Consolidated Sterilizer
Fm: Boston, MA (617) 782-6072
Model: SSR-3A-ADVPB
Serial No.: 101910
Chamber Pressure
in psi
Autoclave Chamber
strainer - cleaned and
checked daily
PUMP
Boiler or Generator
Check Valve
Sight Glass to
evaluate level in
the boiler or steam
generator
Steam Generator
or Steam Boiler
Red Handle valves can
be closed if glass fails ,
so as not to render the
Autoclave useless.
It allows or can be
operated iwth these
valves closed
typ water level
Steam to Jacket
GENERATOR or BOILER
Should be ** 70 - 75psi **
Boiler or Generator
heating elements
Should be measuring
~480Volts for the heating
elements
120VAC
distribution to
Valve Solenoids
to jacket
From Boiler/Generator
removed the
control vavle lower
strainer cover
to jacket
From Boiler/Generator
2nd - remove
4phillips screws
Apollo Model
jacket pressure
safety relief valve
press > 39 psi
releases steam
Parker solenoid
STEAM FLOW
from Jacket to
Chamber
Liquid Cycle
Exhaust Steam
from the Jacket assist in quicker
cooling cycle
LIQUID CYCLE
Solenoid Exhaust
Undo Universal
joint - exhaust side
pipe
Ideal conditions
check valve
steam to chamber
check valve
to chamber
solenoid
from jacket to
chamber
From Jacket
(Steam)
Air Filter
Airflow back into
chamber when done
in Gravity Cycle only
Check Valve
Steam To Chamber
Another view
from back side
check valves
Jacket Steam
direction to the
Back of the
Chamber
BACK OF
AUTOCLAVE
steam into
Chamber
On this one
membrane Tab
should be on this
side since steam
flow is into jacket
notice tab
solenoid Rubber
plunger kit for
Parker valves
steam flow
From Jacket
To Chamber
Flow direction
Liquid Cycle
Cooling Assist
Reassemble the
Parker shaft with
new plunger &
membrane
External Temp
Probe - Chamber
Temperature to
controller
Chamber Exhaust
Other Side - Opposite is steam exhaust trap
Slight by-pass
Check Valve
Creates vacuum
exhaust for pulling
out the chamber
steam
Exhaust temp
sensor to solenoid
cooling water valve
Cooling Water
Exhaust Temp
sensor
Jacket Exhaust Steam Trap Allows water to escape and traps the steam
Chamber Exhaust
Steam Trap
From Jacket
Jacket Steam TRAP
Chamber "Strainer"
Discharge
Steam Blocked
Steam exhaust
to chamber steam
trap
Water discharges thru trap
water Discharged
Generator or Boiler
Safety Relief Valve
Set to 100 PSI
steam direction
Steam
Check Valve
Chamber Discharge
Steam Solenoid
MIX
COOLING WATER
Steam discharge
Cooling Water
Main Discharge
STERLCO Valve
Range: 100degs F - 175 degs F
Series : 55 - T 125PSI Max
RAISE
LOWER
Discharge Water
Temp sensor to valve
MIXING
MAIN Discharge
of Steam / Water
Orifice - Creates
Vacuum, that helps pull
out Steam from
Chamber
Steam Exhaust
MIXING
Cooling Water
EXHAUST
Main Exhaust of
Mixed Steam and
Water
Sterilco Valve
Jacket Steam Trap
Lower
From
Liquid
Cooling
Phase,
Additional
Jacket discharge
from above
from Chamber
Steam Trap
Main Discharge
Line of Steam
mixed with
Cooling Water to
Reduce its Temp
TO DRAIN