CoriolisMaster FCM2000
Mass Flowmeter
Options
Expanded density calibration with temperature
compensation
Pressure-resistant housing
EHEDG-certified
Concentration measurement
D184S068U02
Contents
1
1.2
Model FCM2000-MC2.........................................................................................................................................11
3.1
Specifications ...............................................................................................................................................11
3.2
Dimensions...................................................................................................................................................14
3.3
Ordering information.....................................................................................................................................28
Model FCM2000-MS2.........................................................................................................................................31
4.1
Specifications ...............................................................................................................................................31
4.2
Dimensions...................................................................................................................................................33
4.3
Ordering information.....................................................................................................................................36
Transmitter.........................................................................................................................................................38
5.1
Specifications ...............................................................................................................................................38
5.2
5.3
Inputs/outputs...............................................................................................................................................39
5.4
5.5
Electrical connections...................................................................................................................................43
5.6
Dimensions...................................................................................................................................................52
5.7
Ordering information.....................................................................................................................................54
8
Data Sheet
Installation Requirements...............................................................................................................................6
7.2
7.3
Electrical data...............................................................................................................................................64
Questionnaire ....................................................................................................................................................67
Mass Flowmeter
CoriolisMaster FCM2000
D184S068U02
MC2
MS2
G00315
G00414
G00316
Flowmeter sensor
Model number
Flange DIN 2501/EN 1092-1
Flange ASME B16.5
Threaded pipe connection
conforming to DIN 11851
Tri-Clamp
Aseptic flange DIN 11864-2
"G" threaded pipe connection
NPT threaded pipe connection
Accuracy of mass flowrate
Accuracy of density
Accuracy of temperature
Materials in contact with fluid
Ingress protection acc. to EN
60529
Fluid temperature
(see Section 3/4 of the data
sheet, Section 10 of the
operating instructions)
Approvals
Explosion protection conforming
to ATEX, IECEx
(KEM 08 ATEX 0150X/0151X),
(IECEx KEM08 00.0034X)
Explosion protection conforming
to cFMus
(PID: 3036514)
Other approvals for potentially
explosive areas
Hygienic and sterile
requirements
Transmitter
Model number
Housing
Cable length
Supply power
Current output 1
Current output 2
Pulse output
Ext. output switch-off
Ext. totalizer reset
Forward/reverse flow metering
Communication
Empty pipe detection
Self-monitoring, diagnostics
On-site display/totalization
Field optimized flow/density
Ingress protection acc. to EN
60529
Standard
Hygienic (EHEDG)
Standard
MC2
DN
PN
15 ... 150
40 ... 100
1/2 ... 6
CL 150 ... CL 600
DN 15 ... DN 100 (1/2 ... 4")
MC2_ _ _ 3
DN
PN
1/4
1/4
0.15 % / 0.25 % / 0.4
0.01 kg/l
1K
Stainless steel 1.4435 (316L),
Hastelloy C-22
IP 67
Zone 0/1/2
Dust-ignition-proof
Zone 0/1/2
Dust-ignition-proof
Class I Div. 1
Class I Div. 2
Class I Div. 1
Class I Div. 2
MS2
DN
10 / 15
1/2
PN
40 / 100
CL 150 ... CL 600
DN 10 (3/8")
ME2_/MC23, MC27
ME2_
Separate, field-mount housing/compact housing
Separate, field-mount housing
Up to 50 m (164 ft.);
5, 10, 20, or 50 m
300 m (984 ft.) on request
(16, 32, 65, or 164 ft.)
100 ... 230 V AC, 24 V AC/DC
Active: 0/4 ... 20 mA or passive: 4 ... 20 mA
Passive: 4 ... 20 mA
Active (non-Ex) or passive
Yes
Yes
Yes
HART protocol, PROFIBUS PA, FOUNDATION Fieldbus
Yes, based on preconfigured density alarm < 0.5 kg/l
Yes
Yes
Yes
ME2: IP 65/67, NEMA 4X
MC _ _: IP 67, NEMA 4X
D184S068U02
1.1
Standard/non-Ex
MC23 A, U
Type
Zone 2/21, 22
MC23 M, N
Zone 1/21
MC27 B, E
1. Integral mount
design
Standard/non-Ex
Ex Zone 2/21, 22
Ex Zone 1/21
Type
ME21 A, U
MC21 A, U
ME21 M, N
MC21 M, N
ME26 B, E
MC26 B, E
2. Remote mount
design
Transmitter and
flowmeter sensor
Standard/non-Ex
Ex Zone 1/21
Type
ME21 A, U
ME21 M, N
MC26 B, E
3. Remote mount
design
Transmitter
Standard/non-Ex
Ex Zone 2/21, 22
Flowmeter sensor
Ex Zone 1/21
Type
ME21 A, U
4. Remote mount
design
(small nominal
diameters)
Transmitter
Standard/non-Ex
Ex Zone 2/21, 22
Flowmeter sensor
Ex Zone 2/21, 22
Type
ME22 A, U ...
5. Remote mount
design
(small nominal
diameters)
Transmitter and
flowmeter sensor
Standard/non-Ex
Ex Zone 2/21, 22
Ex Zone 1/21
Type
6. Remote mount
design
(small nominal
diameters)
Transmitter
Standard/non-Ex
Ex Zone 2/21, 22
Flowmeter sensor
Ex Zone 1/21
Fig. 1:
FCM2000 overview
MS21 A, U
ME24/25 A, U ...
ME21 M, N
MC21 M, N
ME27/28 B, E
MS26 B, E
MS26 B, E
G00387
D184S068U02
1.2
Pos: 10 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/bersichten/Gertebersicht DN 1,5 ... DN 150 FM/CSA (SM) @ 30\mod_1245162902390_3101.doc @ 279724 @ @ 1
MC2x
Type
1. Integral mount
design
Standard/non-Ex
Class I Div. 2
Class I Div. 1
Type
2. Remote mount
design
Transmitter and
flowmeter sensor
Standard/non-Ex
Class I Div. 2
Class I Div. 1
Type
3. Remote mount
design
Transmitter
Standard/non-Ex
Class I Div. 2
Flowmeter sensor
Class I Div. 1
MS2x
Type
Standard/non-Ex
Class I Div. 2
Class I Div. 1
MC23 T, X
MC23 O, V, P, W
MC27 C, Y, D, Q
ME21 T, X
MC21 T, X
ME21 T, X
ME22/23 T, X...
MS21 T, X
ME21 O, V, P, W
MC21 O, V, P, W
ME21 O, V, P, W
ME22/23 O, V, P,
W
MS21 O, V, P, W
Housing: XP
ME26 C, Y, D, Q MC26 C, Y, D, Q
MC26 C, Y, D, Q
ME27/28 C, Y,
D, Q
MS26 C, Y, D, Q
4. Remote mount
design
(small nominal
diameters)
Transmitter
Standard/non-Ex
Class I Div. 1
Flowmeter sensor
Class I Div. 2
Class I Div. 1
ME24/25 T, X ...
ME24/25 O, V, P, W
MS26 C, Y, D, Q
5. Remote mount
design
(small nominal
diameters)
Flowmeter sensor
Class I Div. 1
G00899
Fig. 2:
FCM2000 overview
D184S068U02
General information
2.1
Installation Requirements
2.1.1
General information
Pos: 16.3 /Standard-Layout/==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828 @ @ 1
Change from one to two columns
Pos: 16.4 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Einbaubedingungen/Allgemeine Hinweise @ 5\mod_1166449207375_3101.doc @ 52671 @ 55555555 @ 1
Inspection
Before installing the flowmeter sensor, check for physical damage due
to possible improper handling during shipment. All claims for damage
are to be made promptly to the shipper.
Installation position
The FCM2000 operates in any orientation. The optimal installation
orientation is vertical with the flow upwards.
Supports
In order to support the weight of the flowmeter sensor and to ensure
reliable measurements when adverse external effects exist (e. g.,
vibrations), the primary should be installed in rigid pipelines. Two
supports or hangers should be installed symmetrically and stress free
in close proximity to the in- and outlet process connections.
D184S068U02
Zero balance
In order to adjust the zero under operating conditions, it must be
possible to reduce the flowrate "ZERO" while the meter tube remains
completely filled. A bypass line is optimal when the process cannot be
shut down. It is important for accurate measurements that during the
zero adjustment there are no gas bubbles in the flowmeter sensor. It
is also important that the pressure and temperature in the meter tube
be the same as those which exists during operation.
G00311
G00361
Fig. 3:
Fig. 4:
Vibrations
D184S068U02
2.1.2
G00304
Fig. 7:
G00302
Fig. 5:
3
4
G00303
Fig. 6:
G00305
Fig. 8:
1
2
3
4
5
Supply reservoir
Flowmeter sensor
Orifice/pipe constriction
Valve
Product reservoir
D184S068U02
2.1.3
The accumulation of air or gas bubbles in the meter pipe can lead to
increased inaccuracies. Fig. Fig. 9 shows critical installation conditions.
Installations at the highest point in the pipeline (Figure A) can result in
the formation of air pockets which can lead to appreciable
inaccuracies.
Another difficult installation condition is immediately upstream of a
free discharge (Figure B) in a drop line.
G00307
G00306
Fig. 9:
Vertical
Important:
Check whether the coordination flowmeter sensor and transmitter is
correct. Compatible devices have the same end numbers, e. g., X001
and Y001 or X002 and Y002, on the type plate.
Pressure losses
The pressure losses, at the operating conditions, depends on the fluid
and the actual flow. To calculate the pressure loss, the design
program CD-CALC is provided.
G00308
Fig. 10
D184S068U02
45
90
G00371
Fig. 11:
Fig. 14:
G00374
Important
If there are large differences between the fluid and ambient
temperature the sensor must be insulated, to prevent two
phase flow and accuracy effects. This is especially important
for low flowrates.
The sensor must always be completely filled with
homogeneous liquid or a single phase gas, otherwise the
accuracy could be adversely affected.
For air/gas in volatile fluids horizontal installations are
recommended.
0 mm
Vertical
2 mm (0,08)
Fig. 12:
G00372
G00373
G00310
Fig. 13:
Fig. 15
10
D184S068U02
Model FCM2000-MC2
3.1
Specifications
Fig. 16:
Nominal sizes
"E" (DN 20); "F" (DN 25); "G" (DN 40); "H" (DN 50);
"I" (DN 65); "J" (DN 80); "K" (DN 100); "L" (DN 150)
Measuring ranges for flowrate
Nominal size
Max. measuring
range
[Qmax] in [kg/min]
"E"
"F"
"G"
"H"
"I"
"J"
"K"
"L"
DN 20 (3/4")
DN 25 (1")
DN 40 (1
1/2")
DN
50 (2")
DN 65 (2
1/2")
DN 80 (3")
DN 100 (4")
DN 150 (6")
0 ...
0 ...
0 ...
100
160
475
0 ...
0 ...
