Beruflich Dokumente
Kultur Dokumente
Procedure For
Ultrasonic Testing of Weld Joints
(PE-UT and AUT)
EPCMD-2-QM00-JEP-ME-031
01
05.05.2014
ASR
PKD
SNS
00
31.08.2012
PKD
KSS / EMK
SNS
Rev
Date
Revision
Prepared
By
Checked
By
Approved
By
Reliance
Contents
1. Purpose ...3
2. Scope .3
3. References .3
4. Abbreviations and Terms ....3
5. Responsibilities ..4
6. Methodology .4
7. Records
....14
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 2 of 22
Reliance
1.
Purpose
The purpose of this document is to provide the guide lines for the Ultrasonic Testing
of the pipe weld using conventional manual pulse echo and automated phase array.
2.
Scope
The scope of this procedure covers the Ultrasonic testing of
i.
ii.
iii
b.
b.
This procedure may be used for the testing of the weld joints in on line service
having liquid / gases in the pipe line and is applicable for CS, AS and SS material
for all the jobs in Reliance Projects.
3.
References
ASME Sec VIII Div. I
ASME B31.3
Process Piping
ASME B31.1
Power Piping
ASME Sec. V
Non-destructive Examination
CEN/TS 14751
EPCMD-2-QM00-GUI-ME-002
EPCMD-2-QM00-TCR-ME-007
EPCMD-2-QM00-FMT-ME-025
4.
AUT
DAC
dB
Decibel
DTA
HAZ
HOD
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 3 of 22
Reliance
IIW
JERP
MUT
NDT
PE UT
PA UT
QMD
SNT-TC-1A
TCG
TR
TOFD
5.
UFD
UT
Ultrasonic Testing
Responsibilities
6.
Methodology
6.1
6.1.1
Panamatrics
Kraut Kramer
Sonatest
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 4 of 22
Reliance
Omniscan
The use of any other models of UFDs, than the recommended as above, shall have
prior approval from engineer-in-charge.
6.1.2
Probes
Probes used with UFD shall be selected from the following manufacturers,
6.1.3
Panamatrics
Kraut Kramer
Sonatest
Pipe
Dia
Probe Size
Parallel to weld
axis
Circumferential
to weld axis
2"
5-8 mm
45,70,(shoe)
45 (shoe)
5-8 mm
3"
5-8 mm
45,70,(shoe)
45 (shoe)
5-8 mm
4"
5-8 mm
45,70,(shoe)
45 (shoe)
5-8 and mm
6"
89 and
5mm
45,70
89mm
5-8 and 10 mm
> 8"
89mm
45,60/70
89mm
89mm
10 mm
TR
Probe
Couplant
The couplant shall be a manufactured gel product, oil or cellulose paste or other
suitable medium (oil and grease mixture ) which is non toxic and non corrosive. The
couplant must be removed after examination by means of wiping with a cloth with
the help of solvent / detergent.
6.1.5
Standard Blocks
6.1.5.1
For piping job the basic calibration block shall be a section of pipe of same nominal
size, schedule, heat treated, equivalent P number material (as per clause T-434. of
ASME Sec. V).
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 5 of 22
Reliance
Notches shall be prepared on outer as well as on inner surface of the pipe as
specified in fig. T 434.3 of ASME Sec. V. DAC shall be drawn with 3 points.
6.1.5.2
6.1.6
For non piping job, (above 20 diameter non piping welds), the
reference block
shall be used from the same job or a flat basic calibration block as per clause
T - 434.1.7.1. The reference block for non piping welds having diameter equal to
or less than 20dia., shall have curved calibration block and shall be used to scan
the jobs having curvature at the range of 0.9 to 1.5 times the basic calibration
blocks diameter. Non piping calibration blocks shall be as per T- 434.2 and cladding
calibration blocks as per T- 434.4.
UFD Operator
All the technicians shall under go a selection test and qualify the same prior to the
deployment on the job as per Procedure for the selection of UT technicians.
(EPCMD-2-QM00-JEP-ME-038).
6.1.7
6.1.7.1
6.1.7.2
When both sides scanning is not possible, then in that case, at least one side
scanning shall be carried out.
6.1.7.3
For identification of weld joints the optical letter V to be punched on the base
metal at 12 o clock position turning 90 degree (symbolic shown as > ) on the top
of the weld joint and shall be considered as the reference point ( >) or zero point.
