Emotion (2007) / Spirit


CT

Software
System
Installation of SOMARIS/5 VB38/VB39

This document is valid for:


SOMATOM Emotion 6-slice Configuration
SOMATOM Emotion 16-slice Configuration
Siemens, 2012

10165977
10165888
10046789
10046766

Siemens, 2012
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic documents and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are reserved.

Print No.:
Replaces:

CT02-025.816.04.10.02
CT02-025.816.04.09.02

English
Doc. Gen. Date: 05.13
SSME CS
n.a.

2012

Copyright / Version / Disclaimer

1Copyright / Version / Disclaimer

Copyright
Siemens, 2012 refers to the copyright of a Siemens entity such as Siemens Aktiengesellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA,
Siemens Healthcare Diagnostics Inc. - USA and/or Siemens Healthcare Diagnostics Products GmbH - Germany.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents correspond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation as is without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understanding of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 2 of 58
SSME CS

Siemens, 2012

Table of Contents

0Table of Contents

1 _______ General ________________________________________________________ 5


Definitions and Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Vsim only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Software Kit for ICS, IRS, and IES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Software Kit VB38 for ICS Tower 8B, IRS, and IES . . . . . . . . . . . . . . . . . . . . . . . . .
Software Kit VB39 for ICS Tower 8B, IRS, and IES . . . . . . . . . . . . . . . . . . . . . . . . .
Software Kit VB38 SP2 for ICS Tower 9B/9B2, IRS, and IES . . . . . . . . . . . . . . . . . .

10
10
10
10

Images and Raw Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11


Required System Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Host Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Report.htm and License.dat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

12
12
12
12
13

Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Backup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
CD Write Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Creating Backup Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

2 _______ SW Installation ICS _____________________________________________ 17


Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
BHC adaptation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Required Data Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
System Type and Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

17
17
17
18

Installation with Recovery DVD (for SP4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19


Installation without Recovery DVD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Starting the Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Installation running . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
VB38B SP1 and SP3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
VB38B SP1 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
VB38B SP3 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Restore Steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Restore 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Restore 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Completion of Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Importing Reference Images. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setting the Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuring the Keyboard Language and enabling Windows XP Access . . . . . . . . .
BHC Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Additional Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setting Auto Filming, Auto Viewing, or Auto 3D in the IES . . . . . . . . . . . . . . . . . . . .
Siemens Remote Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuring AutoReport Mon_HQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 3 of 58
SSME CS

31
31
31
31
31
32
33
35
35
36

Emotion (2007) / Spirit

Table of Contents

3 _______ SW Installation of the IRS ________________________________________ 37


Installation of the IRS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
IRS Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

4 _______ SW Installation IES _____________________________________________ 38


Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Required Data Media. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Starting the Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Installation with Recovery DVD (for SP4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Installation without Recovery DVD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Starting the Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Installation running . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
VB38B SP1 and SP3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VB38B SP1 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VB38B SP3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Restore Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Completion of Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setting the Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuring the Keyboard Language and enabling Windows XP Access . . . . . . . . .

45
45
45
45

Siemens Remote Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

5 _______ Additional Work Steps ___________________________________________ 48


Software Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
IES. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
ICS and IES. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Monitor Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Tower 9B/9B2 Information Identification Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Final Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Backup at the ICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Disabling Windows XP Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Backup at the IES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Disabling Windows XP Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Changes to previous Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 4 of 58
SSME CS

Siemens, 2012

General

1-

Definitions and Abbreviations

1General

Abbreviations

Definitions

ACR-NEMA

American College of Radiology - National Electrical Manufacturers


Association

BIOS

Basic Input Output System, lowest PC configuration settings

AET

Application Entity Title - Name of DICOM service participating in the


communication either as a user (SCU) or a provider (SCP)

CD

Compact Disk

CDR

Compact Disk Recordable

CT

Computed Tomography

DICOM

Digital Image Communication in Medicine (ACR-NEMA standard)

DHCP

Dynamic Host Configuration Protocol - procedure for assigning a


numeric IP address to nodes in which blocks of addresses are leased
to clients on a first-come, first-serve basis. Typically, DHCP is combined with WINS. This means that an IP address is both dynamically
assigned and automatically associated with a domain name. This can
be combined with DNS, resulting in a dynamically assigned numeric
IP address that is automatically associated with both NetBIOS and
Internet names.

DNS

Domain Name Services - address resolution procedure that associates numeric IP addresses with Internet-style names

FTP

File Transfer Protocol - a client-server application that transfers files


from one computer to another

Gateway

A display that controls traffic between two networks using factors


above the network layer

ICS

Image Control System - control computer for the entire CT scanner,


with interface to the acquisition system; syngo CT Acquisition Workplace.

IES

Imaging Evaluation System; syngo CT workplace

IMS

Imaging System, consisting of ICS, IRS, and console

IRS

Image Reconstruction System - reconstruction computer

LAN

Local Area Network (here, always Ethernet)

LMHOSTS

Address resolution procedure which statically associates numeric IP


addresses with NetBIOS names. The LMHOSTS file is checked to
see if the name is being addressed in this file.

MEP

Market Entrance Phase

MOD

Magneto-Optical Disk

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 5 of 58
SSME CS

Emotion (2007) / Spirit

General
Abbreviations

Definitions

NetBIOS

Network Basic Input Output System - a collection of network services


and programming interfaces. NetBIOS services run on network protocols such as TCP/IP.

NIC

Network Interface Card - an expansion card that provides a physical


connection to the network

NT

"New Technology" Operating System - trademark of Microsoft, Inc.

OI

Operating Instructions

PB

Patient Browser

PPP

Point-to-Point Protocol - a data link and network protocol designed for


use over serial connections (such as dial-up modems)

RAS

Remote Access Services - a Windows NT service that supports


dial-in and dial-out connections using modems and ISDN.

Router

A multiport computer or piece of dedicated hardware that examines


network layer addresses of packets on a network and retransmits
them on a different network, if necessary

SCU

Service Class User

SCP

Service Class Provider

TCP/IP

Transport Control Protocol/Internet Protocol - the most widely used


standard for networking

UI

User Interface

USC

Uptime Service Center

WINS

Windows Internet Name Service - address resolution procedure that


dynamically associates numeric IP addresses with NetBIOS names.
It is possible to combine DNS and WINS.

XP

"eXPerience" Operating System - trademark of Microsoft, Inc.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 6 of 58
SSME CS

Siemens, 2012

General

Safety Information

1.1

Only qualified and trained personnel are allowed to perform CT system installation, service,
maintenance, and quality assurance.
To avoid risk of injury to personnel or damage to the system, the safety instructions must
be read and observed as described in the
General Safety Notes (TD00-000.860.01.xx.xx)
and
Product-specific safety notes (CT02-025.860.02.xx.xx) (CT02-025.860.02 / Product-specific safety notes)
The product-specific safety notes contain important information about:

General safety information about the gantry, patient table, and laser products.
Attachment of screws and nuts, adjustment instructions for torque wrench, and the use
of Loctite.

