Sie sind auf Seite 1von 74

APOLLO EZ

System Integration Guide


Release 7 April 2011 (Rev. 0)

SYSTEM INTEGRATION GUIDE


Contents

Contents
1

INTRODUCTION

CONFIGURATION WITH APOLLO EZ/ CCD N23 / G100R/F-A /


CHEST STAND

2.1

Power supplies connection instruction.................................................... 4

2.2

Hardware interface instruction ............................................................... 5


2.2.1 APOLLO EZ Generator interface ........................................................ 5
2.2.2 Generator I.I. component plate interface............................................ 6
2.2.3 APOLLO EZ I.I. component plate interface ......................................... 7
2.2.4 Monitor interface................................................................................. 8
2.2.5 Generator Three field solid state AEC device 1 ................................... 9
2.2.6 Generator - X-ray tube interface ........................................................ 10
2.2.7 Chest Stand VD3 35x43 Generator interface ................................... 11
2.2.8 Generator Three field solid state AEC device 2 ................................. 12
2.2.9 Generator jumpers and wiring configuration instruction .................... 13
2.2.9.1
2.2.9.2
2.2.9.3
2.2.9.4

2.3

Room interface board jumpers configuration....................... 13


Room interface board jumper wiring ................................... 13
Generator interface PWB jumpers configuration.................. 13
Accessories connection....................................................... 14

Generator set-up programming instruction........................................... 15


2.3.1 Tube selection ................................................................................... 15
2.3.1.1
2.3.1.2
2.3.1.3

2.3.2
2.3.3
2.3.4
2.3.5
2.3.6
2.3.7

X-ray tube present in library .............................................. 15


Loading X-ray tube data from GenWare utility software .... 16
X-ray tube not present neither in library nor in
GenWare utility software .................................................. 16
AEC set-up ....................................................................................... 17
Fluoro set-up .................................................................................... 17
Generator Image receptor programming worksheet ............................ 19
Generator Receptor 1 I/O configuration worksheet............................. 20
Generator Receptor 2 I/O configuration worksheet............................. 21
Generator Receptor 3 I/O configuration worksheet............................. 22

CONFIGURATION WITH APOLLO EZ/ CCD N33 / G100R/F-A /


CHEST STAND

23

3.1

Power supplies connection instruction.................................................. 24

3.2

Hardware interface instruction ............................................................. 25


3.2.1 APOLLO EZ G 100 R/F-A interface.................................................. 25
3.2.2 G 100 R/F-A I.I. component plate interface ..................................... 27
3.2.3 APOLLO EZ I.I. component plate interface ....................................... 28
3.2.4 Monitor interface............................................................................... 29
3.2.5 G 100 R/F-A Three field solid state AEC device 1............................. 30
3.2.6 G 100 R/F-A - X-ray tube interface.................................................... 31

(Rev. 0)

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Contents

3.2.7
3.2.8
3.2.9

3.3

Chest Stand VD3 35x43 G 100 R/F-A interface............................... 32


G 100 R/F-A Three field solid state AEC device 2 ............................ 33
G 100 R/F-A jumpers and wiring configuration instruction................ 34
3.2.9.1
Room interface board jumpers configuration ...................... 34
3.2.9.2
Room interface board jumper wiring................................... 34
3.2.9.3
Generator interface PWB jumpers configuration ................. 34
3.2.9.4
T.D.I. board SW1 Dip-switches setting ............................... 35
3.2.9.5
Accessories connection ...................................................... 35

G 100 R/F-A set-up programming instruction ...................................... 36


3.3.1 Tube selection................................................................................... 36
3.3.1.1
3.3.1.2
3.3.1.3

3.3.2
3.3.3
3.3.4
3.3.5
3.3.6
3.3.7

X-ray tube present in library .............................................. 36


Loading X-ray tube data from GenWare utility software .... 37
X-ray tube not present neither in library nor in
GenWare utility software.................................................. 37
AEC set-up ....................................................................................... 38
Fluoro set-up .................................................................................... 38
G 100 R/F-A Image receptor programming worksheet ........................ 41
G 100 R/F-A Receptor 1 I/O configuration worksheet ........................ 42
G 100 R/F-A Receptor 2 I/O configuration worksheet ........................ 43
G 100 R/F-A Receptor 3 I/O configuration worksheet ........................ 44

CONFIGURATION WITH APOLLO EZ/ DIVA-D / G100R/F-A /


CHEST STAND VD3

45

4.1

Power supplies connection instruction.................................................. 46

4.2

Hardware interface instruction ............................................................. 47


4.2.1 APOLLO EZ G 100 R/F-A interface ................................................. 47
4.2.2 APOLLO EZ DIVA-D interface.......................................................... 49
4.2.3 G 100 R/F-A DIVA-D interface........................................................ 50
4.2.4 G 100 R/F-A DIVA-D serial connection ........................................... 50
4.2.5 APOLLO EZ I.I. component plate interface....................................... 51
4.2.6 DIVA-D I.I. component plate interface ............................................. 52
4.2.7 Monitor interface............................................................................... 52
4.2.8 G 100 R/F-A Three field solid state AEC device 1 ............................ 53
4.2.9 G 100 R/F-A - X-ray tube interface.................................................... 54
4.2.10 Chest Stand VD3 35x43 G 100 R/F-A interface............................... 55
4.2.11 G 100 R/F-A Three field solid state AEC device 2 ............................ 56
4.2.12 G 100 R/F-A jumpers and wiring configuration instruction................ 57
4.2.12.1
4.2.12.2
4.2.12.3
4.2.12.4
4.2.12.5

APOLLO EZ

Room interface board jumpers configuration ...................... 57


Room interface board jumper wiring................................... 57
Generator interface PWB jumpers configuration ................. 57
T.D.I. board SW1 Dip-switches setting ............................... 58
Accessories connection ...................................................... 58

ii

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Contents

4.3

G 100 R/F-A set-up programming instruction ...................................... 59


4.3.1 Tube selection ................................................................................... 59
4.3.1.1
4.3.1.2
4.3.1.3

4.3.2
4.3.3
4.3.4
4.3.5
4.3.6
4.3.7

X-ray tube present in library .............................................. 59


Loading X-ray tube data from GenWare utility software .... 60
X-ray tube not present neither in library nor in
GenWare utility software .................................................. 60
AEC set-up ....................................................................................... 61
Fluoro set-up .................................................................................... 61
G 100 R/F-A Image receptor programming worksheet ........................ 64
G 100 R/F-A Receptor 1 I/O configuration worksheet ........................ 65
G 100 R/F-A Receptor 2 I/O configuration worksheet ........................ 66
G 100 R/F-A Receptor 3 I/O configuration worksheet ........................ 67

This publication can only be reproduced, transmitted, transcribed or translated into


any human or computer language with the written consent of VILLA SISTEMI
MEDICALI S.p.A.
This Manual in English is the original version.
(Rev. 0)

iii

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Contents

THIS PAGE IS INTENTIONALLY LEFT BLANK

APOLLO EZ

iv

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Introduction

INTRODUCTION
This Manual includes the interconnections list, the hardware and
software configuration for a typical radiological room using a generator
G 100 R/F-A.
For the use of this generator with different configurations of room and for
X-ray tube and AEC calibration, please make reference to the
G 100 R/F-A Service Manual and equipment you have to connect with.

(Rev. 0)

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Introduction

THIS PAGE IS INTENTIONALLY LEFT BLANK

APOLLO EZ

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

CONFIGURATION WITH APOLLO EZ/


CCD N23 / G100R/F-A / CHEST STAND
Room configuration

Apollo EZ
I.A.E. X-ray tube type
Three field solid state AEC
Standard I.I. CCD N23.

G 100 R/F-A

Chest Stand VD3 35x43 with motorised foot rest (See Note)
Three field solid state AEC.

Receptor configuration
Receptor 1: Apollo EZ Spot Film Device Tube 1
Receptor 2: Direct X-ray from console Tube 1
Receptor 3: Chest Stand VD3 35x43 Tube 1 (See Note)
NOTE:
When the Generator is supplied with "Falling load" feature, the GalKey
U29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the
"SERIAL" exposure is enable (paragraph 2.3.4 - Image Receptor
Programming Worksheet: "SERIAL=YES"); whenever the Falling Load has
to be activated on the Spot Film Device receptor, the SERIAL parameter
has to be set to "NO".
In case Falling Load feature and SERIAL exposure are required will be
necessary to create for each a dedicated receptor.
NOTE:
Reference to the Chest Stand VD3 35x43 with motorized foot rest is made
in order to allows the installer to record the necessary information for the
hardware interface among Generator and Potter Bucky.

(Rev. 0)

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.1

Power supplies connection instruction

Main
Power Supply

Generator
Terminals

G 100 R/F
ROOM INTERFACE BOARD
TB7.1 (0 Vac)
* INSTALLATION KIT
230Vac AUXILIARY CONTACTOR
V.S.M. cod.4291515200
G 100 R/F
ROOM INTERFACE BOARD
TB10.1 (220 Vac)

R S T

(The auxiliary contactor is already mounted


inside the G 100 R/F-A Generator)

APOLLO EZ
Terminals R-S-T

R S

CHEST STAND
Terminal X1 (**)

(*) If this remote controlled switch is not available use a remote controlled
switch with the following characteristics:

Coil:
220-230 Vac 50 Hz
230-240 Vac 60 Hz

Contact: 4 N.O. contacts, 16 A.


(**) Made this connection only when a motorised foot rest chest stand is present.

