Infralab 710 Setup Procedure


ThissetupprocedureassumesthatthetargetcomputerisaDellOptiplexGX260,
runningwindowsXPservicepack1,ifthecomputerisdifferent,pleasecheckwith
thesoftwaredepartment,beforecontinuing.
TofollowthisprocedurehaveathandSoftwareInstallationCD108/1294301,
ApplicationsCD108/1328901andtheUserGuidePDFfileasdefinebytheorder
forexample108/1317001(tobacco),108/1322901(food).
AlsoensurethatthetwoupperswitchedontheIlabmotherboardareswitchedtothe
righthandsideasviewedfromthefront,labelledRUN.
Forinformation,theschematicbelow,showsthecommunicationsroutes/baudrates
betweenthevariousboardsintheInfralabsampler.
Ethernet

RS232

RS232
19200 baud

Panel

Ethernet/RS2
32 converter

RS232
9600 baud

RS232
9600 baud

710 Gauge

Run

Motherboard

Communication schematic.

PC Software
1. Unpackthecomputerandconnectitaccordingtothemanufacturers
guidelines,usetheopticalvideocable(whiteconnectorsontheend)to
ensurethemonitorgivestheoptimumpicture.
2. Powerupthecomputer.
3. Whenthelicenseagreementscreenappears,pressanykeytoacceptit.
4. OnthewindowsXPwelcomescreen,pressnexttobegintheinstallation.
5. Selectthecorrectcountryoptionsandselectnext.
6. Onthesecondlicenseagreementpage,selectyestoacceptandthenpress
next.
7. EnteracomputernameofINFRALAB710(notethisisallinuppercaseand
therearenospaces),clicknexttomovetothenextpage.

Infralab710SetupProcedure

4
Version:
TS108/1325101
DocumentID:
NigelHooper
Author:
InfraredEngineeringLimited

Issued:1/11/2003
Printed:04/10/2016
Page2of5

8. Enteranadministratorpasswordofmaldon(notethisisallinlowercase)and
clicknexttomovetothenextpage.
9. Skiptheinternetdetection,bypressingtheskipbuttonatthebottomrightof
thepage.
10. OntheregisterwithMicrosoftpage,selectnonotatthistime(notethiswill
needtobedonelaterastheuseronlyhas30daystoactivatethesoftware).
11. Onthepagethatasksfortheyourname,enterinfralab(notethisisallin
lowercase),clicknextandthenfinish.

PC Configuration
1. ConnecttheEthernetcabletothePCandtheAnalyserandswitchthe
Analyseron.
2. Fromthestartmenu,selectcontrolpanel>networkandInternet
connections.
3. Selectnetworkconnections.
4. Rightclickonlocalareaconnectionandselectproperties.
5. SelectInternetProtocol(TCP/IP),selectproperties.
6. SelectUsethefollowingipaddress
7. EnteranIPaddressof10.32.15.5
8. Enterasubnetmaskof255.255.240.0
9. Pressokandwaitforthedialogtoclose
10. ClosetheLocalAreaConnectionPropertiesDialog.
11. Fromthestartmenu,selectrun,typecmdandpressok
12. IntheDOSpromptthatappearsenteripconfigandpressenter.
13. Ensurethattheipaddressandsubnetmaskareshownasabove.
14. ClosetheDOSprompt.
15. Restartthepc.

Analyser Ethernet Configuration


1. FromtheIlabinstallCD,installthexportsoftware,runsetup.exeandwaitfor
ittofinish.
2. RunXPortInstallerSetup.msiandselectRepairXportInstaller
3. EnsurethattheEthernetcableispluggedintoboththeanalyserandthepcand
thattheanalyserison.
4. OpenthexportsoftwarefromthestartmenuselectXPortInstaller>
XPortInstaller>XPortInstaller
5. Maximisethexportsoftware
6. Clicksearchtofindthexportfortheanalyser(ignoretheerrormessagebox
thatappears).
7. Oncethedevicehasbeenfound,makeanoteofthehardwareaddressthis
isinthecolumntotherightoftheIPAddress(Ideallythehardwareaddress

Infralab710SetupProcedure

4
Version:
TS108/1325101
DocumentID:
NigelHooper
Author:
InfraredEngineeringLimited

Issued:1/11/2003
Printed:04/10/2016
Page3of5

shouldbeloggedagainstthegaugeserialnumbersomewhereintheorder
istory?).
8. Leftclickonthedevice.
9. SelectassignIPenterthefollowingvalues10.32.15.4andwaitforthedevice
toreboot.
10. Selectthedevicebyleftclickingonitagain.
11. Intherighthandcolumn(undernetwork),enterasubnetof255.255.240.0
12. Intherighthandcolumn(underOEMConfigurablepins),Changepin2to
DCD.
13. Selectupdatefromthetoolbarandwaitforthedevicetoreboot.
14. Closethexportsoftware.
15. Fromthestartmenu,selectrun,typecmdandpressok
16. IntheDOSpromptthatappears,typeping10.32.15.4
17. Ensurethatthedevicewaspingedcorrectlyitshouldsay0%lossonthe3rd
frombottomline,(ifitdoesntturntheanalyseronandoffandtryagain).
18. ClosetheDOSprompt.
19. OpenupInternetexplorerbyrunningtheshortcutfromthedesktop.
20. Typethevalue10.32.15.4intotheaddressbarandpressenter(thiswillshow
thewebpagerunninginthexport).
21. Oncethewebpagehasfinishedloading,selectspeedandchangeitto19200,
selectupdatesettingsandwaitforthedevicetoreboot.
22. CloseInternetExplorer.

