Sie sind auf Seite 1von 4

Uncertainty Analysis

Parameter:

0-10 Length/Diameter Measure

Method:

Direct

Standards:

ATL1584

A7177

Date:

3/11/2008

UUT:
Range:

54-0150
0-1"

This anlaysis is based on Type B analysis of the parameter in question.


Any historical Type A or pooled uncertainties that may apply will also be included and noted.
Uncertainty Model:

Lu=Su+Ru+Eu+Sm

Measurement Model:

La=Li

Lu
Su
Ru
Eu
Sm

Length Uncertainty
Standard Uncertainty
Resolution Uncertainty
Environmental Uncertainty *
Stability of Measurements

La

Length applied

Li

Length andicated

Uncertainty Analysis Table


Lab Env is 20 deg 1 deg C
Symbol

Estimate

Type

Xi
Sm
Su
Ru
Eu
uc
U

xi
0.E+00
0
0
1

A
B
B
B

Inches

Std. Dev

Degree of Freedom:

Probability
Distribution

Divisor

0.000E+00
In
Normal
1.000
6.100E-06
In
Rec
1.732
2.000E-05
In
Rec
3
5.775E-06
In
U
1.414
Combined Standard Uncertainty (1 Sigma) in Inches
Expanded Uncertainty (k=2) in Inches
Uncertainty:

Sensitivity
coefficient
ci
1
1
1
1

Uncertainty
contribution ui(y)
0.00E+00
3.52E-06
6.67E-06
4.08E-06
8.5749E-06
1.7150E-05

1.71E-05 Inches

Welch-Satterwaite
2245.77754
Uc:
5.40649E-021
Ua:
0
Ub1:
1.53861E-022
Ub2:
1.97531E-021
Ub3:
2.78234E-022

1.00000
1.00000
1.00000
1.00000
1.00000
1.00000
1.00000
1.00000
1.00000
1.00000

Stundent's T:
Mean:

Units
Value

1.00

Observations:
1
2
3
4
5
6
7
8
9
10

Contribution

Infinite

2.00

at 95.45% CL

1.000
0.0000E+00

Atmospheric Pressure/humidity noted sources of uncertainty but found to be negligible in this analysis
Flatness/Parellelism/Bilateral deviation verified with optical flats and found to be of negligible values to include in budget.
User error not found to be a contributing factor for type B uncertainty.
Ru= 10uIn resolution uncertainty + 10uIn paralax error, Divisor of 3 is used for a 1.732 Rec distrubution accounted for twice.
Su= 4.2 +1.9L uIn for Gage blocks
Eu=3^(Ucte(Ut))^3+(Dcte(Ut))^3+(CTE(Ut))^3 *
CTE Coefficient of Thermal Exansion
Ucte Uncertainty of CTE
Dcte Difference of CTE between Gage and Mic
Ut Uncertainty of Temperature

ASME/ANSI Y14.5M 1.4(k) states all temperature correction factors should be referenced to 20 C
is 11.5 for AISI 1050 steel and will be compensated for.
Mitutoyo Certifies compliance to 1050 steel throughout 90% of material(11.5 10 %), = 2.3uIn/In/Deg C max error
ASME B89.1.9 states the CTE of steel should be between 10.5 and 12.5 = 2uIn/in/Deg C
2% capped at 25 deg C resulting in 0.5 deg

Uncertainty Analysis
Parameter:

0-10 Length/Diameter Measure

Method:

Direct

Standards:

ATL1584

A7177

Date:

3/11/2008

UUT:
Range:

54-0150
1-5"

This anlaysis is based on Type B analysis of the parameter in question.


Any historical Type A or pooled uncertainties that may apply will also be included and noted.
Uncertainty Model:

Lu=Su+Ru+Eu+Sm

Measurement Model:

La=Li

Lu
Su
Ru
Eu
Sm

Length Uncertainty
Standard Uncertainty
Resolution Uncertainty
Environmental Uncertainty *
Stability of Measurements

La

Length applied

Li

Length andicated

Uncertainty Analysis Table


Symbol

Estimate

Type

Xi
Sm
Su
Ru
Eu
uc
U

xi
0.E+00
0
0
1

A
B
B
B

Inches

Std. Dev

Degree of Freedom:

Probability
Distribution

Divisor

8.433E-06
In
Normal
1.000
1.370E-05
In
Rec
1.732
2.000E-05
In
Rec
3
2.887E-05
In
U
1.414
Combined Standard Uncertainty (1 Sigma) in Inches
Expanded Uncertainty (k=2) in Inches
Uncertainty:

Sensitivity
coefficient
ci
1
1
1
1

Uncertainty
contribution ui(y)
8.43E-06
7.91E-06
6.67E-06
2.04E-05
2.4392E-05
4.8785E-05

4.88E-05 Inches

Welch-Satterwaite
477.403141
Uc:
3.54009E-019
Ua:
5.05679E-021
Ub1:
3.91463E-021
Ub2:
1.97531E-021
Ub3:
1.73776E-019

5.00000
5.00000
5.00000
5.00000
5.00000
5.00002
5.00000
5.00000
5.00000
5.00002

Stundent's T:
Mean:

Units
Value

5.00

Observations:
1
2
3
4
5
6
7
8
9
10

Contribution

Infinite

2.00

at 95.45% CL

5.000
8.4327E-06

Atmospheric Pressure/humidity noted sources of uncertainty but found to be negligible in this analysis
Flatness/Parellelism/Bilateral deviation verified with optical flats and found to be of negligible values to include in budget.
User error taken into account on stability of measurement, Where different technicians found deviations of 20 uIn.
Ru= 10uIn resolution uncertainty + 10uIn paralax error, Divisor of 3 is used for a 1.732 Rec distrubution accounted for twice.
Su= 4.2 +1.9L uIn for Gage blocks
Eu=3^(Ucte(Ut))^3+(Dcte(Ut))^3+(CTE(Ut))^3 *
CTE Coefficient of Thermal Exansion
Ucte Uncertainty of CTE
Dcte Difference of CTE between Gage and Mic
Ut Uncertainty of Temperature

ASME/ANSI Y14.5M 1.4(k) states all temperature correction factors should be referenced to 20 C
is 11.5 for AISI 1050 steel and will be compensated for.
Mitutoyo Certifies compliance to 1050 steel throughout 90% of material(11.5 10 %), = 2.3uIn/In/Deg C max error
ASME B89.1.9 states the CTE of steel should be between 10.5 and 12.5 = 2uIn/in/Deg C
2% capped at 25 deg C resulting in 0.5 deg

Uncertainty Analysis
Parameter:

0-10 Length/Diameter Measure

Method:

Direct

Standards:

ATL1584

A7177

Date:

3/11/2008

UUT:
Range:

54-0150
1-10"

This anlaysis is based on Type B analysis of the parameter in question.


Any historical Type A or pooled uncertainties that may apply will also be included and noted.
Uncertainty Model:

Lu=Su+Ru+Eu+Sm

Measurement Model:

La=Li

Lu
Su
Ru
Eu
Sm

Length Uncertainty
Standard Uncertainty
Resolution Uncertainty
Environmental Uncertainty *
Stability of Measurements

La

Length applied

Li

Length andicated

Uncertainty Analysis Table


Lab Env is 20 deg 1 deg C
Symbol

Estimate

Type

Xi
Sm
Su
Ru
Eu
uc
U

xi
0.E+00
0
0
1

A
B
B
B

Inches

2
3
4
5
6
7
8
9
10

Std. Dev
Degree of Freedom:

Probability
Distribution

Divisor

9.661E-06
In
Normal
1.000
2.320E-05
In
Rec
1.732
2.000E-05
In
Rec
3
5.775E-05
In
U
1.414
Combined Standard Uncertainty (1 Sigma) in Inches
Expanded Uncertainty (k=2) in Inches
Uncertainty:

Sensitivity
coefficient
ci
1
1
1
1

Uncertainty
contribution ui(y)

9.66E-06
1.34E-05
6.67E-06
4.08E-05
4.4558E-05
8.9116E-05

8.91E-05 Inches

Welch-Satterwaite
1041.44378
Uc:
3.94185E-018
Ua:
8.71111E-021
Ub1:
3.21929E-020
Ub2:
1.97531E-021
Ub3:
2.78292E-018

10.00000
10.00000
10.00002
10.00000
10.00002
10.00000
10.00000
10.00000
10.00002
10.00000

Stundent's T:
Mean:

Units
Value

10.00

Observations:
1

Contribution

Infinite

2.00

at 95.45% CL

10.000
9.6609E-06
9

Atmospheric Pressure/humidity noted sources of uncertainty but found to be negligible in this analysis
Flatness/Parellelism/Bilateral deviation verified with optical flats and found to be of negligible values to include in budget.
User error taken into account on stability of measurement, Where different technicians found deviations of 20 uIn.
Ru= 10uIn resolution uncertainty + 10uIn paralax error, Divisor of 3 is used for a 1.732 Rec distrubution accounted for twice.
Su= 4.2 +1.9L uIn for Gage blocks
Eu=3^(Ucte(Ut))^3+(Dcte(Ut))^3+(CTE(Ut))^3 *
CTE Coefficient of Thermal Exansion
Ucte Uncertainty of CTE
Dcte Difference of CTE between Gage and Mic
Ut Uncertainty of Temperature

ASME/ANSI Y14.5M 1.4(k) states all temperature correction factors should be referenced to 20 C
is 11.5 for AISI 1050 steel and will be compensated for.
Mitutoyo Certifies compliance to 1050 steel throughout 90% of material(11.5 10 %), = 2.3uIn/In/Deg C max error
ASME B89.1.9 states the CTE of steel should be between 10.5 and 12.5 = 2uIn/in/Deg C
2% capped at 25 deg C resulting in 0.5 deg

1 in
5 in
10in

17.1uIn
48.8uIn
89.1uIn

Total deviation from 1-5


Total deviation from 5-10
Total deviation from 1-10

7.925uIn/In
8.06uIn/In
8.0 uIn/In

Total Estimated Uncertainty 0-10 In

17.1uIn + 8uIn/In

Das könnte Ihnen auch gefallen