Beruflich Dokumente
Kultur Dokumente
ABN 18737137544
Gedge Systems expressly disclaims all responsibility for the direct and indirect consequences to each and every
party whatsoever, in respect of the use or application of this product or any other Gedge Systems product if it
is used in any way whatsoever beyond or outside the specified purpose for which it was supplied.
The contents, format and authorship of all material produced or printed by any medium,
by or on behalf of GS Systems Pty. Ltd., is the subject of strict proprietorship of and by
GS Systems Pty. Ltd., and cannot be copied, reproduced, plagiarized or used in part or in
whole either directly or indirectly by any party whatsoever without the prior written
consent of GS Systems Pty. Ltd. This publication, all of its parts and all matters derived
from it, forms part of such material.
Email: gs3100@gedgesystems.com.au
page 2
Table of Contents
5 1 INTRODUCTION & FEATURES.
7 2 OPERATING THE GS3100
7
7
8
9
9
10
2.1
2.2
2.3
2.4
2.5
2.6
Introduction
Applying Power, Retained & Cleared Settings, Serial I/O Considerations.
Messages & their meaning including the 3 methods of locking the GS3100's keys.
Front Panel Display
Front Panel Lamps
Keyboard Controls
13 3.5.1
13 3.5.2
14 3.5.3
15
16
16
17
18
18
3.6
3.7
3.8
3.9
3.10
3.11
Installing an optional serial card into the GS3100 & Setting its Baud Rate.
Selecting Isolated or Non-Isolated Operation of Serial Option Cards 202, 207, 07 & 08.
Setting the baud rate for the GS3100's standard RS-232C serial output.
Connecting Load Cells.
Connecting to the Mains Supply.
Calibration and Setup keys - Access and Use.
18 3.11.1
18 3.11.2
19 3.11.3
General.
Use of keys and access to the settings.
Use of the GS3100's keys for calibration and programming.
3.12.2.1
3.12.2.2
3.12.2.3
3.12.2.4
3.12.2.5
3.12.2.6
3.12.2.7
3.12.2.8
3.12.2.9
3.12.2.10
3.12.2.11
3.12.2.12
3.12.2.13
3.12.2.14
3.12.2.15
3.12.2.16
24 3.13 Device Number - its description and instructions for setting it.
24 3.14 Weight Units Labels
25 3.15 Weight Calibration
25 3.15.1
25 3.15.2
25 3.15.3
General
Zero Out The Dead Load Offset
Span Calibration.
Email: gs3100@gedgesystems.com.au
page 3
Table of Contents
26 3.16 Connecting the Standard RS-232C Serial Output & the Set Point Control Outputs.
27 3.17 Electrical Specification of the GS3100's Set Point Non-Latching Outputs.
28 3.18 Spike and Transient Suppression of Switched Load Noise.
32 6.3
Available Serial Output Options, What they do and Why they are selected.
Physical Installation, Baud Rate Selection, Output Selection, Set up,Format & Isolation.
Optical Isolation, the Position of Links LK1 on the Option Card & the need for a 9VDC Supply.
34 7.3
35 7.4
36 7.5
36
36
37
37
38
7.5.1
7.5.2
7.5.3
7.5.4
7.5.5
38
38
39
40
41
42
45
45
7.5.5.1
7.5.5.2
7.5.5.3
7.5.5.4
7.5.5.5
7.5.5.6
7.5.5.7
7.5.5.8
Available Serial Input/Output Options, What they do and Why they are selected.
Physical Installation, Baud Rate Selection, Input/Output Selection, Set up & Isolation.
Isolation, the Position of Links LK1 on the Option Card & the need for a 9VDC Supply.
Maximum Baud Rate, Delay in Responding and Timing. (Not applicable to MODBUS)
Device Number and Digital Setup Settings for serial I/O. (Not applicable to MODBUS.)
Optional Serial Input/Output Cards - Input/Output and Power Supply Connections.
GS3100's Serial I/O Language.
Communications Flow Control.
Avoiding Multiple GS3100s on the same RS-485 Bus Talking at the same time.
Time Out and Clearing the Command Buffer.
Acquiring Zero and Missing Commands.
Communication Language and Commands.
47 8.3
47 8.4
47 8.4.1
47 8.4.2
48 8.4.3
48
50
50
51
8.5
8.6
8.7
8.8
Introduction.
Serial Transmission Modes, Speeds and Slave Numbers.
ASCII Mode
RTU Mode
52 9 TECHNICAL SPECIFICATION.
53 10 WARRANTY
Gedge Systems - Melbourne - Australia
Email: gs3100@gedgesystems.com.au
page 4
Analog Output using OPTION6HPF plus 100% & 0% Load Simulation for external equipment calibration.
n NOW AVAILABLE - MODBUS ASCII and RTU OUTPUT. The factory fitted Option MODBUS3100"
provides full modbus RTU and ASCII serial I/O from 1200Baud to 19,200Baud with support for
multi-holding register read function 03 and single register write function 06. SLAVE number can be set from
01 to 99. By also selecting Option07 or Option08, users can choose the serial I/O level to suit their
application. Available I/O includes RS-232C; 20mA Current Loop; RS-422A or RS-485 MultiPoint. In
addition, the MODBUS option includes battery backed ram for tare, ID number and zero tracked value.
n NOW AVAILABLE with the release of SOFTWARE VERSION 7 or GREATER - PLC COMPATIBLE 8bit
SERIAL COMMUNICATIONS. You can now easily select the serial word type best suited to your application
- choose either 7bit with Even parity or 8bit with no parity. Word type selection as required by the MODBUS
standard remains fully automatic for MODBUS protocols.
n A World First. The first Digital Weight Indicator to be designed from the ground up as a front end for
Computer/PLC Controlled batching. It has all of the features and none of the hassles of competing products
which have to be shoe horned into the job. It is tailored for the job and fits perfectly.
n Computer/PLCs can talk to factory floor operators, via the GS3100, ACROSS the WORLD. The in-built
advanced serial I/O language, using optional serial I/O cards, allows the GS3100's keyboard and display to be
used as the PLC/Computer operator's interface - whether it be on the factory floor or, via a modem, on the
other side of the World. The remotely situated PLC/Computer can completely control the GS3100. In
addition to weight, set point and ID information, it can also request a vast range of information, including
many of the GS3100's installed settings, its product designation, software version and much more.
n Serial I/O Controlled Stream Weight Data Mode.
n GS3100 as the remote Operator input device for the Computer. All front panel key operations can be sent to
the remote Computer. Immediately the Computer puts the GS3100 into terminal mode, all operations of the
GS3100's front panel keys are sent to the Computer and not acted upon by the GS3100. During all of this
time, the GS3100 continues weighing and set point operations without interruption.
n GS3100 as a remote message display for the PLC/Computer. Immediately the Computer puts the GS3100 into
terminal mode, the GS3100's display is completely under the control of the Computer. It will display any
messages sent to it and so, can be used by the Computer to talk to the on-floor operator. During all of this
time, the GS3100 continues weighing and set point operations without interruption.
n Standard no-extra cost RS-232 Serial Output.
