R22

Q4

C35

R8

R38
R13

L1

C13

C29

R30

C11

R31
R39
R11

U6

R6
R37

Q3

R10

C19

R14

C39

R15
R36
C38 C40

C15
J5

C3

RH3

RH2

T1

C25

ITEM

DWG NO.

DWG.
SIZE

PART NO.

REVISIONS

APV SYM DATE

ASSY OR B/M

UNLESS OTHERWISE
INDICATED:
.
ANGLES =_+ 2
MACHINED FINISH 300/OR
BETTER
.
CHAMFER ANGLE_= + 3
CHAMFER LENGTH_= + .015"
DIMENSIONS ENCLOSED IN A
ARE MORE CRITICAL THAN APPLICABLE
INDUSTRY STANDARDS.
ALL DIMENSIONS_ +.005" UNLESS
OTHERWISE SPECIFIED.

NAME

3000 SERIES POWER SUPPLY


CHANGE

FINISH
PER SPEC.

REQ'D

THIRD ANGLE PROJECTIONS

SUPERSCEDES
DWG.

C5

R20

T4

C20

C22

C26

R27

R5

NOTES: UNLESS OTHERWISE SPECIFIED.

DIMENSIONS IN INCHES.

Package To Prevent Handling


And Stocking Damage

R2

R25

CR20
CR19

R16

CR13
CR12

C32
R33
R17

C7

CR10
CR9

CR21

Q1

R18

C24

5
Q2

6
CR11

7
U2

PART NO.

NAME

PARTS LOCATION

3600-0011

FOR

MODEL 3040/50/80 POWER SUPPLY


SCALE

FULL

DATE

2/23/94

INTERNATIONAL

UNIVERSAL ANALYZERS INC.


SIZE

B
SHEET

DRAWING NO.

E0034

REV.

DWN. CHKD.APVD.ENGR.

MRM

CONT. ON.

10

APPROVED

LINE

F1

NEUT/
LINE 2

RH1

L1

12 AMP
250V

R3
422K

C1
.22
250V

C3
4700pf
250V

C5
4700pf
250V

CL30

C2
.22
250V

TB1

115V
230V

GND

T1

1
3

J1

WHT

1
3

J1

GRN

F2
6 AMP
250V

2
2

J1

BLK
DRAIN PUMP

TB2
NC1 1

J2
J2

TB2
2

TB2
NO1 3
TB2
NC2 4

J2
J2
J2

TB2
NO2 6

115V
230V

RH3
130V
MOV

115V

CR3

J4

CR4

JUMPER

J2

4
5
6
3

RELAY CONTACTS
REQUIRE 24VDC
200mA MAX.
RESISTIVE LOAD
IN CLASS I, GROUP A, B, C, OR D
DIVISION 2 AREAS

J3
J3
J3
J3
J3

BLU

+15V

J5
4

+15V

J5
5

COM

+VREF

J5
2

+VREF

12

J5
6

WET/DRY

1
C37
1000pf
MLC

GRN

BLOWER

CR22
1N4148

MOISTURE
SENSOR

7
5

J6

U7
LM1830N

C41
.1
MLC

1
13
9

2
C43
15
25V
TANT

11

SCHEMATIC
FOR
E

10

4/12
01
9/11
00

REV DATE

REVISION PER
D
SHEET 2
ADD GROUP A&B
C
CLASSIFICATION NOTE

10

TB3
2

Q5
2N4401

14

RED

TB3
1

3). ALL RESISTORS ARE IN OHMS.


2). ALL CAPACITORS ARE IN MFD.
1). ALL RESISTORS ARE 1/4W, 1%, 50ppm/oC.
NOTES: UNLESS OTHERWISE SPECIFIED.

BLK

J6

-160VDC

WHT

R40
71.5K
OHMS

R43
24.9K

C8
15
25V
TANT

6
12
PL5.0-16

JUMPER
SAMPLE PUMP

10K

0VDC

+15V

U2
C7
470
50VDC

10

N/C

R42

7815

4X 1N4005

R41
75
1/2W
3

CR2

CR1

J4
3

GRN

K1

J5

C4
680
200VDC

J4
2

BLK

RH6
250V
MOV
RH7
250V
MOV

TB2
ARM2 5

RELAY
DRIVE

CR21
FPIL
60

+160VDC
R1
100K
1/2W

R2
C6
100K
680
1/2W
200VDC

+15V

AC

WHT

RH4
250V
MOV
RH5
250V
MOV

ARM1

RH2
130V
MOV

TB1

5
J4
6
7

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

FM

TB1

CHANGE

6/7
99
10/15
96
5/16
94

REV DATE

R40 FROM 20K TO


71.5K OHMS
C19 VALUE CHANGE
RELAY RATINGS
CHANGE

NAME:500

WATT POWER SUPPLY

PART NO.3600-0011

UNIVERSAL DATE: 10/4/93


ANALYZERS
C E0024
INC.
SHEET 1 OF

DWNMRM
CHK
APV
ENG

10

1
G

G
+15V
2

+VREF

1
R4

1 MEG
C35

F
+15VDC
+VREF
+24V

C15

R5
2.15K

+24V R27
ADJ 100
1T

P/S 1
CONTROL

R12
499

R30

Q3

24.9K
R28
24.9K

J5

R29
49.9K

2N4401

.1
50V
MLC
R11
10K
C36
.1
MLC

+15VDC

16
R39
10K
2

R32
24.9K

C9
15
25V
TANT

C30
15
25V
TANT

C10
15
25V
TANT

C12
.1
MLC

4.99K
C17
0.1
MLC

20K
C38
0.1
MLC

C13
.1
MLC

+VREF

C14
.1
MLC

1N4148
T2

U3
3525A

50KHz
+15V
11

R21
100

R24
100

12

8
2 U4

CR10

R16

1N4148

24.9

R33
49.9

6
7

C29

CR12
1N5245
R17
15V
10K
CR13
1N5245
15V

Q1
IRFP451R

8
1
2 U5
5 4

R20
120
5W

C22

R36

R38

54.9K
C40

.01
MLC

470pf
2KV

L2

1+

FEP30DP

8T

R18
D.S.

8T

C24
D.S.

18 uH

35T

3+
CR11B

+24V

R22

CR17

R25

2.43

1N4148

24.9

R34
49.9

4420

CR19
1N5245
R26
15V
10K
CR20
1N5245
15V

Q2
IRFP451R

1-

C33
2200pf
MLC

2+

3-

4+

2C25
4700
35V

4-

C26
4700
35V

-160VDC

NOTICE
REVISION OF THIS DOCUMENT
MUST BE AUTHORIZED BY
APPROPRIATE APPROVAL AGENCIES

B
SCHEMATIC
FOR

E
G
F

D.S.

FEP30DP

1N4148
6
7

CR11A

T3
C32
2200pf
MLC

CR15

14T

R13

1
5 4
4420 50V
POLYE

11T

4). D.S. COMPONENTS R15, R18, R38, AND C24


ARE NOT PRESENT.
3). ALL RESISTORS ARE IN OHMS.
2). ALL CAPACITORS ARE IN MFD.
o
1). ALL RESISTORS ARE 1/4W, 1%, 50ppm/ C.
NOTES: UNLESS OTHERWISE SPECIFIED.

TLC3702

15
TANT
25V

11T

+15V
14

5+
U6
6-

100K
D.S.
C39

CR9
3

FM

R37
15
1/2W

+160VDC

APPROVED

C11
.1
MLC

1
R14
24.9K

10

4X 1N4148
CR7
CR8

R8

13

8
2.49K

50T

R7

R15

10

Q4
2N4401

1T

CR5

R6
1K
1/2W

47pf
MLC

CR6

15

C19
330pF
100V
MLC
9
C21
C31
2200pf MLC
1
R19
6
50V
POLYE 11.5K
C23
R10
5
34K
R23
1000pf 5.11
100V
7
MLC

R31

C20
1
400VDC
POLYP

TLC3702

T4

0VDC

U6
3+

4/12
01
9/11
00

REV DATE

C35 FROM 1000pf TO


47pf
REVISION PER
SHEET 1

CHANGE

D
C

6/7
99
10/15
96
5/16
94

REV DATE

R40 FROM 20K TO


71.5K OHMS
C19 VALUE CHANGE
RELAY RATINGS
CHANGE

NAME:

500 WATT POWER SUPPLY

PART NO. 3600-0011

UNIVERSAL DATE: 10/4/93


ANALYZERS
C E0024
INC.
SHEET 2 OF

DWNMRM
CHK
APV
ENG

22-3/4"
CARRYING HANDLE
15-1/4"

WET GAS
SAMPLE INLET
DRY SAMPLE GAS
TO ANALYZER

3/8" TUBING
FITTING
SAMPLE COOLER

SAMPLE PUMP HEAD

DOUBLE HEAD
ADI MINI DIA-VAC
PUMP
26-1/2"
FLOW METER

VISIBLE FILTER
WITH WATER
CARRY OVER
SENSOR

COMPRESSED AIR HEAD

8
,57
65
5,4

78
5,465,5

EDUCTOR

B
AIR FILTER
WATER CARRY
OVER SENSOR
CONNECTOR
FOUR LEG STAND
(ONE LEG NOT VISIBLE IN THIS VIEW)

ISOMETRIC VIEW

FRONT VIEW

12-3/8"

04/17/98

ADD MOUNTING FEET/PATENT NUMBER

REV

DATE

DESCRIPTION
REVISIONS

EAM

3000 SERIES PORTABLE SAMPLE COOLER


ACCESSORY OUTLINE FOR SYSTEMS WITH
EDUCTOR
PART NO.

FOR

UNIVERSAL ANALYZERS INC.


1701 South Sutro Terrace
Carson City, Nevada 89706 USA

DRAWN BY
APVD BY

DATE
12/20/93

JK

DWN APVD

DRAWING NO
HARVEY MITCHELL
P0084
TED BARBEN
SCALE
SIZE
D
1:1

SHEET
1 OF 1

22-3/4"
CARRYING HANDLE
13-5/8"
WET GAS
SAMPLE INLET

DRY SAMPLE GAS


TO ANALYZER

3/8" TUBING
FITTING
SAMPLE COOLER

SINGLE HEAD
ADI MINI
DIA-VAC
PUMP

26-1/2"

FLOW METER

VISIBLE FILTER
WITH WATER
CARRY OVER
SENSOR

8
5,465,57

5,4

65

,57

PERISTALTIC
PUMP HEADS

WATER
CARRY OVER
SENSOR
CONNECTOR
FOUR LEG STAND
(ONE LEG NOT VISIBLE IN THIS VIEW)

ISOMETRIC VIEW

FRONT VIEW

12-3/8"

04/23/98

ADD LEG/FEET/NEW HEAD DESIGN/PAT #

REV

DATE

DESCRIPTION
REVISIONS

EAM

3000 SERIES PORTABLE SAMPLE COOLER


ACCESORY OUTLINE FOR SYSTEMS WITH
PERISTALTIC PUMP
PART NO.

FOR

UNIVERSAL ANALYZERS INC.


1701 S. Sutro Terrace
Carson City, Nevada 89706 USA

DRAWN BY
APVD BY

DATE
12/20/93

DRAWING NO
HARVEY MITCHELL
P0085
TED BARBEN
SCALE
SIZE
D
1:1

JK

DWN APVD

SHEET
1 OF 1

4
1

D
*

8
7

8
6

5,4

65

,57

7
5

ASSEMBLED FRONT VIEW

(ACCESSORIES NOT SHOWN FOR CLARITY)

VIEW AS SEEN LOOKING AT BACK OF SAMPLE COOLER


(ACCESSORIES NOT SHOWN FOR CLARITY)

BILL OF MATERIALS FOR


LEG MOUNT KIT
P/N 9515-0019
ITEM
1
2
3
4
5
6
7
8

*
*

QTY
2
1
1
2
3
4
3
3

DESCRIPTION
LEG SINGLE
LEG DUAL
PLUG ROUND PLASTIC
PLUG SQUARE PLASTIC
SCREW 1/4" x 1-1/2" S.S.
SCREW #6 x 1-1/4" S.S.
WASHER 1/4" FLAT S.S.
NUT 1/4" S.S.

A 04/28/98 ADD FOOT PADS REVISE BOM TABLE


DESCRIPTION
REV DATE

EAM JK
DWN APVD

REVISIONS

P/N
6006-0001
6006-0002
4919-0007
4919-0006
4704-0047
4704-0048
4710-0001
4711-0006

LEG MOUNTING KIT


INSTALLATION FOR 3000
SERIES SAMPLE COOLERS

PART NO.
9515-0019

FOR
3000 SERIES COOLERS

UNIVERSAL ANALYZERS INC.

*NOTE: ITEMS 3 AND 4 ARE FACTORY INSTALLED

1701 South Sutro Terrace


Carson City, Nevada 89706 USA

DRAWN BY
APVD BY

DATE
02/04/1994

DRAWING NO

H. MITCHELL
T. BARBEN II

P0089
SCALE
NTS

SIZE
E

SHEET
1 OF 1

INSTALLATION:
1). INSTALL ITEMS #1 INTO BOTTOM OF
BOTH HEAT EXCHANGERS.

2). INSTALL ITEMS #2 & #3 INTO ITEMS #1


WITH ITEM #2 IN THE FRONT HEAT
EXCHANGER.
3). INSTALL ITEMS #4 INTO TOP OF LIQUID
DRAINS TYPICAL.

FRONT

4). INSTALL ITEMS #5 ONTO ITEMS #2 & #3.

PARTS TABLE
ITEM#
DESCRIPTION
QTY
P/N
1
1/2" TUBE X 3/8"NPT CONNECTOR 2 4954-0058
2
1/2" 316 S.S. X 5" BENT TUBE
1 9011-0004
3
1/2" 316 S.S. X 3" STRAIGHT TUBE
1 9011-0004
4
1/2" TUBE X 3/4"NPT CONNECTOR 2 4954-0135
5
FLOAT DRAIN VALVE
2 4955-0001

4
3

5
DESCRIPTION

REV DATE

DWN APVD

REVISIONS

FLOAT DRAIN INSTALLATION


LAYOUT

PART NO.
9515-0100

UNIVERSAL ANALYZERS INC.

RIGHT SIDE VIEW

1701 South Sutro Terrace


Carson City, Nevada 89706 USA

DRAWN BY
APVD BY

DRAWING NO

E. Musselman
T. Barben II

DATE
09/15/1999

FOR
3020, 3050, 3080

P0360
SCALE
NTS

SIZE
E

SHEET
1 OF 1

UNIVERSAL ANALYZERS INC.


MOISTURE CONVERSION TABLE
DEW POINT,
DEGREES C.
-100
-50
-20
-10
-5
-4
-3
-2
1
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90

% WATER BY VOLUME % WATER BY WEIGHT


AT SATURATION
AT SATURATION
0.00000139
0.00000081
0.00388
0.00241
0.102
0.0633
0.256
0.1596
0.396
0.229
0.432
0.268
0.469
0.291
0.510
0.317
0.555
0.345
0.602
0.364
0.649
0.404
0.696
0.433
0.750
0.466
0.803
0.499
0.861
0.535
0.922
0.573
0.988
0.614
1.06
0.658
1.13
0.702
1.21
0.753
1.29
0.802
1.38
0.860
1.48
0.920
1.58
0.980
1.68
1.044
2.31
1.433
3.13
2.004
4.19
2.64
5.55
3.54
7.28
4.67
9.46
6.12
12.2
8.0
15.5
10.3
19.7
13.3
24.4
16.8
30.7
21.7
38.0
27.8
46.7
35.6
57.2
45.7
69.2
58.4

WATER VAPOR
PRESSURE, mm. Hg
0.0000099
0.0295
0.776
1.950
3.014
3.280
3.569
3.880
4.223
4.579
4.937
5.294
5.70
6.10
6.54
7.01
7.51
8.05
8.58
9.15
9.8
10.5
11.2
12.0
12.8
17.5
23.8

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-A18 REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 1 PROCESS GAS

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

SINTERED METAL FILTER ELEMENT

203495

050-11-20M

BALSTON

0.9

MICROFIBER FILTER ELEMENT

203497

050-11-BQ

BALSTON

0.4

PRESSURE GAUGE

205388

25.500-60-PSI/BAR1/4-LM

NOSHOK

0.8

SOLENOID VALVE

206201

EV8314G301-24VDC

ASCO

0.9

FLOW SWITCH

203427

FS6204CV-1MXSPECIAL

AUTOFLOW

0.8

FLOWMETER 5LPM

024A0037

VFA-21-SSV

DWYER

0.9

FLOWMETER 3.8LPM

203564

KING

1.36

FLOWMETER 42LPM

205421

KING

1.36

PRESSURE GAUGE

205457

NOSHOK

0.8

74C-123G081323810
74C-234G041323810
25.510-30-PSI/BAR1/4-CBM

10 REGULATOR, 0-15 PSIG

025A0003

960-175-203

BELLOFRAM

1.3

11 INLINE FILTER ELEMENT

205973

SS-8F-K4-2

SWAGELOK

0.4

12 PERISTALTIC PUMP TUBING 5 FT

200598

9216-0002

UNIVERSAL ANALYZERS

0.3

13 CHILLER 12 AMP FUSE

200591

3010-0006

UNIVERSAL ANALYZERS

0.2

14 CHILLER 6 AMP FUSE

200590

3010-0005

UNIVERSAL ANALYZERS

0.2

15 O-RING HEAT EXCHANGER

204355

4904-0013

UNIVERSAL ANALYZERS

0.1

16 FILTER ELEMENT WCOF

200587

4980-0006

UNIVERSAL ANALYZERS

0.5

17

FORM 177 REV. C

Page 1 of 2

1
1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-A18 REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 1 PROCESS GAS