920
1,890
0 ...
0 ...
0 ...
2,460
4,160
11,000
11
3.1.1
D184S068U02
Reference conditions
Calibration fluid
Water 25 C (77 F) (+ 5 K/- 5 K)
Pressure 0.5 ... 6 bar (7.3 ... 87.0 psi)
Ambient temperature
25 C (77 F) (+ 10 K/- 5 K)
Supply power
Line voltage as per name plate UN 1 %
Process connections
Flange DIN/ASME
Tri-clamp DIN 32676 (ISO 2852)
- DN 15 ... DN 50 (1/2 ... 2"): series 3
- DN 65 ... DN 100 (2 1/2 ... 4"): series 1
Threaded pipe connection conforming to DIN 11851
The max. permissible operating pressure is determined by the
respective process connection, the fluid temperature, the screws, and
the gasket material.
Warm-up phase
30 minutes
Pressure rating
PN 16, PN 40, PN 100 (up to DN 80 [3"])
CI 150, CI 300, CI 600 (up to DN 80 [3"])
Output calibration
Pulse output
Effect of the analog output on the measurement accuracy
Similar to pulse output 0.1 % of measured value
3.1.2
12
D184S068U02
Threaded
connection
conforming
11851
Tri-Clamp
conforming
32676
3.1.3
pipe
to
DIN
to
DIN
Nominal size
DN
15 ... 40
(1/2 ... 1 1/2")
50 ... 100
(2 ... 4")
15 ... 50
(1/2 ... 2")
65 ... 100
(2 1/2 ... 4")
PSmax
[bar]
40
TSmax
[C]
140
TSmin
[C]
-40
25
140
-40
16
120
-40
10
120
-40
10 000
1 000
Pressure [mbar]
Process connection
100
10
E F
IJ K
1
0,1
120
10
100
1000
10 000
100 000
G00314
1740
PN 100
80
PN 63
60
870
PN 40
40
580
20
290
PN 16
Fig. 19:
1160
-50
-58
0
32
50
122
100
212
150 [C]
302 [F]
Pressure [psi]
Pressure [bar]
100
Viscosity range
Max. dyn. viscosity: 1 Pas (= 1,000 mPas = 1,000 cP)
At higher viscosities, please contact our ABB representatives.
Pos: 22 /Standard-Layout/======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ @ 1
0
G00312
Temperature
Fig. 17:
1740
120
1450
1160
80
870
60
CI 300
580
40
CI 150
20
290
Pressure [bar]
Pressure [bar]
100
CI 600
0
-50
-58
0
32
50
122
100
212
150 [C]
302 [F]
G00313
Temperature C
Fig. 18:
13
D184S068U02
3.2
Dimensions
Pos: 24.2 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Flansch DN20-DN25 (getrennt) @ 5\mod_1167919349281_3101.doc @ 56002 @ 3 @ 1
3.2.1
Design MC21
80 (3,15)
2
B
L -5
G00339
Fig. 20:
Dimensions in mm (inch)
1 Flange DIN 2635 / ASME / ISO 7005 (connecting dimensions for ASME flanges acc. to ASME B16.5 (ANSI))
2 Flow direction
Meter
size
DN
Process
connection
L-5
DIN 2635 DIN 2636 DIN 2637 ASME ASME ASME DIN 11864-2
form A1)
PN 40
PN 64
PN 100 CL 150 CL 300 CL 600
Weight
kg (lb)
DN 15
(1/2)
693
(27.28)
705
(27.76)
705
(27.76)
708
718
730
(27.87) (28.27) (28.74)
672
(26.46)
E (20) DN 20
(3/4)
598
(23.54)
618
628
645
(24.33) (24.72) (25.39)
583
(22.95)
DN 25
(1)
698
(27.48)
735
(28.94)
735
(28.94)
728
738
753
(28.66) (29.06) (29.65)
683
(26.89)
16
(35.3)
DN 20
(3/4)
758
(29.84)
778
788
802
(30.63) (31.02) (31.57)
743
(29.25)
16
(35.3)
(1)
658
(25.91)
693
(27.28)
693
(27.28)
688
698
710
(27.09) (27.48) (27.95)
643
(25.31)
808
(31.81)
840
(33.07)
840
(33.07)
838
855
868
(32.99) (33.66) (34.17)
786
(30.94)
F (25) DN 25
DN 40 (1 1/2)
All dimensions in mm (inch)
1) Antiseptic grooved flange DIN 11864-2, form A, for pipes acc. to DIN 11866
Pos: 24.3 /Standard-Layout/======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ @ 1
14
15
(33.1)
358
127
66
89
(14.09) (5.00) (2.60) (3.50)
358
127
66
89
(14.09) (5.00) (2.60) (3.50)
15
(33.1)
16
(35.3)
19
(41.9)
D184S068U02
Pos: 24.4 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Flansch DN40-DN150 (getrennt) @ 5\mod_1167919375625_3101.doc @ 56023 @ @ 1
80 (3,15)
2
B
-5
G00340
Fig. 21:
Dimensions in mm (inch)
1 Flange DIN 2635 / ASME B16.5 / ISO 7005 (connecting dimensions for ASME flanges acc. to ASME B16.5 (ANSI))
2 Flow direction
15
Meter
size
DN
Process
connection
D184S068U02
L-5
DIN 2633 DIN 2635 DIN 2636 DIN 2637 ASME ASME ASME DIN 11864-2
PN 16
PN 40
PN 64 PN 100 CL 150 CL 300 CL 600 form A1)
Wt.
kg
(lb)
879
(34.61)
914
(35.98)
914
910
922
932
(35.98) (35.83) (36.30) (36.69)
864
(34.02)
20
(44.1)
DN 40 (1 1/2)
780
(30.71)
813
(32.01)
813
810
825
843
(32.01) (31.89) (32.48) (33.19)
761
(29.96)
374
129
64
90
22
(14.72) (5.08) (2.52) (3.54) (48.5)
DN 50
940
(37.01)
967
(38.07)
979
970
980
1001
(38.54) (38.19) (38.58) (39.41)
918
(36.14)
23
(50.7)
DN 40 (1 1/2)
1045
(41.14)
1078
(42.44)
1078
1075 1090 1108
(42.44) (42.32) (42.91) (43.62)
1025
(40.35)
32
(70.5)
DN 50
940
(37.01)
967
(38.07)
979
970
980
1001
(38.54) (38.19) (38.58) (39.41)
918
(36.14)
403
148
80
110
34
(15.87) (5.83) (3.15) (4.33) (75.0)
DN 65 (2 1/2)
1100
(43.31)
1132
(44.57)
1148
1218 1228 1248
(45.20) (47.95) (48.35) (49.13)
1081
(42.56)
38
(83.8)
DN 50
1220
(48.03)
1248
(49.13)
1259
1250 1260 1281
(49.57) (49.21) (49.61) (50.43)
1197
(47.13)
43
(94.8)
DN 65 (2 1/2)
1100
(43.31)
1132
(44.57)
1148
1218 1228 1249
(45.20) (47.95) (48.35) (49.17)
1081
(42.56)
429
164
97
130
48
(16.89) (6.46) (3.82) (5.12) (105.8)
DN 80
1220
(48.03)
1248
(49.13)
1260
1240 1260 1282
(49.61) (48.82) (49.61) (50.47)
1200
(47.24)
50
(110.2)
DN 65 (2 1/2)
1330
(52.36)
1362
(53.62)
1378
1365 1375 1396
(54.25) (53.74) (54.13) (54.96)
1310
(51.57)
56
(123.5)
DN 80
(3)
1220
(48.03)
1248
(49.13)
1260
1240 1260 1282
(49.61) (48.82) (49.61) (50.47)
1200
(47.24)
456
186
108 140
58
(17.95) (7.32) (4.25) (5.51) (127.9)
DN 100
(4)
1480
(58.27)
1494
(58.82)
1530
1500 1520 1568
(60.24) (59.06) (59.84) (61.73)
1463
(57.60)
69
(152.1)
DN 80
(3)
1640
(64.57)
1668
(65.67)
1680
1660 1680 1702
(66.14) (65.35) (66.14) (67.01)
1618
(63.70)
500
215
131 170
84
(19.69) (8.46) (5.16) (6.69) (185.2)
K (100) DN 100
(4)
1450
(57.09)
1480
(58.27)
1494
(58.82)
1530
1500 1520 1568
(60.24) (59.06) (59.84) (61.73)
1463
(57.60)
91
(200.6)
DN 150
(6)
1736
(68.35)
1776
(69.92)
1816
(71.50)
1806 1826
(71.10) (71.89)
120
(264.6)
L (150) DN 150
(6)
2000
(78.74)
2040
(80.31)
2080
(81.89)
2070 2090
(81.50) (82.28)
613
285
190 250
240
(24.13) (11.22) (7.48) (9.84) (529.1)
DN 25
G (40)
H (50)
(65)
J (80)
(1)
(2)
(2)
(2)
(3)
1450
(57.09)
16
D184S068U02
Pos: 24.6 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Rohrverschraubung DN20-DN25 (getrennt) @ 5\mod_1167919527843_3101.doc @ 56086 @ @ 1
Dimensions, remote design, food industry fitting "E" to "F", DIN 11851
80 (3,15)
80 (3,15)
2
g
L -5
G00341
Fig. 22:
Dimensions in mm (inch)
Meter Size
DN
Process connection
DN 15 / 1/2"
"E"
"F"
(20)
(25)
Rd 34 x 1/8
L-5
152 (5.98)
DN 20 / 3/4"
Rd 44 x 1/6
102 (4.02)
DN 25 / 1"
Rd 52 x 1/6
152 (5.98)
DN 20 / 3/4"
Rd 44 x 1/6
162 (6.38)
DN 25 / 1"
Rd 52 x 1/6
112 (4.41)
DN 40 / 1/2"
Rd 65 x 1/6
185 (7.28)
Weight
kg (lb)
13 (28.7)
14 (30.9)
17
D184S068U02
Pos: 24.8 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Rohrverschraubung DN40-DN100 (getrennt) @ 5\mod_1167919559390_3101.doc @ 56107 @ @ 1
80 (3,15)
80 (3,15)
L -5