Email: gs3100@gedgesystems.com.au
page 5
set points can monitor the weight, in-the background, while the Computer/PLC is controlling the batching
process. In the event of a fault with the Computer, or with related equipment, the set points can trigger alarms
or close down the plant.
n Full SIX wire load cell connection. The GS3100 automatically and accurately compensates for long load cell
cable runs typical in weigh/batching applications. Competitors cheap products, which use only FOUR wires
cannot compensate for long cable runs and can produce errors in excess of 10% of reading as a result.
n Assured and Guaranteed Accuracy and Stability. Every GS3100 spends at least ONE DAY in our thermal
chamber cycled between equivalent operating temperatures of -10oC and +40oC. Their accuracy and stability
are continuously monitored and recorded. If they don't meet our specifications 100%, they don't get shipped.
Typically, actual shipped performance is FIVE to TEN TIMES MORE STABLE than our published
specification - you can rely on it.
n High Speed TWENTY PER SECOND operation tuned to provide rapid response. Slower update rates can
14mm display.
n Easy to use, the front panel lamps and descriptions are clear. The method of entering set points and other
Complete
connection to alarm or feeder solenoids driven by a user supplied low cost power supply.
n Not a Panel meter. The GS3100 is a proper full function weight indicator. It is designed and manufactured
by Gedge Systems who has international weights and measures approvals for weight indicators and
controllers.
n Display any weight to 199990.
n Easy to calibrate.
Calibration is
n Easy connections via 100% plug and socket connections. Special short length cable connectors with metal
back shells supplied as standard to reduce the required panel depth. The panel mounting hardware is supplied
as a no extra cost standard item.
n Added security - the GS3100 records attempts to access its settings. Calibration and programming can only be
accessed by a secure key sequence. Every failed attempt to access the GS3100's settings is recorded. So too
is every successful access attempt. This information is displayed when power is applied, and so can be noted
by installation and supervisory personnel - it can also be transmitted to a remote Computer/PLC using the
GS3100's I/O features.
n Quality Assured. All features, operations, displays, outputs as well as standard and options connectors have to
Email: gs3100@gedgesystems.com.au
page 6
2.2 Applying Power, Retained & Cleared Settings, Serial I/O Considerations.
Retained and Cleared Settings & Serial I/O Considerations: When power is removed, the GS3100 retains all
of its settings, including all of the calibration and setup information, the two set point values and the key zero
value. The only exceptions are: the Tare value; any zero tracked value and the ID number, all of which are
cleared and set to zero when power is next applied. In addition, any flags which were set during serial I/O,
will also be cleared. This point is particularly important if the remote computer is locking out the GS3100's
keyboard. The lockout flag will be cleared and keyboard access enabled when power is re-applied. Further, if
the GS3100 was being operated in terminal mode, it resumes operation in its normal mode when power is
re-applied. Remote control schemes should include techniques to frequently poll the GS3100 to determine its
status and, if necessary, re-apply the key locking and terminal mode functions.
{The GS3100 also clears the above mentioned settings following a proper calibration or setup procedure. It does
not clear them, however, following an unsuccessful attempt to gain access to the GS3100's CAL function.)
Messages when power is applied: When power is applied, the GS3100 loads its retained settings, clears the
above mentioned settings and flags, and then verifies its configuration. It then displays the below sequence of
messages which are followed by a display of the gross weight:3100.8A This identifies the product as a GS3100, with software version 8 installed. The GS3100 is type A.
3100.7b This identifies the product as a GS3100, with factory installed MODBUS3100" Option fitted.
Software version 7 installed. The GS3100 is type b.
888888 The GS3100 turns on all the bars of its display and all of its six lamps to verify that they are
working properly.
Blank This is a blank display with nothing shown. The GS3100 turns off all the bars of its display and all
of its six lamps to verify that they are working properly.
Email: gs3100@gedgesystems.com.au
page 7
2.3 Messages & their meaning including the 3 methods of locking the GS3100's keys.
The following messages may be displayed when power is applied, or during operation of the GS3100. The
explanation of the messages contains recommendations that should ONLY be carried out by personnel who are
fully qualified and authorized to do so.
Locd F Locked Key Function. This means that the key just operated has been locked out by the function lock
setting in Digital Setup during installation. See page 21.
Locd L Locked Keys Lockout. This means that the key just operated has been locked out by a remote
Computer or PLC sending a keyboard locking command to the GS3100. See page 40 and the factory installed
MODBUS3100" Option.
Locd.tr Locked Keys in terminal restricted mode. This means that the key just operated has been locked out
by a remote Computer or PLC which has placed the GS3100 into the terminal but does not want the GS3100
to transmit the keyboard key codes. See page 42.
too Hi. Too High. This message is displayed when a set point value is being entered if the key just operated
would cause the value to exceed 120% of the GS3100's installed capacity. The message is displayed for 1
second, followed by a 2 second display of the allowed maximum value for the set point. The value being
entered is then displayed. It should be changed to ensure it does not exceed the maximum allowed value.
Err1 Error 1. The GS3100's factory settings are damaged and the GS3100 cannot operate properly. It must be
returned to the factory for testing and repair. There is no in the field solution for this problem.
--- The gross weight on the scale or in the weigh hopper is greater than the set up and calibrated
capacity of the GS3100. This may occur briefly as the weight on the scale approaches, and, due to impact
loading, overshoots the full capacity of the scale. The scale or system installer should be contacted to remedy
this, as it may indicate the possibility of damage occurring to the load cells due to the load overshoot.
OFF The electrical signal from the load cells is greater than the allowed input to the GS3100. This indicates a
serious problem, because the allowed input to the GS3100 is very large, regardless of how it has been
calibrated. Possible causes are: damaged load cells due to overload, lightning strike or welding. Damaged load
cell cables for whatever reason. Damaged load cell circuit to the GS3100. Incorrect connection to the GS3100
of the load cell cables. A failure of the GS3100's load cell input. If the GS3100 displays weight when a star
bridge or load cell simulator is properly connected to it, the fault will be in the load cell circuit external to the
GS3100. Contact the scale or system supplier to investigate the fault.
-OFF The electrical signal from the load cells is negative and greater than the allowed input to the GS3100. If
the load cells are sitting on a bench, and have no load upon them, this may not be a serious problem, as it
may just mean that the load cells unloaded electrical output is more negative than is usual. Try loading the
load cells to see if the displayed message disappears. Alternatively, try swapping the signal cables as they may
be connected in a way that makes the electrical signal go negative on load rather than positive.
If it is not any of these, then this indicates a serious problem, because the allowed negative input to the
GS3100 is large, regardless of how it has been calibrated. Possible causes are: damaged load cells due to
overload, lightning strike or welding. Damaged load cell cables for whatever reason. Damaged load cell circuit
to the GS3100. Incorrect connection to the GS3100 of the load cell cables. A failure of the GS3100's load cell
input. If the GS3100 displays weight when a star bridge or load cell simulator is properly connected to it the
fault will be in the load cell circuit external to the GS3100. Contact the scale or system supplier to investigate
the fault.
Gedge Systems - Melbourne - Australia
Email: gs3100@gedgesystems.com.au
page 8
The ENTRY lamp will be on as follows:During entry of an ID number or set point value.
