Typical Applications
Sensor
- Analytical instrumentation
- High reproducibility
- Vacuum Drying
- Process industry
- Coated versions for improved
chemical resistance
- General foreline vacuum measurement
11
10
V DC
8
Analog Out
7
6
5
4
3
2
1
0
10-5
10-4
10-3
10-2
10-1
100
Pressure
101
102
Characteristic of the THERMOVAC Transmitters TTR 91 (S)N / TTR 96 N and TTR 911 (C)N / TTR 916 N
leybold
39,30
34,10
12,84
22,28
30
115,00
52,29
33,78
DN 16 ISO-KF
DN 16 CF
33,78
10,29
DN 16 CF
1/8" NPT
15,90
16,65
Measuring,
controlling
40,75
Dimensional drawings for the THERMOVAC Transmitters TTR 91 (S)N / TTR 96 N
30,16
51,16
59,48
62,15
30,11
54,48
34,00
56,65
31,90
15,04
15,04
15,04
30
30
30
40,75
9,20
46
32,40
9,20
12,70
6,34
52,85
Dimensional drawings for the THERMOVAC Transmitters TTR 911 (C)N / TTR 916 N
left: with display, middle: with RS 232, right: with EtherCAT
leybold
Technical Data
TTR 91 N
mbar (Torr)
Measurement range
Repeatability (typical)
1)
1)
Power consumption
mbar
MEMS-Pirani
Analog output
Resolution
Impedance
Update rate
Interfaces
MEMS-Pirani
Coated MEMS-Pirani
Thermal conductivity according to Pirani
V DC
9 - 30
< 1.2
Electrical connection
V DC
bit
Hz
TTR 96 N
mbar
Sensor
Measurement principle
Supply voltage
THERMOVAC Transmitter
TTR 91 SN
FCC68 / RJ45
N / A
Set point
2.7 x 10-4 to 1000 (2.0 x 10-4 to 750)
Range
mbar (Torr)
0 2 2
Relay
1 A @ 30 V AC/ DC, resistive load
Relay contact rating
100
Relay contact resistance, max.
m
Relay contact endurance, min.
100 000
1.0 A@ 30 V DC load
2 000 000
0.2 A@ 30 V DC load
Status indicators
Max. cable length
Overpressure limit
m
bar
LED-ring (360)
100
6
C (F)
C (F)
C (F)
% n.c.
0 to 40 (32 to 104)
-20 to +65 (-4 to 149)
85 (185), non-operating
95
Installation orientation
Any
2.80
170
Protection class
40
CE certification
Controller type
1)
Accuracy and repeatability are typical values measured with Nitrogen gas at ambient temperature after zero adjustment
leybold
Technical Data
THERMOVAC Transmitter
TTR 911 (C/S)N
TTR 916 N
mbar (Torr)
Repeatability (typical)
1)
1)
mbar
mbar
911 N: MEMS-Pirani
Coated MEMS-Pirani
911 NC: Coated MEMS-Pirani
Sensor
Measurement principle
Supply voltage
Power consumption
V DC
9 - 30
< 1.2
FCC68 / RJ45
Electrical connection
Analog output
Resolution
Impedance
Update rate
V DC
bit
Hz
Set point
Range
mbar (Torr)
Relay
Relay contact rating
Relay contact resistance, max.
m
Relay contact endurance, min.
1.0 A@ 30 V DC load
0.2 A@ 30 V DC load
1.0 x 10-4 to 1000 (0.75 x 10-5 to 750) [Profibus]
2.7 x 10-4 to 1000 (2.0 x 10-4 to 750) [RS 232 / EtherCAT / Display]
2 [Profibus / EtherCAT / Display] / 3 [RS 232]
1 A @ 30 V AC/ DC, resistive load
100
100 000
2 000 000
LED-ring (360)
Status indicators
Overpressure limit
Vout = log10 (Pmbar) x 1.286 + 6.143
1.1 to 10
16
100
16
Interfaces
100
m
6
bar
0 to 40 (32 to 104)
C (F)
C (F)
85 (185), non-operating
% n.c.
0 - 95
Any
Installation orientation
Measuring,
controlling
Measurement range
304 stainless steel, Silicon, SiO2, Si3N4, Tin, Gold, Viton, low out gassing epoxy resin
304 stainless steel, Viton, Parylene-HT
2.80
cm3
40
IP
CE certification
Controller type
1)
Accuracy and repeatability are typical values measured with Nitrogen gas at ambient temperature after zero adjustment
leybold
Ordering Information
THERMOVAC Transmitter
TTR 91 (S)N / TTR 96 N
Part No.
