Siemens provides this documentation as is without the assumption of any liability under
any theory of law.
The content described herein shall be used by qualified personnel who are employed by
Siemens or one of its affiliates or who are otherwise authorized by Siemens or its affiliates
to use such documents.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are not entitled to use this documen
tation without prior written authority.
Copyright
Siemens, 2006 refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 2 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Table of Contents
General information
1.1
1.2
1.3
1.4
1.5
1.6
5
7
7
7
8
9
9
10
10
12
13
13
Installation
14
2.1
14
14
14
14
14
14
16
22
24
25
27
28
28
28
31
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.1 SIREMOBIL and monitor trolley . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.2 ARCADIS Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.3 MODULARIS Uro Plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.4 SONOLINE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.5 Unpacking MODULARIS Variostar . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.6 Docking unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.7 Connection, e.g. to ARCADIS Orbic / ISO C . . . . . . . . . . . . . . . . . .
2.1.8 Unit association - labeling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.9 C-arm unit isocenter cross. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.10 Installing the isocenter cross on the C-Arm. . . . . . . . . . . . . . . . . .
2.1.11 External isocenter cross manufacturer . . . . . . . . . . . . . . . . . . . . .
2.1.12 Radiation release option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.13 MODULARIS Uro Table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1.14 ECG triggering option (VARIOSTAR 101 42 974) . . . . . . . . . . . . .
Startup
3.1
3.2
3.3
Siemens, 2006
Restricted
35
Operating elements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1.1 Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1.2 Individual keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1.3 Membrane keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
C-arm unit and monitor trolley . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
MODULARIS Variostar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.3.1 Checking the Line Voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.3.2 Filling the cooling circuit and coupling circuit . . . . . . . . . . . . . . .
3.3.3 Connecting modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.3.4 Connecting the X-ray C-arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.3.5 Connecting MODULARIS Variostar . . . . . . . . . . . . . . . . . . . . . . . .
3.3.6 Checking/correcting the focus position. . . . . . . . . . . . . . . . . . . . .
3.3.7 Checking the isocenter with X-ray . . . . . . . . . . . . . . . . . . . . . . . .
3.3.8 Shockwave release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SPL1-130.814.20.03.02
02.08
Page 3 of 92
CS SD SP/CO
35
35
36
37
39
40
40
41
42
44
48
54
61
61
MODULARIS Variostar
Table of Contents
3.4
3.5
3.6
3.7
85
61
62
63
63
64
73
74
83
84
85
85
85
87
87
87
88
89
90
91
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 4 of 92
CS SD SP/CO
Siemens, 2006
Restricted
General information 1
GEFAHR / DANGER
Tab. 2
WARNUNG / WARNING
Bei einer Gefahr, die bei Nichtvermeidung zum Tod oder zu einer
schweren Krperverletzung fhren kann.
Tab. 3
VORSICHT / CAUTION
Tab. 4
ACHTUNG / NOTICE
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 5 of 92
CS SD SP/CO
MODULARIS Variostar
1 General information
Tab. 5
HINWEIS / NOTE
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 6 of 92
CS SD SP/CO
Siemens, 2006
Restricted
General information 1
1.2.2
Parts under electrical voltage are accessible when the covers are open. To
avoid danger, disconnect the system from the power supply before opening
the covers. Disconnect the power plug.
If work steps must be performed using electrical power, the general safety in
formation according to TD00-000.860.01... must be observed.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 7 of 92
CS SD SP/CO
MODULARIS Variostar
1 General information
1.2.3
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 8 of 92
CS SD SP/CO
Siemens, 2006
Restricted
General information 1
1.3.1
These measurement values as well as the values determined during start-up can be cop
ied from the test certificate to the acceptance certificate.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 9 of 92
CS SD SP/CO
MODULARIS Variostar
10
1 General information
MODULARIS Variostar
Fig. 1: Variostar
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 10 of 92
CS SD SP/CO
Siemens, 2006
Restricted
General information 1
Fig. 2: Overview of complete system
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 11 of 92
CS SD SP/CO
MODULARIS Variostar
11
12
1 General information
SPL1-130.815.20..
*PG Variostar
SPL1-130.891.20..
SPR2-130.814.05..
SPL1-130.620.01..
SPL1-130.620.20..
n.a.
SPL1-130.061.02..
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 12 of 92
CS SD SP/CO
Siemens, 2006
Restricted
General information 1
PC connection cable, 5 m
99 00 440 RE999
84 28,104 RE999
ESD equipment
44 15 899 RV090
99 94 831
97 06 979 Y0526
1.6.1
Spray can or varnish applicator of the color white grained 99 00 705 RE999
MED surface no. 4146 similar to RAL grayish white 9002.