During operation of the GS3100 in terminal mode,
under the complete control of a remote Computer or
PLC. When operated in this way, the ENTRY lamp
illuminates immediately the GS3100 is placed in the
terminal mode by a command from the remote
Computer/PLC. It remains on until a front panel key is
pressed, when it goes off. The key value is then sent
to the remote Computer/PLC which can then send an
acknowledgement message to the GS3100s display.
Receipt of this message by the GS3100 causes the
ENTRY lamp to again be turned on.
This lamp is not directly controlled by the remote
Computer/PLC - its operation is automatic and is
controlled by the GS3100 based on the above
operations.
If the GS3100 is used in the restricted terminal
mode this lamp is always off as the GS3100s keys
are not transmitted to the computer in this mode.
Gedge Systems - Melbourne - Australia
The set point lamps SP1 and SP2 will be OFF all of
the time that the GS3100 is operated in the terminal
mode. At all other times, they will be ON when the
weight is equal to the set point or greater than the set
point and OFF when the weight is lower than the set
point. {The GS3100s control outputs are closed and
conducting current when the set point lamp is ON and
open circuit when the set point lamp is OFF.}
Lamp ON if weight >= setpoint
Lamp OFF if weight < setpoint.
SP1 or SP2 lamp will also be ON while a set point
value is being entered at the keyboard.
Email: gs3100@gedgesystems.com.au
page 9
ENTER, ID#, SP1, SP2 and the numeral keys 0 to 9. These keys are used to display and to enter ID numbers
and set point values. To display a setting, operate the appropriate key - ID#, SP1 or SP2. The current value will
be displayed. If you do nothing for 15 seconds, the GS3100 will automatically return to its weight display. The
GS3100 continues to check its set points and to drive its serial output even while ID numbers and set point
values are being changed. To change a value, first display it, then key CLEAR while the current value is
displayed. This tells the GS3100 that you want to make a change and it immediately clears the display. You then
use the numeral keys to enter the new value and key ENTER again to save it. The GS3100 momentarily blanks
its display to signify a successful operation. If you try to enter a set point value that is too large, the GS3100 will
warn you with a message too Hi. You can quit changing a value without affecting the existing setting by
keying G/N instead of the ENTER. When you key ENTER to save your new set point value, the GS3100
automatically adjusts the smallest increment to suit the GS3100s set up and displays the adjusted value (for
example, it would make 3" a 5" if the GS3100 was set up to count by 5).
Gedge Systems - Melbourne - Australia
Email: gs3100@gedgesystems.com.au
page 10
damage in transit. Only proceed if no damage is found. Check that all of the items are present. Retain the
packing.
n Check that the supply voltage identified on the rear of the GS3100 is correct. Do not connect to the mains
supply yet.
n Read the paragraph on page 12 concerning electrical supply quality. If a mains filter/conditioner is required
obtain it now.
n If an Analog Output Option is to be fitted, install it now. The baud rate links on the motherboard are not used
transmission rate (baud rate) using the switches on the serial option card. Also remove the baud rate link
LK2" situated at the front on the GS3100's motherboard. NOTE:- Operation of the Optional MODBUS serial
I/O feature also requires selection and installation of either an Option07 or Option08 Serial I/O card.
n If the GS3100's standard RS-232C serial output is to be used, instead of an optional serial card, select the
baud rate now by positioning the link LK2" in the 9600 baud or 4800 baud position. LK2" is situated at the
front of the GS3100's motherboard.
n Connect the load cell or cells to the GS3100.
n Connect the GS3100 to the mains supply and apply power. Check to ensure that there are no persistent error
Particularly note the security key sequence needed to access the GS3100's calibration and set up settings.
n Decide on the appropriate Digital Setup settings and then program them into the GS3100.
n If the GS3100's serial output is to be used set the Device Number.
n The GS3100's front panel includes the unit of measure kg. If the units for this installation are not kg
Email: gs3100@gedgesystems.com.au
page 11
3.3 Check the mains supply voltage on the rear of the GS3100.
The GS3100 is designed to work correctly at certain specified nominal voltages and within specified frequencies
and voltage tolerances as contained in the Technical Specification section of this manual. The GS3100 is
usually supplied to operate at 240VAC 50Hz single phase unless specified on the purchase order. Alternative
nominal voltages that can be supplied for 50Hz or 60Hz single phase operation are 100/110/117/200/220VAC.
The rear panel of the GS3100 includes a label which identifies the mains voltage required. Check that this
matches the mains voltage that you intend to use. DO NOT PROCEED IF THE MAINS VOLTAGE IN
YOUR AREA DOES NOT MATCH THE VOLTAGE SPECIFIED ON THE GS3100, AS INJURY OR
DAMAGE MAY RESULT. CONTACT THE SUPPLIER OF THE GS3100 IMMEDIATELY.
Do not apply power to the GS3100 just yet, as there are more installation steps to be carried out before power
should be applied.
Email: gs3100@gedgesystems.com.au
page 12
the GS3100 motherboard, ensuring that the coloured conductor on the edge of the ribbon cable is to the front
of the GS3100 when viewed from the front (this conductor connects to pin 1 of J3). The connector J3 is keyed
but take care not to offset the pins. Also, do not confuse this connector with the six pin connector J4 near the
middle of the motherboard.
n Install the option board with the component side uppermost by screwing it to the GS3100's rear panel using
the jack screws and washers provided. The star washers must be used under the jack screws.
n Connect the ribbon cable connector to the J1 socket on the option board (support the board while doing this)
making sure the coloured conductor on the edge of the ribbon cable is to the front of the GS3100 when
viewed from the front (this conductor connects to pin 1 of J1 on the serial card). The connector J1 is keyed
but take care not to offset the pins.
n Now apply power to the GS3100. There is no need to connect a load cell to the GS3100 in order to adjust the
Option6HPF's analog output. If a load cell is connected there is no need to remove it as it will not affect the
adjustment procedure.
n If not already selected, refer to Digital Setup on page 20 and the Output selection on page 22 and select
Analog Output (if this is not done the Load Simulation will not work).
n Option6HPF is a highly accurate and stable 14bit (1:16,000) D/AC driven 4-20mA and 0-10VDC Output. If
its output is received by external equipment with a 14bit (or greater) A/DC input then the external equipment
should have numeric agreement with the GS3100 in most practical weighing applications.
n Option6HPF is shipped from the factory with the 4-20mA signal adjusted for 4mA at zero load and 20mA at
100% Capacity. It should not require adjustment. If the 0-10VDC signal is to be used instead of 4-20mA it
will require a small adjustment.
n To adjust Option6HPF and calibrate external equipment ensure the GS3100 and Option6HPF have warmed up
for about 30minutes to achieve thermal stability. Now use the GS3100's Analog Load Simulation (described in
the following) to simulate ZERO load and 100% Full Load and use the Zero and Span potentiometers on the
Option6HPF card to obtain the required output. The potentiometers are identified on the card as ZERO and
SPAN. They can be seen on the left of the card in the image on page 14.
n When all adjustments are complete replace the GS3100 top cover to finish the installation.
Email: sales@gedgesystems.com.au
page 13
Option6HPF Connections.
Option6HPF is designed specifically for use in the GS3100 and "short Pin
enclosure" GS products that use a high speed data bus on the
No.
Connection
Motherboard. Its high stability components and 14bit (1:16,000) Digital
to Analog Converter will deliver an analog output capable of numeric
1 Frame (Shield)
agreement between the GS3100 display and remote equipment.
