Sie sind auf Seite 1von 16

measuring

Industrial Resistance
Thermometers according to
DIN

monitoring

analysing

OO

Measuring range:
-198...+600C

OO

 t100 sensor: class A,


P
class B, class 1/3,
class 1/10, or cryogenic

OO

 utput: resistance or
O
analogue 4...20 mA

OO

 ption: head transmitter


O
with HART -protocol
or Profibus /Fieldbus,
display

OO

Protection tube /
measuring insert material:
stainless steel 1.4404,
others on request

1/09-2015

KOBOLD companies worldwide:

ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, CANADA, CHILE, CHINA, COLoMBIA,


CZECHIA, EGYPT, FRANCE, GERMANY, GREAT BRITAIN, HUNGARY, INDIA, INDONESIA, ITALY,
MALAYSIA, MEXICO, NETHERLANDS, PERU, POLAND, Republic of Korea, ROMANIA,
SINGAPORE, SPAIN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, turkey, USA, VIETNAM

T2

MWD

KOBOLD Messring GmbH


Nordring 22-24
D-65719 Hofheim/Ts.
Head Office:
+49(0)6192 299-0
+49(0)6192 23398
info.de@kobold.com
www.kobold.com
1

Industrial Resistance Thermometers according to DIN Model MWD

Applications
OO HVAC
OO Machinery
OO Chemical and petrochemical industries
OO Heating/cooling processes
OO Industrial uses in general
The resistance thermometers with thread and flange
connection are favourably used for the temperature
measurement in liquids, solids and gaseous media. The
reliable water tightness of these installation methods for
gauge pressure and vacuum is an important criteria for
selection.

Head Transmitter
Resistance thermometers with head transmitter are used
whenever a measuring signal must be transported long
distance without any disturbance.
The head transmitter which is encapsulated in epoxies
resin is located right in the connection head and delivers
a temperature-linear output signal of 4-20 mA. The head
transmitter is available with standardised communication
systems just like HART protocol or Profibus/Fieldbus.

www.kobold.com

No responsibility taken for errors;


subject to change without prior notice.

1/09-2015

Description
KOBOLD temperature assemblies comprise a rugged
installation fitting made of stainless steel with thread, flange
or weld-on process connection, a connection head and a
measuring element.
A Pt100 temperature sensor according to IEC 751, class A, B,
1
/3 DIN or 1/10 is fitted in the measuring element. On request,
the measuring element can be removable, thus, allowing
to remove the measuring element without interrupting
the process as the customised thermowell remains in the
installation and seals the process.
Alternatively these sensors can be designed as single or
double resistance thermometers, except the 4-wire version,
which can only be built with one Pt100 due to lack of space.
As an option the resistance thermometers can be supplied
with a head transmitter. Transmitter with a standard 4-20 mA
signal, with HART protocol or with Profibus/Fieldbus are
there to choose from.

Technical Details
Measuring principle: temperature depending resistor
Measuring range: -70...+250 C
-70+400C
-70+600C
-198+100C
Sensor: Pt100 single or double sensor
(1xPt100 or 2xPt100) IEC751
Accuracy:
class A, class B, 1/3 DIN, 1/10 DIN
Ambient temperature: -40...+150 C
with ceramic terminal block
(without transmitter)
-40...+85C (with transmitter)
-20...+70C (with LCD display)
-20...+80C (with LED display)
Connection head: BR screwed cover, with chain,
DIN B cover with two screws,
BUZ, hinged cover
BUZ-H, high model hinged cover and
others
Electrical connection: M20x1.5 standard (others on request)
Materials:
Sensor:
stainless steel 1.4404
Connection head: aluminum, painted, stainless steel,
and PP (others on request)
Terminal block:
ceramic (without transmitter)
Process connection:
Thread: G, G, G, G1 male,
" NPT, " NPT, " NPT,
1" NPT male
DIN-flange:
DN15, 20, 25, 32, 40, 50

PN10, 16, 40
ANSI flange: ", ", 1", 1", 2" class 150,
class 300
Max. pressure:
30 bar
Sensor wiring:
2, 3 or 4-wire
Output:
resistance value
Protection: connection head IP54...68
depending on cable gland

Industrial Resistance Thermometers according to DIN Model MWD

D
2 = 2-wire
3 = 3-wire
43) = 4-wire

18

14

1/09-2015

1)

Only with head transmitter option '0'

2)

G = screw-cap with
chain, aluminum
I = screw-cap with
chain, stainless
steel 1.4401
P = screw-cap with
chain, PP
M1) = mini head
screw-cap with
chain, aluminum
K = head screwcap, stainless
steel 1.4401
B = DIN B screwed
cover, aluminum
Z = BUZ hinged
cover, aluminum
H = BUZ-H high
model with
hinged cover,
aluminum
E1) = aluminium head
with LCD display and HART
transmitter included
D = with LCD
display, stainless
steel 1.4301
R = as D + 2 relay
X = special options

Please specify the measuring range in clear text, while ordering

3)

Head
transmitter

Options

0 = without,
only with
ceramic
terminal
A2) = 5333D
transmitter
4-20 mA
2-wire
B2) = 5337D
D
transmitter
4-20 mA
with HART
protocol
0 = without
2-wire
Y = a
 cc. to
C2) = 5350A
specifitransmitter
cations
4-20 mA
Profibus/
Fieldbus
D = prepared for
subsequent
mounting of
transmitter
ML

