Beruflich Dokumente
Kultur Dokumente
Journal of Biomechanics
journal homepage: www.elsevier.com/locate/jbiomech
www.JBiomech.com
Ecole Polytechnique Fe de rale de Lausanne (EPFL), Laboratory of Movement Analysis and Measurement (LMAM), Lausanne, Switzerland
Ecole de Technologie Supe rieure, Laboratoire de recherche en imagerie et orthope die (LIO). Montreal, Canada
c
Centre Hospitalier Universitaire Vaudois and University of Lausanne (CHUV), Department of Orthopaedic Surgery and Traumatology, Lausanne, Switzerland
b
a r t i c l e in f o
a b s t r a c t
Article history:
Accepted 15 June 2009
Measurement of three-dimensional (3D) knee joint angle outside a laboratory is of benet in clinical
examination and therapeutic treatment comparison. Although several motion capture devices exist,
there is a need for an ambulatory system that could be used in routine practice. Up-to-date, inertial
measurement units (IMUs) have proven to be suitable for unconstrained measurement of knee joint
differential orientation. Nevertheless, this differential orientation should be converted into three
reliable and clinically interpretable angles. Thus, the aim of this study was to propose a new calibration
procedure adapted for the joint coordinate system (JCS), which required only IMUs data. The
repeatability of the calibration procedure, as well as the errors in the measurement of 3D knee angle
during gait in comparison to a reference system were assessed on eight healthy subjects. The new
procedure relying on active and passive movements reported a high repeatability of the mean values
(offseto11) and angular patterns (SDo0.31 and CMC40.9). In comparison to the reference system, this
functional procedure showed high precision (SDo21 and CC40.75) and moderate accuracy (between
4.01 and 8.11) for the three knee angle. The combination of the inertial-based system with the functional
calibration procedure proposed here resulted in a promising tool for the measurement of 3D knee joint
angle. Moreover, this method could be adapted to measure other complex joint, such as ankle or elbow.
& 2009 Elsevier Ltd. All rights reserved.
Keywords:
Knee
3D Kinematics
Inertial sensors
Functional calibration
Ambulatory measurement
Gyroscopes
Accelerometers
1. Introduction
Three-dimensional (3D) knee joint angle measurement is an
important requirement, notably in the orthopaedic and rehabilitation elds. Although several systems, such as optoelectronic
devices, can make this measurement (Andriacchi and Alexander,
2000; Sutherland, 2002), there is a need for an ambulatory system
that can be used for clinical routine examinations. Since a decade,
some authors have proposed wearable systems including accelerometers, gyroscopes and magnetometers (referred as magnetic
and inertial measurement unit: MIMU) (Foxlin, 1996; Bachmann,
2000; Roetenberg, 2006) or wearable systems composed of
accelerometers and gyroscopes (referred as inertial measurement
unit: IMU) (Luinge and Veltink, 2005; Favre et al., 2006) to
measure the orientation of rigid bodies in space. With MIMUs,
these orientations are calculated relative to an earth-xed frame
dened by the gravity and local magnetic eld. However, MIMUs
are sensitive to the magnetic eld distortions and their use in
clinical setting environments is limited (Roetenberg, 2006;
Correspondence to: EPFL-STI-IBI2-LMAM, ELH 131/Station 11, CH-1015 Lausanne, Switzerland. Tel.: +41 21 693 76 06; fax: +41 21 693 69 15.
E-mail address: julien.favre@ep.ch (J. Favre).
0021-9290/$ - see front matter & 2009 Elsevier Ltd. All rights reserved.
doi:10.1016/j.jbiomech.2009.06.025
ARTICLE IN PRESS
J. Favre et al. / Journal of Biomechanics 42 (2009) 23302335
2331
W
The joint coordinate system (JCS) (Grood and Suntay, 1983) was used to
describe the knee joint angles. Fig. 1 shows a joint composed of a proximal (A:
thigh) and distal (b: shank) segment. The BAFs of the segments are respectively IJK
and ijk. The orientation of these two BAFs respective to the xed reference frame
~
~
~
~
(XYZ) are expressed by two quaternions Q
A and Q b . The generalized joint
coordinate e1e2e3 represents the system about which the three clinical rotations
e 2 , and
occur: exion/extension (FE) around ~
e 1 , abduction/adduction (AA) around ~
nally internal/external rotation (IE) around ~
e3.
