Beruflich Dokumente
Kultur Dokumente
Lab. 3211
In conventional flow injection analysis, the sample zone is injected into a flowing carrier
stream, and auxiliary reagents are merged with it on the way to the detector, but
sequential injection is based on a different approach (Figure 1). Using a multiport
selection (rather than injection) valve, a wash solution, sample zone and reagent zone(s)
are sequentially injected (drawn) into a holding coil connected to the common port (CP).
In this way, a stack of well-defined zones is obtained (Figure 2), which is then propelled
through a reactor into a detector. This flow reversal creates a composite zone in which
sample and reagent zones penetrate each other, due to combined axial and radial
dispersion. For a kinetic based assay, a selected section of the interdispersed reagent and
sample zone may be arrested in the observation field of the detector by stopping the flow.
The reaction rate during the stopped flow period can be monitored, and the changing
signal reflects the rate of the chemical reaction. The benefit of the stopped flow
technique is that it extends the reaction time, saves reagents, and eliminates interference
from background signals and lag phases.
The measuring cycle of the SIA technique consists of the following operations: (1)
pumping of carrier/wash solution with the selector valve in position 8, to fill the holding
coil and reactor coil lines and detector; (2) aspiration of a few microliters of sample
solution with the selector valve in position 2, while the pump makes a small reverse step;
(3) aspiration of reagent solution with the valve in position 3, using a small pump reverse
step; (4) a forward pump move with the valve in position 8, with the pumping time
adjusted so that a preselected section of the interdispersed sample/reagent zone is injected
through the reactor into the detector; (5) a stopped flow period for reaction rate
monitoring (Figure 3.); and (6) a forward pump step which expels all sample and reagent
as well as the majority of the carrier/wash solutions from the valve and from the flow
channel. When the chemical reaction is fast and when peak height is measured rather
than reaction rates, step 5 is omitted. If two reagents are used, the optional port 4 and an
appropriate flow programming are used to introduce the second reagent. The total
volume of the dispersing step(s) should be at least 4 times the volume of the reactor coil,
In order to avoid carryover when changing from one sample solution to another, the
auxiliary waste line (position 1) is used to discard the sample solution trapped in the
sampling line by aspirating into the holding coil and then pumping to the auxiliary waste.
The sampling line should be kept short, with an internal diameter of 0.5 mm.
The precise and reproducible choreography of zone sampling, sequencing, merging, and
data collection can be performed only under computer control, using Alitea Instruments
USA custom designed FIAlab for Windows software, which is compatible with both SIA
and FIA modes. For details of operation and computer configuration, consult the FIAlab
for Windows Software Users Guide, provide on the CD that accompanies the instrument.
The SIA technique is a versatile and powerful flow method that allows precise
measurements using only a single line system. Since the flow is not continuous (it only
occurs during the sequencing and measurement steps), reagent consumption and waste
disposal are minimized. In addition to single- and two-reagent chemistries, it is possible
to carry out multireagent chemistries, chemical separations, preconcentration, matrix
removal and multidetector-based assays with SIA.
line to port 4 and auxiliary waste line to port 1 (home position). The sample line should
be as short as possible to minimize the volume of sample required to flush it.
To initialize the system fill all lines with the appropriate reagents. The holding coil and
reactor coil lines are first filled with carrier, followed by filling the individual aspiration
lines at ports 2, 3, and 4, and then flushing the excess solutions from the holding coil to
the auxiliary waste. Use MachineInit.fia program.
In programming the system for measurements, it is necessary to first flush the sample
aspiration line with fresh sample by aspirating sample equal to at least four volumes of
the aspiration line into the holding coil. You can estimate the aspiration line volume from
its dimensions (2 L/cm). Then carrier is pumped to discharge the excess sample in the
holding coil to waste. The system is then ready to sample and measure.
Calibration
Make triplicate measurements on each standard solution by inserting the sample
aspiration line in the appropriate standard solution and running the program
OnePeakPO4.fia". From the measurements, determine the appropriate calibration range.
Prepare a calibration curve from the average of each reading.
Soil Sample Handling
Firstly five grams of the soil sample is weighed. The sample is then put into the oven at
100C and dried for one hour. Sample is cooled down and weighed again. The water
content of the soil is calculated in milligrams per a kilogram of dry soil. Three milliliters
of distilled water is added to dry soil sample and mixed thoroughly. Resulting solution is
then decanted and filtered, drawn into the plastic syringe and microfiltered into the vial.
This sample is then used as unknown solution and measured as calibration solutions.
Questions
1. What steps would you have to go through in phosphate measurement if you were
operating manually?
2. What is the sequence of the sample and reagents drawn in the holding coil? What
should be the sequence if you wanted to do triplicate measurement at once?
1.
v
v
v
3.
v
v
v
v
4. References (5 points)