Signaling
MEDICAL DEVICES
5.1 Magnet indicator
Display
Instruction
6658-0_quick_start_guide_MRI_guard_agilia_eng
Color
No light
Display
Event / Instruction
No light
Blinking
Fix
Blinking
Maintenance mode.
Fresenius Kabi AG
D-61346 Bad Homburg v.d.H
Germany
Phone +49 (0)6172 686-0
www.fresenius-kabi.com
INFORMATION
Before starting to use MRI Guard Agilia, connect the power cord to
the mains power supply for approximately 15 hours without using
the device. Check power status indicator is lit in green.
WARNING
Check the IV lines remain loose between the MRI Guard Agilia and
the patients IV access.
WARNING
2. Approach the MRI Guard Agilia carefully to the MR Scanner and observe the
Magnet Indicator until you reach the desired position in the MRI Scan Room:
Magnet Indicator should remain green.
When the Magnet Indicator flashes yellow, move back
the MRI Guard Agilia until the Magnet Indicator returns
to the green position.
The dangerous area for the pumps regarding the magnetic fiel
starts at 50 mT. Never position the MRI Guard Agilia closer than 1 m
to the MRI Scanner. The Agilia IV infusion pumps may be attracted
towards the MRI scanner when positioned at this distance.
3. Position the MRI Guard Agilia so the front and the Cap User Interface (top)
remains visible at all times from the control room.
4. Once correctly positioned, release
the two handles and push the
brakes of the castors in order to
prevent any movement of the MRI
Guard Agilia.
5. If a wall plug is available, connect the power cable of the MRI Guard Agilia.
WARNING
6. Make sure the door of the MRI Guard Agilia is closed, then start the MR
examination.
WARNING
WARNING
Do not squeeze the infusion lines by closing and locking the door.
The door of the MRI Guard Agilia has to be locked when the MRI
Scanner is imaging but can be opened between different scans
(when the MRI Scanner is not imaging).
In case of an emergency the door can be opened when MRI
Scanner is imaging, but this could lead to artefacts on the MR
Images (not to malfunctions of the pumps).
The pumps should not be removed from MRI Guard Agilia rack.
4
WARNING
Before closing the door, make sure the infusion lines are properly
placed in the outlets to avoid squeezing them and provoking
occlusions.