Safety advice (hazard keys)

Vsim only

NOTE

Switching off the computer without shutting down the system!


Otherwise data loss or system corruption is possible.
To turn off or shut down the computer properly, use Options
> End Session > Shutdown System in the syngo user interface.

NOTE

Mismatch in DICOM conformance can cause incorrect


patient treatment.
Please confirm customer receipt of the validation summary
document as part of the ATP.

NOTE

Connecting non-validated equipment can cause generally


incorrect treatment and/or physical injury!
Persons who connect additional equipment to the medical
device are considered to be system configurators and are
therefore responsible for ensuring compliance with the valid
version of relevant standards (e.g. system standard IEC/EN
60601-1-1 and/or other applicable standards). If you have any
questions, please contact Siemens Customer Service.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 7 of 58
SSME CS

Emotion (2007) / Spirit

General
NOTE

Unauthorized modifications to the system may cause incorrect patient treatment!


Unauthorized modifications to the system hardware, the software or the software configuration can lead to serious malfunctions of the system, to data corruption, or to incorrect
patient treatment. Unauthorized installation and execution of
software other than that specified for use on the scanner with
VSIM may jeopardize patient safety. All software installations
have be performed by authorized personnel.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 8 of 58
SSME CS

Siemens, 2012

General

Overview

1.2

ICS + IRS = syngo CT Acquisition Workplace


In the standard configuration, the ICS is the main console for interchangeable scanning
and imaging functions. The IRS is equipped with a switchable keyboard, mouse and monitor.
IES = syngo CT Workplace
The IES console may be installed in parallel to the main operator ICS console via a shared
database link. The IES is used as a second console for post-processing applications.

NOTE

Siemens, 2012

The IES runs only when the ICS is switched on. If the ICS is
switched off, the shared database cannot be accessed.

CT02-025.816.04.10.02
05.13

Page 9 of 58
SSME CS

Emotion (2007) / Spirit

10

General

Software Kit for ICS, IRS, and IES

1.3

Software Kit VB38 for ICS Tower 8B, IRS, and IES

The software installation kit for SOMATOM Emotion (2007) systems is included in the
delivery volume and comprises the following items required for a complete SW installation:
-

OEMsetup_1 English VB38B CD 1 of 5


OEMsetup_2 English VB38B CD 2 of 5
MUI_CHN_JPN VB38B CD 2A of 5
Som/5 ICS/IES SW/FW VB38B CD 3 of 5
Som/5_DICOM_SR_Viewer VB38B CD 4 of 5
Som/5_ICS/IES_SW/FW VB38B CD 5 of 5
1x Som/5_IRS_VB38_V2.0_64

- CD_ICS_IES_VB38B_SP1 (used for VB38B SP1 & VB38B SP3 software version)
- CD_ICS_IES_VB38B_SP3 (used for VB38B SP3 software version)

Software Kit VB39 for ICS Tower 8B, IRS, and IES

The software installation kit for SOMATOM Emotion (2007) systems is included in the
delivery volume and comprises the following items required for a complete SW installation:
-

OEMsetup_1 English VB39A CD 1 of 5


OEMsetup_2 English VB39A CD 2 of 5
Som/5 ICS/IES SW/FW VB39A CD 3 of 5
Som/5_DICOM_SR_Viewer VB39A CD 4 of 5
Som/5_ICS/IES_SW/FW VB39A CD 5 of 5

- or 1x Som/5_IRS_VB39_V1.0_64

Software Kit VB38 SP2 for ICS Tower 9B/9B2, IRS, and IES

The software installation kit for SOMATOM Emotion (2007) systems is included in the
delivery volume and comprises the following items required for a complete SW installation:
-

Emotion (2007) / Spirit

Som5_ICS_VB38B_SP2_DVD 1 of 1 (only for ICS installation)


OEMsetup_1 English VB38B CD 1 of 5
OEMsetup_2 English VB38B CD 2 of 5
MUI_CHN_JPN VB38B CD 2A of 5
Som/5 ICS/IES SW/FW VB38B CD 3 of 5
Som/5_DICOM_SR_Viewer VB38B CD 4 of 5
Som/5_ICS/IES_SW/FW VB38B CD 5 of 5
CD_ICS_IES_VB38B_SP1
CD_ICS_IES_VB38B_SP3
1x Som/5_IRS_VB38_V2.0_64

CT02-025.816.04.10.02
05.13

Page 10 of 58
SSME CS

Siemens, 2012

General

11

Images and Raw Data


NOTE

Siemens, 2012

1.4

For safety reasons, the customer needs to archive all images


and raw data prior to installing the SW, if possible.

CT02-025.816.04.10.02
05.13

Page 11 of 58
SSME CS

Emotion (2007) / Spirit

12

General

Required System Data

1.5

For installing the software, the following data should always be available:

Service Key

From the syngo user interface, select Options -> Local Service.
Write down the service key.
Service key = 20 characters
ICS

Service key (level=7)

IES 1

Service key (level=7)

IES 2

Service key (level=7)

14 characters

last 6 characters

Host Name

In the main menu of the service software, select Configuration.


Click Next.
Select Local Host -> TCP/IP LAN.
Write down the host name.
ICS

Computer name

IES 1

Computer name

IES 2

Computer name

Regional Settings

From the syngo user interface, select -> Options -> Configuration.
In the configuration panel, double-click the Regional and Language Options icon.
Write down the following data:
Regional Options

Language

Customized

Time

Time separator:

Date

Date separator:

Languages
->details

Emotion (2007) / Spirit

Input language

CT02-025.816.04.10.02
05.13

Page 12 of 58
SSME CS

Siemens, 2012

General

13

Report.htm and License.dat

In case a backup file becomes corrupted, it is useful to have the system configuration and
the license file available. For this reason, you have to save these files.

Open Windows Explorer.


Copy the file report.htm in folder C:\Somaris\service\html\report to, e.g., a floppy.
Copy the file license.dat in folder C:\Somaris\Config\Licensing to, e.g., a floppy.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 13 of 58
SSME CS

Emotion (2007) / Spirit

14

General

Preparations

1.6

NOTE

To ensure successful installation of the software, ensure the


following:
- No restart must be pending due to changes made in the
configuration. Open LocalService > Configuration,
check whether "restart pending" is shown in the message bar. If that is the case, click HOME to start the
missing automatic restart.
- All applications (e.g. Explorer, Patient Browser) must
be closed and stay closed during the entire update
unless requested otherwise in the instructions.
- Walk through the installation step by step and do not try
to save time by working out your own "installation strategy". This is especially important when installing the
Wizard. WAIT when you are asked to wait until the Navigator is ready.
- Thanks!