APOLLO EZ

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.2

Hardware interface instruction

2.2.1

APOLLO EZ Generator interface

APOLLO EZ
Description

Generator

Terminal

Terminal

Description

X27.15

TB6.9

Remote exposure
(Room Interface Board)

X27.3

TB6.7

Remote prep
(Room Interface Board)

X-ray
request
(2 step)

X27.16

TB2.11

Bucky 1 select
(Room Interface Board)

Common
0VS

X0.20

TB6.1

Ale output
(Room Interface Board)

TB6.11

Spare output
(Room Interface Board)

TB6.2

Ale output
(Room Interface Board)

TB3.6

Tomo exposure
(Room Interface Board)

X-ray ON
from
generator

X0.13

Generator
ready

X0.12

TB2.12

Bucky 1 select
(Room Interface Board)

Tomo speed
1-2-3-4

X27.4
X27.1
X27.14
X27.17

TB3.4

Remote tomo select


(Room Interface Board)

SFD
selection

X0.15

TB6.12

Spare output
(Room Interface Board)

Safety
switch

X0.9

J17.1

Safety switch
(Generator Interface PWB)

Safety
switch

X0.11

J17.2

Safety switch
(Generator Interface PWB)

(Rev. 0)

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

Generator I.I. component plate interface

2.2.2

Generator

N23 I.I. component plate

Description

Terminal

Remote Fluoro Exposure


(Room Interface board)

Pcb A2

Description

TB6.5

X2.3

Fluoro

0Vdc
(Generator Interface PWB)

J7.7

X2.1

0Vdc

Proportional dc
(Generator Interface PWB)

J7.12

X2.2

Analog kV auto

APOLLO EZ

Wire n

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

APOLLO EZ I.I. component plate interface

2.2.3

APOLLO EZ

I.I. component plate

Description

Terminal

Common

Wire n

Pcb A1

Description

4 fields

3 fields

X28.1

X4.4

X4.4

Common

Medium field

X28.2

X4.2

X4.1

Medium zoom

Minimum field

X28.14

X4.3

X4.2

Maximum field

X28.15

X4.1

N.C.

Power supply

X0.2

Power supply

Power supply

S3

X0.1

Power supply

X0.3

Ground

Ground

APOLLO EZ
Description

0 Vs common

(Rev. 0)

Maximum
zoom
Minimum
zoom

I.I. component plate


Terminal

Wire n

Pcb A2

Description

X28.18

X3.3

Fluoro request

X28.4

X3.4

Inv

X28.16

X3.5

Inv

X28.3

X3.2

0 Vs common

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

Monitor interface

2.2.4

I.I. component plate

Table and wall version

Description

Pcb A2

Wire n

Plug

Description

0 Vs common

X4.1

Green

0V

0 Vs common

Monitor L.I.H.
visualization

X4.5

White

LI

Monitor L.I.H.
visualization

APOLLO EZ

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.2.5

Generator Three field solid state AEC device 1


NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.

Generator
AEC board
(solid state chamber)
type 737992

AEC device 1
connecting cable

Description Terminal

Wire

Description

Anode (L)

J11.1

Pink

Anode Left

Cath (L)

J11.2

Grey

Cathode Left

Anode (M)

J11.3

White

Anode Middle

Cath (M)

J11.4

Brown

Cathode Middle

Anode (R)

J11.5

Green

Anode Right

Cath (R)

J11.6

Yellow

Cathode Right

Ground

J11.7

Shield

Ground

(Rev. 0)

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

Generator - X-ray tube interface

2.2.6

Generator

X-ray tube

Description

Terminal

Main

Terminal

Description

Tube 1

Principal

Shift

Tube 1

Auxiliary

Common

Tube 1

Common

Ground

Tube 1

Ground

Thermal switch

Tube 1

Thermal switch

Thermal switch

Tube 1

Thermal switch

APOLLO EZ

Wire n

10

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.2.7

Chest Stand VD3 35x43 Generator interface


NOTE:
This hardware interface has to be made only if the Potter Bucky is
present.

Chest stand VD3 35x43


Description

Terminal

X-ray request

Generator
Wire n

Terminal

Description

X1.B1

TB1.11

Bucky 3 select

X-ray request

X1.B2

TB1.12

Bucky 3 select

Potter reply

X1.X

TB2.5

Bucky contacts

Potter reply

X1.X

TB2.4

Bucky contacts

WARNING:
The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the Chest
Stand VD3 35x43 Technical and User's Manual for the Bucky start
voltage setting.

(Rev. 0)

11

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.2.8

Generator Three field solid state AEC device 2


NOTE:
This hardware interface has to be made only if the Potter Bucky with
AEC device is present.
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.

Generator
AEC board
(solid state chamber)
type 737992

AEC device 2
connecting cable

Description Terminal

Wire

Description

Anode (L)

J12.1

Pink

Anode Left

Cath (L)

J12.2

Grey

Cathode Left

Anode (M)

J12.3

White

Anode Middle

Cath (M)

J12.4

Brown

Cathode Middle

Anode (R)

J12.5

Green

Anode Right

Cath (R)

J12.6

Yellow

Cathode Right

Ground

J12.7

Shield

Ground

APOLLO EZ

12

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.2.9

Generator jumpers and wiring configuration instruction

2.2.9.1

Room interface board jumpers configuration

2.2.9.2

2.2.9.3

JW2:

JW4:
JW7:
JW8:

Live contact (this jumper configuration must be made only if


the Potter Bucky is present on Receptor 3)
Dry contact
Dry contact (2-3)
Dry contact.

Room interface board jumper wiring

TB8.1 TB11.5:

TB4.8 TB4.9:

(This wiring must be made only if the 24Vdc


Potter Bucky is present on Receptor 3)
Thermal switch.

Generator interface PWB jumpers configuration


-

JW1:
JW2:
JW3:
JW10:
JW11:
JW12:
JW13:
JW14:
JW15:
JW21:

12
12/
12/
12/
23
Open
Open
12/
12/
2-3

34
34
34

34
34

NOTE:
The configuration of the Jumpers not listed in the previous tables is not
relevant for this system integration.

(Rev. 0)

13

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.2.9.4

Accessories connection

Generator

Accessory

Description

Terminal

Wire n

Description

Room door interlock


(Room Interface board)

TB4.4

Door contact

Room door interlock


(Room Interface board)

TB4.5

Door contact

Room light 0V
(Room Interface board)

TB7

Generator ON lamp
(white lamp)

Room light power supply


(Room Interface board)

(*)

Generator ON lamp
(white lamp)

Room light contact


(Room Interface board)

TB4.11

X-ray ON lamp
(red lamp)

Room light contact


(Room Interface board)

TB4.12

X-ray ON lamp
(red lamp)

(*) TB10 for 220V ac / TB9 for 110V ac / TB8 for 24V dc

NOTE:
Remove jumper JW13 on Room Inteface Board.

APOLLO EZ

14

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.3

Generator set-up programming instruction


After the hardware connections indicated in the previously tables it is
needed to proceed to the software configuration of the Generator.
Refer to the Generator Service Manual, chapter 3, where are described
the interfacing and the programming information to the generator. This
chapter also allows the installer to record the necessary information to
complete the installation, as well as to record the programming values.
Refer to the tables in paragraphs 2.3.2, 2.3.3, 2.3.4, 2.3.5, 2.3.6, 2.3.7 of
this Installation Guide for the programming of the generator set-up
submenu.

2.3.1

Tube selection
Prior to beginning tube calibration, the tube used in the installation must
be properly selected, and generator limits should be programmed.

2.3.1.1

X-ray tube present in library


If the x-ray tube used in the installation is present in the library of the
generator refer to the Generator Service Manual, chapter 3C Programming the generator in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
Generator Service Manual, chapter 2 - Programming the low speed
starter-Programming the dual speed starter to record the necessary
information to complete the configuration.
NOTE:
The tube library, concerning the I.A.E. X-ray tubes, contains only high
speed tube types. When a low speed I.A.E. X-ray tube has to be installed
an high speed tube must be selected informing the generator about the
low speed tube. The generator and the tube limits must be set in
according with the tube type installed.

(Rev. 0)

15

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.3.1.2

Loading X-ray tube data from GenWare utility software


If the X-ray tube used in the installation is not present in the library of
the generator should be present in the utility software. The data
communication with a computer it is allowed using the GenWare utility
software. Refer to the Generator Service Manual, chapter 3C - DATA LINK
to record the necessary information.

2.3.1.3

X-ray tube not present neither in library nor in GenWare utility


software
If the x-ray tube used in the installation is not present neither in the
generator library nor in the GenWare utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the x-ray
tube that is going to be installed. An X-ray tube present in the library or
in the GenWare utility software, as similar as possible to the one is
going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the x-ray tube. Refer to
the Generator Service Manual, chapter 3C - Tube selection and
Generator limits in order to record all the necessary information to
program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
Generator Service Manual, chapter 2 - Programming the low speed
starter - Programming the dual speed starter to record the necessary
information to complete the configuration.

APOLLO EZ

16

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.3.2

AEC set-up
The installation of the AEC devices and their calibration are covered in
the Generator Service Manual, chapter 3E - Automatic exposure control
(AEC) calibration.

