Beruflich Dokumente
Kultur Dokumente
Operator Manual
www.siemens.com/healthcare
Legend
9 Prerequisites
Instructions
Text lists
Cross reference
Caution/
Warning
Cause
Possible consequences
Precautions or remedies
Usability
Original language
Version
2 / 76
AX42-010.621.63.01.02
Operator Manual
Table of Contents
syngo InSpace 3D
InSpace 3D/Cardiac 3D workflow
Cardiac High Contrast 3D workflow
syngo DynaCT and syngo DynaCT Cardiac
DynaCT workflow
DynaCT Cardiac workflow
6
6
7
7
9
10
11
Primary reconstruction
Offline reconstruction
Secondary reconstruction
The InSpace Reconstruction window
Volume data display
Defining the reconstruction volume
Setting parameters for reconstruction
Adjusting cardiac phases for DynaCT Cardiac
Additional functions
Starting reconstruction
12
12
13
13
14
14
17
21
24
26
27
3D display techniques
Multiplanar reconstruction (MPR)
Maximum (MIP) and minimum (MinIP) intensity projection
Volume rendering technique (VRT)
Display in the InSpace task card
Rotating a volume
Changing 3D image display
Visualization of DynaCT images
Visualization of DynaCT Cardiac images
syngo iDentify: Dual volume display
Selecting and loading data sets for syngo iDentify
(dual volume display)
Optimizing dual volume display
Artis C-arm position data
C-arm symbol
Transferring and receiving position data to or from Artis (Automap)
27
27
28
30
31
32
32
34
35
36
37
39
44
45
46
49
49
50
3 / 76
Editing a preset
Creating a new preset
Deleting a preset
Defining defaults
4 / 76
AX42-010.621.63.01.02
50
52
53
53
55
55
56
57
57
58
58
58
62
63
66
67
69
Operator Manual
syngo InSpace 3D, DynaCT and DynaCT Cardiac have been developed for diagnostic purposes in general and for interventional applications in particular.
3D volume data generated by syngo InSpace 3D, DynaCT or DynaCT Cardiac allow
multidirectional viewing of the reconstructed volume.
syngo InSpace 3D, DynaCT and DynaCT Cardiac data provide an additional image
basis for deciding on the therapeutic procedure to be performed immediately after the examination.
At a later stage, syngo InSpace 3D, DynaCT or DynaCT Cardiac data sets are also
excellently suited for the follow-up control of the interventional procedure.
Note
Diagnosis based on syngo Workplace images alone are not a sufficient basis.
Always use the original scenes on the Artis monitors as primary image source
Note
5 / 76
syngo InSpace 3D
InSpace 3D is the basic technique from Siemens to reconstruct a 3D data set
6 / 76
AX42-010.621.63.01.02
Operator Manual
Introduction
structed volume. Those landmarks can be overlaid with the live fluoro image
using iGuide Toolbox.
See the iPilot/iGuide Operator Manual
The step from high contrast imaging to a soft tissue imaging was possible by the
adaptation of CT techniques to the reconstruction.
The most important point is the increase of the number of projections that are acquired during a rotation around the patient and the application of CT reconstruction algorithms to create the volume.
7 / 76
In order to follow the CT techniques, the angio equipment has to acquire in a short
amount of time as many images as possible. The processing step to create crosssectional images from the projection ones has to be adapted as well.
What is DynaCT?
DynaCT enables selected systems of the Artis family to create images that are CTlike images.
What is
DynaCT Cardiac?
What is ACT?
DynaCT involves Angiographic Computed Tomography (ACT) utilizing images obtained through a rotational angiography run on an Artis system.
Image acquisition is performed with a C-arm rotation. The images are transferred
to the syngo Workplace and the 3D volume set is immediately reconstructed using
InSpace Reconstruction.
No patient transfers
DynaCT delivers soft tissue images that make it possible to visualize objects of
10 HU1 and 10 mm in size with a 16 cm CATPHAN phantom. A quality, that enables you to differentiate soft tissue such as hemorrhage and tumors.
Performing an ACT directly on an Artis system, DynaCT or DynaCT Cardiac eliminates unnecessary patient transfers to other modalities like CT or MRI.