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

18
19 FUSE KIT

2500996

SERVOMEX

20 DETECTOR SCRUBBER

S1210939

SERVOMEX

21 CHOPPER BOX SCRUBBER

S2000512

SERVOMEX

22 O-RINGS

S2500982A

SERVOMEX

23
24
25
26
27
28
29
30
31
32
33
34

FORM 177 REV. C

Page 2 of 2

0.4

0.1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-A33.1 REV 1

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 2.1 BUSTLE GAS

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

SINTERED METAL FILTER ELEMENT

203495

050-11-20M

BALSTON

0.9

MICROFIBER FILTER ELEMENT

203497

050-11-BQ

BALSTON

0.4

PRESSURE GAUGE

205388

25.500-60-PSI/BAR1/4-LM

NOSHOK

0.8

SOLENOID VALVE

206201

EV8314G301-24VDC

ASCO

0.9

FLOW SWITCH

203427

FS6204CV-1MXSPECIAL

AUTOFLOW

0.8

FLOWMETER 5LPM

024A0037

VFA-21-SSV

DWYER

0.9

FLOWMETER 3.8LPM

203564

KING

1.36

FLOWMETER 19LPM

204644

KING

1.36

PRESSURE GAUGE

205457

NOSHOK

0.8

74C-123G081323810
74C-214G021423810
25.510-30-PSI/BAR1/4-CBM

10 REGULATOR, 0-15 PSIG

025A0003

960-175-203

BELLOFRAM

1.3

11 INLINE FILTER ELEMENT

205973

SS-8F-K4-2

SWAGELOK

0.4

12 PERISTALTIC PUMP TUBING 5 FT

200598

9216-0002

UNIVERSAL ANALYZERS

0.3

13 CHILLER 12 AMP FUSE

200591

3010-0006

UNIVERSAL ANALYZERS

0.2

14 CHILLER 6 AMP FUSE

200590

3010-0005

UNIVERSAL ANALYZERS

0.2

15 O-RING HEAT EXCHANGER

204355

4904-0013

UNIVERSAL ANALYZERS

0.1

16 FILTER ELEMENT WCOF

200587

4980-0006

UNIVERSAL ANALYZERS

0.5

17

FORM 177 REV. C

Page 1 of 2

1
1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-A33.1 REV 1

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 2.1 BUSTLE GAS

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

18
19 FUSE KIT

2500996

SERVOMEX

20 DETECTOR SCRUBBER

S1200922

SERVOMEX

21 CHOPPER BOX SCRUBBER

S2000511

SERVOMEX

22 O-RINGS

S2500982A

SERVOMEX

23
24
25
26
27
28
29
30
31
32
33
34

FORM 177 REV. C

Page 2 of 2

0.4

0.1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-A37 REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 3 COOLING GAS

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

SINTERED METAL FILTER ELEMENT

203495

050-11-20M

BALSTON

0.9

MICROFIBER FILTER ELEMENT

203497

050-11-BQ

BALSTON

0.4

PRESSURE GAUGE

205388

25.500-60-PSI/BAR1/4-LM

NOSHOK

0.8

SOLENOID VALVE

206201

EV8314G301-24VDC

ASCO

0.9

FLOW SWITCH

203427

FS6204CV-1MXSPECIAL

AUTOFLOW

0.8

FLOWMETER 5LPM

024A0037

VFA-21-SSV

DWYER

0.9

FLOWMETER 3.8LPM

203564

KING

1.36

FLOWMETER 42LPM

205421

KING

1.36

PRESSURE GAUGE

205457

NOSHOK

0.8

74C-123G081323810
74C-234G041323810
25.510-30-PSI/BAR1/4-CBM

10 REGULATOR, 0-15 PSIG

025A0003

960-175-203

BELLOFRAM

1.3

11 INLINE FILTER ELEMENT

205973

SS-8F-K4-2

SWAGELOK

0.4

12 PERISTALTIC PUMP TUBING 5 FT

200598

9216-0002

UNIVERSAL ANALYZERS

0.3

13 CHILLER 12 AMP FUSE

200591

3010-0006

UNIVERSAL ANALYZERS

0.2

14 CHILLER 6 AMP FUSE

200590

3010-0005

UNIVERSAL ANALYZERS

0.2

15 O-RING HEAT EXCHANGER

204355

4904-0013

UNIVERSAL ANALYZERS

0.1

16 FILTER ELEMENT WCOF

200587

4980-0006

UNIVERSAL ANALYZERS

0.5

17

FORM 177 REV. C

Page 1 of 2

1
1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-A37 REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 3 COOLING GAS

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

18
19 FUSE KIT

2500996

SERVOMEX

20 DETECTOR SCRUBBER

S1200922

SERVOMEX

21 CHOPPER BOX SCRUBBER

S2000511

SERVOMEX

22 O-RINGS

S2500982A

SERVOMEX

23
24
25
26
27
28
29
30
31
32
33
34

FORM 177 REV. C

Page 2 of 2

0.4

0.1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-A75 REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 5 PROCESS GAS

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

SINTERED METAL FILTER ELEMENT

203495

050-11-20M

BALSTON

0.9

MICROFIBER FILTER ELEMENT

203497

050-11-BQ

BALSTON

0.4

PRESSURE GAUGE

205388

NOSHOK

0.8

FLOWMETER 42 LPM

205458

KING

1.36

FLOWMETER 400CCM

205459

KING

1.36

PRESSURE GAUGE

205388

NOSHOK

0.8

INLINE FILTER ELEMENT

205973

SS-8F-K4-2

SWAGELOK

0.4

HEATER

205390

400/100 T3 JB TS40

INTERTEC

2.2

PERISTALTIC PUMP TUBING 5 FT

200598

9216-0002

UNIVERSAL ANALYZERS

0.3

10 CHILLER 12 AMP FUSE

200591

3010-0006

UNIVERSAL ANALYZERS

0.2

11 CHILLER 6 AMP FUSE

200590

3010-0005

UNIVERSAL ANALYZERS

0.2

12 O-RING HEAT EXCHANGER

204355

4904-0013

UNIVERSAL ANALYZERS

0.1

13 FILTER ELEMENT WCOF

200587

4980-0006

UNIVERSAL ANALYZERS

0.5

25.500-60-PSI/BAR1/4-LM
74C-234G031325810
74C-103G041425710
25.500-60-PSI/BAR1/4-LM

14
15
16
17

FORM 177 REV. C

Page 1 of 2

1
1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-A75 REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

EQUIPMENT DESCRIPTION:

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

TYPE OR MODEL
NUMBER

MIDREX PURCHASE ORDER:

MFG.

18
19

FORM 177 REV. C

Page 2 of 2

NAME

STREAM 5 PROCESS GAS


15036-3843A-M-IN72

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-D18.1 & AIT- D19 REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 8.1 SEAL GAS

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

SINTERED METAL FILTER ELEMENT

203495

050-11-20M

BALSTON

0.9

MICROFIBER FILTER ELEMENT

203497

050-11-BQ

BALSTON

0.4

PRESSURE GAUGE

205388

25.500-60-PSI/BAR1/4-LM

NOSHOK

0.8

SOLENOID VALVE

206201

EV8314G301-24VDC

ASCO

0.9

FLOW SWITCH

203427

FS6204CV-1MXSPECIAL

AUTOFLOW

0.8

FLOWMETER 5LPM

024A0037

VFA-21-SSV

DWYER

0.9

FLOWMETER 3.8LPM

203564

KING

1.36

FLOWMETER 13LPM

205419

KING

1.36

PRESSURE GAUGE

205457

NOSHOK

0.8

74C-123G081323810
74C-214G021323810
25.510-30-PSI/BAR1/4-CBM

10 REGULATOR, 0-15 PSIG

025A0003

960-175-203

BELLOFRAM

1.3

11 INLINE FILTER ELEMENT

205973

SS-8F-K4-2

SWAGELOK

0.4

12 PERISTALTIC PUMP TUBING 5 FT

200598

9216-0002

UNIVERSAL ANALYZERS

0.3

13 CHILLER 12 AMP FUSE

200591

3010-0006

UNIVERSAL ANALYZERS

0.2

14 CHILLER 6 AMP FUSE

200590

3010-0005

UNIVERSAL ANALYZERS

0.2

15 O-RING HEAT EXCHANGER

204355

4904-0013

UNIVERSAL ANALYZERS

0.1

16 FILTER ELEMENT WCOF

200587

4980-0006

UNIVERSAL ANALYZERS

0.5

17 FLOWMETER 400CCM

202178

74C-103G041423710

KING

1.36

FORM 177 REV. C

Page 1 of 2

1
1

2
1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-D18.1 & AIT- D19 REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 8.1 SEAL GAS

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

18
19 FUSE KIT

2500996

SERVOMEX

20 DETECTOR SCRUBBER

S1210939

SERVOMEX

21 CHOPPER BOX SCRUBBER

S2000512

SERVOMEX

22 O-RINGS

S2500982A

SERVOMEX

0.1

23 FUSE KIT

204629

SERVOMEX

0.4

24 RESTRICTOR KIT

S1910980

SERVOMEX

25
26
27
28
29
30
31
32
33
34

FORM 177 REV. C

Page 2 of 2

0.4

1
1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-H26 REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 9 DUST COLLECTOR

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

SINTERED METAL FILTER ELEMENT

203495

050-11-20M

BALSTON

0.9

MICROFIBER FILTER ELEMENT

203497

050-11-BQ

BALSTON

0.4

PRESSURE GAUGE

205387

25.500-30/15PSI/BAR-1/4-LM

NOSHOK

0.8

SOLENOID VALVE

206201

EV8314G301-24VDC

ASCO

0.9

FLOW SWITCH

203427

FS6204CV-1MXSPECIAL

AUTOFLOW

0.8

FLOWMETER 5LPM

024A0037

VFA-21-SSV

DWYER

0.9

FLOWMETER 3.8LPM

203564

KING

1.36

FLOWMETER 42LPM

205421

KING

1.36

PRESSURE GAUGE

205457

NOSHOK

0.8

74C-123G081323810
74C-234G041323810
25.510-30-PSI/BAR1/4-CBM

10 REGULATOR, 0-15 PSIG

025A0003

960-175-203

BELLOFRAM

1.3

11 INLINE FILTER ELEMENT

205973

SS-8F-K4-2

SWAGELOK

0.4

12 PERISTALTIC PUMP TUBING 5 FT

200598

9216-0002

UNIVERSAL ANALYZERS

0.3

13 CHILLER 12 AMP FUSE

200591

3010-0006

UNIVERSAL ANALYZERS

0.2

14 CHILLER 6 AMP FUSE

200590

3010-0005

UNIVERSAL ANALYZERS

0.2

15 O-RING HEAT EXCHANGER

204355

4904-0013

UNIVERSAL ANALYZERS

0.1

16 FILTER ELEMENT WCOF

200587

4980-0006

UNIVERSAL ANALYZERS

0.5

17

FORM 177 REV. C

Page 1 of 2

1
1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


AIT-H26 REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 9 DUST COLLECTOR

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

18
19 FUSE KIT

2500996

SERVOMEX

20 DETECTOR SCRUBBER

S1210939

SERVOMEX

21 CHOPPER BOX SCRUBBER

S2000512

SERVOMEX

22 O-RINGS

S2500982A

SERVOMEX

23
24
25
26
27
28
29
30
31
32
33
34

FORM 177 REV. C

Page 2 of 2

0.4

0.1

RECOMMENDED SPARE & CONSUMABLE PARTS LIST


MASS SPEC REV 2

EQUIPMENT TAG NO:


VENDOR NAME:

________SERVOMEX_____________

AUTHORIZED SIGNATURE:______________________Date:__________

PART DESCRIPTION

REF. DWG.
OR PART

STREAM 2.2, 4, 8.2, 10.1-10.7

EQUIPMENT DESCRIPTION:

TYPE OR MODEL
NUMBER

15036-3843A-M-IN72

MIDREX PURCHASE ORDER:

MFG.

NAME

RECOMMENDED
APPROX.
Delivery
SPARES QUANTITY
UNIT
Time
WEIGHT
2
(Wks) START
CAPITAL
(Kg.)
UP
YEARS

SINTERED METAL FILTER ELEMENT

203495

050-11-20M

BALSTON

0.9

10

MICROFIBER FILTER ELEMENT

203497

050-11-BQ

BALSTON

0.4

10

PRESSURE GAUGE

205388

NOSHOK

0.8

FLOWMETER 19 LPM

204644

KING

1.36

FLOWMETER 400CCM

202178

KING

1.36

PRESSURE GAUGE

205388

NOSHOK

0.8

INLINE FILTER ELEMENT

205973

SS-8F-K4-2

SWAGELOK

0.4

SOLENOID VAVLE

206201

EV8314G301-24VDC

ASCO

PERISTALTIC PUMP TUBING 5 FT

200598

9216-0002

UNIVERSAL ANALYZERS

0.3

10 CHILLER 12 AMP FUSE

200591

3010-0006

UNIVERSAL ANALYZERS

0.2

11 CHILLER 6 AMP FUSE

200590

3010-0005

UNIVERSAL ANALYZERS

0.2

12 O-RING HEAT EXCHANGER

204355

4904-0013

UNIVERSAL ANALYZERS

0.1

13 FILTER ELEMENT WCOF

200587

4980-0006

UNIVERSAL ANALYZERS

0.5

14 FLOWMETER

205421

74C-234G041323810

KING

1.36

25.500-60-PSI/BAR1/4-LM
74C-214G021423810
74C-103G041423710
25.500-60-PSI/BAR1/4-LM

15
16
17

FORM 177 REV. C

Page 1 of 1

2
4

4
2

4
4

Subject:
Project:
Location:
Order #
Date:

Calibration Fluid Calculations


Midrex
Angul, India
631032096
3/2/2011

t
Vc

Terminology
= cylinder life expectancy (days) (per single cylinder)
= cylinder volume (liters) (1)

Pf

= max filing pressure of cylinder (Bar g)

Pd

= delivery pressure of sample (Bar g)

Qc

= sample flow rate (per calibration) (liters/day)(2)

Q1

= sample leak rate (liters/day)

Calculations
Calibration
Checks/day

Frequency

Volume
required
per
Calibration
Check (3)

10

Liters/Day

Vc(Pf-Pd)

t=

Pd(Qc-Q1)
Vc =

43.8

Pf =

103

Pd =

Qc =

10

Q1 =

0.05

t=

222.30151
Days
7.4100503 Months
Note1 :
Note2 :
Note3 :

Assuming 9 in Diameter X 56 Inch Tall Cylinder for


volume calculations
Assuming 1 calibration check/day required
5 LPM flow to sample probe @ 120
seconds/calibration check per day

Air Dimensions Incorporated


1371 West Newport Center Drive, Suite # 101
Deerfield Beach, FL, 33442
800.423.6464 / 954.428.7333 ph , 954.360.0987 fax
info@airdimensions.com www.airdimensions.com
Standard DIA-VAC

MAINTENANCE AND DISASSEMBLY INSTRUCTIONS


A. General Operations Characteristics
1. The Dia-Vac normally runs at 130 to 150 degrees F. It is not an indication of trouble if the
outer surfaces of the Dia-Vac Pump or motor are hot to the touch.
2. To check pumping efficiency, employ suitably damped gauges connected so as to dead-end
either pressure or vacuum.
NOTE: Check each separately, One or the other port must be open during this test. Use 0-60 PSI
pressure gauge and 0-30 inch hg. vacuum gauge, (or mercury manometer.)
Minimum pressure should be 40 PSIG for the .222 eccentric
Minimum vacuum should be 22 inches hg. For the .222 eccentric

3. The Dia-Vac normally runs quietly, especially when both pressure and vacuum ports are
connected into a closed system. An obvious knock or rattle could indicate a problem. Check
through "Disassembly and Inspection Procedures" with particular attention to the tightness of all
screw fasteners.
4. All bolts and screws except A19005 and A19006 should have one drop of Loctite Removable 242 (or equivalent) on threads before tightening.
B. Maintenance Procedures
1. Motor oiling
No oiling or other lubrication addition is necessary at all with the Dia-Vac pumps. Please note
the Air driven GAST motor does require SAE 10 oil lubrication for proper operation.
2. Diaphragm Replacement (See also Maintenance Procedure Below)
a.) Standard EPDM (part 4002 or complete repair kit 11009).
b.) Teflon coated EPDM (part 4001 or complete repair kit 11005).
c.) Viton (part 4003 or kit complete repair kit 11007).
d.) Teflon/Viton (part 4009 or kit 1 complete repair kit 11006)
e.) All-Teflon (part 4009 or kit complete repair kit 11011)
The degree of usage and condition of operating temperatures or pressure will determine the rate
of replacement of part or parts. For heavy loads (25-75 PSI) and constant operation the diaphragm should be inspected at least every six months.
(0707)

"We Pass Your Gas"

For lighter loads (0-15 PSI or up to maximum vacuum) the diaphragm may operate successfully
for a year or more. The corrosive content of the gas media being pumped can effect the recommended inspection and replacement cycle of the diaphragm.
*Diaphragms require close precision tolerance, therefore only ADI diaphragms should be
used as replacements.
C. Disassembly of Head Section and Service Diaphragm
1. Remove head section by unscrewing the four large bolts. A flat-bladed screw driver may be
needed to gently pry the head free of the service diaphragm. **If you have Teflon coating on
the heads use caution not to scratch the surface.
2. The valve body can then be removed by unscrewing the two smaller screws (also accessible
on the top of the head section). This part may be freed by gently tapping on these two screws
after they have been loosened about three or four turns. When the valve body is removed, check
all internal surfaces for any accumulation of dirt. The two valve discs can be wiped clean and
replaced as long as they appear unaffected by usage. The valve gasket can be easily removed and
should be inspected. As a matter of good practice, the valve discs and valve gasket should be
replaced during any routine maintenance check of the head section. A once a year routine procedure is recommended.
3. The service diaphragm is secured by the single screw in its center. Remove this screw with a
5/32" Allen wrench. The diaphragm and its clamping plate should be easily lifted off. Some
slight adherence to the metal may occur if the diaphragm has been in use for a long period. Before inserting the diaphragm cap screw (19007) apply a small amount of Teflon paste or a Teflon Washer seal (23001), (do not use Teflon tape), then tighten the cap screw.
NOTE: When replacing the service diaphragm, be sure the four projecting studs of the base casting are properly located in the four outer holes provided in the diaphragm before the part is
clamped in place. Be sure the diaphragm plate is firmly replaced with its center screw.
D. Disassembly of the Connecting Rod
1. Remove head section and service diaphragm as described in (C) above.
2. Remove the front plate from the face of the base casting by removing the four retaining
screws.
3. Using a hex socket wrench, remove the hex head bolt on the connecting rod top surface. This
will release the connecting rod cap (3001) which may then be lifted off.
4. The connecting rod assembly on single head units, including the counterweight, is held in
place by the counterweight screw. This can be loosened by a 3/16" Allen wrench. The connecting rod assembly may then be slid off the motor shaft. On the dual and quad head units the assembly is held in place by the set screw. This can be loosened by a 1/8" Allen wrench and the
assembly may then be slid off the motor shaft.

