Beruflich Dokumente
Kultur Dokumente
1.0
INTRODUCTION
Bernoullis Theorem Demonstration (Model: FM 24) apparatus consists of a classical
Venturi made of clear acrylic. A series of wall tappings allow measurement of the static
pressure distribution along the converging duct, while a total head tube is provided to
traverse along the centre line of the test section. These tappings are connected to a
manometer bank incorporating a manifold with air bleed valve. Pressurization of the
manometers is facilitated by a hand pump.
This unit has been designed to be used with a Hydraulics Bench for students to study the
characteristics of flow through both converging and diverging sections. During the
experiment, water is fed through a hose connector and students may control the flow rate
of the water by adjusting a flow regulator valve at the outlet of the test section.
The venturi can be demonstrated as a means of flow measurement and the discharge
coefficient can be determined. This test section can be used to demonstrate those
circumstances to which Bernoullis Theorem may be applied as well as in other
circumstances where the theorem is not sufficient to describe the fluid behavior.
2.0
GENERAL DESCRIPTION
The unit is mounted on a base board which is to be placed on top of the Hydraulic Bench
(Model: FM110). This base board has four adjustable feet to level the apparatus.
The main test section is an accurately machined acrylic venturi of varying circular cross
section. It is provided with a number of side hole pressure tappings, which are connected
to the manometer tubes on the rig. These tappings allow the measurement of static
pressure head simultaneously at each of 6 sections. The tapping positions and the test
section diameters are shown in Appendix A. The test section incorporates two unions, one
at either end, to facilitate reversal for convergent or divergent testing as illustrated in Figure
1 and Figure 2.
Manometer tubes
Unions
Gland Nut
Hypodermic probe
Water inlet
Test section
Adjustable feet
W ater outlet
Flow control valve
Additional tapping
A hypodermic tube, the total pressure head probe, is provided which may be positioned to
read the total pressure head at any section of the duct. This total pressure head probe may
be moved after slacking the gland nut; this nut should be re-tightened by hand after
adjustment. An additional tapping is provided to facilitate setting up. All eight pressure
tapings are connected to a bank of pressurized manometer tubes. Pressurization of the
2
manometers is facilitated by connecting any hand pump to the inlet valve on the
manometer manifold.
The unit is connected to the hydraulic bench using flexible hoses. The hoses and the
connections are equipped with rapid action couplings. The flexible hose attached to the
outlet pipe which should be directed to the volumetric measuring tank on the hydraulics
bench. A flow control valve is incorporated downstream of the test section. Flow rate and
pressure in the apparatus may be varied independently by adjustment of the flow control
valve and the bench supply control valve.
Please familiarize with the unit before operating the unit. The unit consists of the followings:
a) Venturi
The venturi meter is made of transparent acrylic with the following specifications:
Throat diameter
: 16 mm
Upstream Diameter : 26 mm
Designed Flow Rate : 20 LPM
b) Manometer
There are eight manometer tubes; each length 320 mm, for static pressure and total
head measuring along the venturi meter.
The manometer tubes are connected to an air bleed screw for air release as well as
tubes pressurization.
c) Baseboard
The baseboard is epoxy coated and designed with 4 height adjustable stands to level
the venturi meter.
d) Discharge valve
One discharge valve is installed at the venturi discharge section for flow rate control.
e) Connections
Hose Connections are installed at both inlet and outlet.
f) Hydraulic Bench
Sump tank
: 120 litres
Volumetric tank : 100 litres
Centrifugal pump : 0.37 kW, 50 LPM
2.1
Parts Identification
6
2
7
3
4
9
1. Manometer Tubes
6. Discharge Valve
2. Test Section
7. Gland Nut
3. Water Inlet
8. Hypodermic Probe
4. Unions
9. Adjustable Feet
3.0
SUMMARY OF THEORY
3.1
(3.1)
(3.2)
z
ds = dz (the change in elevation)
s
(3.3)
V
ds = dV (the change in speed)
s
(3.4)
gdz = VdV
or
dp
+ VdV + gdz = 0
(3.5)
dp V 2
+
+ gz = constant
2
(3.6)
The relation between pressure and density must be applied in this equation. For
the special case of incompressible flow, = constant, and Equation 3.6 becomes
the Bernoullis Equation.