230035V02
230038V02
230036V02
230037V02
230040V02
230043V02
230045V02
230047V02
see chapter Miscellaneous, para. Oerlikon Leybold Vacuum Calibration Service
Type A
124 26
230012
12427
12428
12429
12431
12432
12433
Accessories, optional
Spiral tube DN 16 ISO-KF
230082
leybold
Ordering Information
THERMOVAC Transmitter
TTR 911 (C/S)N
TTR 916 N
Part No.
TTR 911 N, DN 16 ISO-KF, EtherCAT
230700V02
230701V02
89660V02
89658V02
89654V02
89656V02
Calibration
230650V02
230651V02
See Section Miscellaneous, paragraph Oerlikon Leybold Vacuum calibration service
Type A
124 26
230012
12427
12428
12429
12431
12432
12433
Optional accessories
Spiral tube DN 16 ISO-KF
230 082
leybold
Measuring,
controlling
THERMOVAC Transmitters
TTR 101 (S)N
Typical Applications
Sensor
The THERMOVAC TTR 101 N transmitters can be used in any application that
requires absolute pressure measurement and switching capabilities.
- Analytical Instrumentation
- Autozero of Piezo
- Vacuum Drying
- Coating
leybold
30,16
51,16
59,48
39,30
62,15
56,65
54,48
34,10
15,04
30
12,84
15,04
52,29
30
DN 16 ISO-KF
32,40
DN 16 ISO-KF
12,70
1/8" NPT
15,90
54,49
30
12,70
9,20
1/8" NPT
12,70
6,34
16,65
46
52,85
40,75
Dimensional drawing for theTHERMOVAC Transmitters TTR 101 (S) N (left) and TTR 101 N Display (right)
34,00
31,90
15,04
7
6
5
4
3
30
40,75
2
1
0
10-5
10-4
10-3
10-2
10-1
100
Pressure
101
102
leybold
Measuring,
controlling
30,11
11
10
V DC
8
Analog Out
9,20
Technical Data
THERMOVAC Transmitter
TTR 101 (S)N
Measurement range
mbar (Torr)
1)
mbar
Repeatability (typical)
1)
mbar
Power consumption
V DC
1x FFC68 or 2x FCC68 + 1x D-Sub 15 PIN
Electrical connection
Analog output
Resolution
Impedance
Update rate
1 x 10-3 to 11 2 % of reading
1 x 10-3 to 100 2 % of reading [RS 232 (Pirani) / EtherCAT (Pirani)]
10 to 1067 0.2 % of reading / EtherCAT (Piezo)
11 to 1067 0.2 % of reading
Sensor
101 (S)N
101 N, EtherCAT
101 N, Display
101 N, Profibus
101 N, RS 232
Measurement principle
Supply voltage
V DC
bit
Hz
RS 232 / Display / EtherCAT / Profibus
Interfaces
Set point
Range
mbar (Torr) 2.7 x 10-4 to 1000 (2.7 x 10-4 to 750) / 1.0 x 10-4 to 1000 (0.75 x 10-4 to 750) [RS 232 / Display]
2 / 3 [RS 232]
Relay
1 A @ 30 V AC/ DC, resistive load / [EtherCAT: N/A]
Relay contact rating
Relay contact resistance, max.
m
100 / [EtherCAT: N/A]
Relay contact endurance, min.
100 000 / [EtherCAT: N/A]
1.0 A@ 30 V DC load
2 000 000 / [EtherCAT: N/A]
0.2 A@ 30 V DC load
LED-ring (360)
Status indicators
Max. cable length
Overpressure limit
100
m
2
bar
0 to 40 (32 to 104)
C (F)
C (F)
85 (185), non-operating
% n.c.
0 - 95
Any
Installation orientation
Materials exposed to vacuum for Part-Nos.
230350V02, 230351V02, 230352V02, 230353V02,
230354V02, 230355V02, 230356V02, 230358V02,
230366V02, 230702V02
304 stainless steel, Silicon, SiO2, Si3N4, Tin, Gold, Viton, low out gassing epoxy resin
2.8
Dead volume (DN 16 ISO-KF), approx. cm3
Weight (DN 16 ISO-KF)
Protection class
g 168
40
IP
CE certification
Controller type
1)
Accuracy and repeatability are typical values measured with Nitrogen gas at ambient temperature after zero adjustment
leybold
Ordering Information
THERMOVAC Transmitter
TTR 101 (S)N
Part No.
230350V02
230351V02
230352V02
230353V02
230354V02
230355V02
230356V02
230366V02
230370V02
Replacement sensor
DN 16 ISO-KF
1/8" NPT
230361V02
230362V02
Calibration
883 96
See Section Miscellaneous, paragraph Oerlikon Leybold Vacuum calibration service
Type A
124 26
230012
12427
12428
12429
12431
12432
12433
Optional accessories
Spiral tube DN 16 ISO-KF
230 082
leybold
Measuring,
controlling
10
Loadlock Transmitter
THERMOVAC TTR 200 N
PENNINGVAC PTR 200 N
The TTR 200 N and PTR 200 N combine different measurement technologies in one housing making them the
perfect gauges for load lock applications. The transmitters offer a wide
absolute measurement range and a differential range of -1013 to 1013 mbar
(relative to ambient pressure). The highly accurate differential sensor is ideal
for loadlock control since it is insensitive to changes in ambient pressure
conditions. Efficient loadlock control will
improve throughput due to reduced
cycle time.