(0707)

"We Pass Your Gas"

E. Replacement of the Connecting Rod


1. Replace the connecting rod assembly on the motor shaft, taking care to align set screw or
counterweight screw as squarely as possible on the flat of the motor shaft.
2. Replace the connecting rod cap and connecting rod bolt. Tighten this assembly firmly with the
hex socket wrench.
3. Put a new Service Diaphragm in place taking care to locate it over the four projecting studs on
the top face of the base casting.

F. Related Torque Values


1. Head bolts - 110 inch pounds.
2. Valve body screws and Diaphragm plate screws - 70 inch pounds.
3. Connecting rod bolt - 150 inch pounds.

G. Motor Wiring
1. Please refer to wiring diagram located on motor.
H. Motor Amperage
1. Please refer to data label located on motor for running amperage. Startup amps are 3-4 times
the stated running amps on the motor.

Teflon is a Registered Trademarks of E.I. Du Pont

Loctite is a Registered Trademark of Henkel Technologies Inc.

Dia-Vac is a Registered Trademark of Air Dimensions Inc.

GAST is a Registered Trademark of GAST Mfg. Inc.

Viton is a Registered Trademark of E.I. Du Pont


(0707)

"We Pass Your Gas"

MICRO FLOW SWITCH


Adjustable Gas Series - Range 30 sccm to 20 slpm

FS6200 Series Micro Flow Switch


Available with 1/8 and 1/4 NPT Connections
316L Stainless Steel Construction

Hazardous Location Service


CSA Certified
Class I, Div 1&2
Groups A, B, C, D
Class II, Div 1&2
Groups E, F, G
Class III
Cenelec Certified
EExdII CT3
Class I, Zone I, IIC
CE Compliant
Fully Field Adjustable
Operates in Any Orientation
Operating Temperature
-40F to +220F
(-40C to +105C)
Operating Pressure
3000 psig (207 bar)
316 Stainless Steel construction
Shock and Vibration Resistant
Contamination Resistant
1/8 & 1/4 NPT Connections

FS6200 Series
Micro Flow Switch
available with 1/8 and
1/4 NPT Connections
The FS6200 Series Micro Flow Switch is designed to precisely detect increasing or
decreasing flow rate of gases in critical processes. Typical applications include process analyzer
sample conditioning systems, critical purge monitors, excess flow safety, and oxygen and speciality gas systems. The linear, high resolution adjustment allows for precise settings. The miniature compact construction is ideal for high density system packaging.
revised 11-02

FS6200 Series

MICRO FLOW SWITCH


SPECIFICATIONS
Switch Assembly - SPDT UL 61760
100 VDC, 120 VAC max.
3 Watts resistive max. continuous
0.25 Amp. max. continuous
3.30 VA max. - 100 VDC
2.46 VA max. - 120 VAC
0.20 Ohm max contact resistance
Adjustable Flow Rate Setpoint*

-1 30 sccm to 10 slpm N2
-3 1 slpm to 20 slpm N2
Fluid Media: Gases including
corrosives
Pressure: 3000 psig max. (207 bar) IN
Operating Temperature:
-40F to 220F (-40C to 105C)
Leakage: 1x10-7 cc/sec He max.
Set Point Accuracy: 1%
Response Time: 10 milliseconds max.
Deactuation (reset): less than 12% of
actuation set point
Internal Filter: 40 micron
Materials:
Body and wetted parts: 316 St Stl
Spring: Inconel X
Seals: As specified
Weight: .45 lbs (205 gms)

22 Gauge Lead Wires


x 96 inches long
(244cm)

1/2-14 NPT
Conduit
Connecton
1.66
(42.2mm)
OUT

Flow Rate
Adjustment
1.37
(34.8mm)

Part Number Example:

N.O.

N.C.

Black
White
Blue
Increasing Flow

FS6202 C V - 1 M

Basic Series:
FS6202 - 1/8 NPT
FS6204 - 1/4 NPT

Optional Mounting Bracket

Material:
C = 316 Stainless Steel

Adjustable Flow Range*


-1 30 sccm to 10 slpm
-3 1 slpm to 20 slpm

Seal Material:
V = Viton
B = Buna - N
E = Ethylene Propylene
K = Kalrez

1.50
(38.1mm)

ELECTRICAL SCHEMATIC

COM

ORDERING INFORMATION

Mounting
Bracket

* Adjustable Flow Rate Set Point @ 5 psig (0.34 bar)

Products Co.
15915 So. San Pedro St., Gardena, CA 90248

Ph. (310) 515-2866

Copyright 2002 Autoflow Products Co. Kalrez and Viton are registered trademarks of E.I. du Pont. All rights reserved

FEATURES

Excellent regulation, stability and repeatability


Corrosion-resistant construction (no brass
components, T-51FR & T-51AFR)
NACE Constructed (T-51FRCT Corrosive Tec)
Low droop
Low cost
Small package size
Panel, bracket or pipe mounting
Fluorocarbon pintle seat (T-51FR, T-51AFR & T-51FRCT)
Auto drain option (T-51AFR)

DESCRIPTION

PRECISION CONTROL DEVICE

From
industry
to industry,
Marsh Belloframs
Type 51 Series of Regulators offer a
low-cost, high performance option
for a wide range of applications.
OPTIONS
1
2
3

The Bellofram Type 51 Precision Air Regulator series offers


a high-performance regulator in a compact, low cost package. It operates in output pressure ranges up to 100 psig /
6.9 BAR (120 psig / 8.3 BAR in T-51FR Corrosive Tec), with
a maximum supply pressure of 250 psi (17.3 BAR).
Materials of construction for standard T-51 Series
Regulators (Non-Corrosive Tec): Die-cast aluminum for the
body and dripwell; glass-reinforced thermoplastic polyester
for the bonnet; acetal resin for the internals; BUNA-N for
the diaphragm, gaskets and O-ring, fluorocarbon for the
pintle seat, and aluminum for the drain valve (plated steel
handle).
Materials of construction for Corrosive Tec T-51FRCT:
Aluminum alloy bonnet, body, and filter bowl, 316
stainless steel internals, Inconel alloy range spring, nitrile
diaphragm (fluorocarbon optional), 316 stainless steel
valve assembly, and painted with an epoxy finish. All
metallic parts for this unit conform to NACE material
requirements #MR-01-75 Class III.
These regulators are available standard (T-51R) or as
filter-regulators (T-51FR & T-51FRCT) and are even
available with an automatic drain, for automated flushing
out of contaminants, (T-51 AFR). These versatile regulators
provide excellent regulation for a wide range of applications, including pneumatic instruments, controllers,
chucks, and actuators. They can be through-panel
mounted with the supplied mounting nut, bracket-mounted
with the optional bracket or, due to their light weight,
mounted by their ports. The Corrosive Tec is supplied with
a tapped bonnet vent, to allow for the capture of exhaust air.

= option is available

Fluorocarbon Pintle
Non-Relieving
Knob
Sq. head Adj. Screw
4 5 Micron Filter
5 Epoxy Coating
6 Tapped Vent
Coalescing filter
7 Mounting Bracket
8 Pressure Gauge
9 Tamper - Resistant Cover
Panel Nut Mount
10 Soft Relief Seat
Low Bleed
11 Fluorocarbon Diaphragm

s = option is standard
TYPE 51AFR TYPE 51FRCT

TYPE 51R

TYPE 51FR

s
s

TYPE 51R
Regulator

Filter Regulator

Auto Filter Regulator

TYPE 51AFR

TYPE 51FRCT

250 psig (17.3 BAR)

250 psig (17.3 BAR)

250 psig (17.3 BAR)

250 psig (17.3 BAR)

Output Pressure Range

0-30 psig (0-2.1 BAR)


0-60 psig (0-4.1 BAR)
0-100 psig (0-6.9 BAR)

0-30 psig (0-2.1 BAR)


0-60 psig (0-4.1 BAR)
0-100 psig (0-6.9 BAR)

0-30 psig (0-2.1 BAR)


0-60 psig (0-4.1 BAR)
0-100 psig (0-6.9 BAR)

0-30 psig (0-2.1 BAR)


0-60 psig (0-4.1 BAR)
0-120 psig (0-8.3 BAR)

Supply Pressure Sensitivity

0.20 psig (0.01 BAR )


output change

0.45 psig (0.03 BAR)


output change

0.45 psig (0.03 BAR)


output change

0.45 psig (0.03 BAR)


output change

Sensitivity

1 (2.5 cm) of water

1 (2.5 cm) of water

1 (2.5 cm) of water

1 (2.5 cm) of water

Repeatability
Flow @ 100 psig (6.9 BAR) Supply

0.1 psig (0.01 BAR)


15 SCFM
(425 LPM)
0.1 SCFM
(2.8 LPM)

0.1 psig (0.01 BAR)


20 SCFM
(566 LPM)
0.1 SCFM
(2.8 LPM)

0.1 psig (0.01 BAR)


20 SCFM
(566 LPM)
0.1 SCFM
(2.8 LPM)

0.1 psig (0.01 BAR)


20 SCFM
(566 LPM)
0.1 SCFM
(2.8 LPM)

0 to 125F (-18 to 52 C)

0 to 125F (-18 to 52C)

0 to 125F (-18 to 52 C)

0 to 180F (-18 to 82C)

Maximum Supply Pressure

20 psig (1.4 BAR) outlet

Exhaust Capacity
@ 5 psig (0.4 BAR) above setpoint

Temperature Limits
Air Consumption
Port Size
Materials of
Construction

Corrosive Tec

6 SCFH (2.84 LPM) Maximum 6 SCFH (2.84 LPM) Maximum 6 SCFH (2.84 LPM) Maximum 6 SCFH (2.84 LPM) Maximum
/4 NPT

Aluminum, plated steel,


brass, acetal resin,
Buna-N /polyester,
music wire

/4 NPT

Aluminum, plated steel,


acetal resin,
Buna-N / polyester,
music wire, Fluorocarbon

/4 NPT

Aluminum, plated steel,


acetal resin,
Buna-N / polyester,
music wire, Fluorocarbon

/4 NPT

Aluminum, Stainless Steel,


Inconel, Buna-N / polyester,
Fluorocarbon, acetal,
polyphenylene sulfide

For T-51FR, T-51AFR & T-51R

T-51 SPECIFICATIONS & DIMENSIONS

@ 25 psig / 170 kPa change in supply

TYPE 51FR

ORDERING
INFORMATION
PART NUMBER:

TYPE-51 GENERAL PURPOSE REGULATORS

Part
TYPE
Number
960-222-000
51R
960-223-000
960-224-000
960-175-000
51FR
960-176-000
960-177-000
960-284-000
51AFR
960-285-000
960-286-000
960-303-000
51 FRCT 960-304-000
960-305-000

OPTION ORDERING MATRIX:

Add three digits from table below to end of part number.


OPTION
1 2 3 4 5 6 7

Control Range
psig BAR
0-30 0-2.1
0-60 0-4.1
0-100 0-6.9
0-30 0-2.1
0-60 0-4.1
0-100 0-6.9
0-30 0-2.1
0-60 0-4.1
0-100 0-6.9
0-30 0-2.1
0-60 0-4.1
0-120 0-8.3

REGULATOR
OPTIONS &
ACCESSORIES

9 10 11

Fluorocarbon
Pintle

Non-Relieving

Knob

5 Micron Filter

004 054 064 074 084 094 104 114

Epoxy Coating

005 065 075 085 095 105 115

Coalescing Filter
(T-51AFR only)

Mounting Bracket

Pressure Gauge

Tamper-Resistant Cover

001 021 031 041 051 061 071 081 091 101 111
002 032 042 052 062 072 082 092
003 043 053 063 073 083

112
103 113

006 076 086 096 106 116


007 087 097 107 117
008 098 108 118
009 109 119

Panel mounting Nut


NON RELIEVING
10 Soft Relief Seat
010 110
Used in applications where it is desirLow Bleed
able to relieve pressure downstream
11 Fluorocarbon
011
and not at the regulator. Non-relievDiaphragm
ing regulators should not be used for
low or no flow applications.
TAMPER PROOF COVER
PRESSURE GAUGE
KNOB
Dual scale (psi/kPa) 2 (50mm) A 316 stainless steel hexagonal cover
Replaces the standard square head gauges. Ranges include 0-60 psi (0- placed over the adjusting screw and
adjusting screw. (except T-51FRCT) 4.1 BAR),0-100 psi (0-6.9 BAR) and threaded onto the bonnet of the regu0-120 psi (0-8.3 BAR). When speci- lator with a wrench, prevents ordinary
5 MICRON FILTER
Replaces the 40 micron filter for fied with regulator, the correct range hand adjustments. Supplied with an
O-ring that is designed to seal the
more complete air filtration. (Except will be supplied.
adjusting screws threads in captured
FLUOROCARBON ELASTOMERS
T-51R)
Diaphragm, as well as gaskets and bleed applications. (T-51FRCT Only)
EPOXY FINISH
LOW BLEED DIAPHRAGM
An epoxy paint applied to the body O-rings, are made with a special
and dripwell of the regulator exterior elastomer to prevent deterioration (Soft Relief Seat): Used in applicasurfaces to provide increased from elements in the air supply, such tions where it is desirable to minimize
corrosion resistance. (Standard for as flame retardant synthetic lubri- the standard bleed rate of the
cants normally destructive to the regulator while maintaining the ability
T-51FRCT)
to relieve excess pressure at the
standard BUNA-N material.
MOUNTING BRACKET
regulator. Bleed rate is reduced from
Plated steel bracket for side mountless than 6 SCFH (2.8 LPM) to less
ing. (316 SS for T-51FRCT)
than 0.1 SCFH (0.05 LPM).
COALESCING FILTER
(T-51FRCT only)
Replaces the 40 micron filter for
both moisture and particulate filtration. (T-51AFR only)

USA
State Route 2, Box 305
Newell, WV 26050
800- 727-5646 or (304)387-1200
FAX (304) 387-4417
info@marshbellofram.com

EUROPE
9 Castle Park, Queens Drive
Nottingham NG2 1AH, UK
Tel +44 (0) 115 993 3300
Fax +44 (0) 115 993 3301
bellofram@aol.com

IMPORTANT NOTICE
Our recommendations, if any, for the use of this product are based on
tests believed to be reliable. The greatest care is exercised in the selection of our raw materials and in our manufacturer, no guarantee or warranty, express or implied is made as to such use or effects incidental to
such use, handling or possession or the results to be obtained, whether in
accordance with the directions or claimed so to be. The manufacturer
expressly disclaims responsibility therefor. Furthermore, nothing contained herein shall be construed as a recommendation to use any product
in conflict with existing laws and/or patents covering any material use.

SALES & APPLICATIONS


(800)727-5646 fax:(304) 387-4417
w w w. m a r s h b e l l o f r a m . c o m
LT 0795 5m 5/02

2:

5:

SYS19: BASE2

PDFINFO

50:

95:

98:

JOB: CRMAIN06-0428-6 Name: 5C-428

5C

100:
DATE: JAN 19 2006

Time: 5:25:03 PM Operator: GK COLOR: CMYK

Dimensions Pg. 430


Factory sealed enclosures are installed in a
rigid metallic conduit system for surface
mounting adjacent to or remote from
equipment being controlled and are used:
to prevent arcing of enclosed device from
causing ignition of a specific hazardous
atmosphere or atmospheres external to the
enclosure
in industrial areas such as chemical plants,
oil and gas refineries, paint and varnish
manufacturing plants, gasoline bulk loading
terminals, grain elevators, grain processing
industries, coal processing or handling areas,
or metal handling or finishing areas where
atmosphere may contain hazardous gases
and/or dust
in non-hazardous areas where sturdy,
durable enclosures are required
in conjunction with magnetic starters or
contactors for remote control of motors
Manual motor starting switch enclosures are
used:
for manual starting of small ac or dc motors
to provide manual starting and stopping
and, in the case of units with heaters, motor
running protection

Features:
Factory sealed devices have many distinct
advantages:
reduce installation problems
eliminate external seals
lower installation costs
improve safety
are used with general purpose snap and
pushbutton type switches
standard neoprene covers for front
operated pushbuttons. Prevents
accumulation of dirt and entrance of water
around operating shafts
mounting lugs and taper tapped hubs with
integral bushings
large machine screws for fastening covers
to bodies
lockout provisions on front operated push
button (marked STOP and OFF) and
selector switch covers
lockout hole for padlock having 14 hasp is
provided when used with covers for front
lever and side rocker type operation
close tolerances in machining of wide,
mating flanges and journalled shafts and
bearings for front button operation, produces
flametightness of enclosure joints
on enclosures with front lever and side
rocker type operating handles, threaded type
shafts and bushings are used to ensure
flametightness
dead end (EFS or EDS) or through feed
(EFSC or EDSC) hubs 12 to 1 sizes

Shafts stainless steel


Shaft bushings stainless steel
Rocker handle and pushbutton guards
type 6/6 nylon
Sealing enclosures copper-free aluminum

Standard Finishes:
Feraloy iron alloy electrogalvanized and
aluminum acrylic paint
Copper-free aluminum natural
Type 6/6 nylon black
Stainless steel natural

NEC/CEC:
Class I, Division 1 & 2, Groups B*,C,D
Class II, Division 1, Groups E,F,G
Class II, Division 2, Groups F,G
Class III
NEMA/EEMAC: 3,7B*CD,9EFG
UL Standard: 894, 698
CSA Standard: C22.2 No. 30

Options:
The following special options are available
from the factory by adding suffix to Cat. No.

Suffix to be
Added to
Cat. #

Description

Emergency Stop button (momentary) front operated red mushroom button . . . . . . . . . . S111
Lockout provision on front operated pushbutton cover (standard on buttons
marked OFF and STOP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S153
For 24 VDC operation on pilot lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S300
Three-position selector switches with modified operation:
Momentary contact clockwise operation, spring return to center, maintained
contact counter-clockwise operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S634
Momentary contact counter-clockwise operation, spring return to center,
maintained contact clockwise operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S635
Bodies and covers (single and two gang units) copper-free aluminum. . . . . . . . . . . . . . . . . . . SA
Where indicated in the catalog listings, EDS units suitable for Class I, Division 1,
Group B usage can be supplied, add suffix - GB, EFS units are suitable for Class I, Division 1,
Group B as standard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . GB
Maintained contact mushroom head with lockout and guard. Maximum one per cover
and cannot be used with a pilot light with a transformer or another standard pushbutton.. . S769

EDS bodies and factory sealed cover and device sub-assemblies are available for field
assembly (see pages 394 to 396).