When used in the GS3100, the analog output of Option6HPF is based 13 + Voltage Output
on displayed weight, this will be Gross or Net weight. In the event of an
error mesage being displayed by the GS3100, the analog output will be 11 - Voltage Output (ground)
driven to fullscale (20mA or 10VDC) on Overweight/+Offscale and
9 + Current Output
driven to zero (4mA or 0VDC) on -Offscale.
The table to the right shows the connections on the Options 25 pin D
7 - Current Output (ground)
female connector. A mating male cable connector is supplied with each
option. To minimize noise use a shielded cable with the shield grounded only at the GS3100 end by crimping it
between the metal backshells of the cable connector & connecting to Pin 1 Frame.
Specification
Range:Analog Output:Adjustment Range:Accuracy/Resolution:
External Load:-
Email: sales@gedgesystems.com.au
page 14
the GS3100 motherboard, ensuring that the coloured conductor on the edge of the ribbon cable is to the front
of the GS3100 when viewed from the front (this conductor connects to pin 1 of J3). The connector J3 is keyed
but take care not to offset the pins. Also, do not confuse this connector with the six pin connector J4 near the
middle of the motherboard.
n Install the option board with the component side uppermost by screwing it to the GS3100's rear panel using
the jack screws and washers provided. The star washers must be used under the jack screws.
n Connect the ribbon cable connector to the J1 socket on the option board making sure the coloured conductor
on the edge of the ribbon cable is to the right of the GS3100 when viewed from the front (this conductor
connects to pin 1 of J1 on the serial card). The connector J1 is keyed but take care not to offset the pins.
n Refer to the baud rate selection table and select the baud
600
off
off
off
1200
on
off
off
2400
off
on
off
4800
on
on
off
9600
off
off
on
19200
on
off
on
not used
off
on
on
not used
on
on
on
off
on
off
on
n Remove the GS3100's baud rate link LK2 situated at the front left of the GS3100's motherboard. Failure to
remove this link will not damage the GS3100, however, serial communication will be impossible if it is left in
place. See over page for the location of the link.
n Replace the GS3100 top cover to complete the installation.
Email: gs3100@gedgesystems.com.au
page 15
3.8 Setting the baud rate for the GS3100's standard RS-232C serial output.
If the GS3100's standard serial output is to be used, rather than an optional serial card, the baud rate must be
selected from the two choices available.
The GS3100's standard RS-232C serial output transmits at either 4800baud or
9600baud. The transmission rate is set by the position of a link, LK2, situated at
the front of the GS3100's motherboard. The circuit board is printed with the baud
rate selections. Place the link in the required position to select the baud rate suited
to your application.
Email: gs3100@gedgesystems.com.au
3
4
8
Fig 4 Load cell connector, 9 pin female, cable side.
WARNING. If the load cells are powered by the user's
excitation supply rather than the GS3100's supply the following
apply:-
Connection
Positive Excitation. (+ Exc.)
2
1
6 and 9
Positive Sense.
This connection MUST be
made. If SIX wires are not
used then jumper this pin to
pin 3 positive excitation.
Negative Sense.
This connection MUST be
made. If SIX wires are not
used then jumper this pin to
pin 4 negative excitation.
Positive load cell Signal.
(+ Sig.)
Negative load cell Signal.
(-Sig.)
GS3100 Frame.
(Cable Shield)
2. The user's excitation supply must be isolated from ground and therefore fully floating. If this is not the case
then noise due to ground loops may occur. Even if the user's excitation supply is floating, the additional
distributed wiring and circuitry may introduce noise and it will be the user's responsibility to eliminate this.
3. The user must decide whether to use the GS3100's excitation ratio sense or not. If the sense is not used, the
installation will be sensitive to variations in the excitation supply across the load cell. These are caused by
changes in the resistance of the load cell cables and by time and temperature changes in the excitation supply.
The remote sense can be disabled by placing shorts at the GS3100's load cell connector. Short pins 3-5 and 4-8.
If remote excitation sense is used, the following apply:a. The sense wires must be connected to pins 5 and 8 as identified in the load cell connector table above.
b. The sense voltage level at the GS3100 must be 10VDC at calibration. Variations after calibration of
+0.2VDC to -2VDC is permitted and will result in automatic compensation of the load cell input.
c. Great care must be taken to ensure that the sense input to the GS3100 is an accurate reflection of the actual
excitation across the load cells. If it varies in any way where such variation is not an exact reflection of the
excitation across the load cells, then the automatically compensated load cell input will be in error.
COMMENT. High power radio transmitters in factories and vehicles is making the use of proper load cell
installation techniques MUCH more important than it has been in the past. Failure to strictly follow the
installation techniques provided in this manual for the load cell and any options WILL render the installation
unacceptably sensitive to interference and will void the warranty.
Before connecting the load cell ensure that the jack screws at the load cell connector on the rear of the
indicator have star washers beneath them and that they are screwed up tight. Also ensure that the entire load cell
cable has shield continuity and that a high quality BRAIDED shield is used. Also ensure that the shield is NOT
grounded at the load cell end, although, in some situations a ground at the load cell end has been found to
diminish interference. Do not proceed with the installation until the load cells have been properly wired and
installed.
Email: gs3100@gedgesystems.com.au
page 17
Email: gs3100@gedgesystems.com.au
page 18
Email: gs3100@gedgesystems.com.au
page 19
Comments
d 3000
dP 0.0
Cb 1
UP 20
S 9d
Otr OFF
FL. non
St gg
Display divisions - select the required number for the weight to be displayed.
Decimal places - select the required number for the weight to be displayed.
Count by - select the required number for the weight to be displayed.
Update rate in updates per second.
Standstill (not in motion) condition in display division changes per second.
Zero track rate for automatic zero maintenance in display division changes per second.
Front panel keyboard function lock. Used to disable front panel keys.
Set point test setting. Select gross or net as the weight to be used to operate the set points.
This setting uses the gross weight for both set points.
OC oo
THIS SETTING APPLIES ONLY TO GS3100s fitted with software version 6 or earlier.
Output condition when the set points are tripped. This setting has both set point outputs ON for weight below the set point,
going OFF as the weight rises above the set point.
Out.C1
br. 19.2
SC7b E
Str c1
Un 02
bd 06
bE 02
The type of serial output required or Analog Output. Select printer (P), or continuous (C)
data streams or, if an optional serial I/O card is used, select serial I/O (io1). If the Optional
MODBUS3100" is installed, MODBUS RTU or MODBUS ASCII format can be selected.
MODBUS baud rate. Only displayed if the previous selection is MODBUS RTU or ASCII.
The baud rate switches on the serial I/O card Option08 or Option07 MUST be set to the
value selected for this setup setting. The settings below do not apply to MODBUS serial I/O
and will not be displayed if a MODBUS selection is made.
NEW-THIS SETTING APPLIES TO GS3100s fitted with software version 7 & higher.
Serial Comms word composition. Either 7bit Even parity or 8bit No parity.
Stream data selection only required when the GS3100 is used in serial I/O mode.
Printed units of measure (kg, t etc). Only required for printer outputs.
Block division character inserted between sections of the serial output data string.
Beginning and end character sent at the beginning of a data transmission and at the end.
