Beruflich Dokumente
Kultur Dokumente
ii
iii
iv
Table of Contents
Introduction --------------------------------------------------------------------------------
Methodology -------------------------------------------------------------------------------
Statements
Performance of the Test -------------------------------------------------------------
10
11
11
12
13
References ----------------------------------------------------------------------------------
13
Appendix -------------------------------------------------------------------------------------
14
vi
Perfect consensus is defined as 100% of the voters agreeing with the statement and
none of the voters chose option D.
Very good consensus is defined as 80 to 99% of the voters agreeing with the
statement and there are less than 20% of the voters who chose option D.
Good consensus is defined as either
80 to 99% of the voters agreeing with the statement with more than 20% of
the voters choosing option D
60 to 79% of the voters agreeing with the statement with less than 20% of the
voters choosing option D
Some consensus is defined as 51 to 59% of the voters agreeing with the statement
No consensus is defined as less than 50% of voters agreeing with the statement
The computation of the parameters needed for the level of consensus classification can
be seen in appendix N.
Based on the level of consensus, the statements were modified based on the modal
auxiliary verbs: must, should and may. Statements classified as having either a perfect or a
very good consensus used the modal auxiliary verb must. Must is used in the context that
personnel involved in spirometry testing has a responsibility or an obligation to follow the
statement. Statements classified as having a good consensus used the modal auxiliary verb
should. Should is used in the context that personnel involved in spirometry testing are
advised to follow the statement.
There are 22, 25 & 4 statements with perfect, very good and good consensus
respectively. None of the statements were classified as some or no consensus.
Quality Control
Statements:
(S2-1) Daily instrument calibration must be done (very good consensus).
(S2-2) The interpreter (lab medical director or pulmonary consultant on deck) must
give a feedback to the technician with regards to the quality of the test (perfect
consensus).
(S2-3) A log book must be maintained. It must contain the following (very good
consensus):
1. Daily calibration results
2. Anomalous events or problems encountered with the system
3.Feedback of the interpreter to the technician
Parameters Measured
Statements:
(S3-1) The following must be measured and reported: FEV1/ FVC, FVC and FEV1
(perfect consensus).
(S3-2) If the spirometry machine is capable, the following should be measured and
reported (in order of decreasing importance) (very good consensus):
A. PEFR
B. FEF 25-75
C. FEF 75
D. MIF50/ MEF50
E. FEV1/FV6
F. MVV
G. MIF 50
H. MEF 50
I. FEV6
J. FEF 25
K. FEF 50
Test Result Selection
Statement:
(S4-1) Test result selection must be based on the recommendations set by the ATSERS-TFLFT1 (perfect consensus).
1. Report the highest FVC and FEV1 obtained.
2. The spirogram with the highest sum of FEV1 & FVC is chosen as the best trial.
3. The other parameters reported are obtained from the best trial.
Parts of the Report
Statements:
(S5-1) In the spirometry report, there must be 3 sections for the interpretation of the
results (perfect consensus):
A. Technicians Comments
B. Interpreters comments
C. Final Interpretation
(S5-2) Clinical correlation should not be incorporated in the final interpretation (good
consensus).
General Data
Statement:
(S6-1) The spirometry report must contain a section on general data which contains the
following information (very good consensus):
A. Name of patient
B. Age of patient
C. Gender of patient
D. Measured height
E. Measured weight
F. Race of patient
G. Date of procedure
H. Time of procedure
I. Room temperature at time of testing
J. Room atmospheric pressure at time of testing
K. Name of referring physician
L. Name of the laboratory
M. Phone number of laboratory
For example:
Technicians Comments:
A spirometry test was done to confirm the diagnosis of COPD.
Patient has no history of previous spirometry testing.
The patient is a previous smoker who has a 20-pack year of smoking history.
His last intake of cigarette smoke was 10 years ago.
There were no conditions present in the check list which may alter the results
of this study.
The patient took his Indacaterol 16 hours prior to test
The test done satisfied the reproducibility criteria set by the ATS-ERS.
Semantics in Interpretation
Statement:
(S8-1) In the interpretation of the results, the report must use the phrase ventilatory
defect (not lung defect or ventilatory pattern) (very good consensus).
Interpretation Algorithm
Statement:
(S9-1) Spirometry results must be interpreted based on the ATS-ERS-TFLFT2
suggested algorithm.
Fig. 1 Interpretative strategies for lung function tests: R. Pellegrino, G. Viegi, V. Brusasco, et
al. Eur Respi Respi J 2005; 26: 948-968.
of the flow volume loop must be mentioned in the interpreters comments (very good
consensus).
Example:
Interpreters Comments:
1. The spirometry test is of good quality based on standards set by the ATSERS.
2. The FEV1/FVC is normal (using the 5th percentile of the predicted as the
lower limit of normality).
3. The FVC is normal (using the 5th percentile of the predicted as the lower
limit of normality).
4. Flow volume shows a plateau shape of the expiratory limb which is
reproducible in 3 acceptable trials. The MIF50/MEF50 is greater than one.
These are suggestive of a variable intra thoracic upper airway obstruction.
5. There is no significant response to bronchodilator.
6. The normality of the parameters is based on the Morris Polgar equation
Final Interpretation:
Probable variable intra-thoracic upper airway obstruction
Please correlate results with patients clinical data.
(S10-3) Slowing down of the terminal portion of spirogram must be described as
exaggerated concavity in the distal portion of the expiratory limb of the flow volume
loop. Use of the term scooping is discouraged (perfect consensus).
(S10-4) In cases where in the interpretation uses any of the following: flow volume
loops configuration, FEF25-75 and other spirometry parameters, it must be stated that
these parameters are suggestive but not definitive. Use of the term compatible is
discouraged (very good consensus).
Example:
Interpreters Comments:
1. The spirometry test is of good quality based on standards set by the ATSERS.
2. The FEV1/FVC is normal (using the 5th percentile of the predicted as the
lower limit of normality).
3. The FVC is normal (using the 5th percentile of the predicted as the lower
limit of normality).
4. There is (exaggerated) concavity of the flow volume loop. The FEF 25-75
and FEF 75 are low (using 65% of the predicted as the lower limit of
normality). These are suggestive of obstructive (small airways disease)
ventilatory defect.
5. The severity of the probable obstructive ventilatory defect is mild, based
on the post bronchodilator FEV1 % predicted of 80.
6. There is no significant response to bronchodilator.
7. The normality of the parameters is based on the Morris Polgar equation
Final Interpretation:
Probable mild obstructive ventilatory defect
Please correlate results with patients clinical data.
Grading of Severity
Statements:
(S11-1) Grading of severity must be based on the criteria set by ATS-ERS-TFLFT2
(perfect consensus).
(S11-2) FEV1% predicted (post bronchodilator value if available) must be used both for
obstructive and restrictive ventilatory defect (perfect consensus).