Terminal head

45

A = 1xPt100 class B
(-70+250C)
B = 2xPt100 class B
(-70+250C)
C = 1xPt100 class A
(-70+250C)
D = 2xPt100 class A
(-70+250C)
E = 1xPt100 class B
(-70+400C)
F = 2xPt100 class B
(-70+400C)
G = 1xPt100 class A
(-70+400C)
H = 2xPt100 class A
(-70+400C)
I = 1xPt100 class B
(-70+600C)
J = 2xPt100 class B
(-70+600C)
K = 1xPt100 class A
(-70+600C)
L = 2xPt100 class A
(-70+600C)
M = 1xPt100 class
1 3
/ DIN (-70
+250C)
N = 1xPt100 class
1
/10 DIN (-70
+250C)
O = 1xPt100 class
1
/3 DIN (-70
+400C)
P = 1xPt100 class
1
/10 DIN (-70
+400C)
Q = 1xPt100 class
cryogenic
(-198+100C)
X = special options

HL=150mm

18

000 = none
K15 = compression fitting
G-M
C15 = compression fitting
" NPT-M
XXX = special
options

RTD
wiring

EL

L/C/F

Sensor type/
category

14.5

Process
connection

14.5

4 = tube 4
5 = tube 5
6 = tube 6
8 = tube 8
MWD-A 9 = tube 9
A = tube 10
C = tube 12
X = special
options

HL=130

EL

Protective
tube
diameter (D)

EL

EL

EL

EL

12

12

HL=130

HL=130

11

Model MWD-A compact compression fitting


Specifications:
Sensor element: Pt100 3-wire class B, A, 1/3, 1/10,
cryogenic, and others
10
10
Electrical connection: M20x1.5
Max. pressure:
30 bar (at 20C)
Protective tube: according to DIN 43772,
filled with magnesium oxide (MgO)
Material:
stainless steel 1.4404 (316L)

Model

B/D

A/E

for options A,
B, C, D choose
RTD wiring code
'3'

with 1xPt100 only

Note: Please specify length 'EL' in clear text, while ordering


No responsibility taken for errors;
subject to change without prior notice.

www.kobold.com

EL

EL

EL

D
D
Industrial Resistance Thermometers according to DIN Model MWD

Model MWD-L compact fixed thread


Specifications:
Sensor element: Pt100 3-wire class B, A, 1/3, 1/10,
cryogenic, and others
Electrical connection: M20x1.5
Max. pressure:
30 bar (at 20C)
Protective tube: according to DIN 43772,
filled with magnesium oxide (MgO)
Material:
stainless steel 1.4404 (316L)

EL

EL

HL=150mm

L/C/F

ML

Protective
tube
diameter (D)

Process
connection

Sensor type/
category
A = 1xPt100 class B
(-70+250C)
B = 2xPt100 class B
(-70+250C)
C = 1xPt100 class A
(-70+250C)
D = 2xPt100 class A
(-70+250C)
E = 1xPt100 class B
(-70+400C)
F = 2xPt100 class B
(-70+400C)
G = 1xPt100 class A
(-70+400C)
H = 2xPt100 class A
(-70+400C)
I = 1xPt100 class B
(-70+600C)
J = 2xPt100 class B
(-70+600C)
K = 1xPt100 class A
(-70+600C)
L = 2xPt100 class A
(-70+600C)
M = 1xPt100 class
1 3
/ DIN (-70
+250C)
N = 1xPt100 class
1
/10 DIN (-70
+250C)
O = 1xPt100 class
1
/3 DIN (-70
+400C)
P = 1xPt100 class
1
/10 DIN (-70
+400C)
Q = 1xPt100 class
cryogenic
(-198+100C)
X = special options

RTD
wiring

Terminal head

2 = 2-wire
3 = 3-wire
43) = 4-wire

G = screw-cap with
chain, aluminum
I = screw-cap with
chain, stainless
steel 1.4401
P = screw-cap with
chain, PP
M1) = mini head
screw-cap with
chain, aluminum
K = head screwcap, stainless
steel 1.4401
B = DIN B screwed
cover, aluminum
Z = BUZ hinged
cover, aluminum
H = BUZ-H high
model with
hinged cover,
aluminum
E1) = aluminium head
with LCD display and HART
transmitter
included
D = with LCD display, stainless
steel 1.4301
R = as D + 2 relay
X = special options

14

3 = tube 3
4 = tube 4
5 = tube 5
6 = tube 6
8 = tube 8
MWD-L
9 = tube 9
A = tube 10
C = tube 12
X = special
options

1)

G08 = G -M
G15 = G -M
G20 = G -M
G25 = G1-M
N08 = " NPT-M
N15 = " NPT-M
N20 = " NPT-M
N25 = 1" NPT-M
XXX = special
options

Only with head transmitter option '0'

2)

Head
transmitter

Options

45
18

18

14.5

14.5

Model

Please specify the measuring range in clear text, while ordering

3)

0 = without,
only with
ceramic
terminal
A2) = 5333D
transmitter
4-20 mA
2-wire
B2) = 5337D
transmitter
4-20 mA
with HART
protocol
0 = without
2-wire
Y = a
 cc. to
2)
C = 5350A
specifitransmitter
cations
4-20 mA
Profibus/
Fieldbus
D = prepared for
subsequent
mounting of
transmitter
for options A,
B, C, D choose
RTD wiring code
'3'

with 1xPt100 only

Note: Please specify length 'EL' in clear text, while ordering


4

www.kobold.com

No responsibility taken for errors;


subject to change without prior notice.

