Sie sind auf Seite 1von 41

Mindray BS-120 / BS-200 / BS-300 / BS-380 / BS-400

rev 432.0.2 - 2013-01-15

Chema Diagnostica
Via Campania 2/4
I-60030 Monsano | An Italy
T +39 0731 605064 | F +39 0731 605672
www.chema.com | mail@chema.com

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Mindray BS-120 / BS-200 / BS-300 / BS-380 / BS-400


rev 432.0.2 - 2013-01-15
Warning!
These programs should be used only as guideline.
To check the reagent correct setting on the
instrument, use only good quality control sera
and work accordingly to with good laboratory
practice. Instrumental factor must be checked.
Chema Diagnostica will not be responsible for
instrumental bad programming.

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 3 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

ALBUMIN - Code BC 0100 - 0500 - 1000 - 1500 CH


R1 - LIQUID READY TO USE SINGLE REAGENT
ROOM TEMPERATURE OR REFRIGERATE (2-30C)
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
ALB
*
ALBUMIN
2
ENDPOINT
INCREASE
g/dl
0.1
630
3
400
0
0
17
19
-

BS-300

BS-380

BS-400

605
700
2
300
0
25
-

34
36

44
36
-

8-10

10-12

0.1
6.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 4 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

ALKALINE PHOSPHATASE FL DGKC - Codes AL F080 / F245 / F400 / F600 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
ALP
*
ALP DGKC
2
KINETIC
INCREASE
U/L
1
405
510
5
200
50
3
4
13
15
16
8
-

BS-300

BS-380

BS-400

412
505
5
200
50
5
20
10

41
56

50
69
-

5
2500
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 5 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

ALKALINE PHOSPHATASE FL IFCC - Codes AF F080 / F245 / F400 / F600 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
ALP
*
ALP IFCC
2
KINETIC
INCREASE
U/L
1
405
700
5
200
50
3
4
13
15
16
8
-

BS-300

BS-380

BS-400

412
700
5
200
50
5
20
10

41
56

50
69
-

5
2500
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 6 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

AMYLASE FL - Codes AM F060 / F120 / F245 CH


R1 - LIQUID READY TO USE SINGLE REAGENT
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
AMY
*
AMYLASE
2
KINETIC
INCREASE
U/L
1
405
700
6
250
3
4
13
15
-

BS-300

BS-380

BS-400

412
700
6
250
5
20
-

41
69

50
75
-

3
2000
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 7 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

AMYLASE EPS FL - Codes EA F080 / F245 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
AMY
*
AMYLASE IFCC
2
KINETIC
INCREASE
U/L
1
405
510
5
200
50
3
4
12
14
16
19
-

BS-300

BS-380

BS-400

412
505
5
200
50
5
15
25

41
52

50
64
-

5
1500
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 8 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

PANCREATIC ISOAMYLASE FL - Code PA F080 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
AMYP
*
ISOAMYLASE IFCC
2
KINETIC
INCREASE
U/L
1
405
510
5
200
50
3
4
12
14
16
19
-

BS-300

BS-380

BS-400

412
505
5
200
50
5
15
25

41
52

50
64
-

3
2000
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 9 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

BICARBONATE FL - Codes BR F060 / F245 / F400 CH


R1 - LIQUID READY TO USE SINGLE REAGENT
ROOM TEMPERATURE OR REFRIGERATE (2-30C)
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
CO2
*
BICARBONATE
2
ENDPOINT
DECREASE
mmol/l
1
405
700
3
300
0
0
17
19
-

BS-300

BS-380

BS-400

412
700
2
200
0
25
-

34
36

44
36
-

8-10

10-12

1
50
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 10 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

BILIRUBIN TOTAL FL - Codes DT F125 / F500 CH


R1 - USE LIQUID READY TO USE REAGENT A
R2 - USE LIQUID READY TO USE REAGENT A
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
BILT
*
BILIRUBIN TOTAL
2
ENDPOINT
INCREASE
mg/dl
0.1
510
700
10
200
50
-1
-1
17
19
16
12
-

BS-300

BS-380

BS-400

505
700
10
200
50
-1
25
15

58
60

77
79
-

0.1
20.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 11 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

BILIRUBIN DIRECT FL - Codes DD F125 / F500 CH


R1 - USE LIQUID READY TO USE REAGENT A
R2 - USE LIQUID READY TO USE REAGENT A
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
BILD
*
BILIRUBIN DIRECT
2
ENDPOINT
INCREASE
mg/dl
0.1
546
700
10
200
50
-1
-1
17
19
16
12
-