V2
+ gz
2
constant
(3.7)
Restrictions:
i.
Steady flow
ii.
Incompressible flow
iii.
Frictionless flow
iv.
Flow along a streamline
3.2
Bernoullis Law
Bernoulli's law states that if a non-viscous fluid is flowing along a pipe of varying
cross section, then the pressure is lower at constrictions where the velocity is
higher, and the pressure is higher where the pipe opens out and the fluid stagnate.
Many people find this situation paradoxical when they first encounter it (higher
velocity, lower pressure). This is expressed with the following equation:
p
v2
+
+ z = h * = Constant
g 2g
(3.8)
Where,
p
g
v
z
h*
The terms on the left-hand-side of the above equation represent the pressure head
(h), velocity head (hv ), and elevation head (z), respectively. The sum of these
terms is known as the total head (h*). According to the Bernoullis theorem of fluid
flow through a pipe, the total head h* at any cross section is constant. In a real flow
due to friction and other imperfections, as well as measurement uncertainties, the
results will deviate from the theoretical ones.
In our experimental setup, the centerline of all the cross sections we are
considering lie on the same horizontal plane (which we may choose as the datum,
z = 0, and thus, all the z values are zeros so that the above equation reduces to:
p
v2
+
= h * = Constant
g 2g
(3.9)
3.3
pressure tapping, placed in a region where the flow streamlines are straight, as
shown in Figure 4 (a). The pressure tap is a small hole, drilled carefully in the wall,
with its axis perpendicular to the surface. If the hole is perpendicular to the duct
wall and free from burrs, accurate measurements of static pressure can be made
by connecting the tap to a suitable pressure measuring instrument.
Flow
streamlines
Pressure
tap
Small holes
Flow
Stem
To manometer or
pressure gage
In a fluid stream far from a wall, or where streamlines are curved, accurate static
pressure measurements can be made by careful use of a static pressure probe,
shown in Figure 4 (b). Such probes must be designed so that the measuring holes
are place correctly with respect to the probe tip and stem to avoid erroneous
results. In use, the measuring section must be aligned with the local flow direction.
Static pressure probes or any variety of forms are available commercially in sizes
as small as 1.5 mm (1/16 in.) in diameter. The stagnation pressure is obtained
when a flowing fluid is decelerated to zero speed by a frictionless process. In
incompressible flow, the Bernoulli Equation can be used to relate changes in
speed and pressure along a streamline for such a process. Neglecting elevation
differences, Equation 3.7 becomes
p
v2
= constant
2
(3.10)
If the static pressure is p at a point in the flow where the speed is v, then the
stagnation pressure, Po, where the stagnation speed, Vo, is zero, may be
computed from
0
Vo2 p V 2
+
= +
2 2
po
(3.11)
Therefore,
1
p o = p + V 2
2
(3.12)
(3.13)
Or
V =
2( p o p )
(3.14)
Thus, if the stagnation pressure and the static pressure could be measured at a
point, Equation 3.14 would give the local flow speed.
Flow
Small hole
To manometer or
pressure gage
Flow
Total
head
tube
po
Small holes
Flow
B
C
Stem
p
po
Stagnation pressure is measured in the laboratory using a probe with a hole that
faces directly upstream as shown in Figure 5. Such a probe is called a stagnation
pressure probe (hypodermic probe) or Pitot (pronounced pea-toe) tube. Again, the
measuring section must be aligned with the local flow direction.
We have seen that static pressure at a point can be measured with a static
pressure tap or probe (Figure 4). If we know the stagnation pressure at the same
point, then the flow speed could be computed from Equation 3.14. Two possible
experimental setups are shown in Figure 6.