In the event of a complication during an interventional procedure, time is of absolute essence for the safety of the patient. DynaCT allows you to perform ACT directly in the angio lab, thereby further supporting you in your decision-making.
DynaCT offers you additional information at the point of decision. Consequently,
the need to move the patient out of the sterile environment during the interventional procedure is greatly reduced.
Clinical scope
Neurovascular treatment always bears a risk for complications. The extend of a local bleeding as the result of an aneurysm rupture during a coiling procedure, for
example, is not visible on a conventional angiography system. With DynaCT, you
are able to detect the extend of hemorrhaging directly in the angio lab.
DynaCT allows you to visualize the ventricle system of the brain to aid in diagnosing pathologic processes or to guide and monitor placements during drainage procedures.
Capable of high quality examination without the use of contrast agents, DynaCT
also provides diagnostic and interventional support for a broad range of abdominal procedures, including punctures and drainages.
1
8 / 76
AX42-010.621.63.01.02
Operator Manual
Introduction
Note
The basic difference between InSpace 3D imaging and DynaCT is the number of
projection images acquired and the reconstruction technique.
In InSpace 3D, typically up to 140 images are acquired in 4 to 8 seconds.
In DynaCT, typically 300 images or more are acquired in approximately 6 to 20
seconds.
This is also valid for DynaCT Cardiac with ECG gating.
DynaCT allows a better low contrast solution down to 10 HU.
Acquisition and
reconstruction times
A rotational run on the Artis takes about 5 - 20 seconds for InSpace 3D or DynaCT
and 4 - 8 seconds for DynaCT Cardiac.
The reconstruction on the syngo Workplace using InSpace 3D takes about 20 seconds for a high contrast and about 30 sec. to 1 min. (depending on the selected
acquisition program) for a DynaCT reconstruction.
A DynaCT Cardiac reconstruction takes about 30 seconds without ECG gating,
and about two and a half minutes with ECG gating.
DynaCT workflow
A DynaCT acquisition is performed in the same way as a InSpace 3D acquisition:
1. Selection of a DynaCT acquisition program on the Artis.
2. Rotational acquisition on the Artis.
DynaCT will be in most cases a non-subtracted acquisition.
See the Artis Operator Manual
3. Transfer of image data to the syngo Workplace.
4. Slice reconstruction using InSpace Reconstruction 11
5. Visualization using the InSpace or 3D task cards 34
9 / 76
10 / 76
AX42-010.621.63.01.02
Operator Manual
After a 3D acquisition has been finished on the Artis the transfer of the acquired
data is either started automatically or initiated manually. The transfer of the data
is performed in the background.
When the transfer is started, reconstruction starts automatically on the
syngo Workplace using the reconstruction Preset that is linked to the 3D acquisition program on the Artis or the default reconstruction Preset configured on the
syngo Workplace.
Secondary reconstruction
Note
3D raw data (projection images) remain on the syngo Workplace until they are
overwritten by the next transfer from the acquisition system. You can re-process
these data as often as required.
If the security package has been activated, the screen saver must be set to its maximum value (9999 minutes). Thus it will only get activated if the system remains
untouched for more than 6.9 days, which is highly unlikely.
To activate the screen saver, select Options > Configuration from the main
Note
Offline reconstruction
The licenses for syngo InSpace 3D, DynaCT or DynaCT Cardiac (depending
on the 3D acquisition run) are available.
A 3D realtime reconstruction board (Image ProX) is available on the
syngo Workplace.
Valid calibration data for the acquisition system where the 3D acquisition
run originates from are available.
11 / 76
Primary reconstruction
Once an image series has been loaded via transfer from the acquisition system,
several checks are performed and data is processed using default parameters.
Offline reconstruction
If the original 3D acquisition from the acquisition system is stored in the local database on your syngo Workplace and the calibration data is also available, you can
start a reconstruction from the Patient Browser:
Select the original 3D acquisition data in the Patient Browser window.
Click this icon in the tool bar of the Patient Browser window.
Or
Select Patient > Send To 3D in the menu of the Patient Browser window.