Table 1: Grading of Severity of Ventilatory Defect
Degree of Severity
Mild
70 & above
Moderate
60-69
35-49
Very Severe
<35
Example 1:
Interpreters Comments:
1. The spirometry test is of good quality based on standards set by the ATSERS.
2. Based on a low FEV1/FVC (using the 5th percentile of the predicted as the
lower limit of normality), there is an obstructive ventilatory defect.
3. The FVC is normal (using the 5th percentile of the predicted as the lower limit
of normality).
4. The severity of the obstructive ventilatory defect is moderately severe, based
on the post bronchodilator FEV1 % predicted of 50.
5. There is no significant response to bronchodilator.
6. The normality of the parameters is based on the Morris Polgar equation
Final Interpretation:
Moderately severe obstructive ventilatory defect with no significant response
to Bronchodilator
Please correlate results with patients clinical data.
Example 2:
Interpreters Comments:
1. The spirometry test is of good quality based on standards set by the ATSERS.
2. The FEV1/FVC is normal (using a fixed cut-off value of 0.70 as the lower
limit of normality).
3. The FVC is low (using the 80% of the predicted as the lower limit of
normality) which is probably due to a restrictive ventilatory defect or
obstructive ventilatory defect (with residual volume hyperinflation).
4. The severity of the restrictive ventilatory defect is mild, based on the post
bronchodilator FEV1 % predicted of 70.
5. There is no significant response to bronchodilator.
6. The normality of the parameters is based on the Morris Polgar equation
Final Interpretation:
Probable mild restrictive ventilatory defect with no significant response to
bronchodilator.
We suggest requesting for a lung volume study to confirm presence of a
restrictive ventilatory defect and rule out obstructive ventilatory defect.
Please correlate results with patients clinical data.
Response to Bronchodilator
Statements:
(S13-1) Performance of a post-bronchodilator study must be done based on the
recommendations of ATS-ERS-TFLFT2 (perfect consensus).
Procedure for Performance of Post-bronchodilator Study
1. Assess lung function at baseline
2. Administer salbutamol in four separate doses of 100 mg through a spacer
3. Re-assess lung function after 15 minutes. If you want to assess the
potential benefits of a different bronchodilator, use the same dose and the
same route as used in clinical practice. The wait time may be increased
for some bronchodilators
(S13-2) Interpretation of response to bronchodilator must be based on the criteria set
by the ATS-ERS-TFLFT2 (perfect consensus).
Criteria for Significant Response to Bronchodilator:
An increase in FEV1 or FVC > 12% of control and > 200 mL.
(S13-3) FEF 25-75 must not be used in the interpretation of the post bronchodilator
response (perfect consensus).
(S13-4) To standardize the semantics, response to bronchodilator must be labeled as
significant or no significant response to bronchodilator. The use of other terms such
as poor or good response to bronchodilator are discouraged (perfect consensus).
Statement:
(S14-1) Interpretation must include suggestion/s on need for an additional test
if warranted (very good consensus).
Reference Equations
Statements:
(S15-1) In the section on Interpreters Comments the report must state what
reference equation was used in the interpretation of the results. Statement on the need
to correlate clinically should be written in the section on Final Interpretation (very
good consensus).
Example:
Interpreters Comments:
1. The spirometry test is of good quality based on standards set by the ATSERS.
2. Based on a low FEV1/FVC (using the 5th percentile of the predicted as the
lower limit of normality), there is an obstructive ventilatory defect.
3. The FVC is normal (using the 5th percentile of the predicted as the lower
limit of normality).
4. The severity of the obstructive ventilatory defect is moderately severe,
based on the post bronchodilator FEV1 % predicted of 50.
5. There is no significant response to bronchodilator.
6. The normality of the parameters is based on the Morris Polgar equation
Final Interpretation
Moderately severe obstructive ventilatory defect with no significant response
to bronchodilator
Please correlate results with patients clinical data.
(S15-2) A research study must be conducted by the Council of Diagnostics &
Therapeutics to determine which reference equation is best fitted for the local Filipino
population (very good consensus).
Statements:
(S16-1) In the reports section on Interpreters Comments, it must be stated if the test
satisfies the reproducibility criteria set by the ATS-ERS-TFLFT1 or not (very good
consensus).
(S16-2) If the test does not satisfy one of the acceptability or reproducibility criteria, it
must state the consequences of this deficiency in the parameters (very good
consensus).
Example1.
Interpreters Comments:
1. The spirometry test is of good quality based on standards set by the ATSERS.
2. Based on a low FEV1/FVC (using the 5th percentile of the predicted as the
lower limit of normality), there is an obstructive ventilatory defect.
3. The FVC is normal (using the 5th percentile of the predicted as the lower
limit of normality).
4. The severity of the obstructive ventilatory defect is moderately severe,
based on the post bronchodilator FEV1 % predicted of 50.
5. There is no significant response to bronchodilator.
6. The normality of the parameters is based on the Morris Polgar equation
Example 2:
Interpreters Comments:
1. Despite several attempts, the end of test criteria was not met. This may
underestimate the true FVC of the patient. The FEV1 is however
reproducible and therefore the patients FEV1/FVC may overestimate the
patients real FEV1/FVC
2. Based on a low FEV1/FVC (using the 5th percentile of the predicted as the
lower limit of normality), there is an obstructive ventilatory defect.
3. The FVC is low (using the 5th percentile of the predicted as the lower limit
of normality). This may be due to poor effort of the patient, obstructive
ventilatory (with residual volume hyperinflation) or a concomitant
restrictive ventilatory defect.
4. The severity of the obstructive ventilatory defect is moderately severe,
based on the post bronchodilator FEV1 % predicted of 50.
5. There is no significant response to bronchodilator.
6. The normality of the parameters is based on the Morris Polgar equation
Final Interpretation:
Moderately severe obstructive ventilatory defect with no significant
response to bronchodilator
Please correlate results with patients clinical data.
10
Example:
Interpreters Comments:
1. The spirometry test is of good quality based on standards set by the ATSERS.
2. The FEV1/FVC is normal (using the 5th percentile of the predicted as the
lower limit of normality).
3. The FVC is normal (using the 5th percentile of the predicted as the lower limit
of normality).
4. Flow volume shows a plateau shape of the expiratory limb which is
reproducible in 3 acceptable trials. The MIF50/MEF50 is greater than one.
These are suggestive of a variable intra thoracic upper airway obstruction.
5. There is no significant response to bronchodilator.
6. The normality of the parameters is based on the Morris Polgar equation
Final Interpretation:
Probable variable intra-thoracic upper airway obstruction
Please correlate results with patients clinical data.
Components of the Report: Results in Tabular Graph
Statements:
(S18-1) The spirometry report must contain the measured parameters tabular
graph (very good consensus).
(S18-2) The first 3 rows of the tabular graph must be allotted for the most reliable
spirometry parameter (MRSP) in the following order: {This order will facilitate the
interpretation of the results based on the proposed algorithm of ATS-ERS-TFLFT2}
(very good consensus).