2.3.3

Fluoro set-up
NOTE:
The Fluoro set-up data must be set in according with the X-ray tube
technical specification.
The following tables meets the Fluoro set-up request in according with
most of the X-ray tube supplied by Villa Sistemi Medicali; during the
installation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in the
Generator Service Manual, chapter 3 - Automatic brightness stabilisation
(ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP

FL TIMER:
5 Min
MIN FLUORO KV:
40
MAX FLUORO KV:
120
II MODES:
0
FL ABS:
ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES
SID COMP
FL SETUP

Not Used
Not Used
FL ABS SETUP

ABS CURVE DEF:

(Rev. 0)

17

ABS LEVEL 0:
ABS LEVEL 1:
ABS LEVEL 2:
ABS LEVEL 3:
SIGNAL GAIN
ABS GAIN
ABS DEADBAND
ABD DELAY
ABS START KV

1.00 V
1.60 V
1.10 V
1.60 V
10
20
3%
150 mS
75

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

FL ABS CURVE

FL ABS CURVE 1:
ON
40KV:
0.5 mA
50 KV:
1.0 mA
60 KV:
1.6 mA
70 KV:
2.1 mA
80 KV:
2.7 mA
90 KV:
3.2 mA
100 KV:
3.8 mA
110 KV:
4.4 mA
120 KV:
5.0 mA
5.0mA
125 KV:
* FL MAN DL *
40KV:
0.5 mA
50 KV:
1.0 mA
60 KV:
1.6 mA
70 KV:
2.1 mA
80 KV:
2.7 mA
90 KV:
3.2 mA
100 KV:
3.8 mA
110 KV:
4.4 mA
120KV:
5.0 mA
5.0 mA
125 KV:

FL DOSE LIMIT

* FL ABS DL *

Note: The fluoro dose limit is set at 5 mA in order to


meet the code of federal regulations 21 paragraph
1020. If it is not required to meet these regulations
a higher value can be set in according with the
loading factors of the x-ray tube.

HLF SETUP
PF SETUP

FL II COMP
Not Used
PF:

40KV:

5.0 mA

50 KV:

5.0 mA

60 KV:

5.0 mA

70 KV:

5.0 mA

80 KV:

5.0 mA

90 KV:

5.0 mA

100 KV:

5.0 mA

110 KV:

5.0 mA

120KV:

5.0 mA

125 KV:

5.0 mA

Not Used
OFF

NOTE:
The Pulsed Fluoro set-up is not relevant for this system integration
because the function is not available with the N23 CCD camera.

APOLLO EZ

18

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.3.4

Generator Image receptor programming worksheet

IMAGE RECEPTOR PROGRAMMING WORKSHEET


DATE:
SERIAL#:
FUNCTION

RECEPTOR
1

RECEPTOR
2

RECEPTOR
3

RECEPTOR
4

RECEPTOR
5

RECEPTOR
6

RECEPTOR
NAME
TUBE

NONE

NONE

NONE

TOMO

YES

NO

NO

NO

NO

NO

FLUORO

YES

NO

NO

NO

NO

NO

SERIAL

YES

NO

NO

NO

NO

NO

INTERFACE
OPTS

FUNCTIONAL
OPTS

FLUORO
HANG

30 SEC

30 SEC

30 SEC

30 SEC

30 SEC

30 SEC

RAD HANG

0 SEC

0 SEC

0 SEC

0 SEC

0 SEC

0 SEC

LAST IMAGE
HOLD

40 Ms

40 Ms

40 Ms

40 Ms

40 Ms

40 Ms

MEMORY

NO

NO

NO

NO

NO

NO

REM TOMO
BUT

2500 Ms

2500 Ms

2500 Ms

2500 Ms

2500 Ms

2500 Ms

SF/LF
SWITCH

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

AEC BACKUP

FIXED

FIXED

FIXED

FIXED

FIXED

FIXED

AEC BACKUP
MAS

500

500

500

500

500

500

AEC BACKUP
MS

3200

3200

3200

3200

3200

3200

AEC
CHANNEL

AEC LOCK
EXPS

RECEPTOR
SYM

NOTE:
Receptor 3 has to be programmed only if Chest Stand with Potter Bucky
and AEC device is present.

(Rev. 0)

19

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.3.5

Generator Receptor 1 I/O configuration worksheet

FUNCTIONS

STANDBY

PREP

GEN RDY

RAD EXP

FLUORO EXP

ON
--------OFF
----ON
----OFF
OFF
OFF
OFF
ON
OFF
ON
-----

--------ON
----------------OFF
OFF
----OFF
ON
OFF
ON
-----

ON
OFF
OFF
OFF
OFF
ON
OFF
ON
ON

OFF
OFF
OFF
OFF
OFF
ON
OFF
ON
ON

***INPUTS***
REMOTE EXP
REMOTE PREP
REMOTE FL. EXP
CONSOLE EXP
CONSOLE PREP
TOMO EXP
REM. TOMO SEL
I/I SAFETY
COLL. ITLK
BUCKY CONTACTS
SPARE
THERMAL SW 1
THERMAL SW 2
DOOR ITLK
MULTI SPOT EXP

------------------------ON
------------OFF
ON
OFF
OFF
OFF

OFF
ON
----OFF
OFF
--------OFF
OFF
----OFF
ON
OFF
ON
-----

-------------------------------------------------------------

***OUTPUTS***
BKY 1 SELECT
BKY 2 SELECT
BKY 3 SELECT
TOMO/BKY 4 SEL
TOMO/BKY START
ALE
COLL. BYPASS
ROOM LIGHT
SPARE

APOLLO EZ

OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON

ON
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON

20

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.3.6

Generator Receptor 2 I/O configuration worksheet

FUNCTIONS

STANDBY

PREP

GEN RDY

RAD EXP

FLUORO EXP

ON
--------OFF
----OFF
----OFF
OFF
OFF
OFF
ON
OFF
ON
-----

--------OFF
----------------OFF
OFF
----OFF
ON
OFF
ON
-----

ON
OFF
OFF
OFF
OFF
ON
OFF
ON
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

***INPUTS***
REMOTE EXP
REMOTE PREP
REMOTE FL. EXP
CONSOLE EXP
CONSOLE PREP
TOMO EXP
REM. TOMO SEL
I/I SAFETY
COLL. ITLK
BUCKY CONTACTS
SPARE
THERMAL SW 1
THERMAL SW 2
DOOR ITLK
MULTI SPOT EXP

------------------------OFF
------------OFF
ON
OFF
OFF
OFF

OFF
ON
----OFF
OFF
--------OFF
OFF
----OFF
ON
OFF
ON
-----

-------------------------------------------------------------

***OUTPUTS***
BKY 1 SELECT
BKY 2 SELECT
BKY 3 SELECT
TOMO/BKY 4 SEL
TOMO/BKY START
ALE
COLL. BYPASS
ROOM LIGHT
SPARE

(Rev. 0)

OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

ON
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

21

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N23 /
G100R/F-A / Chest Stand

2.3.7

Generator Receptor 3 I/O configuration worksheet

FUNCTIONS

STANDBY

PREP

GEN RDY

RAD EXP

FLUORO EXP

OFF
--------ON
----OFF
----OFF
OFF
ON
OFF
ON
OFF
ON
-----

--------OFF
----------------OFF
OFF
----OFF
ON
OFF
ON
-----

OFF
OFF
ON
OFF
OFF
OFF
OFF
ON
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

***INPUTS***
REMOTE EXP
REMOTE PREP
REMOTE FL. EXP
CONSOLE EXP
CONSOLE PREP
TOMO EXP
REM. TOMO SEL
I/I SAFETY
COLL. ITLK
BUCKY CONTACTS
SPARE
THERMAL SW 1
THERMAL SW 2
DOOR ITLK
MULTI SPOT EXP

------------------------OFF
------------OFF
ON
OFF
OFF
OFF

OFF
OFF
----OFF
ON
--------OFF
OFF
----OFF
ON
OFF
ON
-----

-------------------------------------------------------------

***OUTPUTS***
BKY 1 SELECT
BKY 2 SELECT
BKY 3 SELECT
TOMO/BKY 4 SEL
TOMO/BKY START
ALE
COLL. BYPASS
ROOM LIGHT
SPARE

OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

NOTE:
The configuration of the Receptor 3 I/O has to be made only if the Chest
Stand with Potter Bucky is present.

APOLLO EZ

22

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

CONFIGURATION WITH APOLLO EZ/


CCD N33 / G100R/F-A / CHEST STAND
Room configuration

Apollo EZ

I.A.E. X-ray tube type

Three field solid state AEC

I.I. H.R. CCD N33.

G 100 R/F-A (T.D.I. Interfaced)

Chest Stand VD3 35x43 with motorised foot rest (See Note)

Three field solid state AEC.

Receptor configuration
Receptor 1: Apollo EZ Spot Film Device Tube 1
Receptor 2: Direct X-ray from console Tube 1
Receptor 3: Chest Stand VD3 35x43 Tube 1 (See Note)
NOTE:
When the Generator is supplied with "Falling load" feature, the GalKey
U29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the
"SERIAL" exposure is enable (paragraph 3.3.4 - Image Receptor
Programming Worksheet: "SERIAL=YES"); whenever the Falling Load has
to be activated on the Spot Film Device receptor, the SERIAL parameter
has to be set to "NO".
In case Falling Load feature and SERIAL exposure are required will be
necessary to create for each a dedicated receptor.
NOTE:
Reference to the Chest Stand VD3 35x43 with motorized foot rest is made
in order to allows the installer to record the necessary information for the
hardware interface among G 100 R/F-A and Potter Bucky.