G00342
Fig. 23:
Dimensions in mm (inch)
Meter Size
DN
Process connection
DN 25 / (1")
Rd 52 x 1/6
64 (2.52)
90 (3.54)
Weight
kg (lb)
218 (8.58)
16 (35.3)
164 (6.46)
18 (39.7)
241 (9.49)
19 (41.9)
DN 40 / (1 1/2") Rd 65 x 1/6
233 (9.17)
28 (61.7)
(50) DN 50 / (2")
177 (6.97)
30 (66.1)
DN 65 / (2 1/2") Rd 95 x 1/6
254 (10.00)
34 (75.0)
DN 50 / (2")
291 (11.46)
40 (88.2)
Rd 78 x 1/6
227 (8.94)
44 (97.0)
Rd 78 x 1/6
(80) DN 80 / (3")
DN 100 / (4")
(100) DN 80 / (3")
DN 100 / (4")
80 (3.15)
97 (3.82)
130 (5.12)
Rd 110 x 1/4 1205 (47.44) 8 (0.31) 456 (17.95) 186 (7.32) 108 (4.25) 140 (5.51) 258 (10.16) 56 (123.5)
Rd 130 x 1/4 1463 (57.60) 10 (0.39)
Rd 110 x 1/4 1618 (63.70) 8 (0.31)
Rd 130 x 1/4 1463 (57.60) 10 (0.39)
110 (4.33)
18
DN 65 / (2 1/2") Rd 95 x 1/6
Rd 78 x 1/6
DN 80 / (3")
J
L-5
D184S068U02
Pos: 24.10 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Tri-Clamp DN20-DN25 (getrennt) @ 5\mod_1167919625125_3101.doc @ 56128 @ @ 1
80 (3,15)
1
L -5
B
G00343
Fig. 24:
Dimensions in mm (inch)
1 Flow direction
Meter Size
DN
Process connection
DN 15 (1/2)
(20)
DN 20 (3/4)
DN 25 (1)
DN 40 (1 1/2)
561 (22.09)
358 (14.09)
127 (5.00)
66 (2.60)
89 (3.50)
621 (24.45)
Weight
kg (lb)
92 (3.62)
12 (26.5)
142 (5.59)
721 (28.39)
DIN 32676
R
140 (5.51)
661 (26.02)
DN 20 (3/4)
(25)
656 (25.83)
DIN 32676
DN 25 (1)
F
L-5
152 (5.98)
358 (14.09)
127 (5.00)
66 (2.60)
773 (30.43)
89 (3.50)
102 (4.02)
13 (28.7)
180 (7.09)
19
D184S068U02
Pos: 24.12 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Tri-Clamp DN40-DN100 (getrennt) @ 5\mod_1167919697953_3101.doc @ 56170 @ @ 1
80 (3,15)
80 (3,15)
1
L -5
G00344
Fig. 25:
Dimensions in mm (inch)
1 Flow direction
Meter Size
DN
G
J
K
(40)
(50)
(65)
(80)
(100)
L-5
R
242 (9.53)
17 (37.5)
374 (14.72)
129 (5.08)
64 (2.52)
90 (3.54)
195 (7.68)
17 (37.5)
278 (10.94)
18 (39.7)
DN 25
(1)
842 (33.15)
DN 40
(1 1/2)
748 (29.45)
DN 50
(2)
913 (35.94)
DN 40
(1 1/2)
1012 (39.84)
DN 50
(2)
913 (35.94)
DN 65
(2 1/2)
1073 (42.24)
DN 50
(2)
1192 (46.93)
DN 65
(2 1/2)
DN 80
(3)
1180 (46.46)
DN 65
(2 1/2)
1302 (51.26)
DN 80
(3)
DN 100
(4)
1448 (57.01)
403 (15.87)
DN 80
(3)
DN 100
(4)
1448 (57.01)
148 (5.83)
164 (6.46)
186 (7.32)
215 (8.46)
80 (3.15)
97 (3.82)
108 (4.25)
131 (5.16)
20
110 (4.33)
130 (5.12)
140 (5.51)
170 (6.69)
Weight
kg (lb)
275 (10.83)
27 (59.5)
225 (8.86)
26 (57.3)
305 (12.01)
27 (59.5)
335 (13.19)
36 (79.4)
275 (10.83)
37 (81.6)
328 (12.91)
38 (83.8)
378 (14.88)
45 (99.2)
296 (11.65)
44 (97.0)
430 (16.93)
46 (101.4)
440 (17.32)
71 (156.5)
365 (14.37)
69 (152.1)
D184S068U02
Pos: 24.14 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Flansch DN20-DN25 (kompakt) @ 5\mod_1167919152250_3101.doc @ 55960 @ 5 @ 1
3.2.2
134
(5,27)
80 (3,15)
2
B
L -5
G00333
Fig. 26:
Dimensions in mm (inch)
1 Flange DIN 2635 / ASME B16.5 / ISO 7005 (connecting dimensions for ASME flanges acc. to ASME B16.5 (ANSI))
2 Flow direction
Meter
Size
DN
20 (E)
25 (F)
G
MC23 MC26/
MC27
89
127
66
470
494
(3.50) (5) (2.60) (18.50) (19.45)
89
127
66
470
494
(3.50) (5) (2.60) (18.50) (19.45)
L-5
Process
connec- DIN 2635 DIN 2636 DIN 2637 ASME
tion DN
PN 40
PN 64
PN 100 CL 150
ASME
CL 300
Weight
kg (lb)
15
(1/2)
693
(27.28)
705
(27.76)
705
(27.76)
708
(27.87)
718
(28.27)
730
(28.74)
672
(26.46)
16
20
(3/4)
598
(23.54)
618
(24.33)
628
(24.72)
645
(25.39)
583
(22.95)
16
25
(1)
698
(27.48)
735
(28.94)
735
(28.94)
728
(28.66)
738
(29.06)
753
(29.65)
683
(26.89)
17
20
(3/4)
758
(29.84)
778
(30.63)
788
(31.02)
802
(31.57)
743
(29.25)
17
25
(1)
658
(25.91)
693
(27.28)
693
(27.28)
688
(27.09)
698
(27.48)
710
(27.95)
643
(25.31)
17
40
(1 1/2)
808
(31.81)
840
(33.07)
840
(33.07)
838
(32.99)
855
(33.66)
868
(34.17)
786
(30.94)
20
21
D184S068U02
Pos: 24.16 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Flansch DN40-DN150 (kompakt) @ 5\mod_1167919205890_3101.doc @ 55981 @ 5 @ 1
134
(5,27)
80 (3,15)
2
B
L -5
G00334
Fig. 27:
Dimensions in mm (inch)
1 Flange DIN 2635 / ASME B16.5 / ISO 7005 (connecting dimensions for ASME flanges acc. to ASME B16.5 (ANSI))
2 Flow direction
22
Meter
Size
DN
D184S068U02
L-5
Process
Weight
conn.
MC23 MC26/
DIN 2633 DIN 2635 DIN 2636 DIN 2637 ASME ASME ASME DIN 11864-2 kg (lb)
DN
MC27
PN 16
PN 40
PN 64 PN 100 CL 150 CL 300 CL 600 form A1)
25
(1)
40 (G)
50 (H)
65 (I)
80 (J)
879
(34.61)
914
(35.98)
914
910
922
932
(35.98) (35.83) (36.30) (36.69)
864
(34.02)
21
(46.3)
780
(30.71)
813
(32.01)
813
810
825
843
(32.01) (31.89) (32.48) (33.19)
761
(29.96)
23
(50.7)
50
(2)
940
(37.01)
967
(38.07)
979
970
980
1001
(38.54) (38.19) (38.58) (39.41)
918
(36.14)
24
(52.9)
40
(1 1/2)
1045
(41.14)
1078
(42.44)
1078
1075 1090 1108
(42.44) (42.32) (42.91) (43.62)
1025
(40.35)
33
(72.8)
50
(2)
940
(37.01)
967
(38.07)
979
970
980
1001
(38.54) (38.19) (38.58) (39.41)
918
(36.14)
35
(77.2)
65
(2 1/2)
1100
(43.31)
1132
(44.57)
1148
1218 1228 1248
(45.20) (47.95) (48.35) (49.13)
1081
(42.56)
39
(86.0)
50
(2)
1220
(48.03)
1248
(49.13)
1259
1250 1260 1281
(49.57) (49.21) (49.61) (50.43)
1197
(47.13)
44
(97.0)
1100
(43.31)
1132
(44.57)
1148
1218 1228 1249
(45.20) (47.95) (48.35) (49.17)
1081
(42.56)
49
(108.0)
80
(3)
1220
(48.03)
1248
(49.13)
1260
1240 1260 1282
(49.61) (48.82) (49.61) (50.47)
1200
(47.24)
51
(112.4)
65
(2 1/2)
1330
(52.36)
1362
(53.62)
1378
1365 1375 1396
(54.25) (53.74) (54.13) (54.96)
1310
(51.57)
57
(125.7)
80
(3)
1220
(48.03)
1248
(49.13)
1260
1240 1260 1282
(49.61) (48.82) (49.61) (50.47)
1200
(47.24)
59
(130.1)
1480
(58.27)
1494
(58.82)
1530
1500 1520 1568
(60.24) (59.06) (59.84) (61.73)
1463
(57.60)
70
(154.3)
1640
(64.57)
1668
(65.67)
1680
1660 1680 1702
(66.14) (65.35) (66.14) (67.01)
1618
(63.70)
85
(187.4)
1530
1500 1520 1568
(60.24) (59.06) (59.84) (61.73)
1463
(57.60)
92
(202.8)
90
129
64
486
511
40
(3.54) (5.08) (2.52) (19.13) (20.12) (1 1/2)
110
148
80
515
540
(4.33) (5.83) (3.15) (20.28) (21.26)
130
164
97
541
566
65
(5.12) (6.46) (3.82) (21.30) (22.28) (2 1/2)
140
186 108
568
593
(5.51) (7.32) (4.25) (22.36) (23.35)
100
(4)
1450
(57.09)
80
(3)
100 (K)
150 (L)
170
215 131
612
637
(6.69) (8.46) (5.16) (24.09) (25.08)
250
285 190
725
750
(9.84) (11.22) (7.48) (28.54) (29.53)
100
(4)
1450
(57.09)
1480
(58.27)
1494
(58.82)
150
(6)
1736
(68.35)
1776
(69.92)
1816
(71.50)
1806 1826
(71.10) (71.89)
120
(264.6)
150
(6)
2000
(78.74)
2040
(80.31)
2080
(81.89)
2070 2090
(81.50) (82.28)
240
(529.1)
23
D184S068U02
Pos: 24.18 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Rohrverschraubung DN20-DN25 (kompakt) @ 5\mod_1167919432562_3101.doc @ 56044 @ 5 @ 1
134
(5,27)
80 (3,15)
2
L -5
Fig. 28:
G00336
Dimensions in mm (inch)
Meter Size
DN
20 (E)
89 (3.50)
25 (F)
89 (3.50)
127 (5.00)
127 (5.00)
66 (2.60)
66 (2.60)
G
MC23
MC26 /
MC27