FEV1/FVC
FVC
FEV1
(S18-3) The succeeding rows maybe be allotted for the other parameters in the
following order of priority (very good consensus):
A. PEFR
B. FEF 25-75
11
C.
D.
E.
F.
G.
H.
I.
J.
K.
FEF 75
MIF50/ MEF50
FEV1/FV6
MVV
MIF 50
MEF 50
FEV6
FEF 25
FEF 50
(S18-4) In the tabular graph, each parameter must have its corresponding (perfect
consensus):
A. Actual Values (pre & post if done)
B. Predicted
C. 95% Confidence Interval Limit (if the spirometry software is capable)
D. % Predicted (pre & post if done)
E. % Change from baseline of post-bronchodilator measurement (if the
spirometry software is capable)
(S18-5) In the tabular graph (if the spirometry software is capable), each parameter
below the 5th percentile should be flagged down (good consensus).
12
5.
6.
Example
Final Interpretation:
Probable mild restrictive ventilatory defect with no significant response to
bronchodilator.
We suggest requesting for a lung volume study to confirm presence of a
restrictive ventilatory defect and rule out obstructive ventilatory defect.
Please correlate results with patients clinical data.
NB: Full document and Executive Summary of the PCCP: Consensus Statement of
Spirometry Testing can be down loaded at the PCCP website
http://www.philchest.org/
References:
1. American Thoracic Society/ European Respiratory Society. Stardization of
Spirometry: M.R. Miller, J. Hankinson, V. Brusasco, F. et al. Eur Respi Resepi J
2005; 26: 319-338.
2. American Thoracic Society/ European Respiratory Society. Interpretative
strategies for lung function tests: R. Pellegrino, G. Viegi, V. Brusasco, et al. Eur
Respi Respi J 2005; 26: 948-968.
13
Appendices
14
APPENDIX A
Age: 55
Room temperature: 30
Gender: Male
Height (inch): 64
Race: Asian
Date: 11/20/09
Technicians Comments:
1. A spirometry test was done to confirm the diagnosis of COPD.
2. Patient has no history to previous spirometry testing.
3. The patient is a previous smoker who has a 20-pack year of smoking history. His
last intake of cigarette smoke was 10 years ago.
4. There were no conditions present in the check list which may alter the results of
this study.
5. The patient took his Indacaterol, 16 hours prior to test.
6. The test done satisfied the reproducibility criteria set by the ATS-ERS.
SGD Charlie Go, RT
Interpreters Comments:
1. The spirometry test is of good quality based on standards set by the ATS-ERS.
2. Based on a low FEV1/FVC (using the 5th percentile of the predicted as the lower
limit of normality), there is an obstructive ventilatory defect.
3. The FVC is normal (using the 5th percentile of the predicted as the lower limit of
normality).
4. The severity of the obstructive ventilatory defect is moderately severe, based on
the post bronchodilator FEV1 % predicted of 50.
5. There is no significant response bronchodilator response.
6. The normality of the parameters is based on the Morris Polgar equation
Final Interpretation:
o Moderately severe obstructive ventilatory defect with no significant
response to bronchodilator
o Please correlate results with patients clinical data.
Diego Tigas, MD
APPENDIX B
Institution
Chinese General Hospital
Lung Center of the Philippines
University of Perpetual-H. D.
Medical Center
Veterans Memorial Medical Center
Resource Person
Eden D. Chua, MD, FPCCP
Shirley C. Panganiban Chua, MD, FPCCP
Luisito F. Idolor, MD, FPCCP
Augusto S. Sablan, Jr. MD, FPCP
Araceli Pascual, RT
Dennis Teo, MD, FPCCP
Aileen M. David-Wang, MD, FPCCP
Rodolfo E. Tamse, MD, FPCCP
Irenee Faustina J. Casio, MD, FPCP
Myrna N. Baares,MD, FPCCP
Raphael Ryan R. Zantua, MD, FPCCP
Ma. Janeth T. Samson, MD, FPCCP
Ma. Bella R. Siasoco, MD, FPCCP
Jubert P. Benedicto, MD, FPCCP
Tim S.Trinidad, MD, FPCCP
Celeste Mae L. Campomanes, MD, FPCCP
Apolonio G. Javier, JR., MD, FPCP
FPCCP
Ricardo M. Salonga, MD, FPCCP
Jose Edzel V. Tamayo, MD, FPCCP
Mary Leslee Tabi,RT
Nio Jessielito N. Doydora, MD
Eloisa S. De Guia, MD, FPCCP
APPENDIX C
-1-
Patient preparation
A.
Patient Details
1)
2)
Name of patient
Patient Identifier (e.g. Date of birth, personal PIN)
YES
YES
NO
NO
NA
NA
3)
4)
Gender of patient
Age of patient
YES
YES
NO
NO
NA
NA
5)
6)
YES
YES
NO
NO
NA
NA
7)
8)
YES
YES
NO
NO
NA
NA
YES
YES
NO
NO
NA
NA
YES
NO
NA
YES
YES
NO
NO
NA
NA
YES
NO
NA
YES
YES
NO
NO
NA
NA
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
NA
NA
NA
YES
YES
NO
NO
NA
NA
YES
YES
NO
NO
NA
NA
YES
YES
NO
NO
NA
NA
YES
YES
NO
NO
NA
NA
Patient Considerations
# of years stopped
3) Medicines taken by patient
-
APPENDIX C
-
II. Methods
A. Procedure
Do you perform the following for recording Forced Vital Capacity:
1) Check the spirometer calibration
2) Explain the test
3) Prepare the subject
a. Ask about: Smoking history
Recent illness
Medication use
Reason for test
b. Measure weight and height without shoes
4) Instruct and demonstrate the test to the subject, to include
a. Correct posture with head slightly elevated
b. Inhale rapidly and maximally
c. Position of the mouthpiece
d. Exhale with maximal force
5) Maneuvers proper (closed-circuit method)
a. Have subject assume the correct posture
b. Attack nose clip, place mouthpiece in month and close lips around the
mouthpiece
c. Inhale completely and rapidly with a pause of <1 sec at TLC
d. Exhale maximally for <6 sec until no more air can be expelled while
maintaining an upright posture
e. Repeat instructions as necessary
f. Coach vigorously
g. Repeat for a minimum of 3 trials. No more than eight trials
h. Check for acceptability
i. Check for reproducibility
6) Perform maneuver (open circuit method)
a. Have subject assume the correct posture
b. Attack nose clip
c. Inhale completely and rapidly with a pause of <1 sect at >LC
d. Place mouthpiece in mouth and close lips around mouthpiece
e. Repeat instructions as necessary
f. Coach vigorously
g. Repeat for a minimum of 3 trials. No more than eight trials
h. Check for acceptability
i. Check for reproducibility
Do you check for the following?