(Rev. 0)

23

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.1

Power supplies connection instruction

Main
Power Supply

Generator
Terminals

G 100 R/F
ROOM INTERFACE BOARD
TB7.1 (0 Vac)
* INSTALLATION KIT
230Vac AUXILIARY CONTACTOR
V.S.M. cod.4291515200
G 100 R/F
ROOM INTERFACE BOARD
TB10.1 (220 Vac)

R S T

(The auxiliary contactor is already mounted


inside the G 100 R/F-A Generator)

APOLLO EZ
Terminals R-S-T

R S

CHEST STAND
Terminal X1 (**)

(*) If this remote controlled switch is not available use a remote controlled
switch with the following characteristics:

Coil:
220-230 Vac 50 Hz
230-240 Vac 60 Hz

Contact: 4 N.O. contacts, 16 A.


(**) Made this connection only when a motorised foot rest chest stand is present.

APOLLO EZ

24

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.2

Hardware interface instruction

3.2.1

APOLLO EZ G 100 R/F-A interface

APOLLO EZ
Description

Terminal

G 100 R/F-A
Wire n

Terminal

Description

X27.15

TB6.9

Remote Exposure
(Room Interface Board)

X27.3

TB6.7

Preparation radiography
(Room Interface Board)

Common
0Vdc

X0.20

TB6.11

Spare Output
(Room Interface Board)

X-ray ON
from
Generator

X0.13

J6.21

X-ray ON
(Table Digital Interface board)

Generator
ready

X0.12

J6.22

Generator ready
(Table Digital Interface board)

Spot Film
Device
selection

X0.15

TB6.12

Spare Output
(Room Interface Board)

Safety switch

X0.9

J17.1

Safety switch
(Generator Interface PWB)

Safety switch

X0.11

J17.2

Safety switch
(Generator Interface PWB)

Alarm
X-ray ON

X27.7
(see Note)

J7.6

X-ray OFF
(Table Digital Interface board)

Fluoro
Man/Auto

X27.20

J6.1

Fluoro Man/Auto
(Table Digital Interface board)

NOTE:
During the auto calibration of the X-ray tube it is required to make the
jumper between J7.6 and J6.2 (force terminal J7.6 at 0Vdc).
It is important to remove the jumper at the end of the X-ray tube
auto calibration.

(Rev. 0)

25

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

APOLLO EZ
Description

Terminal

Vref. Anal.
Fluoro

X27.11

Tomo
Speed 1

X27.4

Tomo
Speed 2

X27.1

Tomo
Speed 3

X27.14

Tomo
Speed 4

G 100 R/F-A
Terminal

Description

J6.7

Analogue Vref Fluoro Manual kV


(Table Digital Interface board)

J6.3

Tomo Speed 1
(Table Digital Interface board)

X27.17

J6.4

Tomo Speed 2
(Table Digital Interface board)

Tomo 7

X27.5

J7.1

Tomo Angle 7
(Table Digital Interface board)

Tomo 20

X27.18

J7.3

Tomo Angle 20
(Table Digital Interface board)

Tomo 30

X27.6

J7.4

Tomo Angle 30
(Table Digital Interface board)

Tomo 45

X27.19

J7.5

Tomo Angle 45
(Table Digital Interface board)

Common 0V

X27.22

J6.2

Signal Ground
(Table Digital Interface board)

Fluoro Man
kV analog 0V

X27.23

J6.23

Signal Ground
(Table Digital Interface board)

J7.2

Signal Ground
(Table Digital Interface board)

J7.7

Chassis Ground
(Table Digital Interface board)

J7.20

Chassis Ground
(Table Digital Interface board)

APOLLO EZ

Wire n

26

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

G 100 R/F-A I.I. component plate interface

3.2.2

G 100 R/F-A

I.I. component plate

Description

Terminal

Remote Fluoro Exposure


(Room Interface board)

Pcb A2

Description

TB6.5

X2.3

Fluoro

L.I.H.
(Table Digital Interface
board)

J6.17

X2.4

Last Image Hold

Proportional dc
(Table Digital Interface
board)

J7.21

X2.1

Analog kV auto

0Vdc
(Table Digital Interface
board)

J7.2

X2.2

0Vdc

SYNC
(Table Digital Interface
board)

J7.24

X4.5

SYNC

P.F. request
(Table Digital Interface
board)

J6.22

X4.6

Pulsed Fluoro

(Rev. 0)

Wire n

27

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

APOLLO EZ I.I. component plate interface

3.2.3

APOLLO EZ

I.I. component plate

Description

Terminal

Common

Wire n

Pcb A1

Description

4 fields

3 fields

X28.1

X4.4

X4.4

Common

Medium field

X28.2

X4.2

X4.1

Medium zoom

Minimum field

X28.14

X4.3

X4.2

Maximum field

X28.15

X4.1

N.C.

Power supply

X0.2

Power supply

Power supply

S3

X0.1

Power supply

X0.3

Ground

Ground

APOLLO EZ

Maximum
zoom
Minimum
zoom

I.I. component plate

Description

Terminal

R.O.I.

X28.8
(see Note)

Fluoro call

Wire n

Pcb A2

Description

X2.15

R.O.I.

X2.11

Recursive Filter X0

X28.18

X2.5

Fluoro call

Vertical image
inversion

X28.16

X2.8

Vertical image
inversion

Horizontal image
inversion

X28.4

X2.9

Horizontal image
inversion

Recursive Filter X4

X28.21
(see Note)

X2.12

Recursive Filter X4

Recursive Filter X8

X28.9
(see Note)

X2.13

Recursive Filter X8

Common 0V

X28.3

X4.10

Common 0V

Common 0V

X28.22

NOTE:
Depending by selection of the Apollo EZ set-up, cell 246 X28.8 output
can drive R.O.I. or Recursive Filter. In the first case X28.21 and X28.9
are disabled, in the second case are enabled.

APOLLO EZ

28

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

Monitor interface

3.2.4

I.I. component plate

Table and wall version

Description

Pcb A2

Wire n

Plug

Description

Monitor L.I.H.
visualization

X4.2

White

LI

Monitor L.I.H.
visualization

0 Vs common

X4.10

Green

0V

0 Vs common

(Rev. 0)

29

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.2.5

G 100 R/F-A Three field solid state AEC device 1


NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.

G100 R/F-A
AEC board
(solid state chamber)
type 737992

AEC device 1
connecting cable

Description Terminal

Wire

Description

Anode (L)

J11.1

Pink

Anode Left

Cath (L)

J11.2

Grey

Cathode Left

Anode (M)

J11.3

White

Anode Middle

Cath (M)

J11.4

Brown

Cathode Middle

Anode (R)

J11.5

Green

Anode Right

Cath (R)

J11.6

Yellow

Cathode Right

Ground

J11.7

Shield

Ground

APOLLO EZ

30

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.2.6

G 100 R/F-A - X-ray tube interface

G 100 R/F-A

X-ray tube

Description

Terminal

Main

Terminal

Description

Tube 1

Principal

Shift

Tube 1

Auxiliary

Common

Tube 1

Common

Ground

Tube 1

Ground

Thermal switch

Tube 1

Thermal switch

Thermal switch

Tube 1

Thermal switch

(Rev. 0)

Wire n

31

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.2.7

Chest Stand VD3 35x43 G 100 R/F-A interface


NOTE:
This hardware interface has to be made only if the Potter Bucky is
present.

Chest stand VD3 35x43


Description

Terminal

X-ray request

G 100 R/F-A
Wire n

Terminal

Description

X1.B1

TB1.11

Bucky 3 select

X-ray request

X1.B2

TB1.12

Bucky 3 select

Potter reply

X1.X

TB2.5

Bucky contacts

Potter reply

X1.X

TB2.4

Bucky contacts

WARNING:
The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the Chest
Stand VD3 35x43 Technical and User's Manual for the Bucky start
voltage setting.

APOLLO EZ

32

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.2.8

G 100 R/F-A Three field solid state AEC device 2


NOTE:
This hardware interface has to be made only if the Potter Bucky with
AEC device is present.
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.

G100 R/F-A
AEC board
(solid state chamber)
type 737992

AEC device 2
connecting cable

Description Terminal

Wire

Description

Anode (L)

J12.1

Pink

Anode Left

Cath (L)

J12.2

Grey

Cathode Left

Anode (M)

J12.3

White

Anode Middle

Cath (M)

J12.4

Brown

Cathode Middle

Anode (R)

J12.5

Green

Anode Right

Cath (R)

J12.6

Yellow

Cathode Right

Ground

J12.7

Shield

Ground

(Rev. 0)

33

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.2.9

G 100 R/F-A jumpers and wiring configuration instruction

3.2.9.1

Room interface board jumpers configuration

3.2.9.2

3.2.9.3

JW2:

JW3:
JW4:
JW8:

Live contact (this jumper configuration must be made only if


the Potter Bucky is present on Receptor 3)
Live contact
Live contact
Dry contact (Spot Film Device selection).

Room interface board jumper wiring

TB8.1 TB11.5

TB4.8 TB4.9:

(This wiring must be made only if the 24Vdc


Potter Bucky is present on Receptor 3)
Thermal switch.

Generator interface PWB jumpers configuration


-

JW1:
JW2:
JW3:
JW4:
JW5:
JW6:
JW7:
JW8:
JW9:
JW10:
JW11:
JW12:

12
12/
12/
Open
Close
12/
12/
12/
12/
12/
23
Open

34
34

3
3
3
3
3

4
4
4
4
4

JW13:
JW14:
JW15:
JW19:
JW20:
JW21:
JW22:
JW24:
JW25:
JW26:
JW27:
JW28:

Open
1 2 / 3 4
12/34
23
Close
12
23
Close
Close
Close
Close
Open

NOTE:
The configuration of the Jumpers not listed in the previous tables is not
relevant for this system integration.