1/09-2015

Industrial Resistance Thermometers according to DIN Model MWD

B/D

Model MWD-B compact threaded or flanged with neck


pipe
Specifications:
Sensor element: Pt100 3-wire class B, A, 1/3, 1/10,
cryogenic, and others
Neck pipe HL:
130 mm
Electrical connection: M20x1.5
Max. pressure:
30 bar (at 20C)
Protective tube: according to DIN 43772,
filled with magnesium oxide (MgO)
Material:
stainless steel 1.4404 (316L)

10

EL

EL

EL

L/C/F

Protective
tube
diameter (D)

Process
connection

G08 = G -M
G15 = G -M
G20 = G -M
G25 = G 1-M

F15 = DN15 PN16


F20 = DN20 PN16
F25 = DN25 PN16
F32 = DN32 PN16
4 = tube 4 F40 = DN40 PN16
D
5 = tube 5 F50 = DN50 PN16
D
6 = tube 6 H15 = DN15 PN40
8 = tube 8 H20 = DN20 PN40
MWD-B 9 = tube 9 H25 = DN25 PN40
A = tube 10 H32 = DN32 PN40
C = tube 12 H40 = DN40 PN40
H50 = DN50 PN40
X = special
options
A15 = "CL150RF
A20 = " CL150RF
A25 = 1" CL150RF
A40 = 1 "
CL150RF
A50 = 2" CL150RF
B15 = " CL300RF
B20 = " CL300RF
B25 = 1" CL300RF
B40 = 1 "
CL300RF
B50 = 2" CL300RF
14

18

14.5

14.5

EL

EL

HL=150mm

N08 = " NPT-M


N15 = " NPT-M
N20 = " NPT-M
N25 = 1" NPT-M

XXX = special
options
1/09-2015

1)

Only with head transmitter option '0'

2)

Sensor type/
category
A = 1xPt100 class B
(-70+250C)
B = 2xPt100 class B
(-70+250C)
C = 1xPt100 class A
(-70+250C)
D = 2xPt100 class A
(-70+250C)
E = 1xPt100 class B
(-70+400C)
F = 2xPt100 class B
(-70+400C)
G = 1xPt100 class A
(-70+400C)
H = 2xPt100 class A
(-70+400C)
I = 1xPt100 class B
(-70+600C)
J = 2xPt100 class B
(-70+600C)
K = 1xPt100 class A
(-70+600C)
L = 2xPt100 class A
(-70+600C)
M = 1xPt100 class
1 3
/ DIN (-70
+250C)
N = 1xPt100
class 1/10 DIN
(-70+250C)
O = 1xPt100 class
1
/3 DIN (-70
+400C)
P = 1xPt100
class 1/10 DIN
(-70+400C)
Q = 1xPt100 class
cryogenic
(-198+100C)
X = special options

RTD
wiring

Terminal head

G = screw-cap with
chain, aluminum
I = screw-cap with
chain, stainless
steel 1.4401
P = screw-cap with
chain, PP
M1) = mini head
screw-cap with
chain, aluminum
K = head screwcap, stainless
D
steel 1.4401
B = DIN B screwed
2 = 2-wire
cover, aluminum
3 = 3-wire Z = BUZ hinged
cover, aluminum
43) = 4-wire
H = BUZ-H high
model with
hinged cover,
aluminum
E1) = aluminium head
with LCD display and HART
transmitter
included
D = with LCD display, stainless
steel 1.4301
R = as D + 2 relay
X = special options

ML

Model

45

18

EL

12

12

HL=130

HL=130

10

HL=130

11

A/E

Please specify the measuring range in clear text, while ordering

3)

Head
transmitter

Options

0 = without,
only with
ceramic
terminal
A2) = 5333D
transmitter
4-20 mA
2-wire
B2) = 5337D
transmitter
4-20 mA
with HART
protocol
0 = without
2-wire
Y = a
 cc. to
C2) = 5350A
specifitransmitter
cations
420mA
Profibus/
Fieldbus
D = prepared for
subsequent
mounting of
transmitter
for options A,
B, C, D choose
RTD wiring code
'3'

with 1xPt100 only

Note: Please specify length 'EL' in clear text, while ordering


No responsibility taken for errors;
subject to change without prior notice.

www.kobold.com

D
D
Industrial Resistance Thermometers according to DIN Model MWD

Model MWD-C compact with removable measuring


insert
Specifications:
Sensor element: Pt100 3-wire class B, A, 1/3, 1/10,
cryogenic, and others
Electrical connection: M20x1.5
Max. pressure:
30 bar (at 20C)
Protective tube: according to DIN 43772
Material:
stainless steel 1.4404 (316L)
Measuring insert:
filled with magnesium oxide (MgO)
ML

D
Sensor type/
category
A = 1xPt100 class B
(-70+250C)
B = 2xPt100 class B
(-70+250C)
C = 1xPt100 class A
(-70+250C)
D = 2xPt100 class A
(-70+250C)
E = 1xPt100 class B
(-70+400C)
F = 2xPt100 class B
(-70+400C)
G = 1xPt100 class A
(-70+400C)
H = 2xPt100 class A
(-70+400C)
I = 1xPt100 class B
(-70+600C)
J = 2xPt100 class B
(-70+600C)
K = 1xPt100 class A
(-70+600C)
L = 2xPt100 class A
(-70+600C)
M = 1xPt100 class
1 3
/ DIN (-70
+250C)
N = 1xPt100 class
1
/10 DIN (-70
+250C)
O = 1xPt100 class
1
/3 DIN (-70
+400C)
P = 1xPt100 class
1
/10 DIN (-70
+400C)
Q = 1xPt100 class
cryogenic
(-198+100C)
X = special options

RTD
wiring

Terminal head

2 = 2-wire
3 = 3-wire
43) = 4-wire

G = screw-cap with
chain, aluminum
I = screw-cap with
chain, stainless
steel 1.4401
P = screw-cap with
chain, PP
M1) = mini head
screw-cap with
chain, aluminum
K = head screwcap, stainless
steel 1.4401
B = DIN B screwed
cover, aluminum
Z = BUZ hinged
cover, aluminum
H = BUZ-H high
model with
hinged cover,
aluminum
E1) = aluminium head
with LCD display and HART
transmitter
included
D = with LCD display, stainless
steel 1.4301
R = as D + 2 relay
X = special options

14

5 = tube 5
6 = tube 6
8 = tube 8
9 = tube 9
MWD-C
A = tube 10
C = tube 12
X = special
options

