Beruflich Dokumente
Kultur Dokumente
QA of Linear Accelerator
Procedure
Daily
Monthl
y
Annuall
y
x-ray
output
Lasers
ODI
Door
Interlock
Audio/Visua
l Monitor
Xray
Output
Electron
Output
Back up
Monitor
Xray CAX
dosimetry
Electron
CAX
dosimetry
X ray
Flatness
Electron
Flatness
X ray and
electron
symmetry
Light/radiat
ion field
coincidence
Gantry
Collimator
Wedge
position
Tray
Position
Applicator
position
Specificati
on
tolerance
-/+
3%
X
X
X
2mm
2mm
Operates
Operates
X
2%
2%
2%
2%
2mm @
depth
2%
3%
3%
2mm or
2%
change
1 deg
X
X
X
2mm or
2%
2mm
2mm
Field Size
Cross hair
Treatment
couch
Jaw
symmetry
Procedure
Field light
intensity
X
ray/electro
n
Calibration
Field size
dependenc
e
Output
factor
constancy
Central axis
parameter
Off axis
factor
constancy
Transmissio
n factor
constancy
for
accessories
Wedge
transmissio
n
Monitor
chambler
linearity
Xray output
constancy
vs. gantry
angle
Electron
output
constancy
X
X
X
2mm
2mm
2mm/1de
g
2mm
X
Daily
Monthl
y
Annuall
y
Specificati
on
tolerance
-/+
Operation
al
2%
2%
2%
2%
2%
2%
2%
1%
2%
2%
vs. gantry
angle
Off axis
factor
constancy
vs gantry
angle
Arc mode
Safety
interlocks
Collimator
rotation
@iso
Gantry
rotation @
iso
Couch
rotation
Coincidenc
e of
gantry ,
collimator,
couch axes
with Iso
Coincidenc
e of
radiation
and
mechanical
iso
Table top
sag
Vertical
travel
2%
Mfrs.
Specs
Operation
al
2mm
diameter
X
X
2mm
diameter
2mm
diameter
2mm
diameter
Respiratory Gating
Beam energy
constancy
Temporal
accuracy of
phase/
2%
100 ms of
expected
2mm
diameter
2mm
2mm
amplitude gate
on
Calibration of
respiratory
phase/
amplitude
Interlock
testing
100ms of
expected
operational
TPS QA
Acceptance Testing
Ct input
Anatomical description
Beam Description
Brachytherapy
Hardcopy output
Imaging artifacts
Finite Voxel size
Contrast agents
MR distortion
Paramagnetic sources
test
Reasons
Image geometry
Text information
Imaging data
Verify accuracy of
grayscale values,
particularly for
conversion of CT number to
electron density
Image unwarping
Methodologies which
modify imaging information
may
leave incorrect data in place
Voxel description
Surface description
Slices
Contours
Reference lines
Points
Density description
Display characteristics
Auto-segmentation
parameters
Structure constructed by
expansion or contraction
from another structure
Tests
Verify type ~e.g., external
surface, internal structure,
in homogeneity! and
capabilities that are
dependent on that type
Verify that correct definition
for relative electron
density! is used: Assigned
bulk density which sets
specified. densities derived
from CT number ~see
density tests in 3.4.3, Table
3-6
Check color, type of
rendering, and type of
contours to be drawn when
displaying structure
Check parameters for auto
contouring and other types
of auto structure definition
for each structure
Reason
. Incorrect attributes may
cause incorrect usage of
the structure
understood. Verify
algorithm with complex
surfaces ~e.g., sharp point,
square corners, convexities,
etc.! Check bookkeeping
issues ~e.g., is expansion
updated upon change of
source structure?!.
Structure constructed from
nonaxial contours
Structure definition
Numerous independent
difficulties can arise
dependent on the
underlying 3-dimensionality
of the data structures and
design of the code.72,80
Contour tests
Manual contour
acquisition
Tests
Define standard
procedures for contour
acquisition. Check and
document separation and
SSDs to AP and lateral
reference points for check
of integrity of digitization.
Check laser alignment
marks
Digitize standard contours
weekly or use other
process-related checks to
check geometric accuracy
Verify the geometric
accuracy of the digitizer
over the entire surface of
the digitizer.
Verify The accuracy of the
contour display with
respect to the image
display. The 3-D location
of the contour in the
coordinate systems! in
which the planning system
calculates dose. The
response of the contouring
algorithm to extreme
situations ~e.g., too many
points entered, looped
contour, .1 distinct closed
contours created!. The
identification of each
contour and its associated
3-D structure.
Tests may include:
Contouring structures on a
scanned phantom and
comparing contours to the
known dimensions of the
phantoms structures.
Reason
. Incorporate standard
checks into the acquisition
of manual contours to
prevent systematic and/or
patient-specific errors.
