MultiParameterPatientMonitor

PRODUCTS
StandardParameter:ECG,NIBP,SpO2,Temp,Resp.
Fullyportabilitywithsteamlinehandle
Highbrightness12.1TFTcolorLCD
Bigfiguredisplay
ECGwaveformof7leaddisplayinonescreen
Compatibleandselfidentify3/5leadECGcable
NIBPdualoverpressureprotection
Isolatedfloating,antidefibrillationprotectedand
antihighfrequencyinterference
72hoursreviewtrend
480hoursreviewtrendwithexternalmemory
Optionalparameters&Configuration:
2IBP,EtCO2,2TEMP,Recorder,12leadECG,
CentralMonitoringnetworkS/W,WirelessNetwork,
VGAoutput,TouchLCD
Builtinrechargeablebattery
Dimension(mm):310(L)x140(W)x263(H)
Weight:3.8Kg

Selectionguide
RestingECG
CARDIPIA800H
CARDIPIA400H
CARDIPIA200
StressECGSystem
VeterinaryECG/Monitor
CARDIPIA400HV
CARDIPIA200
VITAPIA7000KV
HolterECG
ECGSimulator
PatientMonitor

TechnicalSpecifications
Classfication

VITAPIA7000K

Antielectrochocktype

Classlequipmentandinternalpoweredequipment

VITAPIA5000

EMCtype

ClassA

Antielectroshockdegree

ECG(RESP),SPO2,NIBP,IBP,TEMP,CO2 ,CF

VITAPIA5500
CMSS/W
FetalMonitor
FM8000Type8010

Specifications
SizeMonitor

310X140X263mm

WeightMonitor

3.8kg

Temperature

FM8000Type8007
Fetaldoppler

Humidity

DP6000
PulseOximeter
OXY500TypeFB/FBW

Altitude

Working

5~40

Transport/Storage

40~+55

Working

30%~75%

Transport/Storage

95%(nocoagulate)

Working

700hPa~1060hPa

Transport/Storage

500hPa~1060hPa

PowerSupply

OXY500TypeDP

100~240VAC,50/60Hz,Pmax=150VA,FUSET1.6A
Device

12.1in.colorTRT,3LED/800X600dots

TouchLCD(Option)

viewingangle(up:70,down:70,Left:50,Right:60andthebestangleis60)

Messages

8WaveformMaximum
1PowerLED(Green)
3SoungModecorrespondingAlarmMode

ECGOutput

BNC

Amplitude

1mV

Display

SignalInterface Accuracy

1AlarmLED(Yellow/Red)
1BatteryChargeLED(Yellow)

<5%

Impedance

50

SignalDelay

<20ms
Rechargeble3.7A/Hr7.4VLibattery
Operatingtimeunderthenormaluseandfullchargegreaterthan120min.
Operatingtimeafterthefirstalarmoflowbatterywillbeabout5min.

Battery

TrendRecall

Short:1hrs,1secondResolution
Long:72hrs,1minResolution

AlarmEventRecall

71alarmeventsofallparametersand8/16/32secondsofcorrespondingwaveform.

NIBPMeasurement
Recall

Atleast400NIBPmeasurementdata

SDcard

72hrsECGwaveform

480hrsTRENDreview

Leadmode

5Leads(R,L,F,N,CorRA,LA,LL,RL,V)

Leadselection

,,

Waveform

2CH

Leadmode

3Leads(R,L,ForRA,LA,LL)

Leadselection

,,

Waveform

1ch

Gain

x2.5mm/mV,x5.0mm/mV,x10mm/mV,x20mm/mV

HR

MeasuringRange:15~300bpm

AlarmRangle(ADU):15~300bpm

AlarmRange(PED/NEO):15~350bpm

Accuracy:1%or1bpm,whichgreat
Sensitivity:>200uVPP

Resolution:1bpm
Differentalinput
ECG

Impedance

>5M
Monitor:100dB

CMRR
Electodeoffset
potential

Operation:100dB
Diagnosis:60dB
300mV

LeakageCurrent

<10uA

BaselineRecovery

<5sAfterDefi

ECGSignalRange

8mV(Vpp)
Surgery:1~20Hz(+0.4dB,3dB)

