Beruflich Dokumente
Kultur Dokumente
BasedontheGermanStandardDIN68098
1. Introduction
ThisdocumentconstitutesanexcerptofproceduresanddataaspublishedintheGermanstandard
DIN 68098. The aim is to determine absorbed dose to water in the center of small highenergy
photonfieldswithdimensionsof4cmx4cmorsmaller.
Althoughthisdocumentprovidesthereaderwithaconciseoverviewofequationsandfactorsitshall
notreplacepertinentstandards,protocolsandpublications,norisitintendedtogiveallofthedetails
that are important for accurate dosimetry. Also, the procedures outlined in this document are not
theonlyonesdescribedinthereferencedliterature,theyconstituteonlyoneofseveralpossibilities
forabsorbeddosedeterminationinsmallfields.
2. General Instructions
2.1 Measuring phantom
Thisdocumentassumesthatallmeasurementsaremadeinawaterphantom.
D560.210.03/02Sch2014051
Forcylindricalionizationchamberstheeffectivepointofmeasurementisdefinedasapointshifted
by 0.5 from the axis of the chamber towards the focus ( is the inner radius of the measuring
chamber volume) [DIN 68002]. To measure dose at a focus distance of cm, the axis of the
cylindricalchambermustbepositionedatafocusdistanceof cm+0.5 ,i.e.thechambermust
beshiftedawayfromthefocus,i.e.downstream.Thelocationoftheeffectivepointofmeasurement
ofsolidstatedetectorscanbetakenfromthedatasheetofthedetector.
The TRUFIX system
The task of positioning various types of radiation detectors precisely in their effective point of
measurement can be quite challenging. The patented TRUFIX system (see Figure1) facilitates this
task considerably. TRUFIX can be used on automated PTW water phantoms (MP2, MP3 etc.) in
connection with most PTW therapy detectors. A plastic tip lets you easily locate the water surface
where the coordinate system is set to (0,0,0). Then the plastic tip is replaced by a holding device
specifictoeachdetectortype,andtheeffectivepointofmeasurementisautomaticallyplacedatthe
tip's earlier position. The radius of cylindrical chambers, the waterequivalent window thickness of
planeparallel chamber and solid state detector windows as well as the detector centers are
automaticallyaccountedfor.
Figure1:TheTRUFIXdetectorpositioningsystem.
(21)
,
,
and correct for air density, incomplete charge collection,
The correction factors
polarityeffectandhumidity,respectively.Theyaredescribedindetailin[DIN68002].Forsolidstate
detectorssomeofthecorrectionfactorsinequation(21)maynotapply.
Forabsorbeddosedeterminationinsmallfields,additionalfactorsaretobeappliedtothecorrected
reading asdescribedinthefollowingchapters.
D560.210.03/02Sch2014052
(31)
where
absorbeddosetowater
=1 0,03 (replacementcorrectionfactor, istheinnerradiusofthemeasuring
volumeofthecylindricalchamber,givenincm)
calibrationfactorforabsorbeddosetowaterfor60Co
correctedreadingofthedosemeterasdescribedinchapter2.4
D560.210.03/02Sch2014053
PTWchambertype
at6MV
at15MV
PinPoint31015(0.03cm)
1.000
1.001
Semiflex31010(0.125cm)
1.001
1.002
Semiflex31013(0.3cm)
1.004
1.008
Table1:Correctionfactor forionizationchambermeasurementsina4cmx4cmfield,for6MV
and15MV.DataistakenfromTable3andequation(C.11)in[DIN68098].
1.2661
0.0595
(32)
where
10 ,
20
radiationqualityindex
dosevaluesat10cmand20cmdepth,respectively.Sourcetosurface
distance100cm,fieldsizeatthephantomsurface10cmx10cm.
Ifafieldsizeof10cmx10cmcannotbeadjustedforthedeterminationofthequalityindex ,the
qualityindexisdeterminedby[DIN68098]:
(33)
where
20 /
10 at100cmsourcetodetectordistance
TPR20,10
Tissuephantomratio
fielddimensioninonedirection,incm
=19.5cm
A1
=0.625
A2
=0.679
b1
=0.208
b2
=1.213
D560.210.03/02Sch2014054
Hint: DIN 68002 offers an iteration process to increase the accuracy of equation (33).
However, the increase of accuracy is smaller than the uncertainties of the parameters in
equation(33).Therefore,aniterationprocessisnotsuggestedinthisdocument.