THERMOVAC Transmitter TTR 200 N (left) and PENNINGVAC Transmitter PTR 200 N (right)
Typical Applications
Option
11
leybold
30,11
30,11
85,00
82,90
82,90
10,80
15,04
40
30
40,75
40,75
Measuring,
controlling
9,20
9,20
11
10
V DC
8
Analog Out
7
6
5
4
3
2
1
0
10-5
10-4
10-3
10-2
10-1
100
Pressure
101
102
Analog Out
9
V DC
7
6
5
4
3
2
1
0
10-9
10-8
10-7
10-6
10-5
10-4
Pressure
10-3
10-2
10-1
100
101
102
leybold
12
Loadlock Transmitter
TTR 200 N
PTR 200 N
Technical Data
Measurement range (N2 and air) mbar (Torr) 5.0 x 10-5 to 1500 (3.75 x 10-5 to 1125)
absolute
mbar
5 x 10-5 to 1500 [RS 232]
(Torr)
(3.75 x 10-5 to 1125) [RS 232]
differential
mbar
-1013 to 1013 [RS 232]
(Torr)
(-760 to 760) [RS 232]
Measurement uncertainty (typical)
Cold cathode
MEMS-Pirani
Diff. Piezo
Repeatability (typical)
Penning
MEMS-Pirani
Diff. Piezo
1)
mbar
1 x 10-8 to 1 x 10-3 30 % of reading
1 x 10-4 to 1 x 10-3 10 % of reading
1 x 10-3 to 100 5 % of reading
100 to atm. 25 % of reading
mbar 1 x 10-4 to 1 x 10-3 10 % of reading
1 x 10-4 to 1 x 10-3 10 % of reading
1 x 10-3 to 100 5 % of reading
1 x 10-3 to 100 5 % of reading
100 to atm. 25 % of reading
100 to atm. 25 % of reading
mbar -10 to 10 10 % of reading 0.67 mbar - 10 to 10 10 % of reading 0.67 mbar
-100 to -10 8 % of reading
-100 to -10 8 % of reading
-1013 to -100 1 % of reading
-1013 to -100 1 % of reading
10 to 100 5 % of reading
10 to 100 5 % of reading
1)
mbar
mbar
mbar
MEMS-Pirani and Diff. Piezo Cold cathode, MEMS-Pirani and Diff. Piezo
Thermal conductivity, combined with Piezo Thermal conductivity, combined with Piezo
and cold cathode ionization
Sensor
Measurement principle
Supply voltage
V DC
9 - 30
9 - 30
Power consumption
Electrical connection
Analog output
Resolution
Impedance
Update rate
Interfaces
V DC
bit
Hz
D-Sub 15 pin
RS 232
RS 232
Set point
Range
Absolute
mbar (Torr) 1 x 10-4 to 1333 (0.75 x 10-4 to 1000) 1 x 10-8 to 1333 (0.75 x 10-8 to 1000)
-1013 to 133 (-775 to 100) -1013 to 133 (-775 to 100)
Differential
mbar (Torr)
3 3
Relay
1 A @ 30 V AC/ DC, resistive load
1 A @ 30 V AC/ DC, resistive load
Relay contact rating
100
100
Relay contact resistance, max.
m
Relay contact endurance, min.
100 000
100 000
1.0 A@ 30 V DC load
2 000 000
2 000 000
0.2 A@ 30 V DC load
Status indicators
Max. cable length
Overpressure limit
bar 2 2
C (F)
C (F)
0 to 40 (32 to 104)
C (F)
85 (185), non-operating
% n.c.
0 - 95
Installation orientation
304 stainless steel, Silicon,
SiO2, Si3N4, Tin, Gold, Viton,
low out gassing epoxy resin
0 - 95
Any Any
304 stainless steel, 403 stainless steel,
Ceramic (Al2O3), Silicon, SiO2,
Si3N4, Tin, Gold, Viton, Titanium,
low out gassing epoxy resin
2.8 28.6
IP 40 40
Controller type
13
85 (185), non-operating
g 305 321
CE certification
1)
0 to 40 (32 to 104)
Accuracy and repeatability are typical values measured with Nitrogen gas at ambient temperature after zero adjustment
leybold
Ordering Information
Loadlock Transmitter
TTR 200 N
PTR 200 N
Part No.
Part No.
230365V02
230087V02
EK16991V02
20028711V02
230 078
883 96
Measuring,
controlling
leybold