Bodies Feraloy iron alloy; copper-free


aluminum.
Front operated pushbutton and pilot light
covers Feraloy iron alloy
Side operated type pushbutton covers
copper-free aluminum

STIBOINFO((CRH:66008com:5C:428))
CH0 0 5 C- 0

Explosionproof
Dust-Ignitionproof
Raintight
Wet Locations

Certifications and
Compliances:

Suffix S769

Standard Materials:

428

Typedriver Name: TS name csm no.: 100

Cl. I, Div. 1 & 2, Groups B*,C,D


Cl. II, Div. 1, Groups E,F,G
Cl. II, Div. 2, Groups F,G
Cl. III
NEMA 3,7B*CD,9EFG

EFS and EDS Factory Sealed


Control Devices and Manual
Motor Starting Switches

Application:

TCP: 15001

US: 1-866-764-5454

* See Options: suffix GB

CAN: 1-800-265-0502

Copyright 2006 Cooper Crouse-Hinds

Zoom: 100

2:

5:

SYS19: BASE2

PDFINFO

50:

95:

98:

100:

JOB: CRMAIN06-0439-3 Name: 5C-439

DATE: JAN 19 2006

Time: 5:25:16 PM Operator: RB

EDS Factory Sealed General Use


Snap Switches
Dimensions Pg. 430

EDSC2129

COLOR: CMYK

TCP: 15001

Cl. I, Div. 1 & 2, Groups B*,C,D


Cl. II, Div. 1, Groups E,F,G
Cl. II, Div. 2, Groups F,G
Cl. III
NEMA 3,7B*CD,9EFG

Typedriver Name: TS name csm no.: 100

Explosionproof
Dust-Ignitionproof
Raintight
Wet Locations

5C

EDS2229

General Use Snap Switch Front Operated


Single Gang
277VAC
20
20
20
20

Dead End
Cat. #
EDS2129
EDS218
EDS2130
EDS2140

20
20
20
20

EDS3129
EDS318
EDS3130
EDS3140

EDSC3129
EDSC318
EDSC3130
EDSC3140

Amperes

Hub
Size
34
34
34
34

Style
1-pole
2-pole
3-way
4-way

120VAC
20
20
20
20

1
1
1
1

1-pole
2-pole
3-way
4-way

20
20
20
20

Two Gang

Through
Feed
Cat. #
EDSC2129
EDSC218
EDSC2130
EDSC2140

Dead End
Cat. #
EDS2229
EDS2230
EDS3229
EDS328
EDS3230
EDS3240

Through
Feed
Cat. #
EDSC2229
EDSC228
EDSC2230
EDSC2240
EDSC3229
EDSC328
EDSC3230
EDSC3240

ON-OFF standard marking for 1-pole and 2-pole units.


Combinations of switches can be furnished.
AC rated switches are tested for resistive, inductive and tungsten filament loads up to the
full current rating and for motor loads up to 80% of the ampere rating.
* Class I, Group B: All units on this page can be modified for Class I, Group B usage. Add
suffix GB to the Cat. No. Seals must be installed within 112 of each conduit opening in
Division 1. In Canada, for Group B applications consult factory.

US: 1-866-764-5454

STIBOINFO((CRH:66008com:5C:439))
CH0 0 5 C- 9

CAN: 1-800-265-0502

Copyright 2006 Cooper Crouse-Hinds

439

Zoom: 100

2:

5:

SYS19: BASE2

PDFINFO

50:

95:

98:

JOB: CRMAIN06-0556-1 Name: 2A-556

2A

100:
DATE: JAN 19 2006

Time: 5:35:18 PM Operator: RB

with General Use Snap Switches


Front Operated

EDS and EFD enclosures are installed in a


rigid metallic conduit system for surface
mounting adjacent to or remote from
equipment being controlled and are used:
to prevent arcing of enclosed switch from
causing ignition of a specific hazardous
atmosphere or atmospheres external to the
enclosure
in industrial areas such as chemical plants,
oil and gas refineries, paint and varnish
manufacturing plants, gasoline bulk loading
terminals, grain elevators, grain processing
industries, coal processing or handling areas,
or metal handling or finishing areas where
atmosphere may contain hazardous gases
and/or dust
in non-hazardous areas where sturdy,
durable enclosures are required

Features:
Small and compact in design.
Used with snap switches.
Mounting lugs and taper tapped hubs with
integral bushings.
Large machine screws for fastening covers
to bodies.
Lockout hole for padlock having 14 hasp is
provided.
Threaded type shafts and bushings are
used to insure flametightness.

Standard Materials:
Bodies and covers Feraloy iron alloy
Shafts stainless steel
Shaft bushings stainless steel

Standard Finishes:

TCP: 15001

Typedriver Name: TS name csm no.: 100

Cl. I, Div. 1 & 2, Groups B*C,D


Cl. II, Div. 1, Groups E,F,G
Cl. II, Div. 2, Groups F,G
Cl. III
NEMA 3,7B*CD,9EFG,12

EDS and EFD Enclosures

Application:

COLOR: CMYK

Explosionproof
Dust-Ignitionproof
Raintight
Wet Locations

Certifications and
Compliances:
NEC/CEC: Class I, Div. 1 & 2, Groups
B*,C,D
Class II, Div. 1, Groups E,F,G
Class II, Div. 2, Groups F,G
Class III
NEMA/EEMAC: 3,7B*CD,9EFG,12
UL Standard: 894
CSA Standard: C22.2 No. 30

EDS Enclosed
Snap Switch

Options:
Suffix to be
Added to Encl.
Cat. #

Description
Two or three gang bodies can be supplied with
combinations of devices listed for one gang enclosures . . . . Refer to modular listing, section 4C
Class I Group B, NEMA 7B see listing pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . GB
Flush wall mounting cover with 12 overhang single gang only
dull black instrument finish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S173
Complies with U.L. snap switch test requirements as follows:
Type of Test
Overload

Non-Inductive Endurance
Inductive Endurance
Tungsten Filament Lamp Endurance
Temperature Rise
Dielectric Withstand

AC-Rated (only) Switch


Rated Amp. +380%
Power Factor .40-.50
100 cycles, 6-10 cycles per minute
10,000 cycles, 18-24 cycles per minute at
rated current .98 min. P.F.
10,000 cycles, 18-24 cycles per minute
.75-.80 P.F.
10,000 cycles, 6-10 cycles per minute at
rated current and 120 volts
Not to exceed 30C
1500 volts

Feraloy iron alloy electrogalvanized and


aluminum acrylic paint
Stainless steel natural

Dimensions (in inches)


Hub
Size
34
1

Dimensions are approximate, not for construction purposes.


* See suffix GB.

556

STIBOINFO((CRH:66008com:2A:556))
CH0 0 2 A - 4

US: 1-866-764-5454

CAN: 1-800-265-0502

Copyright 2006 Cooper Crouse-Hinds

Dim.
h

Dim.
l

78

1316

1516

Zoom: 100

2:

5:

SYS19: BASE2

PDFINFO

50:

95:

98:

100:

JOB: CRMAIN06-0557-6 Name: 2A-557

DATE: JAN 19 2006

Time: 5:35:22 PM Operator: RB

EDS and EFD Enclosures


with General Use Snap Switches Front
Operated Single Gang and Two Gang

Amperes

Through
Feed
Cat. #

Style

120VAC

277VAC

34

1-pole
2-pole
3-pole
3-way
4-way

20
20

20
20

20
20

20
20

EDS2129
EDS218
EDS2123
EDS2130
EDS2140

EDSC2129
EDSC218
EDSC2123
EDSC2130
EDSC2140

1
1
1
1
1

1-pole
2-pole
3-pole
3-way
4-way

20
20

20
20

20
20

20
20

EDS3129
EDS318
EDS3123
EDS3130
EDS3140

EDSC3129
EDSC318
EDSC3123
EDSC3130
EDSC3140

1
1
1

1-pole
2-pole
3-way

30
30
30

30
30
30

EFD3591
EFD3593
EFD3594

EFDC3591
EFDC3593
EFDC3594

34
34

Explosionproof
Dust-Ignitionproof
Raintight
Wet Locations

2A

Two Gang

Dead End
Cat. #

34

Typedriver Name: TS name csm no.: 100

Through feed

Single Gang

34

TCP: 15001

Cl. I, Div. 1 & 2, Groups B*,C,D


Cl. II, Div. 1, Groups E,F,G
Cl. II, Div. 2, Groups F,G
Cl. III
NEMA 3,7B*CD,9EFG,12

Dead end

Hub
Size

COLOR: CMYK

Dead End
Cat. #
EDS2229
EDS2230
EDS3229
EDS328
EDS3230
EDS3240
EFD3691
EFD3694

Through
Feed
Cat. #
EDSC2229
EDSC228
EDSC2223
EDSC2230
EDSC2240
EDSC3229
EDSC328
EDSC3223
EDSC3230
EDSC3240
EFDC3691
EFDC3693
EFDC3694

*Class I, Group B:
All units listed on this page can be modified
for Class I, Group B usage. Add suffix GB to
the Cat. No.. Example: EDS2129-GB. Seals
must be installed within 112 of each conduit
opening for Group B usage.

See table on page 556 for AC-rated switch information.


ON-OFF standard marking for 1-pole and 2-pole units
15A, 125 VAC; 10A, 250 VAC
Combinations of switches can be furnished.

US: 1-866-764-5454

STIBOINFO((CRH:66008com:2A:557))
CH0 0 2 A - 5

CAN: 1-800-265-0502

Copyright 2006 Cooper Crouse-Hinds

557

Zoom: 100

p148

8/17/07

10:15 AM

Page 1

Series
VFA
&
VFB

Visi-Float
Flowmeters
Used to Indicate or Manually Control Air or Water Flow

MODEL VFA AND VFB


L

Flow

E
D
N

M
I
FULL OPEN

The VisiFloat flowmeter bodies are cut and precision machined from solid, clear acrylic plastic blocks. This construction
not only produces a handsome finished product, but permits
complete visual inspection. As a result, the Visi-Float flowmeters are especially popular for medical and laboratory equipment applications.

DIMENSION-IN INCHES
A
B
C
D
E
F
I
K
L
M
N

Scales are easy to read The front scale location and white
background provides excellent visibility. The direct reading
scales are hot stamped into the plastic and will not wear off.
Midrange calibration is established with a master flowmeter.
Accuracy is 5% of full scale for VFA models, 3% for VFB.
Scales average 2" long on the VFA models, 4" long on VFB.
Durable and attractive construction The machined
acrylic bodies of the VisiFloat flowmeters are practically unbreakable. Fabrication is backed by over 60 years of experience
in acrylic instrument machining. The tapered bore is precision
machined to a smooth surface that provides perfect visibility of
the indicating float. The VFA and VFB models are available
with either brass or stainless steel inlet and outlet connections
and are tapped for 1/8 NPT thread. VFB models 85 and 86
have either 1/4 back or 3/8 end connections. All standard
models employ BunaN O rings for leak proof operation and
are available with either back or end connections for horizontal
or vertical piping. Precision metering valves in brass or stainless steel are available for most VFA and VFB models.
148

VISIT OUR WEBSITES:

www.dwyer-inst.com

Model VFA
4
3 (18 female NPT)
158 (10-32 Thread)
1
2
1316
114
2116 (BV or SSV)
4332
1
3
4 (EC)
3
32 (EC)

Model VFB
612
512 (18 female NPT)
312 (10-32 thread)
1
2
112
114
2116 (BV or SSV)
61932
138
3
4 (EC)
3
32 (EC)

Easy installation All VisiFloat flowmeters have metal


mounting inserts on rear for panel mounting. They can also be
supported directly by system piping.

www.dwyer-inst.co.uk

www.dwyer-inst.com.au

p149

8/17/07

10:16 AM

Page 1

Special Multi-Column Visi-Float Flowmeters


Perfect for OEM applications, Visi-Float flowmeters can be custom made with up to 10 columns in a single block of acrylic plastic. Available with or without valves. Consult factory for more
information.

SPECIFICATIONS
Service: Compatible gases & liquids.
Wetted Materials:
Body: Acrylic plastic.
O-ring: Buna-N (Fluoroelastomer available).
Metal Parts: Brass standard, stainless steel optional.
Float: Stainless steel, black glass, aluminum, K Monel
depending on range.

Flow

Temperature & Pressure Limits:


Without Valve: 100 psig (6.9 bar) @ 150F (65C);
150 psig (10 bar) @ 100F (38C).
With Valve: 100 psig (6.9 bar) @ 120F (48C).
OEM Specials Special flowmeter designs can be supplied to meet a
wide range of requirements and specific applications. These include:
on-off plunger and push-to-test valves, special gas or fluid calibration,
special ranges, scales, name brand or other identification. Pointer flags
can be furnished for instant visual reference. For specific information,
please supply an outline of your requirements.

Accuracy: VFA = 5% of full scale; VFB= 3% of full scale.


Process Connection: 1/8 female NPT. VFB ranges 85 and 86
have 1/4 NPT back connections or 3/8 NPT end connections.
These ranges not available with brass valves.
Scale Length: VFA 2 typical length; VFB 4 typical length.
Mounting Orientation: Mount in vertical position.
Weight: VFA: 4.0-4.8 oz (.11-.14 kg). VFB: 7.2-8.8 oz (.20-.25 kg).

How To Order
SeriesRange No.ValveOption
Example: VFA-9-BV
Series VFA with 20-200 SCFH Air Range & Brass Valve

VFB SERIES

VFA SERIES
Model
VFA-X
VFA-X-SS
VFA-X-BV
VFA-X-SSV
VFA-X-EC
VFA-X-EC-SS

Description
Standard VFA
VFA with Stainless Metal Wetted Parts
VFA with Brass Valve
VFA with Stainless Steel Valve
VFA with End Connections
VFA with End Connections and Stainless
Steel Metal Wetted Parts

Options & Accessories


-PF, Red ABS Plastic Pointer Flag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
-VIT, Fluoroelastomer O-rings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RKA, Pressure Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
POPULAR RANGES

Model
VFB-X
VFB-X-SS
VFB-X-BV
VFB-X-SSV
VFB-X-EC
VFB-X-EC-SS

Description
Standard VFB
VFB with Stainless Metal Wetted Parts
VFB with Brass Valve
VFB with Stainless Steel Valve
VFB with End Connections
VFB with End Connections and Stainless
Steel Metal Wetted Parts

Options & Accessories


-PF, Red ABS Plastic Pointer Flag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
-VIT, Fluoroelastomer O-rings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RK-VFB, Pressure Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
POPULAR RANGES
Model VFB 4" Scale

Model VFA 2" Scale


Range
No.
1
2
3
4
5
6
7
8
9

32
33
34

Range
SCFH Air
.1-1
.2-2
.6-5
1-10
2-20
4-30
5-50
10-100
20-200
CC Water
per min.
6-50
10-100
20-200

CALL TO ORDER:

Range
No.
21
22
23
24
25
26
27

Range
LPM Air
.06-0.5
.15-1
.6-5
1-10
3-25
6-50
10-100

41
42
43
44

Gal. Water
per hour
.6-5
2-10
3-20
8-40

U.S. Phone 219 879-8000

Range
No.
50
91+
51+
52
53+
54+
55+

Range
SCFH Air
.3-3
1-10
2-20
4-40
10-100
10-150
20-200
SCFM Air
.3-3
CC/Min. Air
100-1000

Range
No.
65
66
67
68
69

LPM Air
.2-4
1-10
1-20
3-30
4-40
CC/Min. Water
82
2-30
GPH Water
80+
90
.5-12
83+
1-20
84
60
6-40
81
6-60
GPM Water
85*
.2-2
86*
.6-5
+
For dual range models in English and Metric add D to end of Range No.
* Ranges 85 and 86 consult factory. Not available with brass valve.

U.K. Phone (+44) (0)1494-461707

Asia Pacific Phone 61 2 4272-2055

149

membrane separator

Model 130

Quick Study

The Series 100 Genie Membrane Separators

remove 100% of entrained liquid and particulate in gas samples, which


preserves sample integrity; also, these models only allow gas sample to
flow to analyzers, which protects against liquid damage. The original
Genie Series 100 models are available in several body styles with different
membrane types to accommodate a wide variety of applications. The
Genie Supreme Series 100 models accommodate the same applications,
yet they offer an improved housing design for easy maintenance and the
innovative Liquid Block Technology that prevents liquid from being
forced across the membrane should sample line pressure conditions
become upset.

The Model 130 protects gas systems and related components

Applications


Analyzer protection against liquids in


gas sample conditioning systems
On-line and portable analyzers

GCs, Mass Specs, O2, H2S,

Moisture, and others
Spot, composite, or continuous gas
sampling in any process industry
including natural gas, petrochemical,
and oil refining

where the flow requirement is higher than the Genie Model 101 can
accommodate, higher than 720 cc/min. for the Type 5 membrane and
2,500 cc/min. for the Hi-Flow Backed membrane. It offers a 2 diameter
membrane, which is ideal for removing continuous liquid slugs from
gas samples; it is also perfect for protecting components such as online analyzers, gas chromatographs, or mass spectrometers. Additional
information such as Application Notes is available.