G08 = G -M
G15 = G -M
G20 = G -M
G25 = G1-M
N08 = " NPT-M
N15 = " NPT-M
N20 = " NPT-M
N25 = 1" NPT-M
XXX = special
options

1)

Only with head transmitter option '0'

2)

Options

45

Head
transmitter

18

18

14.5

14.5

Process
connection

Please specify the measuring range in clear text, while ordering

3)

0 = without,
only with
ceramic
terminal
A2) = 5333D
transmitter
4-20 mA
2-wire
B2) = 5337D
transmitter
4-20 mA
with HART
protocol
0 = without
2-wire
Y = a
 cc. to
2)
C = 5350A
specifitransmitter
cations
4-20 mA
Profibus/
Fieldbus
D = prepared for
subsequent
mounting of
transmitter
for options A,
B, C, D choose
RTD wiring code
'3'

with 1xPt100 only

Note: Please specify length 'EL' in clear text, while ordering


6

www.kobold.com

No responsibility taken for errors;


subject to change without prior notice.

1/09-2015

HL=150mm
EL

EL

Protective
tube
diameter (D)

L/C/F

Model

EL

EL

EL

Industrial Resistance Thermometers according to DIN Model MWD

14

1/09-2015

1)

Only with head transmitter option '0'

2)

HL=130

HL=150mm

0 = without (only for


MWD-M)
G = screw-cap with
chain, aluminum
I = screw-cap with
chain, stainless
steel 1.4401
P = screw-cap with
chain, PP
M1) = mini head
screw-cap with
chain, aluminum
K = head screw-cap,
stainless steel
1.4401
2 = 2-wire DB = DIN B screwed
cover, aluminum
3 = 3-wire
43) = 4-wire Z = BUZ hinged
cover, aluminum
H = BUZ-H high
model with
hinged cover,
aluminum
E1) = aluminium head
with LCD display
and HART
transmitter
included
D = with LCD
display, stainless
steel 1.4301
R = as D + 2 relay
X = special options

Please specify the measuring range in clear text, while ordering

45

3)

Options

0 = without,
only with
ceramic
terminal
A2) = 5333D
transmitter
4-20 mA
2-wire
B2) = 5337D
transmitter
4-20 mA
with HART
protocol
D
2-wire
0 = without
C2) = 5350A
Y = a
 cc. to
transmitter
specifi4-20 mA
cations
Profibus/
Fieldbus
D = prepared for
subsequent
mounting
of transmitter

ML

Head
transmitter

Terminal head

18

14.5

EL

A = 1xPt100 class B
(-70+250C)
B = 2xPt100 class B
(-70+250C)
C = 1xPt100 class A
(-70+250C)
D = 2xPt100 class A
(-70+250C)
E = 1xPt100 class B
(-70+400C)
F = 2xPt100 class B
(-70+400C)
G = 1xPt100 class A
(-70+400C)
H = 2xPt100 class A
(-70+400C)
5 = tube 5 000 = none
I = 1xPt100 class B
6 = tube 6 K15 = compres(-70+600C)
sion fittingD
8 = tube 8
J
=
2xPt100 class B
G-M
9 = tube 9
(-70+600C)
MWD-E
C15 = compresA = tube 10
K =
1xPt100 class A
sion fitting
(-70+600C)
C = tube 12
" NPT-M
L = 2xPt100 class A
X = special
XXX = special
(-70+600C)
options
options
M = 1xPt100 class
1/3 DIN (-70
+250C)
N = 1xPt100 class
1/10 DIN (-70
+250C)
O = 1xPt100 class
1/3 DIN (-70
+400C)
P = 1xPt100 class
1/10 DIN (-70
+400C)
Q = 1xPt100 class
cryogenic
(-198+100C)
X = special options

RTD wiring

EL

L/C/F

Sensor type/
category

14.5

Process
connection

18

Protective
tube
diameter (D)

EL

EL

EL

EL

12

12

HL=130

HL=130

11

Model MWD-E compact with removable measuring


insert and compressing fitting
Specifications:
Sensor element: Pt100 3-wire class B, A, 1/3, 1/10,
10
10cryogenic, and others
Electrical connection: M20x1.5
Max. pressure:
30 bar (at 20C)
Protective tube: according to DIN 43772
Material:
stainless steel 1.4404 (316L)
Measuring insert:
filled with magnesium oxide (MgO)

Model

B/D

A/E

for options A,
B, C, D choose
RTD wiring code
'3'

with 1xPt100 only

Note: Please specify length 'EL' in clear text, while ordering


No responsibility taken for errors;
subject to change without prior notice.

www.kobold.com

Industrial Resistance Thermometers according to DIN Model MWD

B/D

HL=130

Model MWD-D compact with removable measuring


insert and extension neck pipe
Specifications:
Sensor element: Pt100 3-wire class B, A, 1/3, 1/10,
cryogenic, and others
Neck pipe HL:
130 mm
Electrical connection: M20x1.5
Max. pressure:
30 bar (at 20C)
Protective tube: according to DIN 43772
Material:
stainless steel 1.4404 (316L)
Measuring insert:
filled with magnesium oxide (MgO)

10

EL

EL

D
Process
connection

EL

EL

HL=150mm

G08 = G -M
G15 = G -M
G20 = G -M
G25 = G 1-M
N08 = " NPT-M
N15 = " NPT-M
N20 = " NPT-M
N25 = 1" NPT-M
F15 = DN15 PN16
F20 = DN20 PN16
F25 = DN25 PN16
F32 = DN32 PN16
F40 = DN40 PN16
F50 = DN50 PN16
6 = tube 6
H15 = DN15 PN40
8 = tube 8
D PN40
H20 = DN20
9 = tube 9
H25 =
DN25
PN40
MWD-D A = tube 10
H32 = DN32 PN40
C = tube 12
H40 = DN40 PN40
X = special
options H50 = DN50 PN40
A15 = " CL150RF
A20 = " CL150RF
A25 = 1" CL150RF
A40 = 1 "
CL150RF
A50 = 2" CL150RF
B15 = " CL300RF
B20 = " CL300RF
B25 = 1" CL300RF
B40 = 1 "
CL300RF
B50 = 2" CL300RF
XXX = special
options