12 / 76
AX42-010.621.63.01.02
Operator Manual
Secondary reconstruction
The InSpace Reconstruction window
In the InSpace Reconstruction window two images (two projections differing by
90) of a scene representing the collected 3D data are displayed.
You can select an image pair, select a Preset, choose the reconstruction volume
and change parameters for reconstruction.
(1)
(4)
(5)
(2)
(2)
(6)
(3)
13 / 76
Example
14 / 76
AX42-010.621.63.01.02
Operator Manual
Note
It is possible to window the projection images via the middle mouse button. 24
The windowing is performed onto one image and applied to all images.
Or
Select the VOI Size on the Recon subtask card.
The VOI is marked with green rectangles in the lateral and frontal views.
and a corner.
The cursor changes its shape.
15 / 76
Adjusting a VOI
Move the mouse cursor on the green rectangle on the center of an edge or on
a corner.
The cursor changes its shape.
Or
Select View > Numeric VOI in the menu of the InSpace Reconstruction win-
dow.
16 / 76
Value
Description
LR
PA
HF
Width
Height
AX42-010.621.63.01.02
Operator Manual
After a value has been modified - via the Return key or if the edit field is
looses the focus or by a click on a spin button - the VOI is updated.
The values are set according to the visible rectangles.
If the rectangle is modified, the values are updated.
Note
Or
Select View > Reconstruction in the menu of the InSpace Reconstruction
window.
17 / 76
The following parameter settings can be changed and are stored in a reconstruction preset:
Preset: Reconstruction preset 50
VOI Size: Small, Medium, or Full
Slice Matrix: resolution of the slices of the volume
Kernel Type: reconstruction kernel
18 / 76
AX42-010.621.63.01.02
Operator Manual
The smaller the voxel size (the larger the slice matrix) the better the spatial resolution of the reconstruction and the longer the reconstruction and storage time
for this 3D volume will be.
Note
injection of iodine.
Otherwise, the HU kernel is recommended. Only the HU kernel results in gray
values that are proportional to X-ray absorption. The HU kernel shall be applied
for quantitative measurements, for data sets with metal (clips, coils), and for
DynaCT data.
19 / 76
With the setting Auto, the sharpness of the reconstruction algorithm is automatically adjusted. The Auto setting balances image quality and artifacts. Due
to the reconstruction principle, it is not possible to avoid artifacts completely.
A special algorithm smears out the artifacts in the Smooth setting. Artifacts
artifact level.
Normal is a compromise between Sharp and Smooth.
Very Smooth optimizes the image contrast.
Note
Do not change the Auto setting unless your specific diagnostic problem requires
sharper or smoother image quality.
In the Nat Fill setting, only the fill images are used for reconstruction.
structed and loaded into the InSpace Viewer. This technique, which we call
syngo iDentify, allows to visualize plaque and stents in the best way.
Note
20 / 76
The settings can be selected only, if the series is available subtracted (mask and fill
images).
AX42-010.621.63.01.02
Operator Manual
(Example)
You can use viewing presets for default display (windowing, segmentation, ...)
which are stored in the Preset Gallery.
Viewing presets 33
Resetting to Preset
If you want to discard your changes, you can reset all parameters to the values defined in the corresponding (reconstruction) Preset.
Select Reconstruction > Reset Preset in the menu of the InSpace Reconstruc-
tion window.
Note
Or
21 / 76
Select View > 3D Card in the menu of the InSpace Reconstruction window.
can be specified (0 ... 100% of the cardiac cycle), which will be used for cardiac
gating.
If more than one cardiac phases are specified, the number N of specified cardiac
phases results in N 3D volumes after the subsequent 3D reconstruction process.
The N 3D volumes cover N different cardiac phases and can be visualized as a 4D
volume.
If two cardiac phases have the same percentage only one volume is reconstructed.
If Ungated is checked only one (ungated) volume is reconstructed from the first
run.
The cardiac phase should be within the Acquired Phase.
22 / 76
AX42-010.621.63.01.02
Operator Manual
Changing scale
You can adjust the scale of the X and the Y axis for better viewing.
Drag the respective slider to the required scale percentage.