BS-300

BS-380

BS-400

546
700
10
200
50
-1
25
15

58
60

77
79
-

0.1
13.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 12 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

CALCIUM ASX - Code CA 0100 - 0500 CH


R1 - LIQUID READY TO USE SINGLE REAGENT
ROOM TEMPERATURE OR REFRIGERATE (2-30C)
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
CA
*
CALCIUM
2
ENDPOINT
INCREASE
mg/dl
0.1
630
700
3
300
0
0
17
19
-

BS-300

BS-380

BS-400

660
700
2
200
0
25
-

34
36

44
36
-

8-10

10-12

0.1
20.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 13 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

CALCIUM - Code CA 0305 - 0505 CH


R1 - INSTALL LIQUID READY TO USE REAGENT B
R2 - INSTALL LIQUID READY TO USE REAGENT A
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
CA
*
CALCIUM
2
ENDPOINT
INCREASE
mg/dl
0.1
578
700
7
200
200
0
0
17
19
-

BS-300

BS-380

BS-400

570
700
5
150
150
0
25
-

38
40

63
67
-

8-10

10-12

0.1
20.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 14 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

CHOLESTEROL FL - Codes CT F100 / F400 / 100F CH


R1 - LIQUID READY TO USE SINGLE REAGENT
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
CHOL
*
CHOLESTEROL
2
ENDPOINT
INCREASE
mg/dl
1
510
700
3
300
0
0
33
38
-

BS-300

BS-380

BS-400

505
700
2
200
0
50
-

58
60

77
79
-

8-10

10-12

20
700
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 15 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

HDL DIRECT FL - Code HD F080 / F245 / F400 CH


R1 - USE LIQUID RFADY TO USE REAGENT A
R2 - USE LIQUID RFADY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
HDL-C
*
HDL-C
2
ENDPOINT
INCREASE
mg/dl
1
578
700
3
225
75
-1
-1
17
19
16
16
-

BS-300

BS-380

BS-400

605
700
2
225
75
-1
25
19

58
60

77
79
-

8-10

10-12

5
220
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 16 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

LDL DIRECT FL - Code DL F080 CH


R1 - USE LIQUID RFADY TO USE REAGENT A
R2 - USE LIQUID RFADY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
LDL-C
*
LDL-C
2
ENDPOINT
INCREASE
mg/dl
1
578
700
3
225
75
-1
-1
17
19
16
16
-

BS-300

BS-380

BS-400

605
700
2
225
75
-1
25
19

58
60

77
79
-

5
400
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 17 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

CHOLINESTERASE FL (DGKC) - Codes CH F096 - F245 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
CHE
*
CHOLINESTERASE
2
KINETIC
DECREASE
U/L
1
405
700
5
200
50
3
4
13
15
16
8
-

BS-300

BS-380

BS-400

412
700
5
200
50
5
20
10

41
56

50
69
-

100
25000
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 18 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

CK-NAC FL - Codes CK F060 / F120 / F245 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
CK
*
CK
2
KINETIC
INCREASE
U/L
1
340
546
5
200
50
10
12
18
23
16
12
-

BS-300

BS-380

BS-400

340
546
5
200
50
15
30
15

51
60

63
80
-

5
2000
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 19 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

CK-MB FL - Code MB F060 / F120 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
CK
*
CK
2
KINETIC
INCREASE
U/L
1
340
546
10
200
50
10
12
18
23
16
12
-

BS-300

BS-380

BS-400

340
546
10
200
50
15
30
15

51
60

63
80
-

3
1000
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 20 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

COPPER - Code CU 0100 CH


R1 - INSTALL LIQUID READY TO USE REAGENT B
R2 - INSTALL LIQUID READY TO USE REAGENT A
ROOM TEMPERATURE (Warning! The reagent A could precipitate if refrigerate)
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
CU
*
COPPER
2
ENDPOINT
INCREASE
mg/dl
1
578
700
20
150
150
0
0
17
19
-

BS-300

BS-380

BS-400

570
700
20
150
150
0
25
-

38
40

63
67
-

8-10

10-12

5
500
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 21 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

CREATININE - Code CR 0500 / 1000 / 1500 CH


R1 - INSTALL LIQUID READY TO USE REAGENT B
R2 - INSTALL LIQUID READY TO USE REAGENT A
ROOM TEMPERATURE OR REFRIGERATE (2-30C)
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
CREA
*
CREATININE
2
FIXED TIME
INCREASE
mg/dl
0.01
510
670
18
180
180
2
2
8
8
16
5
-