Thecorrectionfactor
, istakenfrom[DIN68002].NotethatthenotationinDIN68002is
(34)
Thiscalibrationfactorisvalidfortheradiationqualityandthemeasuringdepthusedforthecross
calibration, and for a field size of 4 cm x 4 cm. Deviations from these calibration conditions are
accountedforbyapplyingcorrectionfactorstothereadingasdescribedinchapter3.4.
Hint:IfPinPointchambersareusedassmallfielddetectors,andifacalibrationfactorat 60Co
as well as , factors are known, a cross calibration may not be necessary down to a
fieldsizeof1.5cmx1.5cm.
(35)
where
DW
absorbeddosetowater
correction factor for the size of the small field detector, causing a volume effect in
smallfields,seeTables3aand3b
correction factor accounting for variation of detector response when changing the
fieldsizefrom4cmx4cmtosmallerfields,seeTables2aand2b
calibration factor for the small field detector under crosscalibration conditions,
seechapter3.3
correctedreadingofthedosemeterasdescribedinchapter2.4
Hint: Due to their size Semiflex chambers with a volume of 0.3 cm or 0.125 cm are not
recommendedfordosimetryatfieldsizessmallerthan4cmx4cm.
D560.210.03/02Sch2014055
PTWdetectortype
Semiflex31010(0.125cm)
=2.5cm
=4cm
=5cm
1.007
1.000
1.007
PinPoint31014(0.015cm)
1.006
1.000
1.007
PinPoint3D31016(0.016cm)
1.005
1.000
1.006
DosimetryDiodeP60016
1.002
1.000
1.003
DosimetryDiode60017/60018(PTW)
1.001
1.000
0.998
DosimetryDiode60017/60018(MC)
1.006
1.000
1.003
DiamondDetector60003
1.000
1.000
1.000
microDiamond60019
1.000
1.000
1.000
Table2a:Correctionfactor
formeasurementsinthecenterofsmallfieldsat10cmdepth.The
field dimensions are designated by , the radiation quality is 6 MV photons. Data is taken from
TableE.1in[DIN68098].AllvaluesarebasedonMonteCarlosimulationswiththeexceptionofthe
line labeled (PTW). The values for the diamond detectors are preliminary as no measurements or
simulationshavebeenpublishedsofar.
PTWdetectortype
Semiflex31010(0.125cm)
=2.5cm
1.006
=4cm
1.000
=5cm
0.996
PinPoint31014(0.015cm)
1.003
1.000
0.996
PinPoint3D31016(0.016cm)
1.003
1.000
0.995
DosimetryDiodeP60016
1.002
1.000
0.991
DosimetryDiode60017/60018(PTW)
1.000
1.000
0.997
DosimetryDiode60017/60018(MC)
1.003
1.000
0.991
DiamondDetector60003
1.000
1.000
1.000
microdiamond60019
1.000
1.000
1.000
Table2b:Correctionfactor
formeasurementsinthecenterofsmallfieldsat20cmdepth.The
field dimensions are designated by , the radiation quality is 6 MV photons. Data is taken from
TableE.2in[DIN68098].AllvaluesarebasedonMonteCarlosimulationswiththeexceptionofthe
linedlabeled(PTW).The valuesforthediamonddetectorsarepreliminaryasnomeasurementsor
simulationshavebeenpublishedsofar.
PTWdetectortype
Semiflex31013(0.3cm)
PinPoint
=1cm
1.570
1.076
=2cm
1.067
1.007
=3cm
1.012
1.001
=4cm
1.004
1.000
PTWdetectortype
Semiflex31013(0.3cm)
PinPoint
=1cm
1.543
1.075
=2cm
1.091
1.009
=3cm
1.023
1.002
=4cm
1.008
1.001
D560.210.03/02Sch2014056
References
[DIN68002]
[DIN68098]
EDIN68098,EntwurfzurDeutschenNorm,KlinischeDosimetrieTeil8:Dosimetrie
kleinerPhotonenBestrahlungsfelder,Mrz2014
PTWFREIBURG
PhysikalischTechnischeWerksttten
Dr.PychlauGmbH
LrracherStr.7
79115FREIBURG
GERMANY
Phone: +49761490550
Fax: +497614905570
info@ptw.de
www.ptw.de
D560.210.03/02Sch2014057