V
RA
QA
J
I
QA
Z
e3
e1
e2
Qb
k
X
Qb
u
v
b
j
Rb
The wearable system was made up of two small IMUs connected to a portable
data-logger (Physilogs, BioAGM, CH). The IMUs were xed to the thigh and shank
segments using an exoskeleton harness (Ganjikia et al., 2000) as shown in Fig. 2a.
The data were recorded at a sampling frequency of 240 Hz. The orientation of the
two IMUs frame UVW and uvw (Fig. 1) were calculated relative to their own static
reference frame using a tracking algorithm described in Favre et al. (2006).
Basically, this algorithm fuses the gyroscope and accelerometer row signals to
calculate an optimised orientation at each time sample. Then, the thigh and shank
IMUs orientation were calculated relative to a common static reference frame XYZ
~0
~0
~
~
and expressed by two quaternions Q
A and Q b (Fig. 1). This alignment was
realised following the functional procedure proposed by Favre et al. (2008), which
estimates the reference frames misalignment comparing the angular velocity
vectors of the thigh and shank segments during a hip AA movement (Fig. 2b). The
combination of the fusion algorithm with the functional alignment procedure was
shown to provide precise short time (up to a few minutes) measurement of thigh
Fig. 2. (a) Measurement systems: the ovals surround the inertial and magnetic sensors. (b) Hip abduction/adduction movement. (c) Passive movement of the shank in its
sagittal plane. (d) Passive movement of the shank in its frontal plane.
ARTICLE IN PRESS
2332
and
~
~0
~
~
~
~
Q
b t Q b t R b
Dsc 2 cos1
~
~ 1
~
R sc
R s ~
!
3
real
where 1 stands for the inverse operation, whereas ||real denotes the real part of the
~
~
quaternion (Kuipers, 2002). ~
R sc corresponds to the calibration quaternion (~
R A or
~
~
~
~
R b ) of the subject s obtained with the movement combination c. R s corresponds,
for each subjects, to the normalized mean calibration quaternion over the eight
movement combinations. DSC represents the absolute angular amplitude of the 3D
~
~
~
rotation aligning ~
R sc with R
s (Spoor and Veldpaus, 1980; Kuipers, 2002).
2.6.2. Repeatability of the 3D knee joint angle
Since the nal purpose of the system is to measure the three knee joint angles,
the propagation of the calibration quaternions repeatability to the knee joint
angles calculation was evaluated. For this, the knee joint angles of the walking
trials (FEsc(t), AAsc(t) and IEsc(t)) were calculated for each subject considering the
~
calibration quaternions (~
R sc ) obtained for each movement combination. Then, the
angles were averaged FE s t; AA s t and IE s t over their eight movement
combinations according to Eq. (4); and similar formulas were considered for AA
and IE:
FE s t
N
1X
FEsc t
N c1
The differences between the eight angles (e.g., FEsc(t)) and their mean (e.g.,
FE s t) were calculated for each subject. The absolute value of the mean differences
(Doffset sc), and the standard deviation of these differences (DSD sc) were then
calculated for every subjects and combinations. Finally, these parameters were
averaged throughout the 8 subjects leading to: mDoffset and mDSD for each knee joint
angle. Moreover, in order to quantify the similarity between the patterns, the
coefcient of multiple correlation (CMC) was calculated between the angles
obtained through the eight combinations for each subject and knee angle (Kadaba
et al., 1989; Kutner et al., 2005). These coefcients were also averaged throughout
the 8 subjects leading to: mDCMC for each knee joint angle. When the patterns are
similar the CMC tends to 1, whereas it tends to 0 if the patterns are totally
dissimilar.