BS-300

BS-380

BS-400

505
700
16
160
160
2
12
5

38
48

45
56
-

0.1
20.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 22 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

CREATININE E FL - Code CE F125 / F375 / F600 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
CREA-E
*
CREAT ENZ
2
ENDPOINT
INCREASE
mg/dl
0.01
546
700
6
180
60
-1
-1
17
19
16
12
-

BS-300

BS-380

BS-400

546
700
6
180
60
-1
25
15

58
60

77
79
-

0.1
20.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 23 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

GAMMA-GT FL - Codes GT F080 / F245 / F400 / F600 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
GGT
*
GAMMA GT
2
KINETIC
INCREASE
U/L
1
405
670
20
200
50
3
4
13
15
16
8
-

BS-300

BS-380

BS-400

412
700
20
200
50
5
20
10

41
56

50
69
-

3
800
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 24 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

GLUCOSE FL - Code GL F400 / 100F CH


R1 - LIQUID READY TO USE SINGLE REAGENT
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
GLU
*
GLUCOSE
2
ENDPOINT
INCREASE
mg/dl
1
510
700
3
300
0
0
33
38
-

BS-300

BS-380

BS-400

505
700
2
200
0
50
-

58
60

77
79
-

8-10

10-12

10
500
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 25 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

GLUCOSE UV FL - Code GL F601 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
GLU
*
GLUCOSE
2
ENDPOINT
INCREASE
mg/dl
1
340
660
3
240
60
0
0
33
38
-

BS-300

BS-380

BS-400

340
660
3
240
60
0
50
-

58
60

77
79
-

8-10

10-12

10
1000
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 26 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

GOT/AST FL - Codes GO F080 / F245 / F400 / F600 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
AST
*
GOT-AST
2
KINETIC
DECREASE
U/L
1
340
405
20
200
50
3
4
13
15
16
8
-

BS-300

BS-380

BS-400

340
412
20
200
50
5
20
10

41
56

50
69
-

3
500
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 27 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

GPT/ALT FL - Codes GP F080 / F245 / F400 / F600 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
ALT
*
GPT-ALT
2
KINETIC
DECREASE
U/L
1
340
405
20
200
50
3
4
13
15
16
8
-

BS-300

BS-380

BS-400

340
412
20
200
50
5
20
10

41
56

50
69
-

3
500
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 28 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

Application:
Preparation:

Storage:
Stability:

IRON FZ - Code FE F245 / F400 CH


R1 - USE LIQUID RFADY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B (preparation: read insert sheet)
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
FE
*
IRON
2
ENDPOINT
INCREASE
mg/dl
1
546
700
30
200
50
-1
-1
17
19
16
12
-

BS-300

BS-380

BS-400

570
700
30
200
50
-1
25
15

58
60

77
79
-

10
1000
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 29 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

LDH FL - Codes LD F060 / F120 / F245 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
LDH
*
LDH
2
KINETIC
DECREASE
U/L
1
340
670
3
240
60
3
4
13
15
16
3
-

BS-300

BS-380

BS-400

340
700
3
240
60
5
20
10

41
56

50
69
-

20
4000
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 30 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

LIPASE FL - Code LP F060 / LP F125 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
LIP
*
LIPASE
2
KINETIC
INCREASE
U/L
0.1
578
670
4
200
50
6
5
10
12
2
8
-

BS-300

BS-380

BS-400

570
700
4
200
50
8
16
10

41
56

50
69
-

0
60
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 31 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

MAGNESIUM - Code MG 0200 / 0500 CH


R1 - INSTALL LIQUID READY TO USE REAGENT B
R2 - INSTALL LIQUID READY TO USE REAGENT A
ROOM TEMPERATURE OR REFRIGERATE (2-30C)
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
MG
*
MAGNESIUM
2
ENDPOINT
INCREASE
mEq/l
0.01
510
700
4
200
200
0
0
17
19
-

BS-300

BS-380

BS-400

505
700
3
150
150
0
25
-

38
40

63
67
-

8-10

10-12

0.1
8.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 32 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

MAGNESIUM XL - Codes MX 0300 / 0500 CH


R1 - LIQUID READY TO USE SINGLE REAGENT
REFRIGERATE (2-8C)
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
MG
*
MAGNESIUM
2
ENDPOINT
INCREASE
mEq/l
0.01
546
700
3
300
0
0
17
19
-

BS-300

BS-380

BS-400

546
700
3
300
0
25
-

38
40

63
67
-

8-10

10-12

0.1
6.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 33 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