NOTE

Emotion (2007) / Spirit

Ensure that the Patient Browser is never opened during the


installation.

CT02-025.816.04.10.02
05.13

Page 14 of 58
SSME CS

Siemens, 2012

General

15

Backup Information
NOTE

1.7

For Chinese and Japanese systems, the system language


must be changed to English before ICS/IES backup.
- From the syngo user interface, select > Options > Configuration
- In the configuration panel, click Regional and Language
Options
- In the Regional Options tab card, select English (United
States)
- Under Location, select the country United States
- In the Advanced tab card, under "Select a language to
match.." select English (United States)
There is no need to restart the system now. The restart can
be performed after the CD Write Check.
Once the installation is finished, change the language back.

Make sure that up-to-date backups of the system are always available.
The CD-Writer (drive [-R-] CD-R) is the standard backup device for the ICS and IES. Alternatively, the MOD may be used, if available.
Make sure a separate backup medium (e.g. "CD Recordable") is on hand for both the
ICS and the IES (if available).

CD Write Check

During previous installations, we've experienced problems writing the backup files to CD.
As a result, this check has to be executed before starting the installation.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 15 of 58
SSME CS

Emotion (2007) / Spirit

16

General

Open Local Service Home Menu and go to Configuration > DICOM > Offline Devices
Check offline device.
Depending on the offline device installed, select either CD-R or CD/DVD Combo
as an offline device in the drop down menu
Check the item: Default device, click Save, and then restart.

Fig. 1:

CD Write Check

Creating Backup Files

If possible, you should generate new backup files before installing the software.

Start the service software and select the Backup & Restore function in the main menu.
Insert the backup CD-R.
From Command, select Backup.
From Drives, select the CD drive, e.g., [-R-] CD-R.
From Packages, select Security-Settings-PXXX and click Go.
BackupRestore Running ... Wait until Ready is displayed.

From Packages, select CT-Settings-PXXX and click Go.


BackupRestore Running ... Wait until Ready is displayed.

From Packages, select SW-Settings02-PXXX and click Go.


BackupRestore Running ... Wait until Ready is displayed.
NOTE

For ICS VB39, it also needs to backup VSim.

From Packages, select VSim and click Go.


BackupRestore Running ... Wait until Ready is displayed.

When "Ready" is displayed, remove the backup medium

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 16 of 58
SSME CS

Siemens, 2012

SW Installation ICS

17

2-

Preparations

2SW Installation ICS

NOTE

If you dont have a current backup, create one now, if possible. Refer to chapter 1 "Creating backup files".

NOTE

To prevent missing data, the customer needs to manually


store reports that have been created as well as all other data
on the H-drive prior to installing the update.

BHC adaptation

From the service software home menu, select Utilities.


Under Source, select Escape to OS and under Command select NT command Interpreter.

Under Parameters, type "Regedit".


Under HKEY_LOCAL_MACHINE\SOFTWARE\Siemens\SOMARIS\Config\Site\Masme\BHC_110KV, record the number, e.g.: the V1 in registry is
KV110_MINUS3

Under HKEY_LOCAL_MACHINE\SOFTWARE\Siemens\SOMARIS\Config\Site\Masme\BHC_130KV.

Under HKEY_LOCAL_MACHINE\SOFTWARE\Siemens\SOMARIS\Config\Site\Masme\BHC_80KV.

Required Data Media

To perform the SW installation of the ICS, the following data media are required:
NOTE

If the system has been updated to VB38B_SP4, software


installation should be performed with the software recovery
DVDs that were created during the update (CT007/11/P or
CT008/11/P).
CD_SOM/5_ICS/IES_BARE_32bit
(p/n
10414847) is used to restore the system.

For ICS Tower 8B


OEMsetup_1 English VB38(or VB39A) CD 1 of 5
MUI_CHN_JPN VB38 (or VB39A) CD 2A of 5
OEMsetup_2 English VB38 (or VB39A) CD 2 of 5
Som/5 ICS/IES SW/FW VB38 (or VB39A) CD 3 of 5
Som/5_DICOM_SR_Viewer VB38 (or VB39A) CD 4 of 5
Som/5_ICS/IES_SW/FW VB38 (or VB39A) CD 5 of 5

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 17 of 58
SSME CS

Emotion (2007) / Spirit

18

SW Installation ICS
CD_ICS_IES_VB38B_SP1 (used for VB38B SP1 & VB38B SP3 software version)
CD_ICS_IES_VB38B_SP3 (used for VB38B SP3 software version)
Backup CD-R

For ICS Tower 9B/9B2


Som5_ICS_VB38B_SP2_DVD 1 of 1
CD_ICS_IES_VB38B_SP1
CD_ICS_IES_VB38B_SP3
Backup CD-R

NOTE

During the software installation of the ICS, the IES has to be


switched off.

NOTE

Unplug the network cable between the ICS and the Ethernet
switch. After the software installation, plug in the cable
again.

System Type and Serial Number

Name

Identifier for
backup files

Material No.

Serial no.

SOMATOM Emotion 6 (2007)

P15F

101 65 888

xxx

SOMATOM Emotion 6 (2007)

P15F

100 46 766

xxx

SOMATOM Emotion 16 (2007) P15G

101 65 977

xxx

SOMATOM Emotion 16 (2007) P15G

100 46 789

xxx

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 18 of 58
SSME CS

Siemens, 2012

SW Installation ICS

19

Installation with Recovery DVD (for SP4)

2.1

Power on the ICS and insert CD_SOM/5__ICS/IES_VB38B_SP4_32bit_BARE (p/n


10414847).
System boots from the CD.

Select Restore system, and press OK.


The following pop-up window appears: "Partition C: will be completely erased!!!..., click
Yes to proceed.

A pop up message Please remove the CTBaRe CD, insert... displays, insert the first
recovery DVD and wait for the Ready of the drive, then click OK.
System is restoring from the DVD.

Insert the 2nd recovery DVD when you are asked for it. Proceed according to the
prompted instructions and then click OK.

When the recovery process is finished, please remove the DVD and press OK to reboot
the system.
After the system restarted, one popup window as shown below appears. Check if the
MOD is plugged into the ICS or not. Press Yes if a MOD is plugged in, press No if no
MOD is plugged in.

Fig. 2:

MOD is plugged in or not

NOTE

During the processing of the system, do NOT reboot the system when prompted for "System settings Change".

NOTE

During the processing of the system,click No to erase the


partitions when prompted for the message Do you want to
keep the partitions?...

Completion of the ICS installation (approx. 25 minutes) is indicated by a pop-up window with the message Please enter license.

Click Click here to start the Local Service Configuration to start the service software
(complete service license required).