* Acceptability Criteria
a. Spirograms are free from artifacts
st
Glottis closure
Early termination
Submaximal effort
Leaks
Obstructed mouthpiece
b. Test results have good starts
-2YES
YES
NO
NO
NA
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
YES
YES
NO
NO
NA
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
YES
YES
NO
NO
NA
NA
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
NA
NA
NA
APPENDIX C
Test Result Selection
Do you select
The average of FVC and FEV1 even if the dont come from the same
curve
-3YES
YES
YES
YES
NO
NO
NO
NO
NA
NA
NA
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
YES
NO
NA
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
YES
YES
NO
NO
NA
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
YES
YES
YES
YES
NO
NO
NO
NO
NA
NA
NA
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
NA
NA
NA
NA
YES
NO
NA
YES
NO
NA
APPENDIX C
-4-
Mild
Moderate
Moderately severe
Severe
Very severe
YES
NO
NA
YES
NO
NA
YES
NO
NA
YES
NO
NA
YES
NO
NA
YES
NO
NA
YES
NO
NA
FEV1 % Predicted
> 70
60 69
50 59
35 49
< 39
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
APPENDIX C
2) Do you have a manual of procedure which contain
the following:
- Calibration procedure
- Test performance procedures
- Calculations
Criteria
- Reference values source
- Action to be taken when panic values are
observed
3) Do you have a record which documents the
following?
- Schedule of instrument calibration
( daily, weekly, monthly, quarterly, others )
- Problems encountered with the system
- Correction action required
- Systems hardware and software upgrades
- Record of anomalous events involving either
patients / subjects or technicians and result of
subsequent evaluation and responses to event
- Record of continuing education of personnel
- Record of results of evaluation and feedback
provided by medical director
- Schedule and procedures for routine
maintenance of all equipment in the facility
- Record of contact phonee numbers of
appropriate maintenance personnel for each
equipment used in the facility
4) Is there a protocol and frequency for calibration?
5) Is there a protocol for testing standard subjects
monthly to supplement spirometry, LV and diffusing
capacity?
6) Is there a schedule for specific quality control
procedures for each procedure done?
7) Are tolerance limits deprived for quality control
checks for each procedure?
8) Is there a plan for corrective action when tolerance
limits are exceeded?
9) Are standard subjects tested monthly?
10) Is there evidence of review by medical director or
head of pulmonary laboratory?
11) Are manuals available for:
- Equipments
- Policies
- Procedures
- Safety
12) Are these manuals readily accessible?
-5-
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
NA
YES
NO
NA
YES
YES
YES
YES
NO
NO
NO
NO
NA
NA
NA
NA
YES
YES
NO
NO
NA
NA
YES
NO
NA
YES
NO
NA
YES
YES
NO
NO
NA
NA
YES
NO
NA
YES
NO
NA
YES
NO
NA
YES
YES
NO
NO
NA
NA
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
YES
YES
NO
NO
NA
NA
YES
YES
NO
NO
NA
NA
APPENDIX C
8) Require potentially infected patients with airborne disease be tested at
end of day or week
9) Require use of disposable mouthpiece of disinfection of tubing or any
parts that come into direct contact with subjects
10) Do you use an open circuit technique? (open circuit means no
rebreathing through mouthpiece of through breathing tube)
If answer is YES:
a) Do you change the mouthpiece between patients?
b) Do you change the breathing tube between patients?
11) Do you use the closed circuit technique (closed circuit means subject
rebreaths into the mouthpiece or through breathing tube and
spirometer)? If answer is YES
a) Do you dispose of or disinfect the mouthpiece between patients?
b) Do you dispose of or disinfect the breathing tube between patients?
12) Do you use a flow sensing system in which no breathing tube is
interposed between subject and device?
a) Is the flow sensing element and interior tubing disinfected between
patients?
b) Is inspiration from the device avoided?
13) Do you use in-line bacterial filters? If answer is yes
a) Do you calibrate machine with the filters?
b) Do you discard the filter after use on a single subject?
c) Do you still do regular cleaning and disinfection despite the filters?
-6YES
NO
NA
YES
NO
NA
YES
NO
NA
YES
YES
YES
NO
NO
NO
NA
NA
NA
YES
YES
YES
NO
NO
NO
NA
NA
NA
YES
NO
NA
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NA
NA
NA
NA
NA
NA
Comments:
_____________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____
Completed by:
Noted by:
______________________________
Name and Position
Laboratory
______________________________
Head of the Pulmonary
LCP
PHC
VMMC
UPHDMC MDH
Viasys
Vmax
Spectra
Microloop
Vmax
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
Spectra
Master
Screen
22
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
NA
YES
YES
YES
NO
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
NO
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NA
YES
YES
YES
YES
YES
NO
NO
NO
YES
YES
YES
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
NO
NA
NA
NA
NA
NA
NA
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NO
NO
NA
NO
YES
YES
NO
YES
YES
YES
YES
PCCP Council on Diagnostics: Descriptive Survey Form for Use and Reporting of Spirometry
SLMC CGH
UST
UP-PGH
Oral or facial pain exacerbated by mouthpiece
YES
NO
NO
NO
Stress incontinence
NA
NO
NO
NO
Dementia or confusional status
YES
NO
NO
NO
6) Position of the patient while performing the test
YES YES
NO
YES
7) Use of restrictive clothing during the test
YES
NO
YES
NO
8) History of doing particular activities immediately prior to test
Smoking within 1 hour
YES
NO
YES
NO
Alcohol intake within 4 hours
YES
NO
YES
NO
Performing vigourous exercise within 30 mins
YES
NO
YES
YES
Eating large meals within 2 hours
YES
NO
YES
NO
Methods
Procedure
Do you perform the following for recording FVC
1) Check the spirometer calibration
YES YES
YES
YES
2) Explain the test
YES YES
YES
YES
3) Prepare the subject
Ask about: Smoking history
YES YES
YES
YES
Recent illness
YES YES
YES
YES
Medication use
YES YES
YES
YES
Reason for test
YES YES
YES
YES
Measure weight and height without shoes
YES YES
YES
YES
4) Instruct and demonstrate the test to the subject, to include
Correct posture
YES YES
YES
Inhale rapidly and maximally
YES YES
YES
YES
Position of the mouthpiece
YES YES
YES
YES
Exhale with maximal force
YES YES
YES
YES
5) Maneuvers proper (closed-circuit method)
Have subject assume the correct posture
YES YES
YES
YES
Attach nose clip, place mouthpiece in mouth and close lips
YES YES
YES
YES
around the mouthpiece
63
Inhale completely and rapidly with a pause of <1sec at TLC
YES YES
YES
YES
64
Exhale maximally for <6secs until no more air can be expelled
YES YES
YES
YES
while maintaining an upright posture
LCP
NO
NO
NO
NO
NO
PHC
NO
NO
NO
NO
NO
VMMC
YES
YES
YES
YES
YES
UPHDMC MDH
YES
YES
YES
YES
YES
YES
YES
NO
YES
NO
NO
NO
NO
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
PCCP Council on Diagnostics: Descriptive Survey Form for Use and Reporting of Spirometry
SLMC CGH
UST
UP-PGH
Repeat instructions as necessary
YES YES
YES
YES
Coach vigorously
YES YES
YES
YES
Repeat for a minimum of 3 trials. No more than 8 trials
YES YES
YES
YES
Check for acceptability
YES YES
YES
YES
Check for reproducibility
YES YES
YES
YES
6) Perform maneuver (open circuit method)
Have subject assume the correct posture
YES YES
YES
YES
Attach nose clip