PHOSPHORUS - Code PH 0100 / 0500 CH


R1 - LIQUID READY TO USE SINGLE REAGENT
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
PH
*
PHOSPHORUS
2
ENDPOINT
INCREASE
mg/dl
0.01
340
700
3
300
0
0
33
38
-

BS-300

BS-380

BS-400

340
700
3
300
0
50
-

58
60

77
79
-

8-10

10-12

0.1
20.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 34 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

PROTEINS HS - Code HS 0100 / 0500 CH


R1 - LIQUID READY TO USE SINGLE REAGENT
ROOM TEMPERATURE OR REFRIGERATE (2-30C)
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
HSP
*
PROTEINS HS
2
ENDPOINT
INCREASE
mg/dl
1
578
700
3
400
0
0
33
38
-

BS-300

BS-380

BS-400

605
700
2
300
0
50
-

58
60

77
79
-

8-10

10-12

5
500
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 35 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

TOTAL PROTEINS - Codes TP 0100 / 0500 / 1000 / 1500 CH


R1 - LIQUID READY TO USE SINGLE REAGENT
ROOM TEMPERATURE OR REFRIGERATE (2-30C)
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
TP
*
PROTEINS TOTAL
2
ENDPOINT
INCREASE
g/dl
0.1
546
700
3
300
0
0
33
38
-

BS-300

BS-380

BS-400

546
700
3
300
0
50
-

58
60

77
79
-

8-10

10-12

1.0
12.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 36 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:







Mindray







Application:
Preparation:
Storage:
Stability:

TRIGLYCERIDES FL - Code TR F100 / F400 / 100F CH


R1 - LIQUID READY TO USE SINGLE REAGENT
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL

BS-120
BS-200
BS-300
BS-380
BS-400


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
TG
*
TRIGLYCERIDES
2
ENDPOINT
INCREASE
mg/dl
1
546
700
3
300
0
0
33
38
-

BS-300

BS-380

BS-400

546
700
2
200
0
50
-

58
60

77
79
-

8-10

10-12

10
1000
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 37 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

UREA UV FL - Codes AZ F080 / F245 / F400 / F600 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
UREA
*
UREA
2
FIXED TIME
DECREASE
mg/dl
1
340
670
4
320
80
2
3
7
9
-

BS-300

BS-380

BS-400

340
700
3
240
60
3
11
-

39
47

47
59
-

1
300
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 38 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

URIC ACID AOX FL - Codes AX F100 / F250 / F600 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
UA
*
URIC ACID
2
ENDPOINT
INCREASE
mg/dl
0.1
546
670
4
320
80
-1
-1
14
15
-

BS-300

BS-380

BS-400

546
700
3
240
60
-1
20
-

58
60

77
79
-

33-35

40-42

0.1
35.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 39 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

URIC ACID T FL - Code AU F250 / F402 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
UA
*
URIC ACID
2
ENDPOINT
INCREASE
mg/dl
0.1
546
670
4
320
80
-1
-1
14
15
-

BS-300

BS-380

BS-400

546
700
3
240
60
-1
20
-

58
60

77
79
-

33-35

40-42

0.1
25.0
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 40 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

Analyzer:








Application:

Preparation:


Storage:

Stability:

Mindray







BS-120
BS-200
BS-300
BS-380
BS-400

ZINC - Code ZN 0125 CH


R1 - INSTALL LIQUID READY TO USE REAGENT A
R2 - INSTALL LIQUID READY TO USE REAGENT B
REFRIGERATE AT 2-8C
AS INDICATED IN THE LABEL


PARAMETER
TEST NAME:
No.:
FULL NAME:
STANDARD No.:
REAC. TYPE:
DIRECTION:
UNIT:
PRECISION:
PRI. WAVE:
SEC. WAVE:
SAMPLE VOLUME (mL):
R1 (mL):
R2 (mL):
REAC. TIME
FROM:
REAC. TIME
TO:
INCUBATION TIME:
BLANKING TIME:
FROM:
LINEARITY RANGE
TO:
SUBSTRATE LIMIT:
FACTOR:

BS-120
BS-200
ZN
*
ZINC
2
ENDPOINT
INCREASE
mg/dl
1
578
670
15
240
60
0
0
33
38
-

BS-300

BS-380

BS-400

570
700
15
240
60
0
50
-

58
60

77
79
-

8-10

10-12

5
1000
CALIBRATION

RULE:
REPLICATES:

two-point linear
3

* user def

IUS.432.0.2

Directions for use

Edition 2013-01-15

page 41 of 41

Das könnte Ihnen auch gefallen