Beruflich Dokumente
Kultur Dokumente
The 3166 Resistor Set tight-tolerance calibrators are more precise than common resistors and can verify meter performance. Six resistors included.
YSI 3161
YSI 3163
YSI 3165
YSI 3167
YSI 3168
YSI 3169
YSI 3401
YSI 3440
YSI 3253, 3403
YSI 3254
0
0.1
1.0
10
100
1,000 S/cm
1
10
100
Conductivity Cell Selection Chart for the 3200
(For the 3100 see specific ranges on back)
1,000 S/cm
10,000 S/cm
100,000 S/cm
1,000 S/cm
10,000 S/cm
50,000 S/cm
0.50% tolerance
0.25% tolerance
0.25% tolerance
1.0% tolerance
1.0% tolerance
1.0% tolerance
1 quart
1 quart
1 quart
8 pints
8 pints
8 pints
1,000
mS/cm
www.YSI.com
Y S I Environmental
amj@ysi.com
Technology
Modes
Resistance Ratio
Forced Current
Conductivity
Conductance
Resistivity
Resistance
Salinity
Total Dissolved Solids
Temperature
Conductivity
Salinity
Conductance
Temperature
Conductance Range
Resistance
0 to 0.9999 S
0.950 to 9.999 S
9.50 to 99.99 S
95.0 to 999.9 S
950 to 9999 S
9.50 to 99.99 mS
95.0 to 999.9 mS
0.95 to 3.00 S
Accuracy
0.30% full scale
0.20% full scale
0.10% full scale
0.10% full scale
0.10% full scale
0.10% full scale
0.30% full scale
1.0% full scale
Resolution
Range (Conductivity) Accuracy
0.0001 S
0 to 49.99 S/cm1
0.50% full scale
0.001 S
0 to 499.9 S/cm
0.50% full scale
0.01 S
0 to 4999 S/cm
0.50% full scale
0.50% full scale
0.1 S
0 to 49.99 mS/cm2
0.50% full scale
1 S
0 to 499.9 mS/cm3
0.01 mS
1 - Do not use K = 10 cm-1 cell.
0.1 mS
2 - Requires K = 10 cm-1 or 1.0 cm-1 cell
0.01 S
3 - Requires K = 10 cm-1 cell
Range
0 to 9.999
0 to 99.99
0 to 999.9
0 to 9.999 k
0 to 99.99 k
100.0 to 999.9 k
1.00 to 9.99 M
10.0 to 29.9 M
Accuracy
0.2% full scale
0.1% full scale
0.1% full scale
0.1% full scale
0.1% full scale
0.2% full scale
0.5% full scale
1% full scale
Resolution
0.001
0.01
0.1
0.001 k
0.01 k
0.1 k
0.01 M
0.1 M
Salinity
0 to 80 ppt (NaCl) 0.1 ppt
Temperature -5 to +100C
0.1C
TDS
0 to 19,999 mg/L 0.50%
Resolution
0.01 S/cm
0.1 S/cm
1 S/cm
0.01 mS/cm
0.1 mS/cm
0.1 ppt
0 to 80 ppt
2% or 0.1 ppt
0.1 ppt
0.01 C
-5 to +95 C
0.1 C + 1 lsd
0.1C
1 mg/L
Temperature compensation
Method
Reference temperature
Temperature coefficient
Cell conguration storage
Data storage
Cell constant
Cell calibration
Output
Alarm & clock
Display
Cell connector
Platinizing
Power
Approvals
Environment
linear, nonlinear
0 to 100C
0 to 10%, nonlinear
6 congurations
100 points
0.001 to 100 cm-1
up to 5 points
RS232
yes
Graphic LCD
7-pin mini DIN
included
115, 220 VAC
UL, CSA, CE
95% RH non-condensing
linear
15 to 25C
0 to 4%
na
na
0.01, 0.1, 1, 10 cm-1
single point
na
na
LCD
7-pin mini DIN
included
115, 220 VAC
UL, CSA, CE
95% RH non-condensing
ISO 9001
ISO 14001
Whos Minding the Planet? Resistance
Ratio Technology and Pure Data for a
Healthy Planet are trademarks of YSI
Incorporated.
2008 YSI Incorporated
Printed in USA 1208 W40-05
Y S I incorporated
Whos Minding
the Planet?
S.I. Cell
Cell Constant Constant
Material
1.0/cm
100/m
ABS plastic
1.0/cm
100/m
Pyrex 7740
1.0/cm
100/m
Pyrex 7740
0.1/cm
10/m
Pyrex 7740
0.1/cm
10/m
Pyrex 7740
Overall
Length
146 mm
178 mm
135 mm
146 mm
159 mm
Max
O.D.
13 mm
13 mm
19 mm
25 mm
25 mm
Chamber
I.D.
10 mm
10 mm
11 mm
21 mm
21 mm
Chamber
Depth Volume
20 mm
51 mm
83 mm 5 mL
76 mm 30 mL
52 mm