470 (18.50)
494 (19.45)
470 (18.50)
494 (19.45)
Process connection
DN
DIN 11851
15 (1/2)
672 (26.46)
20 (3/4)
583 (22.95)
25 (1)
683 (26.89)
20 (3/4)
743 (29.25)
25 (1)
643 (25.31)
40 (1 1/2)
786 (30.94)
24
L-5
Weight
appr. kg (lb)
14 (30.9)
15 (33.1)
D184S068U02
Pos: 24.20 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Rohrverschraubung DN40-DN100 (kompakt) @ 5\mod_1167919468125_3101.doc @ 56065 @ 5 @ 1
134
(5,27)
80 (3,15)
g
L
L -5
G00335
Fig. 29:
Dimensions in mm (inch)
Meter
Size
DN
40 (G)
50 (H)
65 (I)
80 (J)
100 (K)
G
MC23
MC26/
MC27
90
129
64
486
511
(3.54) (5.08) (2.52) (19.13) (20.12)
110
148
80
515
540
(4.33) (5.83) (3.15) (20.28) (21.26)
130
164
97
541
566
(5.12) (6.46) (3.82) (21.30) (22.28)
140
186
108
568
593
(5.51) (7.32) (4.25) (22.36) (23.35)
170
215
131
612
637
(6.69) (8.46) (5.16) (24.09) (25.08)
Process
connection
DN
L-5
25 (1)
864 (34.02)
7 (0.28)
Rd 52 x 1/6
218 (8.58)
21 (46.3)
23 (50.7)
Weight
appr kg (lb)
40 (1 1/2)
61 (29.96)
7 (0.28)
Rd 65 x 1/6
164 (6.46)
50 (2)
918 (36.14)
7 (0.28)
Rd 78 x 1/6
241 (9.49)
24 (52.9)
40 (1 1/2)
1025 (40.35)
7 (0.28)
Rd 65 x 1/6
233 (9.17)
33 (72.8)
50 (2)
918 (36.14)
7 (0.28)
Rd 78 x 1/6
177 (6.97)
35 (77.2)
65 (2 1/2)
1081 (42.56)
8 (0.31)
Rd 95 x 1/6
254 (10.00)
39 (86.0)
50 (2)
1197 (47.13)
7 (0.28)
Rd 78 x 1/6
291 (11.46)
44 (97.0)
65 (2 1/2)
1081 (42.56)
8 (0.31)
Rd 95 x 1/6
227 (8.94)
48 (105.8)
51 (112.4)
80 (3)
1200 (47.24)
8 (0.31)
Rd 110 x 1/4
281 (11.06)
65 (2 1/2)
1310 (51.57)
8 (0.31)
Rd 95 x 1/6
319 (12.56)
57 (125.7)
80 (3)
1200 (47.24)
8 (0.31)
Rd 110 x 1/4
258 (10.16)
59 (130.1)
100 (4)
1463 (57.60)
10 (0.39)
Rd 130 x 1/4
381 (15.00)
70 (154.3)
80 (3)
1618 (63.70)
8 (0.31)
Rd 110 x 1/4
401 (15.79)
85 (187.4)
100 (4)
1463 (57.60)
10 (0.39)
Rd 130 x 1/4
314 (12.36)
92 (202.8)
25
D184S068U02
Pos: 24.22 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Tri-Clamp DN20-DN25 (kompakt) @ 5\mod_1167919667828_3101.doc @ 56149 @ 5 @ 1
134
(5,27)
1
B
L-5
G00338
Fig. 30:
Dimensions in mm (inch)
1 Flow direction
Meter Size
DN
20
25
(E)
(F)
89 (3.50)
89 (3.50)
127 (5.00)
127 (5.00)
66 (2.60)
66 (2.60)
G
MC23
MC26/
MC27
470 (18.50)
494 (19.45)
470 (18.50)
494 (19.45)
Process
connection
DN
L-5
15 (1/2)
656 (25.83)
140 (5.51)
20 (3/4)
561 (22.09)
92 (3.62)
25 (1)
661 (26.02)
142 (5.59)
20 (3/4)
721 (28.39)
152 (5.98)
25 (1)
621 (24.45)
102 (4.02)
40 (1 1/2)
773 (30.43)
180 (7.09)
26
Weight
appr kg (lb)
14 (30.9)
15 (33.1)
D184S068U02
Pos: 24.24 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Mazeichnungen/MC2 Tri-Clamp DN40-DN100 (kompakt) @ 5\mod_1167919728765_3101.doc @ 56191 @ 5 @ 1
134
(5,27)
80 (3,15)
1
L -5
G00337
Fig. 31:
Dimensions in mm (inch)
1 Flow direction
Meter Size
DN
40 (G)
50 (H)
65 (I)
80 (J)
100 (K)
90 (3.54)
110 (4.33)
130 (5.12)
140 (5.51)
170 (6.69)
129 (5.08)
148 (5.83)
164 (6.46)
186 (7.32)
215 (8.46)
64 (2.52)
80 (3.15)
97 (3.82)
G
MC23
MC26/
MC27
486 (19.13)
511 (20.12)
515 (20.28)
541 (21.30)
540 (21.26)
566 (22.28)
593 (23.35)
637 (25.08)
Process
connection
DN
L-5
Weight
appr kg
(lb)
25 (1)
842 (33.15)
242 (9.53)
17 (37.5)
40 (1 1/2)
748 (29.45)
195 (7.68)
17 (37.5)
18 (39.7)
50 (2)
913 (35.94)
278 (10.94)
40 (1 1/2)
1012 (39.84)
275 (10.83)
27 (59.5)
50 (2)
913 (35.94)
225 (8.86)
26 (57.3)
65 (2 1/2)
1073 (42.24)
305 (12.01)
27 (59.5)
50 (2)
1192 (46.93)
335 (13.19)
36 (79.4)
65 (2 1/2)
1073 (42.24)
275 (10.83)
37 (81.6)
80 (3)
1180 (46.46)
328 (12.91)
38 (83.8)
65 (2 1/2)
1302 (51.26)
378 (14.88)
45 (99.2)
80 (3)
1180 (46.46)
296 (11.65)
44 (97.0)
100 (4)
1448 (57.01)
430 (16.93)
46 (101.4)
80 (3)
1598 (62.91)
440 (17.32)
71 (156.5)
100 (4)
1448 (57.01)
365 (14.37)
69 (152.1)
27
D184S068U02
3.3
Ordering information
Separate sensor and integral mount version, DN 15 ... DN 150 (1/2 ... 6")
Additional
order no.
1-3
10
11
12
13
14
15
16
17
18
19
20
FCM2000 MC2 X
Design
Remote mount, ATEX, IECEx Zone 2, FM Div. 2
1
Integral mount, ATEX, IECEx Zone 2, FM Div. 2
3
Remote mount, ATEX, IECEx Zone 1, FM Div. 1
6
Integral mount, ATEX, IECEx Zone 1, FM Div. 1
7
Explosion protection/Cable gland/Ambient temperature
None/cable gland M20 x 1.5/standard
A
None/cable gland NPT 1/2 in./standard
T
None/cable gland M20 x 1.5/extended -40 ... 60 C
1) U
(-40 ... 140 F)
None/cable gland NPT 1/2 in. /extended -40 ... 60 C
1) X
(-40 ... 140 F)
ATEX, IECEx Zone 1/cable gland M20 x 1.5/standard
2) B
ATEX, IECEx Zone 2/cable gland M20 x 1.5/standard
3) M
ATEX, IECEx Zone 1/M20 x 1.5/extended -40 ... 60 C
4) E
(-40 ... 140 F)
ATEX, IECEx Zone 2/M20 x 1.5/extended -40 ... 60 C
2) N
(-40 ... 140 F)
FMus Class I, Div.1, Zone 1/cable gland
4) C
NPT 1/2 in./standard
FMus Class I, Div. 2, Zone 2/cable gland
3) O
NPT 1/2 in./standard
FMus Class I, Div. 1, Zone 1/NPT 1/2 in. /extended
4) Y
-40 ... 60 C (-40 ... 140 F)
FMus Class I, Div. 2, Zone 2/NPT 1/2 in. /extended
1) V
-40 ... 60 C (-40 ... 140 F)
cFM Class I, Div. 1, Zone 1/NPT 1/2 in./standard
2) D
cFM Class I, Div. 2, Zone 2/NPT 1/2 in./standard
3) P
cFM Class I, Div. 1, Zone 1/NPT 1/2 in. /extended
4)
Q
-40 ... 60 C (-40 ... 140 F)
cFM Class I, Div. 2, Zone 2/NPT 1/2 in. /extended
1) W
-40 ... 60 C (-40 ... 140 F)
Certificates
Standard
1
Material certificate 3.1 to EN 10204
2
Material certificate 3.1 to EN 10204 and pressure test to AD3
2000
Pressure test to AD-2000
4
NACE material certificate
5
NACE certificate + pressure test
6
Meter tube material
Stainless steel
1.4435 (AISI 316L SST) EHEDG version
5)
Hastelloy C-4 (2.4610)
6)
Measuring range/Max. measuring range/Size/DN nominal
0 ... 75 kg/min (0 ... 165 lbs/min)/''E''/DN 20 (3/4 in.)
0 ... 125 kg/min (0 ... 275 lbs/min)/''F''/DN 25 (1 in.)
0 ... 365 kg/min (0 ... 803 lbs/min)/''G''/DN 40 (1-1/2 in.)
0 ... 710 kg/min (0 ... 1562 lbs/min)/''F''/DN 50 (2 in.)
0 ... 1,450 kg/min (0 ... 3190 lbs/min)/''I''/DN 65 (2-1/2 in.)
0 ... 1,890 kg/min (0 ... 4158 lbs/min)/''J''/DN 80 (3 in.)
0 ... 3,200 kg/min (0 ... 7040 lbs/min)/''K''/DN 100 (4 in.)
0 ... 8,500 kg/min (0 ... 18,700 lbs/min)/''L''/DN 150 (6 in.)
Continued on next page
1)
2)
3)
4)
28
XX
1
3
4
E
F
G
H
I
J
K
L
D184S068U02
Additional
order no.
1-3
10
11
12
13
14
15
16
17
18
19
20
FCM2000 MC2 X
X X
1
2
2
4
5
6
8
1
1
5
0
5
0
0
5
0
H
F
7)
7)
7)
7)
7)
7)
7)
7)
7)
8)
8)
9)
9)
XX
D
F
G
H
P
Q
R
N
U
V
9)
1
2
1
10)
9)
9)
9)
11)
10)
9)
9)
9)
11)
A
B
C
M
D
E
F
N
G
H
I
O
J
K
L
P
12) G
E
X
X
A
C
Possible combinations of measuring range and process connection nominal diameter: See data sheet.
Not for sizes "K" und "L".
Not for size "L" (DN 150 [6 in.]).
Not for size "E" (DN 20 [3/4 in.])
Not for size "L" (DN 150 [6 in.]) or "E" (DN 20 [3/4 in.]).
Not for devices without Ex approval.
29
D184S068U02
Additional
order no.