All the recordable indications (print outs) shall be given with reference to zero point
in the UT report. Punching and marking on the weld joint shall be done by the NDT
Contractor. Reference points shall be hard punched on CS and AS joints and SS
joints shall be marked with permanent marker.
6.1.7.4
Surface temperatures shall not exceed 50C / subject to couplant and probe service
condition.
6.1.8
DAC Preparation.
6.1.8.1
DAC shall be prepared by using angle probes, taking into consideration the probe
index and the weld edge distance. First point at 80% FSH 5% shall be
established other two points shall be drawn based on the reference block (SDH or
notch).
6.1.8.2
DAC for longitudinal weld scanning shall be made from circumferential notch with
45 and 70 angle probes to detect defect parallel to weld axis. From longitudinal
notch, DAC is to be made from 45 angle probe to detect defect transverse to weld
axis.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 6 of 22
Reliance
6.1.8.3
Every NDE contractor shall establish the DAC on the standard reference block for
his scope of job and get it approved by authorized representative of Reliance before
starting examination on the actual job. Reliance NDT Coordinator shall maintain a
latest list of standard reference block.
6.1.9
6.1.9.1
Angle probe: Calibration range shall be selected as per the job requirement.
6.1.9.2
TR or normal probe: Calibration range shall be selected as per the job requirement.
6.1.10
Sensitivity
6.1.10.1
6.1.10.2
6.1.10.3
Scan Plan: Before actual scanning, scan plan shall be developed by the operator
and shall be approved by Reliance Engineer. The scan plan data shall be recorded in
Form: EPCMD-2-QM00-FMT-ME-231.
6.1.11
6.1.11.1
6.1.11.2
Angle beam probe: (To detect defect parallel to weld axis): All
welds
shall
be
scanned with two different angle beam probes. Angle of the probe is selected based
on the actual job thickness by preparing the weld geometry drawing and the
scanning location /areas. One angle probe is 450 and other angle is selected based
on the base material thickness (90-t, where t is the thickness of the pipe). However
for thickness <15 mm, 70 angle probe shall be used.
6.1.11.3
Angle beam probe (To detect defect transverse to weld axis): Scanning shall be
carried out from the top of the weld reinforcement for the pipes having sizes 10
Inch or less (ground flat to ensure the proper contact) parallel to weld axis. The
probes shall be moved throughout the pipe circumference in such a way that the
entire volume of the weld is covered. Scanning shall be done clockwise and shall be
repeated in anticlockwise direction. Only 45 angle probe shall be used for this
purpose.
For pipe of sizes 12 Inch dia or more, angle beam scanning to detect defect
transverse to weld axis, the examination shall be performed from the base metal on
both side of the weld cap in both weld axis direction, if the weld cap is not
machined or ground flat, with the help of a scanner where two 45 degree miniature
probes are used. One probe used as transmitter and other probe is used as
receiver.
6.1.12
Scanning overlap
Each fresh scan shall have an overlap of 10% of the crystal size over the previous
Scan.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 7 of 22
Reliance
6.1.13
Scanning speed
The probes shall be moved throughout the pipe circumference at a suitable
speed but not more than 150 mm / Sec.
6.1.14
Scanning Sensitivity
6.1.14.1
Angle probe: Ref. dB of DAC + 14dB as per Article T- 481 of ASME Sec V or as
per the job requirement.
6.1.14.2
6.1.15
Evaluation
6.1.15.1
6.1.15.2
6.1.15.3
In case of angle probe, all indications equal to or in excess of 20% of the primary
reference level shall be investigated in order to determine their shape, size, identity
and location.
6.1.15.4
In case of angle probe, any imperfection which produce a response above 50% of
reference level shall be recorded from the uncorrected area with reference from
the reference point marked on the joint.
6.1.15.5
All indications detected with one probe shall be further evaluated using appropriate
combination of probe angle (from other side of the weld joint) so that length,
depth and orientation can be accurately determined and the type of discontinuity
can be established.
6.1.15.6
6.1.15.7
Where ever the indications exceeds the acceptance standard , graph to be plotted
based on probe location to determine shape, size, identity and location of the
defect. Interpretation shall be carried out at reference sensitivity level (dB).