14

18

14.5

14.5

1)

Only with head transmitter option '0'

2)

Sensor type/
category

RTD
wiring

A = 1xPt100 class B
(-70+250C)
B = 2xPt100 class B
(-70+250C)
C = 1xPt100 class A
(-70+250C)
D = 2xPt100 class A
(-70+250C)
E = 1xPt100 class B
(-70+400C)
F = 2xPt100 class B
(-70+400C)
G = 1xPt100 class A
(-70+400C)
H = 2xPt100 class A
(-70+400C)
I = 1xPt100 class B
(-70+600C)
J = 2xPt100 class B
(-70+600C)
K = 1xPt100 class A
(-70+600C)
L = 2xPt100 class A
(-70+600C)
M = 1xPt100 class
1 3
/ DIN (-70
+250C)
N = 1xPt100
class 1/10 DIN
(-70+250C)
O = 1xPt100 class
1
/3 DIN (-70
+400C)
P = 1xPt100
class 1/10 DIN
(-70+400C)
Q = 1xPt100 class
cryogenic
(-198+100C)
X = special options

ML

L/C/F

Protective
tube
diameter (D)

45

Model

18

Terminal head

G = screw-cap with
chain, aluminum
I = screw-cap with
chain, stainless
steel 1.4401
P = screw-cap with
chain, PP
M1) = mini head
screw-cap with
chain, aluminum
K = head screwcap, stainless
steel 1.4401
D B = DIN B screwed
2 = 2-wire
cover, aluminum
3 = 3-wire Z = BUZ hinged
cover, aluminum
43) = 4-wire
H = BUZ-H high
model with
hinged cover,
aluminum
E1) = aluminium head
with LCD display and HART
transmitter
included
D = with LCD display, stainless
steel 1.4301
R = as D + 2 relay
X = special options

Please specify the measuring range in clear text, while ordering

3)

Head
transmitter

Options

0 = without,
only with
ceramic
terminal
A2) = 5333D
transmitter
4-20 mA
2-wire
B2) = 5337D
transmitter
4-20 mA
with HART
protocol
0 = without
2-wire
Y = a
 cc. to
2)
C = 5350A
specifitransmitter
cations
4-20 mA
Profibus/
Fieldbus
D = prepared for
subsequent
mounting of
transmitter
for options A,
B, C, D choose
RTD wiring code
'3'

with 1xPt100 only

Note: Please specify length 'EL' in clear text, while ordering


8

www.kobold.com

No responsibility taken for errors;


subject to change without prior notice.

1/09-2015

EL

12

12

HL=130

10

HL=130

A/E

EL

EL

EL

D
D
D
Industrial Resistance Thermometers according to DIN Model MWD

ML

Model MWD-F compact fixed thread with removable


measuring insert, for mounting in thermowell
Specifications:
Sensor element: Pt100 3-wire class B, A, 1/3, 1/10,
cryogenic, and others
Process connection: threaded
Electrical connection: M20x1.5
Measuring insert: according to DIN 43772,
filled with magnesium oxide (MgO)
Material:
stainless steel 1.4404 (316L)

EL

EL

HL=150mm

L/C/F

Note: F
 or order details of the thermowell see datasheet
TWL-0.
D
D Without thermowell atmospheric pressure.

d
D

Measuring
insert
diameter (d)

Process
connection

Sensor type/
category
A = 1xPt100 class B
(-70+250C)
B = 2xPt100 class B
(-70+250C)
C = 1xPt100 class A
(-70+250C)
D = 2xPt100 class A
(-70+250C)
E = 1xPt100 class B
(-70+400C)
F = 2xPt100 class B
(-70+400C)
G = 1xPt100 class A
(-70+400C)
H = 2xPt100 class A
(-70+400C)
I = 1xPt100 class B
(-70+600C)
J = 2xPt100 class B
(-70+600C)
K = 1xPt100 class A
(-70+600C)
L = 2xPt100 class A
(-70+600C)
M = 1xPt100 class
1 3
/ DIN (-70
+250C)
N = 1xPt100 class
1
/10 DIN (-70
+250C)
O = 1xPt100 class
1
/3 DIN (-70
+400C)
P = 1xPt100 class
1
/10 DIN (-70
+400C)
Q = 1xPt100 class
cryogenic
(-198+100C)
X = special options

RTD wiring

Terminal head

2 = 2-wire
3 = 3-wire
43) = 4-wire

G = screw-cap with
chain, aluminum
I = screw-cap with
chain, stainless
steel 1.4401
P = screw-cap with
chain, PP
M1) = mini head
screw-cap with
chain, aluminum
K = head screw-cap,
stainless steel
1.4401
B = DIN B screwed
cover, aluminum
Z = BUZ hinged
cover, aluminum
H = BUZ-H high
model with
hinged cover,
aluminum
E1) = aluminium head
with LCD display
and HART
transmitter
included
D = with LCD
display, stainless
steel 1.4301
R = as D + 2 relay
X = special options

14

3 = tube 3
4 = tube 4
5 = tube 5
MWD-F 6 = tube 6
8 = tube 8
X = special
options

G08 = G -M
G15 = G -M
G20 = G -M
G25 = G 1-M
N08 = " NPT-M
N15 = " NPT-M
N20 = " NPT-M
N25 = 1" NPT-M
XXX = special
options

