Photo-Therapy 4000
Phototherapy Unit
Revision 3.0
6141.420
9036107
Emergency Care Perioperative Care Critical Care Perinatal Care Home Care
Contents
General
1
Notes
1.1
Function Description
Photo-Therapy 4000 function description
13
Annex
Parts catalog
Test List
K6141420IECIVZ.fm 19.12.05
Technical Information
Contents
K6141420IECIVZ.fm 19.12.05
II
General
Notes
General
This Technical Documentation conforms to the IEC 60601-1 standard.
Read each step in every procedure thoroughly before beginning any test.
Always use the proper tools and specified test equipment. If you deviate from
the instructions and/or recommendations in this Technical Documentation,
the equipment may operate improperly or unsafely, or the equipment could
be damaged.
Use only original Drger parts and supplies.
The maintenance procedures described in this Technical Documentation may
be performed by qualified service personnel only. These maintenance
procedures do not replace inspections and servicing by the manufacturer.
The information in this manual is confidential and may not be disclosed to
third parties without the prior written consent of the manufacturer.
6141.420
General
1.1
Symbols and
Definitions
This symbol indicates a warning.
Maintenance
Repair
Servicing
Inspection
6141.420
Function Description
Photo-Therapy 4000
Photo-Therapy 4000
function description
Function description
The light of the Photo-Therapy 4000 is helpful for malfunctions of bilirubin
metabolism. If a newborn baby develops icterus because detoxification in the
liver has been inadequate, therapy to lower bilirubin concentration may be
required as early as the first day.
The light converts bilirubin into a water-soluble form by photochemical
reaction. This enables natural excretion via bile and urine.
The blue portion of the light spectrum (wavelength: 460 nm) is most effective
in decomposing bilirubin.
Fig.
Distance (cm)
(4 blue light lamps)
30
40
50
Ebi mW/cm2
2.4 +0.4
1.6 +0.3
1.2 +0.2
Effective light intensity for bilirubin dissociation Ebi depends on how far away
the lamp housing (lower edge) is at mains voltage:
6141.420
Function description
Photo-Therapy 4000
If a higher bilirubin-effective intensity of light is required, the white light lamps
may be replaced with blue phototherapy lamps. Six blue light phototherapy
lamps will result in the following values for the bilirubin dissociation Ebi:
Distance (cm)
(6 blue light lamps)
30
40
50
Ebi mW/cm2
3.4 +0.6
2.3 +0.4
1.6 +0.3
Control elements
Fuses
4
3
Fig.
Handle
6141.420
Legend
Photo-Therapy 4000
Function description
4
3
1
2
Fig.
Rating plate
Handle
Legend
6141.420
10
6141.420
Function description
Photo-Therapy 4000
11
12
Photo-Therapy 4000
6141.420
Connecting Diagramm
Figure 1
13
14
6141.420
Annex
Parts catalog
Test List
Technical Information
Parts catalog
Phototherapie 4000
Emergency Care - Perioperative Care - Critical Care - Perinatal Care - Home Care
Parts catalog
Electrical Component
Item
No.
Part No.
Description
Qty.
Qty.u Remark
nit
2M21010
FLUORESCENT LAMP,BLUE
1.000
St
2M21009
FLUORESCENT LAMP,WHITE
1.000
St
2M21383
LAMPHOLDER
1.000
St
Items that are shown in the illustration but are not listed below the illustration are not available as spare parts
6141.420
Parts catalog
Trolley old
Item
No.
Part No.
Description
Qty.
Qty.u Remark
nit
2M18788
CAP (60X30)
1.000
St
Items that are shown in the illustration but are not listed below the illustration are not available as spare parts
6141.420
Phototherapie 4000
Assembly
Description
Parts catalog
Part No.
2M21010
FLUORESCENT LAMP,WHITE
2M21009
LAMPHOLDER
2M21383
2M21010
FLUORESCENT LAMP,WHITE
2M21009
LAMPHOLDER
2M21383
2M21010
FLUORESCENT LAMP,WHITE
2M21009
LAMPHOLDER
2M21383
SUPPLY MAIN
1807331
Housing 230V/127V/90-100V
DISC (PLEXIGLASS)
2M21335
LAMPHOLDER
2M21383
9037157
9037834
9037835
9029150
9029343
9037833
9029345
Products Concerned
PHOTO THERAPY 4000 90-100V
2M22090
PHOTO-THERAPY 4000/110V/127V
2M21700
PHOTO-THERAPY 4000/230V
2M21000
Trolley old
CAP (60X30)
6141.420
2M18788
Revision: 2005-01
Test List
K6141420TL1.fm 19.12.05
Photo-Therapy 4000
Serial no.:
Installation site:
File no.:
6141.420
Edition:
01.97
This Test List does not replace inspection and servicing by the manufacturer
6141.420 Photo-Therapy 4000
Page 1
K6141420TL1.fm 19.12.05
This Test List does not replace inspection and servicing by the manufacturer
6141.420 Photo-Therapy 4000
Page 2
Functional test
3.1
3.2
K6141420TL1.fm 19.12.05
Plug the power plug of the Photo-Therapy 4000 into the AC power
outlet.
Switch on the Photo-Therapy 4000.
Set the switch (blue light) to the ON position.
Four blue phototherapy lamps come on and the hour meter is running.
Set the switch (blue light) to the OFF position.
3.3
This Test List does not replace inspection and servicing by the manufacturer
6141.420 Photo-Therapy 4000
Page 3
4.1
4.2
Set both switches (white light and blue light) to the ON position.