6.2
6.2.1
Operator Qualification
All the technicians shall under go a selection test and qualify the same prior to
the deployment on the job as per Procedure for the selection of UT technicians.
(EPCMD-2-QM00-JEP-ME-038).
6.2.2
6.2.2.1
The ultrasonic Phased Array instrument shall be pulse echo type and shall be
equipped with a calibrated dB gain or attenuation control stepped in increments of
1dB or less. The system shall be capable of generating and displaying linear scan
and sector scan images, which can be digitally stored and recalled for subsequent
review.
6.2.2.2
6.2.2.3
Examination Personnel shall use a real-time linear and or sector scan image with
B / C / D / Sectorial scan presentation during scanning to ensure that proper data
has been collected.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 8 of 22
Reliance
6.2.2.4
The Phased Array system shall be capable of using an external storage device.
A remote portable PC may be used for this purpose.
6.2.2.5
In addition to external data storage, the PC shall also be used by the Data analysis
personnel for analyzing data subsequent to the completion of data collection.
6.2.2.6
6.2.2.7
Any control, which affects the instrument linearity (e.g. Reject) shall be in the off or
minimum position during instrument calibration, system calibration and
examination.
6.2.3
Probe Selection
Phased Array Probe configuration shall be suitable for the inspection of the job.
Phased Array Probe frequency shall be either 1 to 5 MHz or as per the job
requirements.
6.2.4
Couplant
Water, Glycerin or Cellulosic gel equivalent oil grease mixture may be used as
couplant when performing calibration and examinations. Water pump may be used
if required.
6.2.5
Standard Block
The calibration block shall be prepared as per ASME Sec V Fig T-434.2.1 and
T-434.3. IIW Blocks shall be used for time base calibrations. Multiple reference
blocks for calibration may also be used as per the requirements of AUT system.
6.2.6
6.2.6.1
Calibration shall be performed from the surface of the calibration block which
corresponds to the component surface to be examined.
6.2.6.2
System calibration shall include the complete ultrasonic examination system. Time
base calibration shall be suitable for the beam paths which will be used during the
examination, unless otherwise specified.
6.2.6.3
6.2.6.4
6.2.7
6.2.7.1
Verification that the input information is correct and that the phased array system
is working properly shall be checked as follows.
6.2.7.2
Select the Angle Cursor and adjust its position so that is displays A-scan
information for the 450 angle of refraction or the minimum angle that will be used
in the Sector Scan.
6.2.7.3
Using the 4-inch (100 mm) radius of the IIW block, peak the signal shown
A-scan display.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 9 of 22
on the
Reliance
Note: Although the sector scan may indicate different amplitude responses form
other angles. Only use the A-scan response associated with the 450 angle or
minimum angle used for propagation.
6.2.7.4
Mark the beam exit point, on the transducer wedge. This beam exit location is only
valid for the particular angle of propagation.
6.2.7.5
Using the primary angle of beam exit locator, measure the actual angle of
propagation by peaking the response in the A scan display using the plexi-glass
insert as indicated on the IIW block using the beam exit point location.
6.2.8
6.2.8.1
Position the Angle cursor and adjust its position so that it displays A-scan
information for the angle, used for refraction.
6.2.8.2
Place the transducer so that reflections from 2 inch (50mm) and 4-inch (100mm)
radius reflectors on the IIW block are peaked and observed individually on the
A-scan display.
6.2.8.3
Using the A-scan Cursors, measure the distance between the 2 and 4-inch signals
markings. This result shall be 2 0.1 inches.
6.2.8.4
If the measured separation between the signals is too large (grater than 2.1 inch)
decrease the Shear Velocity parameter under the Part Setup Menu. Similarly, if the
measured distance is too short (less than 1.9 inch), increase the velocity value.
Repeat adjustment until an acceptable value is achieved.
6.2.8.5
With the transducer remaining in the peaked position, measure the beam path of
the four-inch radius reflector using a cursor in the A-scan display.
6.2.8.6
The value should measure to be 4 0.1 inches. If this measurement is less than
3.9 inches, increase the value of the Delay parameter until the measurement is
correct. If this value is grater than 4.1 inch, decrease the Delay parameter until the
measurement is correct.
6.2.9
6.2.9.1
The Phased array instrument shall be calibrated for Sensitivity and Wedge Delay as
per the requirements.