Technical Specifications
Maximum allowable working pressure

Maximum recommended
supply pressure
Maximum temperature

Stainless Steel
Polypropylene
Kynar

500 psig
50 psig
50 psig

Lowest possible pressure consistent with application*


*Must not exceed Pressure rating listed above

185 F (85 C) for Type 5 & BTU membrane


212 F (100 C) for Hi-Flow/Hi-Flow Backed membrane*
*Applies to Polypropylene & Kynar housing

302 F (150 C) for Hi-Flow/Hi-Flow Backed membrane*

Benefits





Superior analyzer protection


Helps preserve sample integrity
Reduces analyzer maintenance
Improves analyzer reliability

Features





*Applies to Stainless Steel housing

Maximum recommended
membrane flow rate

5,130 cc/min for Type 5 & BTU membrane*


18,000 cc/min for Hi-Flow/Hi-Flow Backed membrane*

(For higher flow rates contact the factory)

*Maximum flow results in approximately 2 psi membrane


differential pressure

Port sizes

Inlet, Outlet, & Bypass: 1/4 female NPT

Internal volume

Inlet: 2.60 cubic inches


Outlet: 0.30 cubic inches

Wetted materials

Machined parts : 316 stainless steel / NACE compliant


All other metal parts: stainless steel / NACE compliant
O-ring: Viton (other materials available upon request)
Membrane: Inert

Genie Membrane Technology


Low internal volume
Simple design
Optional G.U.T.S. seal

Genie, Genie Membrane TechnologyTM, Genie Membrane Separators,TM are trademarks or registered
trademarks of A+ Corporation, LLC. All other referenced trademarks are the property of their respective owners.

Model Numbering & Additional Part Numbers


Your model number is determined by your specific needs. Choose options below.

Sealing material

0 = Viton

Membrane type

2 = Hi-Flow

Material of construction

SS = Stainless Steel

Mounting bracket accessory

Part # 130-509-SS (sold separately)

How to build the model number:

130

1 = Kalrez

(other materials available upon request)

T = G.U.T.S.

5 = Type 5

6 = BTU

7 = Hi-Flow Backed

PP = Polypropylene

K = Kynar

How to build the replacement membrane kit number:

130

Sealing material
Membrane type
Material of construction

Sealing material
Membrane type

Dimensions
Front View

Top View

2.14

OPTIONAL MOUNTING BRACKET

3.74

1.20

3.09

PORT B

PORT C

1.88
3.20

1.16

2.79
1.40

PORT A

0.95
PORT B

0.54

1.20

Local Distributor:

Manufacturer
A+ Corporation, LLC
41041 Black Bayou Rd.
Gonzales, LA 70737

Call for expert product application assistance:


Phone: (225)-644-5255 Website: www.geniefilters.com
Fax: (225)-644-3975 E-mail: sales@geniefilters.com

2005 A+ Corporation, LLC. All rights reserved.


SCC-130-PS_0206

PR-1 Series
Adjustable Pressure Reducing Regulator
The PR-1 Series is a versatile specialty pressure reducing control regulator
designed to fulfill a wide range of needs in instrumentation sample systems
and other applications such as semiconductor processing gases. The many
features of the PR-1 make it ideal for a wide range of applications controlling
pressures at low to moderate flows in gas or liquid service. 316L material
is used to facilitate welded connections. Stainless steel caps and adjusting
screws prevent atmospheric corrosion and maintain appearance.
Enhanced internal body surface finish of better than 25 Ra plus
electropolishing allows easier cleaning and potentially less particle
contamination in the flow stream.
Five different seat materials, three alternate orifice sizes and seven
pressure control ranges with stainless diaphragms offer the user a wide
spectrum of capabilities for pressure control with inlet pressures up to
6000 psig and standard operating temperatures up to 500 F (260 C).

Features & Specifications


Gas or liquid service
Stainless steel (316L), Inconel,
Teflon and Tefzel (or optimal main
seat material choice) only in flow
stream
Electropolished 316L body with better
than 25 Ra diaphragm cavity surface
finish
Stainless steel cap with SS adjusting
screw

Adjustable outlet pressure ranges of


010, 025, 050, 0100, 0250,
0500 and 0750 psig
Operating temperatures of -40 F
(-40 C) up to +500 F (+260 C)
20 micron filters
Bubble tight shutoff under most
conditions
Cv flow coefficients 0.025, 0.06, 0.20
and 0.50 (0.06 std.)

Inlet pressures of up to 6000 psi

Options
Wetted materials of construction:
Monel, Hastelloy and Titanium

Special diaphragm assembly for water


service

Diaphragm attached poppet

SS inlet pressure gauges

Special fittings

SS outlet pressure gauges

Diaphragm assist spring for vacuum


purging

Base mounting brackets


Captured vent

Panel mount (1 3/8 mounting hole)


Self-relieving
Relief valves

2301 Wardlow Circle


Corona, CA 92880
tel 909.270.6200
fax 909.270.6201
www.goreg.com
sales@goreg.com

PR-1 Series
Adjustable Pressure Reducing Regulator
How to Order
See page 3 for standard configurations. For additional configurations, consult the factory.
See page 4 for port locations.

Maximum Temperature & Operating Inlet Pressures


Seat Material
Tefzel
High Density Teflon
PCTFE (formerly Kel-F-81)
Polyimide
Polyimide
PEEK
PEEK

Maximum
Temperature*
150 F (66C)
150 F (66 C)
175 F (80 C)
500 F (260 C)
175 F (80 C)
500 F (260 C)
175 F (80 C)

@
@
@
@
@
@
@
@

Maximum Operating
Inlet Pressure
3600 psig (24.82 MPa)
3600 psig (24.82 MPa)
6000 psig (41.37 MPa)
3600 psig (24.82 MPa)
6000 psig (41.37 MPa)
3600 psig (24.82 MPa)
6000 psig (41.37 MPa)

* Temperatures in excess of 175 F (80 C) require a metal knob or the tamper proof option.
Tefzel and Teflon are registered trademarks of Dupont Corporation.

Outline and Mounting Dimensions

For flow curve charts, go to www.goreg.com.

Page

PR-1 Series - Pressure Reducing Regulator


Material of Body
1
4
6

SS 316L
Monel
Hastelloy C

Port Configuration (see page 4)


STANDARD BODY "A" (ONE INLET PORT AND ONE OUTLET PORT)

Process port types (gauge port type, if specified)


0
1
4

1/8" FNPT (1/8" FNPT Gauge Ports)


1/4" FNPT (1/4" FNPT Gauge Ports), Standard
3/8" FNPT (1/4" FNPT Gauge Ports)

Surface Finish of Diaphragm Cavity


1

<25 Ra, Standard

Seat Material
A
B
C
H
I
Q

Tefzel
CF Teflon
Polyimide, High Temp. Service (Metal Knob Std.)
PCTFE (formerly Kel-F 81)
High Density Teflon
PEEK

Flow Coefficient (Cv)

Cv

3
5
C
H

0.06, Standard
0.2
0.025
0.5

Outlet Range
C
D
E
G
I
J
W

0 - 10 Psig
0 - 25 Psig
0 - 50 Psig
0 - 100 Psig
0 - 250 Psig
0 - 500 Psig
0 - 750 Psig

Diaphragm Type
1
2
3
4

Standard Diaphragm
Diaphragm Attached Poppet (DAP)
Self Relieving
Vacuum Assist Spring, Standard Diaphragm

Diaphragm Facing / Backing Material


1
6
8
0
J

Teflon / S.S.
Tefzel Ring / S.S.
Teflon / Inconel
Teflon / Hastelloy C
Tefzel Ring / Inconel

Cap Assembly
1
4
5
8
H

Standard, S.S.
Panel Mount, S.S.
Captured Vent, Aluminum
Tamper Proof, S.S.
1/4" NPT Dome Loaded, S.S.

Cap
Assembly

Diaphragm
Material

Diaphragm
Type

Outlet
Range

Flow (Cv)

Seat
Material

Cavity
Finish

Port Style

Port
Config.

Material

P R 1 -

page 3

PORT LOCATIONS (SINGLE STAGE PRESSURE REGULATOR)


OUT

270

OUT

IN OUT

180

IN OUT

IN

90
OUT
IN

OUT

OUT

IN

OUT

OUT

IN OUT

IN

IN

IN

OUT

IN

OUT

OUT

IN

OUT

IN

OUT

IN

IN

OUT

IN

OUT
IN
1/8 NPT
OUT

OUT

OUT

IN

OUT

OUT

IN

OUT

IN

IN

L
OUT

OUT
IN

IN

OUT

OUT

OUT

IN

IN

OUT

IN
OUT
IN

LOCATION OF PORTS FROM


TOP VIEW
4

IN

w w w .in te r te c - in s t.c o m

VARITHERM
1

. . . e n c lo s u re s o lu tio n s

ELECTRIC HEATER

Application

Heating of enclosures and protective cabinets in


hazardous areas for
Freeze protection
Protection against condensation and for
Temperature maintenance

Special Features

Vertical or horizontal installation is possible


High heat transfer capabilities due to the design of
the black anodized aluminium ribbed fins
Fins can be machined or removed for mounting
purposes if necessary
Resettable temperature limiter supplied as standard
Bi-standard: Can be used for conduit installation as
well as for installation according to CENELEC
standards

Description

The VARITHERM is part of a versatile, explosion-proof


heating system. The heater uses fins to heat the air by
convection.
A thermostat limiter for constant surface temperature
and a resettable temperature limiter that disconnects
the heater from the power supply in case of external
heating or technical failure are incorporated in the
heater body. After correcting the cause of the failure
and allowing the heater to cool down, the heater can be
switched on again by resetting the red button with the
reset tool provided.

The effective heating capacity depends on the internal


temperature of the protective housing and the
orientation (vertical or horizontal) of the heater.
Particulary when the required enclosure temperatures
exceed >10C, it has to be taken into consideration that
the heat output can be reduced by the switching
behaviour of the internal thermostat. When a vertical
orientation is used, the heating power is approximately
25% less.
Please contact our engineering department for an exact
calculation of the heat requirement and selection of the
heater.

Cl. 1, Div. 1, Gr. A,B,C,D +

Type of protection

EEx ed IIC T4 / T3
Certificate of conformity

CSA NRTL/C File LR 43764


PTB No. Ex-97.D.1090

Protection degree

NEMA 4X , IP 68

Nominal voltage

120 V AC

Connection cable

EWKF 3x1,0mm 0,29(7,4mm)

Connection cable for ..AM

EWKF 5x1,0mm 0,35(8,8mm)

Length of connection cable

39 (1 meter)

Length, Width

8,7 x 8,4 (220mm x 213mm)

Material of construction

Seawater-proof black anodized


aluminium

6
Type
Nom.
Output
Height
in./mm

Heat Output

Technical Data

Design
200/60

300/80 400/100 500/60

600/80 700/100

200 W

300 W

600 W

2.4/ 60

3.1/ 80 3.9/ 100 2.4/ 60

300 W

500 W

IN S T R U M E N T A T IO N

3.1/ 80 3.9/ 100

Other designs upon request.


Model Number Code:
VARITHERM 400/80 T4R (Standard)
VARITHERM 400/80 T4R AM 3M MB (s.Options)

Options

AM

Alarm opens at <40F (5C)

3M

Connection cable 10 (3 meters) long

120V

Nominal voltage 240V AC

MB

Stainless steel mounting bracket

Hd121-6c.doc
IN T E R T E C

700 W

page 1/ 2
L T D .

P .O . B o x 2 3 0 6 0 , M ilt o n - O n t a r io L 9 T 5 B 4

T e l: ( 9 0 5 ) 8 7 5 - 4 7 2 7

F a x (9 0 5 ) 8 7 5 -4 7 2 9

w w w .in te r te c - in s t.c o m

VARITHERM
8

. . . e n c lo s u re s o lu tio n s

ELECTRIC HEATER
9.3

Wiring Diagram
L
N
PE
e.g. in series with
other alarms

6A

For CENELEC installation is a electronic temperature


controller available. See data sheet TC..Ex EN (HD251)
in the English catalog section.

10

Junction Box

TC...Ex, Electronic Temp.Regulator for


CENELEC installation

Installation of the VARITHERM

When the heater is installed, please make sure that the


label and button of the resettable temperature limiter
remain accessible.
ye/gn

bn

bl

ye/gn

>

>

bk

bk

10.1

<

Using the Zinc Plated Mounting


Brackets

>
Option: KR..Ex,
TC ...Ex, ELPR

Varitherm

Option: AM

bl=blue bn=brown ye/gn=yellow/green

Installation of Thermostats and


Temperature Regulators

9.1

TS, Thermostat (Installation with


conduit)

The assembly of the TS thermostat, heater, and


junction box is shown on following pages.

9.2

KR..Ex EN, Thermostat (CENELEC


installation)

Thermal feedback, necessary for good control, can be


achieved by mounting the KR .. directly to the heater.

The cylindrical spacers are supplied with the KR


thermostat. The mounting bracket can also be installed
vertically, and the difference in thickness can be
compensated by using two washers.

The INTERTEC universal mounting brackets, which are


supplied with the heater, allow versatile installation in
horizontal or vertical position.

10.2

Using the Stainless Steel Mounting


Bracket

The mounting bracket provides easy installation e.g. to


a mounting plate or to the INTERTEC mounting-system
(C-rails) of the protective cabinets and enclosures.
It can also be ordered optionally by using the
designation MB.

Hd121-6c.doc
IN T E R T E C

IN S T R U M E N T A T IO N

page 2/ 2
L T D .

P .O . B o x 2 3 0 6 0 , M ilt o n - O n t a r io L 9 T 5 B 4

T e l: ( 9 0 5 ) 8 7 5 - 4 7 2 7

F a x (9 0 5 ) 8 7 5 -4 7 2 9

NOSHOK 500 Series All Stainless Steel Liquid Pressure Gauge Model 500 2 1/2 Inch Di... Page 1 of 2

Address: 1010 West Bagley Road


Berea, OH 44017
Phone: 440-243-0888
Fax: 440-243-3472
E-mail: noshok@noshok.com

500 Series All Stainless Steel Liquid Pressure Gauge


Model 500 - 2 1/2 Inch Dial Size
ORDERING INFORMATION
For Regularly Stocked Items

(Contact Us for additional gauge offerings)

Size and Model

25-500

2 Inch Dial Size - Model 500

NPT Connection Size

1/4" NPT

316 SS Bottom Connection

Single Scale Dial Units

Single Scale Options:

psi

Outer Scale Options:

psi

Inner Scale Options:

kPa

Outer Scale Options:

Inner Scale Options:

psi

Dual Scale Dial Units


bar

kg/cm

Ammonia Scale
Dial Units1

The Ammonia Dial is regularly Stocked in Ranges Indicated with (1) Below

Pressure Ranges

30" Vac 30 inHg Vacuum to 0 inHg

200 0 to 200 psi

30-0-15 -30 inHg to 0 inHg to 15 psi2

300 0 to 300 psi

30-0-30 -30 inHg to 0 inHg to 30 psi

400 0 to 400 psi

30-0-60 -30 inHg to 0 inHg to 60 psi2

600 0 to 600 psi

30-0-100 -30 inHg to 0 inHg to 100 psi2

1,000 0 to 1,000 psi

30-0-160 -30 inHg to 0 inHg to 160 psi1 2

1,500 0 to 1,500 psi

http://www.noshok.com/pg/500series/model500/order_info2.html

9/25/2007

NOSHOK 500 Series All Stainless Steel Liquid Pressure Gauge Model 500 2 1/2 Inch Di... Page 2 of 2

30-0-200 -30 inHg to 0 inHg to 200 psi2


12

30-0-300 -30 inHg to 0 inHg to 300 psi

2,000 0 to 2,000 psi


3,000 0 to 3,000 psi

15 0 to 15 psi

5,000 0 to 5,000 psi

30 0 to 30 psi

6,000 0 to 6,000 psi2

60 0 to 60 psi

10,000 0 to 10,000 psi

100 0 to 100 psi

15,000 0 to 15,000 psi2

160 0 to 160 psi


1
2

Options

Accessories

This Range is Also Available in the Ammonia F/psi Scale from stock

This Range is Not Regularly Stocked in the kg/cm Inner Scale Option

10-32 UNF 2B Connection

Threaded Brass Orifices

1/4" NPT x 1/4" BSP Brass Adapter

Maximum Indicating Pointer

Sealing Gasket for BSP Adapter

Set Pointer

-40 Service Liquid Fill

Certificate of Conformance

Halocarbon Liquid Fill

Certified Calibration

Silicone Liquid Fill

Cleaning for Oxygen Services

Press Fit Brass Orifices

Rubber Case Protector

Cintered Press-Fit Brass Orifices

Silicone Dampened Movements

Piston Type Pressure Snubbers

Pigtail Steam Syphons

Diaphragm Seals

Print this Page


Go Back

http://www.noshok.com/pg/500series/model500/order_info2.html

9/25/2007

www.swagelok.com

I n s t r u m e n t B a l l Va l v e s

40 Series
On-off, switching, and crossover flow paths
Working pressures up to 3000 psig (206 bar)
Temperatures from 50 to 150F (10 to 65C)
Low-temperature service option from 65 to 150F (53 to 65C)
1/16 to 3/4 in. and 3 to 12 mm end connections
Virtually no dead space

40 Series Ball Valves

Features
Directional handle
indicates position
of orifice.

Capsule seat packing


does not require system
pressure to make a seal
allows bidirectional flow
has virtually no dead space
is easily cleaned and purged
is available in optional materials
for system compatibility.

Top-loaded design
allows adjustment with
the valve in-line.
Panel nut
secures valve to
panel or actuator.
End connections
include gaugeable
Swagelok tube fittings and
a variety of others.

One-piece body
eliminates multiple
seal points.

Stainless steel and alloy 400


Swagelok tube fittings are gaugeable.

One-piece ball stem


ensures alignment of
ball and orifice.

Materials of Construction
1

Valve Body Materials


Stainless Steel
Component

41, 42, 44,


45 series

43 series

Handle
Set screw

Packing bolt

Upper gland

4
5
6
9
8
9

7
10
11
12

4
5
6
7
8
9a
10a
11a
12a

Brass

Alloy 400

Material Grade/ASTM Specification


Nylon with brass insert
S17400 SS/A564
316 SS powdered metal or
316 SS/A276 and A479
316 SS/A240

Brass CDA 360/B16


41, 42, 45 series
brass 260/B36;
43, 44 series
316 SS/A240
PTFE/D1710
Brass CDA 360/B16
PTFE/D1710
Brass CDA 360/B16

Alloy 405/B164

Alloy 400/B127

Bushing
Lower gland
316 SS powdered metal
Alloy 400/B164
Upper packing
Ball stem
316 SS/A276 and A479
Alloy 405/B164
Side rings
Fluorocarbon-coated
Fluorocarbon-coated 316 SS
Fluorocarbon-coated
brass powdered metal alloy 400 powdered metal
powdered metal/B783
Side discs
Lower packing
PTFE/D1710
Panel nut
316 SS powdered metal
Brass CDA 360/B16
316 SS powdered metal
Body
316 SS/A276 and A479
Brass CDA 360/B16
Alloy 400/B164
Lubricant
41, 42, 43 seriessilicone-based; 44, 45 seriessilicone- and fluorinated-based

Wetted components listed in italics.