Required material
n
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 13 of 92
CS SD SP/CO
MODULARIS Variostar
13
14
2 Installation
2.1 Overview
2.1.1
2.1.2
ARCADIS Systems
Refer to the valid installation instructions when installing the ARCADIS system.
2.1.3
2.1.4
SONOLINE
SONOLINE does not require any special installation instructions
Simply unpack the unit according to the "Unpacking procedure" attached to the outside
and attach the accessories including the monitor.
2.1.5
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 14 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Installation 2
NOTICE
Risk of damage
Do not pull on the focus phantom (bracket ( 1/Fig. 3 Page 14)).
Disconnect the yellow transport safety bracket ( 1/Fig. 4 Page 15) that secures the
device from both sides.
Remove all the other transport brackets ( 2/Fig. 4 Page 15) and pull out the wooden
braces from underneath the unit by applying and then releasing the brake
( 3/Fig. 4 Page 15).
Remove the ramp and set it at the rear side, securing it from slippage.
Carefully move the unit off the transport pallet and down the ramp.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 15 of 92
CS SD SP/CO
MODULARIS Variostar
15
16
2 Installation
2.1.6
Docking unit
This chapter first describes mechanical coupling for Variostar in combination with SIRE
MOBIL Compact .
The procedure is the same for all C-arm units. Mechanical installation of the bridge on
the front extension of the C-arm unit must always be adapted. Technical clarification
must be known and in conformance with PG (SPL1-130.890.20...).
For the sake of clarity regarding technical work on all components, the coordinates used
in this document are defined in this diagram: ( Fig. 5 Page 17) (Variostar with shock
wave focus, C-arm unit with isocenter, patient table). For the Variostar and C-arm unit,
the designations "left" and "right" are always from the viewpoint of the operator with the
unit moving forward.
The docking mandrel is pre-installed on the Variostar centrally in relation to the longitudi
nal axis ( 1/Fig. 5 Page 17) and can be adjusted by 30 mm in height (200 to 230 mm
above the floor). The height is determined depending on the installation of the docking
bridge. The docking bridge is installed centrally ( 4 mm) on the front C-arm unit exten
sion ( 2/Fig. 5 Page 17).
The following figures are not displayed to scale, and apply to the Variostar with Sir
emobil ISO C / Compact and the Compact L/ ARCADIS Family. Any other C-arm unit
requires prior clearance by the local project management. The procedure is then
the same.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 16 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Installation 2
Fig. 5: Docking unit
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 17 of 92
CS SD SP/CO
MODULARIS Variostar
17
18
2 Installation
Fig. 7: Docking securing_2
Prepare the Variostar to the left of the C-arm on a level floor (fix C-arm)
Set the C-arm vertically to approx. 1070 mm in the center (( 3/Fig. 5 Page 17)_ Z di
rection)
Remove the top part of the docking bridge.
Unscrew the docking bridge ( 1/Fig. 6 Page 17) or detailed view ( 1/Fig. 7 Page 18)
and pull the docking bridge apart so that the complete adjustment range is available.
Place the bottom part of the docking bridge on the C-arm extension, at least 140 mm
in the X direction ( 1/Fig. 9 Page 19) and centered ( 4 mm).
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 18 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Installation 2
n
Tighten the eight fastening screws ( 1/Fig. 6 Page 17) and ( Fig. 7 Page 18).
Set the Variostar docking mandrel in the horizontal position ( 1/Fig. 5 Page 17).
Secure the bottom part with the Allen screw ( 1/Fig. 11 Page 20).
Position the top part of the docking bridge and secure it ( 1/Fig. 12 Page 21).
Siemens, 2006
Restricted
Test the inward movement and locking of the Variostar. The latch must lock securely.
Correction is performed using the black Allen screw in the latch
( 2/Fig. 12 Page 21).
Finally, perform a rough visual inspection to check whether the central ray (X-ray) can
pass through the focus of the shock wave head.
SPL1-130.814.20.03.02
02.08
Page 19 of 92
CS SD SP/CO
MODULARIS Variostar
19
20
2 Installation
Fig. 10: Variostar connection
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 20 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Installation 2
Fig. 12: Upper part
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 21 of 92
CS SD SP/CO
MODULARIS Variostar
21
22
2 Installation
Fig. 14: Docking bridge
2.1.7
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 22 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Installation 2
For ARCADIS Orbic / ISO C, the docking bridge has to be adapted accordingly
( Fig. 13 Page 21).
n
Attach these blocks to the bridge with 2 screws ( 3/Fig. 13 Page 21).