Email: gs3100@gedgesystems.com.au
page 20
dP0.0000
Email: gs3100@gedgesystems.com.au
page 21
Email: gs3100@gedgesystems.com.au
page 22
Units
kg
g
t
lb
oz
Setting
Un 06
Un 07
Un 08
Un 09
Un 0A
Un 01
Units
lb t
oz t
N
kN
none
Email: gs3100@gedgesystems.com.au
page 23
Email: gs3100@gedgesystems.com.au
page 24
2) Access the CAL display (see page18) and then operate the ZERO key. The GS3100 automatically
zeros out all dead load offsets. Following this operation, which takes 5 seconds, the GS3100 will return to
its normal weighing mode with a gross display of zero. If the required adjustment was outside the GS3100's
range a message is displayed for 5 seconds. In this event the weighing installation should be reviewed.
Messages are :C HI or OFF Load cell input is positive and above the dead load offset adjustment range.
C LO or -OFF Load cell input is negative and below the dead load offset adjustment range.
Do not continue with weight calibration until ZERO has been obtained.
3.15.3 Span Calibration.
Span calibration involves one pass using an accurately known weight. Zero calibration MUST be conducted
immediately prior to span calibration and the normal weight operation key ZERO must not be used to zero the
display prior to span calibration. Operation of the key ZERO rather than the procedure CAL ZERO will cause
zero/span interactivity on span calibration.
The GS3100 exhibits absolutely no interactivity between span and zero. Any failure to return to zero
following span calibration is solely the result of hysteresis or non repeatability in the weighing installation.
1) Place a calibration weight of 50% to 100% of capacity on the scale. The GS3100 will calibrate an
installation using weights as small as 5% of capacity but this is not recommended.
2) Access the CAL display and then operate the SPAN key. The maximum capacity is displayed for
5 seconds. At this point, span calibration may be stopped by keying End Cal. To continue, when 00000 is
displayed, use SD, I and D to enter the required calibrated weight. If a minus sign appears following
operation of the SPAN key the calibration weight may not be on the scale.
3) Operate the SPAN key. The GS3100 now automatically adjusts the display to show the required
calibration weight. This process, which takes 5 seconds will end with the GS3100 resuming normal
weighing and a gross display of the calibration weight. If the required adjustment was outside the GS3100's
range a message is displayed for 5 seconds. In this event the weighing installation should be reviewed.
Messages are :C HI or OFF The load cell input is too high to be spanned to the required weight.
C LO
The load cell input is too low to be spanned to the required weight.
4) Remove the weight from the scale and check the zero return. Calibration is completed.
Check the installation if the display does not return to zero when the weight is removed.
Email: gs3100@gedgesystems.com.au
page 25
Description
Comments.
Frame (shield)
Control & Serial cable shield. Only connect the shield at the
GS3100 end of the cables. Do not ground the other end.
4800baud or 9600baud serial output for short distance (10m)
connection to a PLC/Computer.
Baud rate is selected via a link, LK2, situated at the front on the
GS3100's motherboard.
Direct connection to the user's 5VDC to 45VDC power supply
ground.
These outputs are each connected to the active side of the user's
5VDC to 45VDC power supply through the device (lamp, relay,
solenoid) that they are to drive. They are ON when sinking current
to the power supply ground and OFF when open circuit.
# The operation of these outputs depends upon the way the GS3100 is programmed during installation. See
Digital Setup Set Point Test on page 22 for further details.
NOTE:- Careful spike suppression and noise suppression on the control lines will be required to prevent
interference with the serial output. Keep the length of unshielded cable within the metal back shell to the
absolute minimum. For advice on spike suppression refer to page 28.
Email: gs3100@gedgesystems.com.au
page 26
Email: gs3100@gedgesystems.com.au
page 27
Email: gs3100@gedgesystems.com.au
page 28
The numbers below are the order in which the GS3100 transmits the sub-strings.
1
beg
DNo bd1
ID#
bd2
Wd
Un
Typ
bd
end
beg
DNo bd1
ID#
bd2
Wg
Un
Typ
bd
Wt
Un
Typ
bd
stat
bd
end
beg
DNo bd1
Wd
bd
end
beg
DNo bd1
Wd
bd
stat
bd
end
beg
DNo bd1
Wg
bd
Wt
bd
Wn
10
11
12
13
14
15
16
bd
Wn
Un
Typ
beg
DNo bd1
ID#
bd2
Wd
bd
stat
bd
op
bd
end
beg
DNo bd1
ID#
bd2
Wg
bd
Wt
bd
Wn
bd
stat
bd
op
bd
end
beg
DNo bd1
ID#
bd2
Wg
bd
Wt
bd
Wn
bd
stat
bd
op
S1
bd
17
18
bd
end
S2
bd
19
20
end
++
-
2
space
1
2
2
3
4
.
5
5
3
4
6
6
Email: gs3100@gedgesystems.com.au
page 29
b7
The
SEVEN
segment
equivalent of the ASCII
character sent to the GS3100 GS3100s interpretation of the
(with the parity bit ignored). incoming data.
Useful for checking messages
Normal data - all ok.
to be used in terminal mode. n
E1 RE not enabled.
Hardware fault.
Received Data Byte.
E2 REI not enabled.
This is the last word of 8 bits received by the
Hardware fault.
GS3100. This will depend on the serial comms E3 Overrun error.
setting shown on page 23. It will be either 7 data E4 Frame error.
bits b0 to b6 plus b7 which must have been set or E5 Parity error. Not tested
for 8 bit input.
cleared by the sender to produce EVEN parity OR
it will be 8 data bits b0 to b7 (and no parity bit).
Errors E3 to E5 do not have
conclusive causes as a variety
EVEN parity means there must be an even number of causes can result in any one
of bars illuminated including the decimal point. In of them. Likely causes are:Wrong baud rate.
the example above, which is the numeral 1, there
Wrong word size.
are four bars illuminated including the decimal.
Wrong parity.
Email: gs3100@gedgesystems.com.au
page 30
02
202
207
05
09
Note:- The optically isolated cards Option 202 and 207 require a customer supplied 9VDC, 100mA plug-pak
external power supply to achieve isolation. They can, however, be operated in a non-isolated mode without the
need of this external power supply.
6.2 Physical Installation, Baud Rate Selection, Output Selection, Set up, Format & Isolation.
Physical installation, setting the baud rate, Digital Setup settings, output format selection and isolated or
non-isolated operation are all contained in the preceding sections of this manual, on the pages listed below:-
Email: gs3100@gedgesystems.com.au
page 31
Isolated operation - LK1 in ISOLATED position and 9VDC 100mA supply required.
Only the 20mA output is used - isolated operation - LK1 in NOT ISOLATED position and
no external supply required.
Both the 20mA and RS-232C outputs are used - isolated operation - LK1 in ISOLATED
position and 9VDC 100mA supply required.
6.3 Optional Serial Output Cards - Output and Power Supply Connections.
The table and figure below shows the connections required for serial output option cards 02/202/207/09 and 05.
Serial Output Options 02, 202, 207, 05 and 09 Twenty Five Pin Female Output Connector.
Pin
No.