1/09-2015

1)

Only with head transmitter option '0'

2)

Head
transmitter

Options

45
18

18

14.5

14.5

Model

Please specify the measuring range in clear text, while ordering

3)

0 = without,
only with
ceramic
terminal
A2) = 5333D
transmitter
4-20 mA
2-wire
B2) = 5337D
transmitter 4-20
mA with
HART
protocol
0 = without
2-wire
Y = a
 cc. to
2)
C = 5350A
specifitransmitter
cations
4-20 mA
Profibus/
Fieldbus
D = prepared
for subsequent
mounting
of transmitter
for options A,
B, C, D choose
RTD wiring code
'3'

with 1xPt100 only

Note: Please specify length 'EL' in clear text, while ordering


No responsibility taken for errors;
subject to change without prior notice.

www.kobold.com

Industrial Resistance Thermometers according to DIN Model MWD

B/D

Model MWD-G compact with removable measuring


insert and neck pipe, for mounting in thermowell
Specifications:
Sensor element: Pt100 3-wire class B, A, 1/3, 1/10,
cryogenic, and others
Neck pipe HL:
130 mm
Electrical connection: M20x1.5
Measuring insert: according to DIN 43772,
filled with magnesium oxide (MgO)
Material:
stainless steel 1.4404 (316L)

10

EL

EL

G08 = G -M
G15 = G -M
G20 = G -M
G25 = G 1-M

3 = tube 3
4 = tube 4
5 = tube 5
MWD-G 6 = tube 6
8 = tube 8
X = special
options

N08 = " NPT-M


N15 = " NPT-M
N20 = " NPT-M
N25 = 1" NPT-M

14

Only with head transmitter option '0'

Sensor type/
category

A = 1xPt100 class B
(-70+250C)
B = 2xPt100 class B
(-70+250C)
C = 1xPt100 class A
(-70+250C)
D = 2xPt100 class A
(-70+250C)
E = 1xPt100 class B
(-70+400C)
F = 2xPt100 class B
(-70+400C)
G = 1xPt100 class A
(-70+400C)
H = 2xPt100 class A
(-70+400C)
I = 1xPt100 class B
D (-70+600C)
J = 2xPt100 class B
(-70+600C)
K = 1xPt100 class A
(-70+600C)
L = 2xPt100 class A
(-70+600C)
M = 1xPt100 class
1 3
/ DIN (-70
+250C)
N = 1xPt100 class
1
/10 DIN (-70
+250C)
O = 1xPt100 class
1
/3 DIN (-70
+400C)
P = 1xPt100 class
1
/10 DIN (-70
+400C)
Q = 1xPt100 class
cryogenic
(-198+100C)
X = special options

RTD wiring

2 = 2-wire
3 = 3-wire
43) = 4-wire

18

18

14.5

14.5

XXX = special
options

ML

Process
connection

HL=150mm

L/C/F

Measuring
insert
diameter (d)

45

Model

Note: F
 or order details of the thermowell see datasheet
TWL-0.
Without thermowell atmospheric pressure.

2)

Terminal head

G = screw-cap with
chain, aluminum
I = screw-cap with
chain, stainless
steel 1.4401
P = screw-cap with
chain, PP
M1) = mini head
screw-cap with
chain, aluminum
K = head screw-cap,
stainless steel
1.4401
D
B = DIN B screwed
cover, aluminum
Z = BUZ hinged
cover, aluminum
H = BUZ-H high
model with
hinged cover,
aluminum
E1) = aluminium head
with LCD display
and HART
transmitter
included
D = with LCD
display, stainless
steel 1.4301
R = as D + 2 relay
X = special options

Please specify the measuring range in clear text, while ordering

3)

Head
transmitter

Options

0 = without,
only with
ceramic
terminal
A2) = 5333D
transmitter
4-20 mA
2-wire
B2) = 5337D
transmitter
4-20 mA
with HART
protocol
2-wire
0 = without
C2) = 5350A
Y = a
 cc. to
transmitter
specifi4-20 mA
cations
Profibus/
Fieldbus
D = prepared for
subsequent
mounting
of transmitter
for options A,
B, C, D choose
RTD wiring code
'3'

with 1xPt100 only

Note: Please specify length 'EL' in clear text, while ordering


10

www.kobold.com

No responsibility taken for errors;


subject to change without prior notice.

1/09-2015

d
D

1)

EL

EL

EL

EL

12

12

HL=130

HL=130

10

HL=130

11

A/E

12
EL

EL

12
EL

EL

Industrial Resistance Thermometers according to DIN Model MWD

EL

ML

HL=150mm

L/C/F

Note: F
 or order details of the thermowell see datasheet
TWL-0.
Without
thermowell atmospheric pressure.
D

EL

Dd
Measuring
insert
diameter (d)

Process
connection

Sensor type/
category

N15 = " NPT-M


XXX = special
options

A = 1xPt100 class B
(-70+250C)
B = 2xPt100 class B
(-70+250C)
C = 1xPt100 class A
(-70+250C)
D = 2xPt100 class A
(-70+250C)
E = 1xPt100 class B
(-70+400C)
F = 2xPt100 class B
(-70+400C)
G = 1xPt100 class A
(-70+400C)
H = 2xPt100 class A
(-70+400C)
I = 1xPt100 class B
(-70+600C)
J = 2xPt100 class B
(-70+600C)
K = 1xPt100 class A
(-70+600C)
L = 2xPt100 class A
(-70+600C)
M = 1xPt100 class
1 3
/ DIN (-70
+250C)
N = 1xPt100 class
1
/10 DIN (-70
+250C)
O = 1xPt100 class
1
/3 DIN (-70
+400C)
P = 1xPt100 class
1
/10 DIN (-70
+400C)
Q = 1xPt100 class
cryogenic
(-198+100C)
X = special options