Relocating R peaks
If a R peak has not been detected correctly:
Click a yellow line.
Or
Select Add in the popup menu (right mouse button).
23 / 76
Or
Select Delete in the popup menu (right mouse button).
After ECG editing the identified R peaks will be the base for the subsequent cardiac
gating process.
Additional functions
You can adopt image display in the InSpace Reconstruction window by changing
the window values of the images.
For documentation of the selected VOI, it is possible to store or film an image.
Windowing images
Click in the required image with the middle mouse key.
Hold the mouse key pressed and move the mouse up or down to change the
window level.
Or
Hold the mouse key pressed and move the mouse to the right or left to change
Or
24 / 76
AX42-010.621.63.01.02
Operator Manual
Select Patient > Save As... in the menu of the InSpace Reconstruction win-
dow.
The Save As window appears.
The image is saved in the original resolution including the VOI graphics.
Or
Select Patient > Copy To Film Sheet in the menu of the InSpace Reconstruc-
tion window.
The image is passed to the virtual film sheet.
Or
Select Patient > Close InSpace Reconstruction in the menu of the InSpace
Reconstruction window.
Note
Once the window has been closed, its status remains available until new images
are transferred (or the syngo Workplace is shut down or restarted). You can return
to InSpace Reconstruction at any time:
In the main menu, select Patient > InSpace Reconstruction.
25 / 76
tion window.
Note
You can use this option during examinations when image data is sent from the acquisition system to the syngo Workplace.
In this case the InSpace Reconstruction window will not appear but the reconstructed image is displayed immediately in the InSpace task card.
Starting reconstruction
Note
Or
Select Reconstruction > Run in the menu of the InSpace Reconstruction win-
dow.
Reconstruction is started.
During reconstruction a progress bar is displayed.
Note
26 / 76
When you click on Cancel it will take the program a few seconds to be ready again.
AX42-010.621.63.01.02
Operator Manual
3D display techniques
There are several techniques of displaying volumes:
MultiPlanar Reconstruction (MPR) is used to place the slices in various orienta-
tions through the region of interest. The resulting images are two-dimensional.
Maximum Intensity Projection (MIP) is used for the display of contrast medium
tween organs and tissue structures, and for colored, three-dimensional display
of bone, tissue, and organs.
(1) Transversal/axial
(2) Coronal
(3) Paraxial
(4) Sagittal
(5) Paraxial
(6) Sagittal
In MPR, two-dimensional images are always calculated from the volume. However, you can move through the representation quickly using the 3D cross-hair or the
mouse to gain a three-dimensional impression.
The smaller the distance between the slices or the larger the degree of overlap of
the original images, the greater the resolution in the examination direction.
27 / 76
The minimum slice thickness of an MPR image is one voxel. But you can generate
thicker slices (MPR Thick). The gray-scale mean values are calculated for voxels located one behind the other.
28 / 76
AX42-010.621.63.01.02
Operator Manual
MIP
The voxel with the maximum intensity is displayed in the MIP image.
You can use this method, for example, for examinations of the lung as the structures with the least intensive gray-scale values.
In the MinIP image the voxels with the minimum intensity are displayed.
In order to gain a three-dimensional impression you can generate several MIP/
MinIP images from different viewing directions or rotate the projection using the
mouse or the 3D cross-hair.
You can limit the volume for a projection to a partial area of the volume data set.
In the output type MIP Thin/MinIP Thin you define which expansion the volume
used for the projection should have.
29 / 76
30 / 76
AX42-010.621.63.01.02
Operator Manual
Note
When image data is loaded into a task card and this data is simultaneously altered
using Correct, a refresh is not performed by the task card. Therefore an inconsistent state can arise.
Do not change image information using Correct while image data is loaded.
Note
31 / 76
Rotating a volume
Activate the Manipulate All Objects mode by clicking this icon, if the mode is
tion.
Or
Click this icon.
Keep the mouse button pressed and rotate the volume.
Note
If you release the mouse button while you are rotating the volume, it will continue
to rotate automatically.
Click on the view to stop the volume rotating.
32 / 76
AX42-010.621.63.01.02
Operator Manual
Note
Viewing presets
Or
Select InSpace > Load Modality Preset... in the main menu.