Continue the installtion by jumping to VB38B SP1 and SP3 Installation.(VB38B SP1
and SP3 Installation / p. 27)

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 19 of 58
SSME CS

Emotion (2007) / Spirit

20

SW Installation ICS

Installation without Recovery DVD

2.2

Starting the Installation

Installation for ICS Tower 8B

From the IES, shut down the application and switch off the system.
Insert the OEMsetup_1 English VB38(or VB39A) CD 1 of 5 in the ICS.
From the ICS, select System/End and shut down the system.
Switch off the system at the UPS.

Disconnect the network cable of the ICS from the Ethernet switch.

Switch on the system at the UPS (computer "on").


The computer boots from CD-ROM. After a short while, the following window appears:
(This is only a sample:)

Fig. 3:

Press <Enter> to continue the installation. This reboots the ICS.


Select the distribution type: CUSTOMER (default).
Press Enter, and fill in the machine name (host name) as noted, replacing the default
entry "MedPC". Press Enter.
NOTE

Please note that the computer name must not contain blanks
or underscores. Use the following naming convention for the
ICS computer name: ct<serial
number>
(e.g.
ct123456).

Select Keep data on second harddrive or select Clear data on second harddrive
(deletes patient database).
Use ENTER to change selection.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 20 of 58
SSME CS

Siemens, 2012

SW Installation ICS
NOTE

21
Do not use 'Clear all harddrives 1-2(3)'. This option will format all hard drives only (removes all software). Reinstallation of the SW will not be started.

Use CONTINUE INSTALLATION to confirm the start of the installation.


Press any key.
All necessary files are copied from OEMsetup_1 English VB38 (or VB39A) CD 1
of 5 to the hard drive.

Wait until the CD drive opens, take out OEMsetup_1 English VB38(or VB39A) CD 1 of
5, and insert OEMsetup_2 English VB38(or VB39A) CD 2 of 5 , then press any key
For Chinese and Japanese language users only:
The MUI_CHN_JPN VB38 CD 2A of 5 has to be inserted directly after the OEMsetup_2
English VB38 CD 2 of 5 is ejected. If you miss inserting this CD on time and the computer asks for Som/5 ICS/IES SW/FW VB38 CD 3 of 5, you have to start the installation
from the beginning. No warning is issued for MUI_CHN_JPN VB38 CD 2A of 5, as for
the others if you miss inserting this CD.
Insert the additional MUI_CHN_JPN VB38 CD 2A of 5.

Insert the Som/5 ICS/IES SW/FW VB38 (or VB39A) CD 3 of 5.


The installation continues automatically.
After approx. 4 minutes, the CD is ejected.

Insert the Som/5_DICOM_SR_Viewer VB38(or VB39A) CD 4 of 5 .


The installation continues automatically.
After approx. 3 minutes, the CD is ejected.

Insert the Som/5_ICS/IES_SW/FW VB38(or VB39A) CD 5 of 5.


The installation continues automatically.
After approx. 30 minutes, the CD is ejected
NOTE

If you miss inserting the requested CD in time, a warning is


displayed:
No package has been found. Please insert the CD "........" in
the CD drive to continue the installation.
Insert the requested CD and press any key to continue.

Installation for ICS Tower 9B/9B2

From the IES, shut down the application and switch off the system.
Insert the Som5_ICS_VB38B_SP2_DVD 1 of 1 in the ICS.
From the ICS, select System/End and shut down the system.
Switch off the system at the UPS.

Disconnect the network cable of the ICS from the Ethernet switch.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 21 of 58
SSME CS

Emotion (2007) / Spirit

22

SW Installation ICS
Switch on the system at the UPS (computer "on").
The computer boots from CD-ROM. After a short while, the message Windows is
loading files appears. Wait until a Backup/Restore interface displays.

Fig. 4:Backup/Restore interface

Select Restore the systemdrive item using the drop-down menu under Keyboard, and click OK.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 22 of 58
SSME CS

Siemens, 2012

SW Installation ICS

23

A message The 1st hard disk will be completely erased!! Do you want to continue? pops up. Click OK.

Fig. 5:

The following message appears. Click OK.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 23 of 58
SSME CS

Emotion (2007) / Spirit

24

SW Installation ICS

Fig. 6:

GHOST is running. System is copying files.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 24 of 58
SSME CS

Siemens, 2012

SW Installation ICS

25

Fig. 7:

Wait until the DVD ejects automatically. At the same time, a message Restore of the
system ... ... appears. Remove the DVD and click OK.

Fig. 8:

In about 6 minutes, the system restarts automatically. The system continues with the
installation.
NOTE

In some cases, the message Do you want to keep the partitions? Press ... ... may display. Click No to erase the partition.

Installation running

The system reboots automatically after finishing the installation job.

The following screen appears:

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 25 of 58
SSME CS

Emotion (2007) / Spirit

26

SW Installation ICS

Fig. 9:

(This is only an example)

(This is only an example)

Select the system to be installed (refer to 2-2). Use the drop-down menu to change the
product name.

Click Start.
Completion of the ICS installation (approx. 25 minutes) is indicated by a pop-up window
with the message Please enter license.

Click Click here to start the Local Service Configuration to start the service software
(complete service license required).

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 26 of 58
SSME CS

Siemens, 2012

SW Installation ICS

27

VB38B SP1 and SP3 Installation


NOTE

2.3

Service Pack SP1 is a prerequisite for SP3 and must be


installed first.
If the installation of SP3 is started without having SP1 already
installed, an error message will result but not a malfunction.

VB38B SP1 Installation

Installation at ICS

Insert the CD_ICS_IES_VB38B_SP1 into the CD drive (e.g: [-R])


Select Options > Service > Local Service > Utilities
Under Source, select Escape to OS and under Command, select NT Command
Interpreter.

Under Parameter, type R:\install.bat.


Activate with Return or Go.
Wait a moment. An automatic system shut-down and restart are performed.
During shut-down, open the CD tray and take out the CD.
After restart, the VB38B_SP1 installation continues automatically. Wait until the message window The ServPack1 successfully installed appears.

An automatic system shut-down and restart are performed again.

VB38B SP3 Installation

Installation at ICS

Insert the CD_ICS_IES_VB38B_SP3 into the CD drive (e.g: [-R])


Select Options > Service > Local Service > Utilities
Under Source, select Escape to OS and under Command, select NT Command
Interpreter.

Under Parameter, type R:\install.bat.


Activate with Return or Go.
Wait a moment. An automatic system shut-down and restart are performed.
During shut-down, open the CD tray and take out the CD.
After restart, the VB38B_SP3 installation continues automatically. Wait until the message window The ServPack3 successfully installed appears.

An automatic system shut-down and restart are performed again.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 27 of 58
SSME CS

Emotion (2007) / Spirit

28

SW Installation ICS

Restore Steps
NOTE

2.4

Before restoration, make sure the IRS installation is finished,


is powered on, and connected to the ICS.