APOLLO EZ

34

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.2.9.4

T.D.I. board SW1 Dip-switches setting


-

3.2.9.5

SW1-1:
SW1-2:
SW1-3:
SW1-4:

OFF
OFF
ON
OFF

SW1-5:
SW1-6:
SW1-7:
SW1-8:

ON
OFF
OFF
OFF

Accessories connection

Generator

Accessory

Description

Terminal

Wire n

Description

Room door interlock


(Room Interface board)

TB4.4

Door contact

Room door interlock


(Room Interface board)

TB4.5

Door contact

Room light 0V
(Room Interface board)

TB7

Generator ON lamp
(white lamp)

Room light power supply


(Room Interface board)

(*)

Generator ON lamp
(white lamp)

Room light contact


(Room Interface board)

TB4.11

X-ray ON lamp
(red lamp)

Room light contact


(Room Interface board)

TB4.12

X-ray ON lamp
(red lamp)

(*) TB10 for 220V ac / TB9 for 110V ac / TB8 for 24V dc

NOTE:
Remove jumper JW13 on Room Inteface Board.

(Rev. 0)

35

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.3

G 100 R/F-A set-up programming instruction


After the hardware connections indicated in the previously tables it is
needed to proceed to the software configuration of the G 100 R/F-A.
Refer to the G 100 R/F-A Service Manual, chapter 3, where are described
the interfacing and the programming information to the generator. This
chapter also allows the installer to record the necessary information to
complete the installation, as well as to record the programming values.
Refer to the tables in paragraphs 3.3.2, 3.3.3, 3.3.4, 3.3.5, 3.3.6, 3.3.7 of
this Installation Guide for the programming of the generator set-up
submenu.

3.3.1

Tube selection
Prior to beginning tube calibration, the tube used in the installation must
be properly selected, and generator limits should be programmed.

3.3.1.1

X-ray tube present in library


If the x-ray tube used in the installation is present in the library of the
generator refer to the G 100 R/F-A Service Manual, chapter 3C Programming the generator in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter-Programming the dual speed starter to record the necessary
information to complete the configuration.

APOLLO EZ

36

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.3.1.2

Loading X-ray tube data from GenWare utility software


If the X-ray tube used in the installation is not present in the library of
the generator should be present in the utility software. The data
communication with a computer it is allowed using the GenWare utility
software. Refer to the G 100 R/F-A Service Manual, chapter 3C - DATA
LINK to record the necessary information.

3.3.1.3

X-ray tube not present neither in library nor in GenWare utility


software
If the x-ray tube used in the installation is not present neither in the
generator library nor in the GenWare utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the x-ray
tube that is going to be installed. An X-ray tube present in the library or
in the GenWare utility software, as similar as possible to the one is
going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the x-ray tube. Refer to
the G 100 R/F-A Service Manual, chapter 3C - Tube selection and
Generator limits in order to record all the necessary information to
program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter - Programming the dual speed starter to record the necessary
information to complete the configuration.

(Rev. 0)

37

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.3.2

AEC set-up
The installation of the AEC devices and their calibration are covered in
the G 100 R/F-A Service Manual, chapter 3E - Automatic exposure
control (AEC) calibration.

3.3.3

Fluoro set-up
NOTE:
The Fluoro set-up data must be set in according with the X-ray tube
technical specification.
The following tables meets the Fluoro set-up request in according with
most of the X-ray tube supplied by Villa Sistemi Medicali; during the
installation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in the
G 100 R/F-A Service Manual, chapter 3 - Automatic brightness
stabilisation (ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP

FL TIMER:
5 Min
MIN FLUORO KV:
40
MAX FLUORO KV:
120
II MODES:
0
FL ABS:
ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF:
NONE
FL/RAD KV CURVES
SID COMP
FL SETUP

Not Used
Not Used
FL ABS SETUP

ABS CURVE DEF:

APOLLO EZ

38

ABS LEVEL 0:
ABS LEVEL 1:
ABS LEVEL 2:
ABS LEVEL 3:
SIGNAL GAIN
ABS GAIN
ABS DEADBAND
ABD DELAY
ABS START KV

1.00 V
4.00 V
1.10 V
1.60 V
3
10
30%
150 mS
75

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

FL ABS CURVE

FL DOSE LIMIT

Note: The fluoro dose limit is set at 5 mA in order to


meet the code of federal regulations 21 paragraph
1020. If it is not required to meet these regulations
a higher value can be set in according with the
loading factors of the x-ray tube.

HLF SETUP

(Rev. 0)

FL II COMP
Not Enable

39

FL ABS CURVE 1:
ON
40KV:
0.5 mA
50 KV:
1.0 mA
60 KV:
1.6 mA
70 KV:
2.1 mA
80 KV:
2.7 mA
90 KV:
3.2 mA
100 KV:
3.8 mA
110 KV:
4.4 mA
120 KV:
5.0 mA
5.0mA
125 KV:
* FL MAN DL *
40KV:
0.5 mA
50 KV:
1.0 mA
60 KV:
1.6 mA
70 KV:
2.1 mA
80 KV:
2.7 mA
90 KV:
3.2 mA
100 KV:
3.8 mA
110 KV:
4.4 mA
120KV:
5.0 mA
5.0mA
125 KV:
* FL ABS DL *
40KV:

5.0 mA

50 KV:

5.0 mA

60 KV:

5.0 mA

70 KV:

5.0 mA

80 KV:

5.0 mA

90 KV:

5.0 mA

100 KV:

5.0 mA

110 KV:

5.0 mA

120KV:

5.0 mA

125 KV:

5.0 mA

Not Used

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

PF SETUP

PF:
PF DEFAULT:
PPS DEFAULT:
PF DEFAULT MA:

ON
NONE
NONE
5=40mA
6=60mA
7=80mA
8=99mA

Note: Four options for the settings of the Pulsed Fluoro default mA are possible. It is
suggested to set the value 5 = 40mA as default parameter. The others settings are
possible on line by means the +/- mA keys. If, once set at 40 mA the default mA value
for the pulsed fluoro, is necessary to reduce the dose is possible to reduce the length
of the pulses by means the settings of the jumpers on the A2 board of the I.I. system
(make reference to the I.I. system service manual)

PF ABS SETUP

PF ABS CURVE DEF:


OFF
PF ABS CURVES
PF DOSE LIMITS
PF II COMP
PF PPS SETUP

HPF SETUP
HLF SELECT:
CONSOLE
PF SELECT:
CONSOLE

APOLLO EZ

40

PF DEFAULT Ms:
PF SYNC FREQ:
ABS LEVEL 0:
ABS LEVEL 1:
ABS LEVEL 2:
ABS LEVEL 3:
SIGNAL GAIN:
ABS GAIN:
ABS DEADBAND:
ABS DELAY:
ABS START KV:
PF INTEGRATOR:
1PPS COMP:
3PPS COMP:
6/7.5PPS COMP:
12/15PPS COMP:
25/30PPS COMP:
Not Used
Not Used
Not Used
1 PPS:
3 PPS:
6/7.5 PPS:
12.5/15 PPS:
28/30 PPS:
Not Enable

2
50 HZ
4.0 V
1.10 V
1.20 V
1.90 V
3
5
40%
100 mS
75
OFF
0%
0%
0%
0%
0%

ON
ON
ON
ON
ON

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.3.4

G 100 R/F-A Image receptor programming worksheet

IMAGE RECEPTOR PROGRAMMING WORKSHEET


DATE:
SERIAL#:
FUNCTION

RECEPTOR
1

RECEPTOR
2

RECEPTOR
3

RECEPTOR
4

RECEPTOR
5

RECEPTOR
6

RECEPTOR
NAME
TUBE

NONE

NONE

NONE

TOMO

NO

NO

NO

NO

NO

NO

FLUORO

YES

NO

NO

NO

NO

NO

SERIAL

YES

NO

NO

NO

NO

NO

INTERFACE
OPTS

23

FUNCTIONAL
OPTS

FLUORO
HANG

30 SEC

30 SEC

30 SEC

30 SEC

30 SEC

30 SEC

RAD HANG

0 SEC

0 SEC

0 SEC

0 SEC

0 SEC

0 SEC

LAST IMAGE
HOLD

40 Ms

40 Ms

40 Ms

40 Ms

40 Ms

40 Ms

MEMORY

NO

NO

NO

NO

NO

NO

REM TOMO
BUT

2500 Ms

2500 Ms

2500 Ms

2500 Ms

2500 Ms

2500 Ms

SF/LF
SWITCH

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

AEC BACKUP

FIXED

FIXED

FIXED

FIXED

FIXED

FIXED

AEC BACKUP
MAS

500

500

500

500

500

500

AEC BACKUP
MS

3200

3200

3200

3200

3200

3200

AEC
CHANNEL

AEC LOCK
EXPS

RECEPTOR
SYM

NOTE:
Receptor 3 has to be programmed only if Chest Stand with Potter Bucky
and AEC device is present.