10
11
12
13
14
15
16
17
18
19
20
FCM2000 MC2 X
Outputs
Current output I (active), current output II
(passive), pulse output (active) [Ex not possible]
Current output I (active), current output II
(passive), pulse output (passive)
Current output I (passive, "ia"), current output II
(passive, "ia"), pulse output (passive, "ia")
Not selected/fieldbus (only for remote mount
flowmeter sensors)
Communication
Not selected (only for remote mount flowmeter
None
HART protocol
PROFIBUS PA
FOUNDATION Fieldbus
PROFIBUS PA with M12 plug
Power supply
Not selected (only for remote mount flowmeter
100 ... 230 V AC
24 V AC/DC
Signal cable length
Zero in the case of integral mount version
5 m (16 ft.) with ME21
10 m (33 ft.) with ME21
20 m (66 ft.) with ME21
25 m (82 ft.) with ME21
50 m (164 ft.) with ME21
10 m (33 ft.) with ME26
Language of documentation
German
English
Western Europe/Scandinavia language package
(languages: DE, EN, DA, ES, FR, IT, NL, PT, FI, SV)
Eastern Europe language package (languages: DE,
EL, CS, ET, LV, LT, HU, PL, SK, SL, RO, BG)
Other
FCM2000-MC2 signal cable, price per meter
13)
14)
15)
16)
17)
30
XX
13) A
B
14) D
X
0
1
3
5
15) 7
G
K
16)
16)
16)
16)
16)
16)
17)
C0
C1
C2
C3
C4
C5
C6
M1
M5
MW
ME
MZ
D173D146U01
D184S068U02
Model FCM2000-MS2
4.1
Specifications
Supply power
Line voltage as per name plate UN 1 %
Warm-up phase
30 minutes
Installation according to this specification
No visible gas phase
No external mechanical or hydraulic disturbances, particularly
cavitation
G00315
Fig. 32:
Output calibration
Pulse output
Effect of the analog output on the measurement accuracy
Similar to pulse output 0.1 % of measured value
Nominal sizes
"S" (DN 1.5); "T" (DN 3); "U" (DN 6)
Sensor materials
Parts in contact with fluid
1.4435/316L
Housing 1.4404
"S"
"T"
"U"
DN 1.5 (1/16")
DN 3 (1/10")
DN 6 (1/4")
0 ...
0 ...
0 ...
65
250
1000
Protection class: IP 65
Measured value deviation for flowrate
0.4 % of flow rate 0.02 % of Qmax
0.25 % of flow rate 0.02 % of Qmax
0.15 % of flow rate 0.01 % of Qmax
(deviation from rate + zero error)
Reproducibility of flowrate
0.1 % of flow rate for nom. deviation 0.15 %
0.15 % of flow rate for nom. deviation 0.25 % and 0.4 %
Measuring range for density
0.5 ... 3.5 kg/dm3
Measured value deviation for density
Standard calibration 10 g/l
Temperature range 0 ... 100 C (32 ... 212 F)
Expanded density calibration available upon request
Measured value deviation for temperature
-50 ... 180 C (-58 ... 356 F) < 1 K (1.8 F)
Reference conditions
Calibration fluid
Water 25 C (77 F) (+ 5 K/- 5 K)
Pressure 0.5 ... 6 bar (7.3 ... 87.0 psi)
Fluid temperature
Standard:
-50 ... 180 C (-58 ... 356 F): DN 3 (1/10"), DN 6 (1/4")
-50 ... 125 C (-58 ... 257 F): DN 1.5 (1/16")
-50 ... 180 C (-58 ... 356 F): DN 1.5 (1/16") (optional)
For information about the design for operation in potentially explosive
areas, refer to the corresponding chapter.
Ambient temperature
-20 ... 50 C (-4 ... 122 F)
For information about the design for operation in potentially explosive
areas, refer to the corresponding chapter.
Process connections
G1/4" ISO 228-1
1/4" NPT ASME B1.201
Flange DIN/ASME for DN 6 (1/4")
Threaded pipe connection conforming to DIN 11851 for DN 6 (1/4")
Tri-Clamp conforming to DIN 32676 (ISO 2852) for DN 6 (1/4")
The max. permissible operating pressure is determined by the
respective process connection, the fluid temperature, the screws, and
the gasket material.
Pressure rating
Flange PN 40, PN 100, Cl 150, Cl 600
Thread G 1/4", 1/4" NPT, PN 100 ... PN 410 (for each option)
Installation
For more detailed instructions regarding installation, refer to the
operating instructions.
Ambient temperature
25 C (77 F) (+ 10 K/- 5 K)
31
D184S068U02
Viscosity [cSt]
00
10
0
20
0
50
00
10
10
0
50
0
10
20
0
10
50
10
0
20
50
100
20
10
P [bar]
P [bar]
5
1
1,0
1,0
0,1
0,01
0,1
10
70 [kg/h]
0,1
10
100
G00317
Fig. 33:
G00318
Fig. 35:
Viscosity [cSt]
00
10
0
50
0
20
0
10
20
50
P [bar]
10
10
1,0
0,1
3
Fig. 34:
32
10
100
300 [kg/h]
G00318
1000 [kg/h]
D184S068U02
4.2
Dimensions
4.2.1
Design MS21
D1
B1
D4
D3
H1
D2
77
(3,03)
88 (3,46)
D5
L1
L1
L1
17,5 (0,68)
Size T = DN 3 (1/10)
Fig. 36:
G00345
Size U = DN 6 (1/4)
Dimensions in mm (inch)
33
Nom.
size
DN
Connections
L1
Type
Threaded connector
ISO 228/1-G 1/4
3
(1/10) Threaded connection
ANSI/ASME B1.20.1-1/4"
NPT
Flange DIN 2635
Flange DIN 2637
Flange DIN 2635
Flange DIN 2637
Flange ANSI / ASME B 16.5
Flange ANSI / ASME B 16.5
Flange ANSI / ASME B 16.5
6
(1/4)
L2
L3
H1
B1
D1
D2
D5
100
1/4
100
400
280 75.0 60
1/4
(15.75) (11.02) (2.44) (2.36)
40
DN
560
390 62.0 40
12 17.0 104 90.0 60.0 14.0
10 (22.05) (11.02) (2.44) (1.57) (0.47) (0.67) (4.09) (3.54) (2.36) (0.55)
400
280 75.0 60
(15.75) (11.02) (2.44) (2.36)
21.3 104
(0.84) (4.09)
21.3 104
(0.84) (4.09)
100 (64)
DN
580
390 62.0 40
12 17.0 104
10 (22.83) (11.02) (2.44) (1.57) (0.47) (0.67) (4.09)
40
DN
638
390 62.0 40
12 17.0 104
15 (25.12) (11.02) (2.44) (1.57) (0.47) (0.67) (4.09)
100 (64)
DN
654
390 62.0 40
12 17.0 104
15 (25.75) (11.02) (2.44) (1.57) (0.47) (0.67) (4.09)
1/2
624
390 62.0 40
12 17.0 104 88.9 60.5 15.7
(24.57) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09) (3.50) (2.38) (0.62)
Class 600
1/2
646
390 62.0 40
12 17.0 104
(25.43) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
Class 150
3/4
670
390 62.0 40
12 17.0 104
(26.38) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
Class 600
3/4
693
390 62.0 40
12 17.0 104
(27.28) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
Screwed connection
DIN 11851
40
DN
570
390 62.0 40
12 17.0 104
15 (22.44) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
Tri-clamp,
DIN 32676 (ISO 2852)
16
25
570
390 62.0 40
12 17.0 104
mm (22.44) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
16
25
573
390 62.0 40
12 17.0 104
mm (22.56) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
Threaded connector
ISO 228/1-G 1/4
100
1/4
100
Threaded connection
ISO 228/1-G 1/4
Threaded connection ANSI/
ASME B 1.20.1-1/4" NPT
Weight
appr. kg
(lb)
4
(8.8)
Class 150
DN
532
390 62.0 40
12 17.0 104
10 (20.94) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
D4
Size
40
Measurements in mm (inch)
D3
Pressure PN
Screwed connection
DIN 11851
34
D184S068U02
8
(17.6)
562
390 62.0 40
12 17.0 104
(22.13) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
1/4
562
390 62.0 40
12 17.0 104
(22.13) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
EN1.4435 PN 265
EN2.4602 PN 410
1/4
562
390 62.0 40
12 17.0 104
(22.13) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
EN1.4435 PN 265
EN2.4602 PN 410
1/4
562
390 62.0 40
12 17.0 104
(22.13) (15.35) (2.44) (1.57) (0.47) (0.67) (4.09)
D184S068U02
1/4 NPT
18,8 (0,74)
24 (0,94)
1/4 NPT
ANSI/ASME
B 1.20.1-1983
Fig. 37:
120 (4,72)
NV21
G 1/4
18,8 (0,74)
24 (0,94)
G 1/4
ISO 228/1
9
(0,35)
20 (0,78)
NV21
15 (0,59)
12 (0,47)
15 (0,59)
12 (0,47)
86 (3,38)
152 (5,98)
57 (2,24)
4
(0,16)
10 (0,39)
64,5 (2,53)
78 (3,07)
46 (0,47)
129 (5,07)
20 (1,81)
25 (0,98)
100 (3,93)
G00346
Dimensions in mm (inch)
35
D184S068U02
4.3
Ordering information
Addition
al order
no.
1-3
36
10
11
12
13
14
15
16
A
B
C
O
D
P
1
2
3
3
4
S
T
U
1
3) 2
A
B
C
P
Q
R
I
U
4) V
W
M
N
J
5)
6)
7)
8)
9)
9)
1
2
3
4
5
6
0
10) 1
10) 2
6)
7)
8)
9)
10)
D184S068U02
Addition
al order
no.
10
11
12
13
14
15
FCM2000 MS2 X
1-3
16
XX
A
B
C
G
H
I
11) G
E
X
4)
4)
4)
4)
4)
1
2
3
4
5
M1
M5
MW
ME
MZ
37
D184S068U02
Transmitter
5.1
Specifications
Display
The graphic display has 2 lines and features an LED backlight. Both
lines are freely configurable to display the mass flowrate, volume
flowrate, density, or temperature. Flow count, 7-digit with overflow
counter and physical unit for mass or volume.
When the four mounting screws are loosened, the display can be
installed in 4 positions. This ensures optimal readability.
G00320
Fig. 38:
Measuring range
Freely configurable between 0.01 Qmax and 1 Qmax
Ingress protection
IP 65/IP 67, NEMA 4X
Electrical connections
Cable gland M20 x 1.5 or 1/2" NPT
Max. signal cable length for remote mount design 50 m (longer
lengths available upon request)
Supply power
Supply voltage
100 ... 230 V AC (tolerance -15 % and +10 %), 47 ... 63 Hz
20.4 ... 26.4 V AC, 47 ... 63 Hz
20.4 ... 31.2 V DC
Ripple: 5 %
Power consumption
S 25 VA
Response time
As jump function 0 ... 99 % (corr. to 5 ) 1 s
Ambient temperature
-20 ... 60 C (-4 ... 140 F), optional -40 ... 60 C (-40 ... 140 F)
At operation below -20 C (-4 F), the display can no longer be read
and the electronic unit should be operated with as few vibrations as
possible. Complete operational reliability is achieved at temperatures
above -20 C (-4 F).