1/09-2015

1)

Terminal head

2 = 2-wire
3 = 3-wire
43) = 4-wire

0 = without (only for


MWD-M)
G = screw-cap with
chain, aluminum
I = screw-cap with
chain, stainless
steel 1.4401
P = screw-cap with
chain, PP
M1) = mini head
screw-cap with
chain, aluminum
K = head screwcap, stainless
steel 1.4401
B = DIN B screwed
cover, aluminum
Z = BUZ hinged
cover, aluminum
H = BUZ-H high
model with
hinged cover,
aluminum
E1) = aluminium head
with LCD display and HART
transmitter
included
D = with LCD display, stainless
steel 1.4301
R = as D + 2 relay
X = special options

18

18

14.5
14

3 = tube 3
4 = tube 4
5 = tube 5
MWD-W 6 = tube 6
8 = tube 8
X = special
options

RTD
wiring

45

14.5

Model

Model MWD-W compact with removable measuring


insert and neck pipe (nipple union nipple), for
mounting in thermowell
Specifications:
Sensor element: Pt100 3-wire class B, A, 1/3, 1/10,
cryogenic, and others
Neck pipe HL:
150 mm, (nipple union nipple)
Electrical connection: M20x1.5
Measuring insert: according to DIN 43772,
filled with magnesium oxide (MgO)
Material:
stainless steel 1.4404 (316L)

Only with head transmitter option '0'

2)

Please specify the measuring range in clear text, while ordering

3)

Head
transmitter

Options

0 = without,
only with
ceramic
terminal
A2) = 5333D
transmitter
4-20 mA
2-wire
B2) = 5337D
transmitter
4-20 mA
with HART
protocol
0 = without
2-wire
Y = a
 cc. to
C2) = 5350A
specifitransmitter
cations
4-20 mA
Profibus/
Fieldbus
D = prepared for
subsequent
mounting of
transmitter
for options A,
B, C, D choose
RTD wiring code
'3'

with 1xPt100 only

Note: Please specify length 'EL' in clear text, while ordering


No responsibility taken for errors;
subject to change without prior notice.

www.kobold.com

11

EL

EL

Industrial Resistance Thermometers according to DIN Model MWD

ML

Model MWD-M measuring insert


Specifications:
Sensor element: Pt100 3-wire class B, A, 1/3, 1/10,
cryogenic, and others
Head transmitter:
ceramic terminal block or transmitter
Max. pressure:
atmospheric pressure
Measuring insert: according to DIN 43772,
filled with magnesium oxide (MgO)
Material:
stainless steel 1.4404 (316L)

Dd
Measuring
insert
diameter (d)

Process
connection

18

45

Model

3 = tube 3
4 = tube 4
5 = tube 5
MWD-M 6 = tube 6
8 = tube 8
X = special
options

1)

0 = without

only with head transmitter option '0'

2)

Sensor type/
category
A = 1xPt100 class B
(-70+250C)
B = 2xPt100 class B
(-70+250C)
C = 1xPt100 class A
(-70+250C)
D = 2xPt100 class A
(-70+250C)
E = 1xPt100 class B
(-70+400C)
F = 2xPt100 class B
(-70+400C)
G = 1xPt100 class A
(-70+400C)
H = 2xPt100 class A
(-70+400C)
I = 1xPt100 class B
(-70+600C)
J = 2xPt100 class B
(-70+600C)
K = 1xPt100 class A
(-70+600C)
L = 2xPt100 class A
(-70+600C)
M = 1xPt100 class
1 3
/ DIN (-70
+250C)
N = 1xPt100 class
1
/10 DIN (-70
+250C)
O = 1xPt100 class
1
/3 DIN (-70
+400C)
P = 1xPt100 class
1
/10 DIN (-70
+400C)
Q = 1xPt100 class
cryogenic
(-198+100C)
X = special options

RTD
wiring

2 = 2-wire
3 = 3-wire
42) = 4-wire

Terminal
head

0 = without

Head
transmitter

Options

0 = without, only with


ceramic terminal
A = 5333D transmitter
4-20 mA 2-wire
B = 5337D transmitter
4-20 mA with
HART protocol
2-wire
C = 5350A transmitter
4-20 mA Profibus/Fieldbus
D = prepared for
subsequent
mounting of
transmitter

0 = without
Y = acc. to
specifications

with 1xPt100 only

Note: Please specify length 'ML' in clear text


12

www.kobold.com

No responsibility taken for errors;


subject to change without prior notice.

1/09-2015

38

27

83

41

41

Typ Z

27

38

50
50

64

41

39

38
9

Typ Z

Typ M
91

91
8
4

000.0

000.0

41

28

Typ E

Typ E

41

115
85

55
8
39 91

91

H
Option HTyp
Housing
material: aluminium

Protection:
IP65

Option Z Housing material: Aluminium



Protection:
IP65
116

116

60

Note: IP protection depends on cable gland.


Higher
IP on request.

000.0 000.0
c

Typ E

Typ E

72

Typ D

78
23

www.kobold.com

No responsibility taken for errors;


subject to change without prior notice.