2.7. Evaluation of the 3D knee joint angle
The knee joint angles obtained from the two calibration methods (m M1 or
M2) were compared to the corresponding angles obtained through the reference
system for each subject (e.g., FEs,m(t) vs. FEs,REF(t)). Regarding M2, according to the
results obtained for the Repeatability of the 3D knee joint angle (Section 2.6.2), only
one of the movement combinations (i.e., c 1) was considered. The absolute value
of the mean difference (offsets,m) between the angles (e.g., FEs,m(t) and FEs,REF(t)),
the standard deviation of this difference (SDs,m), as well as the correlation
coefcient (CCs,m) were calculated for each method, subject and angle. Finally,
these three parameters were averaged through the 8 subjects leading to: moffset,m
(accuracy), mSD,m (precision) and mCC,m for each knee joint angle and method.
3. Results
The 3D shank angular velocity measured during the two seated
passive calibration movements are shown for a typical subject in
Fig. 3. The former is expressed in the IMUs frame (which also
corresponds to M1) and in the BAFs which corresponds to M2.
Regarding the functional procedure (M2), the repeatability of
the calibration quaternions (w) was of 2.41 and 2.01 for the thigh
and shank segments, respectively. Table 1 reports the repeatability
of the 3D knee joint angle (mDoffset, mDSD and mDCMC) due to the
propagation of the calibration quaternions precision. These results
showed high repeatability of the patterns for the three knee joint
angles (i.e., small mDSD and high mDCMC). Regarding the inuence of
ARTICLE IN PRESS
J. Favre et al. / Journal of Biomechanics 42 (2009) 23302335
mvt2
200
2333
mvt3
150
sensor frame
sensor frame
100
100
50
0
50
100
100
200
150
0
200
0
150
segment frame
100
100
segment frame
50
0
50
100
100
200
150
0
Time, s
Time, s
Fig. 3. Typical 3D shank angular velocities (1/s) during the two passive calibration movements in sensor and segment frames. mvt2 corresponds to knee FE and mvt3 to
~ or ~
rotation of the shank in its frontal plane. Solid lines corresponds to ~
u or ~
i axes, dashed lines to ~
v or ~
j axes, and dotted lines to w
k axes.
Table 1
Mean (SD) repeatability of the 3D knee joint angle obtained with M2.
FE
AA
IE
mDoffset
mDSD
mDCMC
0.8 (0.9)
0.4 (0.5)
0.6 (0.8)
0.0 (0.0)
0.2 (0.1)
0.1 (0.1)
1.00 (0.01)
0.89 (0.14)
0.91 (0.10)
4. Discussion
In this study, a new functional calibration procedure (M2) was
proposed to measure the 3D knee joint angle according to the ISB
recommendation. In contrary to previous calibration procedures in
literature, this method was suitable for a system only composed of
two IMUs. The performance of this new method was compared to a
simple method (M1) where visual alignment between IMUs and
segments axes was used.
The repeatability (w) of the functional calibration procedure
(M2) was comparable to that reported for calibration methods
relying on bony landmarks and using camera-based systems
(Della Croce et al., 1999, 2003). Unfortunately, none of the
studies proposing calibration procedures for MIMUs had reported
a comparable parameter (ODonovan et al., 2007; Picerno et al.,
2008).
The propagation of the calibration quaternions precision onto the
three knee joint angles was evaluated during gait using three
ARTICLE IN PRESS
2334
60
40
20
0
20
0
20
0
20
40
20
10
0
10
20
Time, s
Fig. 4. 3D knee angles (FE, AA and IE) obtained during the walking task for a typical subject. The reference angles are in black and the angles obtained through the proposed
system in gray (dotted for M1 and solid for M2).
Table 2
Mean (SD) errors of the 3D knee joint angle.
FE
AA
IE
M1
M2
M1
M2
M1
M2
moffset
mSD
11.0
8.1
14.9
6.2
6.3
4.0
4.4
1.3
15.0
2.0
5.2
2.0
(5.1)
(5.4)
(3.5)
(5.1)
(4.3)
(4.7)
mCC
(1.4)
(0.5)
(2.6)
(0.6)
(1.7)
(0.9)
0.96
1.00
0.65
0.76
0.60
0.85
(0.02)
(0.00)
(0.16)
(0.18)
(0.26)
(0.11)
ARTICLE IN PRESS
J. Favre et al. / Journal of Biomechanics 42 (2009) 23302335
References
Andrews, J.G., 1984. On the Specication of joint congurations and motions.