Molybdenum disulfide with hydrocarbon binder coating.
4-way, 5-way, 6-way, and 7-way valves contain stainless steel stem, rings, and discs.
Bodies with VCO end connections have fluorocarbon FKM O-rings.
For valves assembled without lubrication, see Options, page 8.

Important Information about Packed Valves

Testing

Packing adjustment may be required

Valves that have not been cycled for

during the valves service life.


Warning: Failure to periodically
inspect and maintain valve packing
may result in leakage. Service
instructions are shipped with each 40
series ball valve.
Swagelok ball valves are designed to
be used in a fully open or fully closed
position.

a period of time may have a higher


initial actuation torque.
43 series ball valves require an
adapter to adjust the packing bolt.
Ordering number: MS-WK-43
For all other 40 series ball valves,
packing adjustments can be made
with standard wrenches.

Every 40 series ball valve is adjusted


for factory testing at 1000 psig (69 bar)
with nitrogen or at its maximum rated
pressure if less than 1000 psig (69 bar).
Seat tests have a maximum allowable
leak rate of 0.1 std cm3/min.

40 Series Ball Valves

Flow Data at 70F (20C)

On-Off (2-Way) Valves

Pressure Drop to Atmosphere (p), psi


10

StraightPattern
Valve

AnglePattern
Valve

On

Off

On

Pressure-Temperature
Ratings
Pressure ratings for valves with
Swagelok tube fitting ends may be
lower. See Swagelok Tubing Data.

Ordering Information
and Dimensions

Valve Series
41, 42, 41-A, 42-A, 43-A
43

End Connections

Fractional
Swagelok
tube fittings

Size

0.3

0.7

1.0

11

0.6

1.4

2.0

0.50

5.6

15

27

1.6

3.5

5.0

0.60

6.8

18

32

1.9

4.2

6.0

0.90

10

27

48

2.8

6.4

9.0

1.2

14

36

64

3.8

8.5

12

1.5

17

45

80

4.7

1.6

18

48

85

5.0

2.4

27

72

120

7.6

17

24

2.6

29

78

140

8.2

18

26

Pressure
Rating
psig (bar)
2500 (172)
3000 (206)

3.0

34

90

160

9.5

21

30

6.0

68

180

320

19

42

60

6.3

71

6.4

72

20

45

2500 (172)
1500 (103)

12

130

38

85

Off

-43S4

2.4 0.90

0.44
2.58 1.29 (32.8) 1.29 (11.2)
(32.8)

1.5 0.90

64
120

-44S6

3.05 1.52 (38.6) 1.43


6.0 2.00 0.281 7.1 (77.5)
(36.3)
12 4.60
3.92 1.96 (49.8) 1.74
0.406 10.3 (99.6)
(44.2)
6.4 3.80
2.01
0.97
0.2 0.15 0.093 2.4 (51.1) 1.01 (25.7) (24.6)

0.28

1.12

1/4 19/32 1.36

(7.1)

(28.4)

(6.4) (15.1) (34.5) (14.7)

0.38

1.53 3/16 25/32 1.56

(9.7)

(38.9)

0.56 (14.2)

(50.8)

0.69 (17.5)

0.34
2.21 1.10 (27.9) 1.07 (8.6)
0.6 0.35 0.125 3.2 (56.1)
(27.2)
2.39 1.20 (30.5) 1.17
2.4 0.90
(60.7)
(29.7) 0.44
0.187 4.8
2.46 1.23 (31.2) 1.20 (11.2)
1.5 0.90
(62.5)
(30.5)

2.00

1.00 (25.4)

2.06

1.03 (26.2)

(9.5) 1 1/2 2.43 1.50


(38.1) (61.7) (38.1)

1/4 19/32 1.36

(6.4) (15.1) (34.5) (14.7)

0.38

1.53 3/16 25/32 1.56

(9.7)

(38.9)

1 1/8 2.07

2.00
(50.8)

3.00
(76.2)

1.12

1/4 19/32 1.36

(6.4) (15.1) (34.5) (14.7)

(7.1)

0.44

0.38

1.53 3/16 25/32 1.56

(11.2)

(9.7)

(38.9)

0.56 (14.2)

(50.8)

-45F8

0.38

1.53 3/16 25/32 1.56

(9.7)

(38.9)

0.38

1.53 3/16 25/32 1.56

(9.7)

(38.9)

0.69 (17.5)

Male NPT

1/4

-43M4

Male NPT/
Swagelok
tube fitting

1/4

-43M4-S4

2.20 1.00 1.20 1.03 0.44


1.6 0.75 0.187 4.8 (55.9)
(25.4) (30.5) (26.2) (11.2)

1/4

-43F4RT

0.9

3/8

-44F6RT

2.6

2.06 1.03 (26.2)


0.187 4.8 (52.3)
2.50 1.25 (31.8)
0.281 7.1 (63.5)

3.12 1.56 (39.6)


0.406 10.3 (79.2)

-45F8RT

6.3

-42VCO4

1.75 0.88 (22.4)


0.6 0.35 0.125 3.2 (44.4)

1.53 3/16 25/32 1.56

0.56 (14.2)

(50.8)

0.69 (17.5)

3.00
1.12

1.88 0.94 (23.9)


2.4 0.90 0.187 4.8 (47.8)
2.12 1.06 (26.9)
0.6 0.35 0.125 3.2 (53.8)

1.09 0.44

0.38
(9.7)

-44VCR8

2.06 1.03 (26.2)


2.4 0.90 0.187 4.8 (52.3)
2.88 1.44 (36.6)
6.0 0.281 7.1 (73.2)

(27.7) (11.2)

-45VCR8

12

3.12 1.56 (39.6)


0.406 10.3 (79.2)

(76.2)

0.78

(4.8) (19.8) (39.6) (19.8)

(38.9)

2.00

0.78

(4.8) (19.8) (39.6) (19.8)

(9.7)

(9.7)

1/2

(38.1) (61.7) (38.1)

0.38

0.38

-43VCR4

1 1/8 2.07 1.12


3/8 (28.6) (52.6) (28.4)
1 1/2 2.43 1.50

0.44

0.94 0.44

-42VCR4

3.00
(76.2)

0.78

(9.5)

(11.2)

(23.9) (11.2)

1/4
Male VCR
fittings

2.00

0.58

(4.8) (19.8) (39.6) (19.8)

1/2

1.56 (39.6)

1.12

(28.4)

(8.6)

3/8

(79.2)

0.78

3/8 (28.6) (52.6) (28.4)


(9.5) 1 1/2 2.43 1.50
(38.1) (61.7) (38.1)

3.0 1.70
2.50
0.281 7.1 (63.5)
2.6 1.50
6.3 3.50 0.406 10.3 3.12

1.25 (31.8)

0.58

(4.8) (19.8) (39.6) (19.8)

-44F4
-44F6

(52.3)

1.12

(28.6) (52.6) (28.4)

1.12

1.2 0.70

(50.8)

3.00
(76.2)

3/8

(28.4)

-43F2

(41.1)

1 1/8 2.07

2.00

(7.1)

-42F2

-43VCO4

0.78

0.28

3.07 1.53 (38.9) 1.43 0.56 (14.2)


6.0 2.00 0.281 7.1 (78.0)
(36.9)
3.92 1.96 (49.8) 1.74 0.69 (17.5)
12 4.60 0.406 10.3 (99.6)
(44.2)
0.34 0.28
0.5 0.30 0.125 3.2 1.62
0.81 (20.6)

1/4

0.58

(4.8) (19.8) (39.6) (19.8)

0.9 0.75

1/4

(65.5)

2.00 1.00 (25.4) 1.03 0.44


1.2 0.75 0.187 4.8 (50.8)
(26.2) (11.2)

VCO fittings

63

-43F4

1/2

Dimensions shown with Swagelok


tube fitting nuts finger-tight.

2.36 1.18 (30.0) 1.17


(29.7)

(59.9)

-43S6

1/8

Angle-Pattern Valves
Add -A to the ordering number
of a valve with the C
dimension listed.
Example: SS-43S4-A

-42S4

12 mm -45S12MM

Female ISO
tapered

640

16

Dimensions, in. (mm)

(42.7)

0.187 4.8

360

2.21 1.10 (27.9) 1.07


0.6 0.35 0.125 3.2 (56.1)
(27.2)

10 mm -44S10MM

Female
NPT

340

15

11

For angle-pattern valve flow data, see Flow Data, page 4.

A
B
B1
0.052 1.3 1.68 0.84 (21.3)

0.187 4.8

Water Flow, U.S. gal/min

330

190

2.01 1.01 (25.7) 0.97 0.34


0.2 0.15 0.093 2.4 (51.1)
(24.6) (8.6)

6 mm
Metric
-43S6MM
Swagelok
tube fittings 8 mm -43S8MM

1/8 in. (3.2 mm) minimum panel


thickness.

5.3

6.0

-41S2

-42S6MM

B1

3.0

2.3

1/8

-45S8
-45S12

100

1.1

0.1

1/4

50

0.20

Orifice
Cv
Cv Angle in. mm

10

0.10

-41S1

3 mm -41S3MM

E
B

Basic
Ordering
Number

100

Air Flow, std ft3/min

1/16

1/2
3/4
G
max panel
thickness J

(10 to 65)

3/8

H
panel
hole

50 to 150

44, 45
44-A, 45-A

Inlet/Outlet

Add SS for 316 stainless


steel, B for brass, or M for
alloy 400 to the basic ordering
number.
Example: SS-43S4
Dimensions are for reference
only and are subject to
change.

Temperature
Rating
F (C)

50

Cv

0.78

(4.8) (19.8) (39.6) (19.8)

1 1/8 2.07

1.12

3/8 (28.6) (52.6) (28.4)


(9.5) 1 1/2 2.43 1.50
(38.1) (61.7) (38.1)
19/32 1.36

(28.4) 3/16 (15.1) (34.5) 0.78


1.53 (4.8) 25/32 1.56 (19.8)
(19.8) (39.6)
(38.9)

1.12

19/32 1.36

(28.4) 3/16 (15.1) (34.5) 0.78


1.53 (4.8) 25/32 1.56 (19.8)
(38.9)
(19.8) (39.6)

2.00

0.56 (14.2)

(50.8)

0.69 (17.5)

(76.2)

3.00

1 1/8 2.07
3/8 (28.6) (52.6) 1.50
(9.5) 1 1/2 2.43 (38.1)
(38.1) (61.7)

40 Series Ball Valves

Switching (3-Way, 5-Way, and 7-Way) Valves

3-Way Valve

5-Way Valve

Features

7-Way Valve

Flow Data at 70F (20C)

Unique, top-loaded capsule packing allows reliable

switching.
Flow can be switched from a single inlet to multiple outlets
or from multiple inlets to a common outlet.
3-way valve has a center-off position.
5- and 7-way 43 series valves have a spring-loaded detent
for exact port positioning.
Detent handle components:
Handle: nylon with brass insert
Set screw: S17400 stainless steel
Pins, detent plate: nickel-plated steel
Springs: steel/ASTM A228

Pressure-Temperature Ratings
Valve Series
41X, 42X, 43X
44X, 45X
43Z
45Z
43Z6

Flow Pattern

Temperature
Rating
F (C)

3-way
50 to 150
5-way
7-way

(10 to 65)

Pressure
Rating
psig (bar)
2500 (172)
1500 (103)
2500 (172)
1500 (103)
500 (34.4)

Pressure ratings for valves with Swagelok tube fitting ends


may be lower. See Swagelok Tubing Data.

Pressure Drop to Atmosphere (p) psi


10

50

100

Air Flow, std ft3/min

Cv

10

50

100

Water Flow, U.S. gal/min

2-Way Angle Pattern and 3-Way Valves


0.08

0.9

2.4

4.3

0.3

0.6

0.8

0.15

1.7

4.5

8.0

0.4

1.0

1.5

0.30

3.4

9.0

16

0.9

2.1

3.0

0.35

4.0

10

19

1.1

2.4

3.5

0.75

8.5

22

40

2.3

5.3

7.5

0.80

9.0

24

42

2.5

5.6

8.0

0.90

10

27

48

2.8

6.3

9.0

1.5

17

45

80

4.7

11

15

1.7

19

51

90

5.3

12

17

2.0

22

60

100

6.3

14

20

3.5

39

100

180

11

25

35

3.8

43

110

200

12

27

38

4.6

52

140

240

15

33

46

0.07

0.8

2.1

3.7

0.2

0.5

0.7

3.5

39

100

180

11

25

35

0.05

0.6

1.5

2.6

0.1

0.3

0.5

0.07

0.8

2.1

3.7

0.2

0.5

0.7

5-Way Valves

7-Way Valves

40 Series Ball Valves

Ordering Information and Dimensions


Add SS for 316 stainless steel, B for brass, or M for alloy 400 to the basic ordering number.
Example: SS-43XS4
Dimensions are for reference only and are subject to change.
3-Way Valve

5-Way Valve

7-Way Valve

G max panel
thickness

H
panel
hole

F
G
max panel
thickness J

H
panel
hole

G
max panel
thickness J

H
panel
hole

1/8 in. (3.2 mm) minimum panel thickness.


End Connections
Inlet/Outlet

Size

Basic
Ordering
Number

Orifice
Cv

in.

Dimensions, in. (mm)

mm

0.34 (8.6)

1.13 (28.7)

1/4 (6.4)

19/32 (15.0) 1.36 (34.5)

0.58 (14.7)

3/16 (4.8)

25/32 (19.8) 1.56 (39.6)

0.78 (19.8)

1 1/8 (28.7)

2.07 (52.6)

1.12 (28.4)

1 1/2 (38.1) 2.43 (61.7)

1.50 (38.1)

3-Way Valves

Fractional
Swagelok
tube fittings

1/16

-41XS1

0.08

0.052

1.3

1.68 (42.7)

0.84 (21.3)

1/8

-41XS2

0.15

0.093

2.4

2.01 (51.1)

1.01 (25.7)

0.97 (24.6)

-42XS4

0.35

0.125

3.2

2.21 (56.1)

1.10 (27.9)

1.07 (27.2)

1/4

-43XS4

0.90

0.187

4.8

2.39 (60.7)

1.20 (30.5)

1.17 (29.7)

0.44 (11.2)

1.53 (38.9)

3/8

-44XS6

2.0

0.281

7.1

2.89 (73.4)

1.45 (36.8)

1.43 (36.3)

0.56 (14.2)

2.00 (50.8)

1/2

-45XS8

4.6

3/4

-45XS12

3.8

-41XS3MM
-42XS6MM

3 mm
Metric
Swagelok
tube fittings

6 mm

Male NPT/
Swagelok
tube fittings
Female ISO
tapered
Male VCR
fittings

0.406

10.3

3.48 (88.4)

1.74 (44.2)

0.15

0.093

2.4

2.01 (51.1)

1.01 (25.7)

0.97 (24.6)

0.35

0.125

3.2

2.21 (56.1)

1.10 (27.9)

1.07 (27.2)

0.187

4.8

2.39 (60.7)

1.20 (30.5)

1.17 (29.7)

2.46 (62.5)

1.23 (31.2)

1.20 (30.5)

1.45 (36.8)

1.43 (36.3)

3.00 (76.2)

0.34 (8.6)

1.13 (28.7)

1/4 (6.4)

19/32 (15.0) 1.36 (34.5)

0.58 (14.7)

0.44 (11.2)

1.53 (38.9)

3/16 (4.8)

25/32 (19.8) 1.56 (39.6)

0.78 (19.8)

1 1/8 (28.7)

-43XS6MM

0.90

-43XS8MM

0.80

10 mm

-44XS10MM

2.0

0.281

7.1

2.89 (73.4)

0.56 (14.2)

2.00 (50.8)

12 mm

-45XS12MM

4.6

0.406

10.3

3.48 (88.4)

1.74 (44.2)

0.69 (17.5)

3.00 (76.2)

-42XF2

0.30

0.125

3.2

1.63 (41.4)

0.81 (20.6)

0.34 (8.6)

1.13 (28.7)

-43XF4

0.75

0.187

4.8

2.06 (52.3)

1.03 (26.2)

0.44 (11.2)

1.53 (38.9)

-44XF4

1.7

3/8

-44XF6

0.281

7.1

2.50 (63.5)

1.25 (31.8)

0.56 (14.2)

2.00 (50.8)

1.5

1/2

-45XF8

3.5

0.406

10.3

3.13 (79.5)

1.56 (39.6)

0.69 (17.5)

3.00 (76.2)

1/4

-43XS4-S4-M4

0.80

0.187

4.8

2.40 (61.0)

0.44 (11.2)

1.53 (38.9)

1/4

-43XF4RT

0.75

0.187

4.8

2.06 (52.3)

1.03 (26.2)

0.44 (11.2)

1.53 (38.9)

3/8

-44XF6RT

1.5

0.281

7.1

2.50 (63.5)

1.25 (31.8)

0.56 (14.2)

2.00 (50.8)

1/2

-45XF8RT

3.5

0.406

10.3

3.13 (79.5)

1.56 (39.6)

0.69 (17.5)

3.00 (76.2)

-42XVCR4

0.35

0.125

3.2

-43XVCR4

0.90

0.187

4.8

1/4

1/4

2.13 (54.1)

1.20 (30.5)

1.06 (26.9)

1.03 (26.2)

1.09 (27.7)

3/8 (9.7)

0.69 (17.5)

8 mm

1/8
Female NPT

0.81 (20.6)

0.44 (11.2)

1.13 (28.7)
1.53 (38.9)

2.07 (52.6)

1.12 (28.4)

1 1/2 (38.1) 2.43 (61.7)

1.50 (38.1)

1/4 (6.4)

19/32 (15.0) 1.36 (34.5)

0.58 (14.7)

3/16 (4.8)

25/32 (19.8) 1.56 (39.6)

0.78 (19.8)

1 1/8 (28.7)

2.07 (52.6)

1.12 (28.4)

1 1/2 (38.1) 2.43 (61.7)

1.50 (38.1)

3/16 (4.8)

25/32 (19.8) 1.56 (39.6)

0.78 (19.8)

3/16 (4.8)

25/32 (19.8) 1.56 (39.6)

0.78 (19.8)

1 1/8 (28.7)

2.07 (52.6)

1.12 (28.4)

1 1/2 (38.1) 2.43 (61.7)

1.50 (38.1)

3/8 (9.7)

3/8 (9.7)

3/8 (9.7)
3/16 (4.8)

19/32 (15.0) 1.36 (34.5)


25/32 (19.8) 1.56 (39.6)

0.78 (19.8)

5-Way Valves
Female
Swagelok
tube fittings
Female NPT

1/8

-43ZFS2

0.07

0.062

1/8

-43ZF2

0.07

1/2

-45ZF8-ND

3.5

1.6

1.94 (49.3)

0.062

1.6

1.55 (39.4)

0.406

10.3

3.13 (79.5)

0.97 (24.6)
0.78 (19.8)

0.88 (22.4)

1.56 (39.6)

0.44 (11.2)

1.53 (38.9)

5/32 (4.1)

29/32 (23.1) 1.69 (42.9)

0.44 (11.2)

1.53 (38.9)

5/32 (4.1)

29/32 (23.1) 1.69 (42.9)

0.69 (17.5)

3.00 (76.2)

3/8 (9.7)

1 1/2 (38.1) 2.43 (61.7)

0.44 (11.2)

1.53 (38.9)

5/32 (4.1)

29/32 (23.1) 1.69 (42.9)

7-Way Valves
Female
Swagelok
tube fittings

1/16

-43Z6FS1

0.05

0.052

1.3

1/8

-43Z6FS2

0.07

0.062

1.6

1.94 (49.3)

0.97 (24.6)

Dimensions shown with Swagelok tube fitting nuts finger-tight.