YES YES
YES
YES
Inhale completely and rapidly with a pause of <1sec at TLC
YES YES
YES
YES
Place mouthpiece in mouth and close lips around mouthpiece
YES YES
YES
YES
Repeat instructions as necessary
YES YES
YES
YES
Coach vigorously
YES YES
YES
YES
Repeat for a minimum of 3 trials. No more than 8 trials
YES YES
YES
YES
Check for acceptability
YES YES
YES
YES
Check for reproducibility
YES YES
YES
YES
Do you check the following
Acceptability Criteria
a) Spirograms are free from artifacts
Cough during 1st second of exhalation
YES YES
YES
YES
Glottis closure
YES YES
YES
YES
Early termination
YES YES
YES
YES
Submaximal effort
YES YES
YES
YES
Leaks
YES YES
YES
YES
Obstructive mouthpiece
YES YES
YES
YES
b) Test results have good starts
Extrapolated volume <5% of FVC or 0.15L, whichever is greater YES YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
LCP
YES
YES
YES
YES
YES
PHC
YES
YES
YES
YES
YES
VMMC
YES
YES
YES
YES
YES
UPHDMC MDH
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NA
NA
NA
NA
NA
NA
NA
NA
NA
NO
NO
NO
NO
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NA
YES
YES
YES
YES
NA
YES
YES
YES
YES
YES
YES
YES
YES
LCP
YES
PHC
YES
VMMC
YES
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
UPHDMC MDH
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
*
YES
NA
YES
YES
YES
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
YES
YES
YES
YES
YES
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
NA
NO
NO
YES
NO
YES
YES
NO
NO
YES
YES
NO
NO
YES
YES
NO
NO
YES
YES
NO
NO
YES
YES
YES
NO
YES
YES
NO
NO
YES
YES
YES
YES
YES
YES
NO
NO
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
NO
YES
NO
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
PHC
NO
YES
VMMC
YES
YES
UPHDMC MDH
NO
YES
YES
YES
YES
YES
NO
NO
YES
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
NA
NA
NA
NA
NA
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
ATS,
GOLD
NHANES,
0.88 OF
ABS
VALUE
ATS
Morris
Polgar
YES
NO
154 7) Is the criteria that your laboratory results interpreter uses for a
positive response to a bronchodilator as follows: an increase in
FEV1 or FVC >= 12% and 200 ml increase from baseline value
155 8) Does your laboratory test result interpreter routinely comment on
the MVV?
156 8) Does your laboratory test result interpreter routinely comment on
the MVV?
157 9) Is the criteria for the classification of ventilatory dysfunction
suggested by the ATS/ERS followed by your laboratory test
interpreter?
158 Quality Control
159 Laboratory Details
160 1) Name of laboratory
161 2) Address of laboratory
162 3) Phone number of laboratory
163 4) Name of techinican who conducted the test
164 5) Date of procedure
LCP
NO
PHC
NO
VMMC
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
FEF25- FEF 25- FEF25- FEF 2575
75
75,
75, PEF,
FEF75FEF25,
85, FEF
FEF50,
200-1200
FEF75
YES
YES
YES
YES
UPHDMC MDH
NO
NO
YES
YES
YES
YES
NO
NO
YES
YES
YES
NO
YES
YES
YES
YES
YES
NO
NO
NA
NO
NO
NA
YES
YES
YES
NO
NO
NA
NO
NO
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
PCCP Council on Diagnostics: Descriptive Survey Form for Use and Reporting of Spirometry
SLMC CGH
UST
UP-PGH
6) Time of procedure
YES YES
YES
YES
7) Room temperature at time of testing
YES YES
YES
YES
8) Room atmospheric pressure at time of testing
YES YES
YES
YES
10) Possible sequence for undertaking lung function tests
YES YES
YES
1) Who interprets the result
Pulmonary fellow
YES YES
YES
YES
Pulmonary function laboratory director
NA
NO
YES
Pulmonary consultant on deck
YES YES
YES
YES
Patient's attending pulmonary consultant
YES YES
YES
YES
PFT Techniques and Procedures
2) Do you have a manual of procedure which contain the following
176
Calibration procedure
177
Test performance procedures
178
Calculations
179
Criteria
180
Reference values source
181
Action to be taken when "panic" values are observed
182 3) Do you have a record which documents the following
183
Schedule of instrument calibration
184
Problems encountered with the system
185
Correction action required
186
Systems hardware and software upgrades
187
Record of anomalous events involving either patients/subjects or
technicians and result of subsequent evaluation and responses to
event
188
Record of continuing eduction of personnel
189
Record of results of evaluation and feedback provided by
medical director
190
Schedule and procedures for routine maintenance of all
equipment in the facility
191
Record of contact phone numbers of appropriate maintenance
of all equipment in the facility
LCP
YES
NO
NO
YES
PHC
NO
NO
NO
NO
VMMC
YES
YES
NA
YES
UPHDMC MDH
YES
YES
YES
YES
NO
YES
YES
NO
YES
NO
YES
YES
NO
YES
NO
YES
YES
YES
YES
YES
NA
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
NO
YES
YES
YES
NO
YES
NO
NO
YES
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
YES
NO
NO
YES
YES
YES
YES
YES
NO
YES
NO
YES
NA
YES
YES
YES
YES
NO
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
LCP
YES
YES
PHC
YES
NA
VMMC
YES
YES
UPHDMC MDH
NA
YES
NA
YES
YES
NO
YES
NA
YES
YES
YES
NO
NA
YES
YES
YES
NO
NA
YES
NA
YES
NO
NO
YES
YES
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NA
NO
NA
YES
YES
NO
NO
NO
NA
YES
YES
YES
NA
NA
YES
YES
YES
YES
YES
YES
NO
YES
NA
YES
YES
LCP
YES
PHC
NA
VMMC
NA
UPHDMC MDH
NA
YES
YES
YES
YES
NO
NO
YES
YES
NO
YES
NO
YES
YES
YES
YES
YES
NO
YES
YES
NO
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
NO
YES
YES
YES
YES
YES
YES
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NA
NA
YES
YES
YES
YES
YES
YES
Parameter
Do you perform the following for recording FVC
1) Check the spirometer calibration
2) Explain the test
3) Prepare the subject
Ask about: Smoking history
Recent illness
Medication use
Reason for test
Measure weight and height without shoes
4) Instruct and demonstrate the test to the subject, to include
Correct posture
Inhale rapidly and maximally
Position of the mouthpiece
Exhale with maximal force
5) Maneuvers proper (closed-circuit method)
Have subject assume the correct posture
Attach nose clip, place mouthpiece in mouth and close lips around
the mouthpiece
Inhale completely and rapidly with a pause of <1sec at TLC
Exhale maximally for <6secs until no more air can be expelled while
maintaining an upright posture
Repeat instructions as necessary
Coach vigorously
Repeat for a minimum of 3 trials. No more than 8 trials
Check for acceptability
Check for reproducibility
6) Perform maneuver (open circuit method)
Have subject assume the correct posture
Attach nose clip
Inhale completely and rapidly with a pause of <1sec at TLC
Place mouthpiece in mouth and close lips around mouthpiece
Repeat instructions as necessary
Coach vigorously
Repeat for a minimum of 3 trials. No more than 8 trials
SLMC CGH
UST
UP-PGH LCP
PHC
VMMC
UPHDMC MDH
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NA
NA
NA
NA
NA
NA
NA
NO
NO
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
Parameter
Check for acceptability
Check for reproducibility
Do you check the following
Acceptability Criteria
a) Spirograms are free from artifacts
Cough during 1st second of exhalation
Glottis closure
Early termination
Submaximal effort
Leaks
Obstructive mouthpiece
b) Test results have good starts
Extrapolated volume <5% of FVC or 0.15L, whichever is greater
c) Test shows satisfactory exhalation
Duration of >= 6 secs or a plateau in the volume-time curve or if the
subject cannot continue to exhale
Reproducibility Criteria
a) After 3 acceptable spirograms have been obtained, do you apply the
following?