Journal of Biomechanics 17, 155158.
Andriacchi, T.P., Alexander, E.J., 2000. Studies of human locomotion: past, present
and future. Journal of Biomechanics 33, 12171224.
Bachmann, E., 2000. Inertial and magnetic tracking of limb segment orientation for
inserting humans into synthetic environments. Ph.D. thesis, Naval Postgrduate
School, Monterey, CA, p. 176.
Baker, R., 2006. Gait analysis methods in rehabilitation. Journal of Neuroengineering and Rehabilitation 3, 4.
Besier, T.F., Sturnieks, D.L., Alderson, J.A., Lloyd, D.G., 2003. Repeatability of gait
data using a functional hip joint centre and a mean helical knee axis. Journal of
Biomechanics 36, 11591168.
Blankevoort, L., Huiskes, R., Delange, A., 1988. The envelope of passive knee-joint
motion. Journal of Biomechanics 21, 705709.
Blankevoort, L., Huiskes, R., Delange, A., 1990. Helical axes of passive knee-joint
motions. Journal of Biomechanics 23, 12191229.
Cappozzo, A., Catani, F., Croce, U.D., Leardini, A., 1995. Position and orientation inspace of bones during movement-anatomical frame denition and determination. Clinical Biomechanics 10, 171178.
Cappozzo, A., Della Croce, U., Leardini, A., Chiari, L., 2005. Human movement
analysis using stereophotogrammetryPart 1: theoretical background. Gait
and Posture 21, 186196.
Charlton, I.W., Tate, P., Smyth, P., Roren, L., 2004. Repeatability of an optimised lower
body model. Gait and Posture 20, 213221.
Della Croce, U., Cappozzo, A., Kerrigan, D.C., 1999. Pelvis and lower limb anatomical
landmark calibration precision and its propagation to bone geometry and joint
angles. Medical and Biological Engineering and Computing 37, 155161.
Della Croce, U., Camomilla, V., Leardini, A., Cappozzo, A., 2003. Femoral anatomical
frame: assessment of various denitions. Medical Engineering and Physics
25, 425431.
Della Croce, U., Leardini, A., Chiari, L., Cappozzo, A., 2005. Human movement
analysis using stereophotogrammetryPart 4: assessment of anatomical
landmark misplacement and its effects on joint kinematics. Gait and Posture
21, 226237.
Donati, M., Camomilla, V., Vannozzi, G., Cappozzo, A., 2008. Anatomical frame
identication and reconstruction for repeatable lower limb joint kinematics
estimates. Journal of Biomechanics 41, 22192226.
Favre, J., Jolles, B.M., Siegrist, O., Aminian, K., 2006. Quaternion-based fusion of
gyroscopes and accelerometers to improve 3D angle measurement. Electronics
Letters 42, 612614.
Favre, J., Jolles, B.M., Aissaoui, R., Aminian, K., 2008. Ambulatory measurement of
3D knee joint angle. Journal of Biomechanics 41, 10291035.
Foxlin, E., 1996. Inertial head-tracker sensor fusion by a complementary separatebias Kalman lter. In: Proceedings of the VRAIS, Santa Clara, CA,
pp. 185194.
Ganjikia, S., Duval, N., Yahia, L., de Guise, J., 2000. Three-dimensional knee analyzer
validation by simple uoroscopic study. Knee 7, 221231.
Grood, E.S., Suntay, W.J., 1983. A joint coordinate system for the clinical description
of 3-Dimensional motions-application to the knee. Journal of Biomechanical
EngineeringTransactions of the ASME 105, 136144.
Hagemeister, N., Parent, G., Van de Putte, M., St-Onge, N., Duval, N., de Guise, J.,
2005. A reproducible method for studying three-dimensional knee
kinematics. Journal of Biomechanics 38, 19261931.