Geometrical accuracy of
the digitization device can
be quite user- dependent.
Many digitization systems
suffer from positiondependent distortions.
Digitizer behavior can also
be time-dependent
Contouring on CT images
is the basis of most 3-D
planning. Errors in contour
coordinates or display can
lead to incorrect anatomy
being used for planning.
Contour accuracy may be
dependent on image type
or orientation.
Bifurcated structures
DRRs, BEVs
Contours on projection
images
Contouring structures on a
grayscale phantom
constructed in software.
This eliminates any image
acquisition and pixel
averaging errors. A
subset of tests should be
performed for each type of
image, and for each slice
orientation ~sagital,
coronal, axial, oblique!,
since the contouring
features and/or use of the
contours may not be
independent of these
parameters.
Verify proper response of
the tracking algorithm for
various situations ~e.g.,
different grayscale
gradients, different image
types, markers, contrast,
image artifacts!. Tests
may involve scanned
phantoms or simulated
grayscale phantoms as
described above. Partial
volume effects probably
are most easily sorted out
using images which model
the effects of slice
thickness changes on the
grayscale values.
Resolve issues such as:
Can the system maintain
more than one contour
per slice for a particular
structure? Does it form
the 3-D structure
correctly? Check 3-D
surfaces visually and
check DVHs
Check that points defined
on projection images
. Incorrect handling of
contours on projection
Contours on CT
scannograms
CT scannograms have
significant divergence in
the axial direction but
typically negligible
divergence in the sagital
direction
Contour extraction onto
axial and non-axial images
or reconstructions
provides one of the best
ways to quantitatively
check the 3-D description
of anatomical structures.
CT number conversion
Editing
Measurement tools
Bolus tests
Electron density within
bolus
Density measurement
tools
Beam assignment
Dose calculation
method to calculate
monitor units when bolus
is used
Verify that bolus is
displayed properly in all
displays and hardcopy
output Verify that bolus is
properly documented
within the plan and in the
hardcopy output
Verify functionality of
window and level setting
Determine whether
displayed window/level
values agree with those on
scanner/film.
Geometrical accuracy of
slices associated with
images
Region-of-interest analysis
Positional measurements
Beam parameters-Verify
Beam Description
machine
modality
energy Beam Geometry
isocenter location and table position
gantry angle
table angle
collimator angle Field Definition
source-collimator distance source-tray distance
source-MLC distance collimator settings ~symmetric or asymmetric!
aperture definition, block shape, MLC settings
electron applicators skin collimation Wedges
name
type ~physical, dynamic, auto!
angle
Machine/beam accessories
Parameter limitations
Readouts
Wedges
tests
Verify that the library of
available machines and
beams is correct. Clinical
beams should be
segregated from research
or other beams
Verify that the availability
of machine and beamspecific accessories, such
as electron cones or
wedge, is correct
Verify that limitations are
correct for jaws, multiyear
collimator, field sizes for
fields with wedges,
compensators, MLC,
electron applicators. Verify
MU limits, MU/deg. limits,
angle limits ~gantry,
table, collimator!, etc.
Verify correct use and
display of user-defined
names and numbers
reason
. Incorrect beam choice
leads to wrong dose
calculation and monitor
units
Incorrect
numbering/names can
lead to incorrect
treatments due to
confusing documentation
Lack of agreement
between readout
information in RTP system
and machine leads to
systematic machine
treatment errors
Incorrect functioning of
these features will lead to
internal mistakes in
planning.
This can lead to incorrect
wedge use in plan or
during treatment.
Compensators
Internal consistency
Readouts
test
reason
Problems can cause
systematic treatment
errors
Test frequency
Multi-user environment
Block transmission
Electron applicators
Hardcopy output
view display
Verify the accuracy of this
information at the
commissioning of the RTP
system and at each major
software update
Establish a procedure to
ensure consistent beam
information in multi-user
and network
environments.
test
Verify that the system
distinguishes between
island blocks, in which the
aperture delineates the block
shape, and aperture or
conformal blocks, for which
the drawn aperture encloses
the open irradiated area.
Divergent and nondivergent
blocks should also be
considered.
Verify correct specification of
transmission or block
thickness for full blocks and
partial transmission blocks
Document and test all
methods used to fit the MLC
leaves to the desired field
shape.
Verify availability and size of
electron applicators
Check all output showing
beam apertures and/or used
for beam aperture fabrication
~e.g., MLC leaf positions,
BEV plots! for accuracy
against the displays
reason
Could lead to incorrect
identification of blocked or
irradiated areas.
test
Editing apertures
3-D projections
reason
Incorrect block
shape could be
used in plan
Could lead to
incorrect margins
during aperture
design
Could lead to
incorrect aperture
shape
Could lead to
incorrect aperture
shape
This could lead to
incorrect
interpretation of
planned aperture
Might lead to
incorrect aperture
design or choice
of beam direction.