2.1.7.1
Screw the clamping screws from the enclosed material into the fastening block
( 1 und 2/Fig. 13 Page 21).
If the docking bridge is not fastened with screws, it can be removed whenever necessa
ry. This will prevent the C-arm from colliding with the docking bridge in the event of a
"non-lithotripsy application."
Collision warning labels
When the Variostar is used with the Orbic C-arm, the warning labels included with deliv
ery must be affixed to the docking bridge and the table, as shown in the illustrations.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 23 of 92
CS SD SP/CO
MODULARIS Variostar
23
24
2 Installation
2.1.8
(1)
(2)
(3)
(4)
(5)
MODULARIS Variostar
Variostar_serial_number
Manufacturer and type, C-arm unit
C-arm_unit_serial_number
C-arm_unit_serial_number
Variostar_serial_number
SPL1-130.814.20.03.02
02.08
Page 24 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Installation 2
Fig. 19: Label 2
2.1.9
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 25 of 92
CS SD SP/CO
MODULARIS Variostar
25
26
2 Installation
Fig. 22: I.I. 2
Remove the front screw on the I.I. and replace it with the M4 x 16 screw included in
the shipment ( 1/Fig. 23 Page 26).
The crosshairs must be positioned in the center of the mounting ring. Check this with
a tape measure and correct it if necessary.
Affix the arrows (Figure: ( Fig. 21 Page 25)). Note the centering bolts.
Place the supplied "serial number" labels on the isocenter cross ( 1/Fig. 23 Page 26)
and the I.I.
( 2/Fig. 23 Page 26).
Replace the I.I ring if necessary; see following instructions.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 26 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Installation 2
2.1.10
2.1.10.1
Siemens, 2006
Restricted
Remove grid ( 1/Fig. 24 Page 27) and grid ring ( 2/Fig. 24 Page 27) after removing
the attachment screws ( 3/Fig. 24 Page 27).
Mount the grid in the new grid ring. Note the position of the grid's plastic covers
( 1/Fig. 26 Page 27). Use the existing fasteners ( 2/Fig. 26 Page 27) to secure the
grid.
With the Modularis, the switch located in the grid ring need not be connected.
Set the grid ring on the I.I. and secure it using the screws ( 1/Fig. 25 Page 27) inclu
ded with delivery.
SPL1-130.814.20.03.02
02.08
Page 27 of 92
CS SD SP/CO
MODULARIS Variostar
27
28
2 Installation
2.1.11
2.1.12
Install the movable bracket ( 3/Fig. 23 Page 26) in the crosshair extension.
2.1.13
2.1.13.1
The positioning yoke is packed separately in the shipment (part no. 81 25 333).
Attach the yoke to the table with the two screws as shown in
(( 1/Fig. 27 Page 29)). The attachment blocks must be selected according to the
height of the docking bridge.
The positioning yoke must be adjusted so that its height corresponds to that of the
docking bridge (( 1/Fig. 28 Page 29)). Adjustment can be performed by turning the
two rubber stoppers (( 2/Fig. 28 Page 29)).
The rubber stoppers must be secured by the screw (( 1/Fig. 29 Page 30)).
No mechanical connection is created when operating the system for therapeutic pur
poses. The C-arm is simply docked in the yoke.
In the table park position, the yoke must be collapsed to the table base.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 28 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Installation 2
Fig. 27: Positioning_yoke_1
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 29 of 92
CS SD SP/CO
MODULARIS Variostar
29
30
2 Installation
Fig. 29: Securing screw
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 30 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Installation 2
2.1.14
2.1.14.1
Unscrew the rectangular plate ( 1/Fig. 31 Page 31) from the docking station of the
ECG monitor. It is no longer required. The screws will be used to secure the support
arm.
Unscrew the two covers.
Remove the cover from the system trolley.
Attach the power supply unit and the power supply cable of the ECG monitor in the
system trolley to the support, using cable ties. Insert the non-heating apparatus plug
in the power supply ( Fig. 34 Page 32).
Push the 24 V cable of the power unit through the column from below and route it in
the same way as the trigger cable (using a pull wire if necessary).
Route both cables so that a length of approx. 35 cm is available outside the column.
Screw the plate firmly onto the column so that the cables emerge - as shown in .
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 31 of 92
CS SD SP/CO
MODULARIS Variostar
31
32
2 Installation
Fig. 33: ECG monitor installation 1
Install the ECG monitor (including the interface card) and the mounting plate, in ac
cordance with the ECG monitor instructions included with delivery.