In Figure 6(a), the static pressure corresponding to point A is read from the wall
static pressure tap. The stagnation pressure is measured directly at A by the total
head tube, as shown. (The stem of the total head tube is placed downstream from
the measurement location to minimize disturbance of the local flow)
Two probes often are combined, as in the Pitot-static tube shown in Figure 6(b).
The inner tube is used to measure the stagnation pressure at point B, while the
static pressure at C is sensed using the tapping on the wall. In flow fields where
the static pressure variation in the streamwise direction is small, the Pitot-static
tube may be used to infer the speed at point B in the flow by assuming pB =pC and
using Equation 3.14. (Note that when pB pC, this procedure will give erroneous
results)
Remember that the Bernoulli equation applies only for incompressible flow (Mach
number, M 0.3).
Note:
Ma =
u
c
(3.15)
Where,
u = fluid velocity
c = sonic velocity
3.4
Venturi Meter
The venturi meter consists of a venturi tube and differential pressure gauge. The
venturi tube has a converging portion, a throat and a diverging portion as shown in
the figure below. The function of the converging portion is to increase the velocity
of the fluid and lower its static pressure. A pressure difference between inlet and
throat is thus developed, which pressure difference is correlated with the rate of
discharge. The diverging cone serves to change the area of the stream back to the
entrance area and convert velocity head into pressure head.
10
p1
v1
p
v
+ Z1 = 2 + 2 + Z 2
2g
2g
(3.16)
p1 p 2
22
A
1 2
+ Z1 Z 2 =
2 g A1
(3.17)
Ideally,
A 2
Qi = A2V2 = A2 1 2
A1
1 / 2
p1 p 2
+ Z 1 Z 2
2 g
1/ 2
(3.18)
However, in the case of real fluid flow, the flow rate will be expected to be less
than that given by equation (3.18) because of frictional effects and consequent
head loss between inlet and throat. Therefore,
A
Qa = C d A2 1 2
A1
1 2
p1 p 2
+ Z 1 Z 2
2 g
12
(3.19)
Cd =
Qa
Qi
(3.20)
Discharge coefficient, Cd usually lies in the range between 0.9 and 0.99.
11
4.0
12
Note:
The manometer tube connected to the tapping adjacent to the outlet flow
control valve is used as a datum when setting up equivalent conditions for flow
through test section.
15. The actual flow of water can be measured using the volumetric tank with a
stop watch.
4.2
13
5.0
Results:
Volume
(L)
Time
(s)
Qa
(LPM)
hA
14
hB
hE
hF
5.2
Results:
Volume
(L)
Time
(s)
Qa
(LPM)
hA
15
hB
hE
hF
5.3
Results:
Cross
Section
i
hi
(mm)
ViB =
[2*g*(h* - hi )]
(m/s)
Using Continuity
equation
Ai =
ViC =
2
Di / 4 Qav / Ai
(m/s)
(m2)
A
B
C
D
E
F
* Please refer to Appendix C for Cross Section Diameter
16
Difference
ViB-ViC
(m/s)
6.0
It is important to drain all water from the apparatus when not in use. The apparatus
should be stored properly to prevent damage.
2.
Any manometer tube, which does not fill with water or slow fill, indicates that tapping
or connection of the manometer is blocked. To remove the obstacle, disconnect the
flexible connection tube and blow through.
3.
4.
Always wear protective clothing, shoes, helmet and goggles throughout the
laboratory session.
5.
Always run the experiment after fully understand the unit and procedures.
17
7.0
REFERENCES
Applied Fluid Mechanics 5th Edition, Robert L. Mott, Prentice-Hall
Elementary Fluid Mechanics 7th Edition, Robert L. Street, Gary Z. Watters, John K.
Vennard, John Wiley & Sons Inc.
Fluid mechanics 4th Edition, Reynold C. Binder
Fluid Mechanics with applications, Anthony Esposito, Prentice-Hall International Inc.
18