Option 202
Isolated RS-232C
Option 207
Isolated RS-232C &
20mA Passive CL
Option 02
Non-Isolated
RS-232C
Option 05
Non-Isolated 20mA
Active CL
Option 09
Non-Isolated
RS-422A
Frame (Shield)
Frame (Shield)
Frame (Shield)
Frame (Shield)
Frame (Shield)
Serial Out
Serial Out
-Tx 20mA
Serial Return
- Signal
Serial Out
+9VDC 100mA
External Supply
9VDC Supply
Ground
Signal Ground
Serial Out
RS-232C
+9VDC 100mA
External Supply
9VDC Supply
Ground
Signal Ground
RS-232C
+Tx 20mA
Serial Source
-Tx 20mA
Serial Return
+Tx 20mA
Serial Source
+ Signal
Serial Out
Signal Ground
6
7
10
11
Signal Ground
Note :- To minimize interference to the GS3100, and by the GS3100, a braided shield cable MUST be used with
metal back shells for the connectors at each end of the cable. Connect the shield to the metal back shells by
crimping it between the two shells, ensuring that there is not an air gap between the shells. Also connect the
shield to pin 1 of the connector. The shield should only be connected to the metal back shells at the GS3100 end
of the cable. The cable should be marked to identify the end that is connected to the GS3100.
Email: gs3100@gedgesystems.com.au
page 32
7.2 Physical Installation, Baud Rate Selection, Input/Output Selection, Set up & Isolation.
Physical installation, setting the baud rate, Digital Setup settings, input/output selection and isolated or
non-isolated operation are all contained in the preceding sections of this manual, on the pages listed below:-
If you are using serial a serial I/O card ONLY to use XON/XOFF control of its stream data
mode, you will need to select the output format. See page 29.
7.2.1 Isolation, the Position of Links LK1 on the Option Card & the need for a 9VDC Supply.
When the Option cards 07 and 08 are used for Optically Isolated RS-232C, RS-485 or RS-422 operation, the
use of an external 9VDC 100mA power supply is required. Also, the isolating links LK1 on the option card
MUST be placed in the ISOLATED position. See photograph on page 16. Option 07 includes a Passive 20mA
Current Loop I/O which can be selected for use instead of the RS-232C I/O. If the 20mA Passive Current Loop
output is to be used then the situation of the links LK1 and an external supply is not as obvious as the above.
The 20mA Current Loop is automatically isolated and requires no external 9VDC supply to achieve insolation.
In addition, though, it operates in an isolated mode. The links LK1 on the serial card MUST BE PLACED
IN THE NOT-ISOLATED POSITION. In summary, for non-isolated operation of either board, place the links
LK1 in the NOT ISOLATED position. For isolated operation see below:Option 07 - Isolated RS-232C operation - LK1 in ISOLATED position & 9VDC 100mA supply required.
Option 07 - Isolated 20mA CL operation - LK1 in NOT ISOLATED position & no 9VDC supply required.
Option 08 - Isolated operation - LK1 in ISOLATED position & 9VDC 100mA supply required.
Gedge Systems - Melbourne - Australia
Email: gs3100@gedgesystems.com.au
page 33
7.3 Device Number and Digital Setup Settings for serial I/O. (Not applicable to MODBUS.)
The GS3100's device number MUST BE SET in the range 01 to 99. You should note it down, because you will
need it to be able to talk to the GS3100 via serial input. Page 24 shows you how to enter a device number.
The table of Digital Setup settings below identifies those affecting serial I/O. Make your selections and program
them into the GS3100 referring to page 20 and to page 23 for the serial comms word composition setting.
Setup Settings Applicable to Serial I/O
Setting
Output
Example
out.io1
Comment
Select io1 for serial I/O or, if a serial I/O card is being used ONLY for XON/XOFF
flow control, refer to page 29 and make a format selection. This will include making
selections for the Units; Block Division" and Beginning & End settings.
The comments below refer ONLY to the way that the settings are used in serial I/O
Units
Block
Division
Beginning
& End
Un 01
bd 01
bE 01
bE 02
Stream
Output
Str c6
Email: gs3100@gedgesystems.com.au
page 34
Option 07
RS-232C
RS-232C
Frame (Shield)
Serial In
Serial Out
+9VDC 100mA
External Supply
9VDC Supply
Ground
Signal Ground
6
7
20mA Passive CL
Frame (Shield)
Option 08
RS-485
RS-485 Multi-Point
Frame (Shield)
Frame (Shield)
-Signal
Serial In and Out
+9VDC 100mA
External Supply
+Signal
Serial In and Out
Signal Ground
-Signal
Serial Out
+9VDC 100mA
External Supply
+Signal
Serial Out
Signal Ground
9VDC Supply
Ground
9VDC Supply
Ground
-Signal
Serial In
9
+Tx 20mA
Serial Out
-Tx 20mA
Serial Out
+Rx 20mA
Serial In
-Rx 20mA
Serial In
10
11
12
13
RS-422
+Signal
Serial In
Note 1:- Option 07's 20mA current loop I/O is passive and isolated. It includes internal diode bridges so the
polarity of the connection at the option 07 end is immaterial.
Note 2:- To minimize interference to the GS3100, and by the GS3100, a braided shield cable MUST be used
with metal back shells for the connectors at each end of the cable. Connect the shield to the metal back shells by
crimping it between the two shells, ensuring that there is not an air gap between the shells. Also connect the
shield to pin 1 of the connector. The shield should only be connected to the metal back shells at the GS3100 end
of the cable. The cable should be marked to identify the end that is connected to the GS3100.
Email: gs3100@gedgesystems.com.au
page 35
Email: gs3100@gedgesystems.com.au
page 36
Email: gs3100@gedgesystems.com.au
page 37
The one exceptional command is literally NN or nn transmitted to the GS3100 within brackets [nn].
This two character command asks for the GS3100's device number. The reply will comprise two characters,
being numerals in the range 01 to 99. In the examples that follow, a device number of one (01") is used.
When the GS3100 is asked to reply to a command, it replies with the same string structure as the command. For
example, the command to lock the keyboard is [01GPLL]. This command does not ask for a reply as it does
not contain a question mark. If a reply is required, a question mark must be added to this command, which is
then [01GPLL?]. The GS3100 will reply with [01GPLL], the only difference being the absence of the
question mark.
In all cases, a reply means that the GS3100 has understood the command. Generally, this also means that the
GS3100 has been able to carry out the command. There are a number of exceptions, which are noted in the
tables that follow.
For consistency, if the GS3100 is to transmit a value when replying to a command, it will always have the same
structure. It will comprise a sign and six digits with leading zeros and no decimal points.
Examples are +000125"; -000002"; +003285".
The computer can obtain from the GS3100 details of its setup. These include the number of decimal places.
Accordingly, the computer can convert the value sent by the GS3100 into a real number as required.
7.5.5.2 ID Number Command.
The ID number displayed and printed by the GS3100 can be entered via the GS3100's keyboard or a computer.
The ability to transfer this number between the computer and the GS3100 can permit operators to be identified to
the computer or to select recipes within the computer or other procedures.