There you will find several (viewing) presets useful for InSpace 3D / DynaCT /
DynaCT Cardiac images.
33 / 76
Or
Select a segment and then right-click on it.
Recommendation: 3 mm
34 / 76
AX42-010.621.63.01.02
Operator Manual
Further settings:
Disable the Auto-hide MPR lines after delay checkbox if you want to always
simultaneously.
If Maintain orthogonal MPR views is set, changing the orientation of one
MPR plane will also change the other two planes to keep them mutually orthogonal.
If this option is not checked, each MPR plane can be rotated to any desired angulation without affecting the other MPR planes.
Close the dialog box.
Note
The value for MPR image thickness is displayed in the image and always also
filmed and stored.
35 / 76
InSpace EP
Or
A 3D/3D registration has been performed for two 3D data sets.
Please refer to the syngo InSpace 3D/3D Fusion Operator Manual.
36 / 76
AX42-010.621.63.01.02
Operator Manual
Ctrl
37 / 76
38 / 76
AX42-010.621.63.01.02
Operator Manual
39 / 76
You can select the following merging methods of the two volumes:
Addition A+B: Volume A is added to volume B. The two volumes are overlaid
to each other.
Subtraction A-B: Volume B is subtracted from volume A.
Embedded MPR (B+ A MPR): Embed volume A as MPR to volume B.
Clipped (B + A Clipped): The two volumes are overlaid to each other, volume
A is clipped.
Side by Side: The two volumes are displayed side by side in MPR view.
Note
Ensure that the volumes are registered correctly before having confidence to the
ghost cursor of the Side by Side mode!
A Only, B Only: Only one volume is displayed.
Reverse A & B in Merging: Reverse volume A and volume B.
Windowing
For best display, you should adjust image display of the two volumes separately.
For example, you can use a bones preset for the mask images and a vessel preset
for the fill images of a 3D DSA series.
Optimizing the first/second volume
Click Windowing: A Only/B Only.
Or
Select Volumes > [Your first/second volume] in the main menu.
40 / 76
AX42-010.621.63.01.02
Operator Manual
Select InSpace > Load Modality Preset... in the main menu to check/select a
Adjust the ramp for best display (of bones, for example).
41 / 76
42 / 76
AX42-010.621.63.01.02
Operator Manual
Embedded MPR
Click Embedded MPR (B+ A MPR).
43 / 76
You can rotate the view and adjust the MPR plane.
The embedded MPR plane behaves like a clip plane. The thickness of the MPR can
be changed by adjusting the Slab Thickness.
Click this icon on the Clip subtask card with the right mouse button.
Or
Select Type > Set Slab Thickness... from the main menu.
44 / 76
AX42-010.621.63.01.02
Operator Manual
C-arm symbol
When you rotate the image, the C-arm displayed in the top right corner performs
the rotation correspondingly. It moves to all positions, although the C-arm may
not be able to rotate to that position in reality.
In images from biplane systems two C-arms are shown. If plane B is in parking position only the C-arm of plane A is shown.
Angulation data are displayed next to the C-arm (of plane A only).
C-arm display
Description
Biplane configuration:
C-arm of plane A and table, both have blue color:
C-arm position is reachable.
C-arm of plane A has red color and table has blue color:
C-arm position is not reachable.
C-arm of plane B has red color and table has blue color:
C-arm position is not reachable.
45 / 76
C-arm display
Description
Both C-arms have red color:
C-arm position is not reachable.
C-arms collide with each other.
Artis touchscreen
Press this icon.
46 / 76
AX42-010.621.63.01.02
Operator Manual
Note
The patient must be registered in the same position as has been during the acquisition of the 3D volume data set.
Note
If the current 3D image display is not in the possible range of C-arm positions (red
C-arm display) the following situations may occur:
The system tries to move the C-arm near to the required position.
If the required position is too far away, the position data transfer is rejected.
syngo Workplace
Artis touchscreen
Press this icon.
47 / 76
C-arm display
Monitor
icon
Touchscreen
icon
Action
(blue C-arm)
(orange C-arm)
48 / 76
AX42-010.621.63.01.02
Operator Manual
tion window.