6.2.9.2
The Sensitivity Calibration will provide the required gain adjustment for each
refracted angle and sound path used.
6.2.9.3
6.2.9.4
Peak up this signal from the calibration reflector and scan by the phased array
probe through all the different angles or focal laws.
6.2.9.5
Scan back over the calibration reflector through all the refracted angles or
laws.
6.2.9.6
The phased array system will calculate the required gain needed at each focal law
to adjust the required amplitude needed.
6.2.9.7
The Wedge Delay Calibration shall be calibrated for True Depth for the reflector
with the angles used in calibration.
6.2.10
focal
A Distance Amplitude Correction (DAC) Curve or a Time Corrected Gain (TCG) shall
be used to calibrate for attenuation in the material at the sound paths utilized
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 10 of 22
Reliance
during calibration
documented.
6.2.11
and
examination.
This
shall
be
digitally
recorded
and
6.2.12
6.2.12.1
6.2.12.2
With any substitution of power utilized the same type of source (e.g. a change
batteries).
6.2.12.3
6.2.13
6.2.13.1
Perform the following if any point of the DAC curve has decreased 20% or 2 dB of
its amplitude, or any point on the sweep line has moved more than 10% of the
sweep division reading.
6.2.13.2
Void all examinations performed after the last valid calibration verification.
6.2.13.3
6.2.13.4
6.2.13.5
Perform the following if any point on the DAC curve has increased more than 20%
or 2 dB of its amplitude.
6.2.13.6
6.2.14
Recalibration
6.2.14.1
6.2.14.2
6.2.14.3
6.2.14.4
Couplant change.
6.2.15
6.2.15.1
Scanning surface should be free from weld spatter, foreign materials and any
irregularities that would interfere with the free movement of the search unit.
of
The weld surface under examination shall be prepared to a smooth profile with a
section flat at the top of weld to enable good contact with probe to permit
examination.
For pipe of sizes 12 Inch dia or more, angle beam scanning to detect defect
transverse to weld axis, the examination shall be performed from the base metal on
both side of the weld cap in both weld axis direction, if the weld cap is not
machined or ground flat, with the help of a scanner where two 45 degree miniature
probes are used. One probe used as transmitter and other probe is used as
receiver.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 11 of 22
Reliance
In case of spiral and longitudinal weld joints, the top surface as well as protrusion
of weld shall be flushed and both the sides of the weld joint shall be ground to
smooth surface (150 mm (length ) for the examination.
6.2.15.2
Surface temperature shall not exceed 500 C / subject to couplant and probe service
temperature.
6.2.15.3
6.2.16
6.2.16.1
Before actual scanning, scan plan shall be developed by the operator and shall be
approved by Reliance Engineer. The scan plan data shall be recorded in Form:
EPCMD-2-QM00-FMT-ME-231.
6.2.16.2
The specific of the weld volume, weld identification and location shall be identified.
6.2.16.3
The examination volume shall the weld metal and 1 inch from the toe of the weld
both sides. This area includes the heat-affected zone.
6.2.16.4
6.2.16.5
6.2.16.6
Two linear scans shall be performed from the center of the weld from both sides if
required to ensure the weld and HAZ coverage. In case of pipe to fitting joint, two
scan shall be carried out from one side.
6.2.16.7
For thicker materials, multiple liner scans shall be performed to the degree to cover
the required volume of weld and base metal.
6.2.16.8
Scans shall be parallel to the weld using the linear scans both side of the weld axis.
6.2.16.9
The rate of search unit movement shall be limited to a maximum of 3 inches per
second for welds unless calibration has been verified at a higher speed.
6.2.16.10 Perform the examination from both sides of the weld, where practical, or from one
side as a minimum.
6.2.16.11 To detect the defects transverse to weld axis, scanning shall be carried out from the
top of the weld reinforcement (ground flat to ensure the proper contact) parallel to
weld axis. The probes shall be moved throughout the pipe circumference in such a
way that the entire volume of the weld is covered. Scanning shall be done clockwise
and shall be repeated in anticlockwise direction. Only 45 angle probe shall be used
for this purpose.