The 2 largest values of FVC must be within 0.150L of each other
The 2 largest values of FEV1 must be within 0.150L of each other
b) If both criteria are met, conclude test series
c) If both criteria are not met, continue testing until
Both criteria are met OR
A total of eight tests have been performed
Subject can no longer continue
d) Save, as a minimum, 3 satisfactory maneuvers
1) Who interprets the result
Pulmonary fellow
Pulmonary consultant on deck
Pulmonary function laboratory director
Do you perform the following for recording FVC
SLMC CGH
YES YES
YES YES
UST
YES
YES
UP-PGH LCP
YES
NA
YES
NA
PHC
NO
NO
VMMC
YES
YES
UPHDMC MDH
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NA
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
*
YES
NA
YES
YES
YES
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NA
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
NO
YES
YES
YES
YES
YES
NA
YES
YES
YES
Parameter
1) Check the spirometer calibration
3) Do you have a record which documents the following
Schedule of instrument calibration
Problems encountered with the system
Record of anomalous events involving either patients/subjects or
technicians and result of subsequent evaluation and responses to event
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
UP-PGH LCP
YES
YES
PHC
YES
VMMC
YES
UPHDMC MDH
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
NO
YES
YES
YES
NO
NO
NO
YES
YES
YES
YES
YES
YES
NO
YES
NO
YES
YES
NO
NO
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
YES
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NO
NA
NA
NA
YES
YES
NO
NO
YES
NO
NO
NO
NO
YES
NO
NO
NO
93
94
95
96
97
Parameter
SLMC CGH
6) Aside from the FEV1, FVC and FEV1/FVC, does your laboratory test YES YES
results interpreter use the "other parameters" routinely?
UST
NO
UP-PGH LCP
PHC
VMMC
FEF25- FEF 25- FEF25- FEF 2575
75,
75, PEF, 75
FEF75FEF25,
85, FEF
FEF50,
200-1200
FEF75
UPHDMC MDH
NO
NO
YES
YES
YES
YES
YES
NO
NO
YES
YES
YES
YES
YES
NO
NO
YES
98
YES
99
100 6) Aside from the FEV1, FVC and FEV1/FVC, does your laboratory test YES
results interpreter use the "other parameters" routinely?
YES
YES
YES
NA
NO
NO
YES
NO
NO
NO
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
NO
YES
NO
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
101
102 In the interpretation of results are the comments on the following
included
103
Ventilatory pattern of flow volume loop
104
105 Are the following data included in your report?
106 1) Graphical representation of the best trial flow volume loop
107 2) Graphical representation of all trials flow volume loop
108
109 In the interpretation of results are the comments on the following
included
110 9) Is the criteria for the classification of ventilatory dysfunction
suggested by the ATS/ERS followed by your laboratory test interpreter?
111
YES
YES
YES
SLMC CGH
UST
PHC
VMMC
UPHDMC MDH
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
ECCS
Morris
Polgar
ATS,
GOLD
NO
NO
YES
YES
YES
YES
UP-PGH LCP
NA
NO
YES
NO
NO
NHANES, ATS
0.88 OF
ABS
VALUE
NO
NO
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
Morris
Polgar
Parameter
5) Measured height
6) Estimated Height of a patient with limb deformity
7) Measured weight
8) Race of patient
9) Name of referring physician
10) Vitas signs of patient (BP, CR, RR) prior to test
Patient Considerations
1) Symptoms of the patient
2) Smoking history
Duration of smoking (# of pack years)
Smoking status (current or former)
# of years stopped
3) Medicines taken by patient
Short acting B agonist
Inhaled long acting B agonist
Oral bronchodilator
Inhaled steroids
Oral corticosteroids
Antileukotrienes
Anti-IgE antagonist
4) Time the above medications were last taken
5) History of the following particular symptoms
Chest or abdominal pain of any cause
Oral or facial pain exacerbated by mouthpiece
Stress incontinence
Dementia or confusional status
6) Position of the patient while performing the test
7) Use of restrictive clothing during the test
8) History of doing particular activities immediately prior to test
Smoking within 1 hour
Alcohol intake within 4 hours
Performing vigourous exercise within 30 mins
Eating large meals within 2 hours
SLMC
YES
YES
YES
YES
YES
NO
CGH
YES
YES
YES
YES
YES
NO
UST
YES
YES
YES
YES
YES
NO
UP-PGH
YES
YES
YES
YES
YES
NO
LCP
YES
YES
YES
YES
YES
NA
PHC
YES
NO
YES
YES
YES
NO
VMMC
YES
YES
YES
YES
YES
YES
UPHDMC
YES
YES
YES
YES
YES
YES
MDH
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
NO
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NA
YES
YES
YES
YES
YES
NO
NO
NO
YES
YES
YES
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
NO
NA
NA
NA
NA
NA
NA
NA
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
NO
NO
NA
YES
NA
YES
YES
YES
NO
NO
NO
NO
YES
NO
YES
NO
NO
NO
NO
YES
YES
NO
NO
NO
YES
NO
NO
NO
NO
NO
NO
NO
YES
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
YES
YES
YES
YES
NO
NO
NO
NO
YES
YES
YES
YES
NO
NO
YES
NO
NO
NO
NO
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
APPENDIX G
Terms Used by PCCP-PFTI in the Interpretation of Spirometry Results
1
APPENDIX G
APPENDIX G
APPENDIX G
APPENDIX H
Definition of Terms/ Abbreviations
Definition of Terms
Abbreviations
Appendix I
Procedural Steps for Spirometry testing
(using closed circuit method)
o
o
o
o
o
o
o
o
o
APPENDIX J
PCCP Council of Diagnostic & Therapeutic Spirometry Testing Information Sheet
1. Name: __________________________________Age: _________
2. Gender: _________ Height: ____________ Weight: ______________
3. Race of patient: ___________________
4. Date of procedure: _________________Time of procedure: ___________
5. Room temperature: __________Atmospheric pressure: _________
6. Name of referring physician: __________________________
7. Smoking History: Status (Never, Current or Previous): __________________
Pack years: __________________ Last Intake: __________________
8. Working Diagnosis/ Reason for Testing: ____________________
9. Date of Previous Spirometry Testing: _________________
10. Conditions that may influence the results (Place a check if present or x if absent)
Trade Name
Dosage
Salbutamol
Terbutaline
Formoterol
Indacaterol
Salmeterol
Ipratropium Bromide
Tioptropium
Methylxanthines
Formoterol/
Budesonide
Salmeterol/
Fluticasone
Oral Steroid
Roflumilast
Please use the back page for additional listing of medicines if needed
Appendix K: Forms Used for Voting of the Statements/ Recommendations - 1 Name of Institution/ Chapter/ Key Opinion Leader: ____________________________
Name of Person Answering in behalf of Institution/ Chapter: ____________________
For each of the 51 recommendations, kindly write the letter corresponding to your answer/