Restore 1

In the main menu of the service software, select the Backup & Restore function.
Insert the ICS backup CD.
From Command, select Restore.
From Drives, select the corresponding drive.
From Archive, select <hostname>_SW-Settings02-PXXX_date_time_ok.ar.

Fig. 10: (This is only an example)

From Groups, select all except User Settings and Service\Database\DBOffset, and
click Go.
BackupRestore Running ...; wait until Ready is displayed.

From Archive, select <hostname>_CT-Settings-PXXX_date_time_ok.ar.

Fig. 11: (This is only an example)

From Groups, select all except IRS and click Go.


BackupRestore Running ...; wait until Ready is displayed.

From Archive, select <hostname>_SecuritySettings-PXXX_date_time_ok.ar.


From Groups, select all and click Go.
Wait until "Ready" is displayed.
Click Home and restart the system.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 28 of 58
SSME CS

Siemens, 2012

SW Installation ICS

29

Restore 2

In the main menu of the service software, select the Backup & Restore function.
From Command, select Restore.
From Drives, select the corresponding drive.
From Archive, select <hostname>_SW-Settings02-PXXX_date_time_ok.ar

Fig. 12: (This is only an example)

NOTE

During the restore of SW settings, the following red error


message can be ignored
cmd /C C:\Somaris\Config\BaReUp\CapPb\update.bat

From Groups, select User Settings and click Go.


BackupRestore Running ...; wait until Ready is displayed.

From Archive, select <hostname>_CT-Settings-PXXX_date&time.

Fig. 13: (This is only an example)

From Groups, select IRS and click Go.


BackupRestore Running ...; wait until Ready is displayed.
NOTE

Siemens, 2012

For the ICS VB39 software, it also needs to restore VSim.

CT02-025.816.04.10.02
05.13

Page 29 of 58
SSME CS

Emotion (2007) / Spirit

30

SW Installation ICS
From Archive, select VSim and click Go.

Fig. 14: (This is only an example)

BackupRestore Running ...; wait until Ready is displayed.

Take out the backup CD.


Click Home.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 30 of 58
SSME CS

Siemens, 2012

SW Installation ICS

31

Completion of Configuration

2.5

Importing Reference Images

This is mandatory if "Clear data on second hard drive" was selected in the installation
screen.

From the Patient Browser, select Transfer -> Import from Offline.
From the drop-down list, select the path H:\SiteData\ReferenceImages.
Select all images and click OK.
After the import, the images are available in the Patient Browser.

Regional Settings

From the syngo user interface, select -> Options -> Configuration.
In the configuration panel, double-click the Regional and Language Options icon.
In the Regional Options tab card, select a preferred country and click Customize.
Change the date, time, and numbers as noted in the Regional Settings part.(Regional
Settings / p. 12)

In the Languages tab card, click Details and select the default input language type as
noted in the Regional Settings part.(Regional Settings / p. 12).

Setting the Time

In the main menu of the service software, select the Configuration function.
Select Site Info and click >. Check/correct the Time Zone and the Time. Note: Do not
activate the NTP time server or select automatic Daylight Savings Time
If changes were made, click Save. Confirm the message displayed with OK.

Configuring the Keyboard Language and enabling Windows XP Access

Perform the following steps to configure keyboard languages other than English:

From the syngo user interface, select > Options > Configuration.
In the configuration panel, click the Regional and Language Options icon.
Select the Languages > Details tab card
Click Add and select the correct Input Language and confirm with OK.
Perform the following steps to configure keyboard languages other than English (except
Chinese and Japanese):

From the syngo user interface, select > Options > Configuration.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 31 of 58
SSME CS

Emotion (2007) / Spirit

32

SW Installation ICS
Click OK in the remaining windows or close the remaining windows.
For Chinese/Japanese: Open the "DICOM/Character Set" dialog in "Local Service". Select
"allow multiple character set". Next, select the corresponding character set, for Japanese
"ISO 2022 IR 13" and for Chinese "GB 18030".
Perform the following steps to enable Windows XP access at the time of the update.

Enter the Service Key and open the Local Service main menu.
From the service software home menu, select Utilities.
Under Source, select Escape to OS and under Command select NT Command Interpreter.

Under Parameters, type c:\Somaris\Tools\bin\KeyFilter.bat disable.


Activate with Return or GO.
Select System > End > Restart the System.

BHC Configuration

From the service software home menu, select Utilities.


Under Source, select Escape to OS and under Command select NT Command Interpreter.

Under Parameters, type Regedit, activate with Return or click Go.


Under HKEY_LOCAL_MACHINE\SOFTWARE\Siemens\SOMARIS\Config\Site\Masme\BHC_XXXKV, change the related number along with the record.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 32 of 58
SSME CS

Siemens, 2012

SW Installation ICS

33

Under C:\Somaris\Tools\Bin, double-click the BHC_configure.exe file.


for Emotion 6 (2007) e. g. (only an example)

Fig. 15: Configuring BHC

for Emotion 16 (2007) e. g. (only an example)

Fig. 16: Configuring BHC

Click Start.

Additional Steps

NOTE

Siemens, 2012

If a system has a Wizard connected, execute the following


instructions at the Navigator to activate the Wizard.

CT02-025.816.04.10.02
05.13

Page 33 of 58
SSME CS

Emotion (2007) / Spirit

34

SW Installation ICS
Select Start > Programs / Accessories ->CommandPrompt.
In the open command window, type in perl c:\Somaris\utils\determinewizard.pl
Activate with Return.
This script will detect all Wizards that are active and connected to the shared
database and will make them available to the customer in the Configuration
menu.

Press Start. Select Shut down... -> Restart.


NOTE

If a system has a Wizard connected, sometimes in the IES


interface, a message may display Database creation of
exchangeboard failed - please call service.
Please execute the following instructions at the Wizard to
resolve this problem.

From the service software home menu in IES, select Utilities.


Under Source, select Database tools and under Command select DB Remove
Under Parameters, type 1

Fig. 17:DB Remove

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 34 of 58
SSME CS

Siemens, 2012

SW Installation ICS

35

Activate with Return.


A message Are you sure, you want to ... ... appears. Click OK.

Fig. 18:

The system will restore automatically.

Setting Auto Filming, Auto Viewing, or Auto 3D in the IES

If the customer requires Auto Filming in the first or second IES, he or she can select this
under Options/Config/Examination.

Siemens Remote Services

System Management installation

Open the Local Service Home menu and go to Configuration.


At System Mgmt-> Master check and correct the IP addresses.
Click Save and then Next
At System Mgmt-> Agent Control, see whether the checkbox Installed is set.
Click Go
At System Mgmt-> Agent Control set the checkbox Activated.
This may take a few minutes

Click Go
Click Finish.
Set the remote service access level

Go to Options -> Service -> Remote Service


Set the Remote Service Access Control Mode to Limited Access Permanent

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 35 of 58
SSME CS

Emotion (2007) / Spirit

36

SW Installation ICS

Configuring AutoReport Mon_HQ


NOTE

If you activate the Mon_HQ function, the CT system will automatically send event log files to the SRS server.
Please make sure the customer has already agreed to activate this function. Otherwise, please ignore the configuration.