Design
Field-mount housing and compact transmitter unit as alloy
casting, varnished
Mid-section: RAL 7012, dark gray
Cover: RAL 9002, light gray
Paint coat: 80 ... 120 m thick
Forward/reverse flow metering
Signals are shown on the display by direction arrows and by the
optocoupler for ext. signaling.
38
G00322
Fig. 39:
D184S068U02
5.2
5.3
This software calculates on the basis of predefined density-temperature-concentration matrices the present concentration of a 2 phase
liquid. In this software the following matrices are predefined:
Concentration of sodium hydroxyde in water
Concentration of alcohol in water
Concentration of sugar (BRIX)
Concentration of Corn Starch
Concentration of Wheat starch
The user can additionally enter up to 2 variable matrices with up to
100 values for concentration computation.
Calculation of accuracy
The accuracy of the concentration computation depends foremost on
the quality of the matrix data. As density and temperature measurement is the input for the calculation both accuracies define the
accuracy of the concentration measurement.
Example:
Density 0 % alcohol in water (20 C [68 F])
998.23 g/l
Density 100 % alcohol in water (20 C [68 F])
789.30 g/l
100 %
=
208.93 g/l
0.48 %
=
1 g/l
2.40 %
=
5 g/l
The chosen accuracy class of the density measurement effects the
accuracy of the concentration measurement directly.
Inputs/outputs
Load
300
200
100
0
5
10
15
20
25
30
35
G00323
Source voltage
Fig. 40:
1
Permissible range
Y
Y
I max
I max
-100 %
0%
(0 ... 20 mA)
X
100 %
I min
-100 %
0%
(4 ... 20 mA)
X
100 %
G00324
Fig. 41
Important
Failure information acc. to NAMUR recommendation NE43.
39
fmax
Pulse width
Passive
51, 52
16 V UCEH 30 V DC
0 V UCEL 2 V
0 mA ICEH 0.2 mA
2 mA ICEL 220 mA
Active
51, 52
16 V U 30 V DC
Load 150 Ohm
fmax = 5 kHz
Contact output
The following functions can be assigned in the software:
System monitoring: normally open or normally closed contact
Forward/reverse flow indication: closed for forward flow
Max/min alarm: normally open or normally closed contact
Terminals: 41, 42
"Closed"
0 V UCEL 2 V
2 mA ICEL 220 mA
"Open"
16 V UCEH 30 V
0 mA ICEH 0.2 mA
Contact input
The following functions can be assigned in the software:
Ext. output switch-off. When the meter tube empties, all output signals
can be turned off.
Ext. totalizer reset. The internal totalizers can be reset from an
external contact.
Terminals: 81/82
"On"
16 V UKL 30 V
"Off"
0 V UKL 2 V
Internal resistance: Ri = 2 k
All inputs/outputs are electrically isolated from one another.
Pos: 50 /Standard-Layout/======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ @ 1
40
D184S068U02
D184S068U02
5.4
Digital Communication
5.4.1
HART protocol
HART
Bell 202
Modem Box
5.4.2
PROFIBUS PA protocol
G00325
Fig. 42:
HART protocol
Configuration
Transmission
Max. signal
amplitude
Load of
current output
H2-Bus
Cable
Cable
Max. cable length
Baud rate
Display
AWG 24 twisted
1,500 m (4,921 ft.)
1,200 baud
Log. 1: 1,200 Hz
Log. 0: 2,200 Hz
For additional information, see the separate interface description.
System integration
Communication (configuration, parameterization) can be performed
with the DTM (Device Type Manager) available for the unit (software
version B.10 and higher) and the corresponding framework
applications as per FDT 0.98 or 1.2 (DSV401 R2). If you require
integration into different tools/systems (e.g., AMS or Siemens S7), this
is available upon request. DSV401 communication tool for HART, free
90-day test version also available upon request. DTMs are included in
DSV401.
100
1 F
PA+ PA-
PA+ PA-
PA+ PAG00111
PROFIBUS PA
PROFIBUS DP
Bus topology
Voltage/Current consumption
41
5.4.3
D184S068U02
FF interface
Interoperability test
campaign no.
Manufacturer ID
Device ID
Configuration
Transmission signal
Ethernet
HSE-Bus
FF+ FF-
System integration
FF+ FFG00112
Bus topology
Bus address
100
1 F
FF+ FF-
FOUNDATION Fieldbus H1
Voltage/Current consumption
42
D184S068U02
5.5
Electrical connections
5.5.1
Passive
I
0/4 ... 20 mA
E
4 ... 20 mA
+31
-32
12 ... 30 V+
+33
-34
G00326
G00327
I = internal
E = external
Fig. 45:
Switch output
I
E
41
+U
R B*
0V
24 V DC
42
* R B UCE
ICE
Fig. 46:
G00244
Switch output for system monitoring, Max.-Min. alarm for empty tube or forward/reverse signal
Switch input
I
81
+ 16 ... 30 V DC
82
0V
Ri
G00245
Fig. 47:
Contact input for external totalizer reset and external zero return
Pulse Output
active
passive, optocoupler
E
R B*
24 V+
51
52
51
+ 16 ... 30 V DC
52
G00246
* RB
Fig. 48:
UCE
ICE
G00247
43
D184S068U02
98
FF-
PA-
98
PA+
97
R
C
97
1
Fig. 49:
R
C
FF+
G00248
G00328
Fig. 50:
44
Pin assignment
(Front view showing pin insert and pins)
PIN 1 = PA+
PIN 2 = nc
PIN 3 = PAPIN 4 = shield
D184S068U02
5.5.2
ME21
92 91 85 86 87 88 89 90 93 94 95 96
1 2
3 4
5 6
7 8 9 10 11 12
13
13
13
3 4 5 6
SE
7 8 11 12 9 10 2 1
85 86 87 88 89 90 95 96 93 94 91 92
MC21
13
MC26
12 11 10 9
8 7
6 5
4 3
2 1
13
SE
96 95 94 93 90 89 88 87 86 85 91 92
13
G00350
Fig. 51
91 / 92
93 / 94 / 95 / 96
85 / 86
87 / 88
Driver
Temperature
Sensor 1
Sensor 2
1
2
3
4
5
6
7
8
9
10
11
12
13
Red/blue
Gray/pink
White
Brown
Green
Yellow
Gray
Pink
Black
Violet
Blue
Red
"PA" equipotential bonding. The precise position of the ground
terminals may vary according to the device type. However, the
position is appropriately labeled in each case. When connecting
transmitter ME21 to flowmeter sensor MC26, transmitter ME21
also has to be connected to "PA".
45
D184S068U02
ME26
13
13
96 95 94 93 90 89 88 87 86 85
12 11 10 9 8
7 6 5 4
91 92
13
MC26
G00388-01
Fig. 52
91 / 92
93 / 94 / 95 / 96
85 / 86
87 / 88
Driver
Temperature
Sensor 1
Sensor 2
1
2
3
4
5
6
7
8
9
10
11
12
13
Pink
Gray
White
Brown
Green
Yellow
Black
Violet
Blue
Red
"PA" equipotential bonding
Important
For reasons of EMC, the wires must be routed in pairs.
46
D184S068U02
11
11 1 2
3 4
5 6
7 8 9 10
11
ME27
Me28
11
11
MS21
Ms26
Fig. 53
91 / 92
93 / 94 / 95 / 96
85 / 86
87 / 88
Driver
Temperature
Sensor 1
Sensor 2
1
2
3
4
5
6
7
8
9
10
11
G00389
Red
Brown
Green
Blue
Gray
Violet
White
Black
Orange
Yellow
"PA" equipotential bonding. When connecting transmitter to
flowmeter sensor MS26, transmitter also has to be connected to
"PA".
47
D184S068U02
5.5.3
Pos: 58.11 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Elektrischer Anschluss/Ein- / Ausgangssignale ME2 / MC2 @ 50\mod_1312365832409_3101.doc @ 417486 @ 5 @ 1
31 32 33 34 51 52 41 42 81 82
7
1
6
7
ME26
MC26
ME21
2- 1+
51 52 81 82 41 42 31 32 33 34
N L
7
7
4
MC21
G00814
Fig. 54
1 Supply power
Line voltage: UAC 100 ... 230 V AC, frequency 50/60 Hz,
terminals L, N,
Low voltage: UAC 24 V, frequency 50/60 Hz, terminals 1+, 2Low voltage: UDC 24 V
2 Current output 1: can be selected via software
2a: function: active
Terminals: 31, 32; 0/4 20 mA (0 RB 560 ,
MC27/ME26: 0 <= RB <= 300 )
2b: alternate function: passive (option D, not FM)
Terminals: 31, 32; 4 ... 20 mA (0 RB 600 )
Source voltage 12 Uq 30 V
3 Current output 2: can be selected via software
Function: passive
Terminals: 33, 34; 4 20 mA (0 RB 600 )
Source voltage 12 Uq 30 V
Important
You can find the applicable Ex-relevant connection data in the section titled "Ex relevant specifications".