(Rev. 0)

41

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.3.5

G 100 R/F-A Receptor 1 I/O configuration worksheet

FUNCTIONS

STANDBY

PREP

GEN RDY

RAD EXP

FLUORO EXP

ON
--------OFF
----OFF
----OFF
OFF
OFF
OFF
ON
OFF
ON
-----

--------ON
----------------OFF
OFF
----OFF
ON
OFF
ON
-----

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON

***INPUTS***
REMOTE EXP
REMOTE PREP
REMOTE FL. EXP
CONSOLE EXP
CONSOLE PREP
TOMO EXP
REM. TOMO SEL
I/I SAFETY
COLL. ITLK
BUCKY CONTACTS
SPARE
THERMAL SW 1
THERMAL SW 2
DOOR ITLK
MULTI SPOT EXP

------------------------OFF
------------OFF
ON
OFF
OFF
OFF

OFF
ON
----OFF
OFF
--------OFF
OFF
----OFF
ON
OFF
ON
-----

-------------------------------------------------------------

***OUTPUTS***
BKY 1 SELECT
BKY 2 SELECT
BKY 3 SELECT
TOMO/BKY 4 SEL
TOMO/BKY START
ALE
COLL. BYPASS
ROOM LIGHT
SPARE

APOLLO EZ

OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON

42

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.3.6

G 100 R/F-A Receptor 2 I/O configuration worksheet

FUNCTIONS

STANDBY

PREP

GEN RDY

RAD EXP

FLUORO EXP

OFF
--------ON
----OFF
----OFF
OFF
OFF
OFF
ON
OFF
ON
-----

--------OFF
----------------OFF
OFF
----OFF
ON
OFF
OFF
-----

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

***INPUTS***
REMOTE EXP
REMOTE PREP
REMOTE FL. EXP
CONSOLE EXP
CONSOLE PREP
TOMO EXP
REM. TOMO SEL
I/I SAFETY
COLL. ITLK
BUCKY CONTACTS
SPARE
THERMAL SW 1
THERMAL SW 2
DOOR ITLK
MULTI SPOT EXP

------------------------OFF
------------OFF
ON
OFF
OFF
OFF

OFF
OFF
----OFF
ON
--------OFF
OFF
----OFF
ON
OFF
ON
-----

-------------------------------------------------------------

***OUTPUTS***
BKY 1 SELECT
BKY 2 SELECT
BKY 3 SELECT
TOMO/BKY 4 SEL
TOMO/BKY START
ALE
COLL. BYPASS
ROOM LIGHT
SPARE

(Rev. 0)

OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

43

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / CCD N33 /
G 100 R/F-A / Chest Stand

3.3.7

G 100 R/F-A Receptor 3 I/O configuration worksheet

FUNCTIONS

STANDBY

PREP

GEN RDY

RAD EXP

FLUORO EXP

OFF
--------ON
----OFF
----OFF
OFF
ON
OFF
ON
OFF
ON
-----

--------OFF
----------------OFF
OFF
----OFF
ON
OFF
ON
-----

OFF
OFF
ON
OFF
OFF
OFF
OFF
ON
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

***INPUTS***
REMOTE EXP
REMOTE PREP
REMOTE FL. EXP
CONSOLE EXP
CONSOLE PREP
TOMO EXP
REM. TOMO SEL
I/I SAFETY
COLL. ITLK
BUCKY CONTACTS
SPARE
THERMAL SW 1
THERMAL SW 2
DOOR ITLK
MULTI SPOT EXP

------------------------OFF
------------OFF
ON
OFF
OFF
OFF

OFF
OFF
----OFF
ON
--------OFF
OFF
----OFF
ON
OFF
ON
-----

-------------------------------------------------------------

***OUTPUTS***
BKY 1 SELECT
BKY 2 SELECT
BKY 3 SELECT
TOMO/BKY 4 SEL
TOMO/BKY START
ALE
COLL. BYPASS
ROOM LIGHT
SPARE

OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

NOTE:
The configuration of the Receptor 3 I/O has to be made only if the Chest
Stand with Potter Bucky is present.

APOLLO EZ

44

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

CONFIGURATION WITH APOLLO EZ/


DIVA-D / G100R/F-A / CHEST STAND VD3
Room configuration

Apollo EZ

I.A.E. X-ray tube type

Three field solid state AEC

Diva-D.

G 100 R/F-A (T.D.I. Interfaced)

Chest Stand VD3 35x43 with motorised foot rest (See Note)

Three field solid state AEC.

Receptor configuration
Receptor 1: Apollo EZ Spot Film Device Tube 1
Receptor 2: Digital Video Acquisition System Tube 1
Receptor 3: Chest Stand VD3 35x43 Tube 1 (See Note)
NOTE:
When the Generator is supplied with "Falling load" feature, the GalKey
U29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the
"SERIAL" exposure is enable (paragraph 4.3.4 - Image Receptor
Programming Worksheet: "SERIAL=YES"); whenever the Falling Load has
to be activated on the Spot Film Device receptor, the SERIAL parameter
has to be set to "NO".
In case Falling Load feature and SERIAL exposure are required will be
necessary to create for each a dedicated receptor.
NOTE:
Reference to the Chest Stand VD3 35x43 with motorized foot rest is made
in order to allows the installer to record the necessary information for the
hardware interface among G 100 R/F-A and Potter Bucky.

(Rev. 0)

45

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.1

Power supplies connection instruction

Main
Power Supply

Generator
Terminals

G 100 R/F
ROOM INTERFACE BOARD
TB7.1 (0 Vac)
* INSTALLATION KIT
230Vac AUXILIARY CONTACTOR
V.S.M. cod.4291515200
G 100 R/F
ROOM INTERFACE BOARD
TB10.1 (220 Vac)

R S T

(The auxiliary contactor is already mounted


inside the G 100 R/F-A Generator)

APOLLO EZ
Terminals R-S-T

R S

CHEST STAND
Terminal X1 (**)

(*) If this remote controlled switch is not available use a remote controlled
switch with the following characteristics:

Coil:
220-230 Vac 50 Hz
230-240 Vac 60 Hz

Contact: 4 N.O. contacts, 16 A.


(**) Made this connection only when a motorised foot rest chest stand is present.

APOLLO EZ

46

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.2

Hardware interface instruction

4.2.1

APOLLO EZ G 100 R/F-A interface

APOLLO EZ
Description

Terminal

Description

X27.15

TB6.9

Remote Exposure
(Room Interface Board)

X27.3

TB6.7

Preparation radiography
(Room Interface Board)

Common
0Vdc

X0.20

J6.2

0V
(Table Digital Interface board)

Common 0VS

X27.23

J6.23

0V analog
(Table Digital Interface board)

Manual kV
ref.

X27.11

J6.7

Manual kV fluoro input


(Table Digital Interface board)

Fluoro kV
manual/auto
selection

X27.20

J6.1

Man/Auto fluoro
(Table Digital Interface board)

Tomo
speed 1

X27.4

Tomo
speed 2

X27.1

J6.3

Tomo speed 1
(Table Digital Interface board)

Tomo
speed 3

X27.14

Tomo
speed 4

X27.17

J6.4

Tomo speed 2
(Table Digital Interface board)

Tomo 7

X27.5

J7.1

Tomo 7
(Table Digital Interface board)

Tomo 20

X27.18

J7.3

Tomo 20
(Table Digital Interface board)

Tomo 30

X27.6

J7.4

Tomo 30
(Table Digital Interface board)

(Rev. 0)

Terminal

G 100 R/F-A
Wire n

47

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

APOLLO EZ
Description

Terminal

Tomo 45

G 100 R/F-A
Wire n

Terminal

Description

X27.19

J7.5

Tomo 45
(Table Digital Interface board)

Alarm
X-ray ON

X27.7
(see Note)

J7.6

X-ray OFF
(Table Digital Interface board)

SFD selection

X0.15

TB2.12

Bucky 1 select
(Room Interface Board)

DSI selection

X0.14

TB2.1

Bucky 2 select
(Room Interface Board)

X-ray ON
from
generator

X0.13

J6.21

X-Ray on
(Table Digital Interface board)

Generator
ready

X0.12

J6.22

Generator Ready
(Table Digital Interface board)

Safety switch

X0.9

J17.1

Safety switch
(Generator Interface PWB)

Safety switch

X0.11

J17.2

Safety switch
(Generator Interface PWB)

NOTE:
During the auto calibration of the X-ray tube it is required to make the
jumper between J7.6 and J6.2 (force terminal J7.6 at 0Vdc).
It is important to remove the jumper at the end of the X-ray tube
auto calibration.

APOLLO EZ

48

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

APOLLO EZ DIVA-D interface

4.2.2

APOLLO EZ
Description

Terminal

DIVA-D
Wire n

POD pcb

Description

X28.18

20

Fluoroscopy request

X28.6

15

Generator ready

X28.19

21

Exposure request

Common
0Vdc

X28.22

11

0 V digital common

Common

X28.3

12

0 V digital common

X28.4

18

Vertical image revers

X28.16

19

Horizontal image revers

X0.13

14

X-ray ON

1
3
13

X-ray command common


Fluoroscopy command common
0 V digital common

D.S.I.

X-ray ON
from
generator

(Rev. 0)

49

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

G 100 R/F-A DIVA-D interface

4.2.3

G 100 R/F-A

DIVA-D
POD
pcb

Description

J7.22

Pulse fluoroscopy comm.