Select Options > Service > Local Service


In the service software home menu, select AutoReport
Under Report, select Mon_HQ and check the status as shown below.

Fig. 19: Mon_HQ


Pos. 1

Under REPORT, select Mon_HQ

Pos. 2

Active status.

Pos. 3

(mail)Target: SRS_mail is configured as the mail target

Pos. 4

Ensure that the time period (start and end) is set correctly

Pos. 5

Sample source?: source: File

Pos. 6

Sample what?File name: C:\Somaris\log\EventLog\*Evt*20*txt.gz Mode: Time triggered

Pos. 7

Sample when? Continuously, Sample times: 02:00;0700;08:30;10:00

Pos. 8

Send what? Mail: as separate file(s), FTP target: E.g. ftp2lux09505

Pos. 9

Send when? At once

Pos. 10

If you perform any changes <Save> the autoreport

Click Save.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 36 of 58
SSME CS

Siemens, 2012

SW Installation of the IRS

37

3-

Installation of the IRS

3SW Installation of the IRS

To install the software:

Insert the CD Som/5 IRS SW VB38 (or VB39) into the CD drive of the computer.
Switch the computer off and on again.
The computer boots from CD.

When the CD drive opens, take out the CD and reboot the computer again.
NOTE

If the CD is not ejected after 15 minutes, switch off the IRS.


Wait for 15 seconds, then switch it on again to restart the process.

The computer starts the SW installation and will be finished within approx. 30 minutes.

IRS Configuration

After the IRS installation is finished, enter the Service Key in the ICS system and open
the Local Service main menu.

From the service software home menu, select Utilities.


Under Source, select Escape to OS and under Command select NT Command Interpreter.

Under Parameters, type Explorer and click Go.


Log into the C:\Somaris\utils folder, run IrsTablesInstall.bat, and wait until it finishes.
In the service software main menu, select the Backup & Restore function.
Insert the ICS backup CD.
From Command, select Restore.
From Drives, select the corresponding drive.

From Archive, select <hostname> CT-Settings-PXXXXX-.

Fig. 20: (This is only an example)

From Groups, select IRS and click Go.


The "site-specific tune-up tables" are transferred to the IRS.

When "Ready" is displayed, remove the ICS backup CD and restart the complete system.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 37 of 58
SSME CS

Emotion (2007) / Spirit

38

SW Installation IES
4-

Preparations

4SW Installation IES

NOTE

If you dont have a current backup, create one now, if possible. Refer to chapter 1 "Creating backup files".

Required Data Media


NOTE

If the system has been updated to VB38B_SP4, software


installation should be performed with the software recovery
DVDs that were created during the update (CT007/11/P or
CT008/11/P).
CD_SOM/5_ICS/IES_BARE_32bit
(p/n
10414847) is used to restore the system.

To perform the SW installation of the ICS, the following data media are required:
CD OEMsetup_1 English VB38 (or VB39) CD 1 of 5
MUI_CHN_JPN VB38 CD 2A of 5
CD OEMsetup_2 English VB38 (or VB39) CD 2 of 5
CD Som/5 ICS/IES SW/FW VB38 (or VB39) CD 3 of 5
CD Som/5_DICOM_SR_Viewer VB38 (or VB39) CD 4 of 5
CD Som/5_ICS/IES_SW/FW VB38 (or VB39) CD 5 of 5
Backup CD-R

Starting the Installation

Emotion (2007) / Spirit

NOTE

During the software installation of the IES, the ICS has to be


switched on.

NOTE

Unplug the network cable between the IES and the Ethernet
switch.

CT02-025.816.04.10.02
05.13

Page 38 of 58
SSME CS

Siemens, 2012

SW Installation IES

39

Installation with Recovery DVD (for SP4)

4.1

Refer to the procedures for 'ICS installation with recovery DVD'(Installation with Recovery DVD (for SP4) / p. 19).

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 39 of 58
SSME CS

Emotion (2007) / Spirit

40

SW Installation IES

Installation without Recovery DVD

4.2

Starting the Installation

Click Home and terminate the service software by closing the Browser window.
Select System/End and Shut down the system.
Insert the CD OEMsetup_1 English VB38 (or VB39) CD 1 of 5
Power off the IES.
Disconnect the network cable of the IES from the Ethernet switch.
Power on the IES.
The computer boots from CD-ROM. After a short while, the following window
appears:

(This is only a sample:)

Fig. 21:

Press <Enter> to continue installation (the IES is beeping). The IES restarts.
Select distribution type CUSTOMER (default).
Press Enter, and fill in the machine name (host name) as noted by replacing the default
entry "MedPC". Press Enter.
NOTE

Please note that the computer name must not contain blanks
or underscores. Use the following naming convention for the
IES computer name: ctw<serial number> (e.g. ctw12345, IES
connected to ct12345).

'Keep data on other hard drive(s)' is the default setting and may not be modified (only
one hard drive available)
NOTE

Emotion (2007) / Spirit

Do not use 'Clear all hard drive(s)'. This option will only format all hard drives (removes all software). No reinstallation of SW will be started.

CT02-025.816.04.10.02
05.13

Page 40 of 58
SSME CS

Siemens, 2012

SW Installation IES

41

Use CONTINUE INSTALLATION to confirm the start of installation.


Press any key.
All necessary files are copied from CD OEMsetup_1 English VB38 (or VB39) CD
1 of 5 to the hard drive.

When the computer beeps, take out the CD OEMsetup_1 English VB38/VB39 CD 1 of
5, insert the CD OEMsetup_2 English VB38 (or VB39) CD 2 of 5 and press any key
For Chinese and Japanese language users only:
The MUI_CHN_JPN VB38 CD 2A of 5 has to be inserted directly after the OEMsetup_2
English VB38 CD 2 of 5 is ejected. If you miss inserting this CD on time and the computer asks for Som/5 ICS/IES SW/FW VB38 CD 3 of 5, you have to start the installation
from the beginning. No warning is issued for MUI_CHN_JPN VB38 CD 2A of 5, as for
the others if you miss inserting this CD.
Insert the additional MUI_CHN_JPN VB38 CD 2A of 5.

Insert the CD Som/5 ICS/IES SW/FW VB38 (or VB39) CD 3 of 5).


The installation continues automatically.
After approx. 4 minutes, the CD is ejected.

Insert the CD Som/5_DICOM_SR_Viewer VB38 (or VB39) CD 4 of 5.


The installation continues automatically.
After approx. 3 minutes, the CD is ejected.

