CALIBRATION
Each pipette has been checked & calibrated
at factory with procedure conforming to ENISO 8655 standards. It is recommended to
check the calibration at least once a year, for
regularly used pipette.
Important Notes
TROUBLE SHOOTING
0
Trouble
Possible Reason
Correction
Droplets left
inside the tip
Change the
tip cone
Attach firmly.
Unsuitable tip.
Recalibration
Clean the
tip cone.
Change the
O-ring.
Incorrect operation.
Follow instructions
carefully.
Calibration altered.
Recalibrate according
to instructions.
Instrument damaged.
Leakage or
Pipetted volume
too small
Checking calibration
1
Volume Variation
permitted
Volume
Piston contaminated.
Push button
jammed or
moves erratically
Tip ejector
jammed or
moves erratically.
Tip cone
contaminated
from outside.
Remove ejector
collar and clean tip
cones outer surface
with ethanol.
Variation
permitted
1 l
0.15 l
50 l 0.80 l
2 l
0.20 l
100 l 1.50 l
5 l
0.30 l
200 l 2.00 l
10 l
0.30 l
500 l 5.00 l
20 l
0.40 l
1000 l 10.00 l
STORAGE
When not in use, it is recommended that your
pipettor is stored in a vertical position.
Leaving the pipette on its side can cause
liquids to leak into the body of the pipette and
cause corrosion.
PIPETTE DESCRIPTION
These pipettes are available as fixed or
variable, general purpose micropipettes for
sampling and dispensing accurate
amounts of liquid in micro volumes.
PIPETTING TECHNIQUE
A. Forward Pipetting
PIPETTING RECOMMENDATIONS
Aspirate liquid into the pipette only
when a tip is attached to its tipcone.
Increment
0.01 l
0.1 l
0.5 l
1.0 l
5.0 l
0.5 l
5.0 l
0.1 l
1.0 l
50.0 l
200 l
250 l
500 l
1000 l
2000 l
2500 l
5000 l
Increment
0.1 l
0.5 l
0.5 l
1.0 l
Increment
0.1 l
0.5 l
0.5 l
1.0 l
Volume
B. Reverse Pipetting
MAINTENANCE
In-House Maintenance
1. Hold down the
tip ejector.
2. Place the tooth
of the opening
tool between the
tip ejector and
the tip ejector
collar to release
the locking
mechanism.
3. C a r e f u l l y
release the tip
ejector and
remove the
ejector collar.
4. P l a c e t h e
wrench end of
the opening
tool over the tip
cone and turn it
anti-clockwise.