Preparation radiography

J6.24

Preparation command

Start fluoroscopy

J7.23

Fluoroscopy command

X-ray request

J7.24

X-ray command

Description

TDI

Pulses fluoroscopy

Wire n

G 100 R/F-A DIVA-D serial connection

4.2.4

G 100 R/F-A

DIVA-D

Description

Plug

Serial connection
(Generator CPU board)

J15

APOLLO EZ

Wire n

50

Plug

Description

RS232

Serial connection
(Rack)

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

APOLLO EZ I.I. component plate interface

4.2.5

APOLLO EZ

I.I. component plate

Description

Terminal

Common

Wire n

Pcb A1

Description

4 fields

3 fields

X28.1

X4.4

X4.4

Common

Medium field

X28.2

X4.2

X4.1

Medium zoom

Minimum field

X28.14

X4.3

X4.2

Maximum
zoom

Maximum field

X28.15

X4.1

N.C.

Minimum
zoom

Power supply

X0.2

Power supply

Power supply

S3

X0.1

Power supply

X0.3

Ground

Ground

(Rev. 0)

51

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

DIVA-D I.I. component plate interface

4.2.6

DIVA-D

I.I. component plate

Description

POD Pcb

Analog intensifier
field selection
0 V analog

Wire n

Pcb A1

Description

25

X5.1

V Ref. zoom I.I.


(6 - 12 V)

29

X5.2

0 V analog

Monitor interface

4.2.7

I.I. component plate


Description

Pcb A2

Monitor L.I.H.
visualization
0 Vs common

APOLLO EZ

Table and wall version


Wire n

Plug

Description

X4.2

LI

Monitor L.I.H.
visualization

X4.10

0V

0 Vs common

52

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.2.8

G 100 R/F-A Three field solid state AEC device 1


NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.

G100 R/F-A
AEC board
(solid state chamber)
type 737992

AEC device 1
connecting cable

Description Terminal

Wire

Description

Anode (L)

J11.1

Pink

Anode Left

Cath (L)

J11.2

Grey

Cathode Left

Anode (M)

J11.3

White

Anode Middle

Cath (M)

J11.4

Brown

Cathode Middle

Anode (R)

J11.5

Green

Anode Right

Cath (R)

J11.6

Yellow

Cathode Right

Ground

J11.7

Shield

Ground

(Rev. 0)

53

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

G 100 R/F-A - X-ray tube interface

4.2.9

G 100 R/F-A

X-ray tube

Description

Terminal

Main

Terminal

Description

Tube 1

Principal

Shift

Tube 1

Auxiliary

Common

Tube 1

Common

Ground

Tube 1

Ground

Thermal switch

Tube 1

Thermal switch

Thermal switch

Tube 1

Thermal switch

APOLLO EZ

Wire n

54

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.2.10

Chest Stand VD3 35x43 G 100 R/F-A interface


NOTE:
This hardware interface has to be made only if the Potter Bucky is
present.

Chest stand VD3 35x43


Description

Terminal

X-ray request

G 100 R/F-A
Wire n

Terminal

Description

X1.B1

TB1.11

Bucky 3 select

X-ray request

X1.B2

TB1.12

Bucky 3 select

Potter reply

X1.X

TB2.5

Bucky contacts

Potter reply

X1.X

TB2.4

Bucky contacts

WARNING:
The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the Chest
Stand VD3 35x43 Technical and User's Manual for the Bucky start
voltage setting.

(Rev. 0)

55

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.2.11

G 100 R/F-A Three field solid state AEC device 2


NOTE:
This hardware interface has to be made only if the Potter Bucky with
AEC device is present.
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.

G100 R/F-A
AEC board
(solid state chamber)
type 737992

AEC device 2
connecting cable

Description Terminal

Wire

Description

Anode (L)

J12.1

Pink

Anode Left

Cath (L)

J12.2

Grey

Cathode Left

Anode (M)

J12.3

White

Anode Middle

Cath (M)

J12.4

Brown

Cathode Middle

Anode (R)

J12.5

Green

Anode Right

Cath (R)

J12.6

Yellow

Cathode Right

Ground

J12.7

Shield

Ground

APOLLO EZ

56

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.2.12

G 100 R/F-A jumpers and wiring configuration instruction

4.2.12.1 Room interface board jumpers configuration

JW2:

JW3:
JW4:

Live contact (this jumper configuration must be made only if


the Potter Bucky is present on Receptor 3)
Live contact
Live contact.

4.2.12.2 Room interface board jumper wiring

TB8.1 TB11.5

TB4.8 TB4.9:

(This wiring must be made only if the 24Vdc


Potter Bucky is present on Receptor 3)
Thermal switch.

4.2.12.3 Generator interface PWB jumpers configuration


-

JW1:
JW2:
JW3:
JW4:
JW5:
JW6:
JW7:
JW8:
JW9:
JW10:
JW11:
JW12:

12
12/
12/
Open
Close
12/
12/
12/
12/
12/
23
Open

34
34

3
3
3
3
3

4
4
4
4
4

JW13:
JW14:
JW15:
JW19:
JW20:
JW21:
JW22:
JW24:
JW25:
JW26:
JW27:
JW28:

Open
1 2 / 3 4
12/34
23
Close
12
23
Close
Close
Close
Close
Open

NOTE:
The configuration of the Jumpers not listed in the previous tables is not
relevant for this system integration.

(Rev. 0)

57

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.2.12.4 T.D.I. board SW1 Dip-switches setting


-

SW1-1:
SW1-2:
SW1-3:
SW1-4:

OFF
OFF
ON
OFF

SW1-5:
SW1-6:
SW1-7:
SW1-8:

ON
OFF
OFF
OFF

4.2.12.5 Accessories connection

Generator

Accessory

Description

Terminal

Wire n

Description

Room door interlock


(Room Interface board)

TB4.4

Door contact

Room door interlock


(Room Interface board)

TB4.5

Door contact

Room light 0V
(Room Interface board)

TB7

Generator ON lamp
(white lamp)

Room light power supply


(Room Interface board)

(*)

Generator ON lamp
(white lamp)

Room light contact


(Room Interface board)

TB4.11

X-ray ON lamp
(red lamp)

Room light contact


(Room Interface board)

TB4.12

X-ray ON lamp
(red lamp)

(*) TB10 for 220V ac / TB9 for 110V ac / TB8 for 24V dc

NOTE:
Remove jumper JW13 on Room Inteface Board.

APOLLO EZ

58

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.3

G 100 R/F-A set-up programming instruction


After the hardware connections indicated in the previously tables it is
needed to proceed to the software configuration of the G 100 R/F-A.
Refer to the G 100 R/F-A Service Manual, chapter 3, where are described
the interfacing and the programming information to the generator. This
chapter also allows the installer to record the necessary information to
complete the installation, as well as to record the programming values.
Refer to the tables in paragraphs 4.3.2, 4.3.3, 4.3.4, 4.3.5, 4.3.6, 4.3.7 of
this Installation Guide for the programming of the generator set-up
submenu.

4.3.1

Tube selection
Prior to beginning tube calibration, the tube used in the installation must
be properly selected, and generator limits should be programmed.

4.3.1.1

X-ray tube present in library


If the x-ray tube used in the installation is present in the library of the
generator refer to the G 100 R/F-A Service Manual, chapter 3C Programming the generator in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter-Programming the dual speed starter to record the necessary
information to complete the configuration.

(Rev. 0)

59

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.3.1.2

Loading X-ray tube data from GenWare utility software


If the X-ray tube used in the installation is not present in the library of
the generator should be present in the utility software. The data
communication with a computer it is allowed using the GenWare utility
software. Refer to the G 100 R/F-A Service Manual, chapter 3C - DATA
LINK to record the necessary information.

4.3.1.3

X-ray tube not present neither in library nor in GenWare utility


software
If the x-ray tube used in the installation is not present neither in the
generator library nor in the GenWare utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the x-ray
tube that is going to be installed. An X-ray tube present in the library or
in the GenWare utility software, as similar as possible to the one is
going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the x-ray tube. Refer to
the G 100 R/F-A Service Manual, chapter 3C - Tube selection and
Generator limits in order to record all the necessary information to
program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter - Programming the dual speed starter to record the necessary
information to complete the configuration.

APOLLO EZ

60

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.3.2

AEC set-up
The installation of the AEC devices and their calibration are covered in
the G 100 R/F-A Service Manual, chapter 3E - Automatic exposure
control (AEC) calibration.

4.3.3

Fluoro set-up
NOTE:
The Fluoro set-up data must be set in according with the X-ray tube
technical specification.
The following tables meets the Fluoro set-up request in according with
most of the X-ray tube supplied by Villa Sistemi Medicali; during the
installation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in the
G 100 R/F-A Service Manual, chapter 3 - Automatic brightness
stabilisation (ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP

FL TIMER:
5 Min
MIN FLUORO KV:
40
MAX FLUORO KV:
120
II MODES:
0
FL ABS:
ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES
SID COMP
FL SETUP

Not Used
Not Used
FL ABS SETUP

ABS CURVE DEF:

(Rev. 0)

61

ABS LEVEL 0:
ABS LEVEL 1:
ABS LEVEL 2:
ABS LEVEL 3:
SIGNAL GAIN
ABS GAIN
ABS DEADBAND
ABD DELAY
ABS START KV

5.00 V
5.00 V
5.00 V
5.00 V
10
10
3%
150 mS
75

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

FL ABS CURVE

FL DOSE LIMIT

Note: The fluoro dose limit is set at 5 mA in order to


meet the code of federal regulations 21 paragraph
1020. If it is not required to meet these
regulations a higher value can be set in according
with the loading factors of the x-ray tube.