Or
9 InSpace task card is on top.
Select Options > Configuration from the main menu.
This window displays a list of all reconstruction presets and shows its parameters.
49 / 76
Note
The number of entries in the list is limited to 12. If the list is full and you want to
create a new entry, you need to delete an entry first.
It is not possible to delete all entries of the list. At least one entry remains.
Note
Editing a preset
Click the required preset line in the InSpace Reconstruction window.
Double-click an entry.
Or
Select an entry and click on Edit Entry.
50 / 76
AX42-010.621.63.01.02
Operator Manual
Configuring Reconstruction
Reconstruction parameters
See Setting parameters for reconstruction 17
Automatic reconstruction
Automatic Reconstruction: Yes
51 / 76
Automatic visualization
Automatic Visualization: Yes
To create a new preset type in the requested preset-name. The other data
fields are initialized with default values for the parameters.
For example:
VOI Size = Medium
Slice Matrix = 512 x 512,
Kernel Type = HU,
Image Characteristics = Auto,
Automatic Visualization = Yes,
Reconstruction Mode = Subtracted
Window1 Center = 0
Window1 Width = 1000
Window2 Center = 0
Window2 Width = 300
Viewing Preset = <first entry in the list of viewing presets>
Click on Apply to store the setting.
52 / 76
AX42-010.621.63.01.02
Operator Manual
Configuring Reconstruction
Deleting a preset
Click the required preset line in the InSpace Reconstruction window.
Click on Delete Entry.
Defining defaults
The InSpace Reconstruction Preset to be used is configured in the 3D DR, 3D
DSA or 3D CARD acquisition program on the Artis.
The Organ Program Number of the InSpace Reconstruction Preset is used as a
reference for this purpose in the Artis Exam Set Editor (3D Recon Preset in
ACQUISITION TYPE).
53 / 76
54 / 76
AX42-010.621.63.01.02
Operator Manual
(Example, configurable)
Operation works in the same way since the Artis Imaging System can be controlled
by pressing the buttons on the touchscreen and deflecting the joystick.
Refer the Artis Operator Manual for more details.
Most of these functions can also be operated from the syngo Workplace.
If necessary, activate the InSpace task card.
55 / 76
(1)
(2)
(1) Touchscreen
(2) Mouse joystick
For your convenience, some important functions have been mapped to buttons
on the touchscreen with slightly different mouse (joystick) operation.
Activation of buttons
Functions of syngo Workplace, which are explicitly available on the touchscreen
can be activated both on the syngo Workplace and the touchscreen control.
The button of a function which is not active has a gray background.
Press, to activate it.
The button of a function which has been activated on syngo Workplace, has a
white background.
Press, to take control in the examination room.
The button of a function which is active in the examination room has a white
56 / 76
AX42-010.621.63.01.02
Operator Manual
Artis touchscreen
Example
Mouse joystick
Using the mouse joystick you can execute all the mouse functions of the
syngo Workplace.
syngo Workplace mouse control
If you want to take control of the syngo Workplace in the examination room:
Press this icon.
Now you can move the mouse pointer on the monitor of the syngo Workplace
57 / 76
double-clicking.
Select Image > Send to 3D in the main menu.
Artis touchscreen
viewed:
All fill images of the acquired scene will be transferred.
If a 3D CARD series is viewed:
58 / 76
AX42-010.621.63.01.02
Operator Manual
Function
Monitor
icon/Key
Touchscreen
icon
Image display
Short description
Zoom/pan volume
No mouse joystick button pressed:
59 / 76
Function
Monitor
icon/Key
Touchscreen
icon
Short description
Adjust brightness/opacity
No mouse joystick button pressed:
Select orientation
Adjust view
60 / 76
AX42-010.621.63.01.02
Operator Manual
Function
Monitor
icon/Key
Touchscreen
icon
Short description
Measurements
61 / 76
Function
Monitor
icon/Key
Touchscreen
icon
C-arm position
Short description
AXIOM Sensis
syngo iPilot
Monitor
icon
Touchscreen
icon
Short description
62 / 76
AX42-010.621.63.01.02
Operator Manual
Function
Monitor
icon
Touchscreen
icon
Short description
63 / 76
syngo Workplace
Artis touchscreen
Press this icon.