vote. The choices are
A. We disagree with the recommendation.
B. We agree with recommendation and are, in fact, already instituting this.
C. We agree with the recommendation, and we will institute measures to implement
this.
D. We agree with the recommendation but will have a hard time implementing this in the
near future.
_____ 1
_____ 2
(R1-2) A referring physician can either request any of the following types of
spirometry testing: simple spirometry, spirometry with post bronchodilator study
and simple spirometry study with possible post bronchodilator study if initial
result is abnormal
_____ 3
_____ 4
_____ 5
_____ 6
(R2-3) A log book should be maintained. It should contain the following: daily
calibration results, anomalous events or problems encountered with the system
and feedback of the interpreter to the technician.
_____ 7
(R3-1) The following should be measured and reported FEV1/ FVC, FVC and
FEV1.
_____ 8
_____ 9
_____ 10
(R5-1) In the spirometry report, there should be 3 sections for the interpretation
of the results: technicians comments, interpreters comments and final
interpretation.
_____ 11
For each of the 51 recommendations, kindly write the letter corresponding to your answer/
vote. The choices are
A. We disagree with the recommendation.
B. We agree with recommendation and are, in fact, already instituting this.
C. We agree with the recommendation, and we will institute measures to implement
this.
D. We agree with the recommendation but will have a hard time implementing this in the
near future.
_____ 12
(R6-1) The spirometry report should contain a section on general data which
contains the following information: name of patient, age of patient, gender of
patient, measured height, measured weight, race of patient, date of procedure,
time of procedure, room temperature at time of testing, room atmospheric
pressure at time of testing, name of referring physician, name of the laboratory
and phone number of laboratory
_____ 13
_____ 14
(R7-2) The technician comments should state if the test done satisfies the
ATS-ERS-TFLFT (1) reproducibility criteria. If not, it should state which of the
criteria was not satisfied.
_____ 15
(R7-3) The technician comments should state if conditions that may alter the
results are present or not.
_____ 16
_____ 17
_____ 18
(R8-1) In the interpretation of the results, the report should use the phrase
ventilatory defect (not lung defect or ventilatory pattern).
_____ 19
_____ 20
(R10-1) The best trial flow volume (pre and post-bronchodilator if done) should
be part of the spirometry report.
_____ 21
_____ 22
_____ 23
(R10-4) In cases wherein the interpretation uses any of the following: flow
volume loops configuration, FEF25-75 and other spirometry parameters, it
should be stated that these parameters are suggestive but not definitive. Use
of the term compatible is discouraged.
Appendix K: Forms Used for Voting of the Statements/ Recommendations - 3 For each of the 51 recommendations, kindly write the letter corresponding to your answer/
vote. The choices are
A. We disagree with the recommendation.
B. We agree with recommendation and are, in fact, already instituting this.
C. We agree with the recommendation, and we will institute measures to implement
this.
D. We agree with the recommendation but will have a hard time implementing this in the
near future.
_____ 24
(R11-1) Grading of severity should be based on the criteria set by ATS-ERSTFLFT (2).
_____ 25
_____ 26
_____ 27
_____ 28
(R12-3) If the spirometer software is capable (optional), values < 5th percentile
of the predicted should be flagged down (either a change of color or marked).
If the spirometer software is not capable, it is encouraged that the technician or
interpreter derived this value by manual computation or use an online
calculator at http://www.dynamicmt.com/dataform3.html.
_____ 29
(R12-4) The cut-off value for the FEV1/FVC ratio should be <5th percentile of
predicted value. If not available, it should stated in the interpreters comments
what cut off value was used for the lower limit of normality.
_____ 30
(R12-5) Cut off value for the lower limit of normality should be stated in the
interpreters comments.
_____ 31
_____ 32
_____ 33
(R13-3) FEF 25-75 should not be used in the interpretation of the post
bronchodilator response.
_____ 34
_____ 35
_____ 36
(R15-1) In the section on Interpreters Comments the report should state what
reference equation was used in the interpretation of the results. Statement on
the need to correlate clinically should be written in the final interpretation.
Appendix K: Forms Used for Voting of the Statements/ Recommendations - 4 For each of the 51 recommendations, kindly write the letter corresponding to your answer/
vote. The choices are
A. We disagree with the recommendation.
B. We agree with recommendation and are, in fact, already instituting this.
C. We agree with the recommendation, and we will institute measures to implement
this.
D. We agree with the recommendation but will have a hard time implementing this in the
near future.
_____ 37
_____ 38
_____ 39
(R16-2) It the test does not satisfy one of the acceptability or reproducibility
criteria, it should state the consequences of this deficiency in the parameters.
_____ 40
(R17-1) The spirometry report should contain at least, the pre (and postbronchodilator if done) best trial flow volume loop/s
_____ 41
(R17-2) The spirometry report should contain all trials flow volume loops if an
interpretation of CAO or UAO made. If the spirometry software is not capable,
the reproducibility of the flow volume loop should be stated in the section on
Interpreters Comments.
_____ 42
(R18-1) The spirometry report should contain the measured parameters tabular
graph.
_____ 43
(R18-2) The first 3 rows of should be allotted for the MRSP in the following
order: FEV1/FVC, FVC and FEV1.
_____ 44
(R18-3) The succeeding rows maybe be allotted for the other parameters in
the following order of priority: PEFR, FEF 25-75, FEF 75, MIF50/ MEF50,
FEV1/FV6, MVV, MIF 50, MEF 50, FEV6, FEF 25 and FEF 50
_____ 45
(R18-4) In the tabular graph, each parameter should have its corresponding
actual values (pre & post if done), predicted, 95% confidence interval limit (if
the spirometry software is capable), % predicted (pre & post if done) and %
change from baseline of post-bronchodilator measurement.