For pipe of sizes 12 Inch dia or more, angle beam scanning to detect defect
transverse to weld axis, the examination shall be performed from the base metal on
both side of the weld cap in both weld axis direction, if the weld cap is not
machined or ground flat, with the help of a scanner where two 45 degree miniature
probes are used. One probe used as transmitter and other probe is used as
receiver.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 12 of 22
Reliance
6.2.17
Evaluation
6.2.17.1
6.2.17.2
6.2.17.3
6.2.17.4
6.2.17.5
6.2.17.6
6.2.18.
Examination Record
be
Computerized image technique shall have the following data / information recorded
in the system.
6.2.18.1
6.2.18.2
6.2.18.3
6.2.18.4
6.2.18.5
6.2.18.6
6.2.18.7
6.2.18.8
6.2.18.9
Calibration data.
Serial no.
6.3
Acceptance Criteria
Acceptance criteria for nozzle welds and piping welds shall be as per:
ASME Sec VIII Div. I. Appendix 12.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 13 of 22
Reliance
6.4
6.5
Post cleaning
Couplant (grease / oil) applied to weld joint for ultrasonic examination shall be
cleaned properly after completion of the examination by respective NDT agencies.
The couplant must be removed after examination by means of wiping with a cloth
and with the help of solvent / detergent.
7.
Record
7.1
UFD Print out of indication / reports and or graph plotting to be attached along
with the test report for repair. The report shall be in the form of data in the soft
form or hard copy as per the requirements.
7.2
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 14 of 22
Reliance
PE-UT Technique Sheet No. 1
Weld joint
Fit up data
Surface preparation
Diameter
Pipe Dia
Probe Size
Parallel to weld
axis
Circumferential
to weld axis
TR
Probe
2"
3"
5-8 mm
5-8 mm
45,70, (shoe)
45, 70, (shoe)
45 (shoe)
45 (shoe)
5-8 mm
5-8 mm
4"
5-8 mm
45 (shoe)
5-8 mm
6"
89 and
5 mm
5-8 and
10 mm
> 8"
89 mm
45, 70
89 mm
45, 60/70
89 mm
89 mm
10 mm
Scanning technique
(a)
TR probe scanning shall be carried on the area where angle beam scanning is to be
carried out. The both side of the weld joint (up to two skip of the max angle probe to
be used) and top weld shall be scanned with TR probes to identify any reflectors
which may interfere during angle beam scanning.
(b)
Angle beam Scanning(to detect defect parallel to weld axis) to be carried from both
side of the weld side from to 1 skip distance or 1 to 2 skip distance depending
upon the requirement .
(c)
Angle beam Scanning (to detect defect transverse to weld axis) to be carried out as
detailed in clause no. 6.1.11.3 of this procedure.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 15 of 22
Reliance
PE-UT Technique Sheet No. 2
Weld joint Type
Fit up data
Frequency
Pipe
Dia
Probe
Size
58mm
58mm
58mm
89 and
5mm
2"
3"
4"
6"
> 8"
Probe/Shoe
Angle of probe/shoe
TR
Probe
Parallel to weld
axis
Circumferential
to weld axis
45,70,(shoe)
45 (shoe)
5-8 mm
45,70,(shoe)
45 (shoe)
5-8 mm
45,70,(shoe)
45 (shoe)
5-8 mm
45,70 89mm
5-8 and 10 mm
45,60/70
89mm
89mm
10 mm
89mm
Scanning technique
(a) TR probe scanning shall be carried out on one side of the joint on the area where
angle beam scanning is to be carried out.. The scanning to be carried on the
surface ( up to two skip of the max probe angle ) to be used, top weld and heat
effected zone of the fitting side to identify any reflectors which may interfere
during angle beam scanning from
(1) Pipe side in case of pipe to fitting joint
(2) Pipe with lower thickness in case of pipe to pipe joint of unequal thickness
(b)
(c)
(d)
Angle beam scanning (to detect defect parallel to weld axis) to be carried out from
HAZ of fitting side to other side of weld up to 1 skip distance or 1 to 2 skip
distance depending on the requirement.
Angle beam Scanning (to detect defect transverse to weld axis) to be carried from
top of the weld as detailed in clause no. 6.1.11.3 of this procedure.
Scanning shall not be done from the thicker pipe or fitting side where inside
tapering has been carried out.