Push the monitor into the holder, as shown in the illustration: ( Fig. 33 Page 32).
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 32 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Installation 2
Fig. 35: ECG monitor installation 2
Compare the ECG monitor input signal to the signal information at the external ECG
monitor.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 33 of 92
CS SD SP/CO
MODULARIS Variostar
33
34
2 Installation
7. Service tab card: System setup
Handheld remote control displays
Display
item
Description
Start potenti
ometer cali
bration (but
ton)
2: see
( 2/Fig. 36
Page 34)
SW update
Insert the SP card with the new soft
ware.
Select the "disk" symbol and follow
the instructions.
3: see
( 3/Fig. 36
Page 34)
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 34 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
3.1.1
Location
Menu guidance in the display is implemented with tab cards whose arrangement and
content are based on the typical procedure of urological treatment.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 35 of 92
CS SD SP/CO
MODULARIS Variostar
35
36
3 Startup
Fig. 38: Display
(1)
(2)
(3)
(4)
3.1.2
Individual keys
The keys under the display are for releasing radiation and shock waves as well as for se
lecting menus in the display.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 36 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 39: Individual keys
(1)
(2)
(3)
(4)
(5)
3.1.3
Membrane keys
Pre-programmed functions for stone localization and treatment are assigned to mem
brane keys.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 37 of 92
CS SD SP/CO
MODULARIS Variostar
37
38
3 Startup
Fig. 40: Membrane keys
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 38 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Perfect functioning must have already been ensured by the responsible start-up engi
neer.
The release from the planning guide must be available from Siemens project manage
ment. Implementation of the combination by the documentation with the dimension
protocol from planning guide SPL1-130.891.01...
NOTICE
Liability for the complete therapy system.
With the SIREMOBIL Compact and ARCADIS Varic, radiation can be triggered from the litho
module control panel. Use instructions SPR2-130.814.05... and tool kit 7140952 (depend
ing on the SIREMOBIL serial number) for this purpose. With non-Siemens C-arms, radia
tion cannot be released from the litho module.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 39 of 92
CS SD SP/CO
MODULARIS Variostar
39
3 Startup
40
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 40 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
3.3.2
Remove the front cover from the system trolley and then open the right cover panel
door (magnetic holder).
The cooling unit can be pulled partially out of the trolley for better access, after the 4
hex screws on the side have been removed.
(wrench size 10).
Check the bleeder valve at the back, and if closed, open (cap) in the cooling unit with
one turn ( 1/Fig. 41 Page 40) (the valve must remain open at all times).
Open the Velcro strap, remove the water reservoir tank, and fill it with approximately
10 liters of clean water.
Set the shockwave head (50) to the undertable position, see ( Fig. 43 Page 41).
The coupling circuit and cooling circuit are filled automatically in sequence (cooling
circuit pump M1 runs the appropriate period of time, V188 on board D3 illuminates).
Filling the unit without creating bubbles takes approximately 5 minutes. During filling,
check the coupling bellows for bubbles. If necessary, remove the bubbles by pressing
on the coupling bellows. If necessary, start the rinsing cycle again.
Once the system is filled completely, check the fill level of the water reservoir. If it is
not filled to the 10 liter mark, add water until it is.
Finally, check the system for any leaks and reinstall the covers.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 41 of 92
CS SD SP/CO
MODULARIS Variostar
41
42
3 Startup
3.3.3
3.3.3.1
Connecting modules
Setting up the patient table
Adhere to the MODULARIS Uro operator manual and the safety information contained
therein (Safety register).
Move the patient table to the desired position in the treatment room.
Connect the patient table to the line voltage and switch it on.
The MODULARIS Uro patient table and its hand-held remote control are ready for
operation.
Preparing coupling
Before coupling the X-ray C-arm, the patient table must be moved sideways via the MOD
ULARIS Uro hand-held remote control.
MODULARIS Uro hand-held remote control:
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 42 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 44: MODULARIS Uro hand-held remote con
trol
(1)
(2)
(3)
(4)
(5)
(6)
Coupling position
Fig. 45: Coupling position
Move the patient table sideways, away from the table cutout.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 43 of 92
CS SD SP/CO
MODULARIS Variostar
43
44
3 Startup
n
3.3.4
3.3.4.1
Observe the instructions in the operator manual of the X-ray C-arm and the safety infor
mation contained therein.
Installing the crosshair disk
Depending on the X-ray C-arm system being used, different crosshair disks for the image
intensifier are used.
CAUTION
The crosshair disk could fall.