Cross-port flow may occur during switching. If cross-port flow is unacceptable, specify a 0.049 in. ball orifice. Example: SS-43ZF2-049
Cross-port flow may occur during switching. If cross-port flow is unacceptable, specify a 0.093 in. ball orifice. Example: SS-45ZF8-ND-093

40 Series Ball Valves

Crossover (4-Way and 6-Way) Valves


4-Way Valve

6-Way Valve

Features

Flow Data at 70F (20C)

Capsule packing allows crossover of two or three streams.

Pressure Drop to Atmosphere (p), psi

Machined stops provide positive port positioning.

10

Stop plate material: aluminum/ASTM B209 or B211.

50

100

10

Air Flow, std ft3/min

Cv

50

100

Water Flow, U.S. gal/min

4-Way Valves

Pressure-Temperature Ratings
Valve
Series
43Y6
45Y6
43Y6

Flow
Pattern

Temperature
Rating
F (C)

4-way

50 to 150
(10 to 65)

6-way

0.06

0.7

1.8

3.2

0.08

0.9

2.4

4.2

1.6

18

48

85

Working
Pressure
psig (bar)
2500 (172)

0.06

0.7

1.8

3.2

1500 (103)

0.08

0.9

2.4

4.2

0.2
5.0

0.4

0.6

0.5

0.8

11

16

0.4

0.6

0.5

0.8

6-Way Valves
0.2

1500 (34.4)

Ordering Information and Dimensions


Dimensions are for reference only and
are subject to change.
Add SS for 316 stainless steel, B for
brass, or M for alloy 400 to the basic
ordering number.
Example: SS-43YF2

End Connection
Type

1/16
Female
Swagelok
tube fitting 1/8

G
max panel
thickness J

Orifice
Cv

in.

Dimensions, in. (mm)

mm

-43YFS1

0.06 0.052

1.3

-43YFS2
0.08 0.062

1.55

0.78

(39.4)

(19.8)

0.44

0.44

1.53

3/16

29/32

1.68

1.95

0.98

(11.2)

(11.2)

(38.9)

(4.8)

(23.1)

(42.7)

(49.5)

(24.9)

-43YF2

1/2

-45YF8

1.6

0.281

7.1

Female
1/16
Swagelok
tube fitting 1/8

-43Y6FS1

0.06 0.052

1.3

-43Y6FS2

0.08 0.062

1.6

1.55

0.78

0.44

0.44

1.53

3/16

29/32

1.69

(39.4)

(19.8)

(11.2)

(11.2)

(38.9)

(4.8)

(23.1)

(42.9)

3.13

1.56

0.69

0.69

3.00

3/8

1 1/2

2.43

(79.5)

(39.6)

(17.5)

(17.5)

(76.2)

(9.7)

(38.1)

(61.7)

6-Way Valves
1.95

0.97

0.44

0.44

1.53

3/16

29/32

1.68

(49.5)

(24.6)

(11.2)

(11.2)

(38.9)

(4.8)

(23.1)

(42.7)

D
Cross-port flow may occur during switching. If cross-port flow is unacceptable, specify a 0.049 in. ball
orifice. Example: SS-43YFS2-049
Cross-port flow may occur during switching. If cross-port flow is unacceptable, specify a 0.093 in. ball
orifice. Example: SS-45YF8-093

E
B
A

6-Way Valve
F
G
max panel
thickness J

H
panel
hole

D
E
B

1/8 in. (3.2 mm) minimum panel thickness.

1.6

1/8

Female
NPT

H
panel
hole

Basic
Ordering
Number

4-Way Valves

4-Way Valve
F

Size

40 Series Ball Valves


O (S)
N (D)

Pneumatic Actuators
Rack and pinion actuators are compact,
lightweight, easy to mount and can
operate with standard shop air. For
technical data, including actuator types,
materials of construction, and pressuretemperature ratings, see the Swagelok
Rack and Pinion Pneumatic Actuators
catalog.

M
K

F
H

1/8 in.
NPT

1/8 in.
NPT

B
Q
A

4 mounting
holes 0.34
(8.6) dia

S = spring return
D = double acting

Dimensions
Valve Series

Dimensions, in. (mm)

Actuator
Model

41, 41-A, 41X,


42, 42-A, 42X
43, 43-A,
43X, 43Y

-131
-151

-133
-153

2.30
(58.4)

4.13

2.00 2.31 4.07 2.32 1.75 0.75 0.81 0.44 2.16 1.56 5.89 7.86

(104) 0.48 (50.8) (58.7) (103) (58.9) (44.5) (19.1) (20.6) (11.2) (54.9) (39.6) (150) (200) 2.56
(65.0)
4.25 (12.2) 2.19
(108)

(55.6)

Typical Ordering Number

SS-43S4 - 31D

Actuation Modes

-131
(90)
-151
(180)
-133
(90)

-153
(180)

-31

-51

-33

-53

3.72

Factory-Assembled Actuators

Dual

(94.5)

Ordering Information

Single

1.75 3.04 1.73 1.31 0.60 0.52 0.31 1.46 1.25 4.09 4.91 1.80
2.91 0.34 2.00 (44.5) (77.2) (43.9) (33.3) (15.2) (13.2) (7.9) (37.1) (31.8) (104) (125) (45.7)

Required pressures based on valve performance using


pressurized air or nitrogen.

Valve
Series

(71.4)

Actuator Pressure at Maximum System Pressure

Actuator Actuator
Model Designator

(73.9) (8.6) (50.8)

45, 45-A,
45X, 45Y

Double Acting

2.81

43, 43-A,
43X, 43Y
44, 44-A, 44X

Spring Return
Single

Dual

Valve Ordering Number

Minimum Actuator Pressure, psig (bar)

41, 41-A,
42, 42-A

20 (1.4)

35 (2.5)

60 (4.2)

43, 43-A, 43Y

50 (3.5)

80 (5.6)

80 (5.6)

41X, 42X

20 (1.4)

35 (2.5)

60 (4.2)

70 (4.9)

70 (4.9)

For dual-mounted assemblies


(two valves mounted to one
actuator), add DM to the
ordering number.
Example: SS-43S4-31DDM

Actuator Designator
Determine valve series,
then select actuator
model (see Actuator
Pressure table).

43X

50 (3.5)

80 (5.6)

70 (4.9)

43, 43-A, 43Y

20 (1.4)

35 (2.5)

65 (4.5)

75 (5.2)

Actuators for Field Assembly

44, 44-A

25 (1.8)

50 (3.5)

70 (4.9)

90 (6.3)

Typical Ordering Number

45, 45-A, 45Y

60 (4.2)

100 (6.9)

90 (6.3)

43X

20 (1.4)

35 (2.5)

65 (4.5)

70 (4.9)

44X

25 (1.8)

50 (3.5)

70 (4.9)

80 (5.6)

45X

60 (4.2)

100 (6.9)

85 (5.9)

Actuation Mode
D = double acting
C = spring return,
normally closed
O = spring return,
normally open
S = spring return,
3- and 4-way valves

MS-131-DA
Actuator Model
Determine valve series,
then select actuator model
(see Actuator Pressure table).

Actuation Mode
DA = double acting
SR = spring return

Other Actuator Options


For Field Assembly or Factory Assembly
Electric Actuators

For information on electric actuators, see the Electric


Actuators catalog.
Solenoid Valves
attach to the actuator to create an electropneumatically
actuated ball valve assembly.
Limit Switches
indicate actuator position by means of an electrical signal.
They meet a variety of NEMA ratings such as NEMA 4
(weatherproof) and NEMA 7 (explosion proof). For more
information, see the Limit Switches catalog.
Position Indicators
provide visual status of a valve.
ISO 5211-Compliant Actuator Mounting Bracket Kits
For information, see the Actuated Ball Valve Selection Guide.

Mounting Bracket Kits


Valve Series
41, 41-A,
42, 42-A

Actuator Model

41X, 42X

-151

43, 43-A

-131

43X

-151

43, 43-A

-133

43X

-153

43Y
44, 44-A

-131

-131
-133

44X

-153

45, 45-A

-133

45X

-153

45Y

-133

Ordering Number
MS-MB-41

MS-MB-43
MS-MB-43-133
MS-MB-43Y
MS-MB-43Y-133
MS-MB-44
MS-MB-45
MS-MB-45Y

42 series valves with


VCO or VCR end
connections mounted to
an actuator are only
available factory
assembled.
44 series valves with
VCR end connections
require kit
MS-MB-44-VCR.

Options
Handles

Vented Valves

Factory-Assembled Nylon
Black is standard. For other colors,
add a handle color designator to
the valve ordering number.
Example: SS-43S4-BL

Handle Color
Black

Designator
-BK

Blue

-BL

Green

-GR

Orange

-OG

Red

-RD

Yellow

-YW

316 Stainless Steel Bar


Add -SH to the valve ordering number.
Example: SS-43S4-SH

2-Way, Straight-Pattern Valves


When the valve is closed, the downstream port vents to
atmosphere through a vent hole in the side of the valve body.

2-Way, Angle-Pattern and 3-Way Valves


When the valve is closed, the bottom port vents to atmosphere
through a vent hole in the side of the valve body.
To order, insert V into the valve ordering number.
Example: SS-43VS4
Pressure rating for vented valves is 500 psig (34.4 bar).

Valves Assembled Without Lubrication

Nylon Oval (2- and 3-Way)

40 series ball valves assembled without lubricant are cleaned


and packaged in accordance with Special Cleaning and
Packaging (SC-11) and are adjusted for factory testing at
200 psig (13.7 bar). Valves have a pressure rating of 200 psig
(13.7 bar). Brass valves are assembled with stainless steel
rings, discs, and stem. To order, add -1466 to the valve
ordering number.
Example: SS-43S4-1466

Add -K to the valve ordering number.


Example: SS-43S4-K

Low-Temperature Service

Aluminum Bar
Add -BKB to the valve ordering number.
Example: SS-43S4-BKB

Nylon oval handles are available factory assembled only.

Handle Kits
include handle and set screw.
Aluminum
Nylon Directional Stainless Steel
Bar Handle
Bar Handle
Handle Basic
Ordering Number Ordering Number Ordering Number
41, 41-A, 41X, 42, 42-A, 42X
BZ-5K-42
SS-5K-42B
A-5S-42B-BK
Valve
Series

Live-loaded packing option extends the temperature range to


65 to 150F (53 to 65C).
For more information, see the Live-Loaded 40 Series Ball
Valves for Low-Temperature Service catalog.

Oxygen Service
For more information about hazards and risks of oxygenenriched systems, see the Oxygen System Safety technical
report.

43, 43-A, 43X, 43Y

BZ-5K-43

SS-5K-43B

A-5S-43B-BK

43Z

BZ-5K-43Z

A-5S-43ZB-BK

Sour Gas Service

44, 44-A, 44X

BZ-5K-44

SS-5K-44B

A-5S-44B-BK

45, 45-A, 45X, 45Y

BZ-5K-45

SS-5K-45B

A-5S-45B-BK

40 series ball valves with female pipe ends are available for
sour gas service. Materials for wetted valve components are
selected in accordance with NACE Standard MR0175 for
sulfide stress cracking resistant materials. Stem, rings, and
discs are alloy 400. To order, add -SG to the valve ordering
number.
Example: SS-43F4-SG

For a complete ordering number, add a handle color designator to the basic
ordering number. Example: BZ-5K-42-BK

Directional Name Plates


Directional name plates indicate the direction of flow.
A matte surface accepts ink or labels. To order, add -WN1
(blank nameplate) or -WN2 (marked nameplate) to the valve
ordering number.
Example: SS-43S4-WN1

Optional Flow Paths

Safe Product Selection


When selecting a product, the total system design must
be considered to ensure safe, trouble-free performance.
Function, material compatibility, adequate ratings,
proper installation, operation, and maintenance are the
responsibilities of the system designer and user.
Caution: Do not mix or interchange parts with those of
other manufacturers.

40 series ball valves are available with a variety of flow paths


to accommodate many special applications.
For more information, see the 40 Series Ball Valve Flow Path
Options selection guide.

Swagelok, VCR, VCOTM Swagelok Company


2001, 2002 Swagelok Company
Printed in U.S.A., MI
December 2002, R6
MS-01-60

Filters 1
www.swagelok.com

Filters

F W, F, a n d T F S e r i e s
Remove system particulate contaminants
Gas and liquid service
1/8 to 1/2 in. and 3 to 12 mm end connections
Stainless steel and brass materials

Filters

Features
All-Welded In-line Filters (FW Series)
All-welded construction provides reliable fluid containment.

Full-penetration weld between body


and element prevents bypass flow

In-line filters are for use where space is limited.


Filter is easily cleaned by backflushing.
Sintered element is available in 0.5 m nominal pore size;

pleated mesh elements are available in 2, 7, and 15 m


nominal pore sizes.
End connections include Swagelok tube fittings, NPT, and

Pleated element shown;


sintered element available

male VCR face seal fittings.

In-line Filters (F Series)


In-line filters are for use where space is limited.
Replaceable sintered elements are available in 0.5, 2, 7, 15,

60, and 90 m nominal pore sizes; replaceable strainer


elements are available in 40, 140, 230, and 440 m nominal
pore sizes.
End connections include Swagelok tube fittings, NPT, tube

Sintered element shown;


strainer element available

adapter, and male VCR face seal fittings.

Tee-Type Filters (TF Series)


Filter element can be replaced without

removing body from system.


Replaceable sintered elements are available in

0.5, 2, 7, 15, 60, and 90 m nominal pore sizes;


replaceable strainer elements are available in 40, 140, 230,
and 440 m nominal pore sizes.
End connections include Swagelok tube fittings, NPT, and

tube socket or tube butt weld ends.

Sintered element
shown; strainer
element available

Bypass port available;


see page 8

Filter Elements
F and TF Series

FW Series
Sintered

Sintered

Pleated Mesh

Strainer

Magnified 13
Traps particles as small as

0.5 m in diameter
316L SS construction

Retainer
screen

Pleated
mesh
element

Retainer
screen

Offers larger filtration area


Stainless steel construction

Traps fine particles in a

dense matrix
316 SS construction

Magnified 2
Removes larger particles
316 SS construction

Filters

Pressure-Temperature Ratings
For more information about ratings, see the Swagelok Valve Pressure-Temperature
Ratings technical bulletin.
Ratings are based on standard materials of construction. Ratings for TF series filters
with PCTFE gaskets are limited to 200F and 3000 psig (93C and 206 bar). See
page 8.
Series

FW, TF

2F, 4F

Material

6F, 8F

316 SS

Temperature, F (C)

TF
Brass

Working Pressure, psig (bar)

20 (28) to 100
200
300
400

(37)
(93)
(148)
(204)

6000
5160
4660
4280

(413)
(355)
(321)
(294)

3000
2580
2330
2140

(206)
(177)
(160)
(147)

2500
2150
1940
1780

(172)
(148)
(133)
(122)

1000 (68.9)
780 (53.7)
680 (46.8)

500
600
650
700

(260)
(315)
(343)
(371)

3980
3760
3700
2600

(274)
(259)
(254)
(248)

1990
1880
1845
1800

(137)
(129)
(127)
(124)

1660
1560
1540
1500

(114)
(107)
(106)
(103)

750
800
850
900

(398)
(426)
(454)
(482)

3520
3460
3380
3280

(242)
(238)
(232)
(225)

1760
1725
1690
1640

(121)
(118)
(116)
(112)

1460
1440
1410
1360

(100)
(99.2)
(97.1)
(93.7)

2000 (137)
1730 (119)
1470 (101)

Differential Pressure Ratings


Maximum Differential Pressure
psi (bar)
Series

Sintered
Element

Strainer
Element

Pleated
Element

FW

600 (41.3)

100 (6.8)

F, TF

1000 (68.9)

Materials of Construction

Filtration Area

Filter Body Materials


Brass
Component

316 SS

Series Material Grade/ASTM Specification

Series

Sintered
Element
in.2 (mm2)

Strainer
Element
in.2 (mm2)

Pleated
Element
in.2 (mm2)

FW

0.44 (283)

2.25 (1450)

Bonnet nut

TF

Brass/B16

316 SS/A479

2F

0.55 (350)

Bonnet

TF

Brass/B16

316 SS/A479

4F, 2TF, 4TF

1.3 (830)

1.0 (640)

Retainer screens (2)

FW

316 SS

6F, 8F, 6TF, 8TF

2.0 (1280)

1.7 (1090)

0.5 m size
316L SS
FW

2 m size
304 SS

Element

7 and 15 m
size316 SS
F, TF

Sintered316 SS
Strainer316 SS with silver solder

Spring

F, TF

302 SS

Gasket

F, TF

Aluminum/B209

Silver-plated
316 SS/A240

Brass/B16

316 SS/A479

Body

All

Retaining ring

TF

PH 15-7 Mo SS

Lubricant

Silicone based

Wetted components listed in italics.