Insert the CD Som/5_ICS/IES_SW/FW VB38 (or VB39) CD 5 of 5.


The installation continues automatically.
After approx. 30 minutes, the CD is ejected
NOTE

If you miss inserting the requested CD in time, a warning is


displayed:
No package has been found. Please insert the CD "........" in
the CD drive to continue the installation.
Insert the requested CD and press any key to continue.

Installation running

The system reboots automatically after finishing the copy job.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 41 of 58
SSME CS

Emotion (2007) / Spirit

42

SW Installation IES
The following screen appears:

Fig. 22: (This is only an example)

Select syngo CT Workplace in the field "Product Name". Use the drop-down menu to
change to syngo CT Workplace.

Click Start.
Completion of IES installation (approx. 15 minutes) is indicated by a pop-up window with
the message Please enter license.

NOTE

Continue only when the ICS is fully installed and started up.

Click Click here to start the Local Service Configuration to start the service software
(complete service license required).

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 42 of 58
SSME CS

Siemens, 2012

SW Installation IES

43

VB38B SP1 and SP3 Installation


NOTE

4.3

Service Pack SP1 is a prerequisite for SP3 and must be


installed first.
If the installation of SP3 is started without having SP1 already
installed, an error message will result but not a malfunction.

VB38B SP1 Installation

Installation at IES

Insert the CD_ICS_IES_VB38B_SP1 into the CD drive (e.g: [-R])


Select Options > Service > Local Service > Utilities
Under Source, select Escape to OS and under Command, select NT Command
Interpreter.

Under Parameter, type R:\install.bat.


Activate with Return or Go.
Wait a moment. An automatic system shut-down and restart are performed.
During shut-down, open the CD tray and take out the CD.
After restart, the VB38B_SP1 installation continues automatically. Wait until the message window The ServPack1 successfully installed appears.

An automatic system shut-down and restart are performed again.

VB38B SP3 Installation

Installation at IES

Insert the CD_ICS_IES_VB38B_SP3 into the CD drive (e.g: [-R])


Select Options > Service > Local Service > Utilities
Under Source, select Escape to OS and under Command, select NT Command
Interpreter.

Under Parameter, type R:\install.bat.


Activate with Return or Go.
Wait a moment. An automatic system shut-down and restart are performed.
During shut-down, open the CD tray and take out the CD.
After restart, the VB38B_SP3 installation continues automatically. Wait until the message window The ServPack3 successfully installed appears.

An automatic system shut-down and restart are performed again.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 43 of 58
SSME CS

Emotion (2007) / Spirit

44

SW Installation IES

Restore Steps
NOTE

4.4

To perform the following steps, the ICS has to be installed


and running. Switch on the ICS.

Restore

In the main menu of the service software, select the Backup & Restore function.
Insert the backup medium for IES.
From Command, select Restore.
From Drives, select the corresponding drive.
From Archive, select SW-Settings02-Wiz_date_time.ar.
From Groups, select all and click Go.
BackupRestore Running ...; wait until Ready is displayed.

From Archive, select <hostname>_Security-Settings-PXXX_date_time_ok.ar.


From Groups, select all and click Go.
BackupRestore Running ...; wait until Ready is displayed.

From Archive, select CT-Settings-date_time_ok.ar.


From Groups, select all and click Go.
BackupRestore Running ...; wait until Ready is displayed.

Take out the backup CD.


Click Home.
Connect the network cable of the IES to the Ethernet switch.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 44 of 58
SSME CS

Siemens, 2012

SW Installation IES

45

Completion of Configuration

4.5

Regional Settings

From the syngo user interface, select -> Options -> Configuration.
In the configuration panel, double-click the Regional and Language Options icon.
In the Regional Options tab card, select a preferred country and click Customize.
Change the date and time as noted in chapter 1.

NOTE

The ARGUS option allows users to enter the parameters for


"patient height" and "patient weight". For this reason, the
correct measurement units have to be selected.
Select "Regional Options > Customize... > Numbers > Measurement System",
For cm/kg, the setting has to be set to "Metrics".

In the Languages tab card, select the default input language and the keyboard type as
noted on page 1-4.

Setting the Time

In the main menu of the service software, select the Configuration function.
Select Site Info and click >. Check/correct the time zone and the time. Note: Do not
activate the NTP time server or select automatic Daylight Savings Time.

If changes were made, click Save. Confirm the message displayed with OK.

Configuring the Keyboard Language and enabling Windows XP Access

Perform the following steps to configure keyboard languages other than English:

From the syngo user interface, select > Options > Configuration.
In the configuration panel, click the Regional and Language Options icon.
Double-click Regional and Language Options.
Select the tab card Languages > Details.
Click Add and select the correct Input Language and confirm with OK.

Perform the following steps to configure keyboard languages other than English (except
Chinese and Japanese).

From the syngo user interface, select > Options > Configuration.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 45 of 58
SSME CS

Emotion (2007) / Spirit

46

SW Installation IES
Click OK in the remaining windows or close the remaining windows.
For Chinese/Japanese: Open the "DICOM/Character Set" dialog in "Local Service". Select
"allow multiple character set". Next, select the corresponding character set, for Japanese
"ISO 2022 IR 13" and for Chinese "GB 18030".
Perform the following steps to enable Windows XP access at the time of the update.

Enter the Service Key and open the Local Service main menu.
From the service software home menu, select Utilities.
Under Source, select Escape to OS and under Command select NT Command Interpreter.

Under Parameters, type c:\Somaris\Tools\bin\KeyFilter.bat disable.


Activate with Return or GO.
Select System > End > Restart the system.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 46 of 58
SSME CS

Siemens, 2012

SW Installation IES

47

Siemens Remote Services

4.6

System Management installation

Open the Local Service Home menu and go to Configuration.


At System Mgmt-> Master check and correct the IP addresses.
Click Save and then Next.
At System Mgmt-> Agent Control, see whether the checkbox Installed is set.
Click Go.
At System Mgmt-> Agent Control set the checkbox Activated.
This may take as few minutes.

Click Go.
Click Finish.
Set the remote service access level

Go to Options -> Service -> Remote Service.


Set the Remote Service Access Control Mode to Limited Access Permanent.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 47 of 58
SSME CS

Emotion (2007) / Spirit

48

Additional Work Steps


5-

Software Self-test

IES

5Additional Work Steps

Map drive C: of ICS (ICS) at IES


Open Windows Explorer.
Select Tools -> Map Network Drive

Select
Type
Folder:

Drive:

Z:

\\<ICS-Hostname>\C$

In the upcoming window, log on as


the Administrator using the
administrator password of the IES

Fig. 23: MapNetworkDrive

Open Windows Explorer.