Installing the Infralab 710 Software


1. InstallacrobatreaderfromtheInfralab710CDusingthedefaultoptions.
2. InstalltheInfralab710software,fromthesetup.exefileusingthedefault
options,butonthelastpage(aftertheprogressbarshavefinished),removethe
tickfromtheoptionthatsaysLaunchInfralab.
3. CreateashortcutfortheInfralab710softwareinthestartupdirectoryforall
userscopytheshortcutfromthedesktoptoC:\DocumentsandSettings\All
Users\StartMenu\programs\startup\.
4. Restartthecomputeritshouldautologintotheinfralabuseraccountand
automaticallyruntheInfralab710software.

Initial configuration of the Infralab 710 Software


1. IftheInfralabsoftwareisnotalreadystarted,startitviatheshortcutonthe
desktop.
2. Openupthecore,byclickingontheIElogointhesystemtray(bytheclock).
3. Fromtheserversmenu,selectE7Server>configure.

Infralab710SetupProcedure

4
Version:
TS108/1325101
DocumentID:
NigelHooper
Author:
InfraredEngineeringLimited

Issued:1/11/2003
Printed:04/10/2016
Page4of5

4. Intheboxesbelownewsettings,typeintheIPaddressoftheanalyser
10.32.15.4(usetherightarrowkeytomovetothenextbox)
5. ClickOK.
6. Inthewarningboxthatpopsup,clickyes.
7. Fromtheserversmenu,selectE7Server>close
8. Fromtheserversmenu,selectAllServers>E7EthernetServer>Start
andconnect
9. Closethecoreviathecrossatthetoprightofthedialog.
10. FromthemenuintheInfralabsoftwareselectTools>ConnecttoAnalyser.
11. Inthedropdownbox,selecttheE7GaugeServer.
12. WaitforanumbertoappearintheboxbelowtheE7GaugeServerandthen
selectitbyleftclickingonit.
13. Presstheconnectbutton(thiswillconfigurethesoftwaretoautoconnectto
thisanalyserinthefuture).
14. Oncethesoftwarehassuccessfullyconnectedtotheanalyser,thebuttonon
thetoolbarwillchangefromNotReadytoshowingapictureoftheanalyser
andthebuttonatthebottomleftofthescreenwillchangefromDisconnected
toConnected.
15. ClosetheInfralabsoftwareandthenrestartthecomputer.
16. OncetheInfralabsoftwarehasfinishedloading,itshouldhaveautomatically
connectedtotheAnalyser(theirshouldbeapictureoftheanalyserinthe
toolbar).

Applications Installer
TheapplicationsinstallerisusedtoconfiguretheInfralabsoftwareforagiven
application(egmoistureintobacco).
1. IftheInfralab710softwareisrunningcloseitviathecrossatthetopright
handcorner.
2. FromtheinstallCDcopythefile\ApplicationsInstaller\IlabSetup.exe
ontothedesktop.NextrunIlabSetup.exefromthedesktop.
3. InserttheCDwiththeUserguideandfromthefirstdropdownbox/browser
selectthecorrectversionofthemanual(egtobaccoorfood)fromtheCDas
givenbytheorder.
4. Next,inserttheapplicationCDandfromthealgorithmsdropdown
box/browser,selectthedatabasefromtheCDthatmatchesthenameofthenvl
filerunningintheAnalyserasgivenfromtheorder.
5. ClickInstall.
6. Oncetheinstallhasfinished,clickyestoexit.
7. OpentheInfralabsoftwareagain.
8. FromthefilemenuselectRestoreDatabase
9. Enterthelevel1password(thedefaultis1111)andclickok.

Infralab710SetupProcedure

4
Version:
TS108/1325101
DocumentID:
NigelHooper
Author:
InfraredEngineeringLimited

Issued:1/11/2003
Printed:04/10/2016
Page5of5

10. Ensuredirectoryisselectedontherestoredialogandclickok.
11. BrowsetotheApplicationsCDandselecttheBackupsdirectory.
12. Selectthecorrectdirectoryfortheorder,egiftheanalyseristomeasure
Moisture,NicotineandSugarinTobacco,selecttheMNSinTobacco
directory.
13. Selectdisk1,pressokandwaitfortherestoredialogtodisappear.
14. Thesoftwareshouldnowbeconfiguredcorrectly.

Language Selection
ThissectiononlyappliestoorderswherealanguageotherthanEnglishhasbeen
specified.
1. Ifthesoftwareisnotalreadyrunning,startitbydoubleclickingonthe
shortcutonthedesktop.
2. FromtheToolsmenuselectoptions.
3. FromthedropdownboxontheGeneraltabandselectthecorrectlanguage.
4. Pressoktoapplythechangesandclosethedialog.
5. Thesoftwareshouldnowberunninginthenewlanguage,youcanconfirm
thisbylookingatthefilemenu.