HLF SETUP

APOLLO EZ

FL II COMP
Not Used

62

FL ABS CURVE 1:
ON
40KV:
0.5 mA
50 KV :
1.0 mA
60 KV :
2.0 mA
70 KV :
3.0 mA
80 KV :
4.0 mA
90 KV :
5.0 mA
100 KV :
6.0 mA
110 KV :
7.0 mA
120 KV:
8.0 mA
8.0mA
125 KV :
* FL MAN DL *
40KV:
0.5 mA
50 KV:
1.0 mA
60 KV:
1.6 mA
70 KV:
2.1 mA
80 KV:
2.7 mA
90 KV:
3.2 mA
100 KV:
3.8 mA
110 KV:
4.4 mA
120KV:
5.0 mA
5.0mA
125 KV:
* FL ABS DL *
40KV:

5.0 mA

50 KV:

5.0 mA

60 KV:

5.0 mA

70 KV:

5.0 mA

80 KV:

5.0 mA

90 KV:

5.0 mA

100 KV:

5.0 mA

110 KV:

5.0 mA

120KV:

5.0 mA

125 KV:

5.0 mA

Not Used

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

PF SETUP

PF:
PF DEFAULT:
PPS DEFAULT:
PF DEFAULT MA:

ON
NONE
NONE
5=40mA
6=60mA
7=80mA
8=99mA

Note: Four options for the settings of the Pulsed Fluoro default mA are possible. It is
suggested to set the value 5 = 40mA as default parameter. The others settings are
possible on line by means the +/- mA keys. If, once set at 40 mA the default mA value
for the pulsed fluoro, is necessary to reduce the dose is possible to reduce the length
of the pulses by means the settings of the jumpers on the A2 board of the I.I. system
(make reference to the I.I. system service manual)
PF DEFAULT Ms:
PF ABS SETUP

PF ABS CURVE DEF: NONE


PF ABS CURVES
PF DOSE LIMITS
PF II COMP
PF PPS SETUP

HPF SETUP
HLF SELECT:
CONSOLE
PF SELECT:
CONSOLE

(Rev. 0)

63

PF SYNC FREQ:
ABS LEVEL 0:
ABS LEVEL 1:
ABS LEVEL 2:
ABS LEVEL 3:
SIGNAL GAIN:
ABS GAIN:
ABS DEADBAND:
ABS DELAY:
ABS START KV:
PF INTEGRATOR:
1PPS COMP:
3PPS COMP:
6/7.5PPS COMP:
12/15PPS COMP:
25/30PPS COMP:
Not Used
Not Used
Not Used
1 PPS:
3 PPS:
6/7.5 PPS:
12.5/15 PPS:
28/30 PPS:
Not Enable

Not
Used
50 HZ
5.0 V
5.0 V
5.0 V
5.0 V
10
8
10%
150 mS
75
OFF
0%
0%
0%
0%
0%

ON
ON
ON
ON
ON

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.3.4

G 100 R/F-A Image receptor programming worksheet

IMAGE RECEPTOR PROGRAMMING WORKSHEET


DATE:
SERIAL#:
FUNCTION

RECEPTOR
1

RECEPTOR
2

RECEPTOR
3

RECEPTOR
4

RECEPTOR
5

RECEPTOR
6

RECEPTOR
NAME
TUBE

NONE

NONE

NONE

TOMO

NO

NO

NO

NO

NO

NO

FLUORO

YES

YES

NO

NO

NO

NO

SERIAL

YES

YES

NO

NO

NO

NO

INTERFACE
OPTS

20

21

FUNCTIONAL
OPTS

FLUORO
HANG

30 SEC

30 SEC

30 SEC

30 SEC

30 SEC

30 SEC

RAD HANG

0 SEC

0 SEC

0 SEC

0 SEC

0 SEC

0 SEC

LAST IMAGE
HOLD

40 Ms

40 Ms

40 Ms

40 Ms

40 Ms

40 Ms

MEMORY

NO

NO

NO

NO

NO

NO

REM TOMO
BUT

2500 Ms

2500 Ms

2500 Ms

2500 Ms

2500 Ms

2500 Ms

SF/LF
SWITCH

AUTO

AUTO

AUTO

AUTO

AUTO

AUTO

AEC BACKUP

FIXED

FIXED

FIXED

FIXED

FIXED

FIXED

AEC BACKUP
MAS

500

500

500

500

500

500

AEC BACKUP
MS

3200

3200

3200

3200

3200

3200

AEC
CHANNEL

AEC LOCK
EXPS

RECEPTOR
SYM

NOTE:
Receptor 3 has to be programmed only if Chest Stand with Potter Bucky
and AEC device is present.

APOLLO EZ

64

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.3.5

G 100 R/F-A Receptor 1 I/O configuration worksheet

FUNCTIONS

STANDBY

PREP

GEN RDY

RAD EXP

FLUORO EXP

ON
--------OFF
----OFF
----OFF
OFF
OFF
OFF
ON
OFF
ON
-----

--------ON
----------------OFF
OFF
----OFF
ON
OFF
ON
-----

ON
OFF
OFF
OFF
OFF
ON
OFF
ON
OFF

ON
OFF
OFF
OFF
OFF
ON
OFF
ON
OFF

***INPUTS***
REMOTE EXP
REMOTE PREP
REMOTE FL. EXP
CONSOLE EXP
CONSOLE PREP
TOMO EXP
REM. TOMO SEL
I/I SAFETY
COLL. ITLK
BUCKY CONTACTS
SPARE
THERMAL SW 1
THERMAL SW 2
DOOR ITLK
MULTI SPOT EXP

------------------------OFF
------------OFF
ON
OFF
OFF
OFF

OFF
ON
----OFF
OFF
--------OFF
OFF
----OFF
ON
OFF
ON
-----

-------------------------------------------------------------

***OUTPUTS***
BKY 1 SELECT
BKY 2 SELECT
BKY 3 SELECT
TOMO/BKY 4 SEL
TOMO/BKY START
ALE
COLL. BYPASS
ROOM LIGHT
SPARE

(Rev. 0)

ON
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF

ON
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

ON
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

65

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.3.6

G 100 R/F-A Receptor 2 I/O configuration worksheet

FUNCTIONS

STANDBY

PREP

GEN RDY

RAD EXP

FLUORO EXP

ON
--------OFF
----OFF
----OFF
OFF
OFF
OFF
ON
OFF
ON
-----

--------ON
----------------OFF
OFF
----OFF
ON
OFF
ON
-----

OFF
ON
OFF
OFF
OFF
ON
OFF
ON
OFF

OFF
ON
OFF
OFF
OFF
OFF
OFF
ON
OFF

***INPUTS***
REMOTE EXP
REMOTE PREP
REMOTE FL. EXP
CONSOLE EXP
CONSOLE PREP
TOMO EXP
REM. TOMO SEL
I/I SAFETY
COLL. ITLK
BUCKY CONTACTS
SPARE
THERMAL SW 1
THERMAL SW 2
DOOR ITLK
MULTI SPOT EXP

------------------------OFF
------------OFF
ON
OFF
OFF
OFF

OFF
ON
----OFF
OFF
--------OFF
OFF
----OFF
ON
OFF
ON
-----

-------------------------------------------------------------

***OUTPUTS***
BKY 1 SELECT
BKY 2 SELECT
BKY 3 SELECT
TOMO/BKY 4 SEL
TOMO/BKY START
ALE
COLL. BYPASS
ROOM LIGHT
SPARE

APOLLO EZ

OFF
ON
OFF
OFF
OFF
OFF
OFF
OFF
OFF

OFF
ON
OFF
OFF
OFF
OFF
OFF
ON
OFF

OFF
ON
OFF
OFF
OFF
OFF
OFF
ON
OFF

66

(Rev. 0)

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

4.3.7

G 100 R/F-A Receptor 3 I/O configuration worksheet

FUNCTIONS

STANDBY

PREP

GEN RDY

RAD EXP

FLUORO EXP

OFF
--------ON
----OFF
----OFF
OFF
ON
OFF
ON
OFF
ON
-----

--------OFF
----------------OFF
OFF
----OFF
ON
OFF
ON
-----

OFF
OFF
ON
OFF
OFF
OFF
OFF
ON
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

***INPUTS***
REMOTE EXP
REMOTE PREP
REMOTE FL. EXP
CONSOLE EXP
CONSOLE PREP
TOMO EXP
REM. TOMO SEL
I/I SAFETY
COLL. ITLK
BUCKY CONTACTS
SPARE
THERMAL SW 1
THERMAL SW 2
DOOR ITLK
MULTI SPOT EXP

------------------------OFF
------------OFF
ON
OFF
OFF
OFF

OFF
OFF
----OFF
ON
--------OFF
OFF
----OFF
ON
OFF
ON
-----

-------------------------------------------------------------

***OUTPUTS***
BKY 1 SELECT
BKY 2 SELECT
BKY 3 SELECT
TOMO/BKY 4 SEL
TOMO/BKY START
ALE
COLL. BYPASS
ROOM LIGHT
SPARE

OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF

NOTE:
The configuration of the Receptor 3 I/O has to be made only if the Chest
Stand with Potter Bucky is present.

(Rev. 0)

67

APOLLO EZ

SYSTEM INTEGRATION GUIDE


Configuration with Apollo EZ / Diva-D /
G 100 R/F-A / Chest Stand

THIS PAGE IS INTENTIONALLY LEFT BLANK

APOLLO EZ

68

(Rev. 0)

Cod. 6998312603_Rev.0

VILLA SISTEMI MEDICALI S.p.A.


Via Delle Azalee, 3
20090 Buccinasco (MI) ITALY
Tel. (+39) 02 48859.1
Fax (+39) 02 4881844

Das könnte Ihnen auch gefallen