Page 16 of 22
Reliance
PE-UT Technique Sheet No. 3
Weld joint
: Set-On nozzle
Fit up data
Thickness
Surface preparation
Diameter
Frequency
Pipe Dia
2"
3"
4"
6"
> 8"
Probe
Size
5-8mm
5-8mm
5-8mm
89 and
5mm
89mm
Probe/Shoe
Angle of probe/shoe
Parallel to weld
axis
Circumferential
to weld axis
45,70,(shoe)
45,70,(shoe)
45,70,(shoe)
45 (shoe)
45 (shoe)
45 (shoe)
45,70 89mm
45,60/70
89mm
89mm
TR
Probe
5-8 mm
5-8 mm
5-8 mm
5-8 and 10
mm
10 mm
Scanning technique
(a) TR probe scanning to be carried out on branch nozzle to identify any reflectors
which may interfere during angle beam scanning. In addition to that, TR probes
scanning to be carried out from equipment in side on the weld joint wherever
possible.
(b) Angle beam scanning (to detect defect parallel to weld axis) to be carried out with
two angle beam probes from to 2 skip distance based on the requirement
of the branch nozzle...
(c) Angle beam Scanning (to detect defect transverse to weld axis) to be carried
from top of the weld as clause no. 6.1.11.3 of this procedure.
Page 17 of 22
Reliance
PE-UT Technique Sheet No. 4
Weld joint
: Set-In nozzle
Fit up data
: 9 mm and above
Probe
Frequency
Scanning technique
(a) TR probe scanning from nozzle inside and from shell inside/outside on the weld to
cover entire weld volume. Also on shell side up to datum marking to identify any
reflectors which may interfere during angle beam scanning.
(b) Angle beam scanning (to detect defect parallel to weld axis) to be carried out
from shell inside or out side from skip distance to 1 skip distance as per
datum marking.
(c) Angle beam scanning (to detect defect transverse to weld axis) to be carried from
top of the weld as detailed in clause no. 6.1.11.3 of this procedure.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 18 of 22
Reliance
PE-UT Technique Sheet No. 5
Weld joint
Fit up data
Tangential nozzle
Datum mark on shell side .
Surface preparation :
Thickness
9 mm and above
Probe
Frequency
Scanning technique
(a) TR probe scanning from nozzle inside and from shell inside/outside on the weld to
cover entire weld volume. Also on shell side up to datum marking to identify any
reflectors which may interfere during angle beam scanning.
(b) Angle beam scanning (to detect defect parallel to weld axis) to be carried out either
from shell inside or out side from skip distance to 3 skip distance based on
the requirements to cover the total weld volume and as per datum marking.
(c) Angle beam scanning (to detect defect transverse to weld axis) to be carried from
top of the weld as detailed in clause no. 6.1.11.3
of this procedure.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 19 of 22
Reliance
PE-UT Technique Sheet No. 6
Weld joint
Fit up data
Surface preparation :
Thickness
Probe
Frequency
Scanning technique
(a) TR probe scanning from Dollar plate nozzle up to the weld and from shell inside or
outside, top of the weld and heat effected zone to cover entire weld volume to
identify any reflectors which may interfere during angle beam scanning.
(b) Angle beam scanning (to detect defect parallel to weld axis) to be carried out
either from shell inside or out side from skip distance to 1 skip distance as per
requirement.
(c) Angle beam scanning (to detect defect transverse to weld axis) to be carried from
top of the weld as detailed in clause no. 6.1.11.3
of this procedure.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 20 of 22
Reliance
PE-UT Technique Sheet No. 7
Weld joint
Fit up data
Surface preparation :
Thickness
Probe
Frequency
Scanning technique:
(a) TR probe scanning from nozzle inside to cover entire weld volume. Also from shell
inside or outside up to datum marking to identify any reflectors which may
interfere during angle beam scanning.
(b) Angle beam scanning (to detect defect parallel to weld axis) to be carried out at
right angle to the weld axis on shell side with two angle beam probes from to 1
skip distance based on the requirement.
(c) Angle beam Scanning ( to detect defect transverse to weld axis) to be carried from
top of the weld as detailed in clause no. 6.1.11.3 of this procedure.
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 21 of 22
Reliance
EPCMD-2-QM00-JEP-ME-031, Rev:01
Confidential Reliance. All rights reserved
Page 22 of 22