Pos: 58.12 /Standard-Layout/======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ @ 1
48
D184S068U02
Pos: 58.13 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Elektrischer Anschluss/Ein- / Ausgangssignale ME2 / MS2 @ 49\mod_1311943057355_3101.doc @ 416774 @ 5 @ 1
2- 1+
N L
51 52 81 82 41 42 31 32 33 34
7
4
MS21
ME27
ME28
1+ 2L
31 32 33 34 51 52 41 42 81 82
7
1
MS26
G00813
Fig. 55
1 Supply power
Line voltage: UAC 100 ... 230 V AC, frequency 50/60 Hz,
terminals L, N,
Low voltage: UAC 24 V, frequency 50/60 Hz, terminals 1+, 2Low voltage: UDC 24 V
2 Current output 1: can be selected via software
2a: function: active
Terminals: 31, 32; 0/4 20 mA (0 RB 560 , ME27/28: 0
<= RB <= 300 )
2b: alternate function: passive (option D)
Terminals: 31, 32; 4 ... 20 mA (0 RB 600 )
Source voltage 12 Uq 30 V
3 Current output 2: can be selected via software
Function: passive
Terminals: 33, 34; 4 20 mA (0 RB 600 )
Source voltage 12 Uq 30 V
49
D184S068U02
Pos: 58.15 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Elektrischer Anschluss/Profibus PA/FOUNDATION Fieldbus, Hilfsenergie des ME2 / MC2 @ 25\mod_1232376030562_3101.doc @ 240031 @ 5 @ 1
97 98
3
1
2
3
ME26
MC26
ME21
2- 1+
97 98
N L
3
3
MC21
G00816
Fig. 56
1 Supply power
Line voltage: UAC 100 ... 230 V AC, frequency 50/60 Hz,
terminals L, N,
Low voltage: UAC 24, frequency 50/60 Hz, terminals 1+, 2Low voltage: UDC 24 V
2a PROFIBUS PA design conforming to IEC 61158-2 (Profile 3.0)
U = 9 32 V
I = 14 mA (normal operation)
I = 26 mA (in the event of an error / FDE)
Terminals: 97 / 98
Connection example via M12 plug, see Fig. 50
50
D184S068U02
Pos: 58.17 /Standard-Layout/Technische Daten / Datenblatt/Durchfluss/FCM2000/Elektrischer Anschluss/Profibus PA/FOUNDATION Fieldbus, Hilfsenergie des ME2 / MS2 @ 25\mod_1232445241375_3101.doc @ 240114 @ 5 @ 1
1+ 2L
97 98
3
3
MS21
ME27
ME28
2- 1+
97 98
N L
3
3
3
MS26
G00821
Fig. 57:
1 Supply power
Line voltage: UAC 100 ... 230 V AC, frequency 50/60 Hz,
terminals L, N,
Low voltage: UAC 24, frequency 50/60 Hz, terminals 1+, 2Low voltage: UDC 24 V
2a PROFIBUS PA design conforming to IEC 61158-2 (Profile 3.0)
U = 9 32 V
I = 14 mA (normal operation)
I = 26 mA (in the event of an error / FDE)
Terminals: 97 / 98
Connection example via M12 plug, see Fig. 50
51
D184S068U02
5.6
Dimensions
5.6.1
198 (7.79)
.27
167 (6.57)
10 (0.39)
0
7(
14,5 (0.57)
249 (9.80)
265 (10.43)
38 (1.49)
66 (2.59)
min. 62 (2.44)
83,5 (3.28)
M20 x 1,5
min.175
(6.88)
132 (5.19)
4
G00073
Fig. 58:
1
2
3
4
52
Dimensions in mm (inch)
D184S068U02
280,50 (11.04)
1:5
160 (6.30)
241,30 (9.50)
G00843
Fig. 59:
53
D184S068U02
5.7
Ordering information
External transmitter, DSP technology, for remote mount sensor MC2, MS2
Addition
al order
no.
54
10
11
XX
1
2
3
4
5
6
7
8
A
T
O
1) B
C
D
P
2) 3
3) 8
A
C
2) A
B
4) D
X
0
1
3
5
7
G
K
5) G
E
M1
M5
MW
ME
MZ
D184S068U02
Pos: 67.1 /Standard-Layout/berschriften/1/D - F/Ex-relevante technische Daten gem ATEX / IECEx @ 31\mod_1246959440531_3101.doc @ 285023 @ 1 @ 1
6.1
Pos: 67.3 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Sicherheitstechnische Daten ATEX / IECEx - bersicht @ 31\mod_1246866904890_3101.doc @ 284283 @ @ 1
ATEX/IECEx Zone 2
- Current output 1: active
- Current output 2: passive
- Pulse output: active/passive,
switchable
- Contact input and output: passive
ATEX/IECEx Zone 1
- Current output 1: active
- Current output 2: passive
- Pulse output: active/passive,
switchable
- Contact input and output: passive
- Current output 1: passive
- Current output 2: passive
- Pulse output: active/passive,
switchable
- Contact input and output: passive
- Fieldbus communication
(PROFIBUS PA/FOUNDATION Fieldbus)
II
Output option D
in the order number
III
Output option X
and communication
option 3, 5, or 7
in the order number
Fieldbus communication
(PROFIBUS PA/FOUNDATION Fieldbus)
Pos: 67.4 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Version I: Stromausgnge aktiv / passiv @ 49\mod_1311943261715_3101.doc @ 416808 @ @ 1
30
30
(V)
30
(mA)
30
30
30
30
30
30
65
30
65
30
65
30
65
30
10
30
10
All inputs and outputs are electrically isolated from each other and from the supply power.
55
D184S068U02
General
operating values
Ub
Ib
Uo
Io
(V)
20
Ui
(mA)
100
Ii
(mW)
500
Pi
(nF)
217
Ci
(nF)
0
Ci pa
(mH)
3.8
Li
35
(V)
60
(mA)
100
(mW)
500
(nF)
2.4
(nF)
2.4
(mH)
0.17
60
35
30
100
760
2.4
2.4
0.17
65
60
35
15
30
115
2.4
2.4
0.17
30
65
60
35
15
30
115
2.4
2.4
0.17
30
10
60
35
30
60
500
2.4
2.4
0.17
(V)
(mA)
(V)
(mA)
30
30
30
30
60
30
30
30
30
30
65
30
30
65
30
10
Lo
All inputs and outputs are electrically isolated from each other and from the supply power. Only current outputs 1 and 2 are not electrically
isolated from one another.
Pos: 67.5 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Version II: Stromausgnge passiv / passiv @ 49\mod_1311943773162_3101.doc @ 416833 @ @ 1
General
operating values
Ub
Ib
Ui
Ii
(V)
(mA)
(mW)
(nF)
(nF)
(mH)
35
60
300
2000
0,47
0,47
0,17
60
35
60
300
2000
0,47
0,47
0,17
65
60
35
60
300
2000
0,47
0,47
0,17
30
65
60
35
60
300
2000
0,47
0,47
0,17
30
10
60
35
60
300
2000
0,47
0,47
0,17
(V)
(mA)
(V)
(mA)
30
30
30
30
60
30
30
30
30
30
65
30
30
65
30
10
All inputs and outputs are electrically isolated from each other and from the supply power.
Pos: 67.6 /Standard-Layout/======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ @ 1
56
Li
D184S068U02
60
500
32
10
Ui
Ii
Ci pa
Li
(V)
(mA)
(mW)
(nF)
(nF)
(mH)
60
500
7,000
0.17
60
35
General
operating
values
Ub
Ib
Protection type
"ia" FISCO (Zone 1)
Ui
(V)
(mA)
(V)
(mA) (mW)
32
10
60
380
Ii
Pi
5320
Protection type
"ia" (Zone 1)
Ci
Ci pa
Li
Ui
Ii
Pi
(nF)
(nF)
(mH)
(V)
(mA) (mW)
0.17
60
380
5320
Ci
Ci pa
Li
(nF)
(nF)
(mH)
0.17
Special conditions
The output circuits are designed in such a way that they can be
connected to both intrinsically-safe and non-intrinsically-safe circuits.
It is not permitted to combine intrinsically safe and non-intrinsically
safe circuits. On intrinsically safe circuits, equipotential bonding must
be in place along the entire length of the cable used for the current
outputs. The rated voltage of the non-intrinsically safe circuits
is UM = 60 V.
Important
If the protective conductor (PE) is connected in the flowmeter's terminal box, you must ensure that no dangerous
potential difference can arise between the protective conductor (PE) and the equipotential bonding (PA) in the
potentially explosive area.
Pos: 67.12 /Standard-Layout/======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ @ 1
57
D184S068U02
6.1.1
ATEX/IECEx Ex approval
6.1.1.1
Model
Zone 1
Ambient temperature
<=40 C (104 F)
<=50 C (122 F)
<=60 C (140 F)
T1
200 C (392 F)
200 C (392 F)
200 C (392 F)
T2
200 C (392 F)
200 C (392 F)
200 C (392 F)
T3
185 C (365 F)
180 C (356 F)
180 C (356 F)
T4
125 C (257 F)
120 C (248 F)
120 C (248 F)
T5
85 C (185 F)
85 C (185 F)
75 C (167 F)
T6
65 C (149 F)
65 C (149 F)
60 C (140 F)
Temperature class
Model
Ambient temperature
Temperature class
T1
T2
T3
T4
T5
T6
<=40 C (104 F)
Zone 2
<=50 C (122 F)
<=60 C (140 F)
200 C (392 F)
200 C (392 F)
180 C (356 F)
115 C (239 F)
80 C(176 F)
60 C (140 F)
200 C (392 F)
200 C (392 F)
180 C (356 F)
115 C (239 F)
80 C(176 F)
60 C (140 F)
180 C (356 F)
180 C (356 F)
180 C (356 F)
115 C (239 F)
75 C (167 F)
60 C (140 F)
Designation
ATEX
II 3 G Ex nA II T6 ... T2
IECEx
Ex nA II T6 ... T2
Designation
ATEX
II 3 G Ex nA nR II T6 ... T2
58
Ex nA nR II T6 ... T2
No M12 plug
FNICO fieldbus
D184S068U02
Design MC26
Zone 1
Designation
ATEX
DN 40 (1 1/2")
DN 50 (2")
Design MC27
Zone 1
Designation
ATEX
Version II/III
DN 40 (1 1/2")
2 passive analog outputs, outputs "ia"/"e", depending on user
wiring, or FISCO fieldbus
Version I
DN 40 (1 1/2")
Active/passive analog outputs, outputs "ib"/"e", depending on
user wiring
Version II/III
DN 50 (2")
2 passive analog outputs, outputs "ia"/"e", depending on user
wiring, or FISCO fieldbus
Version I
DN 50 (2")
Active/passive analog outputs, outputs "ib"/"e", depending on
user wiring
Version I/II/III
Outputs "e"
Version II/III
Version I
Version III
FISCO fieldbus
Version II/III
Version I
Version I/II/III
Outputs "e"
Version II/III
Version I
Version III
FISCO fieldbus
IECEx
59
6.1.1.2
D184S068U02
Model
Zone 1
-20 ... 50 C (-4 ... 122 F)
Ambient temperature
Temperature class
T1
T2
T3
T4
T5
T6
180 C (356 F)
180 C (356 F)
180 C (356 F)
125 C (257 F)
80 C (176 F)
Designation
ATEX
II 2 G Ex ib IIC T5 ... T3
6.1.1.3
IECEx
60
II 3 G Ex nR II T6
II 3 G Ex nR [nL] IIC T6
No M12 plug
FNICO fieldbus
Ex nR II T6
Ex nR [nL] IIC T6
No M12 plug
FNICO fieldbus
D184S068U02
Designation
ATEX
Version II/III
Version I
II 2 G Ex d e [ib] IIC T6
Version I/II/III
Outputs "e"
Version II/III
Version I
Version III
FISCO fieldbus
Version II/III
Version I
Ex d e [ib] IIC T6
Version I/II/III
Outputs "e"
Version II/III
Version I
Version III
FISCO fieldbus
IECEx
Designation
ATEX
Version II/III
Version I
II 2 G Ex d e [ib] IIC T6
Version II/III
Version I
FISCO fieldbus
61
D184S068U02
Pos: 67.15 /Standard-Layout/berschriften/1/D - F/Ex-relevante technische Daten gem cFMus @ 31\mod_1246959511468_3101.doc @ 285049 @ 1 @ 1
Pos: 67.16 /Standard-Layout/berschriften/1.1/1-spaltig/D - F/Daten fr den Betrieb des MC2x @ 31\mod_1246523675781_3101.doc @ 283556 @ 2 @ 1
7.1
7.1.1
General information
Pos: 67.18 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Allgemeine Daten MC2x (FM/CSA) @ 49\mod_1311939692721_3101.doc @ 416561 @ @ 1
Type of
protection
Explosion-proof
Dust-ignition-proof
Intrinsically safe
Non-incendive
Ex designation
XP-IS/I, II, III/1/BCD/T* TA=*; type NEMA 4x
DIP/II, III/1 EFG/T* TA=*; type NEMA 4x
IS/I, II, III/I/BCDEFG/T* TA = *; type NEMA 4x
NI/I, II, III/2/ABCDFG/T* TA = *; type NEMA 4x
Specific FM coding applies, depending on the design of the flowmeter sensor (integral mount or remote mount design) For detailed information,
please refer to the section titled 1.2 "FM device overview (PID: 3036514)".
