WinCE Upgrading Instruction

1. Download the WinCE Upgrading Package folder and save it C disk.


2. Prepare PC, instrument, power supply cable, USB mouse, RS-232 serial cable (2-3, 3-2, 5-5),
and crossover network line (The network line is crossover cable, not straight cable).
3. Close Fire wall of PC and connect instrument with PC as below figure.
Crossover net cable,
another end to PC

USB MOUSE
RS-232 cable

Power supply cable

Figure 1
4. Enter WinCE Upgrading Package folder.

WinCE software
Figure 2
5. Enter Tftpd32 folder and install the Tftp tool.

Double click to install


Figure 3
6. Click Next.

Figure 4
7. Click Install.

Figure 5
8. Click Close to finish installation of Tftp.

Figure 6
9. Click Tftp icon

in desk to run Tftp tool.

Figure 7
10. Click Browse to change Current Directory to C:\WinCE Upgrading Package,
where the Windows CE software is.
Note: Do not close Tftp tool in this step.

Figure 8
11. Enter Clinet folder.

Figure 9
12. Click

to run Clinet software and change Baud Rate as 115.2k.

Figure 10
13. Click Ok, you will see below figure.

Figure 11
14. Power on the instrument and click Enter in keyboard when you see Hit any key to stop
auto boot within 2 seconds.
Note: This step is important. If you do not click Enter within 2 seconds when you see Hit
any key to stop auto boot, you need to close cliNet and power off the instrument. 2
minutes later, go step 14 again.

Click Enter within 2 seconds

Figure 12

15. Input setenv serverip PC IP address, such as setenv serverip 192.168.1.104 when you
see SJPG2410#.
Note: 192.168.1.104 is just an example and it is IP address of my computer. As a result,
you should enter IP address of your computer here.

Figure 13
16. Then input setenv ipaddr + instrument IP address, such as setenv ipaddr 192.168.1.1
when you see SJPG2410#.
Note: As long as instrument IP address is different to PC IP address, it is ok. It means that
192.168.1 should be the same and 104 should be changed to other number (1-255), such as
192.168.1.1. For example, 202.168.1.1 and 192.168.22.1is wrong.

Figure 14
17. Input run tftp_wince

Figure 15
18. Then you will see below figure.

Figure 16
19. If Figure shows TTTTTT for all the time, please change net card as below setting and
repeat the procedure from step 9. (Properties of my computer Hardware Device
Manager Network Adapter Select the property of the internet card Advanced then
select media tipe Change the link speed to 10mbps half duplex.)
20. When you see SJPG2410# again, input run write_wince

Figure 17
21. The writing of Windows CE software is beginning.

Figure 18
22. Wait for about 3 minutes, the upgrading is finished when you see SJPG2410#.

Figure 19
23. Close cliNet, Tftp, and power off the instrument.
24. 2 minutes later, you can power on the instrument.