Enclosure leakage and patient leakage currents are both recommended under this protocol to measured under single
fault condition, earth open circuit (EOC). The rationale behind this is that any problems are likely to be evident under
this condition and it is not improbable that the fault condition may arise when the equipment is in use.
At the foot of the form, it is recorded whether the equipment has passed or failed in the light of the visual inspection
and the electrical safety test results. The date of the test and the identity of the person who performed the test must
also be recorded.
The comments field below the table is a useful feature of any recording system. It allows any observations to be
recorded, for example, of peculiarities of the equipment under test or concerns about test results. The record should be
referred to by the person performing the next test and inspection on the equipment prior to carrying out the inspection
and test.