Risk of injury!
In the case of LithoShare, only the crosshair disk assigned to the X-ray C-arm in use may
be installed. The serial numbers on the labels on the image intensifier and on the cross
hair disk must match.
With the Lithoreport option (ARCADIS Varic, ARCADIS Orbic) no crosshair disk is installed
because the crosshair is displayed digitally via the imaging system.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 44 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 46: Crosshair disk
Align the triangular marker of the crosshair disk with the one on the image intensifier.
The spring clips of the crosshair disk must snap audibly into the recess in the image
intensifier when pressed.
Siemens, 2006
Restricted
Align the triangular marker on the spring clip of the crosshair disk with the one on the
image intensifier.
Attach the crosshair disk to the image intensifier with the rigid hooks.
Press the spring clip back against the resistance until the crosshair lies fully on the im
age intensifier and then carefully release the spring clip.
Ensure that the crosshair disk is securely in place.
SPL1-130.814.20.03.02
02.08
Page 45 of 92
CS SD SP/CO
MODULARIS Variostar
45
46
3 Startup
3.3.4.2
Position it at approx. 90 with respect to the longitudinal axis of the patient table.
Switching on
n
Connect the X-ray C-arm to the power supply plug of the monitor trolley.
Move the swivel, angulation, and orbital position of the X-ray C-arm to the 0 posi
tion. Lock the brakes again.
When using SIREMOBIL Iso-C, SIREMOBIL Iso-C3D, or ARCADIS Orbic/3D, the triangu
lar marker of the C-arm guide must match the triangular marker on the C-arm ex
actly.
Move fully in the horizontal travel direction and lock the brake.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 46 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
3.3.4.3
CAUTION
Insufficient alignment of the C-arm or coupling of a C-arm with an insufficient Cshape (third-party manufacturer).
Damage to the support arm and the C-arm caused by collision!
Release the foot brakes and move the X-ray C-arm to the adjusting device of the pa
tient table.
Pay attention to exact positioning of the C-arm in relation to the adjusting device. Unfav
orable positioning of the C-arm system and table reduces the safety distances during
lithotripsy operation and can require additional radiation exposure during X-ray localiza
tion operations. The coupling procedure may have to be repeated.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 47 of 92
CS SD SP/CO
MODULARIS Variostar
47
48
3 Startup
Fig. 50: Coupling tunnel
Move the X-ray C-arm and cross-arm to the end stop in the docking tunnel. Use the
adjusting device for fine positioning.
Pay attention to the C-arm height during insertion into the docking tunnel.
Under certain circumstances, it is necessary to raise the ends of the adjusting device so
that they meet the intended position of the docking tunnel.
n
3.3.5
3.3.5.1
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 48 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 51: Preparing Variostar
3.3.5.2
Move the support arm and the shock wave head to the basic position for UT with the
shock wave head in the 50 orbital position.
Check for correct locking in the end positions.
Move the tabletop away from the C-arm and then to the lowest patient entry position.
This prevents damage to the coupling bellows during the coupling procedure.
Siemens, 2006
Restricted
Open the locking lever of the docking tunnel on the X-ray C-arm by turning it counter
clockwise.
SPL1-130.814.20.03.02
02.08
Page 49 of 92
CS SD SP/CO
MODULARIS Variostar
49
50
3 Startup
Fig. 53: Folding out the coupling arm
Open the transparent doors on the front side of the MODULARIS Variostar and fold out
the coupling arm by lifting the lock.
Close the transparent doors.
Use the coupling arm (1) to move the MODULARIS Variostar to the end stop in the
docking tunnel (2) of the X-ray C-arm.
The coupling arm must lock perceptibly into place.
Ensure that the shock wave head does not collide with the tabletop and that the support
arm does not collide with the C-arm.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 50 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 55: Locking
This locks the coupling between the X-ray C-arm and MODULARIS Variostar.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 51 of 92
CS SD SP/CO
MODULARIS Variostar
51
52
3 Startup
n
Lock the locking brake on the MODULARIS Variostar by stepping on the pedal.
3.3.5.3
Press the pedal until the end position is reached (locks in place).
The feet of the device box extend.
When using X-ray C-arms from other manufacturers, MODULARIS Variostar cannot be
connected to the X-ray C-arm. As a result, X-ray images cannot be generated or stored
via the handheld remote control. For this purpose, use the control of the respective Xray C-arm.
Patient table
Fig. 57: Cable connection 1
Coil the cable with the plug marked in gray in the cable compartment of the MODULA
RIS Variostar.
