Beruflich Dokumente
Kultur Dokumente
The standard of comparison must be of the same character as the measurand, and usually
is prescribed and defined by a legal or recognized agency or organization.
Example:
NIST : The National Institute of Standards and Technology.
NBS : The National Bureau of Standards.
ISO : The International Organization for Standardization.
ANSI : The American National Standards Institute.
The engineer is not only interested in the measurement of physical variables but also
concerned with their control. The two function are closely related, however because one
must be able to measure a variable such as temperature or flow in order to control it. The
accuracy of control is necessarily dependent on the accuracy of measurement.
The overall schematic of the generalized measurement systems is quite simple and as one
might suspect the difficult problems are encountered when suitable devices are sought to
fill the requirements of each of the boxes on the schematic diagram.
Definition of Terms
The following terms are often employed to describe the quality of an instruments reading.
Range (Campo de Medida)
The region between the limits within which a quantity is measured, received or
transmitted, expressed by starting the lower and upper range values.
Example: 0 to 150 oF, 20 to 200 psi.
Span (Amplitud o Alcance)
The algebraic difference between the upper and lower range values.
For example:
a) Range 0 to 150 oF , span 150 oF.
b) Range -20 to 200 oF, span 220 oF.
c) Range 20 to 150 psi, span 130 psi.
150
0.5 x 200
= 150 1 or 149 Kpa and 151 Kpa.
100
The difference between the instruments reported values during repeated measurements of
the same quantity. Typically, this value is determined by statistical analysis of repeated
measurement.
Repeatability
Is the ability of an instrument to reproduce the same measurement each time the same set
of conditions is repeated. This does not imply that the measurement is correct, but rather
that the measurement is the same each time.
Example:
The concept of accuracy and repeatability in measurements can be illustrated by the
throw of darts.
In summary
a) Poor Repeatability means poor Accuracy.
b) Good Accuracy means good repeatability.
c) Good Repeatability does not necessarily mean good Accuracy.
Repeatability does not include Hysteresis.
Sensitivity (Sensibilidad)
The change of an instrument or transducer output per unit change in the measured
quantity. A more sensitive instrument reading changes significantly in response to
smaller changes in the measured quantity. Typically an instrument with higher sensitivity
will also have better repeatability and higher accuracy.
Resolution
The smallest increment of change in the measured valve that can be determined from the
instrument readout scale.
Dead Band (Banda Muerta)
In process instrumentation the range through which an input signal may be varied upon
reversal of direction, without initiating an observable change in output signal. Dead band
is usually expressed in percent of span.
Hysteresis
Example 1
A temperature transducer has a range of 20 o to 250oC. A measurement results in a value
of 55oC for the temperature. Compare the errors if the accuracy is:
a) 0.5 % FS.
b) 0.75 % of span.
c) 0.8 % of reading.
What is the possible temperature in each case?
Solution:
a) Error =
0.5%(250 o C )
= 1.25 o C
100%
Thus, the actual temperature is in the range 53.75 oC and 56.25 oC.
b) Error =
0.75%.(250 20) o C
= 1.725 o C
100%
Thus, the actual temperature is in the range 53.275 oC and 56.725 oC.
c) Error =
0.8%(55 o C )
= 0.44
100%
Thus, the actual temperature is in the range 54.46 oC and 55.44 oC.
Example 2
Two pressure instruments with a range of 0 to 100 psi are measuring a process value of
50 psi. The accuracy of both devices is 1% FS. Data obtained from five measurements
are listed in the following table; determine which instrument has a greater degree of
repeatability.
Measurement
1
2
3
4
5
Instrument A (psi)
49.9
49.7
50.1
49.8
50.2
Instrument B (psi)
49.9
49.6
50.4
49.7
50.5
Solution:
From the measurements shown, it can be seen that the instrument A is more repeatable.
The measurements from both instruments are within the tolerance expressed by the 1%
accuracy stated ( 1 psi).
Example 3
A force sensor measures a range of 0 to 150 N with a resolution of 0.1% FS. Find the
smallest change in force than can be measured.
Solution:
0.1%(150 N )
= 0.15 N
100%
Standards
In order that investigator in different parts of the country and different parts of the world
may compare the results of their experiments on a consistent basis. It is necessary to
establish certain standard units of length, mass, time, temperature, pressure, etc. The
National Institute of Standards and Technology (NIST) have the primary responsibility
for maintaining these standards in the United States.
A Dimension defines a physical variable that is used to describe some aspect of a
physical system. The fundamental value associated with any dimension is given by a
Unit.
A Unit defines a measure of a dimension.
Dimension
Length
Mass
Time
Unit
Meter
Kilogram
Second
A Primary Standard defines the value of a unit it provides the means to describe the unit
with a unique number that can be understood throughout the world. The primary standard
then assigns a unique value to a unit by definition.
C = K 273.15
F = oR 459.67
o
F = 1.8 oC + 32
o
Electrical Standards
All electrical units originate from the definition of the Ampere. One ampere is the current
that produce a magnetic force of 2*10-7 N/m on a pair of thin parallel wires carrying that
current and separated by 1 m. The remaining electrical units, such as volts and ohms, can
all be derived from the value of the ampere and the basic units of mass, length and time.
10