_____ 46
(R18-5) In the tabular graph (if the spirometry software is capable), each
parameter below the 5th percentile should be flagged down.
_____ 47
(R19-1) The report should contain a section for the interpreters comments (IC)
of the results.
_____ 48
(R19-2) The IC section should state if the test satisfied the ATS-ERS-TFLFT
(1) reproducibility criteria. If the test did not satisfy the criteria, the interpreter
should state the consequences of this deficiency.
_____ 49
(R19-3) The section should contain the basis for the interpretation. Specifically
it should apply, the ATS-ERS-TFLFT (2) interpretation algorithm, criteria for
severity and response to bronchodilator
Appendix K: Forms Used for Voting of the Statements/ Recommendations - 5 For each of the 51 recommendations, kindly write the letter corresponding to your answer/
vote. The choices are
A. We disagree with the recommendation.
B. We agree with recommendation and are, in fact, already instituting this.
C. We agree with the recommendation, and we will institute measures to implement
this.
D. We agree with the recommendation but will have a hard time implementing this in the
near future.
_____ 50
(R20-1) The spirometry report should contain a section for the final
interpretation (FI) of the results.
_____ 51
Appendix L
List of Voters
Southern Tagalog
Northern Mindanao
Negros
Mindanao
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
(R1-3) In a situation
wherein the referring
physician requested only a
simple spirometry and it
showed abnormal results,
the technician in
coordination with doctor
(interpreter of the results)
should try to ask
permission from the
referring physician to
proceed in performing a
post bronchodilator study
while the patient is still in
the test site. (very good
consensus)
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
B O ANSW
CGH
VMMC
LCP
UP-PGH
MMC
O ANSW
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
SLMC
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
CGH
VMMC
LCP
UP-PGH
MMC
(R10-2) In cases of
possible upper airway
obstruction interpretation,
all of the trials flow volume
loop should be reported. If
the spirometry software is
not capable, reproducibility
of the flow volume loop
should be mentioned in the
interpreters comments.
(very good consensus)
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
C/B
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
C/B
CGH
VMMC
LCP
UP-PGH
MMC
(R13-2) Interpretation of
response to bronchodilator
should be based on the
criteria set by the ATS-ERSTFLFT. (perfect consensus)
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
B/C
ANSW
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
(R14-1) Interpretation
should include suggestion/s
on need for additional test if
warranted. (very good
consensus)
C/B
C/B
A/B
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
Y/A
Y/A
Y/A
Y/A
N/D
Y/A
Y/A
Y/A
Y/A
Y/A
Y/A
Y/A
Y/A
Y/A
Y/A
Y/A
SLMC
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
A/C
SLMC
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
SLMC
CGH
VMMC
LCP
UP-PGH
MMC
CHH MDH
UST
PHC
UPHM
Dr. Roa
Dr. Zotomayor
Dr. Wi
Appendix N: Computation of the Parameters Needed for the Level of Consensus Classification
11
Level of Consensus
64.7
0.0
94.4
47.1
10
94.4
23.5
100.0
23.5
15
100.0
0.0
15
100.0
0.0
100.0
0.0
18
17
100.0
0.0
11
100.0
0.0
Appendix N: Computation of the Parameters Needed for the Level of Consensus Classification
A
Level of Consensus
11
100.0
0.0
11
100.0
0.0
12
100.0
0.0
100.0
0.0
17
100.0
0.0
13
100.0
0.0
17
100.0
0.0
100.0
0.0
17
16
100.0
0.0
16
100.0
0.0
Appendix N: Computation of the Parameters Needed for the Level of Consensus Classification
A
15
Level of Consensus
100.0
0.0
12
100.0
0.0
15
100.0
0.0
17
100.0
0.0
17
100.0
0.0
15
100.0
0.0
100.0
0.0
100.0
5.6
VG
94.4
0.0
VG
18
15
10
Appendix N: Computation of the Parameters Needed for the Level of Consensus Classification
Level of Consensus
14
94.4
11.8
VG
11
100.0
5.6
VG
12
94.4
0.0
VG
94.4
0.0
VG
10
100.0
5.6
VG
11
94.4
0.0
VG
13
100.0
5.9
VG
10
100.0
5.6
VG
(R8-1) In the interpretation of the results, the report should use the
1
phrase ventilatory defect (not lung defect or ventilatory pattern).
(very good consensus)
(R9-1) Spirometry results should be interpreted based on the ATSERS-TFLFT suggested algorithm. (very good consensus)
(R10-2) In cases of possible upper airway obstruction interpretation,
all of the trials flow volume loop should be reported. If the spirometry
software is not capable, reproducibility of the flow volume loop
should be mentioned in the interpreters comments. (very good
consensus)
Appendix N: Computation of the Parameters Needed for the Level of Consensus Classification
A
Level of Consensus
100.0
5.6
VG
10
100.0
5.6
VG
11
94.4
11.1
VG
16
100.0
5.6
VG
10
88.2
6.7
VG
0.0
0.0
VG
94.4
0.0
VG
100.0
11.1
VG
10
Appendix N: Computation of the Parameters Needed for the Level of Consensus Classification
A
(R17-2) The spirometry report should contain the flow volume loops
of all trials, if an interpretation of CAO or UAO made. If the
spirometry software is not capable, the reproducibility of the flow
volume loop should be stated in the section on Interpreters
Comments. (very good consensus)
(R18-1) The spirometry report should contain the measured
parameters tabular graph. (very good consensus)
(R18-2) The first 3 rows of the tabular graph should be allotted for
the MRSP in the following order: FEV1/FVC, FVC and FEV1. (very
good consensus)
(R18-3) The succeeding rows maybe be allotted for the other
parameters in the following order of priority: PEFR, FEF 25-75, FEF
75, MIF50/ MEF50, FEV1/FV6, MVV, MIF 50, MEF 50, FEV6, FEF
25 and FEF 50. (very good consensus)
(R19-2) The IC section should state if the test satisfied the ATSERS-TFLFT (1) reproducibility criteria. If the test did not satisfy the
criteria, the interpreter should state the consequences of this
deficiency.(very good consensus)
(R19-3) The IC section should also contain the basis for the
interpretation. Specifically it should apply, the ATS-ERS-TFLFT (2)
interpretation algorithm, criteria for severity and response to
bronchodilator. (very good consensus)
(R20-2) The FI should contain the following information: presence or
absence of obstructive ventilatory defect and severity if present,
presence or absence of probable restrictive ventilatory defect and
severity if present, significant or no significant response to
bronchodilator if done, suggestion for specific additional test if
indicated, a generic sentence Please correlate results with patients
clinical data. and printed name plus signature of the interpreter/s.
(very good consensus)
100.0
17.6
Level of Consensus
VG
15
100.0
11.1
VG
100.0
5.6
VG
100.0
5.6
VG
10
100.0
11.1
VG
100.0
5.6
VG
15
100.0
5.6
VG
Appendix O
PCCP-DATS Committee on Spirometry Consensus