2335
Jan, S.V., Salvia, P., Hilal, I., Sholukha, V., Rooze, M., Clapworthy, G., 2002.
Registration of 6-DOFs electrogoniometry and CT medical imaging for 3D joint
modeling.
Journal of Biomechanics 35, 14751484.
Kadaba, M.P., Ramakrishnan, H.K., Wootten, M.E., Gainey, J., Gorton, G., Cochran,
G.V.B., 1989. Repeatability of kinematic, kinetic, and electromyographic data in
normal adult gait. Journal of Orthopaedic Research 7, 849860.
Kadaba, M.P., Ramakrishnan, H.K., Wootten, M.E., 1990. Measurement of lowerextremity kinematics during level walking. Journal of Orthopaedic Research 8,
383392.
Kuipers, J.B., 2002. Quaternions and Rotation Sequences. Princeton University
Press, Priceton, NJ, p. 400.
Kutner, M.H., Nachtsheim, C.J., Neter, J., Li, W., 2005. Applied Linear Statistical
Models, fth ed McGraw-Hill, Irwin, USA, p. 1424.
Leardini, A., Chiari, L., Della Croce, U., Cappozzo, A., 2005. Human movement
analysis using stereophotogrammetryPart 3, soft tissue artifact
assessment and compensation. Gait and Posture 21, 212225.
Luinge, H.J., Veltink, P.H., 2005. Measuring orientation of human body segments
using miniature gyroscopes and accelerometers. Medical and Biological
Engineering and Computing 43, 273282.
Mannel, H., Marin, F., Claes, L., Durselen, L., 2004. Establishment of a knee-joint
coordinate system from helical axes analysisa kinematic approach without
anatomical referencing. IEEE Transactions on Biomedical Engineering
51, 13411347.
Marin, F., Mannel, H., Claes, L., Durselen, L., 2003. Correction of axis misalignment
in the analysis of knee rotations. Human Movement Science 22, 285296.
Mills, P.M., Morrison, S., Lloyd, D.G., Barrett, R.S., 2007. Repeatability of 3D gait
kinematics obtained from an electromagnetic tracking system during treadmill
locomotion. Journal of Biomechanics 40, 15041511.
Moglo, K.E., Shirazi-Adl, A., 2005. Cruciate coupling and screw-home mechanism in
passive knee joint during extension-exion. Journal of Biomechanics
38, 10751083.
Most, E., Axe, J., Rubash, H., Li, G., 2004. Sensitivity of the knee joint kinematics
calculation to selection of exion axes. Journal of Biomechanics 37, 17431748.
ODonovan, K.J., Kamnik, R., OKeeffe, D.T., Lyons, G.M., 2007. An inertial and
magnetic sensor based technique for joint angle measurement. Journal of
Biomechanics 40, 26042611.
Picerno, P., Cereatti, A., Cappozzo, A., 2008. Joint kinematics estimate using
wearable inertial and magnetic sensing modules. Gait and Posture
28, 508595.
Roetenberg, D., 2006. Inertial and Magnetic Sensing of Human Motion. Twente
University, Enschede, NL, p. 126 (Ph.D. thesis).
Schache, A.G., Baker, R., Lamoreux, L.W., 2006. Dening the knee joint exionextension axis for purposes of quantitative gait analysis: an evaluation of
methods. Gait and Posture 24, 100109.
Spoor, C.W., Veldpaus, F.E., 1980. Rigid body motion calculated from spatial
coordinates of markers. Journal of Biomechanics 13, 391393.
Sutherland, D.H., 2002. The evolution of clinical gait analysis: Part II Kinematics.
Gait and Posture 16, 159179.
Wilson, D.R., Feikes, J.D., Zavatsky, A.B., OConnor, J.J., 2000. The components of
passive knee movement are coupled to exion angle. Journal of Biomechanics
33, 465473.
Woltring, H.J., 1994. 3-D attitude representation of human jointsa standardization proposal. Journal of Biomechanics 27, 13991414.