Run the cable through the cable gap and out of the cable compartment.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 52 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 58: Cable connection 2
During lithotripsy operation, this connection allows table movement via the handheld remote control of the MODULARIS Variostar.
Siemens C-arm
Fig. 59: Cable connection 3
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 53 of 92
CS SD SP/CO
MODULARIS Variostar
53
54
3 Startup
n
Run the cable with the plug marked in yellow through the cable gap and out of the
cable compartment of the MODULARIS Variostar.
Connect the cable to the front of the Siemens X-ray C-arm.
3.3.5.4
This connection facilitates the generation and storage of X-ray images via the MOD
ULARIS Variostar handheld remote control.
Additionally, for ARCADIS Varic and SIREMOBIL Compact L, connect the cable marked
in blue for the C-arm angulation motor drive (if present).
CAUTION
Control of the X-ray C-arm and the patient table is activated via the MODULARIS
Variostar hand-held remote control.
Unintentional radiation release, C-arm angulation, or table movement possible!
Make sure that the keys of the hand-held remote control are not pressed inad
vertently.
3.3.6
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 54 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 60: Focus position
3.3.6.1
3.3.6.2
Move the support arm and the shock wave head to the basic position for UT with the
shock wave head in the 50 orbital position.
Press the two sliders on the side of the focus phantom in the direction of the arrow
and fold the focus phantom bracket up.
Allow the focus phantom bracket to snap audibly into place and ensure that it is se
cure.
Performing check/adjustment
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 55 of 92
CS SD SP/CO
MODULARIS Variostar
55
56
3 Startup
Fig. 62: Check adjustment 1
Make sure that the orbital movement of the X-ray C-arm is locked in the 0 position.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 56 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 65: Check adjustment 4
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 57 of 92
CS SD SP/CO
MODULARIS Variostar
57
58
3 Startup
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 58 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 71: Check adjustment 8
CAUTION
Move the X-ray C-arm out of the adjustment position.
Patient subjected to unnecessary shock wave application due to an incorrect fo
cus position!
Focus phantom
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 59 of 92
CS SD SP/CO
MODULARIS Variostar
59
60
3 Startup
Fig. 73: Check adjustment 10
Press the two sliders on the side of the focus phantom in the direction of the arrow
and fold the focus phantom bracket back into the locking position.
X-ray C-arm
Fig. 74: Check adjustment 11
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 60 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
3.3.7
3.3.8
Shockwave release
n
3.3.9
Attach the ECG monitor to the docking station in accordance with the operating in
structions.
Connect the power and signal cables to the back of the ECG monitor.
Connect the ECG electrodes to a test subject or connect the ECG simulator to the
the ECG device.
Siemens, 2006
Restricted
Increase the shock wave energy in steps from the lowest energy level to the highest
energy level and release additional shock waves at every energy level.
SPL1-130.814.20.03.02
02.08
Page 61 of 92
CS SD SP/CO
MODULARIS Variostar
61
62
3 Startup
If the Uro table was not included in the delivery volume of the Variostar, the old table
must be installed - for the second table position - in accordance with the following in
structions; refer to the chapter "Retrofitting MODULARIS Uro (old table)".
The following documents are required to start up the MODULARIS Uro:
Start up the table according to the instructions SPL1-130.814.05.01...
NOTICE
The table may only be moved using the motor if it is in the working position.
This means that the foot lever is in the "braked" position (see the MODULARIS
Uro operating instructions).
Connect the cables between the MODULARIS Variostar and MODULARIS Uro.
Move the table into Litho mode and check the table movements with the MODULARIS
Variostar control panel. The following motorized movements are possible from the
MODULARIS Variostar control panel:
Table movement in the Litho mode (green LED on the table control panel goes on)
e.g. key for longitudinal movement pressed.
Table movement approx. 2-3 mm - stops for approx. 2-3 seconds - and then con
tinues movement without pausing.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 62 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
3.5.1
SONOLINE G20
n
Siemens, 2006
Restricted
Attach the connector adapter ( 1/Fig. 76 Page 63) to the serial interface (RS 232) and
secure the screws.
Install software version V1.0.302 or higher in the SONOLINE G20 (see the SONOLINE
G20 operating instructions, chapter 4).
Enable ultrasound localization (refer to the Chapter "Ultrasound localization").
SPL1-130.814.20.03.02
02.08
Page 63 of 92
CS SD SP/CO
MODULARIS Variostar
63
64
3 Startup
3.5.2
3.5.2.1
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 64 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
n
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 65 of 92
CS SD SP/CO
MODULARIS Variostar
65
66
3 Startup
Fig. 82: Installing the shock wave head
The following section outlines the work steps for the old D3, serial number < 3000.