Subsequent measurements may exceed the first-measured value in
the Test List by max. 50 %, but must not exceed the specified test
value.
The equivalent leakage current (IA) is less than/equal to 1.0 mA.
K6141420TL1.fm 19.12.05
Initial value: IA =
mA
Stand
5.1
Height adjustment
Castors 2M 15116 (2x)
This Test List does not replace inspection and servicing by the manufacturer
6141.420 Photo-Therapy 4000
Page 4
5.2
Stand 2M 21190
Height adjustment
Castors 2M 21373 (4x)
Handover
K6141420TL1.fm 19.12.05
Date: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
Name: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
This Test List does not replace inspection and servicing by the manufacturer
6141.420 Photo-Therapy 4000
Page 5
K6141420TL1.fm 19.12.05
This Test List does not replace inspection and servicing by the manufacturer
6141.420 Photo-Therapy 4000
Page 6
Technical Information
2005-11-24
Electromagnetic Emissions
Electromagnetic Emissions
Photo-Therapy 4000 is intended for use in the electromagnetic environment specified below. The user
should assure that is used in such an environment.
Emissions
Compliance
Electromagnetic environment
according to
RF emissions (CISPR 11)
Group 1
Photo-Therapy 4000 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
Harmonic emissions
(IEC 61000-3-2)
Voltage fluctuations / flicker
(IEC 61000-3-3)
Class A
Photo-Therapy 4000 is suitable for use in all establishments other than domestic and those directly connected
to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Not applicable
Not applicable
page 1 of 3
Electromagnetic Immunity
Electromagnetic Immunity
Photo-Therapy 4000 is intended for use in the electromagnetic environment specified below. The user
should assure that is used in such an environment.
Immunity
IEC 60601-1-2 test level
Compliance
Electromagnetic environment
against
level (of PhotoTherapy 4000)
Floors should be wood, concrete or
electrostatic discontact discharge: 6 kV
6 kV
ceramic tile. If floors are covered
charge, ESD
air discharge: 8 kV
8 kV
with synthetic material, the relative
(IEC 61000-4-2)
humidity should be at least 30%.
Mains power quality should be that
electrical fast tran- power supply lines: 2 kV
2 kV
of a typical commercial or hospital
sients / bursts
longer input / output lines:
environment.
(IEC 61000-4-4)
Not
applicable
1 kV
surges on AC
Mains power quality should be that
common mode: 2 kV
2 kV
mains lines
of a typical commercial or hospital
differential mode: 1 kV
1 kV
(IEC 61000-4-5)
environment.
power frequency
3 A/m
Not applicable
In close vicinity to Photo-Therapy
magnetic field
4000, no equipment with extraordi50/60 Hz
nary power frequency magnetic
(IEC 61000-4-8)
fields (power transformers, etc.)
should be operated.
dip >95%, 0.5 periods
>95%, 0.5 per.
voltage dips and
Mains power should be that of a
60%, 5 per.
dip 60%, 5 periods
short interruptions
typical commercial or hospital envi30%, 25 per.
on AC mains input dip 30%, 25 periods
ronment. If user requires continued
dip >95%, 5 seconds
>95%, 5 sec.
lines
operation during power mains inter(IEC 61000-4-11)
ruptions, it is recommended to
power the Photo-Therapy 4000
from an uninterruptible supply or a
battery.
radiated rf
80 MHz 2.5 GHz: 10 V/m
10 V/m
Recommended separation distance
(IEC 61000-4-3)
from portable and mobile rf transmitters with transmission power
PEIRP to Photo-Therapy 4000 including its lines: 1.84 m * PEIRP X1
rf coupled into
150 kHz 80 MHz: 3 V
3V
Recommended separation distance
lines
within ISM bands,
from portable and mobile rf trans(IEC 61000-4-6)
3 V outside ISM bands X2
3V
mitters with transmission power
PEIRP to Photo-Therapy 4000 including its lines: 1.84 m * PEIRP X1
Information re electromagnetic immunity (IEC 60601-1-2: 2001, tables 202, 203, 204)
X1
: For PEIRP the highest possible "equivalent isotropic radiated power" of the adjacent rf transmitter has
interferto be inserted (value in Watt). Also in the vicinity of equipment marked with the symbol
ence may occur. Field strengths from fixed, portable or mobile rf transmitters at the location of PhotoTherapy 4000 should be less than 3 V/m in the frequency range from 150 kHz to 2.5 GHz and less
than 1 V/m above 2.5 GHz.
X2
: ISM bands in this frequency range are: 6.765 MHz - 6.795 MHz, 13.553 MHz - 13.567 MHz, 26.957
MHz - 27.283 MHz, 40.66 MHz - 40.70 MHz.
page 2 of 3
0.001
0.06
0.17
0.003
0.10
0.30
0.010
0.18
0.55
0.030
0.32
0.95
0.100
0.58
1.73
0.200
0.82
2.46
0.250
0.91
2.75
1.000
1.83
5.48
2.000
2.60
7.78
3.000
3.16
9.49
Hint
Information re separation distances (IEC 60601-1-2: 2001, tables 205 and 206)
* 3 V/m distance to transmitters with frequencies from 150 kHz to 2.5 GHz, otherwise 1 V/m distance.
page 3 of 3
Manufacturer:
Drger Medical AG & Co. KG
Moislinger Allee 53 55
D-23542 Lbeck
Germany
Phone: (++49) (0) 1805-3723437
Fax: (++49) 451/882 - 3779