64 / 76
AX42-010.621.63.01.02
Operator Manual
syngo Workplace
Image display on the syngo Workplace monitor is switched from single VRT
to MPR/MPR/MPR/VRT display.
To switch back to VRT only display, press the button again.
View
Ant
from anterior
Post
from posterior
Left
from left
Right
from right
Feet
from feet
Head
from head
65 / 76
syngo Workplace
Artis touchscreen
Press this icon.
66 / 76
AX42-010.621.63.01.02
Operator Manual
syngo Workplace
There you will find several (viewing) presets useful for InSpace 3D / DynaCT /
DynaCT Cardiac images.
Artis touchscreen
Press this icon.
Click Apply.
Note
Bookmarks are assigned to the current study and can only be recalled when the
same study is active again.
67 / 76
Saving a bookmark
Display the image in the way you require.
syngo Workplace
Click on this icon (bottom right) using the right mouse button.
Artis touchscreen
Press this icon.
Loading a bookmark
syngo Workplace
Click on this icon (bottom right) using the left mouse button.
Artis touchscreen
Press this icon.
Choose a bookmark.
68 / 76
AX42-010.621.63.01.02
Operator Manual
Click Ok.
Measurements
syngo Workplace
Artis touchscreen
Press this icon.
Using a grid
It is possible to overlay a grid pattern on the 3D image which makes it possible to
estimate distances between points in the volume.
syngo Workplace
Click on this icon.
69 / 76
Artis touchscreen
Press this icon.
A grid is displayed.
Rotating the grid
Deflect the mouse joystick to move the cursor over a grid axis.
Note
Make sure to move the cursor over a grid axis. Otherwise you will rotate the image.
Measuring distances
It is possible to measure the distance between two points in the display plane.
syngo Workplace
To measure the distance between two points in the displayed plane in 2D, i.e. stay
although the object turns:
Click on this icon.
To measure the distance between two points in the displayed plane in 3D, i.e. turn
with the object:
Click on this icon.
70 / 76
AX42-010.621.63.01.02
Operator Manual
Artis touchscreen
To measure the distance between two points in the displayed plane in 3D:
Press this icon.
Deflect the mouse joystick to move the cursor to the first point.
Press the left mouse joystick button to mark the start point of the distance line.
Deflect the mouse joystick to move the cursor to the second point.
Press the left mouse joystick button to mark the second point of the distance
line.
The distance line is drawn in.
The distance is calculated and displayed.
71 / 76
72 / 76
AX42-010.621.63.01.02
Operator Manual
Index
3D
display techniques 27
load and process data 11
A
Acquisition times 9
ACT 8
Additional color display
selecting the image source 57
Adjust To Patient 63
touchscreen 62
Adjusting cardiac phases 21
Angiographic Computed Tomography
(ACT) 8
Application 5
Artis
C-arm position data 44
touchscreen 55
Auto Rotate 32
Automap 46
Automatic reconstruction
configuring 51
Automatic visualization
configuring 52
B
Bookmarks
touchscreen 67
Brightness
touchscreen 60
C
Cardiac 3D 6
workflow 6
Cardiac High Contrast 3D
workflow 7
C-arm position
touchscreen 62
C-arm position data 44
C-arm symbol 45
Clinical scope 8
Closing
reconstruction 25
Configuring
InSpace Reconstruction 49
Cross-sectional image Reconstruction 11
CT techniques 7
Display
of volume data 14
Distance
touchscreen 70
Dual volume display 36
Dual-volume
windowing 40
Dual-Volume Properties 39
DynaCT 8
visualization 34
with MPR Thick 34
with VRT 34
workflow 9