Wedge tests
Orientation and angle
specifications
2-D display
3-D display
Dynamic wedge69
test
Confirm that wedge
orientation and angle
specifications are
consistent throughout the
planning system, including
the hardcopy output. If
possible, they should
agree with treatment
machine conventions.
Check display of wedges
in different 2-D planes
~parallel, orthogonal, and
oblique! for different beam
directions, collimator
rotations, and wedge
orientations
Check display of wedges
in room view 3-D displays
for situations as described
above
Verify that wedge
orientations and field sizes
not allowed by the
treatment machine are
not allowed in the
planning system. These
limits might be defined
separately for each beam
energy, so they should be
tested for each
energy/wedge
combination.
Confirm that the division
of a field into fractional
open and wedged fields
agrees in the RTP system
and on the treatment
machine
Verify that the
implementation in the RTP
system has the same
reason
Wedge labeling or
orientation conventions
which do not agree with
the RTP system can lead
to confusion in plans and
treatment
Incorrect wedge
orientation leads to large
dose difference
May lead to plans that
cannot be delivered
capabilities, limitations,
orientations, and naming
conventions as on the
treatment machine
Non-axial divergence
BEV/DRR displays
3-D displays
test
Test intersection of
divergent beam and
aperture edges with axial
slices
Test intersection of
divergent beam and
aperture edges with
sagital, coronal, and
oblique slices. For systems
that are not fully 3-D,
there may be 2-D
limitations in the
projections which must be
taken into account.
Verify projection of
contours/structures
defined on axial slices into
BEV-type displays.
Compare with the
grayscale images for DRR
displays. This is most
easily done with a simple
phantom containing only a
few internal structures.
Verify projection of
divergent beam and
aperture edges. Check
at several different SSDs
and projection distances.
Verify that apertures
defined on 2-D planes are
correctly projected in 3-D.
Verify that the
relationships between
structure and beam and
aperture edges are
correct. 2-D limitations
of the system must be
considered ~e.g., a 2-D
system may not correctly
display divergence in the
reason
Incorrect divergence leads
to selection of wrong field
sizes or aperture shape
Incorrect divergence leads
to selection of wrong field
sizes or aperture shape,
especially if 3-D effects
are not completely
understood
. Incorrect projections
lead to selection of wrong
aperture shape, especially
if 3-D effects are not
completely understood.
third direction!
Verify patient orientation
with respect to beam and
orientation annotations.
Verify correctness of
orientations and
annotations for machine
position views or icons
associated with 2-D or 3-D
displays.
Status of density
corrections
Reason
Must calculate dose to
regions which are
important
Misleading dose
distributions, incorrect
Consistency
Dose grids
.. Isodose surfaces
Beam display
Inconsistency
demonstrates algorithm
limitations or problems,
makes evaluations
impossible.
Interpolations done
incorrectly give wrong
dose results, particularly
in penumbra regions.
DVH tests
Volume region of interest
~VROI! identification
Structure identification
Structure volume
DVH calculation
DVH types
Test
Test creation of the voxel
VROI description used to
create DVHs against
structure description.
Test Boolean combinations
of objects ~VROI and DVH
of Normal Tissue-Target!,
and how voxels which
belong to multiple
structures are handle
Verify accuracy of dose
interpolated into each
voxel
Test accuracy of volume
determination with
irregularly shaped objects,
since regular shapes
~particularly rectangular
objects! Can be subject to
numerous grid-based
artifacts.
Verify that appropriate
histogram bins and limits
are used
Test DVH calculation
algorithm with known
dose distributions
Reasons
Misidentification of VROI
leads to incorrect DVH
the DVH
Grid-based artifacts can
cause errors in volume,
dose, DVH, and the
evaluation of the plan
Comparison of DVHs from
different plans depends
critically on bin sizes, etc.
BEV or DRR
DVHs
3-D displays
Test
Source display
Reasons
For source location entry
using a digitizer and
orthogonal or stereo-shift
films, checks should be
made of the data entry
software, the film acquisition
process, source
identification, and other
associated activities. 3-D
seed coordinate
representation after entry
should be confirmed.
Automatic seed
identification and locating
software must be verified.
For source location entry
using CT images, 76 other
tests should be included.
For applicator trajectory
identification, the
appropriate tests described
above should be performed.
In addition, the accuracy of
dwell points or source
locations along the
trajectory should be
confirmed.
Verify accuracy of source
position display on: 2-D
slices, including CT and
reconstructed images and
the arbitrary planes often
used in non-CT
brachytherapy. 3-D views
Special views, such as the
Probes Eye View used in
stereotactic brain implant
planning.77 Dummy
Incorrect functioning of
optimization and evaluation
tools can result in suboptimal or incorrect
treatment