Bandwidth

Monitor:0.5Hz~40Hz(+0.4dB,3dB)
Diagnostic:0.05Hz~75Hz(+0.4dB,3dB),76Hz~150Hz(+0.4dB,4.5dB)

CalibrationSignal

1mV(Vpp),5%Accuracy

STSegment
MonitoringRange

MeasureRange:0.6~+0.8mV
AlarmRange:2.0~+2.0mV
Type:ASYSTOLE,VFIB/VTC,COUPLET,BIGEMINY,TRIGEMINY,RONT,VT>2,PVC,
TACHY,BRADY,MISSEDBEATS,PNP,PNC

ARRDetecting

Alarm:Available
Review:Available

Method

ImpedancebetweenRF(RALL)

DifferentialInput
Impedance

>2.5M

Measuring
ImpedanceRange
RESPARATION

BaselineImpedance
Range

0.3~5.0
0.1~2.5K

Bandwidth

0.3~2.5Hz

Resp.Rate

MeasuringRange:0~120rpm
AlarmRange(PED/NEO):0~150rpm
Accuracy:2rpm

ApneaAlarm

10~40S

AlarmPage(ADU):0~120rpm
Resolution:1rpm

Oscillometric
Mode

Manual,Auto,STAT

Measuring
IntervalinAUTO
Mode

1,2,3,4,5,10,15,30,60,90,120,180,240,480,960min

Measuring
IntervalinSTAT
Mode

5min

AlarmType

SYS,DIA,MEAN

MeasuringandalarmRange
Adultmode

SYS:40~270mmHg
DIA:10~215mmHg
MEAN:20~235mmHg

Pediatricmode

SYS:40~200mmHg
DIA:10~150mmHg
MEAN:20~165mmHg

Neonatalmode

SYS:40~135mmHg
DIA:10~100mmHg
MEAN:20~110mmHg

Resolution

Pressure:1mmHg

Accuracy

Pressure(MaximumMeanerror):5mmHg
Pressure(MaximumStandarddeviation):8mmHg

Overpressure
protection

Adultmode:2973mmHg
Pediatricmode:2403mmHg
Neonatalmode:1473mmHg

NIEP

Measuring

Range

SpO2

0~100%

AlarmRange

0~100%

Resolution

1%

Accuracy
AlarmDelay

70~100%2%
0~69%unspecified
about1sec

Actualizationinterval 10sec
MeasuringandRange:0~250bpm
PulseRate

Resolution:1bpm
Accuracy:2bpm

Channel
Measuringand
AlarmRange
Resolution
TEMPERATURE Accuracy
Actualization
interval
AverageTime
Constant

0~50
0.1
0.2
about1sec
<10sec

Channel

Label

ART,PA,CVP,RAP,ICP,P1,P2

Measuring
range
Alarmrange
IBP(Option)

PressSensor

10~+300mmHg
50~+350mmHg
Sensitivity:5uV/V/mmHg
Impedance:300~3000

Resolution

1mmHg

Accracy

2%or1mmHgwhichgreat

Actualization
Interval
Method
Measuring
mode

about1sec
InfraredAbsorptionTechnique
MainstreamandSidestream

Sidestream
modesampling

50ml/Min.10ml/Min.

pasflowrate
CO2 :0~150mmHg
Measuringrange

INSCO2 :0~150mmHg
AwRR:2~150rpm

CO2
(Option)

Resolution

CO2 :0.1mmHg(0~69mmHg),0.25mmHg(70~150mmHg)
INSCO2 :0.1mmHg(0~69mmHg),0.25mmHg(70~150mmHg)

Accuracy

CO2:2mmHg,0~40mmHg
5%ofreading,41~70mmHg
8%ofreading,71~100mmHg
10%ofreading,101~150mmHg
AqRR:1rpm

Pacemaker
(Option)

Availalbe

Initialization
Time

Mainstrea:Capnogramdisplayedinlessthan15secatanambientTemp.of25,
fullsepcificationwithin2min
Sidestrea:Capnogramdisplayedinlessthan20secatanamblientTemp.of25,
fullsepcificationwithin2min
MainstreamRiseTime:Lessthen60msAdultReusableorSinglePatientUse
AirwayAdaptor

Sidestream
DelayTime

about1sec

Alarmrange

2~3sec

Suffocation
AlarmDelay

CO2 :0~150mmHg

Contentsandspecificationsubjectschangewithoutnotice