Fig. 83: Electronics unit D3 D4
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 66 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 84: Controller 2
The retaining pins are already mounted on the board (AS 03).
Plug the additional MUP D3 board from the conversion kit into position
( 1/Fig. 84 Page 67) and screw it in place.
The following section outlines the work steps for the US connection of the new D3 board
(10142950), serial number 3000 and higher.
After the US localization option is installed, no further configuration is necessary. When
communication is taking place between the US unit and the Variostar, the "US" field in the
tableside control unit is active.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 67 of 92
CS SD SP/CO
MODULARIS Variostar
67
68
3 Startup
Fig. 85: D3 board, new
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 68 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 87: D3 connections_2
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 69 of 92
CS SD SP/CO
MODULARIS Variostar
69
70
3 Startup
Fig. 88: D3_Layout
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 70 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
3.5.2.2
3.5.2.3
Attach the "MODULARIS Litho" label (included with shipment) to the monitor of the
ultrasound unit ( 1/Fig. 89 Page 71) (this label is used to identify which ultrasound
unit is functioning with the MODULARIS Variostar).
3.5.2.4
Select the Calibrate menu and confirm with the OK key (center navigation key).
After you follow the instructions on the tableside control unit field, the calibration is
complete.
Check the US isocenter as described in this document in the Start-up chapter.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 71 of 92
CS SD SP/CO
MODULARIS Variostar
71
72
3 Startup
SONOLINE G20
Fig. 90: SONOLINE G20
Screw the connector adapter ( 1/Fig. 90 Page 72) to the serial interface.
Install software version V1.0.302 or higher in the SONOLINE G20 (see the SONOLINE
G20 operating instructions, chapter 4).
Enable ultrasound localization (Chapter: Selecting the MODULARIS cross on the SO
NOLINE G20" in this document).
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 72 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
3.5.3
3.5.3.1
For ultrasound localization, a mount for an ultrasound probe is installed on the shock
wave head.
Press the
Press the
Siemens, 2006
Restricted
key.
key.
Deselect the screensaver on the SONOLINE G20 (see the SONOLINE G20 operating
instructions).
A cross is displayed when a probe is selected.
SPL1-130.814.20.03.02
02.08
Page 73 of 92
CS SD SP/CO
MODULARIS Variostar
73
74
3 Startup
3.5.4
3.5.4.1
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 74 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Only approved probes may be used for the corresponding SONOLINE. When installing
the probe in the holder, tighten the screws equally (see figure). The holder and probe
may not be offset.
n
Install the mount ( 1/Fig. 92 Page 74) (see also operating instructions).
Insert an ultrasound probe (array probe) ( 2/Fig. 93 Page 74) into the mount
( 3/Fig. 92 Page 74). Carefully lay the cable for the probe in the long slot
( 4/Fig. 92 Page 74) and carefully secure it.
Plug in the cable ( 2/Fig. 92 Page 74) and connect the probe to the ultrasound unit.
Establish the line connection (marked orange) between the MODULARIS Variostar and
the ultrasound unit.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 75 of 92
CS SD SP/CO
MODULARIS Variostar
75
76
3 Startup
Fig. 95: Target_Test_2
Set the value ( 1/Fig. 95 Page 76) to 180 with the "image field/zoom" knob.
Slide the ultrasound probe all the way forward to the shock wave head.
Press the "Measurement" key and select the "Set" measurement marker.
Use the track ball to move to the center of the crosshair. A value of 13 mm 1 mm
( 2/Fig. 94 Page 75) must appear in the display ( 2/Fig. 94 Page 75) .
Move the ultrasound probe all the way in the other direction (acoustic signal).
Use the trackball to position the crosshairs on the upper cross, click the "Measure
ment" button and select the "Set" measurement marker.
Press the "Measurement" key and select the "Set" measurement marker.
Use the track ball to move to the center of the crosshair. A value of 151 mm 1 mm
must appear in the display ( 2/Fig. 95 Page 76).
If this is not the case, the adjustment procedure must be repeated for the potentiome
ter in the ultrasound arm or the ultrasound arm must be replaced.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 76 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 96: Phantom 2
The isocenter (spherical echo) must be checked with every probe type and every time a
probe is replaced.
n
Check whether the isocenter phantom ( 1/Fig. 96 Page 77) is filled with water. If not,
remove the screw ( 2/Fig. 96 Page 77) and fill in water with a syringe
( 3/Fig. 96 Page 77). There may not be any air bubbles. Screw in the screw again.