DynaCT Cardiac 8
cardiac phases 21
visualization 35
workflow 10
E
Editing
R peaks 22
Embedded MPR 43
F
Filming
in InSpace Reconstruction 25
Full screen mode
touchscreen 63
G
Grid
touchscreen 69
I
iGuide
touchscreen 62
iGuide Needle Guidance 63
iGuide Toolbox 63
Image
filming in InSpace Reconstruction 25
saving in InSpace Reconstruction 24
Image characteristics
for reconstruction 20
Image display
touchscreen 59
Image reconstruction 11
InSpace
task card 31
InSpace 3D 6
workflow 6
InSpace Reconstruction
configuring 49
window 13
iPilot 62
touchscreen 62
iPilot Dynamic 62
iPilot Live 62
K
Kernel
for reconstruction 19
L
Linked Contour 63
Linked Marker 63
Linked Pointer 63
Loading a bookmark
touchscreen 68
M
Manipulation mode
touchscreen 59
Maximum Intensity Projection
(MIP) 27, 28
Measurements
touchscreen 61, 69
Minimum Intensity Projection (MinIP) 28
MinIP 29
MinIP Thin 29
MIP 29
display technique 27
MIP Thin 29
Modality preset 33
touchscreen 66
Mouse joystick 55, 57
MPR 27, 28
display technique 27
touchscreen 64
MPR Thick 28
DynaCT 34
Multimodality
image source 57
Multiplanar reconstruction (MPR) 27
N
Needle Guidance 63
O
Offline reconstruction 11
performing 12
73 / 76
Opacity
touchscreen 60
Orientation
touchscreen 60, 65
P
Pan volume
touchscreen 59
Position data
transferring 46
Preset editor 49
Preset gallery 33
Primary reconstruction 11
performing 12
R
R peaks
editing 22
Reconstruction 11
closing 25
filming 25
image characteristics 20
parameters 17
saving 24
slice matrix 19
starting 26
subtracted or unsubtracted 20
viewing 27
windowing 24
Reconstruction kernel 19
Reconstruction presets 50
Reconstruction times 9
Reconstruction volume 14
Resetting visualization
touchscreen 60
Rotate volume
touchscreen 59
Rotating a volume 32
S
Saving
in InSpace Reconstruction 24
Saving a bookmark
touchscreen 68
Secondary reconstruction 11, 58
performing 13
Sensis Report
touchscreen 62
Slice matrix
for reconstruction 19
Soft tissue imaging 8
Standard view
touchscreen 65
74 / 76
Subtracted reconstruction 20
syngo Adjust To Patient 63
syngo iDentify 36
syngo iGuide Needle Guidance 63
syngo iGuide Toolbox 63
syngo iPilot 62
T
Task card
InSpace 31
Touchscreen 56
Artis 55
C-arm position 62
function overview 58
image display 59
measurements 61, 69
orientation 60
view 60
Transferring
for secondary reconstruction 58
W
Window
InSpace Reconstruction 13
Window 1 52
Window 2 52
Window values 52
Windowing
dual-volume 40
in InSpace Reconstruction 24
touchscreen 59
Workflow
Cardiac 3D 6
Cardiac High Contrast 3D 7
DynaCT 9
DynaCT Cardiac 10
InSpace 3D 6
Z
Zoom volume
touchscreen 59
U
Unsubtracted reconstruction 20
V
View
touchscreen 60
Viewing
in InSpace task card 27
Viewing bookmarks
touchscreen 67
Viewing preset 33
configuring 52
Visualization 27
of DynaCT Cardiac images 35
of DynaCT images 34
VOI
numerical 16
pre-defined 15
user-defined 15
VOI (Volume Of Interest) 14
Volume 27
Volume data
display 14
Volume Rendering Technique (VRT) 27
Volume rendering technique (VRT) 30
Volume rotation 32
VRT 30
display technique 27
DynaCT 34
touchscreen 64
AX42-010.621.63.01.02
Operator Manual
Notes
75 / 76
Global Siemens
Healthcare Headquarters
Siemens AG
Healthcare Sector
Henkestrasse 127
91052 Erlangen
Germany
Phone: +49 9131 84-0
www.siemens.de/healthcare
www.siemens.de/healthcare
Order No.: AX42-010.621.63.01.02 | Printed in Germany | 04.2009, Siemens AG
Legal Manufacturer
Siemens AG
Wittelsbacherplatz 2
DE-80333 Muenchen
Germany