Copy the file SomSTSysInfo.info
C:\Somaris\Tools\SelfTest\Communication\Local\
Z:\Somaris\Tools\SelfTest\Communication\Remote\
and
Z:\Somaris\Tools\SelfTest\Communication\Local\
C:\Somaris\Tools\SelfTest\Communication\Remote\

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 48 of 58
SSME CS

Siemens, 2012

Additional Work Steps

49

ICS and IES

Open the Local Service Home menu and go to Test Tools.


Click the Software SelfTest and execute all subtests.
-

Environment
Configuration
Security
Database
Misc
Communication (only if the IES is connected)
Performance
In case of errors, refer to the hints for troubleshooting.

NOTE: If you have to perform troubleshooting, make sure that you recreate the SomSTSysInfo.info and copy and paste it between the ICS and IES as described above. Otherwise, you will get the same error messages even though you have already solved the
problem.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 49 of 58
SSME CS

Emotion (2007) / Spirit

50

Additional Work Steps

Monitor Configuration

5.1

Note: This step has to be performed for all systems.


The configuration steps have to be performed at the ICS and the IES.

Go to the Windows Start Menu,


select Shutdown > Log off
meduser.

Click Yes and press the left Shift key


until the login window appears

Log in as Administrator
Go to the Settings > Control Panel
> Display

Select the Settings tab card.


Go to Advanced (see item 1).

Fig. 24: Display Properties

Select the Monitor tab card.


If a DVI monitor is connected, set the
screen refresh rate to 60 Hertz.

If a VGA monitor is connected, set


the screen refresh rate to 75 Hertz.

Confirm your configuration by pressing OK.

Fig. 25: Plug and Play Monitor

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 50 of 58
SSME CS

Siemens, 2012

Additional Work Steps

51

Tower 9B/9B2 Information Identification Steps


NOTE

5.2

The following identification steps are only valid for ICS Tower
9B/9B2. Please be careful.

Login as Administrator
Open file C:\ DMILOG.txt .
Add one line Chassis Information (1/Fig. 26 / p. 51) to the end of this txt file
Find the line KEY: SMBIOSAssetTag VALUE:... ... (2/Fig. 26 / p. 51), copy the whole
line and paste it to the end of the txt file.

In the last line change KEY:SMBIOSAssetTag VALUE:... ... to Asset Tag:... ...
(3/Fig. 26 / p. 51)

Fig. 26: Sample


Pos. 1

Chassis Information

Pos. 2

SMBIOSAssetTag

Pos. 3

Asset Tag

Save the txt file.


Restart the system.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 51 of 58
SSME CS

Emotion (2007) / Spirit

52

Additional Work Steps

Final Check

5.3

Open the Local Service Home menu and go to Configuration.


Click Next.
Go to Local Host > Technical.
Select Country for Quality:
- USA for USA
- 'other' for all other countries

Click Save and OK in the window that is displayed.


Click Home.
In the Local Service Home menu go to Quality Assurance.
Click Constancy.
Select Normal.
Click Go and proceed according to the program instructions.

Perform scans in the customer mode and expose a film to ensure the camera is functioning properly.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 52 of 58
SSME CS

Siemens, 2012

Additional Work Steps

53

Backup at the ICS

5.4

Start the service software and select the Backup & Restore function in the main menu.
Insert the backup CD-R for the ICS.
From Command, select Backup.
From Drives, select the CD drive, e.g., [-R-] CD-R.
From Packages, select CT-Settings-PXXX and click Go.
BackupRestore Running ...

Fig. 27: (This is only an example)

Wait until Ready is displayed.

Close the CD tray.


From Packages, select SW-Settings02-PXXX and click Go.
BackupRestore Running ...

Fig. 28: (This is only an example)

Wait until Ready is displayed.

Close the CD tray.

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 53 of 58
SSME CS

Emotion (2007) / Spirit

54

Additional Work Steps


From Packages, select Security-Settings-PXXX and click Go.
BackupRestore Running ...

Fig. 29: (This is only an example)

Wait until Ready is displayed.

When "Ready" is displayed, remove the backup medium.


Click Home and terminate the service software by closing the Browser window.
NOTE

The backup files have been saved correctly when ok is added


at the end of the backup file name.
e.g. <hostname>-SW-Settings02-PXXX-date_time_ok.ar

Disabling Windows XP Access

Perform the following steps to disable Windows XP access for the customer. Administrator
privileges are required; you must use the command shell of the service software.

From the service software home menu, select Utilities.


Under Source, select Escape to OS and under Command NT Command Interpreter.
Under Parameters, type c:\Somaris\Tools\bin\KeyFilter.bat enable.
Activate with Return or GO.
Reboot to make the change effective.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 54 of 58
SSME CS

Siemens, 2012

Additional Work Steps

55

Backup at the IES

5.5

Start the service software and select the Backup & Restore function in the main menu.
Insert the backup CD-R for the IES.
From Command, select Backup.
From Drives, select the corresponding drive.
From Packages, select CT-Settings-IES and click Go.
BackupRestore Running ...

Fig. 30: (This is only an example)

Wait until Ready is displayed.

Close the CD tray.


From Packages, select SW-Settings02 and click Go.
BackupRestore Running ...

Fig. 31: (This is only an example)

Wait until Ready is displayed.

Close the CD tray.


From Packages, select Security-Settings-IES and click Go.
BackupRestore Running ..

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 55 of 58
SSME CS

Emotion (2007) / Spirit

56

Additional Work Steps

Fig. 32: (This is only an example)

Wait until Ready is displayed.

Close the CD tray.


From Packages, select InSpace (optional) and click Go.
BackupRestore Running ..

Fig. 33: (This is only an example)

Wait until Ready is displayed.

Close the CD tray.


When "Ready" is displayed, remove the backup medium.
Click Home and terminate the service software by closing the Browser window.

Disabling Windows XP Access

Perform the following steps to disable Windows XP access for the customer. Administrator
privileges are required. As a result, deactivation is performed using the command shell of
the service software.

From the service software home menu, select Utilities.


Under Source, select Escape to OS and under Command NT Command Interpreter.
Under Parameters, type c:\Somaris\Tools\bin\KeyFilter.bat enable.
Activate with Return or GO.
Reboot to make the change effective.

Emotion (2007) / Spirit

CT02-025.816.04.10.02
05.13

Page 56 of 58
SSME CS

Siemens, 2012

Additional Work Steps

57

Changes to previous Version

5.6

Chapter

Change

Reason

2, 4

V9: Add installation with recovery DVD

ECR 576313

V10: Modify the IRS installation process

ECR 585334

Siemens, 2012

CT02-025.816.04.10.02
05.13

Page 57 of 58
SSME CS

Emotion (2007) / Spirit

58

Emotion (2007) / Spirit

Additional Work Steps

CT02-025.816.04.10.02
05.13

Page 58 of 58
SSME CS

Siemens, 2012

Das könnte Ihnen auch gefallen