Pos: 67.19 /Standard-Layout/berschriften/1.1.1/1-spaltig/Temperaturdaten @ 31\mod_1245227663632_3101.doc @ 280223 @ 3 @ 1
7.1.2
Temperature data
Pos: 67.20 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Temp.Daten Messwertaufnehmer / -umformer FM/CSA Div. 1 @ 30\mod_1245222778028_3101.doc @ 280001 @ 5 @ 1
200 C (392 F)
200 C (392 F)
185 C (365 F)
125 C (257 F)
85 C (185 F)
65 C (149 F)
Ambient temperature
50 C (122 F)
Maximum permissible fluid temperature
200 C (392 F)
200 C (392 F)
180 C (356 F)
120 C (248 F)
85 C (185 F)
65 C (149 F)
60 C (140 F)
200 C (392 F)
200 C (392 F)
180 C (356 F)
120 C (248 F)
75 C (167 F)
60 C (140 F)
Pos: 67.21 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Temp.Daten Messwertaufnehmer / -umformer FM/CSA Div. 2 @ 31\mod_1245224842468_3101.doc @ 280051 @ 5 @ 1
62
200 C (392 F)
200 C (392 F)
180 C (356 F)
115 C (239 F)
80 C (176 F)
60 C (140 F)
Ambient temperature
50 C (122 F)
Maximum permissible fluid temperature
200 C (392 F)
200 C (392 F)
180 C (356 F)
115 C (239 F)
80 C (176 F)
60 C (140 F)
60 C (140 F)
180 C (356 F)
180 C (356 F)
180 C (356 F)
115 C (239 F)
75 C (167 F)
60 C (140 F)
D184S068U02
Pos: 67.23 /Standard-Layout/berschriften/1.1/1-spaltig/D - F/Daten fr den Betrieb des MS2x @ 31\mod_1246523723968_3101.doc @ 283580 @ 2 @ 1
7.2
Pos: 67.24 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/1.1.1 Ex-Daten MS2 komplett @ 49\mod_1311939942602_3101.doc @ 416586 @ 335 @ 1
7.2.1
General information
Type of
protection
Explosion-proof
Dust-ignition-proof
Intrinsically safe
Non-incendive
Ex designation
XP-IS/I, II, III/1/BCD/T* TA=*; type NEMA 4x
DIP/II, III/1 EFG/T* TA=*; type NEMA 4x
IS/I, II, III/I/BCDEFG/T* TA = *; type NEMA 4x
NI/I, II, III/2/ABCDFG/T* TA = *; type NEMA 4x
Protection
class
Specific FM coding applies, depending on the design of the flowmeter sensor (integral mount or remote mount design) For detailed information,
please refer to the section titled 1.2 "FM device overview (PID: 3036514)".
7.2.2
Temperature data
Temperature class
T1
T2
T3
T4
T5
T6
Ambient temperature
-20 ... 50 C (-4 ... 122 F)
Maximum permissible fluid temperature
180 C (356 F)
180 C (356 F)
180 C (356 F)
125 C (257 F)
80 C (176 F)
-
63
D184S068U02
7.3
Electrical data
Pos: 67.27 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Sicherheitstechnische Daten FM/CSA - bersicht @ 31\mod_1246524596021_3101.doc @ 283605 @ @ 1
Class I Div. 2
- Current output 1: active
- Current output 2: passive
- Pulse output: active/passive,
switchable
- Contact input and output: passive
II
Output option D
in the order number
III
Output option X
and communication
option 3, 5, or 7
in the order number
Fieldbus communication
(PROFIBUS PA/FOUNDATION Fieldbus)
Class I Div. 1
- Current output 1: active
- Current output 2: passive
- Pulse output: active/passive,
switchable
- Contact input and output: passive
- Current output 1: passive
- Current output 2: passive
- Pulse output: active/passive,
switchable
- Contact input and output: passive
- Fieldbus communication
(PROFIBUS PA/FOUNDATION Fieldbus)
7.3.1
Pos: 67.29 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Version I: Stromausgnge aktiv / passiv (Div. 1) @ 49\mod_1311940456030_3101.doc @ 416611 @ 5 @ 1
Current output 2
passive
Terminal 33/34
Digital output
Terminal 41/42
Digital input
Terminal 81/82
Pulse output
Terminal 51/52
Vmaxo
[V]
Imaxo
[mA]
Po
[mW]
Co
[nF]
Co PA
[nF]
L0
[mH]
20
100
500
217
3,8
VMax
[V]
IMax
[mA]
Pi
[mW]
Ci
[nF]
Ci PA
[nF]
Li
[mH]
60
100
500
2,4
2,4
0,17
30
100
760
2,4
2,4
0,17
15
30
115
2,4
2,4
0,17
30
60
500
2,4
2,4
0,17
15
30
115
2,4
2,4
0,17
All inputs and outputs are electrically isolated from each other and from the supply power. Only current outputs 1 and 2 are not electrically
isolated from one another.
Pos: 67.30 /Standard-Layout/======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ @ 1
64
D184S068U02
Pos: 67.31 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Version II: Stromausgnge passiv / passiv (Div. 1) @ 49\mod_1311941128299_3101.doc @ 416636 @ 5 @ 1
Vmax
[V]
Imax
[mA]
Pi
[mW]
Ci
[nF]
Ci PA
[nF]
Li
[mH]
60
300
2000
0,47
0,47
0,17
60
300
2000
0,47
0,47
0,17
60
300
2000
0,47
0,47
0,17
60
300
2000
0,47
0,47
0,17
60
300
2000
0,47
0,47
0,17
All inputs and outputs are electrically isolated from each other and from the supply power.
Pos: 67.32 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Version III: Feldbuskommunikation (Div. 1) @ 49\mod_1311941376636_3101.doc @ 416661 @ 5 @ 1
Type of protection IS
Inputs and
outputs
VMax
[V]
IMax
[mA]
Pi
[mW]
Ci
[nF]
Ci PA
[nF]
Li
[mH]
VMax
[V]
IMax
[mA]
Pi
[mW]
Ci
[nF]
Ci PA
[nF]
Li
[mH]
Passive fieldbus
Terminal 97/98
60
380
5320
0,17
60
380
5320
0,17
Pos: 67.35 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FEP300/FEH300/Wichtig - Hinweise auf PE und PA @ 28\mod_1241091943984_3101.doc @ 264981 @ @ 1
Important (Notice)
The housing for the transmitter and flowmeter sensor must be connected to the potential
equalization PA. The operator must ensure that when connecting the protective conductor
(PE) no potential differences can occur between protective conductor and potential
equalization (PA).
Pos: 67.36 /Standard-Layout/======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ @ 1
65
D184S068U02
7.3.2
Pos: 67.38 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Version I: Stromausgnge aktiv / passiv (Div. 2) @ 49\mod_1311941573461_3101.doc @ 416686 @ 5 @ 1
Vmaxo
[V]
Imaxo
[mA]
30
30
30
30
30
65
30
10
30
65
Current output 1
Terminal 31/32
Current output 2
passive
Terminal 33/34
Digital output
Terminal 41/42
Digital input
Terminal 81/82
Pulse output
Terminal 51/52
All inputs and outputs are electrically isolated from each other and from the supply power.
Pos: 67.39 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FCM2000/Version III: Feldbuskommunikation (Div. 2) @ 49\mod_1311941731084_3101.doc @ 416711 @ 5 @ 1
Type of protection NI
Inputs and
outputs
Vmax
[V]
Imax
[mA]
Pi
[mW]
Ci
[nF]
Ci PA
[nF]
Li
[mH]
Vmax
[V]
Imax
[mA]
Pi
[mW]
Ci
[nF]
Ci PA
[nF]
Li
[mH]
Passive fieldbus
Terminal 97/98
60
500
7000
0,17
60
500
7000
0,17
Pos: 67.41 /Standard-Layout/Ex-relevante Technische Daten/Durchfluss/FEP300/FEH300/Wichtig - Hinweise auf PE und PA @ 28\mod_1241091943984_3101.doc @ 264981 @ @ 1
Important (Notice)
The housing for the transmitter and flowmeter sensor must be connected to the potential
equalization PA. The operator must ensure that when connecting the protective conductor
(PE) no potential differences can occur between protective conductor and potential
equalization (PA).
Pos: 68 /Standard-Layout/======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ @ 1
66
D184S068U02
Questionnaire
Customer:
Date:
Ms./Mr.:
Department
Phone:
Fax:
Fluid:
Liquid content:
Flowrate:
(min., max., operating point)
Density:
(min., max., operating point)
Dyn. viscosity:
(min., max., operating point)
Fluid temperature:
(min., max., operating point)
Ambient temperature:
kg/h
Pressure:
(min., max., operating point)
Rate of flow:
Batch operation:
Concentration calculation:
Transmitter design:
Explosion protection:
Supply power:
Electrical outputs:
bar
kg/m
mPas/cP
C
C
Steady
Yes
Yes
Compact
Yes
100 230 V, 50/60 Hz
Current output I: 0/4 20 mA
Current output II: 0/4 20 mA
Pulse output, active
Pulse output, passive
Additional specifications:
Pipeline diameter:
Process connection:
Gas content:
Pulsating
No
No
Separate
No
24 V AC/DC, 50/60 Hz
Communication:
HART
.. mm
..
67
ABB Ltd.
Process Automation
Oldends Lane, Stonehouse
Gloucestershire, GL10 3TA
UK
Tel:
+44 (0)1453 826661
Fax:
+44 (0)1453 829671
ABB Inc.
Process Automation
125 E. County Line Road
Warminster PA 18974
USA
Tel:
+1 215 674 6000
Fax:
+1 215 674 7183
ABB Automation Products GmbH
Process Automation
Dransfelder Str. 2
37079 Goettingen
Germany
Tel:
+49 551 905-534
Fax:
+49 551 905-555
www.abb.com
Note
We reserve the right to make technical changes or
modify the contents of this document without prior
notice. With regard to purchase orders, the agreed
particulars shall prevail. ABB does not accept any
responsibility whatsoever for potential errors or
possible lack of information in this document.