Serial String content & position
1
9 10 11 12 13 14 15
[
[
[
n n I D ? ]
n n I D 7 5 4 9 2 ]
n n I D ? 7 5 4 9 2 ]
n n
I D
?
x x x x x
n n I D x x x x x ]
Email: gs3100@gedgesystems.com.au
page 38
The computer should always ask the GS3100 for a reply to a set point setting command. Two issues must be
considered when checking the set point value in the reply. They are:1. The GS3100 converts the set point setting to a value consistent with its installed setup. This involves rounding
the value so its least significant digit is in the same increment size as the GS3100's Count By setting. For
example, if the GS3100's Count By setting is 5, and the set point value transmitted to the GS3100 is 3, it will
round this number to 5. Similarly, a value of 7898 will be rounded to 7900. The computer can obtain the Count
By setting from the GS3100 and can therefore take this into account.
2. The GS3100 will completely ignore the set point value if it exceeds 120% of the GS3100's installed capacity.
The computer can obtain the GS3100's capacity settings and should therefore take this into account. The
GS3100 will also ignore a set point setting value that makes no sense. For example, a number which included
letters 00A2B.
Serial String content & position
1
9 10 11 12 13 14 15
[
[
[
n n S P 1 x x x x x x ]
n n S P 1 ? x x x x x x ]
n n S P 1 ? ]
[
[
[
n n S P 2 x x x x x x ]
n n S P 2 ? x x x x x x ]
n n S P 2 ? ]
]
n n
S P 1
S P 2
x x x x x x
[
[
n n S P 1 x x x x x x C ]
n n S P 2 x x x x x x N ]
Each part of the reply string which is not identical to the components of the command string, is:
C
x x x x x x
Gedge Systems - Melbourne - Australia
Email: gs3100@gedgesystems.com.au
page 39
[
[
n
n
n
n
*
*
*
*
*
*
*
*
]
?
]
n
*
The four character commands are shown below. Commands which do not automatically ask the GS3100 for a
reply are shown with a question mark, which asks for a reply. If they are transmitted without the added
question mark, the GS3100 will not reply:Commands that control the GS3100. (See also Function Lock settings on page 21).
LOCK KEYS - Locks the GS3100's keyboard. Locd L is
G P L L ?
displayed if any of the GS3100's keys are operated and they have
no effect. Reply [01GPLL]
UNLOCK KEYS - Unlocks the GS3100's keyboard.
G P U L ?
Reply [01GPUL]
ACQUIRE ZERO - Zeros the gross input emulating the zero key.
G P A Z ?
The GS3100 may not always acquire zero. See operation of the
ZERO key for the tests applied by the GS3100. A reply ONLY
means the command was understood. Ask for the gross weight
following this command to ensure zero was obtained.
Reply [01GPAZ]
ACQUIRE TARE - Acquires Tare and displays net. The GS3100
G P A T ?
may not always acquire tare. See operation of the TARE key for
the tests applied by the GS3100. A reply ONLY means the
command was understood. Ask for the net weight following this
command to ensure a tare was obtained. Reply [01GPAT]
DISPLAY GROSS - Displays gross weight . Reply [01GPDG]
G P D G ?
Email: gs3100@gedgesystems.com.au
page 40
[
[
n
n
]
]
n
*
The four character commands and example replies are shown below. All of the replies contain a value string of
7 characters comprising a plus sign followed by six digits with no decimal points. The two exceptions are
product and version, however, the string length is still 7 characters.
The above replies can be used to scale the weight and set point information sent to the GS3100, and received
from it. Since all communication with the GS3100 uses value strings which do not include decimal points, this
scaling will be a necessary part of any program written to communicate with the GS3100. Key scale factors are:GS3100's capacity. The installed capacity of the scale or weigh hopper will be the number of full scale divisions
multiplied by the count by setting and adjusted for the number of decimal places. A general formula for this is:Capacity = ( FULLSCALE x COUNT BY ) / ( 10^DECIMAL )
For the example on this page, assuming kg units, the formula becomes
which is
that is
( 3000 x 5 ) / ( 10^ )
( 15000 ) / ( 1000 )
15.000kg
3
If set point 1 is to be set to 15.000kg it will be sent to the GS3100 as [01SP1?15000] with no decimal places.
Email: gs3100@gedgesystems.com.au
page 41
[
[
[
n
n
n
n
*
*
*
*
]
?
n
*
The two character commands, with a question mark and example replies:-
M ?
Terminal Mode ON. This command places the GS3100 in the terminal
mode. All the GS3100's normal functions continue to operate in the
background, but the display and keyboard are completely under the control
of the remote computer. The GS3100 displays information sent to it with a
TD command AND sends ALL keystrokes made on its front panel
keyboard to the computer. The GS3100's keys have no meaning to the
GS3100 during this time, and so keys like ZERO have no effect, apart from
the fact that their key code is transmitted by the GS3100.
Keys that have been locked out in the GS3100's Digital Setup are not
locked out during terminal mode. In addition, keyboard locking using a
serially transmitted LL command also has no effect during terminal mode
operation. All the GS3100's serial I/O functions continue to operate in the
terminal mode. Hint:- Send a display message using the TD command
before turning on the terminal mode.
Reply [01TM] see below for the key codes sent by the GS3100.
Key Codes transmitted by the GS3100. In the terminal mode, the GS3100
transmits one string only, immediately any one of its front panel keys is
pressed. The key must be released before another key identity string will
be transmitted. The string comprises the structure shown on the left, where
nn is the GS3100's device number, K means key code and kk is a
two digit key code number 00 to 14. The code numbers for each key are:00 to 09 for numeral keys 0 to 9; 10 - ENTER; 11 - TARE; 12 - G/N;
13 - PRINT and 14 - ZERO. No code is transmitted if two keys are pressed
at exactly the same time.
Email: gs3100@gedgesystems.com.au
page 42
Terminal Restricted Mode ON. This command is exactly the same as the
TM command above with one important difference. The GS3100's
keyboard is locked out so it cannot send key operations. This is intended for
RS-485 Multi-Point operation where unscheduled transmissions would create
difficulties. The GS3100 displays Locd.tr if a key is pressed.
Terminal Mode OFF. This command turns off the terminal mode. The
GS3100 will immediately display weight. The front panel keyboard resumes
its normal weighing functions and any function locks resume their operation.
This comand does not turn off the stream weight data mode. Reply [01TF]
Terminal Mode - send the GS3100's display. This command asks the GS3100
to send to the computer the information shown on the GS3100's display. The
GS3100 does not need to be in the terminal mode to respond to this
command. In its normal weighing mode, the GS3100 will send the
information on its display, which will generally be the current weight, rather
than a message it may have received from the computer. The ? in the
message is completely optional as this command always results in a reply.
The reply string that the GS3100 sends to represent its display can be longer
than the SIX digits of the display. This is because the GS3100 uses only
ONE digit to display a decimal point and a number (or letter), whereas,
normal ascii uses one character space for a number (or letter) and another
character space for a decimal point. Accordingly, any instance of a number
and decimal on the GS3100's display, will be converted to TWO ascii
characters when the GS3100 sends its display image. There is a further
matter. The GS3100 can display a decimal point at D5, D4, D3, and D2 of its
six digit display. (It cannot display a decimal at D6 (the leftmost digit) or D1
(the rightmost).
If a computer has sent the GS3100 a message containing FOUR decimal
points, eg 12.3.4.5.6, this will be converted to TEN ascii characters in the
GS3100's reply to the T7 command.