FW series filters not available in brass.

Filters

Flow Data at 70F (20C)


FW Series
End Connections
Type
Swagelok tube
fitting,
male VCR fitting

Size

Inlet Pressure, psig (bar)

Nominal
Element
Pore Size
m

5 (0.34)

0.5

0.04 (1.1)

1/4 in.,
6 mm

2, 7, 15

Female NPT

1/4 in.

2, 7, 15

Male NPT, male to


female NPT

1/4 in.

2, 7, 15

10 (0.68)

Pressure Drop, psi (bar)

15 (1.0)

10 (0.68)

Air Flow, std ft3/min (std L/min)

5.6 (150)

0.06 (1.7)

10

0.12 (3.4)

14

(280)

50 (3.4)

(390)

0.01 (0.03)

0.04 (0.15)

0.12 (0.45)

1.7

5.5 (20)

8.3 (31)

(6.4)

4.5 (17)

14

(52)

18

(68)

3.5 (13)

11

(41)

14

(52)

Outlet is discharged to atmosphere.

F Series
Nominal
Element
Pore Size
m

Inlet Pressure, psig (bar)


5 (0.34)

10 (0.68)

Pressure Drop, psi (bar)

15 (1.0)

10 (0.68)

Air Flow, std ft3/min (std L/min)

50 (3.4)

100 (6.8)

Water Flow, U.S. gal/min (L/min)

2F Series
0.5

0.04 (1.1)

0.06 (1.7)

0.12 (3.4)

0.01 (0.03)

0.04 (0.15)

0.12 (0.45)

0.20 (5.6)

0.40 (11)

0.60 (17)

0.08 (0.30)

0.24 (0.91)

0.40 (1.5)

0.50 (14)

0.90 (25)

1.2 (34)

0.10 (0.37)

0.30 (1.1)

0.48 (1.8)

15

0.80 (22)

1.3 (36)

1.5 (42)

0.12 (0.45)

0.36 (1.3)

0.58 (2.1)

60

1.7 (48)

2.2 (62)

2.4 (68)

0.15 (0.56)

0.50 (1.8)

0.70 (2.6)

90

1.8 (51)

2.2 (62)

2.6 (73)

0.20 (0.75)

0.50 (1.8)

0.60 (2.2)

4F Series
0.5

0.12 (3.4)

0.26 (7.3)

0.48 (13)

0.04 (0.15)

0.17 (0.64)

0.29 (1.0)

0.60 (17)

1.4 (39)

2.3 (65)

0.24 (0.90)

0.86 (3.2)

1.3

(4.9)

1.4 (39)

2.9 (82)

4.7 (130)

0.40 (1.5)

1.3

(4.9)

2.0

(7.5)

15

1.2 (34)

2.9 (82)

4.7 (130)

0.50 (1.8)

1.3

(4.9)

2.1

(7.9)

60

3.1 (87)

5.9 (160)

8.5 (240)

0.90 (3.4)

3.3 (12)

4.6 (17)

90

4.1 (110)

7.5 (210)

40, 140, 230, 440

4.7 (130)

8.8 (250)

10

(280)

1.2

(4.5)

4.2 (15)

6.1 (23)

12

(340)

1.7

(6.4)

5.6 (21)

7.8 (29)

6F and 8F Series
0.5

0.36 (10)

0.86 (24)

1.6 (45)

0.09 (0.34)

0.40 (1.5)

0.76 (2.8)

1.4 (39)

2.8 (79)

4.0 (110)

0.26 (0.98)

1.1

(4.1)

1.6

3.5 (13)

(6.0)

1.8 (51)

4.2 (119)

6.8 (190)

0.64 (2.4)

2.2

(8.3)

15

1.8 (51)

4.9 (130)

7.9 (220)

0.84 (3.1)

2.6

(9.8)

60

5.1 (140)

10

(280)

15

(420)

2.0

(7.5)

6.7 (25)

10

(37)

90

6.1 (170)

11

(310)

16

(450)

2.3

(8.7)

7.6 (28)

11

(41)

40, 140, 230, 440

7.2 (200)

14

(390)

20

(560)

4.8 (18)

19

(71)

Outlet is discharged to atmosphere.

100 (6.8)

Water Flow, U.S. gal/min (L/min)

15

(56)

4.1 (15)

Filters

TF Series
Nominal
Element
Pore Size
m

Inlet Pressure, psig (bar)


5 (0.34)

10 (0.68)

Air Flow, std

Pressure Drop, psi (bar)

15 (1.0)

10 (0.68)

ft3/min (std L/min)

50 (3.4)

100 (6.8)

Water Flow, U.S. gal/min (L/min)

2TF Series
0.5

0.04 (1.1)

0.06 (1.7)

0.12 (3.4)

0.04 (0.15)

0.17 (0.64)

0.29 (1.0)

0.20 (5.6)

0.40 (11)

0.60 (17)

0.08 (0.30)

0.24 (0.90)

0.40 (1.5)

0.50 (14)

0.90 (25)

1.2 (34)

0.10 (0.37)

0.30 (1.1)

0.48 (1.8)

15

0.80 (22)

1.3 (36)

1.5 (42)

0.12 (0.45)

0.36 (1.3)

0.58 (2.1)

60

1.7 (48)

2.2 (62)

2.4 (68)

0.15 (0.56)

0.50 (1.8)

0.70 (2.6)

90

1.8 (51)

2.2 (62)

2.6 (73)

0.20 (0.75)

0.50 (1.8)

0.60 (2.2)

40, 140, 230, 440

1.8 (51)

2.3 (65)

2.6 (73)

0.20 (0.75)

0.50 (1.8)

0.60 (2.2)

4TF Series
0.5

0.12 (3.4)

0.26 (7.3)

0.48 (13)

0.04 (0.15)

0.17 (0.64)

0.29 (1.0)

0.60 (17)

1.4 (39)

2.3 (65)

0.24 (0.90)

0.86 (3.2)

1.3

(4.9)

1.4 (39)

2.9 (82)

4.7 (130)

0.40 (1.5)

1.3

(4.9)

2.0

(7.5)

15

1.2 (34)

2.9 (82)

4.7 (130)

0.50 (1.8)

1.3

(4.9)

2.1

(7.9)

60

3.1 (87)

5.9 (160)

8.5 (240)

0.80 (3.0)

2.7 (10)

3.9 (14)

90

4.1 (110)

7.5 (210)

40, 140, 230, 440

4.7 (130)

8.8 (250)

10

(280)

1.1

(4.1)

3.4 (12)

4.9 (18)

12

(340)

1.2

(4.5)

4.2 (15)

5.6 (21)

6TF and 8TF Series


0.5

0.36 (10)

0.86 (24)

1.6 (45)

0.09 (0.34)

0.40 (1.5)

0.76 (2.8)

1.4 (39)

2.8 (79)

4.0 (110)

0.26 (0.98)

1.1

(4.1)

1.6

(6.0)

1.8 (51)

4.2 (119)

6.8 (190)

0.64 (2.4)

2.2

(8.3)

3.5 (13)

15

1.8 (51)

4.9 (130)

7.9 (220)

0.84 (3.1)

2.6

(9.8)

4.1 (15)

60

5.1 (140)

10

(280)

15

(420)

1.50 (5.6)

4.8 (18)

6.7 (25)

90

6.1 (170)

11

(310)

16

(450)

1.70 (6.4)

5.5 (20)

7.6 (28)

40, 140, 230, 440

7.2 (200)

14

(390)

20

(560)

2.40 (9.0)

7.2 (27)

10

(37)

Outlet is discharged to atmosphere.

Testing

Cleaning and Packaging

Every Swagelok filter is factory tested with nitrogen at


1000 psig (69 bar) to a requirement of no detectable leakage
with a liquid leak detector.

Every Swagelok filter with VCR end connections is specially


cleaned and packaged in accordance with Swagelok
Specification SC-11.
Filters with other end connections are cleaned and packaged
in accordance with Swagelok Specification SC-10; special
cleaning and packaging in accordance with Specification
SC-11 is available as an option.

Filters

Ordering Information and Dimensions


Dimensions are for reference only and are subject to change.

FW Series

1 in. (25.4 mm) flat

Add an element designator to the basic


ordering number.

F Series and TF Series


Stainless Steel Filters
Add an element designator to the basic
ordering number.

Example: SS-4FWS-05

Example: SS-2F-2

Brass Filters
Replace SS with B in the ordering
number.

FW Series

Example: B-2F-2

Size

Nominal
Element
Pore Size
m

1/4 in.

0.5

1/4 in.

2, 7, 15

End Connections
Type
Swagelok
tube fitting

6 mm

0.5

6 mm

2, 7, 15

Female NPT

1/4 in.

Male NPT

1/4 in.

Male to
female NPT

1/4 in.

Male VCR
fitting

Dimensions, in. (mm)


Orifice

SS-4FWSSS-4FWSS-6FWS-MM-

A
2.09 (53.1)

0.187 (4.75)

SS-6FW-MM-

2, 7, 15

1/4 in.
1/4 in.

Basic Ordering
Number

2.15 (54.6)
2.13 (54.1)
2.15 (54.6)

Filters with VCR fitting end connections


are not available in brass.

F and TF Series Elements


Elements remove 95 % of particles
larger than the nominal pore size.
Nominal Pore Size
Pore Size Range Element Element
m
m
Type Designator

SS-4FW4-

0.453 (11.5)

1.57 (39.9)

SS-4FW2-

0.281 (7.14)

1.89 (48.0)

SS-4FW5-

0.281 (7.14)

1.72 (43.7)

1 to 4

1.95 (49.5)

5 to 10

2.04 (51.8)

15

11 to 25

0.5

0.5

SS-4FWS-VCR-

2, 7, 15

SS-4FW-VCR-

Dimensions shown with Swagelok tube fitting nuts finger-tight.

0.187 (4.75)

40

0.5 to 2

60

50 to 75

90

75 to 100

FW Series Elements

140

Elements remove 95 % of particles


larger than the nominal pore size.

230

440

Nominal Pore Size


Pore Size Range Element Element
m
m
Type Designator
0.5

0.5 to 2

15

Sintered

05
2

Pleated

7
15

05
Sintered

2
7
15

Strainer
Sintered

40
60
90
140

Strainer

Not available for 2F series.

230
440

Filters

F Series
End Connections

Swagelok
tube fitting

Female NPT

Size

B flat

Dimensions, in. (mm)


Series

Orifice

1/8 in.

SS-2F-

2F

0.094 (2.39)

2.35 (59.7)

9/16 (14.3)

1/4 in.

SS-4F-

4F

0.187 (4.75)

2.95 (74.9)

3/4 (19.0)

3/8 in.

SS-6F-

6F

0.281 (7.14)

3.21 (81.5)

1/2 in.

SS-8F-

8F

0.406 (10.3)

3.49 (88.6)

3 mm

SS-3F-MM-

2F

0.094 (2.39)

2.38 (60.5)

6 mm

SS-6F-MM-

4F

0.187 (4.75)

2.96 (75.2)

3/4 (19.0)

1/8 in.

SS-2F4-

2F

0.094 (2.39)

2.16 (54.9)

9/16 (14.3)

1 (25.4)
9/16 (14.3)

1/4 in.

SS-4F4-

4F

Male NPT

1/4 in.

SS-4F2-

4F

Male VCR
fitting

1/4 in.

SS-4F-VCR-

4F

1/8 in.

SS-2F-T7-

2F

0.094 (2.39)

2.29 (58.2)

9/16 (14.3)

1/4 in.

SS-4F-T7-

4F

0.187 (4.75)

2.91 (73.9)

3/4 (19.0)

Swagelok tube
fitting to tube
adapter

SWAGELOK

Type

Basic
Ordering
Number

2.87 (72.9)
0.187 (4.75)

2.69 (68.3)

3/4 (19.0)

2.82 (71.6)

Dimensions shown with Swagelok tube fitting nuts finger-tight.

TF Series
A
B

3/8 in. butt


weld dia

2 mounting holes,
M5  0.8-6H threads,
0.25 (6.4) deep

1/4 in. socket


weld dia
45

C
D
E
F flat

Bolt circle:
1.00 (25.4) for
2TF, 4TF;
1.13 (28.7) for
6TF, 8TF

Tube socket and tube butt


weld end connections

End Connections
Type

Swagelok
tube fitting

Female NPT

Male NPT
Tube socket
weld and tube
butt weld

Size

Basic
Ordering
Number

Dimensions, in. (mm)


Series

Orifice

1/8 in.

SS-2TF-

2TF

0.094 (2.39)

2.27 (57.7)

1.07 (27.2)

1/4 in.

SS-4TF-

4TF

0.174 (4.41)

2.47 (62.7)

1.06 (26.9)

3/8 in.

SS-6TF-

6TF

0.213 (5.41)

2.84 (72.1)

1.32 (33.5)

1/2 in.

SS-8TF-

8TF

0.250 (6.35)

3.04 (77.2)

1.31 (33.3)

6 mm

SS-6TF-MM-

4TF

0.172 (4.36)

2.46 (62.5)

1.06 (26.9)

8 mm

SS-8TF-MM-

6TF

0.213 (5.41)

2.84 (72.1)

1.38 (35.1)

2.86 (72.6)

1.32 (33.5)

3.04 (77.2)

1.31 (33.3)

10 mm SS-10TF-MM-

8TF

12 mm SS-12TF-MM-

8TF

1/8 in.

SS-2TF4-

2TF

1/4 in.

SS-4TF4-

4TF

1/4 in.

SS-4TF2-

4TF

3/8 in.

SS-6TF2-

6TF

1/2 in.

SS-8TF2-

8TF

1/4 and
SS-4TF-TW3/8 in.

Top

4TF

0.250 (6.35)
0.174 (4.41)
0.174 (4.41)
0.250 (6.35)

0.174 (4.41)

Dimensions shown with Swagelok nuts finger-tight.


Mounting holes not available with 1/4 in. female NPT end connections.

2.00 (50.8)
2.13 (54.1)
2.13 (54.1)
2.38 (60.5)
2.75 (69.9)
1.68 (42.7)

1.00 (25.4)

0.38 (9.7)

1.87 (47.5)

1 (25.4)

1.13 (28.7)

0.46 (11.7)

2.20 (55.9)

1 1/8 (28.6)

1.00 (25.4)

0.38 (9.7)

1.87 (47.5)

1 (25.4)

1.13 (28.7)

0.46 (11.7)

2.20 (55.9)

1 1/8 (28.6)

1.00 (25.4)

1.00 (25.4)

0.38 (9.7)

1.87 (47.5)

1.00 (25.4)

1.00 (25.4)

0.38 (9.7)

1.87 (47.5)

1.25 (31.8)

1.13 (28.7)

0.46 (11.7)

2.20 (55.9)

1 1/8 (28.6)

1.00 (25.4)

1.00 (25.4)

0.38 (9.7)

1.87 (47.5)

1 (25.4)

1 (25.4)

Options and Accessories


All Filters

F Series

Special Cleaning and Packaging

Special Alloys

To order SC-11 cleaning and packaging, if it is not standard,


add -SC11 to the filter ordering number. Filters with VCR end
connections are specially cleaned and packaged in accordance
with Swagelok Specification SC-11 as standard.

Filters of alloy C-276 are available in some sizes. Contact your


independent Swagelok sales and service representative for
more information.

Example: SS-4FWS-05-SC11

TF Series
Bypass Port

F and TF Series
Element Kits
Kits include element
and instructions.

Filter Series

Basic Kit
Ordering Number

2F

SS-2F-K4-

Select a basic kit


ordering number
and add an element
designator.

4F, 2TF, 4TF

SS-4F-K4-

6F, 8F, 6TF, 8TF

SS-8F-K4-

Example:
SS-2F-K4-05

Nominal
Pore Pore Size
Size
Range Element Element
m
m
Type
Designator
0.5

0.5 to 2

05

1 to 4

5 to 10

15

11 to 25

40

Sintered

Strainer

50 to 75

Filter
Series

2TF,
4TF

Example:
SS-2TF-F1-05
6TF,
8TF

7
15

60

The bypass
port at the filter
bottom enables
sampling or
purging. To
order, insert a
designator into
the filter
ordering
number.

90

75 to 100

140

230

440

40

Sintered

Designator

1/8 in. Swagelok


tube fitting

-F1

1/8 in. female


NPT

-F2

1/4 in. Swagelok


tube fitting

-F3

1/4 in. tube weld

-F8

1/8 in. female


NPT

-F4

1/4 in. Swagelok


tube fitting

-F5

3/8 in. Swagelok


tube fitting

-F6

1/2 in. Swagelok


tube fitting

-F7

Overall
Height
in. (mm)
2.36
(59.9)

2.09
(53.1)

2.82
(71.6)

2.21
(56.1)

2.46
(62.5)

3.14
(79.8)

3.20
(81.3)

3.42
(86.9)

60
90
140

Strainer

Bypass Port
End
Connection

Filters Without Elements

230

TF series filters can be ordered without elements. Add LE to


the basic ordering number.

440

Example: SS-2TF-LE

Not available for 2F series.

Gasket Kits
Kits include gasket and
instructions. To order a stainless
steel gasket kit, select a kit
ordering number. For other
gasket materials, replace SS with
A for aluminum or KF for PCTFE
(TF series only).

Filter
Series

Kit Ordering
Number

2F

SS-2F-K3

4F

SS-4F-K3

6F, 8F

SS-8F-K3

2TF, 4TF

SS-4TF-K2

6TF, 8TF

SS-8TF-K2

Example: A-2F-K3

Safe Product Selection


When selecting a product, the total system design must
be considered to ensure safe, trouble-free performance.
Function, material compatibility, adequate ratings,
proper installation, operation, and maintenance are the
responsibilities of the system designer and user.
Caution: Do not mix or interchange parts with those of
other manufacturers.

Swagelok, VCRTM Swagelok Company


PH 15-7 MoTM AK Steel Corp.
1998, 2000, 2002 Swagelok Company
Printed in U.S.A., MI
July 2002, R3
MS-01-92

Das könnte Ihnen auch gefallen