26

72

Typ D

27

115

85
72

116

115
85

104,5

Z 13

39
42
46
Typ
Z
28

42

8
28

Typ BTyp D
Typ B

39

104,5

28Typ

55
8

26
42

Typ D

82
72

38

104,5
72

116
50

26

72

104,5

50

Typ K

55

115
85 72

28

84

Typ B
Typ Option
B
B Housing material: aluminium
Typ G, I, P IP54

Protection:

42

55
110
82,5

83

110
60
82,5

Typ K

33
26

87
72 41

28
39

IP65

8
28

39

23

64

23

55
8

42

27

39

78
23

50
27

83

83
26

42

84

42

41

34

72

50

8
28

M110 60
82,5

27

50

50

72

78
23

27

41
50

41

55

84

84

,5

Protection:

78
23
83

55

Typ Z

60

38

110 8
82,5

8Typ

39

61

83
78
27
23

83
38

4255

,5

39
Option M Housing material: aluminium

Protection:
IP65

K
Typ H
TypTyp
H K stainless steelTyp
Option K 9 Housingmaterial:
1.4401
34

27
38

83

38

Typ M

83

41

41

38

38
9
34

Typ H

23

23

64

33

38

Typ G, I, PTyp G, I, P

Typ H

61

84

Housing material: aluminium


55
55
Housing material:
stainless steel
8 1.4401
8
Housing material: PP 39
8
8
4
4
Protection:
IP65

88

,5

46

27

27

13
38

61

50

87

33
50

87
83

46

60

64

23

82

13

Typ M

83

34

34

Typ M

Typ G, I, P
82

38
9

1/09-2015

,5

,5

38

38

50

Typ

61

61

38

46

13

Typ G, I, P

Terminal Heads

64

23

64

33

87

33

87

50

46
38

Option G
I
P

82

82

Industrial Resistance Thermometers according to DIN Model MWD

13

26

13

72

88

Typ
TypZ Z

3939

Typ
TypHH
6060

50
50

50
50

78
78
23
23

72
72
26
26

2
Industrial Resistance727Thermometers
according to DIN Model MWD

4242
88

4242

2828

88

Terminal Heads (continuation)


TypKK
Option E Typ

2828

Typ
TypBB

110
110
82,5
82,5

000.0
000.0

116
116

104,5
104,5

115
115
8585

Technical Details
Housing material:
aluminium
Supply voltage:
7.5...45 VDC
Output signal:
4...20 mA

9191
Load:
max. (Vsupply 7.5 V)/0.022 A
Signal on alarm: underranging: linear drop to 3.8 mA
overranging: linear rise to 20.5 mA
sensor break, sensor open-circuit:
c c 3.6 mA or 22 mA
Galvanic isolation: 2000 VAC (input/output)
Response time:
1s
Long
term
stability:
RTDs 0.1% of reading or 0.1C
Typ
TypE E
Switch delay:
5 s
Influence of ambient: negligible
Load influence:
negligible
Power supply
influence: negligible
Filter configurating: 0...160 A
Resolution:
0.3 A
Cable entry:
M20 internal thread
Ambient temperature: -40...+85C (-40...185F)
Storage temperature: -40...+100C (-40...212F)
Condensation: allowable
Protection:
IP65
Shock and vibration
resistance:
4g/2 to 150 Hz as per IEC60068-26
LCD-Display,
with backlight: visible range 32.5 x 22.5 mm;
main display: 5-digit, 7-segment,
digit height 8 mm,
range: -19999...99999
additional display: 8-digit 14 segment,
digit height 5 mm for sensor type and
units

Typ
TypDD

Main Features
OO High accuracy 0.02% of span for Pt100 sensor
OO Sensor error detection according to the guidelines in
NAMUR NE 89
OO Backlight LCD-Display

Accessories (for programming the transmitter)


Model

14

HART modem

www.kobold.com

No responsibility taken for errors;


subject to change without prior notice.

1/09-2015

KM-HART
(includes configuration software KM-Soft)

Description

72

50

26

Industrial Resistance Thermometers according to DIN Model MWD

42
8

28

Typ B

Terminal Heads (continuation)


Option D/R
115

85

91

000.0

116

72

Typ E

1/09-2015

Main Features
OO Graphical display with backlight
OO Showing errors and limits by steady or flashing colours
OO Programmable by touch screen
OO Option R: two configurable relay outputs
Technical Details
Input
Housing material:
stainless steel 1.4301
Measuring range: 3.523 mA (normal working range
4...20 mA)
Connections: 2 screw terminals for loop power,
signal 4 screw terminals for relay
outputs
Accuracy:  0.1% of input span within
-10...70C
0.2% of input span within
-30...-10C/70...80C
Loop voltage drop: two levels depending on chosen
backlight brightness at 20 mA:
low bright backlight (<40%):
max. 4 V at 23C
high bright backlight (>40%):
max. 6.5 V at 23C
Sample time:
1 s, typical 0.3 s
Start-up time:
5s
User-configurable data
Measuring range:
4...20 mA
Error/warning
indication: individually configurable display and
backlight indication in white, green
or red colour, steady or flashing
light; configurable limits between 3.5
and 23 mA
Zoom on range:
minimum 2 mA of input span
Damping:
0...30 s
Linearisation table:
2...30 points
Measuring unit
(standard selectable): C, F, K
User defined unit:
8 x 20 pixels matrix
No responsibility taken for errors;
subject to change without prior notice.

Dec. point position: xxxxx, xxxx.x, xxx.xx, xx.xxx, x.xxxx,


.xxxxx, AUTO
Display
Type:
FSTN graphically LCD
Measuring range:
-9999...99999
Digit height:
max. 22 mm
Environmental conditions
Optimal readability:
-10...70C
Operating temperature: -30...80C
Storage temperature: -40...85C
Mechanical data
Material:
polycarbonate plastic
Enclosure:  80 mm housing and front ring
stainless steel, AISI304
Protection: IP10 on terminals
IP67 in 80 mm housing
Relay (option R)
Contacts:
2 solid state relays
Voltage:
60 Vp
Load current:
75 mA

www.kobold.com

15

Industrial Resistance Thermometers according to DIN Model MWD

Electrical Connection

1xPt100, 2-wire

16

2xPt100, 2-wire

1xPt100, 3-wire

1xPt100, 4-wire

2xPt100, 3-wire

www.kobold.com

No responsibility taken for errors;


subject to change without prior notice.

1/09-2015

Das könnte Ihnen auch gefallen