A further point to consider is this. The GS3100 displays some characters
in the same appearance as numbers, for example S looks like 5. This is a
limitation of a SEVEN segment display. When the GS3100 sends its display
image, it may include numbers that are really being used as letters.
Email: gs3100@gedgesystems.com.au
page 43
[
[
n
n
n
n
Comments
10 11 12 13
T D H E L L O
]
T D ? H E L L O
The GS3100 has a SIX digit display. Because decimal points occupy the same space as numerals or letters, it is
able to pack SIX characters plus FOUR decimals into its SIX digit display. An example of this would be the
message 12.3.4.5.6.
An important point must be borne in mind when sending the GS3100 display messages containing multiple
decimal points. The GS3100 cannot display a decimal at Digit 6 or Digit 1 position ( D6 is the leftmost digit).
If it receives a message with decimals that would require placement at D6 it completely ignores the message. If
the message contains a decimal for the D1 location, the decimal is ignored and the message shortened ( the ]
bracket may get displayed in this case.)
The example shown is HELLO . Note the use of a space following HELLO to make up the full six digits.
The message contains SIX characters not including decimal points.
On receipt of the display string, the GS3100 converts the ascii characters to their 7segment display equivalent
and then displays them on its six digit seven segment display. If there is no seven segment equivalent to the
character sent to the GS3100, it displays a bar (-) instead of the character. When a reply is required by the
computer, the GS3100 will send the displayed message as it is shown on the GS3100's display, including any
bars (-) for characters that it could not convert to seven segments. See the table of ASCII to seven segment
characters on page 45.
Reply [01TD?HELLO ] where, instead of HELLO there is a letter, number or decimal as displayed by the
GS3100.
One final note:- Do not use a question mark at Digit 6 location as the GS3100 may interpret it as a request for
a reply rather than a character to be displayed.
Email: gs3100@gedgesystems.com.au
page 44
2
3
4
5
space
U
g
.
?
d
t
L
[
O
]
LF
^S
1
2
space
0D
^Q
&
'
++
,-
<
>
up
left
SUM
+-
pi
tau
none
^Q is XON
^S is XOFF
CR is Carriage Return
LF is Line Feed
Email: gs3100@gedgesystems.com.au
page 45
Email: gs3100@gedgesystems.com.au
page 46
Email: gs3100@gedgesystems.com.au
page 47
1
3
5
10
20
ASCII Mode
4800baud
73
92
110
156
248
9600baud
37
46
55
78
124
RTU Mode
19200baud
18
23
28
39
62
4800baud
35
44
53
76
122
9600baud
17
22
26
38
61
19200baud
9
11
13
19
30
Hex
Hex No
used in
serial I/O
$01
$00
Read
Write
STATUS INFORMATION.
b15 1=Weight Negative Offscale
b14 1=Weight Positive Offscale
b13 1=Weight Inscale
b12 1=Gross Overweight
b11 reserved for future use
b10 1=Net display selected
b09 1= Gross display selected
b08 1=In Motion, 0=Still
Email: gs3100@gedgesystems.com.au
page 48
Read
Write
Decimal
Hex
Hex No
used in
serial I/O
$02
$01
$03
$02
$04
$03
$05
$04
$06
$05
R/W
$07
$06
R/W
$08
$07
R/W
$09
$08
R/W
10
$0A
$09
R/W
11
$0B
$0A
12
$0C
$0B
13
$0D
$0C
14
$0E
$0D
15
$0F
$0E
Email: gs3100@gedgesystems.com.au
page 49
Hex
Hex No
used in
serial I/O
Read
Write
$10
$0F
Email: gs3100@gedgesystems.com.au
page 50
Email: sales@gedgesystems.com.au
page 51
9 TECHNICAL SPECIFICATION.
LOAD CELL INPUT
Sensitivity/Range :- 1.00uV/d to 60uV/d to a maximum
input of 45mV.
Display provides -4% to + 104% Full
Scale.
Linearity :/+ 0.01% Full Scale.
Span Stability :/+ 15ppm/
Zero Stability :+/-500nanoV/ C +/- 30ppm/C
of dead load offset.
Noise :1uV p-p RTI maximum.
Zero Adj. Range :- 1mV to + 20mV
Excitation :10VDC Short Circuit Protected.
Up to 8 x 350ohm Load Cells in
parallel 230mA max.
Excitation Sensing :- True differential remote sensing
using a separate pair of wires.
Source Impedance :- 2kohm maximum.
Input Impedance :- 10Mohm minimum.
Analog Filter :Active 2Hz 2 Pole Low Pass.
Digital Filter :A/D averaging 1 to 8 samples.
RE-CALIBRATION
Re-calibration :Recommended Re-calibration
interval 1 year.
FRONT PANEL
Weight Display :14mm high 7 segment High
Intensity red LED display to a
maximum of 199990 with minus (-)
sign.
Annunciators :6 LED lamps.
Controls :Positive action tactile feedback
key board of 15 keys behind a tough
water resistant membrane.
Rear Panel :-
Front :Body :-
oC
Power :-
OPTIONS
User Installed - A wide range of
serial output and serial input/output
options is available together with a
High Precision Analog Output
Option6HPF. They are supplied loose
for user installation.
Refer to pages 31 and 33 of this
manual for a complete listing and
description of the available serial
options and Page 14 for the Analog
Option.
CONTROL OUTPUTS
2 Outputs :One control output for each
set point. Each is an optically isolated
open collector capable of sinking
0.5Amp 5VDC to 45VDC from a user
supplied external power supply.
STANDARD SERIAL OUTPUT.
RS-232C serial ouput is supplied as
standard. Selectable baud rates are
4800baud or 9600baud.
Email: sales@gedgesystems.com.au
page 52
10 WARRANTY
Definition. In the following, Buyer shall mean the original purchaser of the equipment supplied by Gedge
Systems. Seller shall mean Gedge Systems, the manufacturer of the equipment.
Extent. This express warranty shall extend between Buyer and Seller only and shall cover all claims and costs
during the warranty period between Buyer and Seller only, and, except for warranty of title, is in lieu of all other
warranties, whether expressed or implied, including the implied warranties of fitness for a particular purpose and
merchantability. Seller shall not be liable for any special, incidental or consequential damages resulting from a
breach or alleged breach of this warranty.
Warrants. The Seller warrants to the Buyer that the equipment shall be manufactured by good workmanship of
materials free of defect and that such equipment when used and installed in accordance with the Seller's
instruction manuals shall perform in accordance with the equipment's specifications.
Repair/Replace/Exceptions. The Seller shall repair or replace at its discretion, free of charge, any equipment or
option which is returned to its facility within twelve (12)months of shipment to the Buyer, and which upon
examination proves to be a manufacturing defect.
The warranty does not apply to any equipment that has been :A. Repaired or modified by other than a duly authorized repair facility.
B. Installed or adjusted not in accordance with the equipment's manual.
C. Mishandled, abused or which in the Seller's sole judgment has been exposed to an environment for
which the equipment was not designed.
D. Operated outside of its enclosure or operated in an enclosure other than as supplied by the manufacturer.
Shipping Costs. This warranty applies to equipment received at the Seller's facility, freight prepaid by the
Buyer, and does not include return freight to the Buyer's premises.
Email: gs3100@gedgesystems.com.au
page 53