Fill the water bladder with water (about 60 ml ) ( 1/Fig. 97 Page 77):
1.
Siemens, 2006
Restricted
Fill a syringe with distilled water. Remove any air that may be trapped in the sy
ringe.
SPL1-130.814.20.03.02
02.08
Page 77 of 92
CS SD SP/CO
MODULARIS Variostar
77
78
3 Startup
Insert the syringe needle through one of the valves ( 2/Fig. 97 Page 77) and in
ject the contents of the syringe into the water bladder. The valve has a self-closing
membrane.
Repeat the fill operation until the water bladder is completely filled with water.
After filling, there may not be air bubbles in the water bladder. Remove any air
bubbles with the syringe.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 78 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Fig. 99: Ultrasound arm_
Value:
Adjust using:
German
5
( Fig. 98
Page 78)
English
French
Spanish
Edge
Ct
Contorno
Dyn Rng
DR
Cam Din
Each activation of the "Set" key increases the value by +1. The "Select" controller allows
the value to be changed in both directions.
n
Select Zoom.
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 79 of 92
CS SD SP/CO
MODULARIS Variostar
79
80
3 Startup
Button
But
ton
Using the track ball, move the measurement crosshair to the middle of the spherical
echo. If necessary, reduce the brightness with the 2 D-knobs until the spherical echo
pos. 1./( Fig. 100 Page 80) is clearly visible.
Button
key.
Use the track ball to move the "+2" cross along the longitudinal (vertical) axis to the
center of the the target cross.
Press
the
MODULARIS Variostar
key.
SPL1-130.814.20.03.02
02.08
Page 80 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
The value displayed for "D2" must be between 1 and 5 mm. This is only valid for the
default values, see instructions SPL1-130.038.01... .
Press
the
key.
Press
the
Using the track ball move the "+2" cross onto the center of the spherical echo.
Press
the
key.
Using the track ball move the cross into the center of the target cross.
Button
key.
Use the track ball to select the value during image inclination and increase it with
the "Set" key.
key.
German:
Bildneigung Array
French:
English:
Spanish:
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 81 of 92
CS SD SP/CO
MODULARIS Variostar
81
82
3 Startup
-
Press
Save Button.
Check the deviation; if white area is not in the center of the isocenter, repeat the
procedure.
Also check the central position of the spherical echo to the target cross
After the adjustment check the US arm at the second installation position of the
shock wave head. ( 7/Fig. 92 Page 74) . For each probe, check whether the adjust
ment in this position is also OK. Do not make any change to the image tilt adjust
ment.
Following the check, store the parts in the appropriate transport cases.
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 82 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Startup 3
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 83 of 92
CS SD SP/CO
MODULARIS Variostar
83
84
3 Startup
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 84 of 92
CS SD SP/CO
Siemens, 2006
Restricted
4.1.1
NOTICE
Electrical voltage!
Disconnect the MODULARIS Uro from line voltage prior to working on the ta
ble.
4.1.2
If possible, move the table into Litho position and transversely, so that the accessory
rail will be on the floor when the table is turned over.
Connect the cable connector to the new electronics box ( 2/Fig. 101 Page 86).
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 85 of 92
CS SD SP/CO
MODULARIS Variostar
85
86
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 86 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 87 of 92
CS SD SP/CO
MODULARIS Variostar
87
88
When the start-up is complete, perform a backup and then erase the error memory
(e:show error log) (see Service Instructions).
Delete the therapy data (t:show therapy data).
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 88 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Measurement
device:
Measurement
procedure:
Conditions:
The manufacturer of this product requires information and the legislature demands
proofs that a product delivered free of defects from the factory continues to possess
the required and certified quality properties on installation and start-up. It is therefore
absolutely necessary that the installation report with the installation and start-up data
is sent without delay after completion of the work to the address stated on the instal
lation report. The report is available in the system binder.
File the fax for the release for operation in the system binder. If the C-arm unit is not
from Siemens, include the dimension protocol from the PG.
NOTICE
Fill out the protocols.
engineer should confirm that all the values have been correctly measured and
recorded in the measurement protocols by signing his name and the date
(system, generator and IQ measurement protocol).
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 89 of 92
CS SD SP/CO
MODULARIS Variostar
89
90
MODULARIS Variostar
SPL1-130.814.20.03.02
02.08
Page 90 of 92
CS SD SP/CO
Siemens, 2006
Restricted
Siemens, 2006
Restricted
SPL1-130.814.20.03.02
02.08
Page 91 of 92
CS SD SP/CO
MODULARIS Variostar
91
siemens.com/healthcare